MEDICAL DEVICES SECTOR

Similar documents
MDS - G4 GUIDANCE FOR OVERSEAS MANUFACTURERS

MDS-G25. Guidance on Requirements for Storage, Handling and Transportation of Medical Devices

Dr. Nazeeh S. Alothmany Meshal A. Al-Amri. Saudi Food & Drug Authority

FINAL DOCUMENT. Global Harmonization Task Force

In the huge expanse of Asia, Singapore and

INTERNATIONAL CONFORMITY CERTIFICATION PROGRAM

MEDICAL DEVICE GUIDANCE DOCUMENT REQUIREMENTS FOR LABELLING OF MEDICAL DEVICES

Guide to the in-vitro diagnostic medical devices legislation

Section 3 Company Organisation and Certification Procedures

FINAL DOCUMENT. Global Harmonization Task Force. Title: Principles of Conformity Assessment for In Vitro Diagnostic (IVD) Medical Devices

Medical Devices. LITHUANIA LAWIN Lideika, Petrauskas, Valiunas ir partneriai

GUIDANCE NOTES FOR MANUFACTURERS OF CLASS I MEDICAL DEVICES

Saudi Food & Drug Authority. Khaled Ali Aldahami D.V.M; MSc Head of foreign establishments inspection section

This document is meant purely as a documentation tool and the institutions do not assume any liability for its contents

PROPOSED DOCUMENT. Global Harmonization Task Force. Title: Reportable Events During Pre-Market Clinical Investigations

PROPOSED DOCUMENT. Global Harmonization Task Force. Title: Principles of Conformity Assessment for In Vitro Diagnostic (IVD) Medical Devices

PRODUCT CERTIFICATION REQUIREMENTS

MEDICAL DEVICES SECTOR BUSINESS PROCESS MANAGEMENT MATURITY (BPMM) ASSESSMENT. Friday, November 15, 2013

News of the MDR and IVDR with a focus on innovation, UDI coding and traceability. Roberto Belliato

WORKING DRAFT DOCUMENT

Distinguishing Medical Device Recalls from Product Enhancements and Associated Reporting Requirements

Canada: Consumer Protection Law Overview

MEDICAL DEVICE GUIDANCE DOCUMENT REQUIREMENTS FOR LABELLING OF MEDICAL DEVICES

L 313/16 Official Journal of the European Union

PROPOSED DOCUMENT (for Annex E and F)

Guide to field safety corrective actions for medical devices and in-vitro diagnostic medical devices

GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES

Pure Encapsulations, Inc. United States Minimum Advertised Price Policy Effective November 13, 2017

Medical Devices. LATVIA LAWIN Klavins & Slaidins

Cell and Gene Therapy Medicinal Product Management Act (Draft) General Information

PROPOSED DOCUMENT. Global Harmonization Task Force. Title: Definitions of the Terms Manufacturer, Authorised Representative, Distributor and Importer.

WELMEC Application of module D Quality assurance Under directive 2014/32/EU (MID) or directive 2014/31/EU (NAWID)

DRAFT MEDICAL DEVICE GUIDANCE DOCUMENT

Radio Equipment Directive

Contents. Regulatory Bodies... 13

Korean Medical Devices Regulations

Medical Device Regulatory Framework 9 SEPTEMBER 2015 FUNDISA CONFERENCE JANE ROGERS

WHO GLOBAL BENCHMARKING TOOL (GBT) FOR EVALUATION OF NATIONAL REGULATORY SYSTEM OF MEDICAL PRODUCTS

Qualified Persons in the Pharmaceutical Industry. Code of Practice. March 2008

Medical Device Regulatory Roadmap SAMED Conference 2-3 December 2015

Inspection of the conduct of clinical evaluations on medical devices in the premises of healthcare providers

Section 13 Operating Procedures for Importers

Due diligence in the European medical devices industry

MEDICAL DEVICE GUIDANCE

Legal Obligations of Distributors under the Construction Products Regulation and other Associated European Legislation as from 1 st July 2013

ABC Digital Standards

Medical Devices. The technical requirements also apply when the submitter is not the legal manufacturer 1 (i.e: a distribution company).

