DEVELOPMENT OF DURABLE GOLD VERSUS PLATINUM THERMOCOUPLES

Similar documents
A High-Quality Platinum Resistance Thermometer to 661 C

Thermoelectric properties of currently available Au/Pt thermocouples related to the valid reference function

USING A MINI TRIPLE POINT OF WATER SYSTEM TO IMPROVE RELIABILITY IN A TEMPERATURE CALIBRATION LABORATORY

Available online Journal of Scientific and Engineering Research, 2018, 5(1): Research Article

FIXED POINTS FOR SECONDARY LEVEL CALIBRATIONS

Realization of ITS-90 from K through K: One Company s Approach

REALIZATION OF THE FREEZING POINTS OF INDIUM, TIN AND ZINC USING STAINLESS STEEL CASED CELLS

4 RESISTANCE THERMOMETRY

How Does Temperature Non-uniformity of an Annealing Furnace Affect SPRT Stability?

Temperature Measuring Instrumentation. M H Joshipura Chemical Engineering Department IT, NU

5914A/5915A/5916A 5917A/5918A/5919A Metal Freeze Point Cell. Users Guide

SIB52 THERMO Metrology for thermal protection materials

Overview of Temperature Measurement. The Big Three Thermocouples Resistance Temperature Detectors (RTD s) Thermistors

Selecting the Right Temperature Sensor

Reference thermometer Model CTP5000

Experimental Study and Computer Modeling of the Triple Point of Argon System

Experimental Study and Computer Modeling of the Triple Point of Argon System

Improving the reliability of temperature measurements up to 1550 C

AM1643 Full Immersion Platinum Resistance Thermometer User s Guide

SLIM QUARTZ & METAL CLAD CELLS User Maintenance Manual/Handbook

IMPROVEMENTS RELATING TO THE CALIBRATION OF THERMOMETERS

Temperature calibration

PROGRESS IN HIGH-TEPMERATURE FIXED POINTS USING METAL-CARBON EUTECTICS

ELECTRICAL RESISTIVITY AS A FUNCTION OF TEMPERATURE

SURFACE TEMPERATURE SENSORS

GENERAL INFORMATION ABOUT RESISTANCE THERMOMETERS

How to Calibrate an RTD or Platinum Resistance Thermometer (PRT) 2010 Fluke Corporation Presentation Title 1

Meteorology 432. Thermometry Spring 2013

Guide to the Realization of the ITS-90

Guide to the Realization of the ITS-90

1. TEMPERATURE MEASUREMENT USING ELECTRICAL TECHNIQUES

Primary Standards for Temperature Calibration

Price List. Calibration Fee

3.3 Air temperature. Types of instruments

WORKING STANDARD PLATINIUM RESISTANCE THERMOMETER MODEL 909 User Maintenance Manual/Handbook

DEVELOPMENT OF A LOW DRIFT TYPE K THERMOCOUPLE CABLE FOR AEROSPACE APPLICATIONS

The Measurement of RRR and Resistance for Aluminum Alloy Based on a Cryocooler

Use of W-Re alloys for Manufacturing of High Temperature Thermocouples W-Re5% / W-Re20%

OPERATING MAINTENANCE INSTRUCTIONS

Platinum Resistance Thermometer Pt100 Precision Probe - High accuracy platinum resistance thermometer probe, 6mm diameter x 250mm

Platinum Resistance Thermometer (PRT) Selection Guide

Low Drift Type K And N Mineral Insulated Thermocouple Cable For Aerospace Applications

Lab IV: Electrical Properties

Tin-filled multi-entrance fixed point

Temperature Calibration Technology for Laboratory and Industry

EXPERIMENT FOR VERIFICATION OF GROUND SOURCE HEAT PUMP USING THE DIRECT EXPANSION METHOD

AM1760/1762 Secondary Standard Platinum Resistance Thermometer User s Guide

5. How to Select and Use the Right Temperature Sensor

Midterm Exam. on your way out. Sign - in. Pick up the TAKEHOME Midterm. Complete by Wednesday 8:30am. Turn in beginning of class Wednesday

Appendix 1. Pressure calibration of 12.7 mm diameter piston-cylinder assemblies

General Information, Industrial Thermocouple Section

150 mm. 1/8 BSPP brass male compression fitting with locknut included

FLOW PATTERN IN CO 2 -LUBURICANT TWO-PHASE FLOW AT SUPERCRITICAL PRESSURE

Chapter 11. Measurement of Temparature. Oxford University Press All rights reserved.

