MEDICAL DEVICE REGULATION AND ITS COMPARISON IN EUROPE, AUSTRALIA AND INDIA

Similar documents
Medical Device Regulatory Framework 9 SEPTEMBER 2015 FUNDISA CONFERENCE JANE ROGERS

First Global Forum on Medical Devices Bangkok-Thailand, September Regulatory Control of Medical Devices in Tanzania

GUIDANCE NOTES FOR MANUFACTURERS OF CLASS I MEDICAL DEVICES

Due diligence in the European medical devices industry

Regulation of software a medical device

Ready or Not: The New Medical Device Regulations Are Here!

Laboratorio di Tecnologie Biomediche

In the huge expanse of Asia, Singapore and

ASQ Tappan Zee Section Medical Devices, New Regulations and Standards. 26 th September / V. Fischer / Rev. 01

MEDICAL DEVICE GUIDANCE

Medical Device Regulation Overview

Guide for Class I Manufacturers on Compliance with European Communities (Medical Devices) Regulations, 1994

HOW TO APPLY FOR MEDICAL DEVICE REGISTRATION UNDER MEDICAL DEVICE ACT 2012 (ACT 737)

MEDICAL DEVICE CLINICAL INVESTIGATIONS AND ISO 14155

Post Market Surveillance (including PMCF): common non compliances

Guide for Manufacturers and Sponsors on Clinical Investigations Carried Out in Ireland

The challenges of software medical device regulation.

MEDICAL DEVICE. Technical file.

Medical Devices. LATVIA LAWIN Klavins & Slaidins

2014 India Medical Device Updates

Medical Devices Regulation (MDR) Readiness Review

Guide for Ethics Committees on Clinical Investigation of Medical Devices

TERMS AND DEFINITIONS

Update on the IVDR. Sue Spencer

Update on Regulatory Environment- Europe Experience with 2007/47/EC M5 & Discussions on Possible Recast of EU Medical Device Regulations

Post market Surveillance ISO EU Medical Device Regulation

AUSTRALIAN MEDICAL DEVICES GUIDANCE DOCUMENT NUMBER 5. The Declaration of Conformity

Medical Devices. The technical requirements also apply when the submitter is not the legal manufacturer 1 (i.e: a distribution company).

COUNTRY OVERVIEW: THAILAND. August 2012 SPECIAL REPRINT. By Rarana Phanudulkitti

Surveillance and Medical Devices

FINAL DOCUMENT. Global Harmonization Task Force (revision of GHTF/SG1/N29:2005)

EU MDR Timeline. Dr. Christian B. Fulda Jones Day FDLI Annual Conference Access materials at fdli.org/annual2018

Challenges for South African Medical Device Manufacturers

Journal home page: RESEARCH ARTICLE

Regulatory Framework in Japan - Past and Future - Building a Regulatory Framework Essential Elements of Compliance / Surveillance

The EU s In Vitro Diagnostics (IVD) Regulation:

Korean Medical Devices Regulations

Guide to field safety corrective actions for medical devices and in-vitro diagnostic medical devices

Checklist for the assessment based on the standards

Inspection of the conduct of clinical evaluations on medical devices in the premises of healthcare providers

MDR. Device Classification Conformity Assessment Safety & Performance Requirements Technical Documentation

Di Renzo Regulatory Affairs ROME - MILAN - LONDON

Regulations and Directives Past, Present, Future

Medical Devices: They Matter Too!

VIGILANZA E SORVEGLIANZA POST- COMMERCIALIZZAZIONE

A STUDY ON VARIATIONS IN PHARMACEUTICAL PRODUCTS IN PHILIPPINES AND VARITAION POLICIES IN US, CANADA, AUSTRALIA

Medidée Services SA. Nano-Tera.ch. 05 February 2015 part 8. PMA, 510k, IDE. Pierre-Alain Sommer

Regulations & Guidelines Specific to Ethics Schedule Y & CDSCO-GCP

Medical Device Epidemiology Introduction

GUIDELINES ON MEDICAL DEVICES CLINICAL EVALUATION: A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES

White Paper: Improvements to the Australian Regulatory System for Medical Devices. 23 May 2014

Changes to the Medical Devices Directive and affect on Manufacturers

The European Medical Technology Industry in figures

UNANNOUNCED EU MEDICAL DEVICE AUDITS BY NOTIFIED BODIES: IMPACT ON SUPPLIERS

Application of ISO/IEC 17011:2004 in the Field of Medical Device Quality Management Systems (ISO 13485)

The Medical Device Coordination Group: a new Authority Under EU Device Regulations

WHO Prequalification of In Vitro Diagnostics Programme

Recommendations on Medical Device and IVD Field Safety Corrective Actions and Recalls using Unique Device Identifiers & GS1 Standards

Implications of the new MDR from a Product Testing and Certification Perspective

MEDICAL DEVICES SECTOR

QUESTIONNAIRE FOR COMPETENT AUTHORITIES TO BE USED FOR ASSESSMENT, REASSESSMENT AND SELF- EVALUATION

Regulatory and ethical requirements in medical device studies. Finland

REGULATORY LABELLING REQUIREMENTS AND ITS COMPARISON IN USA AND CANADA

DRAFT MEDICAL DEVICE GUIDANCE DOCUMENT

Flexible, robust solutions from BSI. An In Vitro Diagnostic Notified Body. Expertise and experience. IVD regulatory solutions

IVD Regulation 2017/746

Copyright. Jeremiah J. Kelly (2015). All rights reserved. Further dissemination without express written consent strictly prohibited.

