Sterilization - validation, qualification requirements. Sterilization - Overview

Similar documents
5.1.1 & Changes started in 2008 and were mainly driven under the Chairmanship of Prof Hans van Doorne 11/10/2017

Annex A2. Guidance on Process Validation Scheme for Aseptically Processed Products

STERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation

REUSABLE MEDICAL DEVICE VALIDATION

GUIDE TO INSPECTIONS OF STERILE DRUG SUBSTANCE MANUFACTURERS

Aseptic Process Validation

ISO INTERNATIONAL STANDARD

EQUIPMENT FOR AUTOCLAVING

ISO INTERNATIONAL STANDARD

CAN STERILE REALLY BE STERILE?: EFFECTIVE AUTOCLAVING METHODS

Technical Report No. 48 Moist Heat Sterilizer Systems: Design, Commissioning, Operation, Qualification and Maintenance

Review Validation of aseptic processes for pharmaceuticals

Effective Heat Sterilization in CO 2 Incubators

Sterilisation Monitoring Products

Important Terminology (pg )

NO 2 Gas Sterilization and Decontamination: Keeping Pace, from Start to Finish

GUIDELINE C-17. Non-Incineration Technologies for Treatment of Biomedical Waste (Procedures for Microbiological Testing)

PRODUCT CATALOGUE. for gke Steri-Record Biological Indicators, Process Challenge Devices and Accessories

PRODUCT CATALOGUE. for gke Steri-Record Biological Indicators, Process Challenge Devices and Accessories

Technical Data Monograph. STERIS VHP LTS-V - Low Temperature Surfaces Terminal Sterilization

Doc No.: MHD 12(0333)/ISO :2006 TITLE: Sterilization of health care products Biological indicators Part 1: General requirements

Case Study: Isolator Sanitisation Cycle Development, Validation and Revalidation.

Written By Dr. Robert Ramaley (Univ. NE Med CTR), Mia Ware, Dan Pearson, and Christopher R. Dwyer* (Raven Biological Laboratories)

Biological Monitoring

Sterilization & Disinfection

Nitrogen Dioxide (NO 2 ) PDA: A Global. Sterilization and. Association. Decontamination Systems. November 4, PDA Europe Parenterals 2014

HISTORY AND MILESTONES

Secure. Tested. Certified.

ISO Part 1 and Part 2 Compliance Requirements. Cathriona O Neill

The Parenteral Drug Association

Hot Topics in Drug Product Process Validation: A Reviewer s Perspective

Critical Environment Products and Services

Instructions for Validation of Autoclaves

Product Catalogue. for gke Steri-Record Biological Indicators, Process Challenge Devices and Accessories

Slide 1. Slide 2. Slide 3. Sponsored by Webber Training Inc. For more information contact Paul Webber

2018 North America Lab Services Fee Schedule

VALIDATION OF BUNG PROCESSOR

Chemical Indicators. Biological Indicators. Physical Monitors. Welcome! How do I get a CE Certificate? Learning Objectives

sterilisation monitoring products

Sterility Assurance Level and Aseptic Manufacturing Process in Pharmaceuticals

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

Space product assurance

Advancing the Science of Gamma Irradiation Continuous improvement in radiation sterilization

VALIDATION QUALIFICATION. Pharmaceutical Consultancy Services, All rights reserved. 1. JaapKoster

ISO INTERNATIONAL STANDARD. Sterilization of health care products Chemical indicators Part 1: General requirements

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

Tests to Support Sterility Claim. Imtiaz Ahmed

Chapter 9 Controlling Microbial Growth in the Environment. 10/1/ MDufilho

Validation Guide USTR 2114 (2) Validation Guide for Pall Emflon PFR Filter Cartridges

The Control of Microbial Growth

Annex 1 to the Good manufacturing practices guide Manufacture of sterile drugs GUI-0119

USP Chapter 823 USP 32 (old) vs. USP 35 (new)

Parametric Release An Industry Perspective. October 22, Mike Sadowski Director Sterile Manufacture Support Baxter Healthcare Corporation

Validation Guide Sterisart NF

EH&S. Sheet. Fact. Safe and Effective Use of Autoclaves. What are autoclaves? Factors for effective sterilization. Dry heat cycle - when to use

Meeting the challenges of medical device packaging

Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar

ISO INTERNATIONAL STANDARD

Sterile Compounding elearning Course Curriculum 28 lessons with 33 hours of CE. Fundamentals of Sterile Compounding (8 lessons/8 hours CE)

Chuck Reed at Weiler Engineering, Inc compares aseptic blow-fill-seal technology with traditional aseptic processing of pharmaceutical liquids

Sterile Compounding elearning Course Curriculum 28 lessons with 33 hours of CE. Fundamentals of Sterile Compounding (8 lessons/8 hours CE)

ASTM D10-F02 Workshop FDA Regulatory Perspective. Patrick Weixel FDA Center for Devices and Radiological Health

INTERNATIONAL STANDARDS ON FLUID PREPARATION FOR CONVECTIVE THERAPIES. Richard A. Ward, Ph.D.

