Weighing Indicator YH-T3. User Manual

Similar documents
Weighing Indicator VI 150

User Manual. XK315A1 water proof series Weighing Indicators. Applications: Electronic platform scale, weighing scale, etc.

SE9000 Digital Weighing Indicator. User manual

WS 10 Weighing indicator USER MANUAL

DWP-5500FP (4 x4 ) DWP-11KRFP (5 X5 ) Weighing Indicator. User Manual. Content

CN Series P.N , Software rev. T 8, Revision A1- July 2014

BEFORE USING THE COUNTING SCALE

User s Guide. Weighing Indicator WIDRA - HF L / HF S 2000/02/11 21:21:58. No. TARE NET GROSS TOTAL

Weighing Indicator Version April 2009

PALLET SCALE CPS SERIES OWNER'S MANUAL

MANUAL JCE-SERIES CHAPTER

OP-903 INDICATOR USER S MANUAL

HX-Series. HX-502 (500g x 0.01g) HX-3001 (3000g x 0.1g. Copyright 2014 American Weigh Scales, Inc. All rights reserved. Rev. 1.0

Models TJ30K TJ15K TJ6K TJ3K. Resolution Ratio 1g 0.5g 0.2g 0.1g Calibration graduation 10d 10d 10d 10d

OPERATION MANUAL. Electronic Hanging Scale NINGBO AUDDA WEIGHING CO.,LTD

High Precision Balances. Operation Manual

VWR B-Series Balances Operation Manual

Use and Maintenance Manual

High Precision Balances

Best Selection for Your Business. Electronic Pricing Scale. User s Manual. (Model: TP-31)

Instruction number PTI-16-03/09/06/A

Operating Manual. Professional Weighing Equipment. TB Series intelligent weighing machine. Touch Color Screen HIGH RESOLUTION BALANCE

CSG A.L.C. COUNTING SCALE USER MANUAL

DC-688. When Accuracy Counts

Inscale Measurement Technology Ltd Tel: Fax:

Table of Contents. 3-1 Unpacking and Checking Installing Components Leveling the Scale... 1

CONTENTS. 3-1 Unpacking and Checking Installing Components Leveling the Scale... 1

Professional Weighing Equipment Digital Weight Indicator

INSTALLATION uline.com UNPACKING ASSEMBLY LEVELING H-1096, H-1097 H-1098, H-1099 DELUXE COUNTING SCALE

Operator Manual. OHAUS Aviator 7000

MEDIZINTECHNIK FÜR TIERÄRZTE

OPERATION MANUAL. Model : DC-190 TERAOKA SEIKO CO., LTD. Edition Month Year October 1998

LOW PROFILE FLOOR SCALE

Contents Balance-pan fixing

Splash-Proof Weighing Scale Operation Manual

Manual. AGRETO Hydraulic Scale AGRETO electronics GmbH

INSTRUCTION MANUAL CONTECH

Pennsylvania Scale Company. Model 40SS Multi-Function Digital Indicator. Operation & Calibration Manual Software Version

Adam Equipment LBH SERIES. (P.N , Revision B2, January 2010)

Kilotech weighing scale KWS-CY. Owner s Manual V1.0

CONTENTS Cautions before using 1-2. Specification and characteristics of products

Professional Weighing Equipment. Wheelchair Scale. Operating Manual LWC 800

Operating manual Pallet truck scale

Operation Manual. Professional Weighing Equipment. MRB Series Mid Resolution Balances. MRB 500g MRB 1200g MRB 2500g MRB 6000g MRB 10000g

Manual. Indicator. Operation & Service. Version 0.1 CONTENT. Contents subject to change without notice

PORTION CONTROL SCALE

BC Series Compact Counting Scale INSTRUCTION MANUAL

WeighSouth. WBT Toploader Balance

EGT-AF08 USER GUIDE WHEEL WEIGHING SYSTEMS WITH VARIOUS PLATFORMS

EC Series Toploading Balances

MODEL ASCS-15 CRANE SCALE

Instruction Manual. Model 2831D. 4 ½ Digit True RMS Digital Multimeter

DS-520 Digital Indicator Operation Manual 73366

HOBO Plug Load Logger (UX ) Manual

Operation Manual. Dual Temperature Digital Dry Block Incubator. Catalog No. INCUBATOR Rev C. April 2016

BA110 Electronic Balance 5000g x 1g. Impact Test Equipment Ltd & User Guide. User Guide

Professional Weighing Equipment REMO/REMO-R. Operation Manual

Operation Manual INCUBATOR2 Dual Dry Block Incubator

Certificate of EC type-examination of a measuring instrument Number: UK/0126/0082

Operating Manual. Professional Weighing Equipment SCT SERIES Small Counting Scales. SCT 600g SCT 1200g

