Models TJ30K TJ15K TJ6K TJ3K. Resolution Ratio 1g 0.5g 0.2g 0.1g Calibration graduation 10d 10d 10d 10d

Similar documents
Inscale Measurement Technology Ltd Tel: Fax:

High Precision Balances. Operation Manual

WS 10 Weighing indicator USER MANUAL

Weighing Indicator YH-T3. User Manual

Instruction number PTI-16-03/09/06/A

CONTENTS. 3-1 Unpacking and Checking Installing Components Leveling the Scale... 1

INSTRUCTION MANUAL CONTECH

Model M120 Mini Temperature/Humidity Meter

MEDIZINTECHNIK FÜR TIERÄRZTE

Best Selection for Your Business. Electronic Pricing Scale. User s Manual. (Model: TP-31)

SIWAREX FTC-B Weighing Module for Belt Scales Set-up of SIWAREX FTC with SIWATOOL FTC_B

XSTi Standby UPS. 400VA, 600VA, 800VA Models. User & Installation Manual

Operation Manual. Professional Weighing Equipment. MRW SERIES intelligent weighing scale. WASHDOWN WEIGHING SCALE - Dust and Waterproof Scale IP65

64C Amplification Block. Instruction Manual

OPERATION MANUAL. Electronic Hanging Scale NINGBO AUDDA WEIGHING CO.,LTD

Adam Equipment ADK SERIES

DS-520 Digital Indicator Operation Manual 73366

PMA 31-G. English. Printed: Doc-Nr: PUB / / 000 / 00

C60 Standby UPS. 800VA Models. User & Installation Manual Xtreme Power Conversion Corporation. All rights reserved.

Sby SR Instruments, Inc.

CONTENTS GENERAL Weighing Introduction

GK INDICATOR. (P.N. 9264, Revision K1, July 2011)

Adam Equipment LBH SERIES. (P.N , Revision B2, January 2010)

2% & (+ 2% PSI.

Digital relative humidity & temperature sensor AM2301

COÂNG TY COÅ PHAÀN CAÂN ÑIEÄN TÖÛ THÒNH PHAÙT. Waterproof Weighing Scale. Operation Manual. VMC-WED-en,V

HOBO Plug Load Logger (UX ) Manual

DRS220. DRS220 Digital Refrigerant Scale Limited Warranty. Digital Refrigerant Scale

Operator Manual. brite Advanced Retail scale

Imperial Series. Model IMP-425/525/625/825/1000AP IMD-425/525/625/825/1000AP IMP-1025/1200/1500/2000AP IMD-1025/1200/1500/2000AP

Series. 3 or 12 lb. Capacity. 129 Cramer Road, Jewett, New York USA Fax:

KERN KFB/KFN-TM Version /2010 GB

DWP-5500FP (4 x4 ) DWP-11KRFP (5 X5 ) Weighing Indicator. User Manual. Content

ELCART. Manuale di istruzioni/scheda tecnica. LED Lighting Division APPROVE CHECK DRAW

MANUAL. ENGLISH Antenna Booster Ordercode: D1425. Highlite International B.V. Vestastraat EX Kerkrade the Netherlands

DATA LOGGER Model DL-SA11

KERN HFT Version /2013 GB

PK-110. User Manual. American Weigh Scales. PK-110 (0-44lb x 0.05lb / lb x 0.1lb)

DJ Switch 10F V2 ORDERCODE 50365

DIGITAL PORTION SCALE PS-5 OWNER S Manual

Coating Thickness Gauge Model: AX-CTG10 User s Manual

DMX-BRANCH/4. user manual. ELATION DMX-BRANCH/4 user manual

EGT-AF08 USER GUIDE WHEEL WEIGHING SYSTEMS WITH VARIOUS PLATFORMS

ZK830 COUNTING SCALE. Precision counting scale for industrial use ZK830 COUNTING SCALE

electroplating DC power supply type SIFCO PROCESS POWER PACK 1028

COLow COHigh COLow/H comp NO NOx NO2 CO2 H2S SO2 CXHY LANCOM 4. Portable Flue Gas Monitoring.

