Calcium-EDTA Tutorial for the MicroCal ITC200

Similar documents
Water-Water Titration for the MicroCal ITC200

Human DNA Alu Amplification by Polymerase Chain Reaction (PCR)* Laboratory Procedure

DNA RESTRICTION ANALYSIS

minipcr TM Genes in Space Food Safety Lab: Mars Colony at Risk!

Purpose: Enrich ISPs, perform second Qubit assessment, Load Ion Torrent chip.

User Guide BLF-7510 BUF High-Pass TM DNA Size Selection. Sage Science Part Nos:

PORTABLE ph METER PRICE CODE NUMBER MODEL 100

Affymetrix Processing of ChIP or DamID DNA to GeneChip Drosophila Tiling 2.0R Array

Calcium and Magnesium; Chlorophosphonazo Rapid Liquid Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Pour-Thru Cell

USER MANUAL. ExStik CL200A. Waterproof Total Residual Chlorine Tester. Patent Pending. Additional User Manual Translations available at

LAB 6: Agarose Gel Electrophoresis of Restriction Digested Plasmid DNA

HT RNA Pico Sensitivity LabChip Kit LabChip GX/GXII User Guide

Technical Bulletin No. MK3176. PREPARED: July, 2010

SCIEX Anion Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide

Agilent RNA 6000 Pico

EXPERIMENT GENOMIC DNA ANALYSIS

An Oxidation-Reduction Titration: The Reaction of Fe 2+ and Ce 4+

To whom it may concern, February 21, 2011

1D^2 Sequencing Kit (SQK-LSK308) Pipettes P2, P10, P20, P100, P200, P1000 Freshly prepared 70% ethanol in nucleasefree

Please read and follow the instructions below carefully to ensure proper setup and avoid problems.

ExcelGel 2-D Homogeneous Instructions

Seeplex LabChip Kit LabChip Dx User Guide

Let s Move It! Gel Electrophoresis Using Food Dye Student Guide

Industrial flue gas probes. Instruction manual

Experion RNA StdSens Analysis Kit Instruction Manual

HT RNA LabChip Kit, Version 2 LabChip GX/GXII User Guide

Agilent RNA 6000 Nano Kit Guide

RAINBOW GELS: AN INTRODUCTION TO ELECTROPHORESIS. STANDARDS 3.1.7, , Westminster College 3.3.7, , 3.3.

Agilent Protein 230. Kit Guide. Agilent Technologies

ecosphere water purifier 3in1 USER MANUAL

UCI KCCAMS Facility. Leaching/hydrolysis protocol (carbonate samples) December 26, 2011

Reagent Kit Selection Guide

Maxwell RSC Whole Blood DNA Kit

UltraClean Standard Mini Plasmid Prep Kit Sample (Catalog No S)

Library Loading Bead Kit (EXP-LLB001) Agencourt AMPure XP beads Vortex mixer. Freshly prepared 70% ethanol in nucleasefree

edna PROTOCOL SAMPLE COLLECTION Caren Goldberg and Katherine Strickler, Washington State University Revised November 2015

DNA Profiling with PCR

TWO-DIMENSIONAL GEL ELECTROPHORESIS THE FIRST DIMENSION

Owners Manual Models: 200-RODI4-XXX. US Water Systems 4 Stage RO/DI System Installation and Maintenance Manual.

Electrophoresis 101 Student Worksheet

edna PROTOCOL SAMPLE COLLECTION Caren Goldberg and Katherine Strickler, Washington State University Revised January 2017

Premium Water Filtration Systems

Revised 27 Aug rm (Vers. 4.1)

THERMAL CONDUCTIVITY TESTS OF LIQUIDS USING FOX50 HEAT FLOW METER INSTRUMENT (Amendment the FOX50 Instrument Manual)

Georgia Bureau of Investigation - Division of Forensic Sciences DNA Procedures Manual DNA QUANTIFICATION USING QUANTIFILER HUMAN / Y HUMAN MALE KIT

IDENTIFYING UNKNOWN SUBSTANCES

Continuous Protein Production Protocol for Wheat Germ Cell-free Protein Expression by Dialysis Method

DNA Extraction Kit. Instructions For Use (IFU) English

Maxwell RSC Blood DNA Kit

Protein-Pak Hi Res HIC Column and HIC Protein Standard

BIOLOGY 163 LABORATORY. RESTRICTION MAPPING OF PLASMID DNA (Revised Fall 2017)

Maxwell RSC Blood DNA Kit

Library Loading Bead Kit (EXP-LLB001) Agencourt AMPure XP beads Vortex mixer. Freshly prepared 70% ethanol in nucleasefree

Ligation Sequencing Kit 1D (SQK-LSK108) Ice bucket with ice NEBNext End repair / da-tailing Module (E7546)

FemINDICAtor qpcr Plant Gender Detection Kit on the BIO-RAD CFX96 Touch Real- Time PCR Detection System Page 1 of 11

Special Specimen Collection Procedures for Coagulation Testing

MODEL BE-100HT HEADSET TELEPHONE OWNER S MANUAL PLEASE READ THIS INSTRUCTION MANUAL CAREFULLY.

