Gafchromic XR-RV2 radiochromic films dosimetry in interventional radiology

Similar documents
Patient skin doses in complex cardiac procedures

A dosimetric trial for the clinical follow-up of potential skin injuries on patients undergoing interventional cardiology procedures

Patient Dose Research in Interventional Radiology Suites in Rio de Janeiro, Brazil

Radiation Protection Dosimetry (2008), Vol. 132, No. 1, pp Advance Access publication 25 September 2008

Quality ID #145: Radiology: Exposure Dose or Time Reported for Procedures Using Fluoroscopy National Quality Strategy Domain: Patient Safety

Function of this control

PATIENT DOSIMETRY IN INTERVENTIONAL RADIOLOGY

Investigation of dose distribution to breasts in coronary angiography

Patient doses in lower limb angiography in a recently installed flat detector x-ray system

Evaluation of organ and effective doses for CAG based on in-phantom dosimetry

8/12/2011. Calculating Patient-specific Skin Dose. Call to Action! Fluoroscopy time poor measure of PSD Outline

Establishing a patient safety program in Interventional Radiology. A. Kyle Jones, Ph.D., DABR Assistant Professor MD Anderson Cancer Center

Reference Value Assessment for DAP Measured with a Transmission Ion Chamber in Gastrointestinal Tract Examinations

RADIATION EXPOSURE IN ENDOVASCULAR PROCEDURES - AN ASSESSMENT BASED ON EXAMINATIONS PERFORMED AT SAHLGRENSKA UNIVERSITY HOSPITAL

Radiation Dose Terms Made Simple v1

Radiography Curriculum Analysis

On the Importance of Computation in Clinical Radiology

Fluoroscopy units come in many different physical configurations but all consist of the main components shown in this slide.

Importance of a Patient Dosimetry and Clinical Follow-up Program in the Detection of Radiodermatitis After Long Percutaneous Coronary Interventions

Disclaimer and Notices.

Making the difference with Live Image Guidance Dose management in Allura systems

Types of Radiation Units

Progress in X-Ray & MR

S a f e t y R e p o r t s S e r i e s N o. 5 9

ESTIMATION OF ORGAN DOSES OF PATIENT UNDERGOING HEPATIC CHEMOEMBOLIZATION PROCEDURES

Radiation Physics and Safety Considerations in the EP Lab

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May Dose Reference Levels

DEPARTMENT OF ENVIRONMENT, HEALTH & SAFETY RADIATION SAFETY DIVISION. University of California, Los Angeles RADIATION SAFETY GUIDANCE

DoseWise Portal. Take control of dose management. across your organization

Take control of dose management

Absorbed Dose and Air Kerma Primary Standards

Experience With Patient Dosimetry and Quality Control Online for Diagnostic and Interventional Radiology Using DICOM Services

Brachytherapy Practice Problems EXTRA

Reduction of Backscattered Radiation in Enclosure X-ray Radiography

Protecting Patients 3/5/2014. Educating and Engaging your Staff in Radiation Safety

Artis Q and ArtisQ.zen

Department of Radio Diagnosis

First Experiences with the Ziehm Vision FD Mobile C-Arm with Flat-Panel Detector

PERFORM Operating Document. Radiation Dose Estimates, Calculations and Tracking Methodology

A New Tool for Benchmarking Cardiovascular Fluoroscopes a

2017 ACR Computed Tomography Quality Control Manual FAQS

FLUOROSCOPIC SAFETY AND PATIENT AND OPERATOR RADIATION DOSE REDUCTION: CREATING A FOCUSED FLUOROSCOPIC CURRICULUM FOR RADIOLOGY RESIDENTS

RADIATION EXPOSURE FROM IMAGING GUIDANCE USING THE EXACTRAC IMAGERY SYSTEM ESTIMATED BY PCXMC SOFTWARE. Kathleen Smith EPFL, M.Sci.

