Water Purification System Heal Force leads you to healthier life
Heal Force R NW Series Ultrapure Water System NW series ultrapure water system can be connected directly to tap water source.the system meets all the well known water quality standards like:astm type I,NCCLS type I and ISO 3696 type I. NW Series Product Features Microprocessor Control With self-diagnostic Microprocessor control system, LCD displays system status, including the conductivity inμs/cm or the resistivity in MΩ-cm as well as temperature, Liquid height of water tank and consumable change reminders to prevent a drop in water quality. Easy Maintenance Maintenance on the NW-series ultrapure water systems is very simple and straightforward,cartridge change is very convenient due to quick connectors. CanRex TM Pre-treatment Module Pretreatment is required to operate RO units Pre-treatment Cartridge effectively and economically. ConRex TM water pre-treatment technology can remove suspended matter, such as dirt and pieces of rust, and chlorine from chlorinated supply lines more completely than other pure water systems. 4-stage Resistivity Monitoring System There are 4 resistivity measurements monitors resistivity continuously and indicates water quality in MΩ-cm or μs/cm with automatic compensated temperature which gives the user greater convenience and security. Flow Restriction Design of Reverse-osmosis Heal Force reverse-osmosis system uses a flow resistriction design in order to keep reverse-osmosis membranes surface wet after the device is turned off.incomparably higher are maximum lifetime of Heal Force reverse-osmosis membranes and water quality produced by the reverse-osmosis system. resistivity measurements RO Cartridge NW System NW** NW** UV NW** UF NW** VF Pre-treatment Module Reverse Osmosis Storage Tank 30L Boost Pump Primary Module UV-oxidator UV-sterilizer Ultrafiltration Point-of-use Filter
UV Photo-oxidation Technology In UV and VF models, available with a built-in 185 nm and 254 nm UV lamp for production of low TOC water sensitive applications. Ultrafiltration Technology In UF and VF models ultrafiltration removes endotoxins to make the purified water suitable for critical cell culture applications such as in-vitro fertilization and monoclonal antibody production. Special Water Tank Reverse osmosis module produce Type III pure water for basic lab applications at a flow rate of 10L/h,15L/h,20L/h and 30L/h available Flowsheet NW Ultrapure Water Systems 6 Water Tank 5 Conductivity Measurement UV-oxidator 2 Boost Pump from 30L water tank with anti-contamination design. RS232 data port to enable downloading of RO Module 4 3 Pre-treatment Module Pre-treatment Module Conductivity Measurement 1 Solenoid Valve data to PC or printer. Time-saving fixed volume dispensing program 9 UV Lamp Ultrafiltration Membrane 12 13 Conductivity Measurement allows routine volume to be dispensed 7 automatically. Primary Module Primary Module 11 14 0.22μm point-of-use Filter 8 Conductivity Measurement Requirements Pure Water Quality Ultrapure Water Quality Model Temperature Ionic Rejection NW** NW** UV NW** UF NW** VF Tap Water 1-5 bar 95%-99% Bacteria and Particulates Rejection Conductivity Resistivity at 25 Conductivity at 25 Flow Rate >99% 0-20μs/cm 10L/h, 15L/h, 20L/h, 30L/h 18.2 MΩ-cm 0.055μs/cm 1.0-1.5 L/min TOC Endotoxin <10ppb -- <5ppb -- <10ppb <0.001 Eu/ml <5ppb <0.001 Eu/ml Resistivity Cell Particulates( 0.22μm) Bacteria k =0.01cm -1 <1 per/ml <1 cfu/ml 32Kg 34Kg 36Kg 38Kg Notes. ** stands for hourly output of pure water. For example, output of NW10,NW15,NW20 and NW30 is 10L/h,15L/h,20L/h,30L/h respectively. Typical Applications NW: AAS, Common Analysis,Standard Buffer, GC NW UV: HPLC, IC, GC, GC/MS, TOC Analysis, ICP and ICP/MS NW UF: Cell and tissue culture, monoclonal antibody production, pyrogen sensitive application NW VF: DNASequence, RNase DNase free, DNAfree, PCR, IVF, 2-D-electrophoresis, critical cell culture, pyrogen sensitive application
Heal Force R PW Ultrapure Water Systems Heal Force PW systems produce ultrapure water from pre-treated water at flow rate from 1.0 to 1.5 L/min to meet the basic needs of scientific laboratories. Product Features Control Panel PW Series Microprocessor Control The integrated, microprocessor controlled system monitors the system status to provide high-purity water on demand. A self-diagnostic control feature monitors and LCD displays the system status,including the conductivity in μs/cm or the resistivity in MΩ-cm as well as the temperature, Liquid height of water tank and consumable change reminders to prevent a drop in water quality. Standard Purification Steps All models of PW series employ a 3-step purification process: Step 1: The primary purification modules are tailored to the feedwater source to optimize the performance of downstream purification media to ensure consistent results. Step 2: The polishing purification modules remove ionic and organic contaminants and provides 18.2 MΩ-cm (at 25 ) water for analytical applications and many other ultrapure water requirements. Step 3: Final purification of ultrapure water is carried out at the point of use by a pharmaceutical grade, 0.22μm membrane filter that is recommended for most analytical applications. PW System PW PW UV PW UF PW VF Boost Pump Primary Module UV-oxidator UV-sterilizer Ultrafiltration Point-of-use Filter
