SECTION Manufactured Metal Casework FLEXIBLE LABORATORY FURNITURE SYSTEM (OPTIMA) USA VERSION

Similar documents
SECTION (FORMERLY 11610)

SECTION (FORMERLY 11610)

SECTION RADIATION SHIELDING ENCLOSURE Low Intensity Radiation Protection

SECTION LAGUNA SERIES TOP HUNG SLIDING DOOR SYSTEM

SECTION CRL50 AND CRL51 HEAVY GLASS TOP HUNG SLIDING DOOR SYSTEMS

SECTION [ ] COILING COUNTER DOORS. Display hidden notes to specifier. (Don't know how? Click Here)

DIVISION 10 SPECIALTIES SECTION OPERABLE PARTITIONS

SECTION ACCORDION FOLDING PARTITIONS

SECTION (FORMERLY 11610)

FORM HEADWALL. FORM HEADWALL (floating) (floating) Installation Manual

SECTION HANGERS AND SUPPORTS FOR ELECTRICAL SYSTEMS

SECTION OVERHEAD COILING FIRE DOORS MODEL 662 OVERHEAD COILING COUNTER FIRE DOORS WITH INTEGRAL FRAME

FORM HEADWALL. FORM HEADWALL (floor mount) (floor mount) Installation Manual

SECTION HEAVY GLASS TOP HUNG SLIDING BI-FOLD DOOR SYSTEMS

Northpointe Renovation EPN #1701 F/A Project #17027 September 8, 2017

Section HIGH-TENSION Stretched-Fabric Wall Systems SoftWalls, Inc.

SECTION MODULAR GALVANIZED RF SHIELDING SYSTEM SPECIFICATIONS

SECTION PERSONNEL DUTY LOCKERS. 1. FPPL Series Police Personnel Duty Lockers, including the following:

SECTION DIRECTORIES

GLASS WALL SPECIFICATIONS

SECTION RADIO FREQUENCY SHIELDING ENCLOSURE RF Welded System

Architectural Specifications, Section Phenolic Lockers

GLASS WALL SPECIFICATIONS

A. Submit the following in addition to the standard requirements.

SECTION CLIMBING WALLS

SECTION TOILET COMPARTMENTS. A. Section Metal Fabrications: Concealed steel support members.

MODERNFOLD GWS 100SR-G MOVEABLE GLASS WALLS

SECTION ELEVATED BLEACHERS (Angle Frame Semi-Closed Deck)

PH6000. Key Features

Façade Architectural Wall Panel System - Guide Specification 3/11/2017 SECTION FRAMED DECORATIVE PANEL SYSTEMS PART 1 - GENERAL 1.

Acousti-Seal Encore Paired Panel (10650) Operable Partitions

ALL-ALUMINUM DOOR & WINDOW FRAMING

NORTHWESTERN UNIVERSITY PROJECT NAME JOB #

COLLEGE CLASSROOM & DORMITORY SPECIFICATION SECTION 08100

SECTION FOOD DISPENSING EQUIPMENT- SNEEZE GUARDS AND PARTITION POSTS

SECTION IMPACT RESISTANT WALL PROTECTION

Section PANEL SIGNAGE

SECTION EXTERIOR GRILLS AND SCREENS

GLASSWALL SPECIFICATIONS

The Contractor is responsible for quality control, including workmanship and materials furnished by his subcontractors and suppliers.

1. Hangers and supports for electrical equipment and systems. 2. Construction requirements for concrete bases.

Specifications for FL-100 Soffit Panel

SOLAR SHADING SYSTEMS - Manual Solar Shade SECTION ROLLER WINDOW SHADES

Series 3000 Pocket Doors

SECTION BASIC ELECTRICAL MATERIALS AND METHODS

SECTION STEEL DECKING

SECTION DECORATIVE METAL FENCES AND GATES

SECTION LIGHT DUTY SLIDING AND STACKING GLASS DOOR SYSTEMS

B. Shop Drawings: For metal lockers. Include plans, elevations, sections, details, and attachments to other work.

SPECIFICATION SECTION 05515: LADDERS O Keeffe s Aluminum Ladders

LINEAR METAL CEILINGS Section

PRESTON HEALTH SERVICES ARCHITECT'S NO. 2700

SECTION (Master Template Final-05) ELECTRICAL BOXES. A. Provide boxes for electrical equipment and wiring devices as follows:

