[ Care and Use Manual ]

Similar documents
Protein-Pak Hi Res HIC Column and HIC Protein Standard

Hillary B. Hewitson, Thomas E. Wheat, Diane M. Diehl INTRODUCTION

Ultrahydrogel Columns

Agilent MetaCarb Ca PLUS Carbohydrate Column Ca ++ Form User Manual Part No. A5205

Instructions for Capillary Electrophoresis Peptide Analysis Kit

Chromatography Column Performance and Data Analysis Success Guide. Hints and Tips for Better Purifications

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime

Chromatography Success Guide. Tips and Tricks for Successful Preparative Chromatography and Protein Purification

AFFINITY HIS-TAG PURIFICATION

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

Econo-Pac High Capacity Ion Exchange Cartridges, 5 ml Instruction Manual. Catalog Numbers , ,

Nuvia cprime Hydrophobic Cation Exchange Media

G Sep Size Exclusion Columns

3. Close the bottom end of the column and apply the packing device on top. Pump water through the upper adaptor to remove air.

Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography

Application Note USD Purification of Mouse IgM from Cell Culture Supernatant by Cation Exchange Chromatography on CM Ceramic HyperD F Sorbent

Affi-Gel Protein A MAPS II Kit Instruction Manual

AFFINITY HIS-TAG PURIFICATION

PROCEDURE FOR USE NICKEL NTA Magnetic Agarose Beads (5%)

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Nickel-NTA Agarose Suspension

MassPREP On-Line Desalting Cartridge

ANALYTICAL STANDARDS AND REAGENTS

INSTRUCTIONS The resins are adapted to work mainly in native conditions like denaturing.

[ product solution ] Waters Oasis µ Elution PlateS. Patented Innovation. Elution volume as low as 25 μl. No evaporation and reconstitution

ACQUITY UPC 2 BEH, CSH, and HSS Columns

columns IonPac ICE-AS1 Ion-Exclusion Column Superior Chromatographic Performance Separates weakly ionized acids by differences in pk a

AFFINITY HIS-TAG PURIFICATION

Rev. CO Packed Column for Ion Exchange Chromatography. TSKgel STAT Series INSTRUCTION MANUAL TOSOH CORPORATION

Userguide: SepFast Gel Filtration columns

1 ml gel corresponds to ml of 75% (v/v) Glutathione Agarose suspension.

Phenyl Sepharose CL-4B

On-Line. User s Guide SPE CARTRIDGES. for Rapid Cleanup and Extraction of Analytes

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website: Published in November 2013

1. Connect the column to the chromatography system, syringe or pump.

PESTICIDE SCREENING APPLICATION SOLUTION

5 AMP Sepharose 4B. instructions

Improving Resolution and Column Loading Systematically in Preparative Liquid Chromatography for Isolating a Minor Component from Peppermint Extract

ACQUITY UPLC Protein BEH SEC Columns and Standards

New Core-Shell Technology

ISOLATE II Blood DNA Kit. Product Manual

Size-Select Kit for NGS Library Preparation Product # 53600

Pulsed Field Gel Electrophoresis

Electrophoresis. Ready-to-Run Buffers and Solutions

ProteoMiner Protein Enrichment Technology

AFFINITY GST PURIFICATION

ENSURING QUALITY THROUGH COMPLIANCE [ COMPLIANCE ]

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

TechnicalNOTE METABOLYNX EXACT MASS DEFECT FILTER: RAPIDLY DISCRIMINATE BETWEEN TRUE METABOLITES AND FALSE POSITIVES INTRODUCTION

Analysis of biomolecules by SEC and Ion-Exchange UPLC

PRODUCT MANUAL FOR THE. ELECTROLYTIC ph MODIFIER (EPM) EPM, P/N

E.Z.N.A. Water DNA Kit. D preps D preps D preps

Kit Components Product # (50 samples) Wash Solution A Elution Buffer B

E.Z.N.A. Blood DNA Maxi Kit. D preps D preps

QIAcube HT Your Purification Expert

Product # Specifications. Kit Specifications Maximum Column Binding Capacity (gdna) Maximum Column Loading Volume.

Fast and reliable purification of up to 100 µg of transfection-grade plasmid DNA using a spin-column.

