Lets Talk Indicators Challenging the process Sterilization. ICAN 5 th November Peter Newson

Similar documents
In the Know about.appropriate Use of 3M Attest Rapid Readout Biological Indicators

Biological Monitoring

CERTIFICATE OF CONFORMANCE Mesa Laboratories, Inc. 625 Zoot Way, Bozeman, MT Phone: (303)

STERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation

QUALITY ASSURANCE IN AN MDRD

Hollow, porous, solid,? Testing of Steam Sterilizers

GUIDELINE C-17. Non-Incineration Technologies for Treatment of Biomedical Waste (Procedures for Microbiological Testing)

Welcome! How do I get a CE Certificate?

What Dentistry Can Learn from AAMI ST79. Best Practices for device and instrument reprocessing

FAQFrequently. Asked Questions. Education & Training. 3M Sterilization Assurance Techline

2. Describe correct actions to take to address sterilization process failures. 3. List possible causes of sterilization process failures.

Specialty Testing of Steam Sterilizers

ISO INTERNATIONAL STANDARD

Best Practice Monitoring for Sterile Processing Professionals

ISO INTERNATIONAL STANDARD

New methods to monitor steam penetration into complex medical devices (MD) using Medical Device Simulators (MDS) and Batch Monitoring Systems (BMS)

Disinfection and sterilisation

INSTRUMENT PROCESSING FROM PREP/PACKAGING TO STORAGE. January 2016

ISO INTERNATIONAL STANDARD. Sterilization of health care products Chemical indicators Part 1: General requirements

The Bowie & Dick test is used by manufacturers

Bowie Dick Helix Test

Instructions for Validation of Autoclaves

REUSABLE MEDICAL DEVICE VALIDATION

Steam Sterilization Monitoring in the New CSA Standard (Z )

Choose a category. You will be given the answer. You must give the correct question. Click to begin. Instructions for using this template.

CAN STERILE REALLY BE STERILE?: EFFECTIVE AUTOCLAVING METHODS

Standard Operating Procedure Title: Good laboratory practices (GLP) for microbiology and chemistry laboratories

ISO Part 1 and Part 2 Compliance Requirements. Cathriona O Neill

Steam Sterilizer. SA Series. for hospitals, CSSD and medical centers

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

3M s Sterilization Tech Line FAQs Sandra Velte, BA, CSPDT

GETINGE 633HC STEAM STERILIZERS. Always with you

ISO INTERNATIONAL STANDARD

Sterilization & Chemical Resistance of Healthcare Polymers

Administrative Policies and Procedures. Cross Reference: Date Issued: 2/14 Date Reviewed: 9/14 Date: Revised: 12/14 Attachment: None Page of 1 of 6

Packaging & Sterilization

Slide 1. Slide 2. Slide 3. Sponsored by Webber Training Inc. For more information contact Paul Webber

Your Sterile Processing Partner. Duraholder IPS

Steam Sterilizers. DLOV & DLOM series. more than sixty years of experience in sterilization. DE LAMA s Quality Management System is certified:

POLICY FOR THE DECONTAMINATION OF FLEXIBLE ENDOSCOPES

PRODUCT CATALOGUE. for gke Steri-Record and gke Clean-Record Chemical indicators, PCDs and accessories for cleaning and sterilization

Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar

PRODUCT CATALOGUE. for gke Steri-Record and gke Clean-Record Chemical indicators, PCDs and accessories for cleaning and sterilization

MEDICAL TRADING knows the meaning of sterilization very well: it can be

Chapter 9 Controlling Microbial Growth in the Environment. 10/1/ MDufilho

ISO INTERNATIONAL STANDARD

They can be used on washer-disinfectors, ultrasonic cleaners, surgical instruments and endoscopes. Page 5

Quality Assurance with Rigid Containers Katrina Simpson, M.A., CST, CSPDT


Endoscope Washer Disinfector Cleaning Process Challenge Devices Research & Validation

ISO INTERNATIONAL STANDARD

VALIDATION OF BUNG PROCESSOR

Kevin Lorcheim Project Engineer

European Hygienic Engineering & Design Group

gke Steri-Record Testsets and PCDs for validation, process and batch monitoring of sterilization processes

Partnership. HSG Pre & Post Vacuum - Medium Size Mobile Autoclave. Economic Line

Christina Bradley Endoscope Decontamination A Webber Training Teleclass. Christina Bradley

