Request for Quotations/Proposal

Similar documents
TENDER # SUPPLY OF MOBILE C-ARM RADIOGRAPHY/FLUOROCOPY SYSTEM FOR WESTERN HEALTH

NOTICE INVITING TENDER

Department of Agriculture PROJECT MANAGEMENT UNIT (AIDIP) (Directorate of Horticulture) 2 nd floor "Pant Bhawan" Bailey Road, Patna

SPECIFICATIONS FOR TENDER # SUPPLY OF PROFESSIONAL IDENTIFICATION NAME TAGS FOR WESTERN HEALTH

SPECIFICATIONS FOR TENDER # SUPPLY OF BLOOD GAS ANALYZER FOR WESTERN HEALTH

Department of Radio Diagnosis

SPECIFICATIONS FOR TENDER # SUPPLY OF MCAFEE ANTIVIRUS RENEWAL AND LICENSES FOR WESTERN HEALTH

Invitation for Bid (IFB)

CARESTREAM DIRECTVIEW Max CR System

Vita 25(LE) Vita 45 Vita 65 (XE)

Request For Proposal (RFP) For Media Planning and Booking Television and Radio Air Time For Public Service Announcements (PSAs) on Tuberculosis (TB)

Streamline workflow for higher productivity and patient throughput. DirectView. Classic CR System

ATOMSCOPE HF200A. Operating Instruction Manual. Ver. 3.1

RFP IP-Based Surveillance Camera System

GE Healthcare. Brivo DR-F. Simply digital.

Procurement of Digital Podium (For Contracts value estimated to cost less than Rs. 25 lakhs)

PREQUALIFICATION TENDER FOR SUPPLY OF GOODS, WORKS & SERVICES DESCRIPTION OF THE TENDER

INVITATION FOR BID Copier ST. CHARLES COUNTY AMBULANCE DISTRICT (Herein referred to as District ) 4169 OLD MILL PARKWAY ST.

NOTICE INVITING TENDER

DIVISION OF GENETICS ICAR-INDIAN AGRICULTURAL RESEARCH INSTITUTE PUSA, NEW DELHI

NIT LIMITED NO. NC PH LTN-78 BLUETOOTH SPEAKER & DESKTOP COMPUTER

siemens.com/luminos-select Luminos Select don t compromise, be select.

TENDER FOR INSPECTION AND RADIOGRAPHIC TESTING OF WELDING JOINTS

MNL, SEL DIMENSIONS, SITE AW PLANNING GUIDE ARTWORK SIZE A SHEET 1 OF 1

NORTHEAST TEXAS PUBLIC HEALTH DISTRICT (NET HEALTH) REQUEST FOR PROPOSALS (RFP)

One One Ton Regular Cab 4x4 2 Door Truck Transportation, Infrastructure & Energy

STOCKHOLDING DOCUMENT MANAGEMENT SERVICES LIMITED FORMERLY KNOWN AS SHCIL PROJECTS LIMITED

(To be published in the Gazette of India, Part I, Section 1) RESOLUTION. Tariff based Competitive-bidding Guidelines for Transmission Service

REQUEST FOR QUOTATION (RFQ) Issue Date: December 12, 2017

Elmhurst Community Unit School District 205 Invitation to Bid

A SUPPLIER'S GUIDE TO GOODS PROCUREMENT IN THE GOVERNMENT OF PRINCE EDWARD ISLAND. Procurement Services Office of the Comptroller

TENDER FOR IT TRADE.

Rental Property Registration and Crime Free Housing Software. Request for Proposal

Department of Education, Early Learning and Culture Digital Talking Book Players

Indian Institute of Public Administration I.P. Estate, Ring Road, New Delhi

Request for Quote (RFQ) For. Brown County

Date: Telephone: Company Name: Address: City: State: Zip Code: Contact Person: Title: Authorized Signature:

eurocolumbus SERIES alien e alien I 3030 cardio

Provision of the Printers & Scanners to PSDF

Total Support for SAP HANA Appliances

siemens.com/healthcare Multix Select DR First time. First choice.

Supply of Office Furniture, Fittings, Furnishings & other Equipments.