CORESTA RECOMMENDED METHOD N 61

Average Yields* Yeast DNA Yeast RNA Time to Complete 10 Purifications * Yield will vary depending on the type of sample processed

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

BabyBio S BabyBio Q BabyBio DEAE

ENHANCED MS WITH THE TURN OF A KEY

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

KOBRA CELL Art. No. K01

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Module 16: Gel filtration: Principle, Methodology & applications. Dr. Savita Yadav Professor Department of Biophysics AIIMS, New Delhi

E.Z.N.A. Yeast Plasmid Mini Kit. D preps D preps

SYKAM Amino Acid Analyzer

Tissue & Cell Genomic DNA Purification Kit. Cat. #:DP021/ DP Size:50/150 reactions Store at RT For research use only

Product # 24700, 24750

MAbPac SCX-10 Columns

AnaTag HiLyte Fluor 647 Protein Labeling Kit

ProteoSpin Urine Protein Concentration Maxi Kit Product # 21600

Please read all the information in booklet before using the unit

QIAcube Pure Efficiency

This User Bulletin does not apply to PROCISE clc instruments.

MD60002 MD MD62002

Purification of Recombinant Proteins on Nuvia TM cprime TM Hydrophobic Cation Exchange Media: A Simple Approach to Method Development

Note: for laboratory research use only. RNA High-purity Total RNA Rapid Extraction Kit (Spin-column) Signalway Biotechnology

Cytoplasmic & Nuclear RNA Purification Kit Product # 21000, 37400

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates

Method Development Considerations for Reversed-Phase Protein Separations

Kit Specifications 45 g. 45 g of RNA 8 L

AxyPrep PCR Clean-up Kit AxyPrep-96 PCR Clean-up Kit

[ VION IMS QTOF ] BEYOND RESOLUTION

A RAPID METHOD FOR THE DETERMINATION OF CHLORAMPHENICOL RESIDUES IN BLACK TIGER SHRIMP

RNA Clean-Up and Concentration Kit Product # 23600, 43200

TaKaRa MiniBEST Agarose Gel DNA Extraction Kit Ver.4.0

AccuPrep GMO DNA Extraction Kit Cat. No.: K-3031

Histology FISH Accessory Kit Step-by-Step Procedure

Operating Instructions

MicroElute Cycle-Pure Kit

QIAsymphony SP Protocol Sheet

Fast, easy, reliable, ultra-pure transfection grade plasmid DNA Maxiprep using a microcentrifuge spin-column.

1. Bloomsbury BBSRC Centre for Structural Biology, Birkbeck College and University College London.

HiTrap Heparin HP, 1 ml and 5 ml

Polyvidone Polyvinylpyrrolidone H 2 C H C N

Bivalirudin Purification:

NPTEL VIDEO COURSE PROTEOMICS PROF. SANJEEVA SRIVASTAVA

Transcription:

Amino Acid Analysis Liquid Chromatography Column I. Introduction This manual provides the user with information necessary to obtain the maximum efficiency and life from the Waters Amino Acid Analysis Column. This column has been specifically designed for the separation and analysis of free amino acids. The column is a precision device. Waters Corporation uses very exacting manufacturing procedures to achieve a small average particle size, narrow range of particle size distribution and stable particle structure in the column. These procedures create the high efficiency column-to-column reproducibility and long separation life built into each amino acid column from Waters. However, care must be taken to maintain packing density and protect it from contamination by foreign material. The following information will aid you in obtaining the highest performance and longest life from your column. II. Background Information Although there are some hydrophobic interactions between the column and certain amino acids, the primary mode of separation is that of ion-exchange. The cation-exchange resin utilized has a large number of negatively charged (sulfonate) functional groups which are covalently bound to it. These negatively charged groups, in turn, readily associate with positively charged cations when they are present in solution. Before the amino acid sample is injected onto the column, the resin is equilibrated at a low ph with buffer containing a high concentration of sodium (Na+) ions. This places the column in the sodium form. As amino acids enter the column in this acidic buffer, they are strongly positively charged and readily exchange with the loosely bound sodium ions. As the buffer gradually changes to a higher ph during the gradient run, each amino acid becomes less positively charged reflecting its pk value and differentially elutes accordingly. Contents I. Introduction II. Background Information a. Elution System 1 b. Elution System 2 c. General Preparation Guide for One Liter of Eluent d. Use of Non-aqueous Solvents e. Backpressure f. Sample Considerations g. SEP-PAK C 18 Cartridge Clean Up for Amino Acid Samples h. Preparation Scheme III. Column Care & Use Considerations a. Column Care Considerations b. Column Storage c. Column Regeneration d. Troubleshooting e. Regeneration of Protein and Lipid Contaminated Column f. Regeneration of Heavy Metal Contaminated Column g. Available Service and Applications Information h. Chromatographic Troubleshooting i. Typical Separation IV. Column Performance Checks V. Column End Fittings a. End Fitting Removal Procedure b. Column Hookup c. Storage Appendix A. Amino Acid Analysis Column Specifications