Phase Change Cells. T. Shane Topham, Gail Bingham, Harry Latvakoski, Mike Watson

AM1880 Standard Platinum Resistance Thermometer User s Guide

Users Guide. Gallium Melting Point Cell

AM1660 Precision Industrial Platinum Resistance Thermometer User s Guide

ALLOY DESCRIPTIONS. In the sections that follow, each of the common Ney alloys is described and the general characteristics of each are summarized.

Overview. A thermocouple assembly generally has four major components:

The Stability of Metal-sheathed Platinum Thermocouples

Understanding Extrusion

Temperature measurement. Metrology and Instrumentation Week-15-December 20, 2011 Dr. Belal Gharaibeh

Triple Point of Water Cells

ISA Seminars on the Web Live Experts on Hot Topics

Metal Transitions HIGH TEMPERATURE

5. SHEATH MATERIALS, THERMOWELLS, FITTINGS AND TERMINATIONS

ESTAD2017 Diss Tec seminar Measuring of liquid steel temperature in secondary metallurgy. Torsten Lamp, Marek Cichonski, Harald Fischer

ITS.90THERMOSENSORS. Booklet

The results agree with the manufacturer s specifications for the fixed point cells.

Pt100 (100 0 C) to IEC 751, Class B, 4 wire 3.0, 4.0, 4.5 or 6.0mm diameter 316 stainless steel sheath.

T/C, Noble, Straight Ceramic

New Type of Precision Thermometer Offers Unrivalled Accuracy and Stability

How To Select And Use The Right Temperature Sensor

Progress of Vacuum Interrupter (VI) and Recent Technical Trends

Mineral Insulated Platinum Resistance Thermometer PT100 Sensor - Available in 3mm or 6mm diameter x 150mm or 250mm probe lengths

The Development of Ceramic-Based Thermocouples Rowan Technologies - October 2008

ENGLISH Datasheet Platinum Resistance Pt1000 Class B Sensor with Teflon insulated lead

AM1620 Precision Industrial Platinum Resistance Thermometer User s Guide

HIGH TEMPERATURE STANDARD PLATINUM RESISTANCE THERMOMETER MODEL 96178

Chapter 7 Evaluation of Injection-Molding Phenomena Part 1: Measurement of temperature distribution in the molded materials

6. SHEATH MATERIALS, THERMOWELLS, FITTINGS, AND TERMINATIONS

Application of the Optical Floating Zone Apparatus for Researching Material Sciences. Part-1

Datasheet Platinum Resistance Pt100 Dual Element Industrial Sensor Probe 6mm diameter x 150mm or 250mm probe lengths

ENGLISH Datasheet Platinum Resistance Pt100 4 wire class B Resistance Thermometer with DIN B Head

Thermocouple for Flue Gas Temperature Measurements,

Lab #1a Temperature Measurement

THE NEXT GENERATION OF THERMOCOUPLES FOR THE TURBINE ENGINE INDUSTRY

Endless Possibilities Use of Heat Pipe for Global Warming Reduction

SENSOR TIP OPTIMIZATION FOR A THERMAL ANEMOMETER FOR DETERMINING CONVECTION INTENSITY IN QUENCH BATHS

CHARACTERISTICS OF SELF-POWERED NEUTRON DETECTORS USED IN POWER REACTORS

AM1730 Secondary Reference Platinum Resistance Thermometer User s Guide

Cable Selection Guide

Experimental investigation of a novel heat pipe cold plate for electronics cooling

JLAB-TN /15/2007. JLab Cryomodule Thermal Shield Circuit Piping Design per ASME B31.3 Edward F. Daly, Gary G. Cheng and John Hogan

ENGLISH. 1metre 7/.02mm Teflon insulated twin twisted lead

Thermocouple Quick Reference Guide. Aluminium Temperature Sensing Solutions from the Smelter to the Foundry

Transcription:

XVIII IMEKO WORLD CONGRESS Metrology for a Sustainable Development September, 17 22, 2006, Rio de Janeiro, Brazil DEVELOPMENT OF DURABLE GOLD VERSUS PLATINUM THERMOCOUPLES Masahiko Gotoh 1, Hideaki Oikawa 2 1 Tamagawa University, Tokyo, Japan, mgotoh@eng.tamagawa.ac.jp 2 Chino corporation, Tokyo, Japan, oikawa@chino.co.jp Abstract: The gold versus platinum (Au/Pt) thermocouples are very stable under long term repeated usage at higher temperature. Still there are some unsolved problems that the stress-relieving coil at the top of measuring junction is deformed after thermal cycling. In this study twist pair of thermocouple-element and U shape thin Pt wire at the measuring junction instead of the coil are tested in addition to the ordinary measuring junction. It was found that good stability and immersion characteristic is realized by a thermocouple having ordinary junction with appropriate anchoring of the thermocouple elements. Keywords: Au/Pt thermocouple, new hot junction, stability, immersion characteristic. 1. OVERVIEW OF THE Au/Pt THERMOCOUPL The literatures of Au/Pt thermocouples are extremely sparse. Roeser and Wensel gave a table of Themo-electro motive force (Emf) relative to Pt for Au along with other 27 metallic elements in 1941[1]. McLaren and Murdock found that Au/Pt thermocouples have superior immersion characteristic to Platinum-Rhodium thermocouples [2]. Emf variation measured in an uniform temperature zone of a silver fixed point cell corresponds to 0.1 C in the case of Au/Pt thermocouples whereas its corresponds to 0.5 C in the case of type S thermocouples which was the defining instrument of the IPTS-68 from 630.74 C to 1064 C. And then they performed extensive study on the essential thermoelectric properties of the Au/Pt thermocouple. Their results were published in 1987 including a provisional wire scale derived from a 5th degree polynomial [3]. In such pure element thermocouples as the Au/Pt thermocouple substantial differential thermal expansion occurs and mechanical stress due to it causes the strain. The strain induces parasitic Emf, which leads uncertainty of temperature measurement. McLaren and Murdock put a coil of a few windings thin platinum wire to form the measuring junction. This stress-relieving coil succeeded to reduce parasitic Emf and to improve immersion characteristic to 0.05 C at the silver point. This design is adopted by other works [4,5,6,8] since then. Thereafter, more precise reference functions based on the ITS-90 and covering the range from 0 C to 962 C or 1000 C were also published [4,5]. Bentley precisely discussed annealing process to stabilize the thermocouple [7]. Now a days a few kinds of commercial model of the Au/Pt thermocouples are available in the market. Ripple and Burns published a textbook to provide guidance and to document the method of fabrication and calibration of the Au/Pt thermocouple [6]. They describe that Au/Pt thermocouple is more rugged but slightly less accurate compared to Standard Platinum Resistance Thermometers, defining instrument of ITS-90 in the range from 13.6K to 962 C. Relative to the highest quality industrial platinum resistance thermometers, the Au/Pt thermocouple has similar calibration uncertainties and is more resistant to loss of calibration and has a higher upper temperature limit. The combination of ruggedness, a smaller calibration uncertainty, long-term stability of its calibration and simplicity of electronic measurement system make Au/Pt thermocouple ideal for use as a secondary reference standard. However the stress-relieving coil is very fragile and sometimes do not remain its original shape after repeated exposure to high temperature. Therefore this design is not necessary suitable for practical use. Many kinds of attempts are tried to overcome this problem in this study and results are reported in this paper. 2. THE THERMOCOUPLES Six thermocouples were assembled and studied with various source materials. All the wires have approximately 0.5mm diameter. In the table 2 quality of the wire, insulating material and structure of the measuring junction are summarized. 2.1. Heat treatment The thermocouples were prepared following the recommendations of McLaren and Murdock [3]. The platinum wires were electrically-annealed in air at 1300 C and cooled rapidly (quenched) to room temperature and then vacancy-annealed at about 450 C. Gold wire is too soft to be annealed by self-heating with passing electric current through the wire. Therefore they were annealed in a furnace. The gold wires, 1.5 m or 2m in length, were installed in high purity quartz tubes. These tubes (with the gold wire inside) were heated in a conventional tube furnace that has 100cm effective working length over which the temperature uniformity is within 4.6 C. This furnace was operated in the