Quick Guide. CSN950 MultiHead Dimensioner _v2d_MAN_QG_CSN950MH_EN

MiniCD. User Manual. American Weigh Scales. MCD-100 (100 x 0.01g MCD-500 (500 x 0.1g)

F Series Flow Meter F300. User Manual

1600 Currency counting machines

Operational Manual. Spectrophotometer Model: SP-830 PLUS. Metertech Inc. Version 1.07

TSI Model 3063 Thermal Mass Flowmeter

Owner s Manual & Safety Instructions

Operating Instructions

D Sep ATX324 ATX224 ATX124 ATX84 ATY324 ATY224 ATY124 ATY64

Just a few clicks away from a heating system

AlgaeTorch. The handy measurement instrument for rapid deployment. Fast and simple algae monitoring

7700 TEST KIT OPERATIONS MANUAL

Temperature Controllers: JC Series Digital Indicating Controllers. JC Series. High Performance Controllers...at the lowest prices anywhere!

UniGas Hand-held combustion gas analyzer with 4 sensors

3 PHASE ENERGY ELECTRONIC METERS MKB-363-M INSTRUCTION MANUAL ( M / 00 B

Dual Port isolated USB Charger

Switch-Tek. LV41 & LV42 Series Manual. Sump Float Level Switches

Instruction Manual Dräger MSI EM200-s

Friendly User Interface. Simply AUTO Mode. Support the Universal input voltage

The Ton-Tel Weighbridge

Operating instructions. METTLER TOLEDO B balance line AB PB SB

MAGUIRE PRODUCTS INC. Weigh Scale Dispense System WEIGH SCALE DISPENSE SYSTEM. OPERATIONAL SUPPLIMENT for 6811 Controllers (Red Display)

Precision Automatic Vacuum Control (PAVC+)

Sageon Site Monitor (SSM) Application Note

VENTO WIFI DRONE WITH LIVE STREAMING CAMERA

Rexroth - 440RC Solid De-icer Controller Short Form Configuration & Setup Manual

Cole-Parmer Pro-Spense Disposable Syringe Pump User s Manual. Model # (100ml), # (260ml) Publication REV-A

Instruction Manual HI RH/ C Transmitter. This Instrument is in Compliance with the CE Directives.

RAVAS-310 OWNERS MANUAL

Wireless Freezer Manual Installation Guide

EU-type examination certificate UK 2945 Revision

testo 616 Appareil de mesure de l'humidité

Physikalisch-Technische Bundesanstalt REPORT. No on the. Type Examination of a. Non-Automatic Weighing Instrument

Safety Precautions. 3. Cabinet Assembly 4. Operating buttons / functions.5-6. Machine Operation

VI-D4. Concrete Moisture Meter

USER MANUAL. ExStik CL200A. Waterproof Total Residual Chlorine Tester. Patent Pending. Additional User Manual Translations available at

testo Bedienungsanleitung Instruction manual Mode d emploi Manual de instrucciones Manuale di istruzioni Manual de instruções

Protective Capony RTC-16

Weighloader user manual. Document reference: V2.01

tempmate. -GS Stay Online and Well-Informed to Avoid Unpleasant Surprises: Cold Chain Monitoring 4.0

OM-90 SERIES Portable Temperature and Humidity Data Loggers. Shop online at omega.com

Global Water globalw.com

Global Water Instrumentation, Inc.

RapID Raw Materials Verification System

OPERATOR MANUAL. INDEX page Turning on/off the indicator Weighing Zero weighing Calibration 3

Global Water Instrumentation, Inc.

RJE INTERNATIONAL, INC. U LB-364EL R EV 3.2

600KL Physician Electronic Scale With Integral Digital Height Rod

Global Water Instrumentation, Inc.