Figure 1: Ion Exchange Mechanism of the halide ion requires immediate flushing of the system with distilled water containing 0.2% sodium azide or 0.1% trifluoroacetic acid before system shutdown. Allowing the halide buffers to sit static in the pumps overnight will lead to corrosion. a. Elution System 1: Waters Non-halide Buffer System Buffer Composition Buffer A-ES2 Sodium Citrate Dihydrate: 19.6 g Phenol (Preservative): 1.0 g HCI: to ph 3.00 Total Volume: 1 L Buffer B-ES1 Boric Acid: 1.5 g Sodium Nitrate: 21.0 g 6M NaOH: to ph 9.60 Total Volume: 1 L NOTE: Do not add phenol to Buffer B; it will decompose at this ph. CAUTION: Introduction of improperly prepared or contaminated buffers or samples onto the resin bed could irreversibly poison the column. Heavy metal ions can also poison the column. Column degradation will be indicated by band broadening loss in resolution and/or change in selectivity. All reagents to be used for high performance liquid chromatography (HPLC) analysis of amino acids must be of the highest purity. Water should be distilled and deionized with a purity equivalent to that obtained from a Milli-Q water purifica tion system. Impure buffers will lead to artifact peaks during analysis and shorten column lifetime. There are two elution systems available for analysis: 1) Elution System 1 is the Waters Non-halide Buffer System. It is recommended for use where possible since it employs no potentially corrosive halide ions. b. ELUTION SYSTEM 2: Halide Buffer System Buffer Composition Buffer Composition Buffer A-ES2 Sodium Citrate Dihydrate: 19.6 g Phenol (Preservative): 1.0 g HCI: to ph 3.00 Total Volume: 1 L Buffer B-ES2 Boric Acid: 1.5 g Sodium Chloride: 14.42 g 6M NaOH: to ph 9.80 Total Volume: 1 L 2) Elution System 2 is a halide-containing buffer system and is recommended when maximum sensitivity is needed. Use of chloride enhances orthopthalaldehyde (OPA) fluorescence by approximately twofold as compared to the non halide elution system. The presence 2

c. General Preparation Guide for One Liter of Eluent Dissolve salts in approximately 900 ml of purified water. Adjust ph with appropriate acid or base. Adjust to final one liter volume with water. Filter through the Solvent Clarification Kit using an 0.45 µm aqueouscompatible filter. Transfer buffer to an amber glass bottle and purge for 15 minutes with pure nitrogen. Transfer buffer to analyzer. Buffers prepared in this manner will be stable for at least one week at room temperature. d. Use of Non-aqueous Solvents For additional resolution of the early-eluting amino acids, small amounts of methanol, ethanol, or propanol can be added to Buffer A. This alcohol concentration should not exceed 10% High concentrations of a non-aqueous modifier will cause the resin to swell resulting in high backpressure. e. Backpressure With a new column flowing at 0.4 ml/min in Buffer A at 62 C. typical system pressures will range from 800 psi - 1500 psi. As the column ages, pressure will slowly increase, although this increase is not usually associated with decreased efficiency. The most common causes of high backpressure are: particulates in the sample or buffer, proteinaceous material coating the resin packing, mold formation in the buffers or in the column. Reduced temperature, higher flow rate and high solvent viscosity will all yield higher pressure. To prevent excessive pressure, do not flow through the column with the temperature control module off. f. Sample Considerations Acidic protein hydrolyzates should be lyophilized and reconstituted in Buffer A. Convenient sample concentration range: 0.20-25 nmol amino acid/5 µl - 50 µl injection volume. Sample volume and sensitivity requirements should be optimized for individual needs. Samples that contain protein, lipid or heavy metals (e.g., tissue extracts, food samples, feed samples, etc.) should not be introduced directly on the column. Pre-purification on a strong exchange resin or a Sep-Pak C 18 cartridge is recommended. g. Sep-Pak C 18 Cartridge Cleanup for Amino Acid Samples Prepare the following solutions: Solution 1: 0.1 % trifluoroacetic acid (TFA) in water Solution 2: 0.1 % TFA in H20:MeOH, 80:20 Solution 3: 0.1 % TFA in H2O:MeOH, 70:30 h. Preparation Scheme 1. Activate a new Sep-Pak C 18 cartridge with two 10 ml volumes of methanol. 2. Wash with two 10 ml volumes of Solution 1. 3. Wash with 10 ml of Solution 2. 4. Mix 1 ml sample with 2 ml Solution 3. Sample should be water or acid solubilized and free of particulate matter. 5. Pass the sample through the Sep-Pak C 18 cartridge. 6. Discard the first 1 ml of eluent and collect the next 2 ml. This fraction will contain all amino acids. Lipids and high molecular weight pro teins will be retained on the Sep-Pak C 18 cartridge. 3