Conformity Assessment of Own Brand Labelling

An Introduction to the Worldwide Regulatory Framework for Medical Devices. Elizabeth Malo M.S., Director, Regulatory Affairs

MEDDEV 2.2/3 Rev 3 Use-By Date June 1998

Information for Manufacturers on the Inspection of Manufacturing Site(s) (Assessment of the Quality Management System)

FDA-Regulated Research

State Control of Medicines and Medical Devices in Russian Federation

EMA perspective on increasing focus on review of device components of medicinal products

MedDev and NB-MED Recommendations

MEDICAL DEVICES REGULATORY SYSTEM

Agenzia Italiana del Farmaco

GUIDELINES ON MEDICAL DEVICES GUIDE FOR COMPETENT AUTHORITIES IN MAKING AN ASSESSMENT OF CLINICAL INVESTIGATION NOTIFICATION

Biological evaluation of medical devices --

The interface between Good Clinical Practice and Good Manufacturing Practice

MDR ID: Definition: Applicable:

COMMITTEE FOR PROPRIETARY MEDICINAL PRODUCTS (CPMP) POSITION PAPER ON COMPLIANCE WITH PHARMACOVIGILANCE REGULATORY OBLIGATIONS

Developing a European First-in-Human Study: Three Key Decisions

Office for Human Subject Protection. University of Rochester

ISO INTERNATIONAL STANDARD

Tuesday, 21 October Jang Yong Choi

REGISTRATION OF MEDICAL DEVICE. SUBMISSION & APPLICATION By IDAMAZURA IDRIS MEDICAL DEVICE AUTHORITY

Updated EU Medical Device Regula2ons: What s In, What s Out, and What s Hot?

MedDev Rev 4 Medical Devices Regulation. Clinical Evidence Requirements Key Changes and Clarifications. Alan Eller 21 March 2017

AAMI Quality Systems White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 1

Korea GMP & DMF. April Korea Pharmaceutical Traders Association Deputy director Oh Hyun, Kwon

GENERAL AND ORGANISATIONAL REQUIREMENTS

Manual for ITC's Clients 2016 Conformity assessment of in vitro diagnostic medical devices pursuant to Council Directive 98/79/EC

MEDICAL DEVICES : Guidance document

Prevent Quality System Deficiencies by Conducting Effective Internal Audits. Whitepaper

Laboratorio di Tecnologie Biomediche

Clinical Research: A Multifaceted Discipline

Quality Management System MANUAL. SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702

OVERVIEW OF THE PREQUALIFICATION OF MALE CIRCUMCISION DEVICES ASSESSMENT PROCESS

Transcription:

Page1211 Indo American Journal of Pharmaceutical Research, 2015 ISSN NO: 2231-6876 MEDICAL DEVICE REGULATION AND ITS COMPARISON IN EUROPE, AUSTRALIA AND INDIA Pooja Brijeshkumar Patel, Ms. Krupa Chaitanyakumar Thula *, Dr. Dilip Girishbhai Maheshwari. L. J Institute of Pharmacy, Sarkhej, Ahmedabad 382210, Gujarat, India. ARTICLE INFO Article history Received 21/03/2015 Available online 31/03/2015 Keywords Medical Devices, European Commission, Therapeutic Goods Administration, Central Drug Standard Control Organization, Classification, Labeling Requirements. Corresponding author Ms. Krupa Chaitanyakumar Thula Assistant Professor, Department of Quality Assurance and Pharm Regulatory Affairs L. J. Campus, Between Kataria Motor and Sanand-Sarkhej Circle, S. G. Highway, Ahmedabad-382210, India. krupathula@yahoo.com 09898660011 ABSTRACT Millions of patient s world widely depends on ever widening arrays of medical devices for the diagnosis, treatment and prevention and management of diseases. In Europe medical devices are regulated by European Commission and covered under New Approach of Directives. In Australia medical devices are regulated by Therapeutic Goods Administration (TGA) and in India it is regulated by central drugs standard control organizations (CDSCO). All these countries have slight different classification system and registration process for classifying and registering the medical devices in respected countries. This article mainly includes classification, registration process for approval, labeling requirements, adverse drug reaction reporting system and registration requirement for approval of medical devices in these three countries. This type of evaluation is helpful for newly developing industries for better understanding of all requirements as a regulatory point of view. Time duration limit for review of medical device registration fixed in Australia divergence to India regulation. Medical device registered for five year as per European and Australian regulation whereas in India only for three year. Application fees for registration of medical device in Australia and Europe seen costly as compared to Indian application fees. Please cite this article in press as Pooja Brijeshkumar Patel et al. Medical Device Regulation and its Comparison in Europe, Australia And India. Indo American Journal of Pharm Research.2015:5(03). Copy right 2015 This is an Open Access article distributed under the terms of the Indo American journal of Pharmaceutical Research, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Page1212 INTRODUCTION Medical device is any instrument, implant, in vitro reagent, apparatus or similar or any related article which are used for diagnosis, prevention, or treatment of disease or any other conditions, and which does not achieve its purposes through chemical action within or on the body [1]. Medical devices are becoming more important in the health care sector. This increases the demand for better regulatory frameworks to ensure that products entering the market are safe and efficient. One of the major issues for companies developing and producing medical devices is to be updated on the regulatory requirements and implement them in the process. A company that does not succeed with this may lose thousands of dollars in the delay of marketing the product. The Australian medical devices and diagnostics market was worth $2.9 billion in 2009 and is estimated to grow to $4.3 billion in 2013 at a compound annual growth rate of 10.3%. The global medical device market is worth over US$150 billion, with the United States of America, European Union, and Japan having over 65% of the market share. Regulation of Medical Device, according to the Europe: In Europe Medical Devices are regulated by European Commission. In European Union, European Medicines Agency (EMA) is a decentralized agency which is located in London. EMA is mainly responsible for scientific evaluation of medicines which are developed by pharmaceutical companies for use of it in the European Union [2]. The European Medicines Agency's (EMA) is also responsible for the protection and promotion of public and animal health, by the supervision and evaluation of medicines for human and veterinary use. Definition: Medical devices are any instrument, appliance, apparatus, software, material or other article, whether used alone or in combination, including the software intended by its manufacturer to be used specifically for diagnosis or therapeutic purposes and for its proper application, intended by the manufacturer to be used for human beings for the purpose of: Diagnosis, monitoring, prevention, treatment or alleviation of any disease, Diagnosis, treatment, monitoring, alleviation or compensation for an injury or handicap, Replacement, Investigation or modification of the anatomy or of a physiological process, Control of conception, which does not achieve its principal intended action in or on the human body by pharmacological, immunological or metabolic means, but which may be assisted in its function by such means [2]. Classification: Depending on the level of risk medical devices are classified in to four categories as Class I, Class II, Class III and Class IV [3]. Table 1: Classification according to Europe [3] Classification Risk Level Example Class I Low risk Stethoscope, Conductive Gels Class IIa Medium risk Antistatic Tubing For Anaesthesia Class IIb Elevated risk Ventilators, Infusion Pumps, Anaesthetic vaporizers Class III High risk Implants and dressings made from collagen, Biological Heart valves Regulation: European Medicines Agency (EMA) is a decentralized body of the European Union (EU) which is responsible to protect human and animal health through the supervision and evaluation of medical products for human or animal use [2, 3]. There are main three directives for medical devices: Active Implantable Medical Device (AIMD) Directive - 90/385/EEC Medical Device Directive (MDD) - 93/42/EEC In Vitro Diagnostic Device Directive (IVD) - 98/79/EC Other Directives: General Product Safety Directive 92/59/EEC Machinery Directive 98/37/EC Directive of Low Voltage Electrical Equipment 73/23/EEC Directive of Electromagnetic Compatibility 89/336/EEC