Master Services Attachment for ServiceElite

ALL-IN-ONE CLOUD BASED CITIZEN SERVICE PLATFORM FOR PERMITTING, LICENSING, INSPECTIONS, AND ENFORCEMENT

Oki (High Pressure Boiler Plant)

TNT X-Ray Test Tools

Biogas analysers. Product overview. MRU over 30 years of innovative gas analysis! Biogas, landfill gas, biomethane, coal mine gas, CHP exhaust gas

REQUEST FOR QUOTATION (RFQ # ) FOUR PART FORM ITD 0502 Application For Report Of Sale And Application For Certificate Of Title

Design, Fabrication, Supply, Installation & Commissioning of Scissor Lift for ITMT Building at. CREST, Hosakote. Indian Institute of Astrophysics

PRE-QUALIFICATION FOR PROVISION OF TEAM BUILDING CONSULTANCY SERVICES TENDER NO. KPPF /HRA-B/1-A/22/17-18

Request for Quotation (RFQ) for IT Support Services for India Brand Equity Foundation (IBEF)

SPECIFICATIONS FOR TENDER # SUPPLY OF ONE MULTI-SLICE CT SCANNER WESTERN MEMORIAL REGIONAL HOSPITAL AND SIR THOMAS RODDICK HOSPITAL

Quick Guide. CSN950 MultiHead Dimensioner _v2d_MAN_QG_CSN950MH_EN

CITY OF MOBILE. Purchasing Department and Package Delivery: Government Plaza 4th Floor, Room S Government St Mobile, Alabama 36644

NIT LIMITED NO. NC-311-SBS-P-CMB PERISTALTIC PUMP

City of Pigeon Forge Fire Department Tony L. Watson, Chief

TENDER NOTICE. Printers/Fax 122* 100* 28* 18* 75* 1 *Likely to increase or decrease

Open Tender Enquiry. Exhaust Gas Analyzer (for Diesel & Gasoline version)

PRE-QUALIFICATION OF SUPPLIERS PROVISION OF FUMIGATION, SANITATION AND SANITAL DISPOSAL SERVICES TENDER NO.KPLC1/5BC/9A/PT-ST/113/2010

SECTION I: ADVERTISEMENT/ INVITATION TO TENDER... 3 DISPOSAL NOTICE... 3 CONDITION OF SALE... 3 TENDERING INSTRUCTIONS... 3

NIT LIMITED NO. NC-449-SBS-P-CMB LABWARE CLEANER/DISINFECTOR WITH HOT AIR DRYER

Tel: +91-(5942) /233727/233734/233735/ FAX: +91-(5942) , Gram: astronomy Website: Tender Document FOR

Worth County School District

TENDER DOCUMENT FOR SUPPLY AND INSTALLATION AIR CONDITIONER

teamplay data sheet valid from March 2017 teamplay data sheet siemens.com/teamplay

SPECIFICATIONS FOR TENDER # SUPPLY, INSTALLATION & COMMISSIONING OF ONE DIAGNOSTIC SPECT CT NUCLEAR MEDICINE GAMMA CAMERA FOR WESTERN HEALTH

The. universal 3D Inspection reference. advanced 3D measurement software & solutions

RFP for Diagnostic Connectivity, Afghanistan

Windows Vista Energy Conservation

NIT LIMITED NO. NC MATH VIDEO CAMERA

E-TENDER NOTICE FOR IT SERVICES

Request for Quotations

NIT LIMITED NO. NC-118-PLAN-SPS PULSE PICKER FOR ULTRAFAST FIBER LASER

Compact flue gas analyser: testo 300 M-I

VJ TECHNOLOGIES TECHNOLOGY FOR GLOBAL MARKETS

PLEASE QUOTE PRICES FOB DESTINATION TO BE DELIVERED TO ITEM DESCRIPTION/SPECIAL INSTRUCTIONS QUANTITY UNIT PRICE EXTENDED PRICE