Comprehensive process documentation and sterile supply management. NetBox.2 and EuroSDS systems

ISO INTERNATIONAL STANDARD

8/27/2015. Learning Objectives. Purpose of Sterilization. To destroy all living pathogenic microorganisms

10 Things You Should Do Before You Validate Your Next Package

Steelco Academy Training for success

Entellus Medical Sinuscope INSTRUCTIONS FOR USE

Sterilization & Procedure for proper use of Autoclave and Statim. Thanhvan Chu Vy Tran Libby Ogro Julie Rann

UNIT Two. Lesson 1C Sterilization

Physical and Chemical Control of Microorganisms

Hot Topics in Drug Product Process Validation: A Reviewer s Perspective

EH&S. Sheet. Fact. Safe and Effective Use of Autoclaves. What are autoclaves? Factors for effective sterilization. Dry heat cycle - when to use

Guide for Cleaning, Sterilization and Storage of Introducers for Caldera Medical Implants

HISTORY OF CHLORINE DIOXIDE

BSL3 and BSL4 Autoclaves. Laboratory Line

FDA s Guidance for Industry

Annex A2. Guidance on Process Validation Scheme for Aseptically Processed Products

ISO INTERNATIONAL STANDARD


MANAGING THE CYCLE OF POTENTIALLY INFECTIOUS MEDICAL WASTE


NO 2 Gas Sterilization and Decontamination: Keeping Pace, from Start to Finish

Hygiene Monitoring. Chemical Indicators, BD-Test and Batch- Product Information. CI_01_Eng

Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy!

Reprocessing of Medical Devices: Concerns about Sharps

VHP Sterilization and Prion Inactivation

sterility assurance Prove the power and your processes

Safe to Handle? Comparing Manually and Machine-Washed Medical Devices

Validation of cleaning and sterilization of reusable medical devices

QUALITY ASSURANCE IN ENDOSCOPY

Is contamination of bronchoscopes really a problem?

Recommendations for Physical Space for Decontamination Areas

Water for Instrument Processing

SOP BIO-006 USE OF AUTOCLAVE FOR STERILIZATION OF MATERIALS AND BIOLOGICAL WASTE

PRODUCT CATALOGUE. for gke Steri-Record Biological Indicators, Process Challenge Devices and Accessories

SERIE ENT. the fastest reprocessor designed for small endoscopes

PRODUCT CATALOGUE. for gke Steri-Record Biological Indicators, Process Challenge Devices and Accessories

Microbiological Cleaning Method Validation

INTERNATIONAL STANDARDS ON FLUID PREPARATION FOR CONVECTIVE THERAPIES. Richard A. Ward, Ph.D.

A UT OCL A VE VE OCLA UT A _AutoClave_D1102EN_1703.indd /03/01 13:45

FREQUENTLY ASKED QUESTIONS IPAC Lapses

Testing of routinely sterilized products should

LARGE CAPACITY PHARMACEUTICAL AUTOCLAVES

Monitoring Flexible Endoscope Channels to Assure Cleaning

Transcription:

Lets Talk Indicators Challenging the process Sterilization ICAN 5 th November Peter Newson

Objectives In the context of the Decontamination process what is an indicator? What phases of the process do we want to monitor and in accordance with which standards? Sterilization - Steam daily machine release Sterilization Steam process, load and tray monitoring Conclusion

Indicators - monitoring of which phase of the process Cleaning is the process that physically removes soiling including large numbers of microorganisms and the organic material on which they thrive. Disinfection describes a process that eliminates many or all pathogenic microorganisms with the exception of bacterial spores. Sterilisation refers to a physical or chemical process that completely kills or destroys all forms of viable microorganisms from an object, including spores.. Organic Matter Surface If it's not clean... it can't be sterile (Spaulding) Infectious Microorganisms

Daily testing of your steam sterilizer - What do the standards say?

