Standard Operating Procedure for the Determination of Specific Conductance CCAL 11A.2

Similar documents
Hach Method Luminescence Measurement of Dissolved Oxygen (LDO ) in Water and Wastewater

Standard Operating Procedure for:

User Guide. Conductivity Cell

Instruction Manual HI Portable ph/ec/tds/ C Meters

3.0 DETECTION LIMIT 3.1. See Procedure ALQAP Procedure for the Determination of Uncertainty

Advantages and Disadvantages of Conductivity Measurement

CONDUCTIVITY INSTRUMENT

Ti-Pure TITANIUM DIOXIDE. CONCENTRATED ACID SOLUBILITY OF TiO 2 METHOD: T WP

Total Dissolved Solids Dried at 180ºC. Total Volatile Dissolved Solids at 550 C

Instruction Manual HI 2210 HI Microprocessor-based ph/mv/ C Bench Meters.

Standard Operating Procedure for: Horiba U-22XD Multi-Parameter Water Quality Monitoring System (1050R01 Horiba U-22.doc) Missouri State University

Standard Operating Procedure (SOP)

+ - Conductivity. Conductivity Primer. Basic principles

Environmental Express. Joe Boyd. Solids and BOD Analyses: Pitfalls and Troubleshooting

A24 CALA Checklist for Microbiology Revision 4.1 February 16, 2018

ab Porcine VEGF-A ELISA Kit

Instruction Manual. HI HI Portable ph/ec/tds/ C Meters

Trace Metals in Waters by GFAAS, in Accordance with U.S. EPA and Health Canada Requirements

EC91346S Conductivity Cell Instruction Manual

SOP-C-130 Determination of Total Solids and Percent Solids

PORTABLE ph METER PRICE CODE NUMBER MODEL 100

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Required Documentation for PA State (Chapter 252) Accreditation

Human TACI (TNFRSF13B) ELISA Kit

METHOD 1050 TEST METHODS TO DETERMINE SUBSTANCES LIKELY TO SPONTANEOUSLY COMBUST

Municipal Wastewater. Introduction: Sample Matrix: Wastewater; various locations along the sewage treatment process.

Ensure Accurate and Successful BOD Measurements with Electrochemical or Optical Sensors

Ainvolved in many aspects of project planning, sample collection,

Report No Waters Proficiency Testing Program. Round No Total Solids, Total Suspended Solids, Total Dissolved Solids - March 2017

Fixed (Inorganic Ash) and Volatile Solids by Gravimetric Determination

Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric)

Volatile organic compounds Water No CAS Number assigned

OPERATING INSTRUCTIONS. Portable Turbidity Meter

EU Water Analysis Using the Thermo Scientific icap 7400 ICP-OES Duo

USEPA 1,2 Bicinchoninate Method 3 Method 8506 (CuVer 1) and Method 8026 (CuVer 2) 0.04 to 5.00 mg/l Cu Powder Pillows or AccuVac Ampuls

STANDARD OPERATING PROCEDURES SOP: 2040 PAGE: 1 of 10 REV: 0.0 DATE: 05/17/02

Evaluation copy. Chloride and Salinity. Computer INTRODUCTION

Method 108C - Determination of Arsenic Content in Ore Samples From Nonferrous Smelters (Molybdenum Blue Photometric Procedure)

Cadmium Testing and On Site Calibration for Water Testing Detection Range: ppm

LIFluor EnhanCE. for use with dsdna 1000 and dsdna 20,000 Kits

INTRODUCTION TO ELECTROCHEMISTRY: CURRENT, VOLTAGE, & BATTERIES. Introduction. Electrochemistry Revised 4/28/14

Sample for use a Guide Revised IDEXX COLILERT -18 TEST METHOD FOR THE SIMULTANEOUS DETECTION OF TOTAL COLIFORMS AND E.