horizontal orientation. Following the first heat treatment of the first segment, the gold wire was then shifted and the second section is heated in the same way. Thus gold wires are heat-treated by repeating this process. After cooled in the furnace, they are subjected to vacancy-anneal at 450 C. In this study complete anneal ( mode 1) and simplified anneal ( mode 2) are tried. In the mode 1 post assembly anneal process was added in order to anneal out cold work introduced during assembly process. The annealing procedure is summarized in the Table 2. Mode 1 2 Pt Step 1 Wire 1300 C, 10hours 1300 C, 10hours Pt Step 2 Wire 20hours 5hours Au Step 1 Furnace 960 C, 10hours 970 C, 10hours* Au Step 2 Furnace 20hours Post Assemble Furnace 960 C, 1hours Post Assemble Vacancy 20hours none none none Note*: The Au wire was traveling through a tube furnace having effective heating length 30cm at 970 C at a speed of 3cm/hour. High-purity alumina insulating tube Expansion coil, formed from 4-5 turns of 0.12 mmφ platinum wire (a) Type 1 High-purity alumina insulating tube U shape fine Pt wire Table 2. ing procedure 2.2. Stress-relieving coil The stress-relieving coil recommended by McLaren and Murdock [3] is easily deformed after heat cycles and in an extreme case it is tangled and does not functions properly. In order to improve the problem, three types of measuring junction were tried in this study in addition to the stressrelieving coil. In every type both of the insulating tubes and the capillaries were heated at 960 C for 5 hours prior to assembly. Type 1: The same stress-relieving coil as the one reported by McLaren and Murdock is attached to the top of the measuring junction. A thermocouple with this junction was tested as a reference. Type 2: The structure is the same as Type 1 except that a U shape thin Pt wire bridges the end of Au wire to the end of Pt wire to form the measuring junction. It is expected that the thin platinum wire relieves stress effectively enough to prevent parasitic Emf. Type 3: Measuring junction is formed at the top of thermocouple by direct welding together. Twisted quartz capillary is used for the insulator of the thermo-elements. It is expected that the twisted capillary accounts for the differential thermal expansion and prevents stress induced parasitic Emf. Type 4: At the top of the measuring junction Au and Pt wire is directly welded together. This simplest structure was tried in this study again. In the course of our study it was found that stability and immersion characteristic depend not only on the structure of the measuring junction (stressrelieving coil or U shape thin wire bridge) but also on the anchoring methods of the thermometer elements. (b) Type 2 High-purity Quartz tube (c) Type 3 High-purity alumina insulating tube (d) Type 4 Fig.1 Four types of measuring junctions of the thermocouples 2.2. Assembly of thermocouples Welded junction Welded junction In the preliminary study it was found that long-term stability and immersion profile depends in some way on how thermocouple wire is anchored. In the pioneer work by McLaren and Murdock, they succeeded to get very good performance of stability with a bar clamp attached to the harness just beyond the alumina insulator. The bar clamp