III. Column Care & Use Considerations a. Column Care Considerations ALWAYS filter samples to prevent excessive pressure buildup due to particulate matter. Use a Sample Clarification Kit (Part No. 26865) and, if necessary, centrifugation. For more details on the Waters Sample Clarification Kit call or write for Bulletin D97. ALWAYS filter buffers to remove microparticulate matter. This eliminates high pressures from plugged frits and also helps ensure long column life. Use a Solvent Clarification Kit (Part No. 85113) to filter solvents. Vacuum filtration or sonification may be used to remove dissolved gasses which could affect your solvent delivery system. For more details on the Waters Solvent Clarification Kit call or write for Bulletin D97. ALWAYS use high purity solvents such as Waters solvents. For analytical work choose high quality Waters solvents. For more details on Waters solvents call or write for Bulletin M10. ALWAYS replace plugs in column fittings to prevent the column from drying out when not in use. b. Column Storage The column may be stored 1-2 days in Buffer A without flow. The presence of phenol in the buffer will prevent bacterial growth. For longer storage periods, 0.2 N NaOH is strongly recommended. c. Column Regeneration For protein hydrolyzate samples, regenerate the column at least once a week as follows: d. Troubleshooting This section discusses some possible problems and their solutions (refer to Table 1.) e. Regeneration of Protein and Lipid Contaminated Column Large proteins and some lipids can be removed following alkaline hydrolysis. Reverse direction of buffer flow through column. Elevate column heater temperature to 90-95 C and pump 0.2 N NaOH solution through column for at least 2 hours. Particularly difficult contamination may require overnight treatment. Restore normal flow direction and temperature; reequilibrate column with Buffer A and test column with amino acid standard solution. If problem still exists, repeat the above treatment. Table 1: Typical Column Problems and Solutions Problem Cause Solution Excess pressure buildup Loss of resolution, broad peaks, low plate counts Column frits plugged with particulates or mold is growing in column or in buffers Sample and/or buffer precipitates on column Contaminated column Reverse flow on column. If this does not work, follow procedure for protein removal. If pressure is still high, carefully remove column inlet fitting being sure not to disturb resin. Clean in ultrasonic bath. Always filter solvents and samples. Follow procedure above. Follow procedures for metal and protein removal. Eluent: 0.2 N NaOH Flow: 0.5 ml/min Temperature: 62 C Flush Time: 30 min Crude samples may require more frequent regeneration. 4

f. Regeneration of Heavy Metal Contaminated Column Often metals can be removed by chelation under basic and acidic conditions. Reverse direction of buffer flow through column. Elevate column heater temperature to 90-950C and pump 0.2 N NaOH solution containing 0.25 g EDTA/liter through the column for minimum of 4 hours. Switch buffer tube to second container and pump 2 N HNO, solution containing 0.25 g EDTA/liter through the column for another 2-4 hours. The more advanced Waters HPLC System allows for the generation of a continuous gradient using only two buffers. This is equivalent to an infinite step gradient. Superior resolution is achieved without baseline disturbances. The constant ionic strength, ph gradient (0.2N Na +, ph 3.00 to ph 9.60) allows for rapid reequilibration without NaOH regeneration. i. Typical Separation Figure 2: Amino Acid Analysis Continuous Gradient Restore normal flow direction and temperature. Reequilibrate column with Buffer A and test column with amino acid standard solution; If problem still exists, repeat the above basic/acidic EDTA treatment. CAUTION: Never attempt to clean the column using hydrochloric acid (HCI) or sulphuric acid (H2SO,) solutions. Such acids are harmful to stainless steel. g. Available Service and Applications Information The Waters Corporation staff of trained and e perienced service specialists provides maintenance for Waters instruments on preventive and/or corrective levels. Contact Waters Corporation at 508-478-2000 or your local Waters representative for further details. For answers to specific chromatography questions in areas such as methods development, applications and quality control, contact your local Waters representative or Waters home office applications laboratory. As well as the standard hydrolyzate amino acids, unusual amino acids can be separated by the two-buffer, gradient system. Note cysteic acid, hydroxyproline, methionine sulfoxide, hydroxylysine, and tryptophan. Slight modifications of the ph and/or constituency of the A or B buffers can quickly be used to optimize new separations. h. Chromatographic Troubleshooting When a new column is installed, it is recommended that a separation of standard hydrolyzate amino acids be performed using Elution System 1. A typical separation is shown on the following page (Figures 3 & 4). 5

