Optimizing Purified Sample Recovery on Reverse-Phase HPLC Systems

Similar documents
Automated 2-D HPLC Using Trap Columns for the Fractionation, Isolation and Screening of Natural Products

Automated On-Line SDS Removal from Diminutive Protein Samples Prior to Capillary HPLC Application Note 221

Simultaneous Preparative HPLC Fraction Collection and Column Regeneration

Preparative HPLC: Factors and Parameters that Directly Affect Recovery of Collected Fractions

Fast Preparative Column Liquid Chromatography (PCLC)

PLC Purification Systems

Streamlined Purification of Assay-Ready Oligonucleotides by Automated HPLC

Accessing the E-Gel 96 System via the Quad-Z 215 with Disposable Tips

Reformatting Procedures Employing a Quad-Z 215 with Disposable Tips

Preparative HPLC is still the

Scale-up with the Agilent SD-1 Purification System analytical and preparative runs on a single system

HPLC/FLASH Purification ADVANCE THE PACE OF YOUR DISCOVERY TO ACHIEVE THE RESULTS OF TOMORROW

GPC Clean-up Systems. Benefits to GPC post-extraction clean-up:

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Maximizing Sample Recovery on the ACCQPrep HP125

Agilent SD-1 Purification System. Purify your way SD-1

Simple Techniques for Improving the Isolation of Synthetic Peptides Jo-Ann Jablonski Principal Scientist Waters Corporation

Agilent 218 Purification System. Purify your way

Agilent 1260 Infinity LC & LC/MS Purification Systems. Purify your way

Fully-Automated Workstation for Direct camp Enzyme Immunoassay Kit

The Positive Displacement Pipetting Reference

BÜCHI Labortechnik AG

Solutions Guide. MX Series II Modular Automation for Nano and Analytical Scale HPLC And Low Pressure Fluid Switching Applications

High-Throughput LC/MS Purification of Pharmaceutical Impurities

Hichrom Limited 1 The Markham Centre, Station Road, Theale, Reading, Berkshire, RG7 4PE, UK Tel: +44 (0) Fax: +44 (0)

Optimizing the Purification of a Standard Chiral Compound Utilizing a Benchtop, Multi-Purpose, Semi-Preparative to Preparative HPLC System

Final Report. Characterisation of Sample Report. Job No 2016/11/12-34 AS No. 1234A. Client Example Contact Sample. Signed. Example report.

Techniques for Improving the Efficiency of Large Volume Loading in Preparative Liquid Chromatography

Fast and Comprehensive 2-Dimensional Liquid Chromatography of Block Copolymers through Temperature Gradient Driven Trapping

What differentiates Dynamic Extractions?

Overview of preparative HPLC. Analytical Technologies Limited

Flash Chromatography & Preparative HPLC The right solution for every separation problem

Fundamentals and Techniques of Preparative HPLC. Parto Zist Behboud Tel: 42108

Flexibility with. Preparative HPLC

A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts

Performance characteristics of the 1260 Infinity Quaternary LC system

Preparative Purification of Corticosteroids by HPLC; Scalability and Loadability Using Agilent Prep C18 HPLC Columns Application

Agilent Prep LC Columns for Small Molecules and Biomolecules MAINTAIN RAPID, RELIABLE SEPARATIONS AS YOU SCALE-UP

Introduction Ron Majors is a Senior Scientist at Agilent. Bill Champion is a chemist in Agilent s HPLC Columns tech support group.

Automated fraction re-analysis does it really make sense? Application. Udo Huber. Abstract

Encompassing Method Development, Small Scale Purification, and Post-Purification Analysis with the Investigator SFC System

Determination of Veterinary Drug Residues in Fish by an Automated SPE-HPLC System Joan M. Stevens, Ph.D., Mark Crawford, and Mike Halvorson, Ph.D.

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

Bivalirudin Purification:

Column Watch. Are You Getting the Most Out of Your HPLC Column?

ultra why micro-lc? product note introducing the ExpressLC -Ultra system: UHPLC with all the advantages of micro-lc

Personal Organic Purification

Method Translation in Liquid Chromatography

Gel filtration using ÄKTA start

DIDANOSINE ORAL POWDER Final text for addition to The International Pharmacopoeia

Agilent 1260 Infinity Purification Systems. Infinitely better purity and recovery

Automated alternating column regeneration on the Agilent 1290 Infinity LC

Developing Purification Strategies for the Agilent 1260 Infinity II Preparative LC/MSD System

Automated Purification Software for Agilent OpenLAB CDS

Gold Standard for recovery and purity

Improving Resolution and Column Loading Systematically in Preparative Liquid Chromatography for Isolating a Minor Component from Peppermint Extract

Essentials in " HPLC '88. Poster Presentation. Purification of Synthetic Peptide M by Reverse Phase HPLC. M. Knight, M.P. Strickler, M.J.