PERFORMANCE QUALIFICATION PROTOCOL HVAC SYSTEM

Media Fill A Process Simution. Presented By Shikha Chauhan

Why change is inevitable in aseptic manufacturing?

Kevin Lorcheim Project Engineer

PDA: A Global. Association. Matrix Approach to Media Fills. (c) 2012 Catalent Pharma Solutions. All rights reserved.

DESIGN, FABRICATION AND PERFORMANCE EVALUATION OF CONTINUOUS OPERATING AUTOCLAVE

Lets Talk Indicators Challenging the process Sterilization. ICAN 5 th November Peter Newson

Whitepaper. High temperature HEPA filtration. Dr.-Ing. Marc Schmidt, Dr.-Ing. Lothar Gail, Hugo Hemel MSc. Preview

Temperature mapping and autoclave validation specialists

Validation of Sterilizing Grade Filters

Sterilization & Chemical Resistance of Healthcare Polymers

Specialty Testing of Steam Sterilizers

Glossary of Freeze Drying Terms

2-2 Sterile Product Assessment. Satish Mallya Ph.D Training Workshop CPH, May 2015

Steam Sterilizers. DLOV & DLOM series. more than sixty years of experience in sterilization. DE LAMA s Quality Management System is certified:

Decon 2.0: Emerging Decontamination Technologies

Principles of Chlorine Dioxide Gas as a Decontamination Method

Analytical and formulation attributes

Medline Gemini Sterilization Wrap. Test Studies

Hygiene Monitoring. Biological Indicators for Monitoring Sterilization Processes. Product Information. BI_07_en

Chapter 8 Control of Microorganisms by Physical and Chemical Agents

ANNEX 5 MANUFACTURE OF IMMUNOLOGICAL VETERINARY MEDICINAL PRODUCTS

Microbial Growth and The Control of Microbial Growth (Chapter 6 & 7)

gke Steri-Record Testsets and PCDs for validation, process and batch monitoring of sterilization processes

Guidelines for Process Validation of Pharmaceutical Dosage Forms

For sterile products manufacturing, Lyoseal is an innovative technology. presented by Mr. Tony Bouzerar from BioCorp offers us detailed

Part 1: General requirements

AS/NZS 4187 RELEVANT STANDARDS. Robyn Williams NUM CSSD LBHHS & Standards Australia Chair HE-023

Biological Indicators and H 2 O 2 Decontamination of Isolator Systems. Volker Sigwarth Skan AG Switzerland. Process Development

Advancements on implementation of single use technology in vaccine manufacturing

Medical and Pharmaceutical Packaging Styles

GUIDELINE FOR THE STABILITY TESTING

Quality Control EQUIPMENT CONTROL. Monitor critical performance characteristics of sterilization, decontamination or cleaning equipment

Autoclave Installation and Operation Guidelines

Microbiological Cleaning Method Validation

Guidelines for Validation of Blanching Processes

8/27/2015. Learning Objectives. Purpose of Sterilization. To destroy all living pathogenic microorganisms

Transcription:

Sterilization - validation, qualification requirements Dawn Tavalsky 1 ISPE Boston Chapter Framingham MA 19Sep2013 Sterilization - Overview Objectives Discuss definition of Sterile Briefly describe sterilization methods Describe approaches to be used for the validation of a sterilization process using Moist Heat as an example Describe requirements for routine monitoring and control of sterilization Review issues that are specific to other sterilization processes 2 1

Sterile Products - Overview Certain pharmaceutical products must be sterile injections, ophthalmic preparations, irrigations solutions, haemodialysis solutions Two categories of sterile products those that can be sterilised in final container (terminally sterilised) those that cannot be terminally sterilised and must be aseptically prepared 3 Sterilization - Overview What is the definition of sterile? Free from microorganisms In practice no such absolute statement regarding absence of microorganisms can be proven Defined as the probability of 1 in a million of a container being contaminated (10-6 ) This referred to as the Sterility Assurance Level (SAL) Organisms are killed in an exponential fashion 4 2