SECTION D.3 CHLOROPHYLL AND PHEOPHYTIN

RayBio Rat IL-6 ELISA Kit (For Lysates)

Edition December 2015 Order no inter. Operation Manual GENO-therm analysis case

RayBio Human Caspase-3 ELISA Kit

Engineered Systems & Designs TURBIDITY METER

RayBio Custom ELISA Kit

AGAROSE GEL ELECTROPHORESIS. Assiut University

RELIABILITY OF TOC RESULTS IN ULTRAPURE WATER ANALYSIS

METHOD #: Approved for NPDES (Issued 1971) Oxygen, Dissolved (Modified Winkler, Full-Bottle Technique) ANALYTE: CAS # O Oxygen

METHOD 9013A CYANIDE EXTRACTION PROCEDURE FOR SOLIDS AND OILS

METHODS OF ANALYSIS BY THE U.S. GEOLOGICAL SURVEY NATIONAL WATER QUALITY LABORATORY-- DETERMINATION OF CHROMIUM IN WATER BY GRAPHITE

Process Electrodes and Probes

Total Suspended Solids by Gravimetric Determination

RAJESH NAIR DIRECTOR CALF-NDDB, ANAND, GUJARAT

Safety Services Guidance. Guidance on the safe use and disposal of ethidium bromide

YSI 5905, 5010, 200-BOD, and Pro-BOD Probe Instruction Manual

Rev Experiment 10

Instruction Manual HI RH/ C Transmitter. This Instrument is in Compliance with the CE Directives.

Method 5.9 Syrup: calcium and magnesium by EDTA titration

4. The Second-dimensional SDS-PAGE (vertical) Protocol

AAG PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

Modified Colitag Test Method for the Simultaneous Detection of E. coli and other Total Coliforms in Water (ATP D ) August 28, 2009

STANDARD OPERATING PROCEDURE FOR THE COLLECTION OF CHEMICAL AND BIOLOGICAL AMBIENT WATER SAMPLES

Urine Protein Concentration, Preservation and Isolation Kit 25 Individual Devices Product Insert Product #: 38200

Rat α-melanocyte stimulating hormone (α-msh) ELISA Kit

BENTLEY BACTOCOUNT IBCm INDUSTRY OPERATOR (BCMlO) APPROVAL PROCEDURES PROTOCOL

Total Dissolved Solids

Revised 27 Aug rm (Vers. 4.1)

ImmuLux Human IL-6 Fluorescent ELISA Kit

(aq) + 5e - Mn 2+ (aq) + 4H 2

Routine Cleaning, Disinfecting. and Maintenance of EEG Electrodes. PC-POD-FA-007-v01

ab Plasmin Activity Assay Kit (Fluorometric)

ab Factor IXa Activity Assay Kit (Fluorometric)

Porcine IgM (Immunoglobulin M) ELISA Kit

ort No. 892 Proficiency Testin Roun - Total Solids, Total Suspendedd Solids, Dissolved Solids - January 2015

Bovine Prostaglandin E2 (PG-E2) ELISA Kit

Calcium and Magnesium; Chlorophosphonazo Rapid Liquid Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Pour-Thru Cell

June 8, Submitted to: The Environmental Protection Agency Research Triangle Park, NC. Prepared by:

Practical guidance for switching between anion and cation analysis systems

FOR 3457 Forest Watershed & Water Quality Watershed Management Tools & Methods. Water Quality Lab

Mouse Luteinizing Hormone (LH) ELISA

TESTING THE WATERS HOW GOOD IS THAT BOTTLED WATER AND HOW EFFECTIVE IS YOUR WATER FILTER

Mouse ALP (Alkaline Phosphatase) ELISA Kit

Experiment. Molar Mass of an Unknown Sulfate Salt by Gravimetric Techniques 1

ab VEGF Human ELISA Kit

Human IL-1 alpha ELISA Kit

EXPERIMENT 3 SOLIDS DETERMINATION

SAFETY DATA SHEET according to Regulation (EC) No. 1907/2006

Human DNA Alu Amplification by Polymerase Chain Reaction (PCR)* Laboratory Procedure

NWTPH-Gx. Volatile Petroleum Products Method for Soil and Water

TACS MTT Assays. Cell Proliferation and Viability Assays. Catalog Number: TA tests. Catalog Number: TA tests

METHOD 3 - GAS ANALYSIS FOR THE DETERMINATION OF DRY MOLECULAR WEIGHT. NOTE: This method does not include all of the