X-RAY FOOD INSPECTION SYSTEMS DYMOND X-RAY INSPECTION SYSTEM

Sheffield City School District

DoseWise Portal. Take control of dose management. across your organization

INVITATION FOR QUOTATION. No:-TEQIP-II/2017/iiit/Shopping/8/3857

Ref: Tender Notice No. IIT/VGSoM/ Sever/17-18/10 Date:

INTERNATIONAL FOUNDATION FOR ELECTORAL SYSTEMS (IFES) RFQ/11/055

CRISP Azure Migration Consulting Services. All responses due no later than Friday, July 21 st, at 5pm EST

Take control of dose management

COMMUNITY COLLEGE DISTRICT OF JEFFERSON COUNTY, MISSOURI REQUEST FOR PROPOSAL: WINDOW, DOOR, AND FRAME REPLACEMENT

2017 ACR Computed Tomography Quality Control Manual FAQS

INVITATION FOR BID Bid #872 - Printing Services

QUOTATION ENQUIRY. Frequency Range Output AC Voltage Regulation (Batt. Mode) ±10% Frequency Range (Batt. Mode)

TEACHERS SERVICE COMMISSION TENDER DOCUMENT FOR DISPOSAL OF LPG CYLINDERS WITH GAS (TSC EMPLOYEES ONLY). CLOSING DATE: FRIDAY, 1 ST APRIL 2016.

Emergency ITB document

Request for Quotation (RFQ) for. Conducting audit of Online Campaign for Indian Plantations for IBEF. Request for Quotation [RFQ]

NOTICE INVITING TENDER OF RATE CONTRACT FOR SUPPLY OF RFID TAG FROM APPROVED VENDOR TENDER NO. REIL/MM/16-17/E-PROC./113 DATED

Procurement of Goods, Works and Non- Consultant Services

Get Your Hands on MMP

Dear Sir / Madam, Please quote your best price for the following: S.No DESCRIPTION. 1 Web based Video Conferencing Service 1 License

Request For Proposal Of Printing and Design Services. Marketing Department

DATE: 05/03/2018. REQUEST FOR PROPOSAL: No. RFP/2018/UNHCR/PSP/MENA/001 FOR THE ESTABLISHMENT OF A FRAME AGREEMENT FOR THE PROVISION OF

HOBBS MUNICIPAL SCHOOLS P.O. Box 1030 Hobbs, New Mexico INVITATION TO BID. HMS Nutritional Services Milk

Transcription:

Request for Quotations/Proposal Supply, Delivery and Set-Up of Four Units of Digital Mobile X-Ray Machines under URC- USAID s Regional Health Integration to Enhance Services in East Central (RHITES-EC) Activity Project. University Research Co., LLC (URC), a US-based company working in Uganda since 2005, is implementing the USAID s Regional Health Integration to Enhance Services in East Central Uganda Activity (USAID s RHITES EC). The purpose of this project is to support the implementation of a comprehensive, integrated investment in regional health, HIV/AIDS and nutrition assistance. This fiveyear activity targets 11 districts in East-Central Uganda with an estimated total population of 4.1 million through a district-based integrated package of quality health, HIV/AIDS, nutrition and child development services. USAID s RHITES-EC provides technical support to enable implementation of comprehensive integrated health services and systems strengthening within 11 districts. The project s overall goal is to support the EC districts to increase access to and availability of high quality health care services. One of the obstacles to achieving improved TB Case Notification Rate (CNR) is limited clinical diagnosis of TB due to limited access to supportive screening and diagnostic services like X-ray services. URC is hereby requesting for quotations (RFQ) for Supply, Delivery and Set-Up of Four Units of Digital Mobile X-Ray Machine as per details and specifications indicated here-in. Important To Note: The four units are to be delivered at Iganga Hospital, Kamuli Hospital, Bugiri Hospital and Busia HC IV respectively. The vendor will incur the cost of delivery. These are digital mobile X-Ray units but the vendor must fully set up/assemble the unit at each of the facilities to be ready for use and deployment. The selected vendor will provide comprehensive training to technical staff and support services at the health facilities till familiarity with the system is attained. The vendor shall incur all costs related to offering this training. EVALUATION CRITERIA: Bids shall be evaluated based on the below evaluation criteria. Technical compliance Selection Method shall be used. 1. To establish vendor eligibility, you must submit a current (2018) trading license or an equivalent document to verify your eligibility. 2. Vendor must have experience in provision of supplies of similar nature and value (Attach proof of LPOs, Award letters or Contract Documents) 3. Vendor MUST indicate their delivery period. 4. USAID s RHITES-EC shall evaluate the technical aspects of the bid submitted to confirm that all requirements specified in the statement of requirements below have been met without any deviation or reservation. The Statement of Requirements details the minimum technical requirements. Responsiveness is determined by comparison of the specification offered Office (+256)-(0)414-287-499 (+256)-(0)312-202-045/6 Fax (+256)-(0)414-287-614 www.urc-chs.com urcuganda@urc-chs.com