Definition of Sterile MICROBIAL INACTIVATION N t = N o e -kt 1200 3.50 3.00 1000 2.50 2.00 No of surviving organisms 800 600 400 1.50 1.00 0.50 0.00 No remaining Log No remaining -0.50 200-1.00-1.50 0 0 1 2 3 4 5 6 7 8 9 10 No remaining 1000 368 135 50 18 7 2 1 0 0 0 Log No remaining 3.00 2.57 2.13 1.70 1.26 0.83 0.39-0.04-0.47-0.91-1.34 Time -2.00 5 Definition of Sterile Resistance of an organism is referred as its D-value D-Value D-value - Time (or dose) required to reduce the population of organisms by 1 log (or 90%) 6 3

Definition of Sterile A sterilization process must deliver a Sterility Assurance Level (SAL) of 1 in a million (10-6 ) It is not possible to measure 10-6 The required SAL can be achieved by applying a process that will reduce the number of organisms to zero and then apply a safety factor that will deliver an extra 6 log reduction 7 Definition of Sterile Example For an initial bioburden of 10 2 the sterilization process will need to achieve an 8 log reduction in viable organisms This will require 8 times the D- value (e.g. if the organism has a D value of 2 minutes then 8 x 2 = 16 minutes will be required to achieve an 8 log reduction and an SAL of 10-6 ) (Point Z) 8 4

Sterilization - Overview Commonly used methods of sterilization Moist Heat Dry Heat Gas (Ethylene oxide) Radiation (Gamma or Electron) Filtration Others - UV, Steam and formaldehyde, hydrogen peroxide 9 Saturated steam Common cycles: Moist Heat 121 C for 15 minutes 134 C for 3 minutes Other cycles of lower temperature and longer time may be used (e.g. 115 C for 30 minutes) Used for sterilization of: terminal sterilization of aqueous injections, ophthalmic preparations, irrigation & haemodialysis solutions, equipment used in aseptic processing 10 5

Moist Heat not suitable for non-aqueous/dry preparations preferred method of sterilization 11 Dry Heat Lethality due to oxidative processes Higher temperatures and longer exposure times required Typical cycles: 160 C for 120 minutes 170 C for 60 minutes 180 C for 30 minutes tunnels used for the sterilisation of glass vials may use much higher temperatures (300 ) for a much shorter period 12 6

Dry Heat Used for: glassware and product containers used in aseptic manufacture, non aqueous thermostable powders and liquids (oils) also used for depyrogenation of glassware ( 250 C) (Pyrogens - substances found in cell wall of some bacteria which can cause fever when introduced into the body) 13 Ethylene Oxide Gas Either pure or in mixtures with other inert gases Requires presence of moisture Complex process Typical cycles: 1-24hours 25-1200 mg/l gas 25-65 C 30-85% relative humidity 14 7

Used for: Ethylene Oxide heat labile product containers surface sterilization of powders Adequate aeration to reduce toxic residues 15 Radiation Gamma rays generated by Cobalt 60 or Caesium 137 radionuclides; or Accelerated electrons from an electron generator 25 kilograys (kgy) usual dose dose dependent on bioburden (resistance of organisms not predictable) process must be properly validated used for: dry pharmaceutical products heat labile product containers can cause unacceptable changes 16 8

Filtration Removes organisms from liquids and gasses 0.2-0.22 micron for sterilization composed of cellulose esters or other polymeric materials filter material must be compatible with liquid being filtered used for bulk liquids, gasses and vent filters 17 Validation - Overview Selection of sterilzation process must be appropriate for product terminal sterilization is the method of choice moist heat (autoclaving) is the most common process used for terminal sterilization product must not be affected by heat container/closure integrity must be established items being sterilised must contain water (if sealed) or material must allow for removal of air and penetration of steam for steam (moist heat) sterilization 18 9

Validation - Protocol Requirements for Moist Heat Sterilization Other processes follow similar requirements Validation protocol should include the following details for each sterilization process process objectives in terms of product type, container/closure system, SAL required specifications for time, temperature, pressure and loading pattern description of all equipment and support systems in terms of type, model, capacity and operating range 19 Validation - Protocol Moist Heat continued: performance characteristics of all equipment e.g. pressure gauges, valves, alarm systems, timers, steam flow rates/pressures, cooling water flow rates, cycle controller functions, door closure gasketing and air break systems and filters methodology for monitoring performance of equipment and the process and labatory testing methodology personnel responsible for all stages and final evaluation (should have experience and necessary training and be authorized) 20 10