presses the thermocouple wires over electrical insulator tube and protect the measuring junction from mechanical damage. On the other hand movement of thermocouple wire due to the differential thermal expansion can be prevented with it and stress induced Emf causes instability of the thermocouple. These phenomena are studied by testing two kinds of thermocouples with anchors and without anchors. Anchors were realized with a lock nut (Fig.2) installed in the handle of the thermocouple. The lock nut squeezes the thermo-element by pressing the wire over electric insulator of the wire (thermocouple G61). Thermocouple G51 has exactly same bar clamp as those of McLaren s work. Lock nut Fig.3 Twist pair of quartz insulating tube. Sand blast is applied. Thermocouple designation Gold wire Platinum wire Insulating Tube Measuring Junction type Mode Anchor Handle C2 99.999% 99.999% 1 1 Yes Fig.2 Thermocouple G61 that has a lock nut and twist pair quartz insulating tube. G51 99.999% 1.3924(Note) 2 2 Yes As is mentioned in section 1 of this paper, stress-relieving coil is easily deformed after thermal cycling. In order to solve this problem many attempts are made in this study. Actual differential thermal expansion is calculated in the case of temperature profile of our furnace for silver fixedpoint cell. In the furnace closed type sodium heat pipe is installed. To calculate precisely enough thermocouple wire is divided into 10mm segments in order to take temperature dependent linear thermal expansion coefficient into account. The distance from the bottom of fixed-point cell to the top of the furnace is 50cm and thermocouple wire expands as follows when it is inserted into the cell at 962 C. Thermal expansion of gold wire: 8.6mm Thermal expansion of platinum wire: 4.6mm U shape fine platinum wire is expected to relieve the stress caused by 4mm differential expansion in addition to pre-equalization of Platinum wire illustrated in the Fig. 1(b). Another consideration against the differential thermal expansion is the use of insulation tube made of twisted pair of quartz capillary (Fig.1(c)). The twisted pair of capillary has made of high purity quartz straight tube having outer and inner diameter 2.7mm and 1.5mm respectively. It is expected that considerably large inner diameter spiral tube moderates thermal expansion and reduces stress induced Emf. In the table 2 summaries of studied thermocouples are tabulated with various design features and heat treatments. Except thermocouple G61, 99.7% purity alumina two bore insulating tubes are used for all the thermocouple. The size of them is 4.7mm outer diameter and 1.5mm bore. G61 99.99% 99.99% C1 99.999% 99.999% C5 99.999% 99.999% C6 99.999% 99.999% Twisted quartz capillary Note; Resitance ratio at 100 C to 0 C Table 2. Studied thermocouples 3 2 Yes 4 1 No 4 1 No 4 1 No 3. THE MEASUREMENT SYSTEM AND TESTS Emf is measured with a high precision digital voltmeter. In this study an Agilent technology model 3458A and a Fluke model 8508A are used with a Data proof model 160B low thermal scanner. By reversing the polarity with the scanner parasitic Emf generated in the wiring system is canceled. Two types of the tests are performed. (a) Drift measurement after long-term exposure to nearly 960 C and (b) immersion profile measurements in the silver fixed point cell while freezing plateau is realized. Measurement procedure of immersion profile is as follows. Waiting until the temperature is stabilized after super-cool test thermocouple is inserted to the bottom of the fixed-point cell. Once it attains equilibrium the thermocouple is moved by 2 cm step after taking the Emf value. This process is repeated until the thermocouple reaches to 12 cm from the bottom of the cell. Then the same process is repeated by inserting the thermometer. The thermometers are exposed to nearly 960 C in a vertical conventional furnace and from time to time it is calibrated at Ag fixed point for drift measurement.

4. RESULTS AND DISCUSSIONS Generally speaking uncertainty of Au/Pt thermocouples is an order of magnitude smaller than that of Platinum- Rhodium thermocouples. In most cases Au/Pt thermocouple is used as a secondary standard thermometer. Once it is calibrated the temperature is calculated with the calibration equation. Therefore deviation of Emf at the fixed point is not so significant but drift is more important factor for its uncertainty estimation. So deviation from the initial value is discussed in the following data analysis. 4.1. Drift test In the following two figures Fig.4 and Fig.5 drift measurement results are shown. Thermocouples C1, C5 and C6 prove to be stable within 7mK up to 800 hours as is shown in the Fig.4. All of those have are Type 4 measuring junction (without stress-relieving coil) and have no anchor. Namely the gold and platinum wires are routed to the reference junction without any clamp. At the top of the reference junction these wires are fastened together to a stainless tube. This structure allows thermocouple wire to expand with heating. Up to 200 hours exposure C2 deviates significantly. However after this point deviation decreased. At this time 0.12mm diameter platinum stress-relieving coil was installed instead of 0.2mm diameter coil. Finer coil brought better performance. This proves that fine coil relieves the stress even thermocouple wires are clamped at the handle. This result is consistent with the former works [3,4,5]. Deviation/mK 30.0 25.0 20.0 15.0 10.0 5.0 0.0-5.0-10.0 Drift ( referenced to the final value) 0 200 400 600 800 Exposure time/hours C1 C2 C5 C6 Fig.4 Drift test for the thermocouple C1, C2, C5 and C6. In the Fig.5 results of the drift test for thermocouple G51 and G61 are shown. These two thermocouples received simplified heat treatment ( Mode 2) and also have different measuring junction type (Type 2 and 3 respectively). Therefore G51 exhibits larger initial drift. But after 500 hours it is stabilized within 5mK. Deviation/mK 100.0 80.0 60.0 40.0 20.0 0.0-20.0 Drift (referenced to the final value) 0 200 400 600 800 Exposure time/hours Fig.5 Drift test for the thermocouple G51 and G61. 4.2. Immersion characteristic G51 G61 Measurements of the immersion profile were repeated 9 times at most during the exposure to 800 hours. Every time the measurement concluded one cycle of insertion and withdrawal. Results are shown in the figures 6, 7, 8 and 9 for thermocouples C1, C5, C6 and C2. The first three thermocouples have the same design (Type 4 junction, no anchor). Width of scattering at 6cm level corresponds approximately to 5mK for C5 and C6 and 8mK for C1. Offset Emf/µv (40mK/µv) Offset Emf/µv (40mK/µv) Fig.6 Immersion characteristic for the thermocouple C1 Immersion Characteristic of the Thermocouple C1 - Immersion Characteristic of the Thermocouple C5 - Fig.7 Immersion characteristic for the thermocouple C5

