The manual MGIT system for the detection of M tuberculosis in respiratory specimens: an experience in the University Malaya Medical Centre

Similar documents
Evaluation of the BD BACTEC MGIT 320 for Detection of Mycobacteria and. Drug Susceptibility testing of Mycobacterium tuberculosis

Mycobacterial transport medium for routine culture of fine needle aspiration biopsies

Nichole Hines, 1 Janet B. Payeur, Lorraine J. Hoffman

Mycobacterial Culture

Standard Operating Procedure (SOP) Specimen processing of CSF, lymph nodes and other tissues for Xpert MTB/RIF

Recent Approaches in Detection of Drug- Resistant Tuberculosis. Dr M Hanif Bacteriologist Laboratory Division New Delhi Tuberculosis Centre

TB Intensive Tyler, Texas December 2-4, 2008

Assessing quality-assured diagnoses made by TB laboratories

Evaluation of the MB/BACT automated mycobacteria culture systeim versus culture on Lowenstein medium

Department of Microbiology, M.M. Institute of Medical Sciences and Research 2

Noncommercial culture and drug-susceptibility testing methods for screening patients at risk for multidrugresistant.

LABORATORY METHODS: Tuberculosis Diagnosis. Specimen collection and transport. Collection and transport (2)

ORIGINAL ARTICLE Verification of Bact/Alert 3D System Results for Detection of Mycobacterium Tuberculosis

Cat. no. TB100 TB MODS Kit 100 tests/kit

Automated methods in the Microbiology Lab-Issues and Troubleshooting! Presented By: Dr B. Oboudi

MANUSCRIPT JCM Revised. sputa subjected to long-term storage ACCEPTED

Performance of the new automated Abbott RealTime MTB assay for rapid detection of Mycobacterium tuberculosis complex in respiratory specimens

Evaluation of INNO-LiPA assay for direct detection of mycobacteria in pulmonary and extrapulmonary specimens

Culture negative endocarditis- an approach to laboratory diagnosis. Dr Shradha Subedi Microbiology Registrar Westmead Hospital, Sydney, NSW

Positioning of TB Diagnostics within a Tiered System Integrated Approach from Reference to District Laboratory. Giorgio Roscigno CEO FIND

BBL QualiSwab Quality Control Culture Devices

Stool GeneXpert MTB/Rif Assay

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

MTB Real-TM. For in Vitro Diagnostic Use. For use with RotorGene 3000/6000 (Corbett Research), SmartCycler (Cepheid), iq icycler and iq5 (Biorad)

Procedure Manual. Thin Layer Agar (TLA) Microcolony detection. Rapid culture of Mycobacterium tuberculosis. Anandi MARTIN, PhD

Mycobacterium tuberculosis complex

Mycobacterium tuberculosis End-Point PCR Kit Product# EP42100

Advantages in every dimension. BacT/ALERT 3D TB Culture

INTRODUCTION water-soluble Figure 1.

Mycobacterium tuberculosis complex. genesig Easy Kit for use on the genesig q reaction. Primerdesign Ltd

Does neutralization of gastric aspirates from children with suspected intrathoracic. tuberculosis affect mycobacterial yields on MGIT culture?

Supplement. Guideline for Investigation of Suspected Transfusion Transmitted Bacterial Contamination. Canada Communicable Disease Report

Chapter 02 Tools of the Laboratory: Methods for the Culturing and Microscopic Analysis

Forecast diagnostics for antimicrobial resistance (AMR)

i) Wild animals: Those animals that do not live under human supervision or control and do not have their phenotype selected by humans.