MODEL 6745 BABY SCALE

Sartorius Talent Sartorius Gem, Gold

FUNCTION MODULES ANALYSIS CONTROL MODULE ARCHITECTURE MODULE DESCRIPTION PROCESS LEVEL

This Specification Data Sheet brought to you by: Trinity Blvd. Fort Worth, Texas /

RJE INTERNATIONAL, INC.

Transcription:

Operating Instruction Manual of TJ Series Electronic counting balance ( for following models: TJ3K,TJ6K,TJ15K,TJ30K,TJ3KA,TJ6KA,TJ15KA, TJ30KA) 1. Overview TJ series electronic counting weighing apparatus is quality electronic weighing implement carefully designed and manufactured by the G&G GmbH based on highly accurate sensor, measuring circuit and special MPU system from G&G, featuring advanced technology, excellent materials, exquisite workmanship and strict testing means, thus providing this series electronic counting weighing apparatus with the following perfect advantages: 1.1 High accuracy and sensitivity with fast response. 1.2 Use of highly accurate strain sensor for weighing, eliminating the defect that generally, highly accurate electromagnetic type electronic weighing apparatus can not weigh iron, nickel and other magnetic-conductive substances. 1.3 Advanced mode of circuit and good materials and workmanship, hence the high reliability of products, good resistance to interference, long service life, good stability in long service, and adaptability to poor operating environment and continuous long-term operation. 1.4 Functions of automatic calibration, automatic zero tracing, counting, accumulation, tare and pre-tare, unit conversion, preset quantity alarm and more. 1.5 Use of green LED display allowing distinct display and visible readings. 1.6 The electronic counting weighing apparatus allows options of data output interface and printing button, direct connection with printer for data printing and even direct interface with computer for data acquisition and statistics while the computer controls the operation of the electronic counting weighing apparatus via the interface and conducts real-time remote control of the electronic weighing apparatus. 1.7 Rechargeable battery with large capacity which operates continuously for over 36 hours after one charge, and allows the use of the electronic weighing apparatus during charge, which does not affect the behavior in service of the weighing apparatus. 2. Specification and performance indices Depending on the difference of range and resolutions, TJ series electronic counting weighing apparatus is classified into 8 types as shown in the table: Models TJ30K TJ15K TJ6K TJ3K Capacity 30kg 15kg 6kg 3kg Resolution Ratio 1g 0.5g 0.2g 0.1g Calibration graduation 10d 10d 10d 10d Tare Scope 30kg 15kg 6kg 3kg Calibrating Weight 10kg--30kg 5kg--15kg 2kg--6kg 1kg--3kg Min. weighing 20g 10g 4g 2g Min. sampling value 20g 10g 4g 2g Min. unit weight value 0.5g 0.25g 0.1g 0.05g Dimension of Platform Volume Power supply 300*220mm 320(W)*300(D)*110(H) AC 230V±10% 50Hz±1Hz 4W DC 6V/4Ah/20HR rechargeable battery Operating temperature 0--40