NORTH HARRIS COUNTY REGIONAL WATER AUTHORITY PIPE HANGERS, Section PIPE HANGERS, SUPPORTS, AND RESTRAINTS

SECTION RESIDENTIAL OVERHEAD DOORS SIGNATURE COLLECTION WOOD GARAGE DOORS

SECTION GYPSUM BOARD

SECTION VIBRATION ISOLATION

PASSAIC COUNTY TECHNICAL INSTITUTE CCA 1422 NEW S.T.E.M. BUILDING 2017

SECTION GYPSUM BOARD ASSEMBLIES

SECTION ENTRANCE FLOOR GRILLES STAINLESS STEEL GRATING

SECTION OUTLET, PULL, AND JUNCTION BOXES

SECTION HEALTHCARE CASEWORK (SYNTHESIS CASEWORK COLLECTION BY MIDMARK)

BEARING METAL STUD FRAMING

Page 1 PRODUCT DESCRIPTION SECTION SECTIONAL OVERHEAD DOORS MODEL 900 PRODUCT FEATURES

SECTION PIPE AND TUBE RAILINGS PART 1- GENERAL 1.1 RELATED DOCUMENTS 1.2 SUMMARY 1.3 PERFORMANCE REQUIREMENTS

OVERHEAD SERVICE DISTRIBUTION

B. American Society for Testing and Materials: 1. ASTM F708 - Standard Practice for Design and Installation of Rigid Pipe Hangers.

SECTION Summit Phenolic Lockers (Solid Phenolic Lockers)

UNIVERSITY SERVICES ANNEX James Madison University Harrisonburg, Virginia State Project Code: Architect s Project Number:

B. Polyethylene samples shall be submitted for Owner approval of color and quality.

NORTHWESTERN UNIVERSITY PROJECT NAME JOB # ISSUED: 03/29/2017

DIVISION 15-MECHANICAL SECTION SUPPORTS AND ANCHORS PN PART 1 - GENERAL 1.01 SECTION INCLUDES. A. Pipe and equipment support systems.

Key Features. Glass Locker Series GL5000. ¼ inch tempered glass doors with soft closing hinges. Back-painted glass can be matched to any Pantone color

SECTION ACOUSTICAL PANEL CEILINGS

SECTION SECTIONAL OVERHEAD DOORS

American Metal Specialties, Inc.

University Apartments on Brooklyn PROJECT MANUAL

Division 11 - Equipment Section 11401

ACOUSTICAL WALL TREATMENT SECTION 09840

EDGEWATER, FLORIDA JUNE 14, 2010

SECTION PANEL AND POST AND PYLON SIGNAGE. 1. Non-illuminated Panel and Post signs. 2. Internal illumination Pylon (Cabinet and Post) signs.

SECTION NON-PENETRATING PORTABLE ROOFTOP SUPPORTS AND WALKWAYS. For best results, display hidden notes to specifier.

Model # 800SM-PS3. 8/3/2016 Page 1 of 8

CLINICS. LABS. PHARMACIES. PREP AND PACK.