Key Words: MB BACT, MB7H10 medium, Mycobacterium tuberculosis

Laboratory Procedures in Clinical Microbiology

HardyVal TM CSP RANDOM TEST KIT

Selection of Storage Methods for Maintenance of Different Stock Cultures

BBL Lowenstein-Jensen Medium BBL Lowenstein-Jensen Medium with 5% Sodium Chloride L Rev. 08 January 2007

Serum Ceruloplasmin Albumin Ratio as A Marker of Treatment Response in Pulmonary Tuberculosis

F or many years, the isolation of urinary pathogens has

Comparison of four culture systems for Mycobacterium tuberculosis in the Zambian National Reference Laboratory

Preface. The Global Plan to Stop TB anticipates the availability of new diagnostics, drugs and vaccines by 2015.

Tuberculosis (TB) is an important public

Mycobacterium tuberculosis complex

Sputum processing, Cattamanchi et al, UCSF, 20 August, Cattamanchi A a,b*, Davis JL a,b, Hopewell PC a,b, Steingart KR b

Nontuberculous Mycobacteria

Advances in Medical Microbiology - Are the Patients Better Off?

Mycobacterium tuberculosis complex

ABC. Methods for Determining Bactericidal Activity of Antimicrobial Agents; Approved Guideline. Volume 19 Number 18

Bassett Healthcare The Mary Imogene Bassett Hospital Microbiology Laboratory FUNGAL BLOOD CULTURE PROCEDURE

Biosafety and Risk Assessment for New Molecular Methods

Novel Diagnostic Methods of Tuberculosis

Considerations for Conventional Drug Susceptibility Testing and Molecular Detection of Drug Resistance

320 MBIO Microbial Diagnosis. Aljawharah F. Alabbad Noorah A. Alkubaisi 2017

Line Probe Assays (LiPA)

TITLE: LIVE/DEAD VIABILITY FOR CLINICAL SAMPLES

What is it? Sherlock knows! Microbial Identification System

Interim Technical Note The Use of Cholera Rapid Diagnostic Tests November 2016

Study of Target gene IS 6110 and MPB 64 in Diagnosis of Pulmonary Tuberculosis

the Simple, Efficient, Effective

ORIGINAL ARTICLE /j x

Rapid diagnosis of drugresistant

determine optimum instrument settings for their own instruments and establish their own daily values.

Transmission Electron Microscopic Study of Antibiotic Action on Klebsiella pneumoniae Biofilm

Methods of Culturing Microorganisms. Chapter 3. Five Basic Techniques of Culturing Bacteria. Topics

3B BIOTUB QT MTB/ NTM Real Time PCR

DEPARTMENT OF CLINICAL LABORATORY SCIENCES SCHOOL OF HEALTH TECHNOLOGY AND MANAGEMENT THE UNIVERSITY AT STONY BROOK STONY BROOK, NEW YORK

UltraClean Midi Plasmid Prep Kit

Titration of Fluorochrome-Conjugated Antibodies for Labeling Cell Surface Markers on Live Cells

Cell Growth and DNA Extraction- Technion igem HS

Rapid Assay for Mycobacterial Growth and Antibiotic Susceptibility Using Gel Microdrop Encapsulation

QIAsymphony SP Protocol Sheet

Chapter 03 - Tools of the Laboratory: Methods for the Culturing of Microscopic Analysis of microorganisms

BD GeneOhm MRSA ACP Lysis Kit

GRAM STAIN/CULTURE of CSF and OTHER BODY FLUIDS by HEMATOLOGY LAB (11:30pm 07:30am)

Table of Contents. II. Kit Components III. Materials required but not supplied VII. Experimental Examples IX. Troubleshooting...