to the specification required. You should also indicate whether your specifications are compliant or non-compliant to our minimum required specifications. Vendor should attach relevant technical literature. 5. The bid with the lowest evaluated price, from among those which are eligible, technically compliant and substantially responsive shall be the best evaluated bid. INSPECTION AND VERIFICATION The winning vendor shall be required to organize for the inspection, verification and approval of the unit prior to full delivery. The unit must be verified for specification conformity by the Project Technical Staff. TERMS AND CONDITIONS: 1. Payment Terms: Within 30 days of satisfactory provision of supplies and skills transfer through user training. 2. Quotation validity: 120 days 3. All costs must be quoted in Ugandan Shillings (UGX). CLARIFICATION AND BID SUBMISSION DETAILS: All request for clarification must be sent to RHITES-EC-Procurement@urc-chs.com Address all quotations to USAID s Regional Health Integration to Enhance Services in East Central (RHITES-EC) Activity. Sealed tenders clearly marked; Tender for Supply, Delivery and Set-up of the Digital Mobile X-Ray Units should be addressed and delivered to Chief of Party URC USAID s RHITES-EC Project, Jinja Uganda, Plot 1 East Victoria Close, Mpumudde Division, P.O Box 5053 to reach us by Thursday 29 th March 2018 at 3:00 pm. The sealed envelope shall contain One (1) original and two (2) copies. The original should be clearly marked as original. Late Bids shall be rejected and no electronic bids are allowed. Office (+256)-(0)414-287-499 (+256)-(0)312-202-045/6 Fax (+256)-(0)414-287-614 www.urc-chs.com urcuganda@urc-chs.com

STATEMENT OF REQUIREMENTS (Technical Specifications Details) Item Description and Technical Specifications Technical Specifications offered by the Vendor Comply/Not Comply Digital Mobile X-Ray Unit with fabricated container, generator, printer, and detector. The unit should be completely integrated with the following specifications X-Ray stand on wheels 2-stage hand trigger switch Portable X-Ray with 15-32kw or more GENERATOR Should be microprocessor controlled Constant potential kw ratings: 50KW Generator or more. KVP range in 1 KVP steps accuracy: 40kV to 125-133 or better +5% or better High frequency ripple: 40-60 khz or better ma stations: 25 to 300mA-400mA or more.

Minimum exposure time should be 1ms or less mas range: 0.32 300 ma or more. Should have a digital display of Kvp and mas Should have automatic exposure control (AEC) device X-RAY TUBE Should be a Rotating anode compatible with the generator Two focal spots: Small focus 0.6-0.7mm Large focus 1.2-1.3mm approximately Anode heat storage capacity: 300,000 HU minimum Should have a thermal safety switch for overload protection Target angle: 12-14 Power cord : min. 6 metres Exposure cord : min. 3-4metres

Line regulation: Single-phase line automatic regulation Automatic output power adaption to power line conditions Line power requirements: 110 VAC 240 VAC / 220 VAC 240 VAC DETECTOR Detection method, a-si TFT, Scintillator CsI Dimensions (W x H x D) 384.0-460.2 mm x 383.7 460.0mm x 15.0-15.9 mm (14-18.1 x 15.1-17.0 x 0.6 ), Size: 35 cm x 43 cm (14 x 17 ) cassette size Weight 3-3.3 kg (6.5-7.3lbs) or Less Pixel matrix 1,994-2,800 x 2,430 3,200 pixels or better Pixel pitch 125-175 µm or better A/D conversion 16 bits or better Spatial resolution 3.4-3.6lp/mm or better Should be DICOM 3.0 compatible

