Journal of Chemical and Pharmaceutical Research, 2017, 9(7): Research Article

Similar documents
ISSN India; g,secunderabad. Abstractt. a flow rate. of 1ml/min. di hydrogen. which acts. and chronic. including minimize (5) Figure

Validated Stability-indicating assay method for determination of Ilaprazole in bulk drug and tablets by high performance liquid chromatography

Available Online through (or) IJPBS Volume 2 Issue 3 JULY-SEPT Research Article Pharmaceutical Sciences

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy

Research Paper Development of Stability Indicating Reverse Phase HPLC Method for Aripiprazole from Solid Dosage form

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

RP-HPLC METHOD FOR QUANTITATIVE ESTIMATION OF GLATIRAMER ACETATE FOR INJECTION IN PHARMACEUTICAL DOSAGE FORMS

INTERNATIONAL PHARMACOPOEIA MONOGRAPH ON ARTEMETHER AND LUMEFANTRINE CAPSULES REVISED DRAFT FOR DISCUSSION

International Journal of Pharma Research & Review, Feb 2014; 3(2):11-16 ISSN:

ANALYTICAL METHOD DEVELOPMENT FOR IMPURITIES PRESENT IN HYDROXY CHLOROQUINE SULPHATE FROM ANTIMALARIAL TABLET FORMULATION

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

A green HPLC technique with a 100% water mobile phase for detecting imidacloprid and its metabolite 6-chloronicotinic acid

IJSER. (2E)-2-cyano-3-(3,4-dihydroxy-5-nitrophenyl)-N,N-diethylprop-2-enamide [Entacapone] 7

VICH Topic GL2 (Validation: Methodology) GUIDELINE ON VALIDATION OF ANALYTICAL PROCEDURES: METHODOLOGY

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

QbD approach for analytical method development of anti-pschotic drug

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

Preparative HPLC is still the

International Journal of Medicine and Nanotechnology

Analytical Methods Development and Validation

HPLC METHODOLOGY MANUAL

Analysis of amoxicillin and five impurities on the Agilent 1220 Infinity LC System

N. Tamilselvi *, Dona Sara Kurian. Department of Pharmaceutical Analysis. KMCH college of pharmacy Coimbatore

Fast Preparative Column Liquid Chromatography (PCLC)

Genotoxicity is the property of a compound

Development of Quality Control Method for Dissolution Analysis of Tapentadol and paracetamolin tablet

Certified Reference Material - Certificate of Analysis

Bivalirudin Purification:

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

Development & Validation of RP-HPLC Method for Estimation of Dabigatran Etexilate Mesylate from Capsule Dosage Form

APPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology

CORESTA RECOMMENDED METHOD N 61

Degradation Pathway for Pitavastatin Calcium by Validated Stability Indicating UPLC Method

Phase Appropriate Method Validation

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Saudi Journal of Medical and Pharmaceutical Sciences. Research Article. Available Online: 70

Polyvidone Polyvinylpyrrolidone H 2 C H C N

Peptide Mapping: A Quality by Design (QbD) Approach

HPLC Studies on Degradation Behavior of Brivaracetam and Development of Validated Stability Indicating HPLC Assay Method

Optimizing the Purification of a Standard Chiral Compound Utilizing a Benchtop, Multi-Purpose, Semi-Preparative to Preparative HPLC System

Improving Resolution and Column Loading Systematically in Preparative Liquid Chromatography for Isolating a Minor Component from Peppermint Extract

Heart-cut 2D-LC/MS approach for pharmaceutical impurity identification using an Agilent 6540 Q-TOF LC/MS System

Preparative Purification of Corticosteroids by HPLC; Scalability and Loadability Using Agilent Prep C18 HPLC Columns Application

International Journal of Pharmacy

Novel Rp-Hplc-Pda Method for the Simultaneous Estimation of Metoprolol Succinate and Chlorthalidone in Bulk and Pharmaceutical Dosage Forms

