Technical Focus. Direct PCR Amplification of Buccal Samples on the NUCLEIC-CARD COLOR With the AmpFlSTR Identifiler Direct PCR Amplification Kit

Similar documents
User Bulletin. Veriti 96-Well Thermal Cycler AmpFlSTR Kit Validation. Overview

Investigator Quantiplex HYres Kit

Development of a Global STR Multiplex for Human Identification Analysis. Julio Mulero, Ph. D. R&D Group Human Identification, Life Technologies

Development of a Global STR Multiplex for Human Identification Analysis

Development and Validation of the 3500 Series Genetic Analyzer for Human Identification

AmpF STR NGM PCR Amplification Kit - Overview

Authenticating human cell lines using the Identifiler kits and capillary electrophoresis platforms

GlobalFiler Kit Casework Test Site Results and Other Development Updates

Quantifiler Human DNA Quantification Kit Quantifiler Y Human Male DNA Quantification Kit

HID Validation: The 3500 Series Genetic Analyzers

Maximizing the Performance of Capillary Electrophoresis Systems The Maximizing Data Quality Series Part 4

AmpFlSTR Identifiler PCR Amplification Kit

Procedure for Casework DNA Interpretation

PowerPlex ESX 17 Fast System for Use on the Applied Biosystems Genetic Analyzers

PowerPlex. Y System Validation

Naming Convention and Laboratory Methods: PROVEDIt Database

Data Quality Worth Sharing

AUTOMATION OF DNA ISOLATION, QUANTITATION, AND PCR SETUP

QIAGEN Whole Genome Amplification REPLI-g Eliminating Sample Limitations, Potential Use for Reference Material

Internal Validation of the Promega PowerPlex Fusion System with the Applied Biosystems 3130xl Genetic Analyzer

Testing the Effectiveness of the PrepFiler Kit for DNA Extraction from Forensic Samples (An Overview of Test Site Data)

GenPlex HID Training Class I

GeneMarker HID Software Review

AmpFlSTR NGM SElect PCR Amplification Kit. User Guide

STR Analysis Following Latent Blood Detection by Luminol, Fluorescein, and BlueStar

Development of a Global STR Multiplex for Human Identification Analysis

A*01:02, 68:02 B*15:10, 58:02 DRB1*03:01, 12:01

TECHNICAL MANUAL. GenePrint 24 System. Instructions for Use of Product B1870 6/16 TM465

TaqPath ProAmp Master Mixes

Comparison of the Promega Power Y and AB AmpFlSTR YFiler kits and the Subsequent Validation of YFiler.

Procedure for GeneMapper ID-X for Casework

AmpFlSTR COfiler PCR Amplification Kit

NRAS Codon 61 Mutation Analysis Reagents

Uniparental disomy (UPD) analysis of chromosome 15

Sexual Assault Kit Screening Using the QIAcube HT & Investigator Quantiplex HYres!

Validation of PowerPlex Y23 PCR Amplification System for Forensic Casework

NRAS Mutation Analysis Reagents (Codons 12 and 13)

PowerPlex 5C Matrix Standards, 310

Validation of the MinElute Post PCR Cleanup System

STUDY OF VNTR HUMAN POLYMORPHISMS BY PCR

Relative Fluorescent Quantitation on Capillary Electrophoresis Systems:

Validation of a Multiplexed System for Quantification of Human DNA and Human Male DNA and Detection of PCR Inhibitors in Biological Samples

MightyAmp DNA Polymerase Ver.3

Laboratory Validation. Chapter 16

AccuSEQ Real-Time PCR Detection Software

SeCore SBT Sequence Based Typing

Evaluating Forensic DNA Evidence

PowerPlex 5-Dye Matrix Standards, 310

Pressure Cycling Technology (PCT) Applications for DNA Extractions from Challenging Forensic Samples

Ampli1 WGA Kit. Whole Genome Amplification for Single Cells. USER MANUAL y Version 01 Content version: July I WG R01 50 reactions

