A continuous quality improvement process to reduce excessive CT radiation dose events across a large, multiinstitutional

Similar documents
Contrast Dose Management. For CT and MR

Healthcare. Philips DoseWise Portal. Radiation exposure management software

RADIATION DOSE MONITOR. The DACS for Building a Low-Dose Culture

RADIATION DOSE MONITOR

DoseWise Portal. Take control of dose management. across your organization

Take control of dose management

1. Executive Summary

Joint Commission Medical Imaging Requirements. Laura Gruber, MBA, RT(R), RDMS, RVT Sr. Director Department of Medical Imaging SPR Annual Meeting

General Radiology Improvement Database

2017 ACR Computed Tomography Quality Control Manual FAQS

Stay Ahead of the Curve in Radiation Reduction Standards

UNTAPPED POTENTIAL PART 1 OF 3 How Radiology Administrators made the most of their CT investments

Implementing Optimal Kidney Stone CT Protocols Using Outreach from a National Database. Purpose

Update on Radiation Dose Analytics: Meaningful Use or Useless Meaning?

Protecting Patients 3/5/2014. Educating and Engaging your Staff in Radiation Safety

Handzettel 1. CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014

Effective Date: April 19, 2018 TITLE: GUIDELINE FOR RADIATION SAFETY COMMITTEE REVIEW OF HUMAN RESEARCH STUDIES. Table of Contents

Installation of a regional CT dose registry with a dose management software

Siemens Ct Scanner Manual

Donna Newman B.A. RT ( R) CNMT Director of Professional Practice ISRRT

Eric W. Abelquist, Ph.D., CHP President, Health Physics Society. August 2017

Quality ID #145: Radiology: Exposure Dose or Time Reported for Procedures Using Fluoroscopy National Quality Strategy Domain: Patient Safety

PHARMACY, MEDICINES & POISONS BOARD

Interventional CT: More Procedures, Reduced Dose, Improved Speed at Minimal Cost

Advanced Process and Quality Control in Hot Rolling Mills Using Eddy Current Inspection

Future bound. Philips Ingenuity Core

GENESIS Edition Transforming CT

Using Application Response to Monitor Microsoft Outlook

GENESIS Edition. Transforming CT

ADDENDUM 2. I. The following are revisions to the general information provided in the RFP 744-R1715 posted March 21, 2017:

MONTAGE. drives COMPETITIVE ADVANTAGE

PRINCE Update. Changes to the manual. AXELOS.com. April 2017 PUBLIC

DENOMINATOR: All final reports for patients, regardless of age, undergoing a CT procedure

IMV 2006 CT Market Summary Report Featuring results from 2,565 hospital and non-hospital CT imaging sites in the United States

Leading Canadian Pediatric Hospital Continues to Drive Dose Down for Sick Children

8/2/2017. Key attributes of scientific excellence: rigor, innovation, and relevance. Medical Physics 3.0. Key Attributes of Scientific Excellence

TRAINING AND DEVELOPMENT POLICY AND PROCEDURE ISO 9001:2008 Clause 6.2.2

Welcome to Module 2. In this module, we will be providing an overview of the TransCelerate Methodology process and toolkit, comparing off-site,

PERFORM Operating Document. Radiation Dose Estimates, Calculations and Tracking Methodology

Overview of IMV Data: Perspectives on Large Scale Data Collection. December 8, 2011

Boris Brkljačić. Overview of Oncologic Imaging in Europe

CONTRACTING CONSIDERATIONS

Fluoroscopy Testing Workshop

Guidance for Industry and FDA Staff Procedures for Handling Post-Approval Studies Imposed by PMA Order

Pathology Image Exchange. Paul J van Diest, MD, PhD UMC Utrecht

Aims and objectives. Page 2 of 13

On the Importance of Computation in Clinical Radiology

Medical Radiation Reporting: A Manufacturer s Perspective

ITIL Foundation v V1. Module 4: Service Transition. Reader s Note QAI India Ltd. I

WINNING THE ED THROUGHPUT BATTLE: TENET HEALTHCARE S EMERGENCY DEPARTMENT FLASHBOARD

VIRIDOR WASTE MANAGEMENT LIMITED. Lostock Waste Treatment Plant. Airborne Pollutant Management Plan. November 2009

Create interoperability in a MEDITECH environment

IT Administration including SIMS Support

FORM 5 University of California, Davis Health System Health Physics Office HUMAN RADIATION USE RESEARCH APPLICATION

Health Job Family: Radiographer Progression

Computed Tomography: Optimization of acquisition protocols & Justification of clinical referrals. Koos Geleijns, medical physicist

MRI Conditional Devices and Patient Safety. Outline. MRI Labeling: Objects 8/3/13 1. David W. Jordan, Ph.D. Senior Medical Physicist

b i n a r y H e a l t h c a r e The DICOM Conformance Black Paper What Your Vendors Don t Want You To Know

Inside View Features and Specifications Software Version 3.0

Bed Management Solution (BMS)

