Analysis of Intact Monoclonal Antibodies Using an M3 MicroLC with the TripleTOF 6600

Similar documents
Improving Sensitivity in Bioanalysis using Trap-and-Elute MicroLC-MS

Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF

PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC

IgG Purity/Heterogeneity and SDS-MW Assays with High- Speed Separation Method and High Throughput Tray Setup

Application Note # ET-20 BioPharma Compass: A fully Automated Solution for Characterization and QC of Intact and Digested Proteins

Thermo Scientific Solutions for Intact-Protein Analysis. Better, Faster Decisions for. Biotherapeutic Development

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Proteins. Patrick Boyce Biopharmaceutical Marketing Manager Waters Corporation 1

Antibody Analysis by ESI-TOF LC/MS

Rapid Peptide Catabolite ID using the SCIEX Routine Biotransform Solution

Utilizing novel technology for the analysis of therapeutic antibodies and host cell protein contamination

MultiQuant Software Version 3.0

MasterView Software. Turn data into answers reliable data processing made easy ONE TOUCH PRODUCTIVITY

Cell Clone Selection Using the Agilent Bio-Monolith Protein A Column and LC/MS

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

Detecting Challenging Post Translational Modifications (PTMs) using CESI-MS

Enabling Systems Biology Driven Proteome Wide Quantitation of Mycobacterium Tuberculosis

N-Glycan Profiling Analysis of a Monoclonal Antibody Using UHPLC/FLD/Q-TOF

A Complete Workflow Solution for Intact Monoclonal Antibody Characterization Using a New High-Performance Benchtop Quadrupole- Orbitrap LC-MS/MS

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

A Comprehensive Approach for Monoclonal Antibody N-linked Glycan Analysis from Sample Preparation to Data Analysis

Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles

Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

MassPREP On-Line Desalting Cartridge

Investigating Biological Variation of Liver Enzymes in Human Hepatocytes

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column

Analysis of Monoclonal Antibodies and Their Fragments by Size-Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer

Bioanalysis of Intact Therapeutic Monoclonal Antibodies from Non-Human Primates Using MSIA D.A.R.T. S Technology

2012 Waters Corporation 1

Rapid Soft Spot Analysis using the SCIEX Routine Biotransform Solution

Generating Automated and Efficient LC/MS Peptide Mapping Results with the Biopharmaceutical Platform Solution with UNIFI

Application Note. Kwasi Antwi, Amanda Ribar, Urban A. Kiernan, and Eric E. Niederkofler Thermo Fisher Scientific, Tempe, Arizona

A Sub-picogram Quantification Method for Desmopressin in Plasma using the SCIEX Triple Quad 6500 System

High Levels of Sensitivity and Robustness for Complex and Demanding Matrices SCIEX TRIPLE QUAD 5500 SYSTEM

Analysis of Illegal Dyes in Food Matrices Using Automated Online Sample Preparation with Liquid Chromatography-Mass Spectrometry

Trastuzumab Glycan Batch-to-Batch Profiling using a UPLC/FLR/Mass Spectrometry Platform

CQA Assessment by Peptide Mapping Using LC/MS with an AdvanceBio Peptide Plus Column

mabs and ADCs analysis by RP

Accelerate mab Characterization Using Automated Sample Prep

Thermo Scientific Peptide Mapping Workflows. Upgrade Your Maps. Fast, confident and more reliable peptide mapping.

A Unique LC-MS Assay for Host Cell Proteins(HCPs) ) in Biologics

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Pharmacokinetic Analysis of the mab Adalimumab by ELISA and Hybrid LBA/LC/MS: A Comparison Study

Application Note LCMS-87 Automated Acquisition and Analysis of Data for Monitoring Protein Conjugation by LC-MS Using BioPharma Compass

SMART Digest Kit User Manual. Version 2

Development of Analysis Methods for Therapeutic Monoclonal Antibodies Using Innovative Superficially Porous Particle Biocolumns