PHYSICS OF DIAGNOSTIC RADIOLOGY: MDPH

Prevention of radiation injury from medical imaging

Donna Newman B.A. RT ( R) CNMT Director of Professional Practice ISRRT

An optimal protocol for full-spine radiography

CTA Throughout the Ages

Introductory Presentation: EURAMED s Vision on Medical Radiation Protection (Research)

University of Cape Town

Making the difference

Quality Assurance for particle beam therapy

} } 8/3/2016. Accounting for kv Imaging Dose. kilovoltage imaging devices/techniques. Current imaging dose determination methods

Ensuring Uniformity of Measurements of Absorbed Doses of Intensive Photon and Electron Radiation in Radiation Technologies

Verification of dose limitation of the general public and determination of lead equivalence of x-ray rooms at Karolinska University Hospital Huddinge

1. Executive Summary

Quality Estimation of Filtration of Diagnostic X-Ray at Federal Medical Centre and Bishop Murray Hospital Makurdi using Half-Value Layer (HVL)

Madan Rehani. ICRP Committee 3

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

TENDER # SUPPLY OF MOBILE C-ARM RADIOGRAPHY/FLUOROCOPY SYSTEM FOR WESTERN HEALTH

R/F. Experiences Using SONIALVISION safire and the Utility of Tomosynthesis. 1. Introduction. 2. Basics of Tomosynthesis.

Employer s Unit of Competence Radiographic testing of materials, products and plant

Low-dose and High-resolution Cardiovascular Imaging with Revolution* CT

X-ray Equipment Safety Orientation. May 4, 2018

25 TEXAS ADMINISTRATIVE CODE (h) Radiation Safety Requirements for Accelerators, Therapeutic Radiation Machines and Simulators Used

Optima IGS 330 Versatility within your reach. gehealthcare.com

PATIENT RADIATION DOSE RANGES FOR PROCEDURES IN UNIVERSITAS HOSPITAL VASCULAR LABORATORIES

Clinical Applications. ImagingRite. Neuro Intervention. 1 ImagingRite

Environmental Regulator's RWA Syllabus mapped to PHE's RPTS courses. RPTS coverage EA. No

Fluoroscopy Testing Workshop

Authors: (see complete list in Appendix A) History and comments: (see complete details in Appendix B) Preface

New York State Department of Health Bureau of Environmental Radiation Protection

Clinical Applications. ImagingRite. Interventional Radiology

Vascular. Improvement of Functions Supporting Neuro-IVR. 1. Introduction. 2. Current State of Functions Supporting Neuro-IVR

ABSTRACT INTRODUCTION

Request for Quotations/Proposal

Growth Strategy for Medical Systems Business

Optimize efficiency with a universal system.

A New Standard for the Future

Research Article Shielding Effects on Low-KVIORT Forward Projecting Applicators

How Tracking Radiologic Procedures and Dose Helps: Experience From Finland

Streamline workflow for higher productivity and patient throughput. DirectView. Classic CR System

Deviation Value for Conventional X-ray in Hospitals in South Sulawesi Province from 2014 to 2016

Radiation Safety Issues in Fluoroscopy During Percutaneous Nephrolithotomy

Virtual Mold Room. The Future of Custom Radiation Therapy Devices

7/31/2017. Implementing a Clinical Practice Guideline; Lessons From An Early Adopter of MPPG 5.a. Disclosures

UNIVERSITY OF CALGARY. An X-ray Source Model and Characterization Method. for Computing kv Radiation Dose. Yannick Poirier A THESIS

New Material for Shields in X-Ray Diagnostic Procedures

3/6/2017 TOMOSYNTHESIS GUIDED BREAST BIOPSY LEARNING OBJECTIVES DISCLOSURES

Computed Tomography: Optimization of acquisition protocols & Justification of clinical referrals. Koos Geleijns, medical physicist

Handzettel 1. CARE Right Computed Tomography, committed to the right dose Ivo Driesser Austin, July 22 nd 2014

IAEA-TECDOC Dose Reduction in CT while Maintaining Diagnostic Confidence: A Feasibility/Demonstration Study

Radiation protection CRISTER CEBERG

Vascular and Interventional Radiology Original Research

Act anything done, being done, or to be done; the process of doing. Synonymous with procedure and clinical services.