Odyssey Conductivity & Temperature Logger.

SOP: Applikon ez-control Bioreactor Controller Operation

4. The Second-dimensional SDS-PAGE (vertical) Protocol

Applied Biosystems QuantStudio 12K Flex Real- Time PCR System: OpenArray Plate

Alkaline Lysis Large Scale Plasmid Preparation

Mycoplasma bovis Staphylococcus aureus Streptococcus agalactiae Prototheca

Cadmium Testing and On Site Calibration for Water Testing Detection Range: ppm

Library Loading Bead Kit (EXP-LLB001) NEBNext FFPE Repair Mix (M6630) Magnetic rack. NEBNext End repair / da-tailing Module (E7546)

Thank you for choosing bq. This manual will help you use your new 3D printer correctly. Welcome to the world of Witbox!

How to Set Up and Run Gel Electrophoresis

Fast, easy, reliable, ultra-pure transfection grade plasmid DNA Maxiprep using a microcentrifuge spin-column.

Instruction Manual ODY-1765

Performing the CTA Carryover PVT Procedure on the LXi or DXC 600i System

Human Myostatin, ELISA Kit (MSTN)

Standard Operating Procedure RIB Operation and Safety Boat Duty

mag mini kit Description Kit uses

Recommended Procedure for Sampling Gastrointestinal Flexible Endoscopes

Guidelines for QualiTru Line Sampling

The Biology CRO. Protein Staining System. For electric staining of proteins in mini polyacrylamide gels. Version 1.0 April 3, 2010.

AFFINITY HIS-TAG PURIFICATION

N Scale Double Track Lift Bridge v1.2

Cells and Tissue DNA Isolation Kit (Magnetic Bead System) 50 Preps Product # 59100

Introduction. Principle

Lab 5: Protein electrophoresis

Reagent Kit Selection Guide

E.Z.N.A. Cycle Pure Kit

LIFluor EnhanCE. for use with dsdna 1000 and dsdna 20,000 Kits

ab Plant Chromatin Extraction Kit


MOLECULAR WEIGHT DETERMINATION & BY ELECTROPHORESIS and MALDI-TOF MS

For in vitro Veterinary Diagnostics only. DNA Extraction and PCR Detection Kit for Pasteurella multocida.

Lab 1 Flow Chart : Learning basic laboratory skills

Single-Use RNA Preparation

Instructions for Use Reprocessed Balloon Inflation Devices

DNA Size Selection System Operations Manual

HT RNA LabChip Kit User Guide

Title: Hemoglobin HemoCue Hb 301 method

3. Add 0.4 ml of. 7. Use a TenSette

Cells and Tissue DNA Isolation 96-Well Kit (Magnetic Bead System) Product # 62500

EASY-Spray Columns. Guidance for column set up and installation Tips to maximize column lifetime

TREE I.V. Micro-Infusion System

Instruction Manual HT 24 BM-D2400 Homogenizer

DNA Size Selection System Operations Manual

Transcription:

Calcium-EDTA Tutorial for the MicroCal ITC200 Getting Started 1. The computer and the ITC200 are normally left on. If they have been switched off for any reason makes sure that you boot them in the order shown below. If you don't, you may encounter communication problems between computer and ITC. Turn on the ITC 200 computer. Turn on the Micro-Cal ITC200. Launch the ITC 200 with Origin software. 2. The software will establish a connection with the instrument, and the instrument will then heat the cell to 30 C. If you are planning on using a different temperature, change the temperature setting right away. This will save you time later. Under the Instrument Controls tab (see figure below) change the Set Point temperature and click "Set Jacket Temp". Solutions Needed For This Experiment Name Composition EDTA solution 0.4 mm EDTA, 10 mm MES ph 5.6 Calcium solution 5 mm Calcium Chloride, 10 mm MES ph 5.6 MES buffer ph 5.6 10 mm MES ph 5.6 3. Please follow the separate instructions to make the solutions. You will only get good data if your concentrations are precisely what you think they are and if all solutions have perfectly matched buffer. 1

2

Loading Sample Cell and Syringe Preparing the ITC 200 for sample loading 1. Remove the white cap from the cell and insert the cleaning apparatus into the sample cell. Make certain that the cleaning apparatus is pushed firmly into the sample cell. 2. Attach the fill adapter to the injection syringe. To do this the syringe should be firmly seated in the wash position. Align the inner and outer ring so that you can see the hole the connector is going to be screwed into. Stabilize the aligned assemblies with your fingers and screw the connector in. It should fit snugly. If it is too loose the syringe dry might not work correctly leaving Methanol in the syringe and the sample will not be drawn into the syringe right. If you over tighten it, you will break the syringe! Note: This is how the instrument should have been left by the previous user) 3

