S4 T STAR Biopharmaceutical production: Rapid nutrient and contaminant analysis of cell culture media by TXRF spectroscopy

Similar documents
Composition Analysis of Animal Feed by HR ICP-OES

Trace Element Analysis of Industrial Wastewater and Sewage with TXRF. Bruker Nano GmbH Berlin, Germany, February 23 rd, 2011

On site heavy metal testing and monitoring using monochromatic XRF methods

Core Analysis with the Tracer

S2 RANGER LE: Analysis of Light Elements in Cement, Slags and Feldspar

ANALYTICAL PERFORMANCE OF A HANDHELD EDXRF SPECTROMETER WITH MINIATURE X-RAY TUBE EXCITATION

Analysis of an OTC Multivitamin Tablet Using X-ray Fluorescence Spectroscopy and Atomic Absorption Spectroscopy

SPECTRO XRF Report. SPECTRO xsortxhh03. Analysis of Solid Metal Samples. Summary

S2 PUMA for Analysis of Geological Materials. Innovation with Integrity. Precise and Reliable Grade Control of Geological Materials EDXRF

XRF DRIFT MONITORS DATA CALIBRATION MATERIAL

Layer Thickness Analysis of Thin Metal Coatings with. Bruker Nano Analytics, Berlin, Germany Webinar, June 8 th 2017

DETERMINATION OF ELEMENTS IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS 10500, & USING THE AGILENT 5100 ICP-OES

How to Increase Productivity in Mining with Automated Simultaneous XRF Analysis. Arkady Buman and Dr. Kai Behrens Bruker AXS Inc.

Analysis of steel and its alloys using the GB/T standard and an Agilent 5100 ICP-OES in dual view mode

The image part with relationship ID rid4 was not found in the file. Welcome

New Developments in Measuring Low Silicon in Process Piping Using Handheld X-ray Fluorescence (XRF)

STUDY OF ELEMENT FINGERPRINTING IN GOLD DORÉ BY GLOW DISCHARGE MASS SPECTROMETRY

The Agilent Atomic Spectroscopy portfolio for Environmental applications AA, MP-AES, ICP-OES, ICP-MS & ICP-QQQ

Applied Comparison in Coating Analysis Micro-XRF and Handheld XRF

The Determination of Trace Elements in Stainless Steel by Forked Platform GFAAS

AUTOMATED WATER ANALYSIS

Evaluation of Energy-Dispersive X-Ray Fluorescence for Mobile Analysis of Hazardous Metals in Transuranic Waste

Analysis of Stainless Steel by Dual View Inductively Coupled Plasma Spectrometry

SPECTRAL INTERFERENCE IN X-RAY FLUORESCENCE ANALYSIS OF COMMON MATERIALS

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES

Creating Value in Mining: How XRF Delivers Accuracy and Precision in Daily Routine

Analysis of Minor Elements and Metals in Hog Manure by Field-portable Nearinfrared Spectroscopy: Results for the Zeiss Corona Spectrometer

ISO INTERNATIONAL STANDARD. Iron ores Determination of various elements by X-ray fluorescence spectrometry Part 1: Comprehensive procedure

Rapid Extraction of Therapeutic Oligonucleotides from Primary Tissues for LC/ MS Analysis Using Clarity OTX, an Oligonucleotide Extraction Cartridge

Analysis of Trace Metals in Surface and Bottled Water with the Optima 7300 DV ICP-OES

Cadmium Testing and On Site Calibration for Water Testing Detection Range: ppm

Chapter 6: Microbial Growth

Analysis of Cast Iron Using Shimadzu PDA-7000

CERTIFICATE OF ANALYSIS REVISED REPORT

New Advances using Handheld XRF Technology for the Prevention of Flow-Accelerated Corrosion (FAC)

for production of 2-[ F]-fluoro-2-deoxy-D-glucose with the

Chemical quality of household water in Bangladesh

Reduction of the Sample Size in the Analysis of Rock by EDXRF

Application Note LCMS-87 Automated Acquisition and Analysis of Data for Monitoring Protein Conjugation by LC-MS Using BioPharma Compass

Analysis of 14 Trace Elements in Stainless Steel

CERTIFICATE OF ANALYSIS

Analysis of Trace Elements in Seawater Using the Thermo Scientific icap 7000 Series ICP-OES Duo

Review on Agriculture and Rural Development vol. 2. (1) ISSN

Application Note # ET-20 BioPharma Compass: A fully Automated Solution for Characterization and QC of Intact and Digested Proteins

SCOPE OF ACCREDITATION TO ISO GUIDE 34:2009

quantitate host cell DNA with high sensitivity and throughput resdnaseq Host Cell DNA Quantitation Systems: CHO, E.