2.2.2 GUIDE TO THE IMPLEMENTATION OF DIRECTIVES BASED ON THE NEW APPROACH AND THE GLOBAL APPROACH

UNICEF TECHNICAL REQUIREMENTS FOR MEDICAL DEVICES (MD)

AUDIT REPORT. Audit of Traceability in Food and Food Products of Animal Origin

(Information) INFORMATION FROM EUROPEAN UNION INSTITUTIONS, BODIES, OFFICES AND AGENCIES EUROPEAN COMMISSION

MEDICINES CONTROL COUNCIL

Proposed Regulatory Requirments National Health Act, Republic of South Africa CEASA Breakfast Meeting

Terms of Audit Engagements

CE marking is regarded as a passport for products to enter and circulate freely within the 30 counties forming part of the EEA.

NOTICE. Re: Guidance Document GD211: Guidance on the Content of Quality Management System audit reports

Saudi FDA Drug Approval Process. Dr. Mohammed A. Alquwaizani Consultant- Chief of the scientific office Saudi Food and Drug Authority

State Institute for Control of Drugs-/ Státní ústav pro control léčiv (SÚKL) Regulatory and ethics bodies involved in approval process

(Draft for comment) Administrative Measures for Recall of Consumer Products

DRAFT GUIDANCE DOCUMENT ON THE LOW VOLTAGE DIRECTIVE TRANSITION

Health Products and Food Branch Inspectorate

MEDICINES CONTROL COUNCIL

Operational Policy CODE OF ADVERTISING STANDARDS 1. STATEMENT

Manufacturers Guideline to the Construction Product Regulation and its Implementation

CABINET OF MINISTERS OF UKRAINE

QUESTIONNAIRE FOR COMPETENT AUTHORITIES TO BE USED FOR ASSESSMENT, REASSESSMENT AND SELF- EVALUATION

Medical Device Regulatory Roadmap SAMED Conference 2-3 December 2015

FINAL DOCUMENT. International Medical Device Regulators Forum. Medical Device Regulatory Audit Reports

Medical Device Postmarket Compliance in the United States

Risk Management Plan Guidance

closer look at Definitions The General Data Protection Regulation

AMF Position Compliance function requirements

An introductory guide to the medical device regulation (MDR) and the in vitro diagnostic medical device regulation (IVDR)

Inspections, Compliance, Enforcement, and Criminal Investigations

Quality Management Exporter Guideline. IRAQ ICIGI Program Rev 1.1, 18 April 2016

Regulation of Medical Devices in Hong Kong. Dr Terence CHEUNG Principal Medical and Health Officer Department of Health, Hong Kong SAR

Marking of Conformity Assessment

2.1 Reliance Courier Distribution undertakes to deliver to an address specified in the despatch documentation, not to a specific person.

Measures on Open Environmental Information (for Trial Implementation)

Abdul Aziz AL JUAITHEN

UPDATES ON THE ASEAN MEMBER STATES. Washington, September 2014

EXPORTER AND IMPORTER GUIDELINES KSA CAP EXPORTER AND IMPORTER GUIDELINES

PROPOSED DOCUMENT. International Medical Device Regulators Forum. Title: Common Data Elements for Medical Device Identification

Medical Device Regulation Overview

Guideline on Good Pharmacovigilance Practices Module V- Pharmacovigilance System Master File

An Introduction to the Directive 2014/31/EU The new NAWI Directive

EU MDR Timeline. Dr. Christian B. Fulda Jones Day FDLI Annual Conference Access materials at fdli.org/annual2018

DECREE - 8,077, OF AUGUST 14TH, 2013

Implementation of the Food and Drug Administration Food Safety Modernization Act

FINAL DOCUMENT. Global Harmonization Task Force (revision of GHTF/SG1/N29:2005)

GUIDELINES ON THE PRINCIPLES OF GOOD DISTRIBUTION PRACTICE OF ACTIVE SUBSTANCES FOR MEDICINAL PRODUCTS FOR HUMAN USE

TRADE ACCOUNT MINIMUM ADVERTISED PRICE POLICY

PROCEDURE FOR PREPARING AN EMISSION SUMMARY AND DISPERSION MODELLING REPORT [GUIDELINE A-10]