High-Throughput Peptide Mapping with the Vanquish UHPLC System and the Q Exactive HF Mass Spectrometer

New Approaches to Quantitative Proteomics Analysis

AB SCIEX TripleTOF 5600 SYSTEM. High-resolution quant and qual

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

A highly sensitive and robust 150 µm column to enable high-throughput proteomics

Asish Chakraborty, St John Skilton, Weibin Chen and John C. Gebler Waters Corporation, Milford, MA, U.S. INTRODUCTION METHODS. Sample preparation

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

Xevo G2-S QTof and TransOmics: A Multi-Omics System for the Differential LC/MS Analysis of Proteins, Metabolites, and Lipids

Tony Mire-Sluis Vice President, Corporate, Product and Device Quality Amgen Inc

Monoclonal Antibody Characterization on Q Exactive and Oribtrap Elite. Yi Zhang, Ph.D Senior Proteomic Marketing Specialist Oct.

rapiflex Innovation with Integrity Designed for Molecules that Matter. MALDI TOF/TOF

A Generic Kit-Based Approach for Quantifying Monoclonal Antibody Drugs Through Direct Digestion of Discovery Study Samples

Featuring Analyst software under Windows NT for enhanced performance and ease of use. API 150EX. LC/MS System

On-line SPE-LC/MS/MS to Detect Organonitrogen and Triazine Pesticides at 10ng/L in Drinking Water

Made for Routine Food, Environment, and Forensic Testing. The X500R QTOF System

Rapid Peptide Mapping via Automated Integration of On-line Digestion, Separation and Mass Spectrometry for the Analysis of Therapeutic Proteins

Rapidly Characterize Antibody- Drug Conjugates and Derive Drug-to-Antibody Ratios Using LC/MS

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

Exploring Extra Sensitivity Using ionkey/ms with the Xevo G2-XS Q-Tof HRMS for Small Molecule Pharmaceutical Analysis in Human Plasma

Characterization of Glycosylation in the Fc Region of Therapeutic Recombinant Monoclonal Antibody

mab Glycopeptide Profiling with V-Tag Adding reliable glycoprofiling to your peptide mapping workflow with ease and simplicity

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

Method Development Considerations for Reversed-Phase Protein Separations

Faster, easier, flexible proteomics solutions

Advanced QA/QC characterization MS in QC : Multi Attribute Method

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

The LC-MS/MS Workhorse. SCIEX 4500 Series Mass Spectrometers

Advantages of Ion Mobility QTOF for Characterization of BioPharma Molecules

Fast and High Resolution Analysis of Intact and Reduced Therapeutic Monoclonal Antibodies (mabs)

FPO. BioPharma Compass 2.0. Innovation with Integrity. Accelerate Biopharmaceutical Analysis. Software

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

CaptiveSpray nanobooster

PAVING THE WAY FOR ASSESSING IN VIVO DYNAMICS OF MULTIPLE QUALITY ATTRIBUTES FOR PROTEIN THERAPEUTICS

Analysis of biomolecules by SEC and Ion-Exchange UPLC

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

Improved Peptide Maps Using Core-Shell Media: Explaining Better Biopharmaceutical Applications With Kinetex C18 Columns

Separation of Recombinant Human Erythropoietin (rhepo) using the European Pharmacopoeia Method on the PA 800 plus Pharmaceutical Analysis System

Peptide Mapping of Glycoprotein Erythropoietin by HILIC LC/MS and RP-LC/MS

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

A Robust and Generic Method for Quantification of Monoclonal Antibody Therapeutics in Biological Matrices

ichemistry SOLUTIONS Integrated chemistries to boost mass spectrometry workflows

Rapid Extraction of Therapeutic Oligonucleotides from Primary Tissues for LC/ MS Analysis Using Clarity OTX, an Oligonucleotide Extraction Cartridge

Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass

Conjugated protein biotherapeutics

SMART Digest compared to classic in-solution digestion of rituximab for in-depth peptide mapping characterization