Operating humidity 80%R.H Models TJ30KA TJ15KA TJ6KA TJ3KA Capacity 30kg 15kg 6kg 3kg Resolution Ratio 2g 1g 0.5g 0.2g Calibration graduation 10d 10d 10d 10d Tare Scope 30kg 15kg 6kg 3kg Calibrating Weight 10kg--30kg 5kg--15kg 2kg--6kg 1kg 3kg Min. weighing 40g 20g 10g 4g Min. sampling value 40g 20g 10g 4g Min. unit weight value 1g 0.5g 0.25g 0.1g Dimension of Platform Volume Power supply 300*220mm 320(W)*300(D)*110(H) AC 230V±10% 50Hz±1Hz 4W DC 6V/4Ah/20HR rechargeable battery Operating temperature 0--40 Operating humidity 80%R.H 3. Operating method 3.1 Prior to the use of the electronic weighing apparatus that has been just purchased or laid idle for long time, power should be switched on for battery charging and leveling. Tune-on the power switch and the weight display window displays F----1 to F----9 and then 0 after holding for a period of time. Prior to the use, switch on the power for 15 minutes of warming-up. 3.2 Tare 3.2.1 If the weight display deviates from the zero with nothing on the weighing platform, press the TARE key to reset the weight display to zero. 3.2.2 If the weight of the article for which the packaging is to be removed is unknown, place the packaging container on the weighing platform, wait till the weight display is stable before pressing the TARE key. The zero is displayed for the weight. Then place the article to be weighed in the container. Now, the weight window displays the net weight of the article. When the article and its container are removed, the weight display window displays the negative value of the weight of the container. Nevertheless, press the TARE key so that zero is displayed for the weight. 3.2.3 If the weight of the article for which the packaging is to be removed is known (pre-tare), key-in the weight (in g) of the container by means of number keys, then press the TARE key and the weight display window displays the net weight of the article excluding the weight of the container. Remove the article and container, the display window displays the negative value of the weight of the container. Nevertheless, press the TARE key and the Packaging Removal function is canceled and zero is displayed for the weight. 3.3 Calibration When laid idle for long time or just purchased, the electronic weighing apparatus should be subject to calibration. With nothing on the weighing platform, fully warm up the electronic weighing apparatus (for 15 minutes or more),press and hold the CE

key followed by pressing the UNIT W.T. key and the weight display window displays C--XXX. The unit weight display window displays -CAL- and it is in the calibration state (XXX refers to the weight of the calibration required. For instance, 10.000 means that standard weight of 10kg is desired.). Now, simply place the calibration weight on the weighing platform, wait till stability is achieved and the weight display window of the electronic counting weighing apparatus displays the value of the weight. The indicator for the sign kg of stable weight lights, the unit weight display window displays 0 and thus ends the calibration. Now, normal weighing and counting is possible. Or when the unit weight display window displays -CAL-, use the number keys to enter the weight value before calibration by placing sufficient counter poise (the entered value of the weight should be between 1/3 of the range and full range). In the calibration state, press the TARE key and you will come out of the calibration state. When the weight display window displays C----F during the calibration, it means that the zero is unstable. You may press the TARE key for the weight display to reset before the calibration. 3.4 Counting 3.4.1 When the unit weight of the article to be weighed is unknown, place the article to be sampled on the weighing platform, key-in the quantity of the article to be sampled, press the SMPL key and the unit weight display window displays the unit weight of the article to be sampled, the quantity display window displays the quantity of the article to be sampled, thus ends the sampling and it is in the counting state. The larger the quantity of the article to be sampled, the more accurate the calculated unit weight value is. 3.4.2 When the unit weight of the article to be weighed is known, key-in the unit weight value of the article to be weighed, then press the UNIT W.T. key, (If no key is pressed for entering after 5s when numbers are keyed in, the entered number in the unit weight window is deemed as the unit weight value of the article to be weighed.) and thus ends the setting and it is in the counting state. 3.5 Weight Accumulation Press the M+ key to enter the accumulation state and the accumulation indicator lights. The max. number of operations for accumulation is 99 and the max. number to be displayed in the quantity display window is 99999. When weight data display is stable in the weight display window, press the M+ key and the weight display window displays the total weight with the unit weight window displaying the accumulated number of operations and the quantity display window displaying the total weight. After about 5s, you are again in the counting state where the weight display window displays the current weight, the unit weight window displays the unit weight and the quantity display window displays the current quantity. When the weight display window displays zero, press the M+ key and you will view again the accumulated value. Press the MC key and you can clear the accumulated value and are out of the accumulation state. 3.6 Presetting of quantity For counting, the upper limit can be preset. For each of the subsequent counting, alarm is actuated when this limit is exceeded and --H-- flickers in the unit weight display window. Key-in the upper limit of number to be set, press the QTY PST key and thus ends the presetting of the quantity. When you press the QTY PST key without data entered, the quantity display window displays the current value of preset quantity. To clear the upper