Advance the potential of your cell culture Cedex Bio and Cedex Bio HT Analyzers

High-Purity Acids & Reagents

Elemental impurities Expectations for APIs and Excipients in the EU

icap 7000 Plus Series ICP-OES Streamlined performance and ultra-low ICP-OES detection limits for routine laboratories

Nitrogen, Simplified TKN (s-tkn ) DOC

High matrix tolerance of the Thermo Scientific icap 7000 Plus Series ICP-OES with the radial Ceramic D-Torch and the sheath gas adaptor

EU Water Analysis Using the Thermo Scientific icap 7400 ICP-OES Duo

EPSILON 4 PHARMACEUTICALS & COSMETICS

Agilent 200 Series AA Systems AGILENT ATOMIC ABSORPTION SPECTROMETERS

ICP-OES. Varian Vista-MPX CCD Simultaneous ICP-OES

Agilent 5100 ICP-OES DUAL VIEW ICP-OES MINUS THE WAIT

This method is applicable to the following parameters: Nitric Hydrochloric acid digestion, Instrumental analysis.

The Determination of Trace Elements in Waters using the US EPA Method and the Thermo Scientific icap 6500 ICP-OES (Duo)

High Purity Acids Trace Elemental Analysis. Detect as low as 1 to 100 ppt

DRPK-005

GB Translated English of Chinese Standard: GB

Determination of Elemental Bad Actors (As, Hg and Si) in Petrochemical Samples Using ICP-AES and ICP-MS

National Food Safety Standard

Blank Optimization Using Ultrapure Water Suitable for Trace Ion Analysis.

Chapter 1.6. Polished Single-Crystal Silicon, Prime Wafers (all numbers nominal) Wafer Specification Table. Diameter 100 mm 4-inch 150 mm 6-inch

Application Note. Authors. Abstract

Thermo Scientific icap 7000 Plus Series ICP-OES. Gain more power. experience more performance

Assaying and Analysis Choices in Practice

The unofficial TRITON METHOD for dummies guide

ATI FLAT ROLLED PRODUCTS Brackenridge, PA for technical competence in the field of

Test sticks and test papers for semi-quantitative determinations

X-ray Fluorescence Spectrometry X-ray Fluorescence Spectrometry (XRF) is a non-destructive, quantitative technique for determining chemical

Cell Growth and DNA Extraction- Technion igem HS

Good Diffraction Practice Webinar Series

Analysis of Minor Elements and Metals in Hog Manure by Fieldportable. Spectroscopy: Results for the Textron/Case NH ProSpectra Spectrometer

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

ICP Expert software. Technical Overview. Introduction

Preparing Cell Cultures of Human Embryonic Kidney

Water Analysis Using ICP-OES with an Ultrasonic Nebulizer

The Determination of Toxic Metals in Waters and Wastes by Furnace Atomic Absorption

SUB-Programs - Calibration range Fe Base for "PMI-MASTER Pro" Spark - mode Fe 000

Trace Moisture Contamination in Ultra-High Purity Phosphine: Techniques for Measurement and Control

Total Histone H3 Acetylation Detection Fast Kit (Fluorometric)

Annex to the Accreditation Certificate D-PL according to DIN EN ISO/IEC 17025:2005

14 th Lecture Biogas and Biohydrogen

rapiflex Innovation with Integrity Designed for Molecules that Matter. MALDI TOF/TOF

Application Note 6: On-line Reaction Monitoring of Alcoholic Fermentation

INDEX DETAILED. STP292-EB/Jan. 1962

Locked Cycle Leaching Test and Yellowcake Precipitation

The Development of a Bio-inspired, Chemically Defined Media Supplement for Cell Culture

1. Bloomsbury BBSRC Centre for Structural Biology, Birkbeck College and University College London.

Convoy TM Transfection Reagent

Trace Metals in Waters by GFAAS, in Accordance with U.S. EPA and Health Canada Requirements

Method (0.1 to 8.0 mg/l Cu) TNTplus 860

Strong Acid Leachable Metals (SALM) in Soil - Prescriptive

Optical Emission Spectrometers

Analysis of Illegal Dyes in Food Matrices Using Automated Online Sample Preparation with Liquid Chromatography-Mass Spectrometry

See what s really in your sample

Water Vapor and Carbon Nanotubes

Transcription:

DMEM medium Homogenization S4 T-STAR Lab Report XRF 468 Low detection limits S4 T STAR Biopharmaceutical production: Rapid nutrient and contaminant analysis of cell culture media by TXRF spectroscopy Cell cultures offer a wide range of applications in research and industry. In particular mammalian cells are used in most industrial cell culture systems. Cell culture Biopharmaceuticals (pharmaceutically active proteins and nucleic acids) are produced by mammalian cell culture systems. In research, tissue cells can be cultured for therapeutic purposes. Elementary cellular processes can be studied on cells without having to carry out classical animal experiments. Total Reflection X-Ray Fluorescence (TXRF) spectroscopy is a well established method for the analysis of biological and environmental samples [1]. However, a systematic investigation of cell culture media has not been published yet. This lab report describes different preparation methods for cell culture media (CCM) and the rapid and accurate TXRF element analysis of nutrients and contaminants. Fig. 1 Cell cultures in research

Mass Spectra Cell culture media In order to ensure the function of all physiological processes of the cell culture also ex vivo, the natural environment of the cells has to be imitated in defined media. Standardized media, such as Dulbecco s Modified Eagle s Medium (DMEM), are used for cell cultures. Cell culture media (CCM) contain amino acids for protein biosynthesis, vitamins and inorganic salts for metabolic processes and carbohydrates such as glucose (typically 1 g/l or 4.5 g/l) for energy production. Nutrient elements (P, Zn, Cu, Fe) as well as trace elements (Se, Mn) serve as nutrient sources for the cells or are required to maintain the membrane potential and the osmotic pressure. Some ions also serve as co-factors in enzymatic reactions. In order to achieve reproducible results with a cell line, the composition of the medium must always be kept constant and free from contamination. Nutrients and trace elements are fed via highly concentrated feed solutions. Small fluctuations in the added components and minimal contaminations may reduce the yield of the biopharmaceutical product and, in the extreme case, leads to necrosis of the cells. Objective The aim of the present study was to develop a preparation method for the rapid and accurate element TXRF analysis of cell culture media (CCM) in order to control nutrient elements and to identify undesirable contaminants. Sample preparation DMEM The measurement of cell culture media with high sugar concentrations requires the development of a dedicated preparation method. Commercially available DMEM (1 g/l or 4.5 g/l glucose) was used to optimize the preparation and to determine reproducibility and detection limits for trace elements. To optimize the preparation of sample layers on quartz carrier discs different smoothing agents (1% Triton X-100, ultrapure ethanol, or 0.3% polyvinylalcohol PVA) were added to the DMEM. Four different sample series with cell culture media dilutions from 1:5 up to 1:100 of the media were prepared and tested. Procedure 1: In a micro reaction tube 1 ml of a mixture with the following components were prepared: DMEM: 10 to 200 µl Smoothing agent: 100 µl Internal standard Sc: 10 µl Filling up with water to 1 ml (resulting in dilution factors from 1:5 to 1:100) After careful homogenization 10 µl of each sample were prepared on a quartz disc. The samples were dried on a heating plate at 30 C. Procedure 2: In a micro reaction tube 1 ml of a mixture with the following components were prepared: DMEM or HP-CCM: 700 µl Ethanol: 290 µl Internal standard V (1 g/l): 10 µl After careful homogenization 10 µl were transfered to a quartz disc. The samples were dried on a heating plate at 30 C. HP-CCM Modern high performance cell culture media (HP-CCM) are used for mammalian cells of the following types: Human Embryonic Kidney (HEK) Chinese Hamster Ovary (CHO) Baby Hamster Kidney (BHK) Invertebrate (IKZ) In high performance cell culture media the amount of glucose is significantly higher, e.g. 8 g/l or 20 g/l in feed media, respectively. The HP-CCM samples were prepared according procedure 2. Spiked HP-CCM For an accurate and reliable quantification of trace metals in the low ppb range spiked samples were prepared. Three micro reaction tubes with 1 ml of a HP-CCM were each spiked with five concentrations of different elements given in Table 1. Table 1 Element concentrations of the prepared solutions for spike experiments in µg/l. Pb Mn Se Ni Cr Cd 5 20 5 10 15 300 10 50 10 20 30 500 20 100 20 50 50 1000 50 200 50 100 100 2000 100 500 100 200 200 4000