Abstract. Authors. BioPharma Compass. Antibody drug conjugate (ADC) mabs Native MS Subunits Intact protein analysis Drug antibody ratio (DAR)

A High-Resolution, Accurate-Mass Approach for Ultra-High Throughput Screening Plasma Protein Binding

High Resolution Accurate Mass Peptide Quantitation on Thermo Scientific Q Exactive Mass Spectrometers. The world leader in serving science

LabChip GXII: Antibody Analysis

Size Exclusion Chromatography/ Mass Spectrometry Analysis of Antibody Drug Conjugates Using the Agilent 1260 Infinity II Bio-Inert LC

Transcription:

Procedures Analysis of Intact Monoclonal Antibodies Using an M3 MicroLC with the TripleTOF 66 Robust and Sensitive Workflow for Qualitative and Quantitative Analysis of Biotherapeutic IgGs Khatereh Motamedchaboki, Remco van Soest, Daniel Warren and Jenny Albanese SCIEX, USA Overview Who Should Read This: Senior Scientists, Lab Directors Focus: Analysis of intact monoclonal antibodies (mabs) using microflow liquid chromatography-mass spectrometry (LC-MS) with on-column desalting. Goal: Determine accuracy and sensitivity of the microflow LC-MS method for characterization and quantitation of intact monoclonal antibodies. Problem: Traditional LC-MS methods for analyzing intact monoclonal antibodies and similar biotherapeutics require large amounts of sample and frequently deliver insufficient sensitivity. Microflow LC-MS, in spite of its inherent sensitivity advantage, has not been used extensively because the necessary off-line desalting and sample clean up result in possible sample loss and long sample preparation cycles. Results: The SCIEX M3 MicroLC coupled with a SCIEX TripleTOF 66 mass spectrometer provided accurate and highly sensitive characterization and quantitation of intact monoclonal antibodies. On-column desalting simplified and minimized sample preparation. Key Challenges: Quantitation of intact mabs over a wide subnanogramto-nanogram dynamic range often requires greater sensitivity than can be achieve with LC-MS at standard LC flow rates. Microflow LC-MS can achieve the necessary sensitivity, but requires extensive desalting and sample cleanup. Traditional off-line (on-membrane) desalting methods are time consuming and result in sample loss. Key Features: Single-run characterization and quantitation of intact mabs and similar biologics Robust quantitation over a linear dynamic range of 3 orders of magnitude (.1-1 ng on column) Quantitation at 5x lower concentration than can be achieved with standard-flow LC-MS High throughput intact mass analysis of mabs (~14 samples/day) Fewer sample preparation steps, eliminating off-line desalting (~3 min process) and removal of carbohydrate (overnight digestion process) Reduced solvent, desalting, enzyme and operation costs Fast-Growing Field: Protein biotherapeutics such as immunoglobulin G (IgG)-derived monoclonal antibodies occupy a rapidly increasing share of the pharmaceutical industry. Understanding the primary structure, heterogeneity, and posttranslational modifications of these biologics are essential to understanding function, to developing novel therapeutics and to ensuring product safety and quality. Modifications (Figure 1) can directly affect protein activity, immunogenicity and stability. LC-MS analysis has become an essential tool for the identification, characterization and quantification of intact mabs) and similar high-molecular-weight proteins. 1, 2 In this note, we describe a robust and sensitive workflow using a M3 MicroLC coupled to a TripleTOF 66 mass spectrometer for characterization and quantitation of intact mabs. The method p1