User Manual. Precision Balances XSR models

Operating instructions Analytical balance

KERN EMS Version /2012 GB

Adam Equipment ADK SERIES

PD- OWNER S MANUAL. ECR Interface Scale 9002-PD

X320 IP69K. Digital Indicator Operator Manual 003X M02

Moving money forward UV MG

Professional Weighing Equipment. K-Series Scales. Operation Manual KHR 6000 KHR 123 KHR 502 KHR 3001

AP5K-C Precision AC Double-Pulse Spot Welding Machine User s Manual Shenzhen Will-Best Electronics Co., Ltd

Operating manual Counting balances

Operating manual Counting balances

POWER PINNER BENCH HAND WELDER 7250 OPERATOR S MANUAL

EC Type-Approval Certificate

LC-VS 60 Operation Manual

Jingzhe SLK-5 Electric Multi-rotor agricultural drone User Manual

QUALITY INSTRUMENTS. Annexure A1: Digital Tape Measures & magnetic V Block

Professional Weighing Equipment !!!! Weigh Bars. Dust and Waterproof Scale IP65 !! 1

SERVICE MANUAL. Model: DS-425. Edition Month Year 1 st September nd December rd April th January 2002

PD6770 VANTAGEVIEW LOOP-POWERED PROCESS METER

DIGITAL PORTION SCALE PS-5 OWNER S Manual

4 CONNECTION of L/C and JWI-586: 1 E+ 2 E- 3 S+ 4 S- 5 SHIELD LOAD CELL CONNECTION. 50k~500k ohm

2% & (+ 2% PSI.

Coefficient of Friction Tester

D Sep ATX324 ATX224 ATX124 ATX84 ATY324 ATY224 ATY124 ATY64

Contents

WELMEC European cooperation in legal metrology Guide for the testing of automatic catchweighing instruments

Adam Equipment. PGW..M SERIES (P.N. 8723, Revision B1, July 2011) Software rev.: 3.38 or above

High Precision Balance

15 15A 20 20A 30 30A 40 40A 60 60A VAC VAC VDC 24VAC VAC. 1 Single-phase. SRH VAC Zero cross turn-on

XPI SERIES GRID SOLAR POWER SYSTEM

Contents

TRICOR. TRICOR Series. Technical Datasheet. Mass Flow Meters MISURATORE MASSICO - CORIOLIS MASS FLOW METER CORIOLIS TECHNOLOGY

WELMEC European cooperation in legal metrology Guide for the Testing of Automatic Catchweighing Instruments

PD662 LOOP-POWERED METER

Operating instructions and installation information. METTLER TOLEDO MultiRange ID7-Form-XP application software

FIXED-WINDOW POWER DEMAND CONTROLLER BY ENERGY PULSES CVM-R8D-CPP INSTRUCTION MANUAL

Temperature Controllers: JC Series Digital Indicating Controllers. JC Series. High Performance Controllers...at the lowest prices anywhere!

Operating manual Pallet scale

KERN HFT Version /2013 GB

Transcription:

Weighing Indicator YH-T3 User Manual

Content CHAPTER 1 MAIN SPECIFICATION......-2- CHAPTER 2 INSTALLATION......- 3-2.1 FRONT AND BACK VIEW OF INDICATOR... - 3-2.2 CONNECTING LOAD CELL TO INDICATOR...- 4-2.3 SERIAL COMMUNICATION INTERFACE (OPTIONAL)..- 4- CHAPTER 3 OPERATION...... - 5-3.1 TURNING ON..... - 5-3.2 OPERATION....... - 5- CHAPTER 4 CALIBRATION.....- 7- CHAPTER 5 ERRORS INDICATION......-9- CHAPTER 6 CHARGEABLE BATTERY... -9- CHAPTER 7 MAINTENTANCE.....-10- YH-T3-1 -

Chapter 1 Main Specification 1. Analog: Input Signal: 0 ~ 10mV Converting Speed: 40 times/second A/D code: 360000 Non-linearity: 0.03% Drift: 0.03% Stimulating Voltage: DC+5V Load Cell: 1 350Ω 2. Display Display: 0 ~ 999999 Interval: 1/2/5/10/20/50 (optional) 3. Serial Communication Interface (Optional) Signal: RS232C Transmission Distance: < 20m 4. Ambient Environment Power Supply: AC 220v ( ~15% ~ +10% ); 50HZ (-2% ~ +2%) Ambient Temperature: 0 ~40 Storing Temperature: -25 ~55 Ambient Temperature: 85%RH Warm-up Time: 10~15minutes Fuse: 500Ma 5. Weight: 2Kg ( excluding battery ) YH-T3-2 -

2.1 Front and back view of the Terminal Chapter 2 Installation Front View of the Terminal Back View of the Terminal YH-T3-3 -