Inertsil ODS-3 2 µm. Base Silica Physical Properties and Chemical Modification. Base Silica: High Purity Silica Gel % Surface Area: 450 m 2 /g

Integrated IC Systems Installation Qualification Dionex Corporation

Product Guide for LudgerSep TM N2 High Resolution Amide HPLC Columns for Glycan Analysis

Instructions for Capillary Electrophoresis Peptide Analysis Kit

A Simple Rapid and Sensitive Method Development for Quantification of Quetiapine Fumarate in Bulk and Dosage Forms Using RP-HPLC

Effective Scaling of Preparative Separations Using Axial Compression Columns

Lifetime stability of size exclusion chromatography columns for protein aggregate analysis

On-Line. User s Guide SPE CARTRIDGES. for Rapid Cleanup and Extraction of Analytes

MPLC. Simply Efficient! Watchers Flash Flash Chromatography by ISU. Ver

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations

Chromatography Column Performance and Data Analysis Success Guide. Hints and Tips for Better Purifications

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

Bio-Scale Mini Affi-Prep Protein A Cartridges, 1 and 5 ml. Instruction Manual. Catalog # ,

Experiment 11. DNA Composition by HPLC 1,2

High Throughput Sub-4 Minute Separation of Antibodies using Size Exclusion Chromatography

Reveleris Technologies

Agilent 6430 Triple Quadrupole LC/MS System

Practical Aspects of Preparative HPLC in Pharmaceutical Development and Production

Application Note. Rapid optimisation and development of an automated two-step purification procedure for monoclonal IgG antibodies.

Qualification of High-Performance Liquid Chromatography Systems

AUTOMATIC PEROXYACETYL NITRATE (PAN) ANALYZER

TSK-GEL BioAssist Series Ion Exchange Columns

Application Note. Fast online SPE purification of Stevia plant extracts. Summary. Introduction. Experimental: Sample extraction

Customized systems to meet your lab s needs

Agilent LC and LC/MS Purification Solutions. Purify your way

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

ITSP Extraction of Cortisol from Plasma

Solubility Studies Utilizing a Phase Monitor

Prominence Preparative HPLC System

Fast Agilent HPLC for Large Biomolecules

Agilent 1260 Infinity Bio-inert Quaternary LC System. Infinitely better for bio-molecule analysis

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions

Impurity Control in the European Pharmacopoeia

CryoTrap. For Agilent Technologies 7890/6890/6850 GC Systems. CryoTrap

Preparative HPLC System

SAFETY AND QUALITY CONTROL

Econo-Pac High Capacity Ion Exchange Cartridges, 5 ml Instruction Manual. Catalog Numbers , ,

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

How do you maximize purity and yield? 2 Continuous counter-current chromatography

Application Note. Authors. Abstract. Introduction. Pharmaceutical

Protein-Pak Hi Res HIC Column and HIC Protein Standard

Transcription:

Optimizing Purified Sample Recovery on Reverse-Phase HPLC Systems Application Note 209 Joan Stevens, PhD, Alan Hamstra, Luke Roenneburg, Tim Hegeman (Gilson, Inc.) Craig Esser, PhD, Regina Black, Derek Von Langen, PhD (Merck & Co., Inc.) Introduction Today s researchers process hundreds of thousands of compounds per month. They face many challenges when trying to accomplish this daunting task. Two factors that directly affect the outcome of this task are purity and recovery. Acceptable limits for purity and recovery need to be established before the compounds can be processed. If time is taken to develop, optimize, and validate an analytical method, then a robust analytical method can be scaled to preparative when processing the samples. Many columns allow the direct transfer from analytical to preparative chromatography, since the same packing media is used in both types of columns. Although transferring to a preparative mode may seem linear, there are many factors that need to be addressed in order to achieve optimum purity and recovery. The purpose of this application is to present the factors that can directly affect purity and recovery when processing compounds on a preparative level. The following parameters will be addressed: Characteristics of short, fast preparative columns General gradients for classes of compounds Determination of collection valve switch time Determination of detector time constant Injection techniques that optimize sample loading Accurate determination of delay volume from detector to fraction collection valve Determination of software response speed Materials & Methods Instruments and Accessories Gilson 215 Liquid Handler, equipped with: 175-mm Z-arm, 819 Injection Module with 5.0-mL stainless steel loop, and beveled-tip probe (269 x 1.3 x 0.8 mm ID) Gilson VALVEMATE Valve Actuator, equipped with: 7610 Rheodyne valve (10-port, 2-position) Gilson 322 HPLC Pump, equipped with H2 pump heads (flow rate up to 30 ml/min.) August 2003 Page 1 of 8 319309-01