Division MEDICAL EQUIPMENT SUPPORT SYSTEMS PART 1 GENERAL

SECTION HANGERS AND SUPPORTS FOR PLUMBING PIPING AND EQUIPMENT. A. ASME B31.1 (American Society of Mechanical Engineers) - Power Piping

A-Series Aluminum Gantry Crane

SECTION HANGERS AND SUPPORTS FOR HVAC PIPING AND EQUIPMENT

SECTION TURBINE VENTILATORS

C. Concrete Inertia Bases: Drawings showing required concrete steel reinforcing, construction details, and dimensions.

Post Style with ABS Real Tuff

2000 POCKET DOOR TECHNICAL DATA

DESIGN AND CONSTRUCTION STANDARDS. Section 2: Standards for Specific Categories of Work. Laboratory Standards

3000 SERIES POCKET DOOR

B. Shop Drawings: Include plans, elevations, sections, details, and attachments to other work.

HANGERS AND SUPPORTS FOR HVAC PIPING AND EQUIPMENT SECTION HANGERS AND SUPPORTS FOR HVAC PIPING AND EQUIPMENT

1. Vertical reciprocating conveyor, Hydraulic straddle VRC.

SECTION SLOPED TRANSLUCENT METAL SKYLIGHT SYSTEM

This product specification is written according to the Construction Specifications Institute MasterFormat, 2014 Update.

SECTION SOLID COLOR REINFORCED COMPOSITE TOILET PARTITIONS

Transcription:

SECTION 12 31 00.4 Manufactured Metal Casework FLEXIBLE LABORATORY FURNITURE SYSTEM (OPTIMA) USA VERSION PART 1 GENERAL Summary: This Specification identifies the minimum material and construction standards that are required to deliver a quality installation of the flexible laboratory furniture system. Laboratory furniture shall be supplied in accordance with the requirements of this Specification. The laboratory furniture identified in this Specification shall include the miscellaneous metal panels and other related components as identified on the drawings and that are necessary for the complete installation. 1.1 SECTION INCLUDES A. Modular Support Structure B. Structural Table Base C. Mobile Base / Wall Cabinets D. Shelves E. Fixtures F. Service Connections 1.2 RELATED SECTIONS A. Division 06 Section 10 00, Rough Carpentry B. Division 06 Section 40 00, Architectural Woodwork C. Division 11 Section 53 00, Laboratory Equipment D. Division 12 Section 36 00, Countertops E. Division 12 Section 35 53, Manufactured Metal Casework F. Division 12 Section 32 00, Manufactured Wood Casework MOTT/OPTIMA SPEC REV-4 05/13 12 31 00.4-1

G. Division 22 Section 40 00, Plumbing Fixtures H. Division 26 Section 05 00, Common Work Results for Electrical I. Related Work To Be Performed By Others: 1. Final installation of all plumbing, service and electrical fixtures attached to service carriers. 2. Final connection to service lines of all plumbing, service and electrical fixtures attached to service carriers. 1.3 REFERENCES A. SEFA 8: Laboratory Furniture Casework, Shelving and Tables Guidelines Science Equipment and Furniture Association (SEFA) B. ISO 9001:2008 Quality Management International Standards Organization (ISO) C. ADA (ATBCB ADAAG) Americans with Disabilities Act Accessibility Guidelines Americans with Disabilities Act (ADA) 1.4 SUBMITTALS Refer to Section 01 33 00, Submittal Procedures, for requirements, procedures, etc. A. Product Data: Drawings shall include data and details for construction of the laboratory furniture as well as information regarding the name, quantity, type and construction of materials (such as hardware, gauges, etc), that will be used to complete the project. B. Shop Drawings: 1. The laboratory furniture manufacturer shall furnish shop drawings illustrating the layout and placement of all laboratory furniture, casework and fume hoods as well as any products included in this section. 2. Indicate the type and location of all service fittings and associated supply connections. 3. Preparation instructions and recommendations. 4. Storage and handling requirements and recommendations. 5. Installation methods. MOTT/OPTIMA SPEC REV-4 05/13 12 31 00.4-2