Material Safety Data Sheet

Transcription Factor ELISA Kit

QuantiFluor ONE dsdna System

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

ort No. 925 Proficiency Testin Roun - Total Solids, Total Suspendedd Solids, Dissolved Solids - August 2015

Transcription:

Standard Operating Procedure for the Determination of Specific Conductance CCAL 11A.2 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall Corvallis, Oregon Prepared by Kathryn Motter And Cameron Jones Revised

Standard Operating Procedure for the Determination of Specific Conductance CCAL 11A.2 Specific Conductance/CCAL 11A.2 Page 2 of 11 Table of Contents 1.0 Scope and Application... 3 2.0 Summary of Method... 3 3.0 Definitions... 3 4.0 Interferences... 4 5.0 Safety... 4 6.0 Equipment and Supplies... 4 7.0 Reagents and Standards... 5 7.1 Preparation of Standards... 5 8.0 Sample Handling and Storage... 5 9.0 Quality Control... 5 10.0 Calibration and Standardization... 6 11.0 Procedure... 6 11.1 Calibration... 6 11.2 Sample Testing Procedure... 6 11.3 System Notes... 7 12.0 Data Analysis and Calculations... 7 13.0 Method Performance... 7 14.0 Pollution Prevention... 7 15.0 Waste Management... 8 16.0 References... 8 17.0 Tables, Diagrams, Flowcharts, and Validation Data... 9 17.1 ph, Alkalinity and Conductivity Data Sheet... 9 18.0 Document Revision History... 11

Standard Operating Procedure for the Determination of Specific Conductance CCAL 11A.2 Specific Conductance/CCAL 11A.2 Page 3 of 11 1.0 Scope and Application 1.1 Specific Conductance is a numerical expression of the ability of a water sample to carry an electric current. This number depends on the total concentration of ionized substances dissolved in the water and the temperature at which the measurement is made. The mobility of each of the various dissolved ions, their valence, and their actual and relative concentrations affect conductivity. The practical range of determination is 0 µs 3 S. 2.0 Summary of Method 2.1 Specific Conductance is determined with a YSI 3200 with temperaturecompensation. The meter is calibrated with a single point calibration. The samples are measured using a glass probe with thermistor. A consistency standard is analyzed once each analysis batch to monitor instrument drift over time. 3.0 Definitions 3.1 DI water: Water that has been through a deionization system to produce water similar to ASTM Type I reagent with 16.7 Mohms resistivity (ASTM) (Reference 16.3). 3.2 Method Detection Limit (MDL): The minimum concentration of an analyte that can be measured and reported with 99% confidence, based on a one-sided 99% confidence interval (t-value at a significance level of 0.01 and n-1 degrees of freedom) from at least seven repeated measurements of a low concentration standard measured within an analysis run. MDL = ts Where, t = Student s t value at a significance level of 0.01 and n-1 degrees of freedom

Page 4 of 11 s = standard deviation of at least seven repeated measurements of a low level standard 4.0 Interferences 4.1 Electrolytic conductivity (unlike metallic conductivity) increases with temperature at a rate approximately 2% per degree C. Significant errors can result from inaccurate temperature measurements. Although the YSI model 3200 meter compensates for variability in temperature, samples are still allowed to warm to at least 15º C to minimize temperature effect. 4.2 Sources of contamination include dust, carryover and salts from hands. The probe must be washed thoroughly between samples and gloves should be worn during the procedure. 4.3 Air bubbles trapped inside the probe will result in erratic measurements. Shake the probe to remove trapped bubbles before each measurement. 4.4 Build-up on the conductivity cell may contaminate solution and alter conductivity reading. The cell should be cleaned as a part of regular maintenance. 4.5 The platinum black on the electrodes is extremely important to cell operation. The cell should be cleaned and the electrodes replatinized if wear or flaking is observed. 5.0 Safety 5.1 The toxicity or carcinogenicity of each reagent has not been precisely determined; however, each chemical should be regarded as a potential health hazard. Exposure to these chemicals should be reduced to the lowest possible level. Cautions are included for known extremely hazardous materials. 6.0 Equipment and Supplies Note: Brand names, suppliers and part numbers are for illustrative purposes only. No endorsement is implied. Equivalent performance may be achieved using apparatus and materials other than those specified here, but demonstration of equivalent performance that meets the requirements of this method is the responsibility of the laboratory. 6.1 YSI 3200 Conductivity Instrument

