Analytical Instrument Qualification According to USP <1058>: Requirements and Examples for the Agilent 1290 Infinity LC System

Similar documents
New Advances in UHPLC -Resolution, Speed & Sensitivity

Qualification of High-Performance Liquid Chromatography Systems

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Analysis of amoxicillin and five impurities on the Agilent 1220 Infinity LC System

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

Controlling Agilent 1200 Series Rapid Resolution LC systems through Waters Empower chromatography data software. Technical Overview.

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Analysis of TCM injections using the Agilent 1290 Infinity LC system

Application Note. Author. Abstract. Small Molecule Pharmaceuticals. Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany

easy. HPLC has never been so Agilent 1120 Compact LC Our measure is your success.

ACQUITY UPLC. I-Class PLUS WHAT SEPARATES YOU FROM EVERYONE ELSE

Application Note. Author. Abstract. Small Molecule Pharmaceuticals & Generics. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

C196-E067A. Prominence UFLCXR. Shimadzu Ultra Fast Liquid Chromatograph

Method Translation in Liquid Chromatography

Application Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Peptide Mapping: A Quality by Design (QbD) Approach

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Agilent 1220 Infinity II LC GET ON THE FAST TRACK TO HIGHEST EFFICIENCY AT A PRICE YOU CAN AFFORD

Technical Note. Introduction

Validation Strategies for Equipment from Multiple Vendors

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

HPLC to UPLC Method Migration: An Overview of Key Considerations and Available Tools

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

BioHPLC columns. Tim Rice Biocolumn Technical Specialist

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

IC-D3160 Ion Chromatograph

Minimize Method Development Time for Pesticide Screening Using LC/MS Application Kit

Agilent Infinity LC Instrument Component Software Version 1.0 for Empower Software

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Preparative Purification of Corticosteroids by HPLC; Scalability and Loadability Using Agilent Prep C18 HPLC Columns Application

APPLICATIONS TN Overview of Kinetex 2.6 µm Core-Shell Technology

SFC Säulen für analytische und preparative Anwendungen

Heart-cut 2D-LC/MS approach for pharmaceutical impurity identification using an Agilent 6540 Q-TOF LC/MS System

Agilent 1260 Infinity II LC ASPIRE TO NEW LEVELS OF EFFICIENCY EVERY DAY IN EVERY WAY

Update to USP <621> Chromatography: What does this mean? USP 37 NF 32 1S Current as of August 2014

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS

Analysis of Monoclonal Antibody Digests with the Agilent 1290 Infinity 2D-LC Solution

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

Gas Chromatography Assignment Chem 543/443

High Resolution Analysis of Heparin and Heparin-like Impurities on Glycomix TM an Anion Exchange Column

Agilent 6430 Triple Quadrupole LC/MS System

Overview of preparative HPLC. Analytical Technologies Limited

BÜCHI Labortechnik AG

Method Development Considerations for Reversed-Phase Protein Separations

Protein-Pak Hi Res HIC Column and HIC Protein Standard

Part II: Column Temperature and Refractive Index Detection

Forced Degradation of Ibuprofen in Bulk Drug and Tablets

Agilent 7820A GC System THE EASY CHOICE FOR ROUTINE ANALYSES

On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts

Preparative HPLC System

Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns

Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants

Genotoxicity is the property of a compound

Combining High Temperature and Small Particles: The Advantages of Zirconia

Improved SPE for UPLC/MS Determination of Diquat and Paraquat in Environmental

Quantitative determination of residual 2-(2-chloroethoxy) ethanol (CEE) in quetiapine fumarate by gas chromatogaraphy

Agilent 6500 Series Q-TOF LC/MS System

AZURA Bio purification

Enhancing your HPLC Separations using Temperature Programming

2017 by Agilent Technologies Agilent CrossLab Compliance Services

2017 by Agilent Technologies Agilent CrossLab Compliance Services

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime

Analysis of Monoclonal Antibody Digests with the Agilent 1290 Infinity 2D-LC Solution Part 2: HILIC RPLC-MS

Prominence Preparative HPLC System

mabs and ADCs analysis by RP

Fast Preparative Column Liquid Chromatography (PCLC)

Certified Reference Material - Certificate of Analysis

Puzzled About LCMS? Sample. Sensitivity in Mass Spec Analysis. Prep. Adapting LC-UV. Optimizing and Maintaining Your Mass Spec LCMS

Agilent 1260 Infinity High Performance Autosampler

Elute LC series. Innovation with Integrity. UHPLC Precision for MS applications LC-MS

AZURA Bio LC. 2 FPLC and Purification Solutions

Application Note. Abstract. Authors. Introduction. Environmental

Fundamentals and Techniques of Preparative HPLC. Parto Zist Behboud Tel: 42108

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph

The Theory of HPLC Gradient HPLC

Identification, expression and characterization of a R- -transaminase from Capronia semiimmersa.