C. Selection Samples: Submit the following: 1. One complete set of color chips representing the manufacturer s full range of available colors. Minimum sample size 2 inches by 2-1/2 inches (50mm x 76mm). D. Quality Assurance/Control 1. Design Data/Test Reports: Manufacturer shall submit test data and design criteria which are in compliance with the project specifications. 2. Certificates: All certifications required in the specifications shall be submitted with the original submittal package under separate cover. Certificates must be provided with the signature of a qualified individual of the supplier. 3. Manufacturers Instructions: Provide manufacturer s instructions for installation and maintenance of all products provided and installed within this section. Instructions will be in bound form, tabbed and organized by section number. 1.5 QUALITY ASSURANCE A. Manufacturer Qualifications: The following list of information will be provide to the Architect at least ten (10) days prior to the bid opening: 1. List of manufacturing facilities; 2. Construction details depicting the materials, sizes and methods of construction; B. Mock-Ups 1. Area mockups shall be as indicated on the shop drawings. Post bid mockup areas must be priced for disassembly and reassembly and used within the project. 2. Do not proceed with remaining work until installation is approved by Architect. a) Install service carrier post assembly with specified hardware. b) Install service carrier body with specified hardware. c) Install all supports and cover panels as required. 1.6 DELIVERY, STORAGE AND HANDLING A. Packaging, Shipping, Handling and Unloading MOTT/OPTIMA SPEC REV-4 05/13 12 31 00.4-3

1. Packaging: Products shall have packaging adequate enough to protect finished surfaces from soiling or damage during shipping, delivery and installation. 2. Delivery: Casework delivery shall only take place after painting, utility rough-ins and related activities are completed that could otherwise damage, soil or deteriorate casework in installation areas. 3. Handling: Care such as the use of proper moving equipment, experienced movers, etc., shall be used at all times to avoid damaging the casework. Until installation takes place, any wrapping, insulation or other method of protection applied to products from the factory will be left in place to avoid accidental damage. B. Acceptance at Site: Casework will not be delivered or installed until the conditions specified under Part 3, Installation section of this document have been met. C. Storage: Casework shall be stored in the area of installation. If, prior to installation, it is necessary for casework to be temporarily stored in an area other than the installation area, the environmental conditions shall meet the environmental requirements specified under the Project Site Conditions article of this section. D. Waste Management and Disposal: The supplier of the laboratory casework is responsible for removing any waste or refuse resulting from the installation of, or work pertaining to laboratory casework; thereby leaving the project site clean and free of debris. Trash container(s) to be provided by others. 1.7 PROJECT SITE CONDITIONS A. Building must be enclosed (windows and doors sealed and weather-tight); B. An operational HVAC system that maintains temperature and humidity at occupancy levels must be in place; C. Adjacent and related work shall be complete; D. Ceiling, overhead ductwork and lighting must be installed; E. Site must be free of any further construction such as wet work ; F. Required backing and reinforcements must be installed accurately and the project must be ready for casework installation. 1.8 WARRANTY MOTT/OPTIMA SPEC REV-4 05/13 12 31 00.4-4

A. Furnish a written warranty that work performed under this section shall remain free from defects as to materials and workmanship for a period of two (2) years from date of shipment. Defects in materials and workmanship that may develop within this time are to be replaced without cost or expense to the Owner. Defects include, but are not limited to: 1. Ruptured, cracked, or stained coating 2. Discoloration or lack of finish integrity 3. Cracking or peeling of finish 4. Slippage, shift, or failure of attachment to wall, floor, or ceiling 5. Weld or structural failure 6. Warping or unloaded deflection of components 7. Failure of hardware B. The warranty with respect to products of another manufacturer sold by Mott Manufacturing is limited to the warranty extended by that manufacturer to Mott Manufacturing. PART 2 PRODUCTS 2.1 MANUFACTURER A. Acceptable Manufacturer: Mott Manufacturing Ltd.; 452 Hardy Rd. Brantford, ON, Canada N3T 5L8. Tel: (519) 752-7825. Fax: (519) 752-2895. Email: inquire@mott.ca, www.mott.ca. B. Substitutions: 1. Must meet all specification requirements and have prior approval. 2. Must meet the minimum design and performance requirements of SEFA and UL 962. C. Requests for substitutions: All requests will be considered in accordance with provisions of Section 01 60 00. 2.2 MATERIALS A. Sheet Steel: Mild steel, cold rolled furniture grade to requirements of ASTM A1008/A1008M, Grade C or higher, with smooth surfaces to furniture quality. MOTT/OPTIMA SPEC REV-4 05/13 12 31 00.4-5