limit of preset quantity, press the PST CE key 3.7 Unit conversion Press kg/lb key and the displayed value for weight display window switches between kg (kilogram)and lb (pound)and the indicator for relevant unit lights. The unit conversion i n the weight window does not affect the unit in the unit weight window (which is always in g). 3.8 Data output For the electronic weighing apparatus, standard RS232 data output interface may be used, which may be directly connected to printer. When you are not in the number input state, press. key to print counting data. Also, it may be directly connected to the computer interface. 4. Illustration of keys 0-9 Keys for entering numbers. In the number-entering state, it is a key for decimal. When not in the number-entering state, it is the printing key. CE TARE kg/lb UNIT W.T. SMPL In the number-entering state, it clears the entered numbers. When it is not in the number-entering state, press and hold the CE key followed by pressing the UNIT W.T. key and you are in the calibration mode. In the number-entering state, press the TARE key, packaging pre-removal is provided for the current entered value. When it is not in the number-entering state, pressing the TARE key clears the current weight. In the calibration state, press the TARE key and you are out of the calibration state. The unit in the weight window between kg(or g) and lb. In the number-entering state, press the UNIT W.T. key and the current entry is set as unit weight. In the number-entering state, press the SMPI key and the current entry is set as number of pieces. To set the number of pieces, the displayed data in the weight field must be more than 0. Otherwise the set number of pieces is 0. M+ When the displayed data in the weight field is more than 0, accumulated weight value and MC QTY PST PST CE value for number of pieces are re-established after the unit weight window displays the accumulated number of operations for 5s, and the next accumulation is possible only when the weight returns to 0. Allowing you to get out of the accumulation state and clear the accumulated value. In the number-entering state, press the QTY PST key to set the current entry as the upper limit of the quantity. When it is not in the number-entering state, the quantity window displays the preset quantity and returns after 5s. It clears the quantity preset value. 5. Information displayed on the counting weighing apparatus 5.1The weight field displays F----1 Data exceeds the range of display.

F----2 F----3 F----5 F----L F----H C----F C----H E----- Error in A/D conversion circuit Indicating that keys are available Indicating fault in the sensor Zero of sensor signal is too low Weight exceeds the range Error in calibration. It is not at zero or the data are not stable during the calibration. Error in calibration. Article is placed on the weighing platform or the zero of the sensor is too high during the calibration. Error in storing the calibration data. Service is possible only after re-calibration. 5.2 The unit weight field displays the following information: -CAL- You are in the calibration state X The displayed value is the number of operations for accumulation. The weight field displays the total weight and the number-of-pieces field the total weight. H Meaning the preset quantity is exceeded and an audible warning is given. 5.3The quantity field displays the following information: F---1 The value of the number-of-pieces exceeds the range of display 5.4 Indicators kg(g) lb SAMP UNIT Under-voltage Charging Fully charged Lights when the data are stable. The current weight unit is in kg(g) Lights when the data are stable. The current weight unit is in lb. Lights when the total sampling amount is less than the min. sampling value. Now, add the number of samples till the indicator is out. Set by re-sampling to ensure the accuracy. Lights when the average or set unit weight is less than the min. unit weight value. Now, the counting weighing apparatus remains serviceable but error may happen during the counting. It is recommended that counting weighing apparatus with high distinguishability is used. Lights when the voltage of the battery is insufficient along with audible alarm, charge the battery. Otherwise, the battery operates only 1-2 hours. The counting weighing apparatus operates normally during the charge. Lights when the counting weighing apparatus is in the charging state. It takes about 8 hours for one charge. Slowly lights when the counting weighing apparatus is under charge. When the charging lamp dies out, it indicates that it is fully charged. Now, it allows further charging. 6.Setting of adjustable parameters Start the unit by holding down the QTY PST key till it comes to F----3 and you are in the setting mode. Press the QTY PST key to change the parameters to be set and press the TARE key to change the parameter value. C1---- for setting the sensitivity at 0 1 2 3 4. The higher the value and the poorer the sensitivity, the better the stability is. The factory setting is 2.