3. Click on the Instrument Controls tab. (See Figure above) 4. Under the Instrument Controls Washing Module, click on the Cell and Syringe Wash button. The Wash Module will first flush the cell with water. Once it is done with the cell it will sound a small beep. After that it will flush the syringe with water and methanol and dry it with air. Loading the Sample Cell: 5. After the beep, while the syringe wash is going on you can start filling the cell. 6. Use the dedicated loading syringe to fill the cell. It has a protective sleeve over the needle to avoid scratching the sample cell. Do not use any other syringe!!!! 7. Fill the sample cell with the EDTA solution. To fill the cell slowly, draw up ~300μl of EDTA solution into the loading syringe. Avoid air bubbles. If you inject air bubbles into the sample cell your experiment will not work! Slowly insert the syringe into the cell until the tip gently touches the bottom of the cell. Pull the syringe up slightly, so that it is no longer pressing against the bottom. Begin to inject the EDTA solution into the cell. Once you have injected ~150 μl into the cell, inject the rest of solution in short pulses to dislodge any bubbles that may be clinging to the walls of the cell. Continue in this manner until you see the EDTA solution well up around the syringe. 4

Draw the syringe out of the cell and gently place it on the lip between the plastic collar and metal of the cell. Draw up any excess liquid. Soak the cell in the EDTA solution for 2 minutes. This allows any water remaining on the cell walls to diffuse into the EDTA solution. 8. Remove the EDTA solution from the sample cell using the loading syringe and discard the solution. 9. Refill the sample cell with fresh EDTA solution in the same way as described above. Loading the Titration Syringe 10. Fill a 200 μl PCR tube with 80 μl titrant (5 mm Ca solution). Do not close the lid. Insert the filled, open tube into the tube holder. Make sure it sits all the way down with the lid fitting into the slot provided (see photo below). 11. To fill the syringe, go to the Instrument Controls tab and click Syringe Fill. 12. Follow the prompts. You will be asked to place the syringe pipette into the rest position. After that the plunger of the syringe will be moved down. Next you will be asked to place the syringe into the loading position. Be careful when doing this and make sure that the needle does not get bent when inserted into the PCR tube with the titrant. 5

Next, the washing module will fill the syringe. Observe the process carefully. If a large air bubble remains on the top of the solution you can restart the loading process but leave the syringe in the loading position for all steps. This will most likely eliminate the bubble. 13. Disconnect the fill adapter from the syringe. 14. Insert the syringe pipette into the cell. Rest position PCR tube with titrant Load position 6

Setting Run Parameters 15. Click on the Advanced Experimental Design tab and enter the parameters shown below in the indicated fields. Don't forget to change the filename. 7

16. Click the Setup tab and change the path for data storage to your folder. Note: If you want to save time you can set the temperature of the cell right after you loaded it to the desired start temperature. To do this click Instrument controle and set the jacket temperature to your desired value. Starting the Run: 17. Switch off the overhead light if you used it. 18. Double check that the fill adapter is disconnected and press the start button. 19. The system will adjust to the chosen temperature. 20. Once the temperature is reached, the inner assembly of the titration syringe will begin to spin. 21. The system will adjust the reference power. 22. Once the reference power is stabilized the run will begin. Note: If it takes much longer than 10 min to adjust the reference power, something might not be quite right. 23. Check that you switched off the overhead light (it will heat the cell from above, making it harder to stabilize the reference power). 8

24. You can consider reloading the cell, since air bubbles might affect the stabilization of the reference power. To do that, stop the run, return the syringe pipette into load position, and remove the sample from the cell using the sample loading syringe. Reload the cell, following the procedure above. Clean Up After the Run: 25. Remove the syringe pipette from the cell and insert firmly into the wash position. Note: 26. Attach the fill adapter to the syringe pipette as described above. 27. Remove the solution from the cell using the loading syringe. You can keep the solution to double check the concentration. 28. Rinse the cell with 3 x300 ul water using the loading syringe. 29. Rinse the cell with 5% Contrat. 30. Rinse again with 3x water. 31. Insert the long needle of the cell cleaning apparatus into the sample cell and push down carefully until the O-ring has sealed. 32. Go to the instrument control tab and perform a "cell and syringe wash". If you have very high protein concentration, liposomes, or you suspect that your sample could have precipitated please perform a detergent soak and rinse by click the "detergent soak and rinse" button and following the prompts. After the detergent soak and rinse is complete and the cell has cooled down, remove the detergent solution. Rinse the cell with 3 x300 ul water using the loading syringe. Insert the long needle of the cell cleaning apparatus into the sample cell and push down carefully until the O-ring has sealed. Go to the instrument control tab and perform a "cell and syringe wash". 33. Leave the instrument in this configuration. 9

If you run into any problems or have questions about ITC let Dr.Erbse know. Annette Erbse, Office C316, phone 2-0528, erbse@colorado.edu 10