B. Galvanized Sheet Steel: Commercial quality galvanised sheet steel to ASTM 653, Designation Z275. C. Stainless Steel: 1. Sheet: ASTM A240, type 304 and 316 alloy. 2. Finish: Unless otherwise indicated, AISI No. 4 brushed finish. D. Glass: Glass, ½ thick with steel frame work to protect edges. 2.3 DESIGN REQUIREMENTS: *Basis of design: Mott Manufacturing Optima Series Table system* A. Modular system shall be made of tubular style framing combined with D shape or tubular formed steel uprights. 1. To be used in island, wall or peninsula situations. B. Tubular Frames: Table supports. 1. Table supports to be adjustable height in 1 increments. 2. Table support frames to have levelers equipped. C. Rear frame to be used for carrying services and electrical conduit. 1. Rear upright supports to be equipped with slots for adjustable shelving and levelers. 2. All services must terminate at the top of the rear tubular support frame. D. Assembled frame to be self supporting without needing to be anchored to the building. E. The modular system must ship complete from the factory with minimal on-site assembly. 2.4 2100 SERIES CONSTRUCTION A. Rear Support Structure: 1. Nominal rear frame dimensions: a) Width: 48, 60, 72, 96 b) Depth: 2" diameter c) Height: 84 MOTT/OPTIMA SPEC REV-4 05/13 12 31 00.4-6

2. Rear and Center Uprights: a) 2 outside diameter 11ga. powder coated cold rolled steel or stainless steel b) Center uprights apply to units 60, 72 or 96 wide to accommodate split shelving c) 2 diameter nylon leveling glide 3/8 x 2-1/2 long 3. Upper and Lower Cross Rails: a) 11ga. powder coated cold rolled steel or stainless steel. 4. Load Capacity: Rear Upright to support up to 3 shelves loaded to a maximum of 180lbs per 12 deep shelf. The total load capacity for the Rear Upright is 540lbs. 5. Uprights to house services, electrical and data cables. a) High voltage cabling to be in a separate upright from gas piping. 6. Lower cross rail shall house an electrical circuit raceway. 7. Rear posts have slots punched on 1 increments starting at nominal 59 above the finished floor, to the top of the post. B. Tubular Table Assembly: 1. Nominal table assembly dimensions: a) Width: 48, 60, 72, 96 b) Depth: 29 or 35 c) Height: Adjustable from 29 36 (not including work surface) 2. Tubular Table Legs: a) 2 outside diameter, 12ga. powder coated cold rolled steel or stainless steel outer leg b) 1-¾ outside diameter, 11ga. powder coated cold rolled steel or stainless steel inner telescoping leg c) 2 diameter nylon leveling glide 3/8 x 2-1/2 long 3. Capable of vertical height adjustment in 1 increments. 4. Primary table assembly shall be fastened to the rear upright with four (4) hex 3/8 socket head bolts 5. Secondary table assembly shall be a four leg table designed to mate with the contoured uprights of the main table to form the appearance of one complete assembly with two work surfaces, while remaining independent and self-supporting. Interlocking hooks shall be provided to ensure that the entire assembly is stable. It shall be possible to move the secondary table to another location as use as an independent four leg table. 6. Leveling Bolt: Frame shall be fitted with a leveling bolt which will allow the legs to be adjusted for proper alignment of work surface height. MOTT/OPTIMA SPEC REV-4 05/13 12 31 00.4-7