Page1213 Placing medical devices on the market: According to EMA, market approval for medical devices is achieved via a decentralized procedure of CE marking [2, 4, 5]. If any manufacturer wants to sell medical devices in the EU, there is a need of only submitting one single marketing authorization application to the EMEA. The documentation shall be in English, French or German. A manufacturer who does not have a registered place of business in the EU shall appoint a single authorized representative in the European Union [2, 3, 4]. The devices shall have a GMDN code. All medical devices except class I medical devices require the involvement of a Notified Body. Medical devices must comply with the essential requirements of Annexure I of the Directive 93/42/EEC [2, 4, 5]. Harmonized Standards are used to meet compliance with the essential requirements. Clinical trials are required for Active Implantable Medical Devices, Invasive Medical Devices, class III devices and for long-term use of class IIa and IIb medical devices. For class I and IIa devices instructions for use are not necessary if they can be used safely without them. A registration of a product is valid for 3 years. The European market approval system consists of national authorities for separate registration and certification of medical devices. The decentralized procedure mainly depends on the designated notified bodies that regulate CE marking. Regulatory process for approval of Medical Device [6] : Figure 1: Regulatory process for approval of Medical Device. CE-Marking: CE mark is a conformity mark, which must place on all European medical devices before they can be marketed. It is a declaration by the manufacturer that the product meets all the appropriate provisions of the relevant legislation including those related to safety and quality [2, 3]. The CE mark also means that the product can be freely marketed anywhere in the European Economic Area (EEA) without further control [2]. Competent Authorities (CA), Notified Bodies (NB) and authorized representatives are involved in the CE marking process. Competent Authorities are exists in each European Member State and are nominated by each government to monitor and ensure compliance with its provisions of the MDD [2-5]. The Competent Authority designates a Notified Body to ensure that conformity assessment procedures are completed according to the relevant requirements. The Authorized Representative is designated by the manufacturers and they are legally responsible for compliance with the regulations of medical device directives. Manufacturer of the medical devices are responsible to ensure that their product complies with the essential requirements of the relevant EU legislation. CE-certification includes the following steps: Decision whether or not a product is a medical device Medical Device classification by the manufacturer Contact to Notified Bodies, preliminary discussions and answering of specific questions, confirmation of device classification by the Notified Body Time and cost estimation for different certification routes Formal application and certification contract Submission of documents to the Notified Body Evaluation of the submitted documents and report Audit of the manufacturer s operations if applicable Decision about the certification and issuing of the relevant certificate Surveillance audits, Full re-audit and issuing of a new certificate (usually after five years)

Page1214 A CE mark expires after three years. Upon CE certification in the first year, unannounced surveillance audits follow. As part of vigilance, the safety and intended use of the medical device is monitored after its introduction on the market post-marketing surveillance (PMS) and post-marketing clinical follow-up (PMCFU). In the fourth year, renewal of the CE mark can be applied [2-4]. The manufacturer or the authorized representative is requested to keep copies of the technical documentation at least 5 years for IVDs and at least 10 and 15 years for AIMDs, after the last product has been placed on the market. Notified Bodies: For the conformity assessment of certain products the European new approach directives require the involvement of third parties and these are the national authorities of the member states [2, 4]. Under surveillance by national notifying authorities Notified Bodies have the authority to withdraw or modify the notification as soon as the conditions of notification are no longer met. Notified Bodies are free to offer their conformity assessment services for which they are notified to any manufacturer established either inside the EU or in third countries. They may carry out these activities on the territory of other countries either with their own personnel or with subcontractors. Quality Systems: Effective quality management systems (QMS) are key regulatory consideration for allowing permitting medical device manufacturers to market their products around the world. ISO 13485 is a harmonized standard for the quality management system requirements for medical device manufacturers. The Medical Device Directive requires that the manufacturers of medical devices keep a product-related, efficacious and adequate quality system. The application of the quality system must ensure that the products comply with the provisions of the MDD [2, 3]. All the requirements and provisions adopted by the manufacturer for his quality system must be documented in a systematic manner in the form of written policies and procedures such as quality plans, quality manuals, quality programs and quality records [2]. Additional aspects to be covered by the quality management system include: The technical documentation. Reference to the essential requirements according to Annex I of the Medical Device Directive. Information about medical device regulations and harmonized standards. instructions for use and Labeling. Risk Analysis Different languages used Post-marketing surveillance (PMS) Reporting of ADR under the vigilance system Retention of certain documents Depending on the class, status of QMS and the technical file cost and duration of the certification procedure itself are varied. The minimum duration of a certification procedure with a Notified Body is twelve weeks. Documents required: Name and address of manufacturer and identification of products European Agents name and address if applicable WHO-GMP compliance certificate Declaration of Conformity and/or certificate for regulated product Description of product (Name, Type, Model) Test reports Technical File Labeling Requirements: Each device must be accompanied by the information needed to use it safely and properly by which users can understand it easily and also be identified by the manufacturer. This information comprises all the details of the product i.e. instructions for use, intended purpose and storage condition of the product. Instructions for use must be included in the packaging for every device. Except Class I or Class IIa devices there is no need such instructions for use, if they can be used safely without any such instructions. Where appropriate, this information should also include specific form of symbols or identification color. It is necessary to conform that any identification color or symbol used are complies with the harmonized standards. In such where no standards exist, the symbols and colors used must be described in the documentation supplied with the device.