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental

Determination of Iron Content in Different Hemoglobin Samples from Some Patients by UV-Visible Spectrophotometer

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis

Forced degradation studies and validated stability indicating HPTLC method for determination of miconazole nitrate in soft lozenges

Qualification of High-Performance Liquid Chromatography Systems

Analysis of Illegal Dyes in Food Matrices Using Automated Online Sample Preparation with Liquid Chromatography-Mass Spectrometry

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Research Article. Optimization and validation of rapid and simple method for determination of Isoniazid and Pyrazinamide in plasma by HPLC-UV

Method development and validation for the quantitative estimation of cefixime and ofloxacin in Pharmaceutical preparation by RP- HPLC

Gas Chromatography Assignment Chem 543/443

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Fundamentals and Techniques of Preparative HPLC. Parto Zist Behboud Tel: 42108

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

AdvancedTools in HPLC methoddevelopment

Quality by Design (QbD) Based Development of a Stability Indicating HPLC Method for Drug and Impurities

Method Development and Validation for Online UV-Dissolution Methods Using Fiber-Optic Technology

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

Case study 2: Parenteral Drug Product

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

USP Chapter 823 USP 32 (old) vs. USP 35 (new)

Developing and Validating Dissolution Procedures for Improved Product Quality

Determination of Adenovirus pviii (31K) Concentration for Estimation of the Empty Capsid Concentration of the Adenovirus Reference Material (ARM)

Lifetime stability of size exclusion chromatography columns for protein aggregate analysis

Application Note. Author. Abstract. Small Molecule Pharmaceuticals. Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Method Translation in Liquid Chromatography

Verification of Compendial Methods

HEPARIN CALCIUM. Heparinum calcicum

Determination of Sulfite in Food and Beverages by Ion Exclusion Chromatography with Pulsed Amperometric Detection

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates

JPMA 06, Dec., Case Studies of Established Conditions in Analytical Procedures. Biopharmaceutical Committee Technology Working Committee

A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts

HPLC to UPLC Method Migration: An Overview of Key Considerations and Available Tools

New Core-Shell Technology

Overview of preparative HPLC. Analytical Technologies Limited

Forced Degradation of Ibuprofen in Bulk Drug and Tablets

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions

API Stability Protocols and. Chris Byrne Tasmanian Alkaloids

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

C196-E067A. Prominence UFLCXR. Shimadzu Ultra Fast Liquid Chromatograph

ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER

Bioanalytical method validation: An updated review

Venkatesh S et al. IRJP 1 (1)

PAHs in Surface Water by PDA and Fluorescence Detection

Formulas: See Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1. METHOD: 2555, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Kalyani G. et al.; International Journal of Pharmamedix India, 2013, 1(2),

Determination of organochlorine pesticide residues in eggs by gel-permeation chromatography (GPC) cleanup and GC-ECD

BÜCHI Labortechnik AG

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website: Published in November 2013

Validation of a concentration assay using Biacore C

DETERMINATION OF MYCOTOXINE DEOXYNIVALENOL IN WHEAT FLOUR AND WHEAT BRAN

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

Transcription:

Available online www.jocpr.com Journal of Chemical and Pharmaceutical Research, 2017, 9(7):55-61 Research Article ISSN : 0975-7384 CODEN(USA) : JCPRC5 A Validated RP-HPLC Method for the Determination of 2-Chloroadenosine as Process Related Impurity in Regadenosen Parenteral Dosage Form Y Rajesh Babu 1, VVSS Appala Raju 2 and N Appala Raju 1* 1 Department of Pharmaceutical Chemistry, Sultan Ul-Uloom College Of Pharmacy, Hyderabad, Telangana, India 2 Deparment of Chemistry, Faculty of Pharmacy, MAHSA University, Koula lumpur, Malaysia ABSTRACT Regadenoson is an A 2A adenosine receptor agonist that is a coronary vasodilator. Regadenoson is chemically described as adenosine, 2-[4[(methylamino) carbonyl]-1h-pyrazol-1-yl]-, monohydrate. Regadenoson is a low affinity agonist (Ki 1.3 μm) for the A2A adenosine receptor, with at least 10-fold lower affinity for the A 1 adenosine receptor (Ki > 16.5 μm), and weak, if any, affinity for the A 2B and A 3 adenosine receptors. Activation of the A 2A adenosine receptor by regadenoson produces coronary vasodilation and increases coronary blood flow (CBF). A simple and precise reverse phase high performance liquid chromatography (RP-HPLC) method has been developed and validated which can separate and accurately quantitate 2-chloro adenosine related compounds. The capability of the method was demonstrated through adequate separation of all potential 2-chloro adenosine related compounds from active pharmaceutical ingredient and also from each other that are present in aged and stress degraded 2-chloro adenosine stability samples. Successful separation of 2-chloro adenosine from its one process impurity and degradation impurities formed under stress conditions was achieved by using a isocratic elution at a flow rate of 1.0 ml/min on a symmetry C18-3v column (250 mm 4.6 mm, 5 µm particle size, 100 Å pore size) at ambient temperature. Mobile phase is the mixture of 1:1 v/v of acetonitrile and methanol; (ph-adjusted to 3.2 with 10% v/v ortho-phosphoric acid) in gradient mode was found to be best suitable for separation of impurity. UV detection at 205 nm was employed to monitor the analytes. The method was successfully validated in accordance to ICH guidelines acceptance criteria for system suitability, specificity, linearity, range, precision, accuracy, limits of detection and quantification for the impurities, and robustness, following the ICH guidelines. Therefore, the proposed method was suitable for the simultaneous determination of 2-chloro adenosine and its process related impurities. Finally, the applicability of the method was evaluated in commercial dosage form analysis as well as in stability studies. Keywords: 2-chloro adenosine; Degradants; Validation; Stability indicating method INTRODUCTION Regadenoson is an A 2A adenosine receptor agonist that is a coronary vasodilator. Regadenoson is chemically described as adenosine, 2-[4[(methyl amino) carbonyl]-1h-pyrazol-1-yl]-, monohydrate [1,2]. Regadenoson has the empirical formula C 15 H 18 N 8 O 5 H 2 O and a molecular weight of 408.37. Lexiscan contains Regadenoson as active component, is a sterile, nonpyrogenic solution for intravenous injection. The solution is clear and colorless. Each 1 ml in the 5 ml pre-filled syringe contains 0.084 mg of regadenoson monohydrate, corresponding to 0.08 mg regadenoson on an anhydrous basis, 10.9 mg dibasic sodium phosphate dihydrate or 8.7 mg dibasic sodium phosphate anhydrous, 5.4 mg monobasic sodium phosphate monohydrate, 150 mg propylene glycol, 1 mg edetate disodium dihydrate, and Water for Injection, with ph between 6.3 and 7.7. Lexiscan (Regadenoson) injection is a pharmacologic stress agent indicated for radionuclide myocardial perfusion imaging (MPI) in patients unable to undergo adequate exercise stress [3-5]. The novel synthetic method has been developed by concerning the yields and 55