7. Load Capacity: Table frame shall support the work surface plus 100lbs/linear ft of table length up to a maximum load rating of 800lbs. C. Shelves: 1. Nominal shelf dimensions: a) Width: 30, 36, 48 b) Depth: 13 Top, 12 or 15 for remaining shelves c) Thickness: 1 2. Shelf requirements: a) Shelves shall be constructed of powder coated cold rolled steel, stainless steel, steel framed 6mm laminated safety glass, or a selected wood species. b) Shelves shall have a 1 overhang behind the face of the rear tubular posts. c) Shelf brackets to be constructed of 14ga. powder coated cold rolled steel or stainless steel. d) Shelves shall have a rear 1 high retaining lip. Top shelf assemblies do not require retaining lip. e) Vertical shelf adjustment in 1 increments. f) Optional 1 tall x 5/16 dia. shelf retainer rods, available in #304 stainless steel or #304 powder coated stainless steel. D. Suspended/Mobile Base Cabinets: 1. Design and construction shall be as in section 12 35 53 - Laboratory Metal Casework. 2. Mobile cabinets shall have casters in lieu of a toe kick. Casters shall be rated for 165lbs minimum each and shall be locking type. Cabinet height must ensure 2-½ of clearance under the table frame. 3. Suspended base cabinets: Provide a system of cold-rolled steel hanger rails attached to the casework frames. Installation and removal to be accomplished without the use of tools. E. Plumbing/Fixtures: 1. Rear upright structure to support a maximum of three plumbing fixtures. 2. Fixtures shall be needle valve style with a single serrated hose end. 3. Plumbing lines shall be ¼ copper tubing running the length of the upright. MOTT/OPTIMA SPEC REV-4 05/13 12 31 00.4-8

4. All burning gas tubing shall be specified as stainless steel. 5. All plumbing shall have quick disconnect at the top of the upright. 6. Plumbing shall be arranged in such a fashion that they services can not be intermixed. 7. All service valves and quick disconnects shall be keyed and color coded. Only plug and body connects of the same key will couple and allow flow. F. Service Connections: 1. Electrical, data and plumbing services shall terminate at the top of the rear support upright. 2. Electrical services shall have a 20 amp cord extending 4 above the top of the upright. 3. Data services shall have a male plug extending 10 above the top of the upright. G. Ceiling Service Panels: 1. Panels shall be compatible with most T-grid acoustical suspended ceiling structures. 2. Panel shall provide a means to mount and disconnect quick connect service fixtures, electrical and data outlets. 3. Panel shall accommodate single sided and back to back bench configurations. 4. Panels ship with cover plates. Data outlets, electrical outlets, junction boxes and service fixtures shall be ordered separately. 5. Panels shall be 23-¾ x 23-¾ x 1, 14 gauge cold rolled steel with a powder coated finish. 2.5 2500 SERIES CONSTRUCTION A. Rear Support Structure: 1. Nominal rear frame dimensions: a) Width: 48, 60, 72, 96 b) Depth: 6 c) Height: 84 2. Rear and Center Uprights: a) 2 x 6 14ga. powder coated cold rolled steel or stainless steel b) Removable end panels shall be ordered separately c) Center uprights apply to units 60, 72 or 96 wide to accommodate split shelving d) 2 diameter nylon leveling glide 3/8 x 2-1/2 long MOTT/OPTIMA SPEC REV-4 05/13 12 31 00.4-9

3. Upper and Lower Cross Rails: a) 11ga. Powder coated steel or stainless steel 4. Load Capacity: Rear Upright to support up to 3 shelves loaded to a maximum of 180lbs per 12 deep shelf. The total load capacity for a single sided Rear Upright is 540lbs. The total load capacity for a double sided Rear Upright is 1080lbs. 5. Uprights to house services, electrical and data cables. a) High voltage cabling to be in a separate upright from gas piping. 6. Lower cross rail shall house an electrical circuit raceway. 7. Rear posts have slots punched on 1 increments starting at nominal 59 above the finished floor, to the top of the post. B. Tubular Table Assembly: 1. Nominal table assembly dimensions: a) Width: 48, 60, 72, 96 b) Depth: Single sided assembly -32 or 38 (not including work surface) Double sided assembly -58 or 70 (not including work surface) c) Height: Adjustable from 29 36 (not including work surface) 2. Tubular Table Legs: a) 2 outside diameter, 12ga. powder coated cold rolled steel or stainless steel outer leg b) 1-¾ outside diameter, 11ga. powder coated cold rolled steel or stainless steel inner telescoping leg c) 2 diameter nylon leveling glide 3/8 x 2-1/2 long 3. Capable of vertical height adjustment in 1 increments. 4. Table assembly shall be fastened to the rear upright with four (4) hex 3/8 socket head bolts 5. Leveling Bolt: Frame shall be fitted with a leveling bolt which will allow the legs to be adjusted for proper alignment of work surface height. 6. Load Capacity: Table frame shall support the work surface plus 100lbs/linear ft of table length up to a maximum load rating of 800lbs. C. Shelves: 1. Nominal shelf dimensions: a) Width: 30, 36, 48 b) Depth: 14 Top, 12 or 15 for remaining shelves c) Thickness: 1 2. Shelf requirements: MOTT/OPTIMA SPEC REV-4 05/13 12 31 00.4-10