EVALUATION OF THEPRECISIONID SYSTEM FOR TARGETED SEQUENCING OF DNA AND RNA MARKERS FOR HUMAN IDENTITY AND BODY FLUID IDENTIFICATION

Short Tandem Repeat (STR) Analysis

John M. Butler and Peter M. Vallone National Institute of Standards and Technology

E.Z.N.A. MicroElute Genomic DNA Kit. D preps D preps D preps

JS 190- Population Genetics- Assessing the Strength of the Evidence Pre class activities

Standard for Forensic DNA Interpretation and Comparison Protocols

For in vitro Veterinary Diagnostics only. Real-Time PCR Detection Kit for detection of Histomonas meleagridis.

CRISPR/Cas9 Gene Editing

Latest Tools for Human Identification

Chapter 7 DNA Fingerprinting By the end of this chapter you will be able to:

BigDye Direct Cycle Sequencing Kit

Evaluation of the Autogen FlexSTAR automated DNA extractor, for suitability in routine diagnostic use

UF Center for Pharmacogenomics. Explanation of Services. UF Center for Pharmacogenomics Services

The ABC s of DNA. Jerome F. Buting BUTING, WILLIAMS & STILLING, S.C. Wisconsin State Public Defender Annual Conference November 12, 2015

SYBR Premix DimerEraser (Perfect Real Time)

AMPURE PCR PURIFICATION PAGE 1 OF 7

Automated Protocol for Extract-N-Amp Blood PCR Kits Using the Sciclone ALH 3000 Workstation (Caliper Life Sciences)

Applied Biosystems Real-Time PCR Rapid Assay Development Guidelines

PERFORMANCE MADE EASY REAL-TIME PCR

One Step SYBR PrimeScript RT-PCR Kit II (Perfect Real Time)

Developmental and Internal Validation of a Mitochondrial DNA Direct Amplification Kit for Forensic Reference Samples

Certificate of Analysis of the Holotype HLA 24/7 Configuration A1 & CE

Reagent Kit Selection Guide

QIAsymphony SP Protocol Sheet

SNP GENOTYPING WITH iplex REAGENTS AND THE MASSARRAY SYSTEM

E.Z.N.A. Blood DNA Mini Kit. D preps D preps D preps

Perform an HRM Methylation Study

Developmental Validation of the AmpF STR Ò Identifiler Ò Plus PCR Amplification Kit: An Established Multiplex Assay with Improved Performance

Quantitative analysis of PCR fragments with the Agilent 2100 Bioanalyzer. Application Note. Odilo Mueller. Abstract

Development of a fast PCR protocol enabling rapid generation of AmpFlSTR Identifiler profiles for genotyping of human DNA

PrimeScript RT reagent Kit (Perfect Real Time)

A genetic database for DNA-based forensic analysis in Brunei Darussalam

PAXgene Blood DNA PAXgene Blood DNA Tube (IVD) for clinical use PAXgene Blood DNA System (RUO) for research use. Explore more at

DNA Genotyping from Human FFPE Samples Reliable and Reproducible

DNA and RNA Clean-up Guide

TWGDAM Validation of AmpF STR PCR Amplification Kits for Forensic DNA Casework

2x PCR LongNova-RED PCR Master Mix

hamilton firefly nimbus 96

Product Name : Simple mirna Detection Kit

Evaluation of the genesig q16 quantitative PCR unit

qpcr Quantitative PCR or Real-time PCR Gives a measurement of PCR product at end of each cycle real time

1Genomic DNA purification, modification, and amplification

Company workshop. Forensica 2008

Effects of ph on Binding of DNA Using the PrepFiler BTA Forensic DNA Extraction Kit on the AutoMate Express Forensic DNA Extraction System

STR Profiling Matching Criteria: Establishment and Importance of a Cell Line Database

