Maximum Yield Mini System. Protocol Book YPD100 // YPD300. Ver

Similar documents
Fast Ion. Plasmid Midi Advanced Kit (incl. Thimble) Plasmid Midi Advanced Kit. Protocol Book. Ion Exchange-High Yield, High Purity, Transfection-grade

FastGene Plasmid Mini Kit For isolation of high copy and low copy plasmid DNA

HiYield TM Genomic DNA Extraction Kit

Presto Mini gdna Bacteria Kit

(Spin Column) For purification of DNA and RNA from formalin-fixed, paraffin-embedded tissue sections. Instruction for Use

E.Z.N.A. mirna Kit. R preps R preps R preps

Plasmid Midiprep Plus Purification Kit. Cat. # : DP01MD-P10/ DP01MD-P50 Size : 10/50 Reactions Store at RT For research use only

For simultaneous purification of genomic DNA and total RNA from the same animal cells or tissues

StrataPrep Plasmid Miniprep Kit

Genomic DNA Extraction kit Protocol Book

Specifications. Kit Specifications. Alcohol Precipitation: Up to 100 ml Column Purification: Up to 5 ml Column Binding Capacity 25 µg

Cells and Tissue DNA Isolation Kit (Magnetic Bead System) 50 Preps Product # 59100

Spin Micro DNA Extraction Kit

Blood DNA Extraction Kit

TIANpure Mini Plasmid Kit II

SPEEDTOOLS DNA EXTRACTION KIT

Presto Soil DNA Extraction Kit

Average Yields* Yeast DNA Yeast RNA Time to Complete 10 Purifications * Yield will vary depending on the type of sample processed

Urine DNA Isolation Maxi Kit (Slurry Format) Product # 50100

Cells and Tissue DNA Isolation 96-Well Kit (Magnetic Bead System) Product # 62500

E.Z.N.A. Bacterial RNA Kit. R preps R preps

Fast, easy, reliable, ultra-pure transfection grade plasmid DNA Maxiprep using a microcentrifuge spin-column.

Plasmid DNA Purification

The fastest, simplest method for purifying the highest quality endonuclease-free plasmid DNA.

Kit Components Product # (50 samples) Wash Solution A Elution Buffer B

foodproof Sample Preparation Kit III

Life science unlimited

Geneaid DNA Isolation Kit (Yeast)

Presto Mini RNA Bacteria Kit

User's Manual. MagCore HF16 Nucleic Acid Extraction Kit CONTENTS. Kit Contents, Description, Applications, pretreatment, Protocol

EZ-10 SPIN COLUMN GENOMIC DNA MINIPREPS KIT HANDBOOK

GeneJET Gel Extraction Kit #K0691, #K0692. }!tc!!!#!!h~ PA-K069

FOR RESEARCH USE ONLY. NOT FOR HUMAN OR DIAGNOSTIC USE.

UltraClean Midi Plasmid Prep Kit

Genomic DNA Extraction Mini Kit (Blood and Urine)

ISOLATE II Blood DNA Kit. Product Manual

Manual innuprep DNA Micro Kit

E.Z.N.A. PF Micro RNA Kit. R preps R preps

E.Z.N.A. Stool DNA Kit. D preps D preps D preps

Product # 24700, 24750

Presto Stool DNA Extraction Kit

E.Z.N.A. Tissue RNA Kit. R preps R preps

ULTRAPrep PLASMID DNA KIT

For Research Use Only Ver

Lysis Buffer A Wash Solution A Elution Buffer B Mini Filter Spin Columns 50 Collection Tubes 50 Elution tubes (1.7 ml) 100 Product Insert 1

Note: for laboratory research use only. RNA High-purity Total RNA Rapid Extraction Kit (Spin-column) Signalway Biotechnology

Product # Kit Specification. Kit Specifications Maximum Column Binding Capacity 50 µg Maximum Column Loading Volume 650 µl

UltraClean Endotoxin-Free Mini Plasmid Prep Kit

EZ-10 SPIN COLUMN GENOMIC DNA MINIPREPS KIT HANDBOOK

E.Z.N.A. Yeast Plasmid Mini Kit. D preps D preps

TIANamp Yeast DNA Kit

MD60002 MD MD62002

Instructions for Use

GenElute Plasmid Miniprep Kit

E.Z.N.A. Stool DNA Kit. D preps D preps D preps

GenepHlow Gel Extraction Kit

TIANamp Marine Animals DNA Kit

EasyPrep TM Total RNA Extraction Miniprep Manual

Biospin Plasmid DNA Maxi Extraction Kit

Plasmid DNA Purification

Tissue & Cell Genomic DNA Purification Kit. Cat. #:DP021/ DP Size:50/150 reactions Store at RT For research use only