a) Shelves shall be constructed of powder coated cold rolled steel, stainless steel, steel framed 6mm laminated safety glass, or a selected wood species.. b) Shelves shall have a 1 overhang behind the face of the rear tubular posts. c) Shelf brackets to be constructed of 14ga. powder coated cold rolled steel or stainless steel. d) Shelves shall have a rear 1 high retaining lip. Top shelf assemblies do not require retaining lip. e) Vertical shelf adjustment in 1 increments. f) Optional 1 tall x 5/16 dia. shelf retainer rods, available in #304 stainless steel or #304 powder coated stainless steel. D. Suspended/Mobile Base Cabinets: 1. Design and construction shall be as in section 12 35 53 - Laboratory Metal Casework. 2. Mobile cabinets shall have casters in lieu of a toe kick. Casters shall be rated for 165lbs minimum each and shall be locking type. Cabinet height must ensure 2-½ of clearance under the table frame. 3. Suspended base cabinets: Provide a system of cold-rolled steel hanger rails attached to the casework frames. Installation and removal to be accomplished without the use of tools. E. Plumbing/Fixtures: 1. Rear upright structure to support a maximum of three plumbing fixtures. 2. Fixtures shall be needle valve style with a single serrated hose end. 3. Plumbing lines shall be ¼ copper tubing running the length of the upright. 4. All burning gas tubing shall be specified as stainless steel. 5. All plumbing shall have quick disconnect at the top of the upright. 6. Plumbing shall be arranged in such a fashion that they services can not be intermixed. 7. All service valves and quick disconnects shall be keyed and color coded. Only plug and body connects of the same key will couple and allow flow. F. Service Connections: MOTT/OPTIMA SPEC REV-4 05/13 12 31 00.4-11

1. Electrical, data and plumbing services shall terminate at the top of the rear support upright. 2. Electrical services shall have a 20 amp cord extending 4 above the top of the upright. 3. Data services shall have a male plug extending 10 above the top of the upright. G. Ceiling Service Panels: 1. Panels shall be compatible with most T-grid acoustical suspended ceiling structures. 2. Panel shall provide a means to mount and disconnect quick connect service fixtures, electrical and data outlets. 3. Panel shall accommodate single sided and back to back bench configurations. 4. Panels ship with cover plates. Data outlets, electrical outlets, junction boxes and service fixtures shall be ordered separately. 5. Panels shall be 23-¾ x 23-¾ x 1, 14 gauge cold rolled steel with a powder coated finish. 2.6 STEEL FURNITURE FINISH 1. Metal finish to be as in Appendix 1 - Laboratory Steel Furniture Finish. PART 3 EXECUTION 3.1 INSTALLATION 1. Install casework within system, align and set level with levelling devices, in accordance with shop drawings. 2. At wall locations secure wall cabinets to face of finished walls and partitions, applying selftapping screws through wall finish material into each concealed stud flange. 3. Install components to effect a secure, neat and complete installation. END OF SECTION MOTT/OPTIMA SPEC REV-4 05/13 12 31 00.4-12