BenchSmart 96. Semi-automated Pipetting Higher Accuracy, Greater Flexibility

Simplified Low Copy Number Dna Analysis By Post Pcr Purification

Transcription:

Technical Focus Direct PCR Amplification of Buccal Samples on the COLOR With the AmpFlSTR Identifiler Direct PCR Amplification Kit Forensic News 2013 Alessandro Agostino 1, Annalisa Gervasoni 2, Alice Squassina 2, Jill Muehling 1 1 Life Technologies Corporation 2 Copan The Copan COLOR system is designed for collection and preservation of single-source buccal samples used in forensics. AmpFlSTR Identifiler Direct PCR Amplification kit or other Life Technologies direct amplification kits represents a reliable and cost-effective solution to process single-source buccal samples for forensic applications. Figure 1 & 2 COLOR is available as a stand-alone collection card that can be used with 4N6FLOQSwabs, or in a convenient Collection Device with the 4N6FLOQSwabs Lollipop Swab prepackaged with the card. The use of Copan COLOR with 4N6FLOQSwabs for buccal sample collection generates high-quality genotypes when amplified with the AmpFlSTR Identifiler Direct PCR Amplification kit. Introduction In forensic investigations, buccal swab samples are some of the most challenging samples to process in terms of reproducibility and success rate when DNA is amplified for STR profiling in direct mode (i.e., without sample extraction and quantification). High-throughput laboratories are seeking a solution to help reduce the cost and increase the first pass success rate of buccal samples, in order to lower the backlog and make the reference profiling of these samples more reliable and efficient. Copan COLOR is an indicating sample collection system that has been specifically designed to facilitate processing of colorless samples while maintaining the standard features of in terms of cell lysing capability, long term DNA integrity, and preservation. COLOR can be used in conjunction with the 4N6FLOQSwabs Regular size tip (Figure 1) or is combined with a 4N6FLOQSwabs Lollipop Swab and packaged as the Collection Device (Figure 2). The COLOR changes from a pink background to white when buccal samples are deposited onto the card. The change in background allows visualization of where the sample was deposited making it easier for manual and automated punching and increasing the probability of getting full STR profiles. COLOR combined with the Figure 1. COLOR and 4N6FLOQSwabs Regular size tip. Figure 2. Collection Device. Experiments and results Materials and methods Studies were designed to compare the Collection Device, the COLOR, and the EasiCollect device (Whatman) with foam swab and Indicating FTA card (Whatman). Buccal samples were collected based on each of the manufacturer s recommendation and air dried for one hour prior to storing at room temperature. With the exception of the samples used in the stability study, samples were stored for three days prior to processing. 1.2 mm punches were collected and amplified with the AmpFlSTR Identifiler Direct PCR Amplification Kit using the recommended protocol. PCR was performed on the Veriti PCR Thermal Cycler, capillary electrophoresis was performed on an Applied Biosystems 3130 Genetic Analyzer, and the analysis was performed with GeneMapper ID-X v1.3. Analysis was conducted using thresholds of 50 and 150 RFU, and data were evaluated for First Pass (full profile obtained from one amplification and one CE injection), Mean Peak Height, Intracolor Balance, and Intralocus Balance. A detailed description of the experiments performed and the purpose of the data assessment for that experiment is outlined in Table 1. 1