2.2 Connecting load cell to Terminal The 9 pin plug is for load cell, please connects the line as following graphs -V Shield End of Indicator +V End of load cell Stimulating Power Supply + - V +IN +V -IN +IN -IN Shield Stimulating Power Supply - Signal Output + Signal Output - Connection of load cell You must cut off power when connecting load cell, the connection must be reliable. After connecting load cell, you should fasten plug with screw-driver. Load cell and Terminal are static sensitive devices, you must adopt anti-static measures during operation. You must mount lightening rod in thunderstorm frequently happening area. 2.3 Serial Communication Interface (Optional) 1. Terminal communicate with computer by serial communication interface. 2. Connection Terminal communication interface adopts 5 pin plug, pin definition is as follows 5 pin ------TXD signal (serial communication data) 1 pin ------ ground line 2 pin ------ shield line The communication interface is 2pin plug. 1 pin ---------signal (serial communication line) 2 pin ---------ground line for communication 3. Parameter 1 Signal: RS232C 2 Baud Rate: 1200/2400/4800/9600 (Ref. Chapter 4 Calibration) 3 Data format:= Weight data (including decimal),ascii Code The low bit is prior to high bit and symbol. Negative symbol is - ; Positive symbol is 0. For example: While Terminal displays 500.00Kg, serial output is 00.005- While Terminal displays 500.00Kg, serial output is 00.0050 YH-T3-4 -

Chapter 3 Operation 3.1 Turning on 3.1.1 Turn on the power, the Terminal performs self-checking and displays current voltage of battery and go into weighing mode. 3.2 Operation 3.2.1 Press for zero return, Terminal returns to zero within the zero range 4%FS. To make sure the stabilization light is on when zero operation. 3.2.2 At weighing mode, press to deduct displaying weight while it is positive and stabilization light is on. Terminal displays net weight 0, tare light is on, press to deduct tare, tare light is off, indicator displays gross weigh. 3.2.3 At weighing mode, press to enlarge the resolution of weight to 10 times. Press again to get back. 3.2.4 At weighing mode, press to lock the weight value. Holding light is on..the weight value will be hold even if you take object off or put it on. Press again to get back 3.2.5 Manual accumulation At weighing mode, while the weight value is large or equal to 20 counts and the value is stable, press to perform manual accumulation operation. Meanwhile, Terminal first shows the total accumulating times (two steps) :[n=] (that means the times displaying), after 1 second later, it displays accumulating times [ ****]; and then to display the total value of accumulation (two step):[ totl =] (that means to display the total value of accumulation), after 1 second later, it displays accumulating value [******], accumulating light is on. Note: the max. accumulation times are 99999 (under the condition of accumulating amount 999999); Prior to the delete operation,the accumulating amount is recorded even power off. YH-T3-5 -

3.2.6 Automatic accumulation At weighing mode, press + together, Terminal performs automatic accumulation, and accumulating light flashes. At automatic accumulation mode, while the weight value is large or equal to 20 counts and the value keeps stable for 1-2 seconds, indicator performs one time automatic accumulation, displaying accumulation times and amount. Note: The accumulating amount is not recorded under automatic accumulating mode when power off. 3.2.7 At weighing mode, press to delete the accumulation times and amount. 3.2.8 Kg/lb diversion At weighing mode, press + together, Terminal performs unit diversion; while the unit is lb, the final decimal of the displaying value is lighted. Note: once you operate accumulating and fail to delete the accumulation amount at one of the unit, unit diverting is not allowed,unless you delete the accumulation amount 3.2.9 preset tare At weighing mode, press + together, Terminal performs tare presetting. The procedure is as follows: Step Operation display Explanation 1 [ ****** ] Weighing display 2 Press + [P00000] Remind to input preset tare 3 Input preset tare, eg 6000 [ 6000] Press 6 times to change the lowest digit to 6, and then press 3 times to set the value as 6000, and press to confirm 4 [ ******] Turn back to weighing mode, displaying the net weight of deducting tare 3.2.10 Automatic dormancy Terminal keeps at stable mode for 2 minutes and no any operating and weighing, it will turn to automatic dormancy, meanwhile the final decimal is lighted. Any operating and weighing will arouse it to working mode. YH-T3-6 -