Validation - Calibration Laboratory testing should be performed by a competent laboratory, methodology should be documented All instruments must be calibrated e.g. temperature recorders and sensors thermocouples pressure sensors for jacket and chamber timers conductivity monitors for cooling water flow metres for water/steam water level indicators when cooling water is used thermometers including those for thermocouple reference, chamber monitoring and laboratory testing 21 Validation - Calibration Indicators should be calibrated physical and chemcial indicators should be tested to demonstrate acceptable response to time and temperature biological indicators should be tested for count and time/temperature exposure response for commercial indicators - test certificate with count and D-value and exposure response should be available. Results acceptable if verified in house periodically. In house indicators must be fully characterized (Dvalue, identification) and appropriate for sterilization process All indicators should be appropriately stored and within expiry 22 11

Validation - Cycle Development Concept of F o Lethality factor equivalent to time at 121 C 1 minute at 121 C is equivalent to F o of 1. Lethality can accumulate during heat up and cool down phases Typical temperature profile of a heat sterilization process What would be the Fo of a cycle at 121 C for 15 minutes? 23 Validation - Cycle Development F o is calculated using the following equation: F o = tσ10 (T-121/Z) where: t is the time interval between measurements of temperature (T) T is the temperature of sterilised product at time (t) Z is a temperature coefficient which measures the number of degrees required to change the D-value of an organism by 1 log 24 12

Validation - Cycle Development The minimum F o required by a sterilzation process is related to the resistance of the bioburden (D-value) F o = D 121 (LogA - Log B) where: D 121 is equal to the time at 121 C to reduce the population of the most resistant organism in each product container by 90% (or 1 log) A is the number of microoganisms per container B is the maximum acceptable probability of survival (Sterility Assurance Level, 10-6 ) 25 Validation - Cycle Development 26 Two approaches to sterilization Overkill Probability of survival Overkill approach used when the product can withstand excessive heat treatment without adverse effects Cycle should deliver an F o of at least 12 This will achieve a 12 log reduction of microorganisms with a D-value of 1 minute (Assuming each product unit contains 10 6 organisms a 12 log reduction will result in 10-6 organisms per unit or probability of survival (SAL) of 1 in a million) (Normal bioburden is usually much lower and the organisms normally much less resistant than this) 13

Validation - Cycle Development Biological Indicators device consisting of a known number of microorganisms, of a known resistance to a particular sterilization process in or on a carrier and enclosed in a protective package. Organisms are in the form of endospores (not vegetative state) as these are most resistant to sterilization 27 Validation - Cycle Development Spore Strips - a narrow strip of fibrous paper impregnated with a bacterial spore suspension contained in a glassine envelope Spore Dots - Circular pieces of fibrous paper impregnated with the spore suspension Spore Suspensions - pure spore suspension of the desired challenge organism which can be inocluated onto the surface of a material Self contained units containing spore strips or suspensions and the media in which they are to be incubated (simple/convenient to use) 28 14

Validation - Cycle Development Probability of Survival approach used used for heat labile products The process is validated to achieve a destruction of the presterilization bioburden to a level of 10 0 (Point Y), with a minimum safety factor of an additional six-log reduction (Point Z) Determination of the minimum F o required is based on the bioburden and its heat resistance What F o would be required for a bioburden of 10 2 (100) if D-value was 1? 29 Validation of Sterilization 30 Basic Principles Installation Qualification (IQ) Ensuring equipment is installed as per manufacturer s specification Operation Qualification (OQ) Ensuring equipment, critical control equipment and instrumentation are capable of operating within required parameters Performance Qualifcation (PQ) Demonstrating that sterilizing conditions are achieved in all parts of sterilization load Physical and microbiological 15

Validation - Equipment Installation Qualification Ensuring equipment is installed as per manufacturer s specification considerations for new and existing equipment specifications for the type of autoclave, construction materials, power supplies and support services, alarm and monitoring systems with tolerances and accuracy requirements for existing equipment documented evidence that the equipment can meet process specifications 31 Validation - Equipment Operational Qualification Ensuring equipment, critical control equipment and instrumentation are capable of operating within required parameters Three or more test runs which demonstrate controls, alarms, monitoring devices and operation indicators function chamber pressure integrity is maintained chamber vacuum is maintained (if applicable) written procedures accurately reflect equipment operation pre-set operation parameters are attained for each run 32 16