Rapid molecular diagnosis of TB and drug-resistant TB

Pallavi Sinha 1, Anamika Gupta 1, Pradyot Prakash 1, Shampa Anupurba 1*, Rajneesh Tripathi 1 and G. N. Srivastava 2

Bio-Rad Laboratories, Inc Alfred Nobel Dr. Hercules, CA USA (510) Rev C

Mycobacterium Tuberculosis

21.4 Recombinant DNA technology Calculation worksheet. AQA Biology. Calculating the efficiency of DNA transfer during genetic engineering

Providing clear solutions to microbiological challenges TM. cgmp/iso CLIA. Polyphasic Microbial Identification & DNA Fingerprinting

Aurum Plasmid Mini Kit. Instruction Manual. Bio-Rad Laboratories, Inc Alfred Nobel Dr. Hercules, CA USA (510)

Evaluating Diagnostic Tests in the Absence of a Gold Standard

Evaluation of a new automated, rapid, colorimetric culture system using solid medium for laboratory diagnosis of tuberculosis and determination of

Determination of Pseudomonas aeruginosa by Biochemical Test Methods Test, a Modified Biochemical Test for

MICROBIAL GROWTH. Dr. Hala Al-Daghistani

Acinetobacter baumannii

Abstract. Mary Jane Ferraro, PhD, MPH Jana M. Swenson, MMSc

Supplementary Appendix

Project 5: Urine Cultures and Identification

Should IGRAs replace the TST?

Utility of the Germ Tube Test for the Identification of Candida albicans Directly from Positive Blood Culture Bottles. ACCEPTED

LESSON ASSIGNMENT. After completing this lesson, you should be able to: Identify principles for maintaining a "working" stock culture.

UltraClean Endotoxin-Free Mini Plasmid Prep Kit

Transcription:

Malaysian J Pathol 2009; 31(2) : 93 97 ORIGINAL ARTICLE The manual MGIT system for the detection of M tuberculosis in respiratory specimens: an experience in the University Malaya Medical Centre FADZILAH Mohd Nor MBBS, Kee Peng NG MBBS, PhD, Yun Fong NGEOW FRCPath, MD Department of Medical Microbiology, Faculty of Medicine, University of Malaya, Kuala Lumpur, Malaysia Abstract A prospective study was conducted on 510 respiratory specimens for the presence of M. tuberculosis detected by direct acid-fast bacilli (AFB) smear examination, culture in the Manual Mycobacteria Growth Indicator Tube (BBL MGIT, Becton-Dickinson) and culture on Lowenstein-Jensen (LJ) medium. From positive BBL MGIT tubes, Ziehl-Neelsen and Gram stains were performed and subcultures were put up on LJ medium. A total of 101 (19.8%) specimens were positive by the BBL MGIT, 60 (11.8%) by primary LJ medium culture, 31 (6.1%) by direct smear examination and 29 (5.7%) by all three methods. Using primary LJ culture as the gold standard, the sensitivity and specificity of the BBL MGIT were 90% and 89.6% respectively but the sensitivity of AFB smear microscopy was only 48.3%. About half (51.1%) of the BBL MGIT false positives were due to contamination by non-afb bacteria. The remaining false positives comprised specimens that were AFB microscopy positive but LJ culture negative. Of the AFB isolates obtained on LJ primary and sub-cultures, almost all (93.3%) were identified as Mycobacterium tuberculosis complex. The mean time-to-detection was significantly shorter (p<0.0001) for the BBL MGIT than for LJ culture. For the former, positive results were available within 14 days for both AFB smear-positive and AFB smear-negative specimens. On the average, positive results were obtained 1.8 days earlier for direct AFB smear-positive samples than for AFB smear-negative samples. On the other hand, positive growth on LJ medium appeared after at least 33 days of incubation. These findings suggest that the BBL MGIT system will be a suitable alternative to LJ culture for the routine diagnosis of pulmonary tuberculosis, but a combination of liquid and solid cultures is still required for the highest diagnostic accuracy. Keywords: Mycobacterium tuberculosis diagnosis, manual MGIT system INTRODUCTION Mycobacterium tuberculosis, the aetiological agent of tuberculosis (TB), is one of the most successful bacterial pathogens in the history of mankind. Despite the availability of antituberculosis chemotherapy for the past 5 decades, this organism continues to cause extensive morbidity and mortality, infecting one third of the world s population and causing an estimated 8.8 million new cases and approximately 2 million deaths each year. 1 TB is also the primary cause of death among individuals infected by the human immunodeficiency virus. The World Health Organization (WHO) in 1993 declared TB to be a global public health emergency and reported that more than 95% of the estimated annual new cases and deaths occur in the developing world. 2 The M. tuberculosis complex comprises human and bovine strains of M. tuberculosis as well as M. microti, M. africanum, M. canettii, M. caprae and M. pinnipedii which are very closely related genetically but can be differentiated by biochemical and molecular characteristics. Currently, all commercially available tests for TB detect the M. tuberculosis complex as a single entity. The Mycobacterium Growth Indicator Tube (MGIT) system from Becton-Dickinson was introduced about a decade ago as a means to speed Address for correspondence and reprint requests: Professor Ng Kee Peng, Department of Medical Microbiology, Faculty of Medicine, University of Malaya, 59603 Kuala Lumpur, Malaysia.Tel: 603 79492418. Email: kpng@ummc.edu.my 93