mild reaction conditions to synthesize regadenoson from stable starting material such as 2-chloro adenosine as stable starting material. Besides the simple route of synthesis for regadenoson, we have observed that 2-chloroadenosine is present as potential impurity in our synthetic process and also appears in finished product. The aim of this paper was to develop validate a simple and reliable RP-HPLC method for the determination of 2-chloroadenosine as process related impurity in regadenoson parenteral dosage form. Hence, a RP-HPLC method for determination of 2- choloroadenosie was developed and validated as per International Conference on Harmonization (ICH) guidelines [6-9]. This manuscript describes the development and validation, in accordance with ICH guidelines, of a rapid, economical, precise, and accurate reversed-phase HPLC method in gradient mode for analysis of 2-chloro adenosine as process related impurity [10]. To the best of our knowledge, no analytical methods are currently available to separate and quantitate the known process related compounds of Regadenoson. Furthermore, there is no RP-HPLC method reported in the literature that can adequately quantitate impurities present in finished pharmaceutical dosage forms. Synthesis of regadenosone from 2-chloro adenosine is depicted in Figure 1. Figure 1: Synthesis of regadenosone from 2-chloro adenosine EXPERIMENTAL SECTION Materials and Reagents Regadenosone as active pharmaceutical ingredient, 2-Chloro adenosine (Regadenoson impurity A), 2-Methoxy adenosine, 2, 6-Dichloropurine, 2-Amino adenosine and 2-Chloro adenosine were obtained from Gland Pharma as reference standards for process related substances. Instrumentation and Software Shimadzu 2010CHT liquid chromatograph separation module equipped with PDA detector (for specificity and degradation studies) with LC solutions software was used for the analysis. The chromatographic and the integrated data were recorded using HP-Vectra (Hewlett Packard, Waldbronn, Germany) computer system. Chromatographic separation of 2-chloro adenosine as process related impurity was achieved Symmetry C-8 (250 4.6 mm, 5 µ particle size). The ph measurements were carried out with Elico, model LI 120, ph meter equipped with a combined glass-calomel electrode calibrated using standard buffer solutions of ph 4.0, 7.0 and 9.2. Chemicals and Reagents All the reagents were of analytical-reagent grade unless stated otherwise. Glass-distilled and de-ionized water (Nanopure, Barnstead, USA), HPLC-grade acetonitrile, potassium dihydrogen ortho phosphate, and orthophosphoric acid (S.D.Fine chem, Mumbai, India) were used. Pure sample of 2-chloro adenosine (reference standard), and other process related substances viz. 2,6-dichloro adenosine, 2-methoxy adenosine and 2-amino adenosine were a kind of gift from Gland Pharma Pvt. Ltd, Hyderabad. Acetonitrile (HPLC grade) and potassium dihydrogen orthophosphate were obtained from Loba Chem, Mumbai, and Sd fine-chem ltd, Mumbai, respectively. All other chemicals used in the analysis were AR grade. Chromatographic Conditions Separation was achieved on a Symmetryc-C18-column (250 4.0 mm, 5 μm) using a gradient mode by the mobile phase consists of a mixture of Buffer and acetonitrile- methanol as equimolar mixture at 1:1 v/v with ph-3.2 adjusted with 10% o-phosphoric acid in water. The column was then re-equilibrated for 10 min with mobile phase. The flow rate of the mobile phase was 1.0 ml/min and the total elution time, including the column re-equilibration, was approximately 55 min. The UV detection wavelength was carried at 205 nm. The injection volume was 20 μl 56