Cytoplasmic & Nuclear RNA Purification Kit Product # 21000, 37400

GenUP Virus DNA/RNA Kit

E.Z.N.A. Blood DNA Mini Kit. D preps D preps D preps

PureColumn DNA Extraction Kit

BIO BASIC. Gel Extraction Kit. 9K s (10 preps) 9K (100 preps) 9K (200 preps) For Research Use Only

Product # Specifications. Kit Specifications Maximum DNA Input. 200 g Maximum DNA Volume Input Final Endotoxin Levels

RNA Clean-Up and Concentration Kit Product # 23600, 43200

E.Z.N.A. MicroElute Genomic DNA Kit. D preps D preps D preps

Purification of cytoplasmic RNA from animal cells using the RNeasy Mini Kit

OneKit Genomic DNA Purification Handbook

For Research Use Only Ver

QIAamp Fast DNA Stool Mini Handbook

E.Z.N.A. Plant RNA Kit. R preps R preps R preps

GenElute Plant Genomic DNA Miniprep Kit

PowerSoil DNA Isolation Kit

Plant/Fungi Total RNA Purification Kit Product # 25800, 31350, 25850

E.Z.N.A. Water DNA Kit. D preps D preps D preps

Manual innuprep Micro RNA Kit

Size-Select Kit for NGS Library Preparation Product # 53600

Plasmid Mini Kit for high-copy plasmid DNA purification. version 0617

Gel Extraction & PCR Purification Combo Kit (Spin-column)

Introduction. Principle

Thermo Scientific GeneJET NGS Cleanup Kit

HiPer Yeast Genomic DNA Extraction Teaching Kit

QIAfilter Plasmid Midi Kit (Cat #: 12243)

Kit Specifications 45 g. 45 g of RNA 8 L

Mag-Bind Universal Pathogen 96 Kit. M x 96 preps M x 96 preps

Aurum Plasmid Mini Kit. Instruction Manual. Bio-Rad Laboratories, Inc Alfred Nobel Dr. Hercules, CA USA (510)

FosmidMAX DNA Purification Kit

Table of Contents. II. Kit Components III. Materials required but not supplied VII. Experimental Examples IX. Troubleshooting...

Contents. I. Overview 1. II. Kit Components 1. III. Storage 2. IV. Intended Use 2. V. Safety Warnings and Precautions 2. VI. Warranty and Liability 2

HighPrep Blood & Tissue DNA Kit

E.Z.N.A. Total RNA Kit II. R preps R preps R preps

E.Z.N.A. Plasmid DNA Mini Kit I

Genomic DNA Purification from Tissue

Bioneer Corporation 8-11, Munpyeongseo-ro, Daedeok-gu, Daejeon 34302, Republic of Korea Tel: Fax:

Kit for total DNA isolation from yeast

Product # Specifications. Kit Specifications Maximum Column Binding Capacity (gdna) Maximum Column Loading Volume.

TIANamp Soil DNA Kit. For isolation of genomic DNA from soil.

Transcription:

HiYield Plasmid Kit Mini Maximum Yield Mini System Protocol Book YPD100 // YPD300 Ver. 2017-1

Precautions I) Handling Requirements Do not use a kit after its expiration date has passed. Some reagents contain the hazardous compounds guanidine thiocyanate or guanidine hydrochloride. Do not let these reagents touch your skin, eyes, or mucous membranes. If contact does occur, wash the affected area immediately with large amounts of water. If you spill the reagents, dilute the spill with water before wiping it up. Do not allow reagents containing guanidine thiocyanate to mix with sodium hypochlorite solution or strong acids. This mixture can produce a highly toxic gas. II) Laboratory Procedures Handle all samples and the resulting waste as if potentially infectious, using safe laboratory procedures. As the sensitivity and titer of potential pathogens in the sample material varies, the operator has to optimize pathogen inactivation by the Lysis Buffer or take appropriate measures according to local safety regulations. RBC Bioscience does not warrant that samples treated with Lysis Buffer are completely inactivated and noninfectious. After sample processing is completed, remove and autoclave all disposable plastics, if you worked with potentially infectious sample material. Do not eat, drink or smoke in the laboratory work area. Wear protective disposable gloves, laboratory coats and eye protection when handling samples and kit reagents. Do not use sharp or pointed objects when working with the reagent cartridge, in order to prevent damage of the sealing foil and loss of reagent. Do not contaminate the reagents with bacteria, virus, or ribonuclease. Use disposable pipettes and RNase-free pipette tips only to remove aliquots from reagent bottles. Use the general precautions described in the literature. Wash hands thoroughly after handling samples and test reagents. III) Waste Handling Discard unused reagents and waste in accordance with country, federal,state and local regulations.