Page1215 The label may bear following requirements [6] : Name, address and trade name of the manufacturer Address of the manufacturing site Name and address of the authorized representative (if the manufacturer does not have a registered place of business in the Community) Details to identify the device and the contents of the packaging especially for the users If medical device is sterile than use the word STERILE If medical device is used for single use than indicate it as SINGLE USE If any device is custom-made, than use the word custom-made device If any device is intended for clinical investigation than use the word exclusively for clinical investigations Batch code preceded by the word LOT, or the serial number; If appropriate any special storage or handling conditions; If appropriate any special operating instructions; Any warnings and precautions to take, if required; If applicable give method of sterilization; Manufacturer clearly states the instruction for use on the label (with its intended purpose, any undesirable side-effects and contraindication); If medical device is reusable, give information about the appropriate processes to allow reuse; Any devices which emit radiations used as medical purposes than give the details of the nature, type, intensity and distribution of this radiation. Vigilance system: The main purpose of the Vigilance system is to protection of health and safety of patients, users and others by reducing the adverse incident being repeated in different places at different times. This is to be achieved by the evaluation of reported incidents and, where appropriate, dissemination of information which could be used to prevent such repetitions, or to alleviate the consequences of such incidents. The response of the vigilance and market recall is to determine the action to be taken once we receive information concerning an incident directly or if the Competent Authority receives it) after the device has been placed in the market. Time scale for initial reporting is as follows:serious public health threat immediately, not later than 2 days Death or unanticipated serious deterioration in state of health - not later than 30elapsed calendar days. Medical device regulatory system in Australia: In Australia medical devices are regulated by TGA (Therapeutic Goods Administration). The Therapeutic Goods Administration (TGA) is part of the Australian Government Department of Health and Ageing, which are responsible for regulating medicines and medical devices. According to the Therapeutic Goods 1989 Act, therapeutic goods (medicines and medical devices) supplied in Australia must meet acceptable standards of safety, efficacy, quality and performance. On the basis of clinical and scientific expertise to decision-making, TGA ensures any risk associated with the use of medicines and medical devices. The TGA depends on the healthcare professionals, public and industries to report problems with medicines or medical devices. TGA investigate the received reports and determine any necessary regulatory action [7]. Definition: On the basis of Therapeutic Goods Act 1989, medical devices are defined as [7] : Any appliance, instrument, apparatus, material or other article (whether used in combination or alone, and including the software necessary for its proper application) intended by the person under whose name it is to be supplied, to be used for human beings for the purposes of one or more of the following: Diagnosis, treatment, monitoring, prevention or alleviation of disease, Diagnosis, monitoring, treatment, alleviation of or compensation for an injury or handicap, Modification, Investigation or replacement of the anatomy or of a physiological process, Control of conception, and does not achieve its principal intended action in or on the human body by pharmacological, immunological or metabolic means, but which may be assisted in its function by such means; or an accessory to such an instrument, apparatus, appliance, material or other article. Classification of medical device as per Australia: Based on the level of risk and the intended purpose of the Medical Device, they are classified in to 5 classes in accordance to a set of 22 classification rules. The higher the class, the more regulatory control is required [7, 8]. Depending on the level of risk, medical devices are classified as following:

Page1216 Table 2: Classification of Medical Device as per Australia Classification Risk level Example Class I Low risk Urine Collection Bottle, Non-sterile Dressings, Plaster Bandages, Handheld dental mirrors, Scissors, Surgical microscopes Class IIa Low-Medium risk Anesthesia Breathing Circuits, Intravenous tubing, Blood transfusion sets, Adhesive for topical use, Contact lenses, Nasopharyngeal airways, Oxygen Tubing and Mask, Suturing needles and Clamps Class IIb Medium-High risk Artificial Eyes, Dressing for Chronic Extensive Ulcerated Wounds, Condoms and Nebulizers, Personal insulin injectors, Bone cements, Lung ventilators Class III High risk Antibiotic bone cements, Cardiovascular Catheter, Prosthetic Heart valves, Cardiac Output Probes AIMD (Active Implantable Medical Device) Cardiac stents, Pacemaker, Nerve stimulator, Cardiac defibrillator Registration of medical device as per Australia: The sponsor is responsible for registration of the medical devices in the Australian Register of therapeutic Goods (ARTG) database. Before registration of the medical devices in the ARTG, the medical device must be classified according to the Australian system and suitable risk analysis has been done and quality management systems (QMS) must be applied for complying with the Essential principles according to required criteria [7]. Sponsor may use the Devices Electronic Application Lodgment system (DEAL system) for the applications and registering of medical devices. Sponsors are responsible for all activities concerning medical devices while manufacturers have obligations to fulfill the requirements. After registration medical devices have Global Medical Device Nomenclature (GMDN) codes [7]. A quality management system (QMS) is required for medical devices class IIa, IIb, III and AIMDs to get a conformity assessment certificate. Australia has its own standard orders but the international ISO standards can be used. According to the Australian Essential principles which ensure the safety and efficacy of the therapeutic goods, manufacturer is required to have made a documented risk analysis of the product. For Class III medical devices and AIMDs conformity assessment certificate is required and manufacturer shall submit a Design Dossier whereas for class I medical devices conformity assessment certificate is not required. The registration of medical device is valid for five years [7]. [7, 8] Figure 2: Registration process of medical device in Australia [7]