Model Temperature Requirements Resistivity at 25 Conductivity at 25 Ultrapure Water Quality TOC Endotoxin Conductivity/Resistivity Cell Particulates( 0.22μm) Bacteria PW PW UV PW UF PW VF Pre Purified Water -0.3~3.0 bar 18.2 MΩ-cm 0.055μs/cm 1.0-1.5 L/min <10ppb <5ppb <10ppb <5ppb -- -- <0.001 Eu/ml <0.001 Eu/ml k =0.01cm -1 <1 per/ml <1 cfu/ml 32Kg 34Kg 36Kg 38Kg UV Photo-oxidation: PW UV and PW VF systems have a dual wavelength UV-oxidizer for oxidation of organic contaminants at 185nm plus bacteria control at 254nm. The purified water has extremely low levels of inorganic and organic contaminants for the most demanding HPLC, GC-MS and other ultratrace organic applications. Ultrafiltration: PW UF and PW VF Systems include a ultrafiltration module for the production of pyrogen- and nuclease-free ultrapure water (<0.001 Eu/ml), particularly important for life science or molecular biology applications. UV Sterilizer UF Cartridge Typical Applications Flowsheet PW ultrapure water systems PW: AAS, Common Analysis,Standard Buffer, GC PW UV: HPLC, IC, GC, GC/MS, TOC Analysis, ICP and ICP/MS PW UF: Cell and tissue culture, monoclonal antibody production, pyrogen sensitive application PW VF: DNASequence, RNase DNase free, DNAfree, PCR, IVF, 2-D-electrophoresis, critical cell culture, pyrogen sensitive application 2 Primary Catridge Primary Catridge 1 Solenoid Valve Ultrafiltration Membrane 3 UV Lamp 6 4 TOC Measurement 5 Conducttivity Measurement 7 8 0.22μm point-of-use Filter
Heal Force R ROP Purewater System The ROP system combines new reverse-osmosis technology with deionization resins that instantly provides deionized water (resistivity >10 MΩ-cm) directly from tap water. Suitable for users requiring purified water output 10 to 60 L/h. ROP(RO) Series Product Features The ROP system produce high quality pure water of high resistivity and low TOC (<30 ppb). The system's product water is stored in a water tank with anti-contamination design and is ideal for routine laboratory applications. Combining several purification technologies including Reverse Osmosis, Deionization and UV-sterilizer, the ROP system removes all types of water contaminants (typically 95%- 99 %). The built in UV lamp is ideal for applications that require low bacteria levels. The consumable includes the pre-treatment, reverse osmosis membrane and ion-exchange resin. LCD display shows the system status,including the conductivity in μs/cm or the resistivity in MΩ-cm, temperature and water level in water tank. ROP(RO) 30L Water Tank Typical Applications of ROP SYSTEM Feed for laboratory ultrapure water systems General chemistry Laboratory washing machines,including final rinse Feed for autoclaves and climate chambers Buffer preparation Photometry Spectrophotometry Media preparation
Model Feed water Temperature Requirements Resistivity at 25 Rejection Rate for Bacteria Pure water quality Flow Rate TOC Resistivity Cell Particulates( 0.22μm) Bacteria Power Consumption ROP10 ROP15 ROP20 ROP30 Tap Water 1-5 bar 10-15 MΩ-cm >99% 10L/h 15L/h 20L/h 30L/h 1-1.5L/m 1-1.5L/m 1-1.5L/m 1-1.5L/m <30 ppb k =0.01cm -1 <1 per/ml <1 cfu/ml 32Kg 34Kg 36Kg 38Kg 0.05kW/h Product Features Heal Force R RO Series Pure Water System Automatic micro-computer control and intelligent control Big LCD display to show temperature, conductivity/resistance, water level of water tank, quantitative outlet water and other operating states or related information High-precision resistance sensor Automatically rinsing the reverse osmosis membrane RS232 interface which conforms to GLP standard ROP(RO) Control Panel Typical Applications of RO SYSTEM Feed for laboratory ultrapure water systems General chemistry Laboratory washing machines Feed for autoclaves and climate chambers Buffer preparation Feed water Requirements Pure water quality Model Temperature Ionic Rejection Bacteria Rejection Size RO10 RO15 RO20 RO30 Tap Water 1-5 bar 95%-99% >99% 10L/h 15L/h 20L/h 30L/h 340X540X500mm(DXWXH) 18Kg 19Kg 18Kg 19Kg
Heal Force R CROB Analyser Feed System Heal Force is an acknowledged specialists in water purification for Clinical Biochemical Analyzer worldwide. We offer Water Purification Systems for Chemical Analyzers Specifically designed to produce NCCLS Type I reagent water for feed to clinical analyzers, Heal Force's AFS systems are used to support a variety of on board applications from reagent reconstitution, probe rinsing and cuvette washing. It is important to use an analyzer feed system for tap water purification to provide perfect Result to Chemical Analyzers. CROB (10,15,20,30) Inlet water Output Water Quality Model Temperature Resistivity at 25 Flow Rate TOC Bacteria Particulates( 0.22μm) Power Consumption CROB10 CROB15 CROB20 CROB30 CROB50 CROB70 Tap Water 1.0-4.0 kg/cm 2 10 MΩ-cm 10L/h 15 L/h 20 L/h 30 L/h 50 L/h 70 L/h 1-1.5L/m 1-1.5L/m 1-1.5L/m 1-1.5L/m 1-1.5L/m 1-1.5L/m <30 ppb <1 cfu/ml <1 per/ml 0.1kW/h 28Kg 28Kg 30Kg 30Kg 65Kg 65Kg CROB (,50,70) Nison Instrument(Shanghai)Limited International Sales Office:Floor 15th,No.2 Hua Shan Road,Shanghai 200040,PR China Tel:+86 21 62728646 Fax:+86 21 62710529 E-mail:export@hf.healoo.com Website:www.healforce.com