MRI Conditional Devices and Patient Safety. Outline. MRI Labeling: Objects 8/3/13 1. David W. Jordan, Ph.D. Senior Medical Physicist

Similar documents
Magnetic Resonance Imaging (MRI) Safety for Boston Scientific Products. 1. WALLSTENT Iliac Endoprosthesis with Unistep Plus Delivery System

MRI Information. MRI Safe The following AMS product does not contain metallic components and therefore is considered MRI Safe.

MR Safety Evaluation of Orthopedic Devices: Why It's Important and How to Select Test Device Configuration

MRI Safety. How can we ever forget?? July Missile Effect. Have we lost sight of our MR Safety responsibility?

Chapter 2. Magnetic Resonance Safety. Andrew Simmons and Kristina Hakansson. Abstract. 1. Legislation, Guidance and Best Practice

MRI Safety aspects. for the anaesthesiologist

Magnetic Resonance Imaging at 7T in Glasgow. A unique opportunity!

12 pericardial bioprostheses. 7300TFX Carpentier-Edwards bioprosthetic valved conduit

Replacement Heart Valve Product Description (Stented Tissue) Models Reference

Quality Control Concerns for New Generations of CT, MRI, & PET

Guidance Document Magnetic Resonance Imaging Effective Review Delegated & Full Board Version Date February 1, A. Preamble

Special thanks to the University of Utah for allowing us to adapt their materials. MRI Safety and Training

MRI Systems Market to 2018

Replacement Heart Valve Product Description (Stented Tissue) Models Reference

Current Issues in MRI Safety

Physicists Quality Control for MR Equipment

Siemens Ct Scanner Manual

Ingenia MR-RT. MR Systems. The comprehensive MR-sim solution to fit your planning

DuoFLEX. MRI Coil Suite. Flexible coils that contour to your patient and your budget.

MRI Safety & Site Planning. Lawrance Yip Senior Radiographer, DR, QMH 16 June 2007

RePneu (Lung Volume Reduction) Coil System Instructions For Use

Joint Commission Medical Imaging Requirements. Laura Gruber, MBA, RT(R), RDMS, RVT Sr. Director Department of Medical Imaging SPR Annual Meeting

Healthcare. Philips DoseWise Portal. Radiation exposure management software

Implants for surgery Active implantable medical devices. Part 3: Implantable neurostimulators

Future bound. Philips Ingenuity Core

Implementing Optimal Kidney Stone CT Protocols Using Outreach from a National Database. Purpose

MRI safety in practice: The EU directive on work in electromagnetic fields practical and clinical aspects

DELIVERING FAST EFFICIENT SPONGE DETECTION SITUATE DELIVERY SYSTEM

New Hope for Thalassemia Patients

Disclaimer and Notices.

Testing Methods and Standards for Magnetic Resonance (MR) Safety and Compatibility of Medical Devices - Update

New PET/CT from Siemens helps more patients benefit from premium technologies

ClinScan. think forward. Small Animal MRI Solution for Molecular Imaging and Translational Research. Preclinical MRI

SELF-REGULATORY INITIATIVE FOR MEDICAL IMAGING EQUIPMENT STATUS REPORT 2010

Approval of subjects for measurements at ultra-high-field MRI

Reduction of total energy consumption in hospitals

Stay Ahead of the Curve in Radiation Reduction Standards

Last Name First Name Middle Initial Previous Last Name(s)

Vantage Titan 1.5T MR. The Intelligent Choice.

AAPM IGRT Recommendations

Declaration of Relevant Financial Interests or Relationships

Follow this and additional works at: Part of the Physical Sciences and Mathematics Commons

2017 ACR Computed Tomography Quality Control Manual FAQS

Uncompromised accuracy in neuroimaging and treatment

MRI in Clinical Practice

UNC s Biomedical Imaging Research Center (BRIC) Roger Sit, Ph.D., CHP NCHPS Spring 2013 Meeting February 28, 2013

How GE Research Developed MRI (Magnetic Resonance Imaging) Guided Surgery By Kirby Vosburgh

10 Best Practices for Electrosurgical Unit Testing: Tips to quickly & effectively test

DoseWise Portal. Take control of dose management. across your organization

Take control of dose management

Fluoroscopy Testing Workshop

WHITE PAPER. Everything You Need To Know About the FDA UDI Rule

Eric W. Abelquist, Ph.D., CHP President, Health Physics Society. August 2017

MEDICAL PHYSICS (MED PHYS)

Press Presse Press Presse

siemens.com/avanto MAGNETOM Avanto The landmark in 1.5T imaging.

Niccolo Machiavelli (1523)

HHS Public Access Author manuscript Abdom Imaging. Author manuscript; available in PMC 2016 February 01.

Recharge your MR experience.

Simple, intuitive and accessible MRI solution for preclinical research. M-Series Compact MRI Systems

Virtual Mold Room. The Future of Custom Radiation Therapy Devices

8/2/2017. Key attributes of scientific excellence: rigor, innovation, and relevance. Medical Physics 3.0. Key Attributes of Scientific Excellence

A Guide to the Safety, Health and Welfare at Work (Electromagne c Fields) Regula ons (S.I. No. 337 of 2016)

7T MRI research in neurology: initial results and future research applications

FDA initiatives to facilitate evaluation of novel percutaneous valve technologies within the US

IMPLEMENTATION REALITY SESSION Unique Device Identification (UDI)

IMV 2006 CT Market Summary Report Featuring results from 2,565 hospital and non-hospital CT imaging sites in the United States

Progress in X-Ray & MR

1. Buying Imaging Equipment for Interoperability and Compliance: Beyond DICOM 2. Informatics for the Acceptance Testing of Imaging Equipment

The Role and Activities of ICNIRP and ICNIRP Exposure Limits and Guidelines

CODING GUIDELINES. Radiation Therapy. Effective January 1, 2018

Power-class. Expect more performance. MAGNETOM Symphony.