Table 1 Study Type Sample Type/Substrate Data Assessment Sensitivity Performance Inhibition Stability 30 saliva donors on Indicating FTA with EasiCollect 30 saliva donors on COLOR 38 saliva donors on Indicating FTA with EasiCollect 30 saliva donors x3 on COLOR with 4N6FLOQSwabs Regular 30 saliva donors x2 on Collection Device 4x Indicating FTA with PBS + 4 ng 9947 Control DNA 4x COLOR with PBS + 4 ng 9947 Control DNA 2 lots COLOR x 3 saliva donors x 4 temperatures (4 C, 22 C, 30 C, 55 C) x 5 time points (0, 2, and 6 months and simulated 24 and 72 months) 2 lots COLOR with donor sample deposited after each x 4 temperatures (4 C, 22 C, 30 C, 55 C) x 5 time points (0, 2, and 6 months and simulated 24 and 72 months) 26 and 27 cycles tested to evaluate optimum cycle number for subsequent experiments Performance comparison of Indicating FTA and the COLOR with both types of 4N6FLOQSwabs Comparison of possible inhibition due to indicating dye and other reagents by punching various locations on the Indicating FTA and the COLOR Evaluation of performance of samples deposited onto COLOR and then aged Results Sensitivity study When amplified for 27 cycles and analyzed with a 50 RFU threshold, the 4N6FLOQSwabs samples used with COLOR produced full profiles with 3 observations of off-scale alleles, while 29/30 buccal swab samples on Indicating FTA CARDS produced full profiles with 11 off-scale alleles seen. When analyzed at a peak amplitude threshold of 150 RFU, both COLOR and Indicating FTA CARDS had 29/30 samples produce full profiles. No reproducible artifacts or peaks indicative of contamination were observed for any of the two card types tested. 27 cycles was selected as the optimum cycle number, as it produced an excellent first pass success rate while minimizing null or partial profiles when compared to 26 cycle protocol. Accordingly, all samples in the subsequent experiments were amplified for 27 cycles. Results are summarized in Table 2. Performance study A total of 188 buccal samples were tested: 60 samples collected with the Collection Device, 90 samples obtained with COLOR and 4N6FLOQSwabs Regular size tip, and 38 buccal samples spotted in FTA CARDS with the EasiCollect device. 149/150 samples on COLOR and 37/38 samples on Indicating FTA CARDS produced full profiles at a threshold of 50 RFU, while 145/150 samples on Table 2 Analysis at 50 RFU Analysis at 150 RFU Observations Substrate: Indicating FTA CARDS COLOR Indicating FTA CARDS COLOR PCR cycles N samples Number of Full Profile First Pass Number of Full Profiles First Pass Homozygote Off-Scale Alleles 26 30 29/30 96.7% 28/30 93.3% 4 26 30 28/30 93.3% 26/30 86.7% 1 27 30 29/30 96.7% 29/30 96.7% 11 27 30 30/30 100.0% 29/30 96.7% 3 2

Table 3 Analysis at 50 RFU Analysis at 150 RFU Substrate: N samples Full Profile First Pass Full Profiles First Pass Indicating FTA CARDS 38 37/38 97.4% 36/38 94.7% COLOR with 4N6FLOQSwabs Regular Collection Device 90 90/90 100.0% 87/90 96.7% 60 59/60 98.3% 58/60 96.7% Figure 3 Figure 4 Figure 3. Peak height comparison of the Collection Device, COLOR and Indicating FTA Cards generated in the Performance evaluation. Figure 4. Intra-color balance of the Collection Device, COLOR and Indicating FTA Cards obtained in the Performance evaluation. COLOR and 36/38 samples on Indicating FTA CARDS produced full profiles at a threshold of 150 RFU (Table 3) Average peak heights were higher for the Collection Device as compared to the COLOR and 4N6FLOQSwabs Regular size tip and the EasiCollect device with Indicating FTA paper. No differences were observed when the samples were collected with the COLOR and 4N6FLOQSwabs Regular size tip and the EasiCollect device (Figure 3). Intra-color balance results were comparable between the three sample collection methods (Figure 4). No differences were observed in the intra-locus alleles ( COLOR Collection Device PHR = 0.912±0.085; COLOR, 0.905±0.087; FTA EasiCollect 0.907±0.070). Inhibition study 40 μl of PBS was spotted onto 4 different COLOR and 4 different Indicating FTA cards. 1.2 mm punches were taken in different areas of the card and amplified with 4 ng of 9947A to assess the possibility of inhibition from the indicating dye (Figure 5). In the area where the strongest concentration of dye is seen (Area 2), some inhibition was seen with both card types, while more allele dropout was seen with the Indicating FTA cards. No differences were observed between the COLOR and the Indicating FTA cards in Areas 1 and 3. 3