Offset Emf/µv (40mK/µv) Immersion Characteristic of the Thermocouple C6 - Offset Emf /µv (40mK/µV) Immersion Characteristic of the Thermocouple G51 1 0.5 0 0 4 8 12-0.5-1 -1.5-2 -2.5 Fig.8 Immersion characteristic for the thermocouple C6-3 Among the thermocouple group of C1, C5, C6 and C2 only C2 has anchor mechanism. This difference is demonstrated in the immersion profile of the Fig. 9. Width of scattering at 6cm level corresponds approximately to 13mK for C2. Because this is slightly worse than C5 and C6 anchoring the wire at the handle possibly disturb stressrelieving mechanism. Offset Emf/µv (40mK/µv) Immersion Characteristic of the Thermocouple C2 with 0.12mm coil - Fig.10 Immersion characteristic for the thermocouple G51 5. CONCLUSIONS (1) Without stress-relieving coil good long time stability and good immersion characteristics is attainable by not anchoring the thermocouple wire (Thermocouple C1, C5, and C6). To fabricate the proper reference junction is necessary to prevent any movement of the thermocouple wire through reference junction side (Fig.11). Thus we have succeeded to remove fragile stress- relieving coil without any degradation of the performance of the thermocouples. They are stable within 7mK. Immersion characteristic is within 5mK for C5 and C6 Fig.9 Immersion characteristic for the thermocouple C2 G51 has U shape fine platinum wire to form the measuring junction (Type 2 Junction). Fig.10 is immersion profile of the thermocouple. Width of scattering at 6cm level corresponds approximately to 50mK. Although profile is getting better with exposure time it is suspicious that the U shape junction properly relieves differential expansion. Fig.11 Picture of the thermocouple. In the handle of Ice point probe, thermocouple wire is anchored. Flexible stainless bellows connecting thermocouple and reference junction protects thermocouple wire from any stress for the thermocouple. (2) Thermocouple C2 has the same stress-relieving coil and anchoring construction. This behaves exactly same as the one reported in the other studies [3,4,5] in

terms of immersion characteristics and drift. Drift is within 5mK after initial aging. Immersion characteristic is within 13mK. (3) Validation of Type 2 and Type 3 junction has not yet completed (Thermocouple G51 and G61). For practicality and further mechanical ruggedness it is preferable that thermocouple wires are anchored in the handle. Further study is necessary to prove Type 2 and Type 3 junction to work properly. REFERENCES [1] Roeser Wm.F. and Wensel H.T., In Temperature Its measurement and control in science and industry Vol.1 Reinhold Publishing Co., New York, Table 12 Appendix, p 1309, 1941 [2] McLaren E. H., Murdock E. G., Properties of some noble and base metal thermocouples at fixed points in the range 0-1100 C In Temperature Its measurement and control in science and industry Vol.5 American Institute of Physics. New York, p 953-975, 1982 [3] McLaren E. H., Murdock E. G., The Pt/Au Thermocouple, NRCC Report 27703, 1987 [4] Gotoh M., Hill K. D., Murdock E. G., A Gold/Platinum thermocouple reference table Rev. Sci. Instrum, 1991, 62, 2778-2791 [5] Burns G. W., Strouse G. F., Liu B. M., Mangum B. W., Gold versus platinum thermocouples: performance data and an ITS-90 based reference function, In Temperature: Its Measurement and Control in Science and Industry, Vol. 6 (Edited by J. F. Schooley), New York, American Institute of Physics, 1992, pp.531-563 [6] Gotoh M., Iwata N., Funayama K., Long term stability test for Gold/Platinum thermocouples. In TEMPMEKO 96:Proc. 6 th International Symposium on Temperature and Thermal Measurements in Industry and Science Torino, 1996,p 207-211 [7] Bentley, R.E., The use of elemental thermocouples in high-temperature precision thermometry. Measurement Vol.23 1998,P35-46 [8] Dean C. Ripple, George W. Burns, Standard Reference Material 1749:Au/Pt Thermocouple, NIST Special Publication 260-134, 2002