Malaysian J Pathol December 2009 up the isolation of slow-growing mycobacteria. This is done by liquid culture in a modified Middlebrook 7H9 broth and detection of growth by a ruthenium pentahydrate fluorescent oxygen sensor embedded in silicon at the bottom of the culture tube. Oxygen reduction induced by metabolising bacteria reduces fluorescence quenching. Hence, the level of fluorescence that the tube emits corresponds to the amount of oxygen consumed by organisms in the tube, and is proportional to the number of bacteria present. This test is now automated for continuous monitoring in the BACTEC MGIT 960 (Becton Dickinson) but the older manual system requires tubes to be examined for fluorescence under a Wood s lamp or with some other long wave UV light source. Though requiring more technical time, the manual system is a lot more affordable for laboratories with a smaller budget. In this study, the manual MGIT system (BBL MGIT) from Becton Dickinson was evaluated against conventional culture on Lowenstein-Jensen (LJ) agar slopes and microscopical examination for acid-fast bacilli (AFB) in Ziehl-Neelsen (ZN) stained smears. MATERIALS AND METHODS Clinical samples The clinical samples studied comprised sputa, bronchoalveolar lavages, pleural fluids and tracheal aspirates collected from patients suspected to be having pulmonary or pleural TB. These were all consecutive samples received in the TB and Fungus Laboratory, University Malaya Medical Centre, from March 28, 2007 until September 26, 2007. All specimens were processed for AFB smear microscopy and culture on the day of collection. Microscopy Smears were screened by Auramine-O fluorescence microscopy under 40x magnification. Positive smears were re-examined with ZN staining for AFB under oil immersion. Culture on Lowenstein-Jensen medium Regardless of microscopy results, all specimens were decontaminated and digested using the BBL MycoPrep Reagent (Becton Dickinson) containing N-acetyl L-cysteine- 2% sodium hydroxide (NALC-NaOH). For each specimen, equal volumes (about 1 ml each) of reagent and specimen were pipetted into a centrifuge tube, vortexed briefly and allowed to stand at room temperature for 15 minutes. The BBL MycoPrep Phosphate Buffer (ph6.8) was added to the 10 ml mark on the centrifuge tube and the resulting mixture was centrifuged at 3,000 rpm for 15 minutes. The supernatant fluid was discarded. A small amount of phosphate buffer was added to resuspend the sediment which was used to inoculate a slant of LJ medium (Oxoid). All cultures were incubated at 37 0 C and examined daily until the appearance of colonies resembling mycobacteria. ZN staining was carried out on the colonies to identify AFB. Confirmed AFB were sent to the TB Reference Laboratory at the National Public Health Laboratory, Sg. Buloh, Selangor, for speciation. The time-to-detection for LJ culture was defined as the time between inoculation and the earliest reading when colonies resembling mycobacteria were noted on the agar surface. Specimens were considered culture negative if there were still no colonies formed after 8 weeks of incubation. Culture by the manual MGIT system (BBL MGIT) Specimens were decontaminated and digested as described above for LJ culture. A portion (0.5ml) of the processed specimen was inoculated into each BBL MGIT tube together with BACTEC MGIT Growth supplement (oleic acid-albumindextrose-catalase) and BBL MGIT PANTA antibiotic mixture (polymyxin B, amphotericin B, nalidixic acid, trimethoprim, and azlocillin). The BBL MGIT tubes were incubated at 37 0 C and examined daily in a 365nm wavelength UV light source fluorescence detector (Bactec TM MicroMGIT device, Sparks, MD, Maryland, USA) up to 8 weeks of incubation. A positive and a negative control tube were included with every batch of test tubes examined. In addition, bacterial growth was detected by the presence of a non-homogenous turbidity or small grains or flakes in the culture medium. When a tube was found to be positive for bacterial growth, a portion of the tube content was removed and used to prepare two smears, one for ZN staining and one for Gram staining. If AFB were present in the ZN smear, the tube content was subcultured onto a slant of LJ medium. If organisms grew on subculture and were identified as M. tuberculosis complex, the BBL MGIT culture was considered a true positive. However, if AFB were not seen but the Gram stained smear showed other bacteria or fungi, the BBL MGIT culture was considered contaminated. Time-to-detection was defined 94