takes advantage of on-column desalting to decrease sample preparation time and increase throughput. Three different IgG1 molecules were used for method evaluation. Microflow Liquid Chromatography: The on-column desalting and separation of all three mabs was performed on an M3 MicroLC (SCIEX) using Waters ACQUITY UPLC M-Class Protein BEH C4 column (3Å, 1.7 μm, 3 μm X 5 mm) at a 15 µl/min flowrate. All tests were performed in direct inject mode using the LC gradient in Table 1. The total LC run time was 1 minutes, of which the first 3 minutes was on-column desalting followed by 7 minutes of linear gradient separation from 2% B to 8% and back to 5% for maximum binding. The column temperature was maintained at 6 C. Five microliters of each serial dilution was loaded on column for each of the 3-4 replicate analyses. Mobile phase A was 1% water with.1% formic acid and.1% trifluoroacetic acid (TFA, Thermo Scientific P/N 2894). Mobile phase B was 1% acetonitrile with.1% formic acid and.1% TFA. Figure 1. IgG1 Structure Heterogeneity. Therapeutic antibodies are generally complex, heterogeneous and subject to a variety of enzymatic or chemical modifications during expression, purification and long-term storage. These modifications are listed in blue. Results for all three mabs showed improved sensitivity with a high level of accuracy and a good coefficient of variation (CV) for intact antibody quantitation across 3 orders of magnitude linear dynamic range (.1-1 ng), and successful characterization of the glyco-isoforms at low nanogram levels down to 2.5 ng. The sensitivity gained with the M3 MicroLC, along with the high resolution and mass accuracy of the TripleTOF 66, was essential for this comprehensive analysis. Experimental Design Sample Preparation: Intact mab Mass Check Standard (Waters, P/N 1866552), trastuzumab and adalimumab (Myoderm, NDC #5242-134-68, NDC# 74-3799-2 ABB) were serially diluted with 1 pm/µl of BSA (bovine serum albumin) in.1% FA (formic acid, Thermo Scientific P/N 2895) from 2 µg/ml down to.2 µg/ml. All antibodies were desalted by our on-column desalting strategy for maximum recovery. The microflow LC utilizing on-column desalting method was compared to off-line on-membrane desalting of adalimumab on 1K cut off Amicon Ultra Centrifugal filters (Sigma-Aldrich, P/N Z67718-96EA). The adalimumab was desalted on membrane six times with 1% acetonitrile in.1% FA and then recovered with.1% FA. Serial dilutions of.2-1 µg/ml concentration were made for desalted adalimumab with 1 pm/µl of BSA in.1% FA. p2 Time(min) % Solvent B 2 3 2 5 8 6 8 6.5 2 7 2 7.5 5 1 5 Table 1. Microflow LC Gradient Used for Intact Antibody Analysis. Mass Spectrometry: MS analyses were performed using a TripleTOF 66 equipped with a DuoSpray Source and 25 μm I.D. electrode (SCIEX). The MS method was built in two periods: 3 min with ion spray voltage floating (ISVF) set to during on-column desalting to avoid spraying salt into MS followed by 7 min with ISVF set to 55 for sample analysis. The detailed MS and source parameters are shown in Table 2. A minimum of three replicate injections were performed for each serial dilution. Data Processing: Analysis of intact mabs, including spectral deconvolution, mass reconstruction and analysis of glycans and other post-translational modifications (PTMs), was performed using BioPharmaView (version 1.4.917) software. mabs quantitation analysis, including calibration curve, CV and accuracy, was performed using MultiQuant (version 3..2) software. The reproducibility of data was assessed visually with PeakView (version 2.2..11391) software as well as quantitatively using MultiQuant software. The method was optimized by analyzing the data using a combination of BioPharmaView and PeakView.