CERTIFICATION SCHEME FOR TYPE APPROVAL OF PRODUCTS

Conformity Risk Assessment To Comply With RE Directive 2014/53/EU

REGULATION REGARDING EQUIPMENT PROCUREMENT PROCESS AND APPROVAL OF MANUFACTURERS FOR NUCLEAR FACILITIES

Definitions of Suspect Product and Illegitimate Product for Verification Obligations Under the Drug Supply Chain Security Act Guidance for Industry

Guidance for Industry Product Name Placement, Size, and Prominence in Advertising and Promotional Labeling

DRUG DRAFT GUIDELINE FOR REGISTRATION OF MEDICAL DEVICES FOR HUMAN USE IN UGANDA, SEPTEMBER 2009

Corporate trademark graphic standards

Transcription:

MEDICAL DEVICES SECTOR MDS - G4 GUIDANCE FOR OVERSEAS Version 2

Our mission is to ensure the safety of food. the safety, quality and efficacy of drugs. and the safety and effectiveness of medical devices, by developing and enforcing an appropriate regulatory system. SFDA Our mission is to ensure safety, effectiveness and quality of medical devices and their performance according to their intended purpose and to ensure the safety of related electronic products. Medical Devices Sector I

Table of Contents Preface Medical Devices Interim Scheme Guidance Documents Part I: Preparing for Medical Device A. Designation of Authorized Representatives Part II: Applying for Medical Device Marketing Authorization B. General Provisions and Application Procedure Part III: Activities after the Device Receives Marketing Authorization C. Post-Marketing Responsibility for Advertising and Marketing Material D. Storage, Handling and Transport of Medical Devices E. Post-Marketing Surveillance of Medical Devices III III III 1 1 5 5 7 7 9 10 II

Preface Medical Devices Interim Scheme The Kingdom of Saudi Arabia (KSA) has adopted an interim regulatory scheme comprising a Medical Devices Interim Regulation together with eight Implementing Rules. The Interim Regulation Decree number 1-8-1429 is dated 27 th December 2008 and was published in Umm Al-Qura Journal year 86 Issue No 4249 dated 17 th April 2009. It specifies the overall framework of the regulatory approach for the Saudi marketing authorization and the post-marketing surveillance of medical devices. It places responsibilities on organizations responsible to the importation and distribution of medical devices within the KSA, on authorized representatives of overseas manufacturers and on local manufacturers. Eight Implementing Rules specify and complete the general provisions of the Decree. Guidance Documents The Medical Devices Sector of the Saudi Food and Drug Authority (SFDA/ MDS) is the Regulatory Authority responsible for regulatory scheme. It has issued a number of guidelines to assist an organization operating in the field of medical devices to understand its obligations under the Interim Regulation and the Implementing Rules. These guidelines provide general, informative guidance and should not be regarded as an authoritative statement of the law. If such is required, please refer to the Medical Devices Interim Regulation together with the eight Implementing Rules adopted by the SFDA/MDS. An English translation of the Regulation and the Implementing Rules may be found on the SFDA website at http://www.sfda.gov.sa/en/medicalequipments/topics/interim+e.htm Further information is available from Medical Devices Sector Saudi Food and Drug Authority SFDA 3292 North Ring road Al Nafel Area Unit (1) Riyadh 13312 6288 KSA Tel +966 1 2038222 Fax: +966 1 2757245 www.sfda.gov.sa mds@sfda.gov.sa I I I

Part I: Preparing for medical device marketing authorization A. Designation of authorized representatives Exerts from chapter four of the MEDICAL DEVICES Interim Regulation Article eleven When the manufacturer is located outside the KSA he shall appoint an authorized representative to act on his behalf. Exerts from implementing rule MDS-IR5 Licensing of Authorized representatives Article five: general A. Establishment licensing is intended to ensure that an authorized representative Has been properly appointed to represent a particular manufacturer in the KSA; Possesses a written mandate describing the activities for which it acts on the manufacturer s behalf and these are sufficient to ensure the proper application of the relevant provisions of the Medical Devices Interim Regulation; and Has set up appropriate procedures to comply with the mandated activities. B. Only an authorized representative holding a valid establishment license, issued by the SFDA, is permitted to legally act on behalf of the manufacturer in relation to the relevant provisions of 1