Gilson 307 HPLC Pump, equipped with 10 ml/min. pump head (regeneration pump) Gilson 155 UV/VIS Dual-wavelength Detector (0.2-mm pathlength, 1.0 AUFS) MetaChem Inertsil ODS-2 Column (50 x 21.2 mm, 5 µ) YMC ODS Column (50 x 50 mm, 10 µ) Purity vs. Recovery Purity and recovery are used to define the capabilities of a system to collect fractions. Although each describes a unique feature of fraction collection, the two concepts are related. Purity is usually determined by means other than UV detection (e.g., MS, DAD). This allows for the evaluation of the component that is responsible for the absorbance peak in the chromatogram. Determination of whether one compound or more than one compound is responsible for the peak is what defines the level of purity. Recovery is based on the amount of product collected; the purer the compound, the greater the recovery which leads to enhanced synthesis and assay development. Key Factors of Fast Preparative Reverse-Phase HPLC Cycle time per sample is important What purity is required? (80, 85, 90, 95, >95%?) How to define purity? (UV, ELSD, MS,?) What recovery is acceptable? (80, 85, 90, 95, >95%?) Acceptable limits of purity and recovery may limit cycle time, and, therefore, productivity Key Factors of Fast Preparative Methods Develop general gradients for classes of compounds For more polar molecules: For average polarity molecules: For more lipophilic molecules: 3 50% AcCN/water 10 95% AcCN/water 40 100% AcCN/water Use short columns (50 100 mm), small diameter packing material (5 or 10 µ), and high flow rate Cycle time per injection of 3 minutes to 10 minutes Approximately 100-mg injections. Based on 21.2-mm ID columns. Sample purity requirements may limit injected sample mass Key Factors of Fast Preparative Collection Applications Conventional prep peak widths are from 1 to 20 minutes Fast preparative chromatography may have total run times of 3 to 10 minutes per sample Fast preparative peak widths are from 10 to 60 seconds Time resolution of all hardware components becomes significant August 2003 Page 2 of 8 319309-01

Figure 1: Preparative System Hardware Figure 2: Fast Preparative Chromatogram Collection Valve Switch Time Switch time should only be a few milliseconds A switch time of 1 or more seconds is too long At 40 ml/min. and less than 5-minute run times, the peak widths can be between 10 and 20 seconds August 2003 Page 3 of 8 319309-01

Detector Time Constant Most detectors have a detector noise filter which smooths the data output signal. This may be labeled: Peak Width Time Constant Rise Time Detector signal smoothing takes time; set the time constant to the fastest setting With fast flow rates and narrow peaks, the sample may reach the fraction collector before the software receives the signal from the detector A delay of 2 or 3 seconds can result in the loss of 50% of the sample, relative to a peak width of 5 seconds for a given peak With fast chromatography, as the value for the detector time constant increases, mass recovery decreases Detector Time Constant (sec.) Yield (%) 0 >95 25 93 50 77 75 47 Table 1: Detector Time Constant Using a DMSO Solvent Sandwich Using a small amount of a strong solvent to buffer the sample from the highly aqueous mobile phase (in the plumbing before the column, at initial conditions) can be beneficial in preventing precipitation. However, a sandwich solvent can cause degradation of resolution. Therefore, care should be taken with how much additional strong solvent is injected with the sample. The use of a sandwich solvent should not affect mass recovery. Sandwich Solvent (µl) Yield (%) 0 84 50 89 100 >90 150 85 200 84 250 84 Table 2: Using a DMSO Solvent Sandwich Delay Volume (I) For example: 5 feet of 0.030 inch internal diameter tubing contains approximately 695 microliters. At a flow rate of 20 ml/min., the delay time is 2.1 seconds. 10 feet of 0.040 inch internal diameter tubing between the detector and fraction collector contains 2,470 microliters. At a flow rate of 40 ml/min., the delay time is 3.7 seconds. August 2003 Page 4 of 8 319309-01