C2---- for setting the filtering intensity at 0 1 2 3. The higher the value and the slower the response speed, the better the stability is. The factory setting is 1 or 0. C3---- for setting the baud rate at 1(300)2(600) 3(1200) 4(2400) 5(4800) 6(9600) 7(19200). The factory setting is 6. C4---- for setting the communication number. The TARE key changes the low-value and the kg/lb key changes the high-value. The data are the first data to be received during the communication. The factory setting is 27. Start the unit by pressing the TARE key and all set parameters and calibrated data return the factory state. Before service, perform re-calibration. 7.Serial-port communication It is in operating mode 1 where there is no calibration at the stop bit Bit 1 of 8 data bits. Receiving data: Communication number + command printing 70H( ACSⅡ code p) calibration 71H(ACSⅡ code q) accumulation 72H(ACSⅡ code r) unit conversion 73H (ACSⅡ code s) packaging removal 74H(ACSⅡ code t) Transmitting data: WT 12345.6 kg 15 bits UW 1234.5 g 12 bits QT 12345 Pcs 14 bits To display accumulation: MW 12345.6 kg 15 bits MN 12 12 bits MQ 12345 Pcs 14 bits Example: To set C3---6 C4 27 (factory state), the communication procedure for VB is: MSComm1.Settings = "9600,n,8,1" MSComm1.Output = Chr(&H1B) + Chr(&H70) sending the printing command or MSComm1.Output = Chr(27) + Chr(112) or MSComm1.Output =Chr(27) +"p" Do DoEvents Loop Unti MSComm1.InBufferCount = 41 a = MSComm1.Input Print a 8. Structure (1) TARE (2) kg/lb

(3)CE (4)Numbers (5)Windows (6)Level indicate (7) Level adjust (8)Power switch (9) Power socket (10) Scale pan (11) Data output 9. Installation 10. Cautions for operation 10.1 The electronic scale is a precision instrument which is to be gently handled during the weighing of articles. Avoid impact and heavy load on top (never exceed its max. weighing limit). 10.2 Place the scale on a firm table for operation, and make sure that the operating environment of the balance is free of great vibration and interference of power supply, and of corrosive gases and liquids. 10.3 Ensure the warm-up time after power-on. 10.4 When the under-voltage indicator lights, it means the battery needs charging. No charge at the moment may lead to inaccuracy or instability in weighing. 11. The Warranty 11.1 We provide warranty within the scope of the statutory regulations, however we are entitled to either repair or replace the defective product. 11.2 Within two years after purchase, the producer offers a free repair service to all G & G Product, which become defect under proper usage without human damage. The customers are kindly requested to mail the product as well as its original proof of purchase to the producer. Notification of defects and complaints shall be made in writing after the delivery has been received at the point of delivery. For legitimate notifications of defects and warranty cases we are entitled at our own discretion to either replace the delivery at the prices which were valid at the time of the complaint or to rectify the defect.

11.3 We have the right to try to replace the defective product at least twice. There will be a slight charge for the repair service of those products, by which the warranty period expires. Defective goods may only be returned to us by the customer with our express approval and at the customer's cost and risk. The notification of the defect does not release the customer from his obligation to pay. 11.4 Please attach the address, postal code and telephone number of the buyer, so that we can mail the repaired product back in time. 12. Trash Meaning of crossed out wheeled dustbin: Do not dispose of electrical appliances as unsorted municipal waste, use separate collection facilities. Contact you local government for information regarding the collection systems available. If electrical appliances are disposed of in landfills or dumps, hazardous substances can leak into the groundwater and get into the food chain, damaging your health and well-being. When replacing old appliances with new ones, the retailer is legally obligated to take back your old appliance for disposals at least free of charge. 13. Contact Address: G&G GmbH Koenigstr.34 41460 Neuss Germany Tel: +49-2131-609-7271 Fax: +49-2131-609-7273 Email: info@gandg.de Website: www.gandg.de