Page1217 Documents required [7] : Name and Address of Manufacturer and there relevant details All the details of Client Conformity assessment certification including the certificates details and if there are any restrictions on the scope Classification and classification rule of the device Conformity assessment procedure Conformity assessment body GMDN code - The Global Medical Device Nomenclature (GMDN) code is a collection of terms, each with a unique code number, to describe and catalogue medical devices. Australian declaration of certificate Labeling Requirements [7] : Name of Therapeutic device Name and address of manufacturer and sponsor or Registered trademark of manufacturer or sponsor of therapeutic devices Instruction for use Batch number and serial number preceded by words Batch, Batch Number, Batch No., Lot, Lot Number, Lot No. or Lot Code, Serial Number, Serial No., or by words having a similar meaning, or by the symbol B, (B), or B, SN, S/N, or FABR. Date of manufacture may be used as the batch number or serial number (if clearly identifiable as a date). If any devices used in examination of specimens taken from the body of a person or animal in connection with the diagnosis of a disease, injury or any defect than preceded as diagnostic goods for in vitro use. Package containing one or more therapeutic devices, means the outermost level of packaging on that number of therapeutic devices which is supplied for use is preceded as Outer package. Size describes the size or shape of device and may include: physical measurement of device Descriptive expression including: small, medium or large. If any medical device is sterile than represented as sterile therapeutic device. If medical devices are used in one occasion only than preceded as single use or use only once. Vigilance system: The purpose of medical device vigilance is to protect the health and safety of patients, users, and others by reducing the adverse events being repeated. This can be achieved by: Evaluating reported adverse incidents Disseminating information that could be used to prevent or minimize the consequences of adverse events, where appropriate Modifying the therapeutic goods (Medicines and medical devices) [7, 8] Removing the medical device from the market Action is undertaken by the TGA and the sponsor or manufacturer may aware of information about medical devices which are supplied in Australia, such as: Malfunctions Adverse event reports Results of testing and any other information Manufacturer, sponsor and user may responsible for reporting adverse drug reaction of medical devices. Implantable medical device tracking system (IMDTS) is tracking system for ADR reporting. Manufacturer may report death and serious injury within 10 days, non-adverse event or event not result in death or serious injury within 30 calendar days (1 month) and any serious public health threats require prompt remedial action within in 48 hours (2 days) [9]. Medical Device Regulatory system in India: In India medical devices are regulated by Central Drugs Standard Control Organization (CDSCO) [10]. The Central Drugs Standard Control Organization (CDSCO) is the Central Drug Authority for discharging functions assigned to the Central Government under the Drugs and Cosmetics Act. CDSCO has six zonal offices, four sub-zonal offices, 11 port offices and six laboratories under its control [10]. In India medical devices are regulated by Directorate General of India (DCGI), Ministry of Health and Family Welfare (MHFW) [10, 11]. In India Drug Technical Advisory Board (DTAB) has recommended strict implementation of Indian Conformity Assessment Certificate (ICAC) for manufacture, import and marketing the medical devices on Indian market [11]. Indian Medical Device Regulatory Act (IMRDA) had been set up for their opinion and recommendation [11]. The importing, manufacturing, distribution and sale of medical devices in India are overseen by drug and cosmetic act (1940) and rules (1945) and Inspections are carried out by CDSCO.

Page1218 Definition: Medical devices are any apparatus, instrument, implant or other similar or related article, which are intended for use in diagnosis, disease, cure, mitigation, prevention or treatment of disease or intended to affect the structure or any function of the body and which does not achieve its primary intended purposes through its chemical action within or on the body [10]. Classification of medical devices as per India: Depending on the level of risk medical devices are classified in 4 categories in to Class A, Class B, Class C and Class D [10]. Table 3: Classification of Medical device as per India [10] Classification Risk level Example Class A Low risk Tongue Depressors, Thermometers Class B Low-Moderate risk Suction Equipment, Hypodermic Needles Class C Moderate-High risk Lung ventilator, Bone Fixation Plate Class D High risk Pacemaker, Heart valves, Implantable Defibrillator Registration procedure of medical device as per India: New medical Devices and Notified Medical Devices are classified under drug rules which require marketing Authorization from DCGI. Other medical devices such as Non-notified medical devices do not require manufacturing, sales, import registration and can be marketed freely [10]. The applicant can be the manufacturer, the importer or the responsible agent in India [10, 11]. The registrations of medical devices are done according to Rule 24A of the Drugs and Cosmetic Act and Form 40 shall be filed. Manufacturers of medical devices conduct suitable tests to prove the product quality and quality systems shall concern design, development and manufacture. If quality systems for medical devices do not exist, CE-marked or FDA approved products are preferred because of their quality, performance and safety. Medical devices also require risk management in form of ISO 14971. Manufacturer or authorized agent of medical devices shall have documented procedures for distribution records, complaint handling, adverse incident reporting and product recall. A registration of a medical device is valid for five years [11]. Figure 3: Registration procedure for India. Documents required [12] : Covering Latter Authorization latter TR6 (Treasury) challan Power of Attorney Free Sale Certificate Name and Address of Manufacturer Name and Address of Authorized Agent Declaration of certificate Wholesale License