Figure 5 Figure 5. Area 1, 2, and 3 after spotting PBS on a COLOR (left) and on indicating FTA card (right). Peak height comparison between NUCLEIC CARD COLOR and indicating FTA cards after amplification with AmpFlSTR Identifiler Direct. Area 1, COLOR Area 1, from Indicating FTA card Area 2, COLOR Area 2, from Indicating FTA card Area 3, COLOR Area 3, from Indicating FTA card 4

Stability study Samples from 3 different donors were deposited onto 2 different lots of COLOR and stored for 2 to 6 months away from direct light exposure. Based on broadly-accepted accelerated stability protocols, incubation for 2 and 6 months at 55 C correspond to a real aging of approximately 24 and 72 months. Differences were seen with the average peak heights between samples and the different length of time in storage that was assumed to be normal variation seen in buccal samples for different donors (Figure 6). Even after simulated storage at 72 months, full profiles with no artifacts were generated for all cards processed. No differences were observed in terms of ICB in all the stability scenarios (Figure 7). Real time stability studies are ongoing to confirm the results obtained with the accelerated aging conditions. Figure 6 Forensic News 2013 Conclusions In this comparative study, buccal samples collected from different donors, using the Collection Device and the COLOR with the 4N6FLOQSwabs Regular size tip and amplified with the AmpFlSTR Identifiler Direct PCR Amplification kit generated full and high-quality profiles. The COLOR products consistently produced peak heights and intra-color balance that were comparable or sometimes higher than those generated from samples collected with the EasiCollect device using Indicating FTA cards. Stability tests demonstrated the robustness of the COLOR showing lysing and good profile generating capabilities across time and temperatures. Together with Copan 4N6FLOQSwabs, the COLOR and the AmpFlSTR Identifiler Direct PCR Amplification kit demonstrated to be a consistent and successful processing tool for direct profiling of buccal cell samples. References 1. http://rtsf.msu.edu/assets/files/genomics/assays_stability.pdf 2. http://www.norgenbiotek.com/product_resources/saliva_dna_collection_preservation_ and_isolation_kit_long_term_stability_of_dna_stored_in_norgens_saliva_dna_ collection_preservation_and_isolation_kit 35075_1051.pdf 3. http://www.stabilitysystem.com/arrheniusequation.pdf 4. Genes IX, Benjamin Lewin Oxford University Press (2007) 5. Oxygen effect in heat-induced DNA damage, A. V. Chernikov, S. V. Gudkov, I. N. Shtarkman and V. I. Bruskov, Biophysics 2007. 6. N. Oldroyd, J.Schroeder, (2012) Direct PCR amplification of blood samples on Copan 7. http://www.whatman.com/easicollect.aspx 8. http://www3.appliedbiosystems.com/cms/groups/applied_markets_support/ documents/generaldocuments/cms_065522.pdf How to Cite This Article Figure 6. Peak height after long term storage of the COLOR. Figure 7 Agostino A, Gervason A, Squassina A, Muehling J (2013) Direct PCR Amplification of Buccal Samples on the COLdentifiler Direct PCR Amplification Kit. Forensic News; 2013. Available from http://www.lifetechnologies.com/forensicnews 2013 Life Technologies Corporation. All rights reserved. The trademarks mentioned herein are the property of Life Technologies Corporation and/or its affiliate(s) or their respective owners.all Life Technologies products found herein are For Forensic or Paternity Use Only. NUCLEIC-CARD and 4N6FLOQSwabs are trademarks of Copan Italia S.P.A. FTA and EasiCollect are trademarks of Whatman Ltd. Figure 7. Intra-color balance after long term storage of the NUCLEIC- CARD COLOR. 5