and experiments were conducted at ambient temperature. Gradient program for acceptable resolution was given below (Table 1) Table 1: Gradient program for acceptable resolution Time (min) Solvent- A (%) Solvent B (%) 0.01 95 5 20 83 17 35 55 45 40 50 50 42 95 5 55 95 5 55.01 Controller Stop Preparation of System Suitability Solution (Standard Solution) Standards of 2-Chloro adenosine as process related impurity (20 mg each) was accurately weighed, transferred into 20 ml volumetric flasks, dissolved in diluents and made up to the mark with the mobile phase to get 1000 ppm of 2- Chloro adenosine primary standard solution. Transfer the 2.0 ml of the above stock solution and was adequately diluted up to 20 ml in volumetric flask with mobile phase to get the concentration of 80 microgram/ml as working standard solution. Preparation of Sample Solutions Samples of Regadenoson powder for injection containing 2-chloro adenosine as process related impurity were prepared as by dissolving equivalent amount of 20 mg is dissolved in 25 ml volumetric flask containing diluents solution. Transfer the 2.0 ml of the above stock solution and was adequately diluted up to 20 ml in volumetric flask with mobile phase to get the concentration of 80 microgram/ml as working sample solution. Method Validation Validation of the developed method for the determination of 2-Chloro adenosine present as process related impurity in Regadenosone sterile powder for injection was performed according to the ICH guidelines with standards, and related substance. Thus, system suitability along with method selectivity, specificity, linearity, range, precision, accuracy, limits of detection and quantification for the impurities, short term and long term stability of the analytes in the prepared solutions and robustness were demonstrated. System suitability: The system suitability was conducted using 1.5% (w/w) of the impurity spiked to 2-chloro adenosine (80.0 g/ml) and evaluated by making three replicate injections. The system was suitable for use if the tailing factors for 2-chloro adenosine and its impurity were 1.55 (observed value is 1.06) and the resolution was 1.90 (Observed value is 7.33). Specificity: Specificity of a method can be defined as the absence of any interference at retention times of peaks of interest, and is normally evaluated by observing the chromatograms of blank samples and samples spiked with the pure sample of process related impurity in the presence of other related substance and impurities in the mobile phase and the sterile powder for injection of Regadenoson. The known impurities were injected at specification limit and the retention times were tabulated along with retention time of 2-chloro adenosine. It was observed that there is no interference at retention time of 2-chloro adenosine peak. A separate solution of blank, standard and batch samples of 2-chloro adenosine were evaluated along with impurity solutions (Figure 2). 57

Figure 2: Estimation of specificity Linearity and range: The linearity of 2-chloro adenosine was studied by preparing standard solutions at five different levels ranging from 40 µg/ml to 120 µg/ml. The data were subjected to statistical analysis using a linear-regression model; the regression equations and coefficients (r 2 ) are given in calibration curves of 2-chloro adenosine and its impurity in Figure 3. Figure 3: Calibration graphs of 2-chloro adenosine as process related impurity Precision: Precision as measured by multiple injections of a homogeneous sample (prepared solution) indicates the performance of the HPLC instrument under the chromatographic conditions and day tested. The system precision was conducted using 80.0 micrograms/ml of the 2-chloro adenosine (i.e., 100% level) and evaluated by making six replicated injections. The method precision was conducted by using 10% (w/w) (i.e., 8 µg/ml) of the impurity spiked to 2-chloro adenosine (80 mg/ml) (i.e.,-100% level) and evaluated by making duplicate injections (Table 2). 58

Table 2: Precision studies of 2-chloroadenosise as impurity in regadenosine No. of Injections Peak Area Retention time (min) Inj-1 5256405 15.21 Inj-2 5255765 15.2 Inj-3 5255536 15.2 Inj-4 5254927 15.21 Inj-5 5253552 15.2 AVERAGE 5254945 15.20125 SD 993.7226307 0.0025 % RSD 0.02 0.02 Accuracy: To determine the accuracy of the test method samples were prepared by spiking 2-chloro adenosine raw material with the equivalent amount of placebo at 50%, 75%, 100%, 125% and 150% of the target concentration. Six samples were prepared at lower and higher concentration levels and remaining levels in triplicate. The average % recovery of 2-chloro adenosine was found to be within the limits. The results were summarized in Table 3 (Figure 4). Figure 4: Accuracy studies of 2-chloroadenosine as process related impurity in regadenosen Table 3: Accuracy studies of 2-chloroadenosine as process related impurity in regadenosen Concentration in (MG/ML) Dilution Level (%) Spike Area Stnd. Area Amount Added Amount Found Recovery 2880241 5255237 0.039525 0.041 102.1662343 2897168 5255237 0.039525 0.041 102.772536 0.04 50% 2881458 5255237 0.039525 0.041 102.2098256 2892466 5255237 0.039525 0.041 102.6041169 2882482 5255237 0.039525 0.041 102.2465038 2897995 5255237 0.039525 0.041 102.802158 4621005 5255237 0.06324 0.065 102.4487402 0.06 80% 4626713 5255237 0.06324 0.066 102.5765231 4624701 5255237 0.06324 0.065 102.5314812 5747431 5255237 0.07905 0.081 102.9325 5759667 5255237 0.07905 0.082 102.1516382 0.08 100% 5748976 5255237 0.07905 0.081 102.9601698 5757604 5255237 0.07905 0.082 102.1146913 5748371 5255237 0.07905 0.081 102.9493347 5754368 5255237 0.07905 0.081 102..0567368 6740123 5255237 0.09486 0.095 100.5924372 0.1 120% 6741562 5255237 0.09486 0.095 100.6139135 6742171 5255237 0.09486 0.095 100.6230024 8473809 5255237 0.118575 0.12 101.1733584 8472305 5255237 0.118575 0.12 101.1554013 0.12 150% 8474348 5255237 0.118575 0.12 101.1797938 8459743 5255237 0.118575 0.12 101.0054169 8462886 5255237 0.118575 0.12 101.0429428 8455534 5255237 0.118575 0.12 100.9551634 59