MRI Conditional Devices and Patient Safety. Outline. MRI Labeling: Objects 8/3/13 1. David W. Jordan, Ph.D. Senior Medical Physicist

IAEA HUMAN HEALTH SERIES

Transcription:

Gafchromic XR-RV2 radiochromic films dosimetry in interventional radiology A. Neocleous *1, E. Yakoumakis 1, G. Gialousis 2, A. Dimitriadis 1, N. Yakoumakis 3 and E. Georgiou 1 1 Medical Physics Department, Medical School, University of Athens, 75 Mikras Asias Str., Goudi 11527, Athens, Greece 2 IASO Hospital, Medical Physics Department, 37-39 Kifisias Str., Maroussi 15123, Athens, Greece 3 Medical Physics Department, Medical School, University of Patras, Rio, Greece *presenting author: ask4areti@gmail.com

1. Introduction Patient skin injuries in interventional procedures are fully documented in the scientific literature. Research into new methods for patient dose measurement should be encouraged. There is a need to define and group IR procedures in order to account for their complexities since a number of parameters like fluoroscopy duration and x-ray beam orientation affect the differences in peak skin dose measurement. Mapping skin doses, with the use of a dosimetry film, is useful to determine peak skin dose and to detect areas of overlapping radiation fields. a b c (a) Condition of patient s back, six to eight weeks following multiple coronary angiography and angioplasty procedures. (b) Appearance of skin injury approximately 16 to 31 weeks following the procedures with small ulcerated area present. (c) Appearance of skin injury approximately 18 to 21 months following procedures evidencing tissue necrosis. Thomas B. Shope, Ph.D. Radiation-induced skin injuries from fluoroscopy., Center for Devices and Radiological Health. FDA, Rockville, 2010

2. Purpose The aim of the work was: to use the new Gafchromic XR-RV2 film and / to measure patient skin dose during interventional radiology procedures to evaluate dose distributions using the film for some frequent procedures, identified in three categories: 1. Nefrostomies (PCN) and urinary stenting 2. Biliary stenting and PTBD 3. Aorta stent grafting

3. Materials and Methods Materials: Gafchromic XR-RV2 film Number of patients: 45 interventional procedures Tips: The film was placed in the area of interest between the patient and the exam table Dosimetric data such as fluoroscopy duration and dose-area product for fluoroscopy and cine radiography were recorded Exposed film after an interventional procedure

3.1 Characteristics and calibration of the Gafchromic XR-RV2 film Cafchromic XR-RV2 film Measures doses up to 50 Gy from photons with energies ranged from 30 kev to 30 MeV Self-developing film media Handled in normal room light with no hazardous chemicals or waste to dispose of Can be cut to any size No dose fractionation and dose rate dependence were recorded Principal absorption peak at 675 nm The colour of the film is orange and becomes darkened according to radiation exposure. Darker film areas represent higher absorbed doses. Gafchromic XR-RV2 films exposed to radiation. The last piece is unexposed.