MS Parameters Values Electrode ID 25 um Curtain Gas 25 Collision Energy 25 IonSpray Voltage 55 Temperature ( C) 3 Ion Source Gas 1 35 Ion Source Gas 2 35 Declustering Potential 24 Polarity Positive Mass Range 1-5 Accumulation Time (sec) 1. Time Bins to Sum 8 Scan Type TOF MS Intact Protein Mode On Advantages of On-Column Desalting Commercial mabs require buffering salt for stabilization in solution. However, to gain sensitivity in LCMS analysis of intact mabs, it is critically important to remove the salt before LC-MS analysis. This is typically done either by off-line desalting or a trap-elute workflow. Instead, we used on-column desalting to minimize sample preparation time and avoid sample loss. Ion spray voltage floating (ISVF) was set to during the on-column desalting period to avoid spraying salt into the MS. To confirm the viability of on-column desalting, we compared this approach against on-membrane desalted adalimumab (5 ng load on column). We observed less sample loss and greater sensitivity using the on-column desalting method (Figure 3). Table 2. MS and DuoSpray Source Parameters. High Retention Time and Sensitivity Reproducibility High reproducibility of retention time (RT) and sensitivity were observed. The M3 MicroLC-MS method provided highly reproducible retention time, spectra, peak height and peak area for all three antibodies. Figure 3. Advantage of On-Column Desalting Method. TIC chromatogram and TOF MS spectra for intact adalimumab (5 ng load on column) clearly shows the gain in sensitivity and efficient desalting using an on-column approach (pink spectra) vs on-membrane desalting (blue spectra). Quantitation of mabs with Increased Dynamic Range Figure 2. TIC Chromatogram for Waters Intact mab Standard. Overlays of microflow LC replicates for mab at 5-1 ng load on column shows high reproducibility of RT and intensity (STDEV.P of.1). Peak height corresponding to.1-1 ng mab load on column is shown in Figure 2a. The spectra for +51 charge species of Waters mab in.1-1 ng range are shown in Figure 2b. Quantitation of intact proteins was evaluated based on linear dynamic range of the method and the lower limit of detection (LLOD) and quantitation reproducibility and accuracy. Three different intact antibodies: Waters Intact mab Standard, adalimumab (Humira) and trastuzumab (Herceptin), were used to evaluate the analytical performance of our integrated M3 MicroLC coupled to a high-resolution TripleTOF 66. The MS spectra were analyzed with MultiQuant using protein charge states peak area with linear regression and 1/x weighting (Figure 4). The same MS data were used for characterization of glycans and their abundances using BioPharmaView software. The M3 MicroLC provided increased sensitivity for analysis of intact proteins as compared to standard-flow LC due to improved p3

electrospray ionization and sampling efficiency of protein charged species and decreased matrix suppression effect. Quantitation over 3 orders of magnitude dynamic range (.1-1 ng/column) was achieved for the Waters Intact mab Standard (Figure 4a), with extended linear response for 5-1 ng (Figure 4b). Quantitation over 3 orders of magnitude linear dynamic range (.1-1 ng/column) was achieved for intact adalimumab (Figure 4c) and 2.5 orders of magnitude linear dynamic range (.5-1 ng/column) was achieved for trastuzumab, (Figure 4d). Results showed improved sensitivity with high accuracy and CV for intact antibody quantitation across 3 orders of magnitude linear dynamic range (.1-1 ng). The sensitivity gained using the M3 MicroLC coupled with the high resolution and mass accuracy of TripleTOF 66 was essential for this comprehensive analysis. In addition, we compared the linear regression with a Wagner logarithmic regression 3 which worked better for the nature of intact proteins at higher concentration and allowed quantification over 4 orders of magnitude dynamic range. a) Waters Intact mab Standard.1-1 ng on column b) Waters Intact mab Standard 5-1 ng on column c) Adalimumab.1-1 ng on column d) Trastuzumab.5-1 ng on column Figure 4. Linear Quantitation Curve for mabs. Calibration curves for 3-4 replicate analysis of different amount of Waters Intact mab Standard (Figure 4a and 4b) with r =.99 (charge state +48). Similar data and r values were obtained for quantitation of adalimumab (sum of charge states +46 to +55) (Figure 4c) and trastuzumab (charge state +55) (Figure 4d). p4