the Medical Device Interim Regulation and the corresponding implementing rules. C. The authorized representative shall have separate licenses for each manufacturer it represents within the KSA. D. Where a manufacturer intends to make available more than one category of medical device to the KSA market, it may designate a different authorized representative for each category. Article six: Activities to be performed by an authorized representative A. The mandate established between the manufacturer and the authorized representative shall be in writing, specify the mandated activities and be subject to the laws of the KSA. B. The mandate shall, at a minimum, allow the authorized representative to: 1. Represent the manufacturer in its dealings with the SFDA. 2. List each medical device category or generic device group intended to be supplied to the KSA market. 3. Access the electronic application form available on the MDMA portion of the SFDA website and provide the SFDA with all necessary supporting documentary evidence, required by CHAPTER II of Implementing Rule MDS IR 6 Marketing Authorization. 4. Cooperate with the SFDA on evaluations and actions taken during market surveillance and/or vigilance procedures described in implementing Rule MDS - IR7 On Post-Marketing Surveillance. 5. Make the following information available to the SFDA when so required in relation to its market surveillance activities. 2

The marketing authorisation issued by the SFDA for the listed medical devices. The documentation which was used to demonstrate compliance with the regulation of the relevant GHTF founding member jurisdictions. The documents approved by the SFDA demonstrating compliance with the specific saudi provisions referred to in Article 6 of implementing Rule MDS IR 6 on Marketing Authorisation. 6. Inform the SFDA of any incidents that have occurred outside the KSA but have consequences for medical devices that have been authorized to be placed on the market of the KSA. The authorized representative shall explain the circumstances and provide information on the corrective action the manufacturer has taken or intends to take. 7. Inform the SFDA of all field safety corrective actions resulting from post-market follow-up investigations performed by the manufacturer for medical devices that have been authorized to be placed on the market of the KSA. The authorized representative shall explain the reason for the corrective action and provide information on the action the manufacturer has taken or intends to take. 8. Cooperate with parties involved in distribution activities, installation and maintenance of medical devices that have been placed on the KSA market under its mandate. C. The manufacturer may, under his sole responsibility, authorize his representative to perform tasks other than those specified in paragraph B of this article. 3

COMMENTS 1. Every manufacturer established outside the KSA (i.e. an overseas manufacturer) is required to appoint an authorized representative resident within the KSA. 2. An authorized representative may represent more than one manufacturer provided it has the necessary mandates and licenses. 3. There is no requirement in the Interim Regulation for an overseas manufacturer to be registered with, or licensed by, the SFDA. However, the authorized representative must be both registered and licensed before it may act on the overseas manufacturer s behalf. A guideline Entitled Guidance For Authorized Representatives is available on the SFDA website and provides an overview of the registration and licensing process. One important requirement is to provide the SFDA with a copy of the written mandate between the overseas manufacturer and authorized representative. The minimum content of the mandate is described in article six of Implementing rulemds-ir5 Licensing of Authorized Representatives (see exert above). 4

Part II: Applying for Medical Device Marketing Authorization B. General Provisions and Application Procedure Exert from chapters two of the MEDICAL DEVICES Interim Regulation Article four: Medical devices may be placed on the market and/or put into service only if they comply with the applicable provisions of this Interim Regulation, as signified by the SFDA issuing the manufacturer with a written marketing authorisation. Article five: Accessories of medical devices shall, for the purpose of this Interim Regulation, be treated as if they are medical devices in their own right and shall comply with all relevant provisions of the Interim Regulation. Article six: To obtain marketing authorisation, medical devices shall comply with the relevant regulatory requirements applicable in one or more of the jurisdictions of Australia, Canada, Japan, the USA and the EU/EFTA, and additionally with provisions specific to the KSA concerning labeling and conditions of supply and/or use. COMMENTS 1. From 14 th February 2011 medical devices that have a SFDA marketing authorization may be placed on the market within the KSA. 2. After 14th August 2011 only medical devices that have a SFDA marketing authorization may be placed on the market within the KSA. 3. Marketing authorization is required for all medical devices whatever their risk class; 5