Preparative HPLC: Factors and Parameters that Directly Affect Recovery of Collected Fractions

Development of Analysis Methods for Therapeutic Monoclonal Antibodies Using Innovative Superficially Porous Particle Biocolumns

Everyone can be an Expert. Electrochemical Detector. Autosampler. Complete HPLC-ECD

ISSN India; g,secunderabad. Abstractt. a flow rate. of 1ml/min. di hydrogen. which acts. and chronic. including minimize (5) Figure

Antibody Analysis by ESI-TOF LC/MS

MAbPac SCX-10 Columns

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

A green HPLC technique with a 100% water mobile phase for detecting imidacloprid and its metabolite 6-chloronicotinic acid

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates

Transcription:

Analytical Instrument Qualification According to USP <1058>: Requirements and Examples for the Agilent 1290 Infinity LC System 1) USP <1058>: Scope, approach, requirements 2) Qualification Examples for LC 1290 3) Selected performance tests Dr. Edgar Naegele Application Scientist Agilent Technologies, Germany 1

Design Qualification Selected HPLC specifications for design qualification. Selected specifications of the Agilent 1290 infinity it autosampler 2

Design Qualification Selected criteria i for vendor assessment. All requirements of this custom vendor assessment template are fulfilled by Agilent Technologies! X X X X X X X X 3

Installation Qualification 1) Template and examples to document completeness of shipment for IQ. 3) Equipment documentation for IQ. yes yes 2) Verification of construction requirements for IQ. 4

Installation Qualification Testing for Installation Qualification Verification of correct system installation for IQ. For this purpose p Agilent uses a certified standard! 5

Operation qualification - an example procedure HPLC - System Pump Injector Oven Detector Computer Control& Evaluation Flow rate Inj. Vol. Temperature Signal precision accuracy accuracy and/or noise Specifications of the Agilent 1290 infinity LC system Flow rate precision Inj. Vol. accuracy Temperature accuracy Signal and/or noise 0.07% 07% RSD or <025RSD 0.25 +/- 08 C 0.8 <+/ +/- 3x10-6 AU 0.005 SD at Precision 0.2 5.0 ml/min +/- 0.05 C from 5 20 µl < 0.50 RSD from 2 5 µl < 0.70 RSD from 1 2 µl at 230/4 nm, slit width 4nm, TC 2 s ASTM 6

Operation qualification - an example procedure Computer Control & Evaluation Generated Automatically Header Signed and dated Summary Test/Limit/Result/Status Configuration Module/FW/Serial No. & Software Verification Log Book Test Reports Sequence/Log Method Chromatograms/Spectra/Reports Results/Sample Summary 7

Gradient test depending on the volume of the mixers (blue-35µl, µ, red-135µl, µ,g green-300µl) µ ) mau Gradient (stepwise): Time % A % B 600 000 0.00 100.00 000 0.00 3.00 100.00 0.00 3.01 90.00 10.00 500 6.00 90.00 10.00 6.01 52.00 48.00 400 9.00 52.00 48.00 9.01 48.00 52.00 12.00 48.00 52.00 300 12.01 10.00 90.00 15.00 10.00 90.00 15.01 0.00 100.00 200 18.00 0.00 100.00 18.01 80 48.00 800 52.00 100 21.00 48.00 52.00 21.01 52.00 48.00 Influence of the mixer 24.00 52.00 48.00 0 24.01 100.00 0.00 27.00 100.00 0.00 0 5 10 15 20 25 Gradient Testing: Mobile Phase A: Water Mobile Phase B: Water + 0,5% Aceton (v/v) min