Measurements Diluted CCM on quartz carrier discs All measurements were performed for 1000 s with a S4 T STAR spectrometer at the conditions shown in Table 2. The spiked samples were measured with Mo-K (Mn, Se, Pb, Ni), W-L (Cr), and W-Brems (Cd) excitation. Table 2 Measurement parameters of the S4 T STAR X-ray tube 1 Mo target, 50 kv, 1000 µa X-ray tube 2 W target, 50 kv, 1000 µa Excitations Detector Measurement time Mo-K 17.5 kev; W-L 8.4 kev XFlash SDD, 60 mm 2, energy resolution < 145 ev 1000 s Results for DMEM Selection of the smoothing agent Cell culture media contain high amounts of sugars and other nutrients. During the drying process of such a medium on a TXRF carrier unwanted crystallization may occur. The formation of a homogenous flat sample layer can be achieved by addition of a smoothing agent like Triton X-100, ethanol or polyvinylalcohol. In a first experiment the effect of different smoothing agents was compared. After dilution with water and drying of the DMEM on quartz sample carriers the formation of large crystals was observed in an optical microscope (Fig. 2, top). The addition Fig. 2 10-fold magnified images of DMEM on quartz sample carriers after dilution with water (top) and ethanol (bottom). Comparison of smoothing agents Fig. 3 Recovery of the nutrient elements K and Ca (measured with Mo-K excitation) and S and P (measured with W-L excitation) in DMEM with different smoothing agents and dilution factors.

of ethanol as smoothing agent leads to smaller crystals and a more homogeneous layer (Fig. 2, bottom). The TXRF analysis of CCM at different dilutions from 1:5 to 1:100 does not provide satisfying recovery rates in case of water (Fig. 3). The use of 1% Triton X-100 or polyvinylalcohol as smoothing agents is well established for sample preparation and leads to better recovery rates. Comparing all smoothing agents and all dilutions, the DMEM sample prepared with 1:10 Ethanol delivers the most accurate results with standard deviations typically better than 5% (< 10% at 1:100). While the Mo-K excitation works well for the quantification of Ca and P, only the W-L excitation in combination with the smoothing agent Ethanol shows reproducible results for the light elements P and S. Fig. 4 demonstrates the improved reproducibility after dilution of the high glucose media. Typically, standard deviations were better than 3% for nutrients and better than 10% for traces at low ppb concentrations. Limit of quantification Since ethanol has proven to be the most suitable smoothing agent providing the best results for the quantification of nutrient elements, the preparation procedure 2 was also applied for the DMEM samples used for the determination of the limits of detection (LOD) and quantification (LOQ). The results are shown in Fig. 5 and given in Table 3. Limits of detection and quantification During these experiments scandium was selected as internal standard element (see procedure 1). Due to the overlap of Sc with Ca, the following LOQ and reproducibilty experiments were performed with DMEM samples prepared according to procedure 2 with vanadium as internal standard. Reproducibility The reproducibility for the measurement of nutrient elements was proven with the two different methods described in the Sample preparation section: Procedure 1: dilution factor 1:10, 100 μl sample + 100 μl ethanol + 790 μl H 2 O + 10 µl internal standard Sc Procedure 2: no dilution, 700 μl sample + 290 μl ethanol + 10 µl internal standard V. Reproducibility Fig. 5 Limits of detection (LOD) and quantification (LOQ) for nutrient elements in cell culture media. Table 3 Limit of quantification (LOQ) of nutrient elements in DMEM in mg/l. Element LOQ Magnesium* 122.7 Phosphor 4.165 Sulfur 1.205 Potassium 0.200 Calcium 0.122 Iron 0.016 * measured with W-L excitation Fig. 4 Reproducibility of the measurement of nutrient elements in DMEM. The acceptable recovery limit of ± 20 % is defined with the yellow frame. Although the matrix content of the media induces a higher background, a low LOQ value of 16 µg/l for a trace metal like iron could be achieved. For the measurement of light elements the W-L excitation was tested the first time. While the Mo excitation did not provide reliable data for Mg, the use of W-L excitation leads to a LOQ of about 123 mg/l.

Results for high performance media (HP-CCM) High performance media samples for the growth of mammalian cell cultures were prepared according to procedure 2 and measured with the S4 T STAR using Mo-K excitation at 17.5 kev and W-L excitation at 8.4 kev. Between two and eight samples of each cell culture type (HEK - Human Embryonic Kidney, CHO - Chinese Hamster Ovary, BHK - Baby Hamster Kidney, and IKZ - Invertebrate) were analyzed. Various micro nutrients and contaminants were detected in all media in concentrations from 1 ppb to about 10 ppm as displayed in Fig. 6. While the trace element concentrations within one cell line were almost identical (data not shown), significant variations between the cell lines were observed. Therefore, all media are clearly distinguishable by the amount of trace elements. Limits of detection Fig. 7 shows the detection limits for several nutrient and trace elements for two different high performance cell culture media (CHO and feed) and the two applied excitation modes. HP-CCM contain high amounts of sugars, in case of the CHO medium about 8 g/l. Nonetheless, the ethanol preparation method (procedure 2) leads to limits of detection down to 1.5 ppb for CHO and Mo-K excitation. Even feed media with 20 g/l glucose could be analyzed efficiently with TXRF down to 7 ppb. High performance cell culture media Fig. 6 Concentration of micro-nutrients and contaminants in different types of high performance cell culture media. Detection limits Fig. 7 Detection limits of micro-nutrients and contaminants in CHO and feed of high performance cell culture media.