TECHNOLOGY brief: Event Management. Event Management. Nancy Hinich-Gualda

Trial oversight SOP for HEY-sponsored CTIMPs

Automated Quality Assurance System for Clinical CT Systems

PURSUING A CAREER IN MEDICAL IMAGING LEADERSHIP by AHRA

Excellence in Control of Radiation Exposures

TrialStat Corporation: On Schedule with High Quality and Cost Savings for the Customer*

Soteria Strains Safe Patient Handling and Mobility Program Guide

Practical Steps To Implement The Balanced Scorecard A Case Study

Effective Date: March 8, 2017 TABLE OF CONTENTS... 1 LIST OF TABLES... 7 INTRODUCTION... 8

T.W.I.S.T. IDEAL Radiological Imaging Services Presented by Adrienne Coya & Sandra Marcelin

Report on 2011 NPCC Culture of Compliance Survey Initiative

Data Collection Tools Functions, Indicators & Sub-Indicators

Session Number: 5 Proactive Remote Service Support

Ethics Application Overview

PRINCE2 walkthrough and Roadmap. Dave Litten

Report on Human Resources Surveys in Medical Imaging & Radiation Therapy for the Canadian Association of Medical Radiation Technologists (CAMRT)

Overcome the Top Challenges of Handling OOS Results by Knowing FDA Observations

Creating the future of Healthcare through meaningful innovation. Gene Saragnese, CEO Healthcare Imaging Systems

PhUSE Paper RG06. Quality Risk Radar for the Outsourcing of Clinical Trials The Best Guide to Success

CUSTOMER AND SUPPLIER ROLES AND RESPONSIBILITIES FOR 21 CFR 11 COMPLIANCE ASSESSMENT. 21 CFR Part 11 FAQ. (Frequently Asked Questions)

TECHNOLOGY POLICY SUMMARY FOR THIRD PARTY SUPPLIERS

ISC: UNRESTRICTED AC Attachment. Virtual Desktop Information Technology

Operations Manager Gastroenterology

Procurement of Project Management Training Services-

ITIL: Operational Support & Analysis (OSA) (Revision 1.6)

Summary of posters. General papers and clinical audit for. National experiences and. diagnostic radiology. expectations:

A technical discussion of performance and availability December IBM Tivoli Monitoring solutions for performance and availability

Job Fact Sheet Sections

Stand Alone Viewers A COMPLETE DIAGNOSTIC VIEWER WITH CONSOLIDATED WORKLIST DESIGNED TO OPTIMIZE REPORTING TIME

Implementing the New Pharmacovigilance Legislation

GUIDELINES FOR PROPER TRADEMARK USE INTRODUCTION

WELSH HEALTH CIRCULAR

GE Healthcare. Introducing Veo * on Discovery * CT750 HD Great care by design

ANNUAL SAFETY PROGRAM AUDIT FORM FOR FISCAL YEAR 2014 (JULY 2014 TO JUNE 2015)

Synapse RIS SOLUTIONS

IBM Infrastructure Security Services - Managed Security Information and Event Management (Managed SIEM)

4.513 Medical Examinations and Fitness for Duty

Fergus Sweeney, European Medicines Agency

Implementing ITIL Best Practices

ENERGY PERFORMANCE PROTOCOL QUALITY ASSURANCE SPECIFICATION

Transcription:

A continuous quality improvement process to reduce excessive CT radiation dose events across a large, multiinstitutional academic center Matthew E. Zygmont, MD, Rebecca Neill, MS, Pratik Rachh, MD, Shalmali Dharmadhikari, MS, PhD, Phuong-Anh Duong, MD Department of Radiology and Imaging Sciences Emory University School of Medicine Background Medical radiation is the greatest source of radiation exposure in the US and has increased 600% from 1980s to 2006. Computed tomography (CT) is the greatest source of medical radiation (49%) prompting greater scrutiny and increased focus on CT dose in the last several years. Early 1980s 2006 National Council for Radiation Protection, report 160 1

Background: optimizing CT dose in a large healthcare system Emory Healthcare is a large multi-institutional system with 26 scanners spread across 13 inpatient and outpatient locations. Over 150 subspecialty protocols are used across the system. Systematically optimizing protocols and standardizing them across a system with 3 scanner platforms is a major challenge. Background: New Regulations The Joint Commission Requirements (effective July 1, 2015): The organization documents the radiation dose index on every study produced during a diagnostic CT examination Diagnostic CT imaging protocols are reviewed and kept current with input from interpreting radiologist, medical physicist and lead technologist The organization reviews and analyzes incidents where the radiation dose index from diagnostic CT exams exceeded expected dose ranges identified in the imaging protocols. These incidents are then compared to external benchmarks 2

Aims To develop a process to identify excessive CT radiation exposures and continuously refine and optimize CT protocols to reduce overall computed tomography (CT) radiation dose by 5% by July 2016. To engage technologists in the process of recognizing and identifying solutions for excessive dose events. Baseline State Department of Radiology and Imaging Sciences installed radiation dose tracking software in November 2014 (Radimetrics, Bayer Healthcare, Whippany NJ ). 26 inpatient and outpatient CT scanners at all EHC sites were configured to send dose reports to Radimetrics server after 11/2014. On average, 10,556 exams were performed monthly generating 59.6 dose alerts. Dose thresholds were established for over 150 protocols using benchmark data from American College of Radiology Dose Index Registry (ACR DIR). Alerts for examinations exceeding the threshold were sent to the site QC technologist and a radiologist. 3