MGIT SYSTEM FOR M TUBERCULOSIS as the interval between the incubation of an inoculated tube and a positive reading in a tube whose content was also AFB smear positive. RESULTS During the 6-month study period, a total of 510 respiratory specimens collected from 332 patients were processed. The specimen types consisted of 457sputum samples (90.0%). 37 pleural fluids (7.0%), 10 bronchoalveolar lavages (2.0%), and 6 tracheal secretions (1.0%) The results showed 101 (19.8%) positives by the BBL MGIT, 60 (11.7%) by LJ medium culture, 31 (6.1%) by direct AFB smear examination and 29 (5.7%) by all three methods (Tables 1 & 2). From the speciation tests carried out in the National Reference TB Laboratory, Sg. Buluh, 54 (93.3%) of the 60 isolates on LJ agar were identified as M. tuberculosis complex. Members of the M. fortuitum complex made up the remaining 6 isolates. Sensitivity and specifi city The BBL MGIT was positive for all the LJ culture positive specimens except for 6 specimens that grew M. tuberculosis complex, thus giving a sensitivity of 90% (54/60) against LJ culture. In contrast, the sensitivity of the AFB microscopy was only 48.3% (29/60). Of the 54 BBL MGIT positives, 29 (53.7%) were ZN smear positive and 25 (46.3%) were ZN smear negative. The specificity of the BBL MGIT was calculated to be 89.6%, as this test was negative in 403 of the 450 LJ culture negative specimens. Of the 47 false positives defined by the absence of mycobacterial growth on LJ agar, 24 were due to obvious contamination by other aerobic bacteria, thus giving an overall breakthrough contamination rate of 4.7% (24/510) for the BBL MGIT system. In the other 23 BBL MGIT positive tubes, AFB were seen in the ZN stains done on the tube contents but the LJ subcultures were negative for mycobacteria. The corresponding primary LJ cultures of these specimens were also negative. Two specimens were positive by direct AFB smear examination but negative by both culture systems. Clinical information was not available to determine whether or not these specimens were taken after the start of anti-tb treatment. Repeat samples Of 107 patients who gave repeat samples (2 to 3 samples each), the BBL MGIT test was negative in 80 and positive in 27 patients. Discrepant results from consecutive cultures were seen in only 6 (5.6%) patients. TABLE 1. BBL MGIT and LJ culture results in direct AFB microscopy positive specimens BBL MGIT + BBL MGIT Total LJ culture + 29 0 29 LJ culture 0 2 2 Total 29 2 31 + Positive result Negative result TABLE 2. BBL MGIT and LJ culture results in direct AFB microscopy negative specimens excluding specimens with negative results in all 3 tests BBL MGIT + BBL MGIT Total LJ culture + 25 6 31 LJ culture 47* 0 47 Total 72 6 78 + Positive result Negative result * Made up of 23 tubes that were positive in the AFB stain but negative in the LJ subculture and 24 tubes with Gram-stain showing contamination by other bacteria. 95