Page1219 Labeling Requirements [13] : Product description Name of manufacturer or authorized representative Address of premises Number of units of content Batch number preceded by words Batch No, B. NO., Batch or Lot No. Manufacturing license number preceded by word Manufacturing License Number or Mfg. Lic. No. or M.L. Date of expiration Vigilance system: Manufacturer is responsible for reporting adverse drug reaction of any of medical devices. There is no specific tracking system for reporting of adverse drug reaction of medical device. Manufacturer may report death or serious injury or a serious public health threat within 10 calendar days and other reportable event not later than 30 calendar days. Comparison of medical device regulation in Europe, Australia and India: Table 4: Comparison of medical device regulation in Europe, Australia and India. Parameters Europe Australia India Regulatory Therapeutic Goods Authority Administration (TGA) Legislation and guideline follows Classification European Commission European Medicines Agency (EMA) 3 Directives: Active Implantable Medical Device (AIMD) Directive - 90/385/EEC Medical Device Directive (MDD) - 93/42/EEC In Vitro Diagnostic Device Directive (IVD) - 98/79/EC 4 categories depends on level of risk: Class I - Low Risk Class IIa - Medium Risk Class IIb Elevated Risk Class III - High Risk Australian Regulatory Guideline for Medical Devices (ARGMD) TGA 1989 Act TGA Regulation 2002 5 categories depends on level of risk: Class I - Low Risk Class IIa - Low to medium Risk Class IIb - Medium to high Risk Class III - High Risk Active Implantable Medical Devices (AIMD) Central Drugs Standard Control Organization (CDSCO) CLAA is a main regulatory body for medical device By CDSCO Directorate General of Health of India (DCGI), Ministry of Health and Family Welfare Gov. of India IMRDA (Indian Medical Device Regulatory Act) Drug and Cosmetic Act (1940) and rules (1945) 4 Categories depends on level of risk: Class A - Low Risk Class B - Low to medium Risk Class C - Medium to high Risk Class D - High Risk Rules Total 18 rules Total 22 rules Not specified Essential Compulsory Compulsory Not applicable Principles Applicant Manufacturer/Authorized representative Application in form of Documents Required Manufacture,/Australian based sponsor/agent on behalf of Manufacturer and sponsor Manufacturer/Authorized representative Technical file Technical file Technical file Name and address of manufacturer and identification of products European Agents name and address if applicable WHO-GMP compliance certificate Declaration of Conformity and/or certificate for Manufacturer s details including name and address Client reference and details Conformity assessment certification including the certificates details and if there are any restrictions on the scope Class of the device Conformity assessment procedure Covering Latter Authorization latter TR6 challan Power of Attorney Free Sale Certificate Manufacturer s detail Declaration of certificate Wholesale License

Page1220 Indicative mark required Registration steps regulated product Description of product (Name, Type, Model) Test reports Technical File Conformity assessment body GMDN code - The Global Medical Device Nomenclature (GMDN) code is a collection of terms, each with a unique code number, to describe and catalogue medical devices Australian declaration of certificate CE mark ARTG number CE mark TGA refers CE mark as EC- European community certificate Step 1- Confirm classification of the device Step 2 - Authorized representative and notified bodies are appointed Step 3 - Submission of documents and Evaluation of documents Step 4 Provide CE certification Step1-Manufacturer obtain conformity assessment certificate from TGA and gives to the sponsor. Step 2 Sponsor submits the application in to DEAL system with appropriate charges. Step 3 - TGA review and accepts the application Step 4 TGA includes the device in ARTG. ICAC marking Indian Conformity Assessment Certificate CE mark and FDA approval is required Step1- Appoint Authorized agent. Step 2- Submit application with appropriate fees (in TR6 challan) Step 3- Compilation of registration dossier as per regulatory guideline Step 4- Submit product registration application at CDSCO Delhi Headquarter Registration done for Five years Five years Three years Fees Certification Fees charged by Notified Body. Application fees Fees paid through TR6 (Treasury) challan or through electronic Official fees charged by government competent authorities. Class I measuring 0 AUD$ clearance system (ECS) Registration fees 1500 US$ In UK registration and notification fees are about 133 US$ Conformity Assessment Post market surveillance activity Vigilance system ADR reporting by In Spain registration and notification fees are about 705 US$ Compulsory performed by Authorized Notified Body ADR reporting Field safety notice Investigation enforcement Post market clinical followup records Adverse Incident Tracking System (AITS) Manufacturer/Authorized representative/competent authority Class I sterile - 730.00 AUD$ For registration of single product 1000 US$ Class IIa - 730.00 AUD$ For additional products - 1000 US$ Class IIb - 730.00 AUD$ For duplicate copy 300 US$ Class III - 960.00 AUD$ AIMD - 960.00 AUD$ Compulsory Prepared by authorized notified body and manufacturer and sponsor ADR reporting Vigilance exchange program Enforcement activities Distribution record Audits Implantable Medical Device Tracking System (IMDTS) Manufacturer/Sponsor/User For class A performed by manufacturer For other class by Authorized Notified Body ADR for importers Complaint handling ADR reporting procedure for distribution of records Recall procedure No tracking system Manufacturer