Page 5 of 11 6.2 YSI 3256 Glass probe with thermistor 6.3 Nitrile gloves 7.0 Reagents and Standards 7.1 Preparation of Standards 7.1.1 QCCS - Dilute Demal s Solution (KCl Standard Solution): Dissolve 745.6 mg anhydrous KCl into 1000 ml of DI water at 25ºC. A working solution is prepared by diluting 30 ml of Demal s solution to 1-L total volume. The resulting KCl standard reference solution has a specific conductance of approximately 40 µs/cm at 25ºC and is satisfactory for most waters when using a cell constant between 1 and 2. 7.1.2 Calibration Check Standards/Calibration Standards: Standards are purchased from a vendor that provides traceability to NIST standards. 10 and 100 µs/cm standards are used daily to check calibration of the instrument. 8.0 Sample Handling and Storage 8.1 Unfiltered samples are stored at 4 C in the dark. Samples are analyzed within 7 days to ensure sample integrity. 9.0 Quality Control 9.1 Blank: DI water run before the calibration. 9.2 Quality Control Check Sample (QCCS): Run once per analysis batch. 9.3 Accuracy of calibration is checked daily with 10 and 100 µs/cm standards. If results fall outside acceptable precision limits, the instrument is recalibrated. 9.4 Method Detection Limit (MDL): Established for each analyte. Based on a one-sided 99% confidence interval (t-value) from at least seven repeated measurements of a low concentration standard. The t-distribution value is multiplied by the standard deviation of the population (n-1) to obtain the MDL.

Page 6 of 11 9.5 Analytical Duplicate: Separate analysis from the same sample aliquot. Run a minimum of once every analysis set. 10.0 Calibration and Standardization 10.1 Balances: calibrated yearly by external vendor. 11.0 Procedure 11.1 Calibration 11.1.1 Refer to the YSI model 3200 Operations Manual to properly configure the instrument. This will include calculating and storing the cell constant. Temperature compensation should be set to "natural". The instrument will need to be reconfigured each time a new cell is necessary. 11.1.2 At least 1 hour before initiating conductivity analysis fill the glass electrode chamber in the conductivity probe with fresh DI water. 11.1.3 Press the power switch and allow the conductivity meter to warm up and equilibrate for at least 15 minutes. 11.1.4 Rinse the probe with DI water and shake-out any excess. 11.1.5 Fill the probe with fresh DI water. 11.1.6 Allow the instrument to equilibrate (generally 1-2 minutes) before recording the readout on the data sheet. The reading for DI water should be approximately 0.4-0.7 µs/cm. Rinse probe and shake dry. 11.1.7 Analyze the QCCS solution. Rinse the electrode chamber with enough solution to completely rinse the electrodes and discard the rinse. Fill the electrode chamber with solution, allow the instrument to equilibrate (1-2 minutes) and record the measurement. Rinse probe with DIW and shake dry. 11.1.8 Analyze the 10 and 100 µs/cm standards following the same rinseand-fill procedure as above. Record results. If results are outside precision limits (+/- 0.1 µs/cm for 10 µs/cm standard or 2% for 100 µs/cm standard), the instrument is recalibrated. 11.2 Sample Testing Procedure 11.2.1 Rinse the electrode chamber three times with DI water and shakeout any excess after the last rinse. 11.2.2 Rinse the electrode chamber with enough sample to completely rinse the electrodes and discard the rinse. Then fill the electrode