1290 Infinity Binary Pump Degasser High resolution pump di drives Silicon Carbide piston Channel A Solvent selection valve Purge valve Jet Weaver Mixer Multi layer heat exchanger Channel B Next Generation Technology 1200 bar @ 2.0 ml/min 800 bar @ 5.0 ml/min Minimum dwell volume: 10 µl / 45 µl Minimum Mixing Noise Maximum RT repeatability Automatic Purge Automatic correction of von micro leaks 9

1290 Infinity binary Pump Minimum dead volume and mixing noise 1200 Series Solvent A 1200 Series Solvent B Passive Damper 0-300 µl Passive Mixer 400 µl Low Noise ~ 800 µl dead volume 1290 Inifinity Solvent A 1290 Inifinity Solvent B Active Damping: 10 µl Jet Weaver Mixer: 35 µl Firmware Hardware Agilent microfluidic, Multilayer Technologie + Pump Pump + = 10 µl / 45 µl DV: 80x 20x / smaller 20x kleiner Niedriges Lower Noise Noise < 50% 10

Example: Performance Qualification Documentation ti of PQ tests. t ChemStation Report with detailed statistical information Amount S.D. : 0.035 RSD: 0.13 11

Performance qualification an example 300 0.889 0.818 400 0.685 0.617 500 0.713 0.761 0.451 600 0.551 mau AU 200 100 0 0 0.2 0.4 0.6 0.8 1 RSD % RT (n=10) RSD % Area (n=10) 1 0.01600 0.401 2 0 00769 0.00769 0 393 0.393 3 0.00517 0.429 4 0.00575 0.470 5 0 00577 0.00577 0 398 0.398 6 0.00538 0.463 7 0.00585 0.437 8 0.00571 0.518 9 0.00797 0.555 Mean 0.00725 0.450 min Sample: Check Check-out out sample with 9 phenones 12 Peak Column: 2.1 x 50 mm, 1.8 µm Gradient: 5 to 95 % B in 0.5 min Flow rate: 1.2 ml/min Injection volume: 1µL

Analytical Instrument Qualification According to USP <1058>: Requirements and Examples for the Agilent 1290 Infinity LC System 1) USP <1058>: Scope, approach, requirements 2) Qualification Examples for LC 1290 3) Selected performance tests and performance data 13

Application depended Performance Data and System Suitability - Experimental conditions for the Agilent 1290 infinity LC system Configuration of the 1290-System G4220A 1290 Infinity Binary pump with integrated vacuum degasser and different solvent mixers G4226A 1290 Infinity Autosampler G1316C 1290 Infinity Column Compartment G4212A 1290 Infinity Diode Array Detector Software: Chemstation B.04.02 Reference samples A liquid formulation of metoclopramide hydrochloride (1 mg/ml) comprising 0.01% concentration of impurities was used for all tests. Setup for testing 1. Determination of pump performance depending on dwell volumes by gradient tests 2. Establishment of a chromatographic separation to achieve data for long time evaluations 3. Similar peak pattern and resolution according to selected column with respect to particle size and column dimension and adapted to gradient shape and flow 4. Precision of areas must be < 2 % RSD. 5. Precision of retention times must be < 0.5 % RSD. 6. Linearity should be given at least with R2>0.999 Sample setup for testing Sample Purpose Injections Blank solution Verify baseline stability and identify artifacts 3 Suitability sample Verify precision of areas and 10 retention times for reference solution Calibration Verify linearity 3 for each level Highest concentration ti Verify Carry over 3 of each sample and Blank solution 14 Confidentiality Label

Separation of Metoclopramide and its impurities on different columns. A B C D A) Agilent TC C18, B) Eclipse Plus C18, C) Eclipse Plus C18, D) Waters BEH C18, 250*4.6mm, 5µm 150*2.1mm, 3.5µm 150*2,1mm, 1.8µm 150*2.1mm, 1.7µm Flow rate 1.058ml/min 1.058ml/min 0.221ml/min 0.221ml/min Gradient 0-25min 5-57% B 0-15min 5-57% B 0-15min 5-57% B 0-15min 5-57% B Temperature 37 C 37 C 37 C 37 C Injection 8µl 2µl 1µl 1µl volume Detection DAD, Signal 275/4, Reference 400/60, standard cell Data rate 2 Hz 2 Hz 40 Hz 40 Hz Maximum 138 bar 145 bar 310 bar 435 bar pressure Mobile phase A : 0.25% w/w Ammonium acetate in water Mobile phase B : Acetonitrile