Intact mab Heterogeneity Characterization and PTM Analysis The sensitivity gained using the M3 MicroLC coupled with the high resolution and mass accuracy of TripleTOF 66 enabled comprehensive analysis of intact mabs at low ng level. We demonstrated successful characterization of Waters Intact mab Standard in low nanogram amounts (2.5 ng) and characterized different PTMs, including N-terminal glutamate to pyroglutamate conversion and different glycosylations (M*+GF+GF, M*+GF+G1F, M*+G1F+G1F, M*+G1F+G2F, M*+G2F+G2F). Characterization of mabs glycoforms and other PTMs using BioPharmaView provided routine molecular weight determination and characterization of mabs (Figure 5). Results show the potential application of this workflow, for both quantitation and characterization in low nanogram amounts, to other mabs and IgG classes. Additional modifications such as oxidation, deamidation, pyroglutamic acid formation at the N- terminus and N-terminal lysine loss were identified on adalimumab (Figure 6) and trastuzumab (Figure 7). Our data identified potential major glycoforms. Deeper characterization may be further evaluated with glycopeptides analysis at MS/MS and MS/MRM and level. 4 % of Total Area by Multiplicity 8 7 6 5 4 3 2 1 Adalimumab [Modifications] [Deamidated] [G] [GF] [GF-GlcNAc] [G1F] [G1F-GlcNAc] [G2F] [Oxidation] [Protein Terminal Lys-loss] Figure 6. Characterization of Adalimumab by BioPharmaView. Intact adalimumab mass analysis including spectral deconvolution, mass reconstruction and analysis of glycans and other PTMs was performed using BioPharmaView. Figure 5. Quantitative Multiplicity Characterization of mab Glycoforms. Intact mabs mass analysis, including spectral deconvolution, mass reconstruction and analysis of glycans and other PTMs was performed. We demonstrated successful characterization using only 2.5 ng of Waters Intact mab Standard. This workflow was developed to confirm the intact mass of the target biotherapeutic antibodies and to profile product-related variants such as N-glycosylation of major glycan isoforms and antibody drug conjugates (ADCs) or other PTMs. The SCIEX M3 MicroLC coupled with a SCIEX TripleTOF 66 mass spectrometer and aided by the advanced, easy-to-use MultiQuant and BioPharmaView software provided accurate quantitation of intact monoclonal antibodies at subnanogram-to-nanogram levels. The method and instrumentation also demonstrated successful characterization of multiple PTMs and glycosylation states on intact antibodies at very-low-nanogram levels. Oncolumn desalting simplified and minimized sample preparation, enabling the use of the more sensitive microlc-ms technique. p5

References 1. Johansen, E., Shin, B-H., and C. Hunter. Simultaneous Quantitative and Qualitative Analysis of Proteolytic Digests of Therapeutic Monoclonal Antibodies Using a TripleTOF System. SCIEX Technical Note, Document Number 746213-1. 2. Eric Johansen, Jenny Albanese and Christie Hunter. Analysis of Intact and Reduced Therapeutic Monoclonal Antibodies using the TripleTOF 56 System. SCIEX Technical Note, Document Number 422211-1. 3. Harvey M. Wagner. Linear Programming Techniques for Regression Analysis, American Statistical Association Journal, March 1959, p 27-212. % of Total Area by Multiplicity 8 7 6 5 4 3 2 1 Trastuzumab [Modifications] [Deamidated] [G] [GF] [G1] [G1F] [G2] [G2F] [Glu > pyro-glu] 4. Jenny Albanese, Christie L. Hunter and Ningombam Sanjib Meitei, Profiling the Distribution of N-Glycosylation in Therapeutic Antibodies using the QTRAP 65 System. SCIEX Technical Note, Document Number RUO-MKT-2-259-A Figure 7. Characterization of Trastuzumab by BioPharmaView Intact adalimumab mass analysis including spectral deconvolution, mass reconstruction and analysis of glycans and other PTMs was performed using BioPharmaView. AB Sciex is doing business as SCIEX. 218 AB Sciex. For research use only. Not for use in diagnostic procedures. The trademarks mentioned herein are the property of the AB Sciex Pte. Ltd. or their respective owners. AB SCIEX is being used under license. RUO- MKT-2-4948-C Headquarters 5 Old Connecticut Path, Framingham, MA 171, USA Phone 58-383-78 sciex.com International Sales For our office locations please call the division headquarters or refer to our website at sciex.com/offices