Figure 3: Amino Acid Analysis Special Amino Acids IV. Column Performance Checks 1. Cysteic Acid 2. Methionine Sulfoxide 3. Hydroxy Proline 4. Aspartic Acid 5. Threonine 6. Serine 7. Glutamic Acid 8. Proline 9. Glycine 10. Alanine 11. Cystine 12. Valine 13. Methionine 14. Isoleucine 15. Leucine 16. Tyrosine 17. Phenylalanine 18. Histidine 19. Hydroxy Lysine 20. Tryptophan 21. Lysine 22. Ammonia 23. Arginine Column: CAT EX Resin Column (0.4 cm x 25 cm) Sample: 1 NMOL AA STDS Injection Volume: 10 µl Flow Rate: 0.4 ml/min Mobile Phase: A. 0.2 N Na citrate, ph 3.0 B. 0.2 N Na borate, ph 9.8 0 100% B, curve 6, 48 min Detector: Model 420 ex: 338 nm em: 425 nm gain: 4 Table 2 provides performance check information. Use the table in conjunction with the following procedure to determine the condition of your column. If you feel it has become contaminated, refer to Troubleshooting section for appropriate column cleaning procedures. Table 2: Performance Check for Columns Column: Buffer: Flow rate: Waters Amino Acid Analysis Column (Part No. 80002) A-ES1 Temperature: 62 C Chart Speed: Detection: 0.4 ml/min 0.25 cm/min Fluorescence (o-phthalaldehyde) Sample: Sample Amount: Results: Amino acid mixture contain ing equalmolar concentra tion aspartic acid, threonine, serine, and glutamic acid 1-2 n mole of each amino acid in 5-20 µl volume Chromatogram should show a minimum resolution of 85% for all amino acids as determined by the following equation: Figure 4: Resolution Variations in results as compared to the expected chromatogram can usually be explained by improper buffer preparation (weighing/volume error) or improper ph. The following guidelines describe the effect of ph: 1. If Buffer A is 0.20 ph units too high, there will be a loss of THR-SER resolution and OH-PRO will coelute with ASP. 2. If Buffer A is 0.30 ph units too low, there will be broadening of the neutral amino acid peaks and ASP will elute later than expected. 3. If Buffer B is 0.20 ph units too high, ARG will coelute with NH. If Buffer B is 0.20 units too low, ARG will elute 10-15 minutes later than expected. Increased column temperature will lead to a decrease in the elution volume for all amino acids and a decrease in the THR-SER resolution. 6