CONTENTS HiYield Maximum Yield Mini System Plasmid Kit Mini Cat.No. YPD100//YPD300 High Copy Number Protocol Low Copy Number Protocol Trouble Shooting 1 3 5 8

www.rbcbioscience.com Maximum Yield Mini System HiYield Plasmid Kit Cat.No. YPD100//YPD300 Kit Contents Cat.No. YPD100 100 mini preps / kit PD1 Buffer(B008)...25ml PD2 Buffer(B009)...25ml PD3 Buffer(B010)...40ml W1 Buffer...50ml Wash Buffer (concentrated)*...25ml Elution Buffer...10ml RNase A (50mg/ml)(RN050)...50µl 2ml Collection Tube...100pcs PD Cloumn...100pcs Cat.No. YPD300 300 mini preps / kit PD1 Buffer(B046)...65ml PD2 Buffer(B047)...75ml PD3 Buffer(B048)...100ml W1 Buffer...130ml Wash Buffer (concentrated)**...40ml Elution Buffer...30ml RNase A (50mg/ml)(RN130)...130µl 2ml Collection Tube...300pcs PD Cloumn...300pcs PD1,2,3 Buffers and RNase A can be purchased separately. (Cat.No. B008~B010 / B046~B048, RN050/RN130) Sample Source: Plasmid DNA from Bacteria Sample Size: 1-4ml of LB broth overnight incubate bacterial cultures Typical Plasmid Yield : Low Copy Number : 0.5-5µg // High Copy Number : 10-20µg (max 30µg) Preparation time : 20min Add provided RNase A to PD1 Buffer and store at 4 C. If precipitates have formed in PD2 Buffer, warm the buffer at a 37 C waterbath to dissolve. * Add 100ml ethanol (96-100%) to Wash Buffer prior to initial use. ** Add 160ml ethanol (96-100%) to Wash Buffer prior to initial use. 1 HiYield Plasmid Kit

Description The HiYield Plasmid Mini Kit is designed for rapid isolation of plamid or cosmid DNA from 1-4 ml of bacterial cultures. The modified alkaline lysis method followed by RNase treatment is utilized to obtain cleared cell lysate with minimal genomic DNA or RNA contamination. Silica spin technology coupled with chaotrophic salt provides a reliable DNA binding and elution system. Purified DNA is ready for restriction digestion, ligation, PCR and sequencing reaction. Quality Control The quality of HiYield Plasmid Mini Kit is test on a lot-to-lot basis. The kits are tested by isolation of plasmid DNA from 3ml culture of E.coli DH5 transformed with the plasmid puc19(a 600 >2units/ml). More than 20μg of plasmid DNA was quantified with spectrophotometer. Reference Birnboim, H. C., and Doly J. (1997) Nucleic Acids Res. 7, 1513. (2) Vogelstein B., and Gillespie D. (1979) Proc. Natl Acad Sci. USA 76,615. Note * For research use only. Not for use in diagnostic or therapeutic procedures. * PD3 buffer contains guanidine hydrochloride which is harmful and irritant. During operation, always wear a lab coat, disposable gloves and protective goggles. HiYield Plasmid Kit 2

www.rbcbioscience.com High Copy Number Protocol Harvesting 1. Transfer 1.5ml of bacterial culture to a microcentrifuge tube (not provided). 2. Centrifuge for 1 min at full speed (13,000 rpm) in a microcentrifuge and discard supernatant. (If more than1.5 ml of bacteria culture is used, repeat the Harvesting Step. For over 4ml, use multiple columns.) Resuspension 3. Add 200μl of PD1 Buffer (RNase A added ) and resuspend the cell pellet by vortexing or pipetting. Lysis 4. Add 200μl of PD2 Buffer and mix gently by inverting the tube 10 times. Do not vortex, avoid shearing genomic DNA. 5. Allow mixture to stand for 2 minutes at room temperature until lysate clears. Neutralization 6. Add 300μl of PD3 Buffer and mix immediately by inverting the tube 10 times. Do not vortex. 7. Centrifuge for 2 min at full speed (13,000 rpm). DNA Binding 8. Place a PD Column in a 2ml Collection Tube. 9. Apply the clear lysate (supernatant) from Step 7 to the PD Column. 10. Centrifuge at full speed (13,000 rpm) for 30 seconds. 11. Discard the flow-through and return the PD Column back to the 2ml Collection Tube. 3 HiYield Plasmid Kit