Page1221 CONCLUSION Medical devices are serving an increasing role in improving patient s health and quality of life and in clinical practice. All three countries follow the GHTF regulations for medical devices but all have their own guidelines. All the three countries have different regulatory approval process for registration of medical device. By comparing all the parameters it has been seen that Europe and Australia has more stringent regulations as compared to India and according to the application fees Australian registration fees are most costly as compared to all the three countries. Medical devices cover a vast range of equipment, from simple tongue depressors to haemodialysis machines. Like medicines and other health technologies, they are essential for patient care at the bedside, at the rural health clinic or at the large, specialized hospital. Surprisingly, regulatory controls for medical devices are scarce in the developing world, even though implementation of national medical device regulations will often address the very issues raised in countries as major concerns for patient safety. As an important input to the health care system, medical devices should be properly managed and utilized in order to produce an efficient health intervention. ACKNOWLEDGEMENT The authors are thankful to Dr. Pundarikakshidu, Director of L. J. Institute of Pharmacy, Ahmedabad, India for providing all facilities and encouragement to carry out the work. ABBREVIATION TGA: Therapeutic Goods Administration CDSCO: Central Drugs Standard Control Organization EMA: European Medicines Agency EU: European Union AIMD: Active Implantable Medical Device MDD: Medical Device Directive IVD: In Vitro Diagnostic Device Directive ARGMD: Australian Regulatory Guideline for Medical Devices EC: European Commission GMDN: Global Medical Device Nomenclature ISO: Indian Standard Organization EEA: European Economic Area CA: Competent Authority NB: Notified Body PMS: post-marketing surveillance PMCFU: Post-marketing clinical follow-up QMS: Quality Management System DEAL System: Device Electronic Application Lodgment IMDTS: Implantable Medical Device Tracking System ADR: Adverse Drug Reaction ICAC: Indian Conformity Assessment Certificate IMRDA: Indian Medical Device Regulatory Act FDA: Food and Drug Administration DCGI: Directorate General of Health of India REFERENCES 1. World Health Organization. Definition of medical device. October 2013. http://www.who.int/medical-device/en. (26 Jan 2015) 2. European Medicines Agency. Basic information about the European Directive 93/42/EEC on Medical Devices. November 2009. http://www.mdc-ce.de (26 Jan 2015) 3. European Medicines Agency. European medical device classification and classification rules. October 2007. (28 Jan 2015) 4. European Commission. Placement of Medical Device in European Market. August 2009. http://eurlex.europa.eu/lexurisev/lexurisev.do?uri=conslg:1998l0079:20090807:en:pdf (28 Jan 2015) 5. European Commission. The medical device regulatory approval process. 2013. www.emergogroup.com/files/medcal-deviceregulatory-process-europe.pdf (28 Jan 2015) 6. European Commission. Labeling Requirement for medical devices. 2007. http://www.consleg-1993loo42-20071011-en- TXT.pdf (28 Jan 2015) 7. Therapeutic Goods Administration. Australian Regulatory Guideline for Medical Devices (ARGMD)..May 2011. www.tga.gov.au/sites/default/files/devices-argmd-01.pdf (28 Jan 2015) 8. Therapeutic Goods Administration. Classification of medical device. 25 January 2005. http://www.tga.gov.au/docs/pdf/devguid25.pdf (28 Jan 2015) 9. Therapeutic Goods Administration. Post-market activities. http://www.tga.gov.au/pdf/devguid11.pdf (28 Jan 2015) 10. Central Drugs Standard Control Organization. Organogram of Medical Devices Division. Dec. 2 2014. http://www.cdsco.nic.in/writereaddata/medical%20device%20organo.pdf (28 Jan 2015)

Page1222 11. Central Drugs Standard Control Organization. Guidance document on common submission format for Import License in Form-10 of Notified Medical Device in India. 1 Jan. 2013 http://www.cdsco.nic.in/writereaddata/final%20guidance_doc_import_licence_31.10.2012.pdf (28 Jan 2015) 12. Gupta R., Review of marketing authorization of medical device in India, IJDRA 2014; 2(4):25-32. 13. Gupta P., Janodia M., Puralea C., Udupa N. Medical device vigilance systems: India, US, UK and Australia, Medical Device: Evidence and Research 2010; 3:67-79. 54878478451150336