Project Team The CT Quality and Safety Committee organized and led the effort. This multidisciplinary team includes physicists, radiologists, and technologists from each imaging site (13 locations). Baseline CT dose (11/2014 5/2015) 4

Intervention 1: Update protocols (May 2015 ongoing) Technologists were trained to compare scanner protocols to an approved protocol repository in Radimetrics. Result: Some improvement in dose and protocols but many discrepancies were still found. Partnered with technologists to encourage review of protocols on scanner. Intervention 1 begins As protocols became available online in the protocol book, there was a decrease in the dose. baseline shift Shift: 7 or more consecutive points below the median indicates special cause variation 5

Intervention 2: dose alerts Dose alert levels were selected for over 150 protocols Alerts were set up to be e-mailed to the quality lead technologists, responsible radiologist, and physicist Technologists and radiologists unsure what to do with alerts 6

DP1 Intervention 2: Physicist reviewed dose alert results from first six months with technologists Reviewed the 6/2015-12/2015 dose alerts with technologists one-on-one at every site and discussed how to identify, respond, and mitigate dose alerts. Some sites were able to respond quickly to alerts and two sites subsequently decreased alerts to 0 in the next period. Dose alerts from July 2015 to December 2015 were reviewed and categorized. Majority of alerts were triggered due to scanner protocols not matching the master protocols Reasons for dose alerts 7

Slide 13 DP1 Will will Duong, Phuong-Anh, 10/10/2016

Intervention 2 begins Dose alerts turned on, some supervisors made changes based on alerts baseline shift Intervention 2: results and reflection Dose alerts alone were insufficient. A standard process was not established for resolving dose alerts and preventing future alerts. Staff were reluctant to make changes and use new protocols that were available in Radimetrics. A survey of technologists revealed that older protocols were still used because, I know it works. 8

Intervention 3 Step 1: Multi-step process created to address dose alerts Step 2: Reviewed dose alert data with managers to get additional support for implementing process Step 3: Implemented protocol-of-the-week program Step 1: process created for responding to alerts After an initial 6 month review of dose alerts, it became clear that a process was needed to guide team members in responding to dose alerts. Technologists were trained how to investigate the apparent cause. Protocols that continued to have dose alerts were reviewed by a radiologist and physicist. Some protocols were amended to reduce the rate of excessive dose. All protocol changes were logged and documented in Radimetrics. 9

Step 1: Create new process Technologist apparent cause analysis To review dose alerts, we developed a checklist for technologists to use when first evaluating dose alerts. Results were documented using the form below and sent by e-mail or added directly to the dose alert comments in Radimetrics. 10

Step 2: Reviewing dose alerts with managers (May 2016) Continued to have persistent non-response to alerts at some locations. Managers were included in the dose alert review and trained on how to help with intervention. Step 3: Protocol of the week (June 2016) High yield protocols are selected for update and review. A radiologist reviews the protocol for best practices and optimal parameters. The medical physicist distributes the master protocol. Technologists provide feedback and, if necessary, update the CT scanner using the master protocol Protocol of the Week: VN4A & VN4C Due Date: 10/7/2016 Site Scanner EUHM Brightspeed EUHM Emotion 16 EUHM Flash ESJH Definition AS ESJH Sensation 64 Checked Scanner with Protocol Book? Description of Changes Made Protocol Feedback Given 11

Managers assisted in adoption of new process and protocolof-the-week program targeted specific protocols Intervention 3 begins baseline shift Dose alert analysis 2 nd period: 1/2016 6/2016 Total number of alerts decreased from 372 to 335 (10% reduction) Most common reason is still incorrect protocol 12

Results Median exam doses decreased from 8.77 msv to 7.73 msv, an 12% drop in an health system with over 100,000 CT examinations per year. Dose alerts decreased by 10% as well from 373 to 335. The percentage of alerts due to protocol deviation from protocol book dropped from 70% to 44%. Given the large number of protocols in the system (150+), focusing on one high-yield protocol a week may be helpful. Reflection Because of the need for training in a multi-center institution, the interventions occurred over a long time period and overlapped. Therefore, the effects of such changes were not immediate. Initial largest drop in dose prior to implementation of dose alerts can be explained by updates to protocols occurring once the master protocol book became available. Some of the protocols-of-the-week were initiated because of concerns about image quality. Average exam doses were occasionally increased which may explain the slight increase during the final period during Sept. 2016. 13

Future Directions We have recently changed our alert levels, based on new ACR DIR data. Dose thresholds will be continuously adjusted based on benchmark data. ACR DIR thresholds will likely decrease as more participating institutions improve dose. The CT quality team will continue running a protocol-of-the week as a component of our annual protocol review. Dose alert reviews are being planned for shorter intervals (3 months). Individual sites can receive monthly reports on response rate and unauthorized protocol use to guide in service training. 14