In the context of the EN ISO 17665 standard what is a Bowie and Dick test? The Bowie Dick test is a daily steam penetration test that is performed at the commencement of each day that the steam sterilizer is used. It is a requirement of EN ISO 17665-1 to perform this test. ( if an alternative Bowie and Dick test is used this must comply to all criteria as stipulated in EN ISO 11140-4)

Daily test - Bowie and Dick Test

Alternative Bowie and Dick tests BSI EN ISO 11140-4 11140-4 requires that alternative Bowie Dick products have equivalent performance to the towel pack Reference pack (towel pack) and alternative are tested alternately in both good pass cycles and corrupted fail cycles

VACUUM PULSE REMOVING AIR AIR AIR AIR MASS AIR

STEAM SURROUNDS THE MASS MASS

ATTRACTION OF STEAM THROUGH CONDENSATION STEAM STEAM CONDENSATE MASS STEAM STEAM

VACUUM PULSE REMOVING AIR, STEAM AND CONDENSATE AIR, STEAM & WATER MASS AIR, STEAM & WATER AIR, STEAM & WATER AIR STEAM & WATER

SUCCESSFUL PENETRATION OF STEAM TO CENTRE OF MASS STEAM MASS STEAM STEAM STEAM

Insufficient air removal, air leaks and poor steam quality AIR/GAS POCKET STEAM MASS STEAM STEAM STEAM

Four Key Requirements for a Bowie Dick Test Porosity Density Insulation Mass

Sterility Assurance The Sterility Assurance Level, or SAL, is a measure of the confidence in the attainment of sterility a probability of 1 unsterile load in 1,000,000

No. of Micro-organisms 10 6 10 5 10 4 10 3 10 2 10 1 10 0 10-1 10-2 10-3 10-4 10-5 10-6 Safety Time Assumed Bioburden of 10 6 Micro-organisms 0 1 2 3 Mins. SAL 10-6 Minimum of 1 Min. Safety Time

No. of Micro-organisms Biological indicators 10 6 10 5 10 4 10 3 10 2 10 1 10 0 10-1 10-2 10-3 10-4 10-5 10-6 Assumed Bioburden of 10 6 Micro-organisms 0 1 2 3 Mins.

Classification of indicators recent changes in thinking EN ISO 11140-1

ISO11140 - Classifications Class 1: Process indicators Class 2: Indicators for use in specific tests Class 3: Single parameter indicators Class 4: Multi-parameter indicators Class 5: Integrating indicators Class 6: Emulating indicators

Class 3 and Class 4 Class 3 indicators Prove that one or more parameters of the sterilization process such as temperature or time were present Class 4 indicators Prove that 2 or more of the parameters of, temperature, steam and time were present. Tolerances are: Time +0%, to - 25% Temperature +0, to -2C

ISO11140 - Class 5 Integrating Indicators Also known as chemical biological indicators Must react to all critical parameters of a given process Follow the death curve of a given spore population, e.g. G. stearothermophilius in steam as per ISO 11138 part 3 For steam the tolerances are; time +0%, - 15% temperature +0, -1C

No. of Micro-organisms Class 5 indicators at 134 C 10 6 10 5 10 4 10 3 10 2 10 1 10 0 10-1 10-2 10-3 10-4 10-5 10-6 Assumed Bioburden of 10 6 Micro-organisms 0 1 2 3 Mins.

Class 5 endpoint DRAIN PACK

Class 6 A Class 6 indicator would prove that all parameters of a given process were present as per values stated on the indicator. Monitoring all 3 critical parameters of Time, Temperature and Steam Cycle emulating For steam the tolerances are; time +0%, - 6% temperature +0, -1C Some examples :134 C for 3.5, 4, 5.3, 7, and 18 m - 121 C for 12, 15 and 20 min

No. of Micro-organisms Class 6 10 6 10 5 10 4 10 3 10 2 10 1 10 0 10-1 10-2 10-3 10-4 10-5 10-6 Assumed Bioburden of 10 6 Micro-organisms 0 1 2 3 Mins. SAL 10-6 Minimum of 1 Min. Safety Time

What does this mean? Class 5 endpoint Class 6 endpoint 5.3 min 134 C DRAIN PACK

Classifications and the FDA The FDA recognises Process indicators - Class 1 Air removal indicators for test packs - Class 2 Chemical integrators - Class 6 The FDA does not recognise Class 3 single parameter indicators Class 4 Multi parameter indicators Class 5 integrating indicators

In conclusion Do we need to monitor the decontamination process? Simply because it is a process with all the variables and contributing factors that can influence a process it is imperative that you consistently provide end-users and your facilities with the evidence of the efficacy of the reprocessing of your hospitals instruments Thank you very much!!!