LOD and LOQ: The LOD and LOQ were determined for 2-chloro adenosine and its process related impurity based on the standard deviation of (SD) of the response and slope (S) of the regression line as per ICH guidelines according to the formulae give below [11,12]. RESULTS AND DISCUSSION Optimization of Chromatographic Conditions To optimize the chromatographic conditions, different combinations of methanol water (90:10, 80:20, 70:30, 60:40, 50:50, 40:60, 30:70), acetonitrile methanol (90:10, 80:20, 70:30, 60:40, 50:50, 40:60, 30:70), and acetonitrile potassium dihydrogen phosphate buffer (90:10, 80:20, 70:30, 60:40, 50:50, 40:60, 30:70) were tested. Acetonitrile: methanol (1:1 v/v )with phosphate buffer (ph 3.5) was preferred because it resulted in a greater response to 2-chloro adenosine and also to its related impurity after several preliminary investigatory runs compared with the other mobile phases. The composition, ph, and flow rate of the mobile phase were changed in gradient mode to optimize the separation conditions. Increasing the organic modifier content resulted in a decrease in the retention time of the analyte. The effect of the flow rate was studied in the range 0.8 to 1.2 ml/min and higher acetonitrile content high flow rate resulted, as expected, in rapid analysis before dead volume. A high buffer concentration was therefore used at a flow rate of 1.0 ml/ min, keeping in mind the possibility that potential minor degradation products could appear after stress studies and might co-elute with the drug because of the reduced RT if the flow rate was increased. High flow rates also reduce the life time of both column and pump. Changing the buffer ph from 3.5 to 5.5 (by addition of potassium dehydrogenate phosphate (0.01 M) or ortho-phosphoric acid) affected the RT and shape of the 2-chloro adenosine impurity peak. The effect of ph on analyte elution was related to the degree of ionization. Reducing the ph resulted in a shorter RT of process-related impurity because of ionization of its basic site. A ph of 3.2 was regarded as optimum because at this ph the analyte peak was sharp and well-defined. The preferred brand of HPLC column should be selected primarily based on the long term stability and lot-to-lot reproducibility. Preliminary development trials have performed with various octadecyl columns (C18 columns) of different types and dimensions from different manufacturers were tested for the peak shape and the number of theoretical plates of 2-chloro adenosine raw material at 500 µg/ml concentration. Finally by switching to Symmetry C18-3V column (250 mm 4.6 mm, 5 μ) column there was a substantial increase in the theoretical plates (~50000) with a significant improvement in the peak shapes with 1.14 tailing factor. It also produced adequate resolution between 2-chloro adenosine and its two process related impurity. As a result, it was selected as an optimum one and used throughout this investigation. Linearity The calibration curve showed good linearity in the range of 40-120 μg/ml, for 2-chloro adenosine API and its related substance with correlation coefficient R 2 =of 0.9998 and 1.0000 respectively indicating strong correlation (Figure 3). Precision The results of system precision (% RSD=0.02 for peak area 0.47 for retention time of 2-chloro adenosine as process related impurity respectively), method precision (% RSD=1.4 and 1.0 for 2-chloro adenosine and its process related impurity respectively) are found within the prescribed limit of ICH guidelines (% RSD <1%, and % RSD <2% respectively in case of system precision and method precision. Accuracy The accuracy of the method was determined by measurement of recovery. The results of accuracy studies from standard solution and process related impurity were shown in Table 3; recovery values demonstrated that the method was accurate within the desired range. The results revealed an excellent correlation between the amount added and the amount found. 60