contact lenses for cosmetic as well as for medical purposes; and laser surgical equipment intended for cosmetic as well as medical purposes. It is not required for medical devices designed and constructed by health facility staff for internal use within that health facility, alone. 4. While an accessory to a medical device is not of itself a medical device, the Interim Regulation applies equally to it (see Article 5 above), and the manufacturer must obtain a marketing authorization to supply it to the KSA. 5. Before an overseas manufacturer may apply, through its authorized representative, for KSA marketing authorization for one of the medical devices it manufacturers, it must. comply with the regulatory requirements that would permit it to place the device on the market of one of the GHTF Founding Member jurisdictions (namely, Australia, Canada, Japan, the USA, or the EU). The manufacturer may choose which of these jurisdictions it uses as the basis of its authorized representative s subsequent SFDA marketing authorization application (MDMA). comply also with provisions specific to the KSA concerning, for example, labeling, storage/installation and advertising. 6. The overseas manufacturer shall provide its authorized representative with the information it requires to complete the electronic application forms found on the Medical Device Marketing Authorization (MDMA) portion of the SFDA s website. For further information on this subject, including language requirements for the documentation to be provided. the overseas manufacturer should refer to: Implementing Rule MDS-IR6 Marketing Authorisation; the electronic application forms found on the Medical Device Marketing Authorization (MDMA) portion of the SFDA s website; and MDS G5 Guidance on Marketing Authorization Procedures. 6

Part III: Activities After the Device Receives Marketing Authorization C. Post-Marketing Responsibility for Advertising and Marketing Material Exerts from CHAPTER FOUR of the MEDICAL DEVICES INTERIM REGULATION Article Thirty Nine: Advertising A. The advertising of a medical device for which the SFDA has not issued a marketing authorisation is prohibited. B. All advertisement material must be approved by SFDA. C. The advertising material shall not mislead the user regarding the performance of the medical device as specified by the manufacturer. D. The advertising to the general public, including on the internet, shall avoid misleading lay persons. E. Any advertising to persons qualified to use medical devices shall include the relevant information compatible with their specific needs. F. Medical sales representatives shall have sufficient knowledge to be able to provide appropriate information about the medical devices they promote. Exerts from Implementing Rule MDS-IR 6 Marketing Authorization Article Eight: Documentary evidence that the medical device complies with the National Provisions of the KSA ~ 7

The applicant shall provide the SFDA with information and documentary evidence, as follows: D. A copy, in electronic form, of the advertising and marketing materials that will be used in the KSA if any. Article Nine: Language requirements for the documentation to be provided to,or kept available for inspection by, the SFDA F. Advertising and marketing information shall be in English and, where justified, in Arabic. COMMENTS 1. When submitting information for a medical device MDMA, the overseas manufacturer s authorized representative is required to provide any existing advertising or marketing material that is intended to be used in the KSA after the medical device has been authorized to be placed on the market. In addition, the overseas manufacturer s authorized representative is responsible for ensuring any subsequent advertising and marketing materials (either newly prepared material or revised versions of the material submitted through the MDMA) are approved by the SFDA before they are used by the overseas manufacturer, its authorized representative or its distributor/s. 2. Once the medical device has been approved by the SFDA for marketing in the KSA, the overseas manufacturer s authorized representative shall write to the SFDA asking for the previously provided advertising and marketing material, if any, to be approved. The letter shall include the Medical Device National Listing Number of the relevant medical device, the proposed target recipient (e.g. nurse, paediatrician, oncologist, radiographer, biomedical engineer, lay person) and intended placement of any advertisement (e.g. newspaper, professional magazine, television, radio, internet, exhibition material, and the like). The overseas manufacturer shall provide its authorized representative with the required information. 8