All peaks should be sharp and symmetrical. Backpressure should be less than 2,000 psi. If you suspect that the quality of your column has decreased, the above test is a rapid method of determining the efficiency of your separation. Peaks which are not symmetrical could in dicate partially plugged frits are at fault or foreign material has accumulated on the resin packing. Generally this phenomenon is also ac companied by increased backpressure (refer to Table 2 for corrective action). Alternatively, resin packing material may have shifted, voided or become uneven. Careful removal of top column packing nut and visual inspection of the resin bed is the best diagnostic procedure to determine whether this is the cause. Since removal of the nut could, itself, disturb the packing, this should be a last resort measure. If resin bed is not smooth and completely full, please contact your nearest Waters representative for instructions. If backpressure is normal but resolution is poor or peaks are very broad to nonexistent, the problem could be heavy metal contamination of the resin (refer to Table 1 for corrective action). It should always be kept in mind that column efficiency is affected by the liquid chromatography (LC) system as well as buffers and reagents. Care should be taken to make sure the LC system is working normally, that tubing and post-column reaction coil(s) are properly connected and leak free, and that buffers and reagents are fresh, filtered and tree from contamination. Whenever problems persist after making the above tests and checks on your system, con tact your local Waters representative for further assistance. V. Column End Fittings The column is symmetrical, having the same kind of fittings and frits at both ends. Flow of buffer can occur in either direction without harmful effects. However, the end bearing and column label should be considered the inlet since this was the direction of flow during packing and testing procedures. Use of the same flow direction for analysis is encouraged, and it will prove helpful during any necessary cleaning procedures. Both end fittings are made up of a smaller and a larger hex nut. The larger hex nuts contain a non-removable stainless steel frit to both retain the resin packing and to catch particulates before they reach the resin. In normal operation particulates from the sample or buffer should be filtered out using a 0.45 µm membrane filter (refer to Section III d.). If the backpressure of the column should rise above 2500 psi, reversing the direction of flow on the column for a few minutes will usually flush particulates away. If flushing and cleaning procedures (see Troubleshooting, Section III d.) do not work, the frits may be cleaned ultrasonically following their removal from the column. a. End Fitting Removal Procedure CAUTION: Removal of end fittings may disturb the resin packing material and cause skewed or broad peaks. Therefore, this procedure should be used only as a last resort measure. 1. Work only in a clean area where dust, dirt and debris will not contaminate the exposed packing material. 2. Secure column in a vice. Hold in an upright position by clamping lightly (do not over tighten) onto the smaller hex fitting. 3. Remove only one fitting at a time to avoid loss of packing material. Always begin by removing column inlet (end with column label) first. 4. Use 5/8 inch (16 mm) wrench, turning counterclockwise slowly until fitting is removed. 5. With a clean spatula or similar tool, remove any resin packing which has adhered to the frit and place smoothly and carefully back on to resin bed at end of column tube. 6. Fritted fitting may now be cleaned in ultrasonic cleaner using appropriate solvent. 6 N nitric acid (HNO,) often works well but be very cautious to avoid contact with skin. 7. Column fittings should be reassembled as soon as possible after inspection and cleaning to avoid drying and contamination of resin. 7

b. Column Hookup 1. Direction of flow is indicated on the column. 2. Connect one adaptor to the inlet and one adaptor to the outlet of the column. 3. Connect the tubing from the injector to the column inlet union. Be sure to use 0.009-0.010 inch (0.23 mm) i.d. tubing for this purpose. 4. Connect the outlet adaptor to Mixing Tee #1 of the post column reaction system. c. Storage If you do not intend to use this column for a period of a week or longer, it is best to store it in 0.2 N sodium hydroxide (NaOH). Using your HPLC, pump the filtered NaOH solution through the column at 0.4 ml/min for at least 30 min. To reduce solution viscosity, column heater should be on at normal analysis temperature (60-70 C). At the end of this period turn off pump and disconnect inlet and outlet tubing. Replace the plug fittings on the column, screwing them in tightly to prevent evaporation and drying out of the resin. Appendix A: Amino Acid Analysis Column Specifications Column Size Length: 25 cm Diameter: 0.46 cm i.d. Packing Material Styrene - divinyl benzene copolymer with sulfonic acid functionality in Na+ form Particle Size 9 ± 0.5 Nm in diameter Separation Method Strong Cation Exchange ph Range 1-14 Pressure Limit 3000 psi Shipping Solvent 0.2 M NaOH (Na + form) The HPLC should now be flushed out with filtered water before it is stored or used for any other pur pose. A stainless steel jumper tube which takes the place of the column will prove useful for this purpose. 8

Sales Offices: Austria and European Export (Central South Eastern Europe, CIS and Middle East) 43 1 877 18 07 Australia 61 2 9933 1777 Belgium 32 2 726 1000 Brazil 55 11 5094 3788 Canada 1 800 252 4752 x2205 China 86 10 8586 8899 CIS/Russia +7 495 3367000 Czech Republic 420 2 617 1 1384 Denmark 45 46 59 8080 Finland 09 5659 6288 France 33 1 30 48 72 00 Germany 49 6196 400600 Hong Kong 852 29 64 1800 Hungary 36 1 350 5086 India and India Subcontinent 91 80 2837 1900 Ireland 353 1 448 1500 Italy 39 02 27 421 1 Japan 81 3 3471 7191 Korea 82 2 820 2700 Mexico 52 55 5524 7636 The Netherlands 31 76 508 7200 Norway 47 6 384 60 50 Poland 48 22 833 4400 Puerto Rico 1 787 747 8445 Singapore 65 6273 1221 Spain 34 93 600 9300 Sweden 46 8 555 11 500 Switzerland 41 62 889 2030 Taiwan 886 2 2543 1898 United Kingdom 44 208 238 6100 2007 Waters Corporation. Waters and Sep-Pak are registered trademarks of Waters Corporation. Milli-Q is a registered trademark of Millipore Corporation. October 2007 WAT085999 Rev 02 VW-PDF Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com