Malaysian J Pathol December 2009 Time to detection The mean time to detection by culture was determined for AFB smear-positive versus AFB smear-negative specimens. With the AFB smear-positive specimens, the mean times to detection were 11.5 days (SD 3.4 days; median, 12.0 days) with the BBL MGIT system, and 31.2 days (SD 5.8 days; median 32.0 days) with LJ culture. For AFB smear- negative specimens, the corresponding times were 13.2 days (SD 5.0 days; median 13.0 days) with the BBL MGIT system, and 35.3 days (SD 5.3 days; median 34.0 days) with LJ culture. The difference between BBL MGIT and LJ culture was statistically significant (p<0.0001). DISCUSSION TB in its typical clinical presentation is often diagnosed and treated on suspicion. Laboratory confirmation is however, very important not only for the optimal management of patients and their close contacts, but also for the generation of accurate epidemiological data to be used as the foundation for appropriate and effective control strategies. The strength of the laboratory network in a country is often a direct reflection of the success of its TB control programme. Improvements in the laboratory diagnosis of TB have focused on the reduction of turn-around time for detection and drug susceptibility testing, to allow for earlier detection of drug resistance and treatment failure. Light microscopy is rapid, inexpensive, technically simple, and specific for AFB, but will only detect open infectious cases of TB. The overall sensitivity of AFB microscopy has been reported to be 22 to 78% for respiratory specimens. While 10 6 AFB/ml of sputum usually results in a positive smear, the detection rate drops to 60% if 10 4 AFB/ml are present and to only 10% if less than 10 3 AFB/ml are in the smear. 3,4 Microscopy is frequently negative for non-respiratory specimens because of the lower bacterial load, and for NTM, because some species have cell walls that do not stain well in AFB stains. 5 Furthermore, microscopy will not discriminate between M. tuberculosis and other mycobacteria, nor viable from dead bacilli. Hence, microscopy is not very useful for the diagnosis of early TB, post-treatment TB and some NTM infections. However, it is of epidemiologic and clinical interest to be able to diagnose TB in patients who are AFB smear-negative, because at this stage they are less contagious and have lower morbidity and mortality. 6,7 M. tuberculosis culture is still the cornerstone on which definitive diagnosis of TB relies. It can be performed on a variety of respiratory and non-respiratory samples, is much more sensitive than microscopy and allows the recovery of the bacteria for other studies, such as the identification of mycobacterial species, drug susceptibility testing and genotyping. Liquid cultures generally require a shorter incubation time than culture on solid media for the detection of M. tuberculosis. 8 Automated systems such as the BACTEC MGIT 960, ESP Culture System II, and MB/BacT ALERT 3D system enable large numbers of specimens to be processed rapidly with positive results appearing usually within 10-12 days of incubation. However, these systems have similar problems of contamination, crosscontamination and the need for subculture onto solid media for the confirmation of identity and purity. Cruciani et al., 9 in a meta-analysis of 10 studies, reported that the BACTEC 960/MGIT and BACTEC 460 systems showed a sensitivity and specificity in detecting mycobacteria of 81.5 and 99.6% and 85.8 and 99.9%, respectively, but when combined with solid media, the sensitivity of the two systems increased to 87.7 and 89.7%, respectively. Both systems had a higher sensitivity for smear-positive specimens. Rates of breakthrough contamination were 12.8% with solid media, 8.6% with the BACTEC 960/MGIT and 4.4% with the BACTEC 460. In this study, the sensitivity (90%) and breakthrough contamination rate (4.7%) of the manual BBL MGIT were better than those of the automated BACTEC systems described in Cruciani s meta-analysis but the specificity (89.6%) was significantly lower. This inferior specificity could be due to the large number of specimens that were considered false positives in the BBL MGIT because they were negative in the ZN smear and LJ primary and subcultures. Although technical error could not be ruled out, some of these apparent false positives could be due to the growth of non-tuberculous mycobacterial species that require special growth factors not provided in LJ medium, or grew better in liquid than solid media. 10 With the exclusion of these contentious results, the BBL MGIT specificity would be 92.8%. As reported by other workers 9,10 the BBL MGIT was positive more often for AFB smear-positive than smear-negative specimens and the mean time-to-detection was significantly shorter for the BBL MGIT than for LJ culture, for both AFB smear-positive and AFB smear-negative 96