3. Any newly prepared or revised advertising or marketing material is submitted manually by the overseas manufacturer s authorized representative to the SFDA with a covering letter that includes the Medical Device National Listing Number of the relevant medical device, the proposed target recipient (e.g. nurse, paediatrician, oncologist, radiographer, biomedical engineer, lay person) and intended placement of any advertisement (e.g. newspaper, professional magazine, television, radio, internet, exhibition material, and the like). The overseas manufacturer shall provide its authorized representative with the required information. 4. The SFDA shall review the submitted advertisements and marketing materials to confirm any performance or safety claims are consistent with the information provided with the original MDMA. 5. The SFDA may ask the applicant to provide additional information and documentation before it reaches a decision. 6. Once satisfied, the SFDA will approve the submitted material in writing and it is only then the overseas manufacturer, its authorized representative or its distributor/s may use it. 7. The overseas manufacturer shall, through its authorized representative, supply its distributor(s) with a copy of all approved advertising and marketing materials (including any submitted through the MDMA) and indicate the proposed recipient (e.g. nurse, paediatrician, oncologist, radiographer, biomedical engineer, lay person) and placement (e.g. newspaper, professional magazine, television, radio, internet) of any advertisement. D. Storage, Handling and Transport of Medical Devices 1. The overseas manufacturer shall ensure the medical devices it manufactures are properly packed, handled, stored and transported as they are transferred from the manufacturing site to the customer via the supply chain. To do so, the manufacturer shall take account of the likely modes of transport, the environmental 9

conditions to be encountered both within and outside the KSA (e.g. temperature, humidity, vibrations), and provid adequate protection from physical damage. It shall also: a. provide written instructions on the required transportation, handling and storage conditions, in both the Arabic and English language, to: organizations responsible for transporting its devices; importers; distributors; and users. b. ensure the packaging of a consignment of medical devices is clearly identified and that an individual medical device within the consignment is accompanied by all relevant documentation, such as, the instructions for installation, maintenance and use, in the language required by the Interim Regulation. E. Post-Marketing Surveillance of Medical Devices Exerts from CHAPTER EIGHT of the MEDICAL DEVICES INTERIM REGULATION Article Thirty Where the SFDA has reason to believe a manufacturer, an authorized representative or another party in the supply chain of a medical device has made a misleading or fraudulent claim of the medical device, it shall investigate and take action as appropriate to the circumstances. Article Thirty One If during its market surveillance activities, the SFDA comes across a noncompliance that has implications for public health, it shall alert patients, users or other persons, as appropriate. 10

~ Article Thirty Three The SFDA shall review adverse events reported to its NCMDR and take appropriate action to safeguard public health. Article Thirty Four The SFDA shall establish a mechanism to issue Field Safety Notices to medical device users and, where relevant, patients. Before issuing such a Notice, its text shall be discussed with the organizations responsible for manufacturing the device and supplying it to the KSA COMMENTS 1. CHAPTER EIGHT of the Medical Device Interim Regulation requires the SFDA to take all appropriate measures to ensure that medical devices authorized to be placed on the KSA market are subject to post-marketing surveillance (see exert from the Interim Regulation, above). It comprises two activities, namely medical device adverse event management, of which a medical device vigilance system is an integral part, and market control. Together these help to ensure and maintain a high level of patient health and safety with respect to medical devices. 2. Implementing Rule MDS-IR7 Post-Marketing Surveillance and the associated guidance published on the SFDA website describe the manufacturer s responsibilities in respect of this procedure. These include requirements to: report to the SFDA s National Centre for Medical Devices Reporting (NCMDR), any Field Safety Corrective Action that may affect medical devices supplied to the KSA; and investigate any adverse event of which it becomes aware. If the manufacturer confirms a malfunction or deterioration in the characteristics and/or performance of the medical device, as well as any inadequacy in the labeling or the instructions for 11

use, has led, or might have led, to the death of a patient, user or third person, or to a serious deterioration in the state of health of a patient, use or third person, it shall submit an adverse event report to the SFDA and agree a corrective action plan. 3. An important aspect of effective post-market surveillance is the need of the SFDA, health authority and/or overseas manufacturer to identify an individual medical device so that it may be traced, examined, or recalled after is has been put into service. Therefore, the manufacturer is required to label the medical devices it manufactures with an unambiguous identification, such as batch code / lot number, or serial number, preceded by the word LOT or SERIAL NUMBER (or an equivalent symbol) as appropriate. In general, consumable and single-use devices have a batch code while powered medical devices have individual serial numbers. 12