While the common X-ray tube with Mo target is most suitable to detect the elements in the range from Mn to Y, W-L radiation efficiently excites light elements. Using W-L excitation the detection limits for the element range from Mg to Cr can be improved by a factor of 2 or more as shown in Fig. 7. Due to the sensitivity of the S4 T STAR TXRF spectrometer, for all possible contaminants measurement times (here: 1000 s) can be significantly reduced to increase the sample throughput. Results for spike experiments To detect trace metals in the low ppb range with high accuracy, spiked HP-CCM samples were prepared and measured with the S4 T STAR (see Table 1). For this series of experiements the element Scandium was chosen as internal standard. A typical spectrum of a HP-CCM spiked with Cr and measured with the W-L excitation is shown in Fig. 8. In addition to Cr (spike concentration 100 µg/l) traces of Mn, Fe, Cu, and Ni were detected. The comparison of measured values with spiked concentrations for the element Mn is demonstrated in Fig. 9. The original Mn concentration in the medium can be calculated by the intercept of the y-axis. The concentrations of all elements quantified by the spike experiment are summarized in Table 4. For Ni and Se the results of both TXRF measurements are similar and at the same order of magnitude as the ICP values. In case of Mn the spike experiment is much closer to the ICP data. The ICP result for Cr is much lower than the value of both TXRF measurements and is therefore highly doubtful. Table 4 Comparison of TXRF measurements in triplicate of a HP-CCM with TXRF results of the spike experiment and ICP-MS data. Element TXRF (µg/l) TXRF Spike (µg/l) ICP-MS (µg/l) Cr 24.7 14.7 1.9 Mn 15.8 4.4 2.6 Ni 5.3 7.7 2.1 Pb 7.6 17.0 Se 3.4 3.6 5.2 Spectrum of a cell culture medium Fig. 8 Typical spectrum of a HP-CCM spiked with 100 µg/l Cr and measured with W-L excitation.

Spike experiment Fig. 9 Calculation of the Mn concentration in HP-CCM by comparison of measured values (blue) and spiked concentrations (red). Conclusion Cell culture media are of great importance during the production of biopharmaceutical products. Up today the optimal supply with micro-nutrients and the impact of contaminants is not fully understood. The systematic investigation of nutrients and contaminants in cell culture media by TXRF is described in this lab report for the very first time. A rapid sample preparation method for TXRF measurements of cell culture media with high sugar concentrations of up to 20 g/l was developed. High purity ethanol was used as smoothing agent for optimal sample layer formation on quartz sample carriers. The measurement of nutrient elements after a 1:10 dilution provides a reproducibility < 10%. For the control of contaminants, detection limits in the one digit ppb range can be achieved. All measurements were performed with different excitation methods of the advanced TXRF spectrometer S4 T STAR. While Mo excitation is suitable for most metals, a W-L excitation improves the detection of light elements up to Cr by a factor of two or more. In addition to the outstanding analytical performance the S4 T STAR is a plug and play system with low operation costs dedicated for research, routine or QC laboratories.

All configurations and specifications are subject to change without notice. Order No. DOC-L81-EXS068. 2018 Bruker Nano GmbH. Literature [1] ISO 18507 Surface chemical analysis Use of Total Reflection X-ray Fluorescence spectroscopy in biological and environmental analysis, First edition, 2015-07-15. Authors Sophia Tessaro, University of Applied Sciences, Berlin, Germany Armin Gross, Global Product Manager TXRF, Bruker Nano GmbH, Berlin, Germany Bruker Nano GmbH Berlin Germany Phone +49 (30) 670990-0 Fax +49 (30) 670990-30 info.bna@bruker.com Bruker AXS Inc. Madison, WI USA Phone +1 (608) 276 3000 Fax +1 (608) 276 3006 info.baxs@bruker.com www.bruker.com www.s4tstar.com