Determination of the precision of areas and retention times Impurities in Metoclopramide p Determination of the precision of areas and retention times Retention times Areas Linearity Mean RSD Mean RSD R 2 Impurity 1 6.927 0.030 3.765 1.36 0.9994 Impurity 2 7.386 0.025 2.170 1.49 0.9997 Metoclopramid 8.170 0.039 10190.182 0.58 0.9998 Impurity 3 8.714 0.035 1.707 1.48 0.9993 Impurity 4 9.395 0.013 1.496 1.58 0.9993 Impurity 5 10.018 0.039 1.023 1.55 0.9990 Impurity 6 10.551 0.039 1.971 1.46 0.9993 16

Agilent 1290 Infinity Autosampler Flow-through design Advantage of the flow-through design No sample loss Flexible injection volume with high precision from 0.1 to 20 µl No need to exchange the sample loop Compromise of the flow-through design higher dead volume of 85 µl For the 1290 Agilent infinity LC system 125 µl dead volume if loop is in main path 17

Detection of carry over. A) Injection of sample with highest concentration of metoclopramide. B) Injection of a blank solvent sample. C) Carry-over specification. A Metoclopramide mau 1.5 1.25 1 DAD1 A, Sig=257,4 Ref=360,100 (F:\CHLORHEX100\WATER1200000014.D) DAD1 A, Sig=257,4 Ref=360,100 (F:\CHLORHEX100\CHLOREX30020013.D) Chlorhexidine 0.75 B 0.5 0.25 0-0.25-0.5 0.0028% carry over 0.25 0.5 0.75 1 1.25 1.5 1.75 2 2.25 min Column: 2.1x50mm Eclipse Plus C18, 1.8µm Sample: Chlorhexidine (1200ng/µl) dissolved in Water+0.1%TFA Mobile Phases: water+0.05%tfa, Acetonitrile+0.045%TFA Isocratic: 64 % A + 33 % B. Flow rate: 0,5ml/min, Stop time: 2.5min Injection volume: 1µl, external needle wash, water + 0.1%TFA, 10sec Column temperature: 50 C DAD: 257/4nm, Ref 360/100nm, 20Hz

Speed optimized separation of Metoclopramide and its impurities Zorbax Eclipse Plus C18 150*2 1mm 18µm Zorbax Eclipse Plus C18, 150 2.1mm, 1.8µm Flow rate : 0.44ml/min Gradient : 10 minutes

1290 Infinity Diode Array Detector Optofluidic Waveguides Non-coated fiber (fused silica) High light transmission by Total-Internal Reflection (TIR) Principle Advantage High Sensitivity at Lowest Cell Volume No RI (Refractive Index) and Thermal Effects (Solvent Temperature) Data Rate up to160 Hz (Multi Wave Length + Spectra) One Flow Cell for all Applications Cartridge Design for easy Handling 20

Conclusion The new Agilent Infinity LC is designed to provide highest speed, resolution and sensitivity. A new power range allows you to operate with any particle type, any column dimensions, or any mobile and stationary phase. The Agilent 1290 Infinity HPLC system is can be qualified according to the USP <1058> chapter and will meet highest requirements for every LC application. A selection of tests that can be established to evaluate a LC system were shown: For instance gradient testing, the determination of precision of areas and retention and the determination of flinearity. it All results show explicitly the applicability of the 1290 Infinity LC system for quality control testing as well as in the FDA regulated environment. Agilent Services qualifies your system according to USP1058! 21

References/Further reading United States Pharmacopiea, Chapter <1058>, Analytical Instrument Qualification, USA, 2008 Analytical Instrument Qualification and System Validation, January 1, 2009, Agilent Publication Number 5990-3288EN Increasing productivity in the analysis of impurities iti in metoclopramide hydrochloride hl formulations using the Agilent 1290 Infinity LC system, Agilent Technologies Pub.-Number: 5990-3981EN Analytical instrument qualification and system validation according to USP Chapter <1058> for the Agilent 1290 Infinity LC system, Agilent Technologies Pub.-Number: 5990-4781EN 22

Thank you for your attention! Do you have Questions? 23