Film Calibration Set up & Measurements Five 6 cm x 6 cm sections of the film were irradiated using a radiografic x-ray tube (Philips Integris 5000 C-arm). Each piece of film was placed at 5 cm plexiglas thickness centered in the radiation field (10 cm x 10 cm) at a distance SSD=40 cm in order to include backscattered radiation in dose measurement. Electronic dosimeter with a solid state detector (Solidose Model 400, R100, RTI Electronics) was used to measure the exposure rate in air. Backscattered factor was used to correct the dose reading (BSF=1,30). The peak potential of the x-ray unit was adjusted to 100 kvp (HVL=3,6 mm Al) and the film pieces were repeatedly irradiated to reach high doses (0-7000 mgy). Processing & Results Exposed films were scanned by a flat bed scanner (Microtek ScanMaker 9800XL, TMA 1600) on the reflective mode and red pixel values were extracted from the RGB image using Matlab routine. The film response (FR) was read in its red component. FR=log 10 {(pixel value) D0 /(pixel value) Di } The calibration curve of dose values vs film response was generated. X-ray tube D (mgy) FR 8000 Calibration curve Dosimeter 750,57 0,186 7000 6000 y = 135394x 3-62505x 2 + 11812x R 2 = 0.9891 Film 40 cm 1438,60 2814,65 0,267 0,371 dose (mgy) 5000 4000 3000 2000 Plexiglas (d=5 cm) 4878,73 7005,35 0,459 0,490 1000 0 0 0.1 0.2 0.3 0.4 0.5 0.6 Film Response

3.2 Measuring skin dose during interventional procedures Three days after the procedure, the film was scanned by a flat bed scanner on the reflective mode and the RGB image was stored as 48 bit in.tiff format for further analysis. Matlab routine has been chosen for image processing analysis. After the extract of the red component from the original image, red pixel values were converted firstly to film response (FR) and secondly to dose according to the calibration curve. A numerical matrix of dose values was generated for subsequent processing and peak skin dose (PSD) was defined. Dose matrix 2170.93 2076.36 2158.84 2244.44 2358.55 2021.95 2234.66 pixel 2182.87 2253.80 2188.53 2050.11 2236.28 2209.51 2228.40 2209.05 2189.21 2188.76 2204.47 2237.45 2172.96 2177.23 2067.64 2166.89 2177.68 2044.65 2166.67 2122.16 2481.51 2189.44 2103.88 2143.95 2141.74 2149.49 2248.17 2048.85 2231.87 2302.48 2246.30 2084.70 2457.53 2088.78 2152.16 2026.10 2150.38 2188.31 2185.14 2188.31 2181.97 2123.48

4. Results and Conclusion Mean and standard deviation for fluoroscopy duration, fluoroscopy and cine DAP and patients skin doses are summarized in table. For each procedure the number of patient reports is given in parenthesis. Fluoroscopy duration (min) Fluoroscopy DAP (Gy.cm 2 ) Cine radiography DAP (Gy.cm 2 ) Peak Skin Dose (mgy) Mean Stand. Mean Stand. Mean Stand. Mean Stand. Dev. Dev. Dev. Dev. Nefrostomies (PCN) & Urinary stent (19) 6,8 2,8 60 26 6 5 534 119 PTBD & Biliary stent (12) 24,8 16,6 99 40 35 25 1339 631 Aorta stent grafting (14) 24,0 16,6 223 109 1062 240 2434 569

Aorta stent grafting procedure involve fluoroscopy and cine radiography mode in use. Digital acquisition contributes in great amount to dose delivered to patient. Nefrostomy and urinary stenting is almost a routine procedure in an interventional radiology department with relatively low duration and dose to patient. STENT GRAFT STENT GRAFT 223 1062 BILIARY STENT & PTBD BILIARY STENT & PTBD 35 99 PCN & URINARY STENT PCN & URINARY STENT 60 6 0 5 10 15 20 25 tim e duration (m in) FLUORO CINE 0 500 1000 1500 DAP (Gy.cm ^2)

Gafchromic XR-RV2 film is suitable for complex high dose interventional procedures guided by fluoroscopy. Dose in complex interventional procedures like aorta stent grafting is above the threshold of 2 Gy for erythema. Those procedures should be recorded and distinguished with reference to patient s benefit. Long time fluoroscopy combined with cine radiography make large x-ray dose distributions and overlapped radiation fields result entrance doses near the erythema dose threshold. Those areas of the patient s skin that received high dose should be identified and noted in the patient file record.