CONCLUSION The HPLC method was developed using Symmetry C8 (250 mm 4.6 mm ID 5 µm) column with mobile phase and Transfer 2.0 ml of 85% ortho phosphoric acid into 1000 ml of water. Adjust the ph of the solution to 4.0 ± 0.05 with Triethylamine. Flow rate was 1.0 ml/min, with injection volume 50 µl detection done by using UV detector at 250 nm. The run time was 35 minutes which enables rapid quantitation of many samples in routine and quality control analysis of tablet formulations. The proposed method was found to be simple, sensitive, rapid and economical for the determination of Regadenosine parenterals. The developed method was checked for the performance characteristics and has also been validated. The method was validated by using various validation parameters like linearity, accuracy, precision, specificity, solution stability and robustness. The method was found to be linear (R 2 >0.999) precise (%RSD: 0.02) and accurate (mean % recovery fields 99.1% - 103.00%). The proposed HPLC method was simple, precise because of commonly used buffer and shorter run time. The mean percentage recovery above 95% indicates the reproducibility and accuracy of new developed method compared. The result of study include the proposed method is highly accurate, simple, precise and specific. The simple recoveries in all formulations were in good agreement with their respective label claims. After validating proposed method as per ICH guidelines and correlating obtained values with the standard values, satisfactory results were obtained. From the specificity studies it was concluded that the developed method was specific for 2-chloro adenosine. ACKNOWLEDGEMENTS The authors are grateful to M/s GLAND PHARMA, Hyderabad for the supply of 2-chloro adenosine and its process related impurity A as a gift sample. The authors also thankful to Sultan-Ul-Uloom college of Pharmacy, Hyderabad for providing the necessary facilities to carry out the research work. REFERENCES [1] GJ Zoghbi, AE Iskandrian. Pharmacological stress testing. In: Nuclear Cardiac Imaging: Principles and Applications. Garcia EV, Iskandrian AE (Eds). Oxford University Press, NY, USA, 2008, 293-315. [2] AS Gemignani; BG Abbott. J Nucl Cardiol. 2010, 17(3), 494-497. [3] ME Olah; GL Stiles. Annu Rev Pharmacol Toxicol. 1995, 35, 581-606. [4] RC Hendel; T Jamil; DK Glover. J Nucl Cardiol. 2003, 10(2), 197-204. [5] Z Gao; Z Li; SP Baker; RD Lasley; S Meyer; E Elzein; V Palle; JA Zablocki; B Blackburn; L Belardinelli. J Pharmacol Exp Ther. 2001, 298(1), 209-218. [6] ICH. Q1A (R2) Stability Testing of New Drug Substances and Products. Geneva, February. 2003. [7] ICH. Q1B Stability Testing: Photo stability Testing of New Drug Substances and Products Geneva, November, 1996. [8] ICH. Q6A: Specifications: Test Procedures and Acceptance Criteria for New Drug Substances and New Drug Products: Chemical Substances Geneva, October, 1999. [9] ICH. Q3A (R2) Impurities in New Drug Substances Geneva, October, 2006. [10] ICH. Q3B (R2) Impurities in New Drug Products Geneva, June, 2006. [11] ICH. Validation of Analytical Procedures: Text and methodology Q2 (R1): International Conference on Harmonization, IFPMA, Geneva, 2005. [12] United States Pharmacopoeia (USP), XXVI. Validation of compendial methods. United States Pharmacopoeial Convention Inc.; Rockville, MD, USA, 2003. 61