Wash 12. Add 400μl of W1 Buffer in the PD Column. 13. Centrifuge at full speed (13,000 rpm) for 30 seconds. 14. Discard the flow-through and return the PD Column to the 2ml Collection Tube. 15. Add 600μl of Wash Buffer (ethanol added) to PD Column. 16. Centrifuge at full speed (13,000 rpm) for 30 Seconds. 17. Discard the flow-through and return the PD Column to the 2ml Collection Tube. 18. Centrifuge again for 3 min at full speed to dry the column matrix. DNA Elution 19. Transfer the dried PD Column to a clean 1.5ml microcentrifuge tube (not provided). 20. Add 50μl of Elution Buffer or ddh 2 O (ph 8.0-8.5) directly onto the centre of the membrane. Avoid residual buffer adhering to the wall of the column. 21. Allow to stand for 2 min until the liquid is absorbed. 22. Centrifuge for 2 min at full speed (13,000 rpm) to elute plasmid DNA. HiYield Plasmid Kit 4

www.rbcbioscience.com Low Copy Number Protocol Add ethanol and RNase A to buffers according to component instructions. The typical yield is about 0.5-1.0 μg per 1ml culture when preparing low-copy-number plasmid from overnight bacterial culture in LB medium. If the plasmid is larger than 30 kb, preheat the Elution Buffer to 70 C prior to the Elution Step. Harvesting 1. Harvest up to 10ml of overnight culture by centrifugation. Resuspension 2. Add 400μl of PD1 Buffer (RNase A added) to the tube and resuspend the cell pellet by votexing or pipetting. Lysis 3. Add 400μl of PD2 Buffer and mix gently by inverting the tube 10 times. Do not vortex to avoid shearing genomic DNA. 4. Allow mixture to stand for 2 min at room temperature until lysate clears. Neutralization 5. Add 600 μl of PD3 Buffer and mix immediately by inverting the tube 10 times. Do not vortex. 6. Centrifuge for 3 min at full speed (13,000 rpm). 5 HiYield Plasmid Kit

DNA Binding 7. Place a PD Column in a 2ml Collection Tube. 8. Apply 750μl of the clear lysate (supernatant) from step 6 to the PD Column. 9. Centrifuge at 10,000 xg (13,000rpm) for 30 seconds. Discard the flow-through and return the PD Column to the 2ml Collection Tube. 10. Apply the remaining clear lysate to the same PD Column. 11. Centrifuge at 10,000 xg (13,000rpm) for 30 seconds. 12. Discard the flow-through and return the PD Column to the 2ml Collection Tube. Wash 13. Add 400μl of W1 Buffer in the PD Column. 14. Centrifuge at 10,000 xg (13,000rpm) for 30 seconds. 15. Discard the flow-through and return the PD Column to the 2ml Collection Tube. 16. Add 600μl of Wash Buffer (ethanol added) in the PD Column. 17. Centrifuge at 10,000 xg (13,000rpm) for 30 seconds. 18. Discard the flow-through and return the PD Column to the 2ml Collection Tube. 19. Centrifuge again for 2 min at full speed to dry the column matrix. DNA Elution 20. Transfer the dried PD Column onto a clean microcentrifuge tube (not provided). 21. Add 50μl of Elution Buffer to the center of the column matrix. If plasmid DNA is larger than 10kb, use preheated Elution Buffer (70 C) during Elution Step to improve the elution eciency. 22. Allow to stand for 2 min until the liquid is absorbed by the matrix. 23. Centrifuge for 2 min at full speed to elute plasmid DNA. HiYield Plasmid Kit 6

www.rbcbioscience.com Spin Down Bacterial cells Cell Lysis Spin DNA Binding Wash Step Elution Step 7 HiYield Plasmid Kit

Troubleshooting Problem Low yield Eluted DNA does not perform well in downstream applications Possible Reason/Solution Bacterial cells were not lysed completely Too many bacterial cells were used. If using more than 10A 600 units of bacterial culture, separate into multiple tubes. Following PD3 Buffer addition, break up the precipitate by inverting to ensure higher yield. Incorrect Wash Buffer Check to ensure Ethanol was added to Wash Buffer prior to use. Incorrect DNA Elution Step Ensure that Elution Buffer was added and absorbed to the center of PD Column matrix. Incomplete DNA Elution If plasmid DNA is larger than 10kb, use preheated Elution Buffer (70 C) during Elution Step to improve the elution efficiency. Residual ethanol contamination After wash step, dry PD Column with additional centrifugation at top speed for 5minutes or incubation at 60 C for 5 minutes. RNA Contamination Prior to using PD1 Buffer, ensure that RNase A was added. If RNase A added PD1 Buffer is out of date, add additional RNase A. Too many bacterial cells were used, reduce sample volume. Genomic DNA contamination Do not use overgrown bacterial culture. During PD2 and PD3 Buffer addition, mix gently to prevent genomic DNA shearing. Nuclease contamination If host cells have high nuclease activity(e.g.,enda + strains), perform this Optional Wash Step to remove residual nuclease. After DNA Binding Step, add 200ml of PD3 Buffer to PD Column and incubate for 2 min at room temperature. Centrifuge at 6000 xg (8,000rpm) for 30 seconds. Continue from standard Wash Step. HiYield Plasmid Kit 8