There are risks and costs to a program of action. But they are far less than the long-range risks and costs of comfortable inaction

Guidance for the Physics Aspects of Clinical Trials

QIBA Profile: Diffusion-Weighted Magnetic Resonance Imaging (DWI)

Quality ID #262: Image Confirmation of Successful Excision of Image-Localized Breast Lesion National Quality Strategy Domain: Patient Safety

Magnetic Resonance Imaging 23 (2005)

REFERENCE CODE GDME1123CFR PUBLICAT ION DATE DECEMBER 2014 MAGNETIC RESONANCE IMAGING SYSTEMS APAC ANALYSIS AND MARKET FORECASTS

1. Executive Summary

Journal of Clinical Urology (JCU): Declaration Guidelines for Authors

Understanding Clinical Equivalence

Leading Canadian Pediatric Hospital Continues to Drive Dose Down for Sick Children

International Standards and EU regulation of medical device software an update

Magnetic Resonance Safety Update 2002: Implants and Devices

Interventional CT: More Procedures, Reduced Dose, Improved Speed at Minimal Cost

Handzettel 1. CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014

TITLE: Research Pharmacy Standard Policy SOP #: INV-100 Page: 1 of 5 Effective Date: 8/31/17

Status of MedAustron

Safety Management for MR-Guided Interventions

Quality Assurance for particle beam therapy

Laboratorio di Tecnologie Biomediche

Pulse of MR. Spring ISMRM Academic Issue. Imagine what

UDI 101: What You Need to Know About Unique Device Identifiers

REQUIREMENTS FOR COMMERCIAL PARTICLE ACCELERATORS IN RADIATION THERAPY

Oktober 4, 2011 Els C.M. van der Wilden-van Lier, MD, MPH Board of Governors / CMO ZGT. GS1 in Dutch Hospitals

Transcription:

MRI Conditional Devices and Patient Safety David W. Jordan, Ph.D. Senior Medical Physicist 8/3/13 1 Outline MRI Conditional Overview Spatial Gradient Hazards RF SAR Hazards Device Registries MRI Labeling: Objects MR Safe: Completely nonmetallic MR Conditional: can be safe in MR environment under certain known conditions MR Unsafe: demonstrated attractive forces in magnetic field ACR Guidance Document on MR Safe Practices: 2013 1

???? www.magmedix.com www.alimed.com Conditions of Concern Static Field Strength (B 0 ) Static Spatial Gradient SAR (RF heating) Gradient Strength (time varying) Gradient Speed or Slew Rate Not all implants/devices will specify limits for all of these! A device is MR conditional if: Potential harm to patient; (or) Potential harm to staff or other personnel; (or) Potential damage to device; (or )Potential damage to MR imaging system; (or) Potential negative impact on MR imaging For ANY condition(s) of MR environment 2

GRADIENT CONDITIONS Spatial Gradients Field Gradient Contour Plots Siemens Healthcare 3

Field Gradient Contour Plots Siemens Healthcare Field Gradient Contour Plots Philips Healthcare Field Gradient Contour Plots Philips Healthcare 4

Product Labeling Example Non-clinical testing has demonstrated that the XXX is MR Conditional. A patient with this XXX can be scanned safely immediately after placement under the following conditions: Static Magnetic Field: Static magnetic field of 3.0 Tesla or less. Highest spatial magnetic gradient field of 720 gauss/cm Are the conditions met in this scanner? Siemens Healthcare RF CONDITIONS 5

RF SAR Specific Absorption Rate W/kg (think of as a thermal dose rate ) MRI RF amplifier outputs rated in kw Modern designs up to 30 kw RF Heating Dependence on: MRI output power Patient mass Within transmit coil volume Frequency Size and shape of implant Conductivity Heat sink / physiological cooling Product Labeling Example Non-clinical testing has demonstrated that the XXX is MR Conditional. A patient with this XXX can be scanned safely immediately after placement under the following conditions: MRI-Related Heating: Maximum whole-body-averaged specific absorption rate (SAR) of 2 W/kg for 15 minutes of scanning (i.e., per scanning sequence) 6

Estimating SAR Scanner may show % (of what?) IEC / FDA Limits for Heating: Normal Mode: 0.5 o C temperature rise, or 2 W/kg whole body First Level Controlled Mode: 1.0 o C temp. rise, or 4 W/kg whole body Second Level Controlled Mode: >1.0 o C or > 4 W/kg (IRB approval only) Localized SAR Limits Head normal mode: 3.2 W/kg averaged over head mass Torso normal mode: 10 W/kg over any 10 grams of tissue Extremities normal mode: 10 W/kg over any 10 grams of tissue (no First Level limits for localized modes) DEVICE REGISTRIES 7

Tool for assessing device labeling vs. magnet, scan conditions Specific conditions must be met Acceptance varies from magnet to magnet, protocol to protocol Device Registries Device Registries Enhanced device information e.g. The List at www.mrisafety.com Must still be assessed for each scanner, sequence/protocol Do not rely on what others have done Device Registries Cheat sheet for each device for: YOUR staff YOUR magnets YOUR scan protocols Share experiences within institution/ system as new devices emerge 8

Presence and Identity of Devices Establish registry with surgeons Make, model, serial number, date of implant MRI safety notes where appropriate Summary Assess device manufacturer s detailed MRI Conditional requirements Spatial gradient and RF SAR conditions are most challenging for technologists Most likely to be denied unnecessarily Local/institutional device registry can save time and duplication of effort, avoid errors. Thanks! / Questions? Dave Jordan david.jordan@uhhospitals.org (216) 286-6911 8/3/13 27 9