Page 7 of 11 chamber with sample, allow the instrument to equilibrate (1-2 minutes) and record the measurement. 11.2.3 Repeat steps 1 and 2 for all samples. 11.3 System Notes 11.3.1 Although the YSI model 3200 meter compensates for variability in temperature, blank, standard and samples are still allowed to warm to at least 15º C to minimize temperature effect. 11.3.2 If there is a shift in order of magnitude in response between two samples, the meter is allowed to stabilize with the new sample; the sample is discarded and poured fresh for determination. 11.3.3 The electrode should always be immersed with DI water when not in use. This keeps the electrode hydrated and provides for more stable response. 11.3.4 Care must be taken not to insert anything in the electrode chamber that would cause any of the black platinum coating from the electrode to chip-off. 11.3.5 The instrument automatically sets the range for the analysis. All analyses are performed in conductance mode. Be certain to correctly record the proper unit magnitude (µs/cm, ms/cm, S/cm). Readings should be recorded to the nearest tenth of a unit for the µs/cm scale, to the nearest unit for the ms/cm and S/cm scales. 11.3.7 Keep a record of analyzed values for the QCCS over time. Large deviations show that the meter is not functioning or the solution has deteriorated. At 25 o C the standard reference solution has a specific conductance of approximately 40 µs/cm. 12.0 Data Analysis and Calculations 12.1 YSI model 3200 data are direct reading and require no calculation. 13.0 Method Performance 13.1 This method was validated through inter-laboratory studies. The CCAL Water Analysis Laboratory participates in the USGS Standard Reference Water QA program and the National Water Research Institute s (NWRI) Environment Canada Proficiency Testing (PT) Program. 14.0 Pollution Prevention

Page 8 of 11 14.1 The chemicals used in this method pose little threat to the environment when properly managed. 14.2 All standards and reagents should be prepared in volumes consistent with laboratory use to minimize the volume of disposable waste. 14.3 For further information on pollution prevention consult Less is better: Laboratory Chemical Management for Waste Reduction, available from the American Chemical Society s Department of Government Relations and Science Policy, 1155 16 th Street NW, Washington D.C. 20036, (202) 872-4477. 15.0 Waste Management 15.1 It is the laboratory s responsibility to comply with all federal, state and local regulations governing waste management, and to protect the environment by minimizing and controlling all releases from fume hoods and bench operations. Compliance with all sewage discharge permits and regulations is required. 15.2 For further information on waste management, consult The Waste Management Manual for Laboratory Personnel, and Less is Better: Laboratory Chemical Management for Waste Reduction, both available from the American Chemical Society s Department of Government Relations and Science Policy, 1155 16 th Street NW, Washington DC, 20036. 16.0 References 16.1 Standard Methods For The Examination of Water and Wastewater, Method 2510-Conductivity. American Public Health Association. 21 st Edition, 2005. 16.2 Code of Federal Regulations. Protection of Environment. Section 40, Appendix B to Part 136. Definition and procedure for the determination of the method detection limit. Revision 1.11. Revised July 1, 1990. Office of the Federal Register, National Archives and Records. 16.3 ASTM. American Society for Testing and Materials. Standard Specifications for Reagent Water. D1193-77 (Reapproved 1983). Annual Book of ASTM Standards, Vol. 11.01. ASTM: Philadelphia, PA, 1991.

Page 9 of 11 16.4 YSI 3200 Conductivity Instrument Manual. YSI Environmental, YSI Incorporated. 17.0 Tables, Diagrams, Flowcharts, and Validation Data 17.1 ph, Alkalinity and Conductivity Data Sheet

ph, Alkalinity and Conductivity Specific Conductance/CCAL 11A.2 Page 10 of 11 page of Project: Alkalinity Volume (ml): Filtered Unfiltered Analyst: Date: QCCS: T( C): Remarks: Buffer Calibration: ph 4.0 ph 7.0 ph 10.0 Cond Std Checks: 10 100 Cup # Sample # µs/cm Observations Smplr # QPHB1 QPHB2 QPHM1 QPHM3

Page 11 of 11 18.0 Document Revision History Original Document: March 2006 Version: 11A.0 Title: Standard Operating Procedure for the Determination of Conductivity Edit Date: February 2010 New Version: 11A.1 Address update Change text from Conductivity to Specific Conductance Section 13.1: add Environment Canada Proficiency Testing Program participation. Edit Date: New Version 11A.2 Section 7.1: revised standards Section 8.0: changed holding time for analysis Section 11: revised and reorganized for accuracy and clarity Section 17.2: updated data sheet General editing