3.3 At calibration and preset tare mode, some of the keys perform as follows 1. : perform 1 adding, press to add 1 on the lowest digit. 2. : perform moving, press to move the lowest digit to left. 3 : perform inputting, press to input setting value to the indicator 4 : perform exiting, press to exit calibration Chapter 4 Calibration 4.1 Turn on the Terminal and connect load cell correctly. the indicator warms up for 15-30 minutes at no load At weighing mode, press + together for calibration 4.2 Calibration steps Step Operation Display Explanation 1 [ ****** ] Weighing display Setting of interval: Terminal displays the 2 Press + [ d * ] original interval, which can be updated by operator Input updated interval Eg: 1 Parameter range:1,2,5,10,20,50, 3 Press once to change the [ d 1 ] Press to confirm the updated one is value to 1 as same as the original one Press for confirmation 4 [dp *] Input updated decimal Eg: 3 5 Press three times to [dp 3] change the value to 3 Press for confirmation Decimal setting: Terminal displays the original decimal position, which can be updated by operator. Parameter range:0~5, 0 means no decimal, 1~5 separately means the decimal position from 1~5 Press to confirm the updated one is as same as the original one. YH-T3-7 -

6 [F *****] F.S setting: Terminal displays the original F.S,, which can be updated by operator. Input undated F.S. Eg: 6000 Press six times to change If there is no need for zero and F.S 7 the value to 6, and then [ 6000] calibration, press to step 13, press three times to move otherwise you must input F.S value to step 8 and 9. to 6000, press for 8 9 10 11 confirmation Zero calibration: Terminal self-checks the zero position, provided indicator [ nload ] didn t calibrated zero before, it displays the zero inner code, otherwise it go to step 9 [ LoAd] F.S calibration: Zero position is confirmed, you can load weights for the preparation of F.S calibration, that should be >1/2 of the F.S Finish of weight loading, press [ ******] Terminal displays the inner code of the loaded weight [ ****** ] At the stabilization of the value, Terminal will display the weight of loaded weight Input the actual weight of [ 3000] loaded weight Press to step 13, if the inputted Eg: 3000 weight is as same as the loaded weight. 12 Press three times to change the value to 3, and press for confirmation 13 14 [bt *] Baud rate: Terminal displays the original baud rate parameter, which can be updated by operator [ ****** ] Calibration is over, exit to weighing mode YH-T3-8 -

Chapter 5 Error Indication ERRORO The zero inner code is over 260000 when calibration. ERROR1 The loaded inner code is less ; load cell capacity is more. ERROR2 The loaded inner code is more ; load cell capacity is less. ERROR 3 Calibration exceeding HHHHHH The zero lever is too high or load on the platform. LLLLLLL The zero lever is to high or no pan on the platform. Chapter 6 Chargeable Battery 6.1 Turing on the AC power, the Terminal will charge the battery automatically. So if you don t use battery frequently, you should take battery out. Note: red end is +, black end is -. Wrong connection will destroy Terminal. Note: The built-in battery should be fully charged before it is used for the first time. 6.2 Only when you turn off the AC power, and push start key, battery works. Displaying [LouoL] means the insufficient of voltage, it needs charge. 6.3 When you use the battery first time, you should charge the battery for 20 hours in order to prevent low voltage resulted from the self leakage of the battery. 6.4 If you don t use battery for a long time, you should charge the battery for 10-12 hours for each 2 month to prolong using life of battery. 6.5 The battery is easily exhausted products. And it is not granted free guarantee. YH-T3-9 -

Chapter 7 Maintenance 7.1 To guarantee Terminal clarity and using life, the indicator shouldn t be placed directly under sunshine and should be set in the plain space. 7.2 The Terminal can t be placed into the place where the dust pollution and vibration are serious. 7.3 Load cell should connect with Terminal reliably, and the system should be well connected into ground. The Terminal must be protected from high electrical field and high magnetic field. In order to protect the operator, Terminal and relevant device, you should mount lightning rod in thunderstorm frequently happening area. Load cell and Terminal are static sensitive device, you must adopt anti static measures. 7.4 It is strictly forbidden to clean the case of Terminal with intensive solvents (for example: benzene and nitro oils) 7.5 Liquid and conducting particle should not be poured into the Terminal, otherwise the electronic components will be damaged and electric shock is likely to happen. 7.6 You should cut off power supply of Terminal and relevant device before you pull-in and out the connecting line of Terminal and external device. You must cut off power supply of Terminal, before pulling out connecting line of load cell. 7.7 During operation, if trouble occurs, operator must pull off the power supply plug immediately, and user should return this Terminal to our company for repair. Non-weighing manufacturer should not repair it, or by yourself, otherwise further destruction may happen. 7.8 The storage is not granted the free repair guarantee, because it is easily exhausted products. In order to prolong using life, please charge the cell fully before using it. If you don t use the Terminal for a long time, you must charge the cell every two month and for eight hours/each charging time. Moving or installation must be carefully taken and must avoid strong vibration, impact and bump in order to protect the storage cell from being damaged. 7.9 From invoice date, the Terminal has a one-year free repair period. If any non-artificially obstacle about the Terminal happens under correct using conditions within this period, the user is allowed to send the product with its guarantee card (of the correct number) back to our corporation for free repair. The Terminal shouldn t be taken apart, otherwise free guarantee will be cancelled. YH-T3-10 -