MGIT SYSTEM FOR M TUBERCULOSIS specimens. The highest recovery rate is also obtained with a combination of liquid and solid media. Overall, the findings in this study indicate a useful role for the manual BBL MGIT system in the routine diagnosis of pulmonary TB, especially in laboratories with a small number of specimens for TB detection. Although it is more costly and laborious, it has distinct advantages over the conventional LJ culture with respect to faster growth and better growth of fastidious mycobacterial species. Nevertheless, as shown by the 6 LJ culture positive but BBL MGIT negative specimens, a combination of solid and liquid culture systems is still required for the highest recovery of mycobacteria from clinical specimens. Furthermore, as shown by the inconsistent detection in some repeat samples, the sensitivity of culture will increase with more number of specimens cultured. 9. Cruciani M., Scarparo C, Malena M, Bosco O, Serpelloni G, Mengoli C. Meta-analysis of BACTEC MGIT 960 and BACTEC 460 TB, with or without Solid Media, for Detection of Mycobacteria. J Clin Microbiol. 2004 May; 42(5): 2321 5. 10. Vibeke stergaard Thomsen VØ, Dragsted UB, Bauer J, Fuursted K, Lundgren J. Disseminated Infection with Mycobacterium genavense: a Challenge to Physicians and Mycobacteriologists. J Clin Microbiol 1999;37:3901-5 ACKNOWLEDGEMENT The authors thank Becton Dickinson Malaysia for the supply of BBL MGIT kits and technical support. REFERENCES 1. Dye C, Scheele S, Dolin P, et al. Consensus statement. Global burden of tuberculosis: Estimated incidence, prevalence, and mortality by country. WHO Global Surveillance and Monitoring Project. JAMA 1999;282:677-86 2. Sudre P, Dam GT, Kochi A. Tuberculosis: a global overview of situation today. Bull.WHO 1992;149-59. 3. Palomino JC. Nonconventional and new methods in the diagnosis of tuberculosis: Feasibility and applicability in the field. Eur Respir J 2005;26;339-50 4. Pfyffer GE, Brown-Elliot BA, Wallace Jr RJ. Mycobacterium: General characteristics, isolation and staining procedures. Manual Clin Microbiol 2003;532-59 5. Lipsky BA, Gates J, Tenover FC, Plorde JJ. Factors affecting the clinical value of microscopy for acid-fast bacilli. Rev Infect Dis 1984;6:214-22 6. Menzies D. Issues in the management of contacts in patients with active pulmonary tuberculosis. Can J Public Health 1997;88:197-201 7. Behr MA, Warren SA, Salamon H, et al. Transmission of Mycobacterium tuberculosis from patients smear-negative for acid-fast bacilli. Lancet 1999;353:444-9 8. Morgan MA, Horstmeier CD, De Young DR, Robers GD. Comparison of radiometric method (BACTEC) and conventional culture media for recovery of mycobacteria from smear-negative specimens. J Clin Microbiol 1983;18:384-8 97