Delay time the time it takes the sample to travel from the detector to the fraction collector dispense head Figure 3: Delay Volume or Time It is important to accurately determine the delay volume or time to optimize the collection of desired fractions. Method Used to Determine Delay Time 1) 1 mg of Coumarin 314 was injected. A gradient from 50 100% AcCN/water was used to elute the sample. Detector wavelength was set at 436 nm. 2) The signal readout on the front of the UV detector was carefully monitored. 3) A stopwatch was started as soon as the signal appeared on the UV detector. 4) The stopwatch was stopped as soon as the dye reached the fraction collector valve. 5) Using a minimal amount of tubing (1.1 ml) between the detector and fraction collector, at 20 ml/min., the delay time is 3.3 seconds. 200 mv 100 mv 0.0 mv Graph 1: Delay Time Setting Confirmed by Reanalyzing Collected Fractions Individual fractions collected by time. Approximately 20 fractions per minute. Software Processing Speed At a flow rate of 20 ml/min., it takes a sample 2.1 seconds to travel through 5 feet of 0.030 inch internal diameter tubing. It may take the software data controller longer than 2 seconds to process the following data: Is a peak being eluted from the column? Does this peak meet set peak identification parameters? What is the delay from the detector to the fraction collector? Start collection August 2003 Page 5 of 8 319309-01

Delay Volume (II) It may be necessary to add tubing between the detector and fraction collector to increase the delay volume. This will overcome the inherent delay times associated with the detector time constant, fraction collector valve switch time, and data processing. One of the problems associated with additional tubing lengths is peak spreading. However, at these increased flow rates, peak spreading through 10 feet of tubing is negligible. Graph 2: Example of the Peak Spreading in 10 ft. of 0.040 in. ID Tubing at a Flow Rate of 40 ml/min. Peak spreading was determined by placing two detectors in series within an HPLC system with 10 ft. of 0.040 in. ID tubing between them. The sample was injected and the response from each detector recorded. Column used: Metachem Inertsil ODS-2 (50 x 21.2 mm, 5µ). 200 mv 100 mv 0.0 mv Graph 3: Closeup view of the Peak Spreading in 10 ft. of 0.040 in. ID Tubing at a Flow Rate of 40 ml/min. August 2003 Page 6 of 8 319309-01

Delay Volume (III) 10 feet of 1/8 inch outer diameter and 0.062 inch inner diameter tubing contains 5,937 microliters. At a flow rate of 150 ml/min., the delay time is 2.4 seconds. Graph 4: Peak Spreading from 10 ft. of 0.062 in. ID Tubing at a Flow Rate of 150 ml/min. The procedure is the same as desribed for Graph 2. Column used: YMC ODS (50 x 50 mm, 10 µ). Graph 5: Peak Spreading from 10 ft. of 0.062 in. ID Tubing at a Flow Rate of 150 ml/min. Graph 6: Recoveries without 10 ft. of Tubing to Increase Delay Volume Tubing ID: 0.040 inches; flow rate: 40 ml/min. August 2003 Page 7 of 8 319309-01

Graph 7: Recoveries with 10 ft. of Tubing to Increase Delay Volume Tubing ID: 0.040 inches; flow rate: 40 ml/min. Conclusion Paying attention to system parameters directly affects purity level and percent recovery of compounds processed via fast preparative HPLC. The following system parameters were evaluated: Collection valve switch time Detector time constant delay DMSO sandwich solvent Delay volume from detector to collection valve Software response Time restraints and sample throughput may not allow optimization of all of the system parameters evaluated in this application. This application represents the guidelines and avenues for testing the parameters, whether one or all are required. Gilson, Inc. World Headquarters 3000 W. Beltline Hwy., P.O. Box 620027, Middleton, WI 53562-0027 USA Telephone: (1) 800-445-7661 or (1) 608-836-1551 Fax: (1) 608-831-4451 Gilson S.A.S. 19, avenue des Entrepreneurs, BP 145, F-95400 VILLIERS LE BEL France Telephone: (33-1) 34 29 50 00 Fax: (33-1) 34 29 50 20 www.gilson.com sales@gilson.com, service@gilson.com, training@gilson.com August 2003 Page 8 of 8 319309-01