Environmental Conditions Operating Temperature: 10 C to 30-40 C (50 F to 86 F) Humidity: 35 % to 80 % (non-condensing) Atmospheric pressure: 70 kpa to 106 kpa Battery Type: Energy Storage via Capacitor /Battery Operated Charging time empty to full: min. 30-90minutes Number of exposable images: 200 images / 6 hours or better Duration in standby status: approx. 16 hours Power supply will be ensured through an inverter with battery rack that ensures minimum 4 hours uninterrupted autonomous operation. CAPTURE CONSOLE Mobile Computer as Acquisition Station Optical Driver for Patient CD

SOFTWARE Software controls detector and generator / auto detector control. Software does image acquisition and optimization. Organ selection menu with organ buttons leading to the right image processing settings as well as to the right generator settings. Dose area product measurement device with printer. X-Ray journal with kv, mas and dose area product. Combination of a DICOM Modality (Send/Store, Work list for communication with a central RIS/PACS) and a local PACS. Sending to the central PACS via DICOM Send/store should be either activated permanently or executed by choice of the operator. The local PACS has to be integrated into the DR Panel acquisition software. For viewing the images during the acquisition process and for viewing older images, the same viewer software has to be used. The Viewer has to be fully functional for medical image viewing and diagnosis. Patient CD burning capability. Wireless communication between Flat Panel Detector and computer. All kv- and ma-settings of the software are shown at the control panel of the X-Ray. The detector is communicating directly with the mobile computer.

Compatible with DR Panel. CHEST STAND DICOM software for DICOM printing and DICOM send for CAD4TB compatibility and for uploading/cloud storage Laser X-ray film printer size 14 x 17 to be included with hardware devices Suitcase for Detector and mobile Computer Detector Size: 14x17 based CsI scintillator All components are installed, the suitcase is ready for use with detector Direct Radiography detector (DR) with excellent image quality even on low dosage. Fabricated Container for X-Ray Machine with generator as per the requirement of X-Ray Machine (The container will be placed on a Truck) The container should be provided with; Rust protection through a 2-component paint Floors finished with wear resistant coating The container should be strong enough to hold all the parts of the X-Ray Unit for easy and safe transportation on the truck.

SOFTWARE SUPPORT FOR ALL X-RAY MACHINES Capture console acquisition station with 21" monitor All required software for printing and computer aided diagnosis for TB compatibility and for uploading/cloud storage Computer Aided detection for TB software The supply of software should include the required licenses for each unit and must include off-line analysis and manipulation of acquired data. The software shall help to decrease false negative readings and enabling the clinical staff with minimal expertise in radiography to analyze TB suspect images The software shall provide markers and color overlay around suspect regions on digital radiographic images and these must be consistent with TB. The software shall include reference images for teaching and learning. The software shall be installed on all the systems and must be supplied with all the licenses. Laptop/ computer device for results 17 screen Operating system; windows 8 or higher Processor capacity; 8 GB Internal (RAM) memory DVD drive Printer Dry laser colour printer, minimum 300 DPI suitable for printing of digital images.

All bids to include Lead Aprons for radiation protection, 4 sizes (S, M, L, XL) Full 0.5mm Pb Protection or equivalent, max weight of each Apron 5KG Small = 2, Medium=4 Large= 6, Extra Large=4 Lead Apron Unisex SERVICE Minimum 18 months warranty Service possibilities within East Africa Remote control diagnostic access; enabling diagnosis and service from a distance

Please note: Issuance of this solicitation does not in any way obligate URC-CHS or USAID to award a contract, nor does it commit URC-CHS or USAID for cost incurred in the preparation and submission of a proposal. Office (+256)-(0)414-287-499 (+256)-(0)312-202-045/6 Fax (+256)-(0)414-287-614 www.urc-chs.com urcuganda@urc-chs.com