Validation - Performance 33 Performance Qualification Demonstrating that sterilizing conditions are achieved in all parts of sterilization load Physical and microbiological Physical Heat distribution studies on empty chamber maximum and minimum cycle times and temperatures to identify heat distribution patterns including slowest heating points mulitple temperature sensing devices should be used (thermocouples) location of devices should be documented and ensure that heat distribution is uniform Validation - Performance 34 Performance Qualification - Physical (2) Heat distribution of maximum and minimum chamber load configurations multiple thermocouples throughout chamber (not inside product containers) to determine effect of load configuration on temperature distribution temperature distribution for all loads using all container sizes used in production should be tested position of thermocouples should be documented Slowest to heat/cold spots in each run should be documented, inlcuding the drain repeat runs should be performed to check variability temperature distribution profile for each chamber load configuration should be documented 17

Validation - Performance Performance Qualification - Physical (3) Heat penetration studies to detect the maximum and minimum temperature within all loads all parts of each load must be on contact with steam need to determine lowest and highest temperature locations and slowest and fastest to heat locations (measured inside product containers) need to consider all variables such as container size, design, material, viscosity of solution and fill volume. Container with maximum fill volume and slowest to heat solution should be used maximum and minimum load configurations for each sterilization cycle using routine cycle parameters 35 Validation - Performance Performance Qualification - Physical (4) Heat penetration (2) May be necessary for container mapping for larger volumes - cold spot then used for penetration studies Need to consider effects of packaging e.g. overwrapping Three runs performed once cold spots have been identified to demonstrate reproducibility 36 18

Validation - Performance Performance Qualification - Microbiological Biological challenge studies used when Probability of Survival approach is used may not be necessary when cycle is > 121 C for 15 minutes (except US and Australia) biological indicators (BI) containing spores of Geobacillus stearothermophilus are most commonly used (considered worst case ). BIs containing other organisms may be used performance studies based on product bioburden require a considerable amount of work indicators should be placed throughout the load, adjacent to thermocouples, at cold spots and slowest to heat locations (identified during heat penetration studies) any growth is unacceptable unless processing errors demonsrated 37 Validation - Performance Validation report must demonstrate requirements in Validation protocol have been met, any deviations must be justified Requalification must be repeated on an annual basis (usually one run is acceptable) Any changes or modifications must be evaluated may just require requalification any changes to loading patterns, new container/closure systems or cycle parameters require full validation 38 19

Routine Production Issues considered for routine production Manufacturing environment should be controlled Procedures in place to minimize the presterilization bioburden bioburden limits specified (although not so important when overkill cycle used) Time between filling and sterilization should be specified Integrity of container/closure system should be periodically verified Periodic leak testing of chamber (if vacuum is part of cycle) 39 Routine Production Cooling water should be sterile Differentiation between sterilized and not-yet sterilized product Physical separation (double ended autoclave) Labelling and use of visual indicators (e.g autoclave tape) Periodic testing of containers to verify integrity of container/closure system Quality of steam should be defined and periodically tested for contaminants 40 20

Routine Production Each sterilization cycle must be monitored temperature, time and pressure recorded temperature recorder independent from cycle controller second independent temperature recorder drain temperature should be recorded chemical and biololgical indicators (if applicable) Sterilisation cycle records should form part of batch records 41 Other Sterilization Processes Sterilization using other processes should follow a similar approach as that described for moist heat Validation protocol Equipment calibration Determining the process that will deliver the desired SAL (10-6 ) IQ, OQ, PQ Requirements for routine monitoring and control 42 21

Other Sterilization Processes Dry Heat Should have air circulation in the chamber Positive pressure in the chamber to prevent entry of non-sterile air HEPA filtered air supplied Biological indicators containing Bacillus atropheus (if used) removal of endotoxin is usually sufficient When removing pyrogens need to validate process using challenge tests 43 Other Sterilization Processes Radiation Usually performed by contracting service (need to ensure validation status, responsibilities) Based on bioburden of product being sterilised Biological indicators may be used as additional control but may not be as resistant as naturally occuring bioburden Method defined in International Standard ISO 11137 44 22

Other Sterilization Processes Radiaton (2) Correct dose of radiation (~25 kgy) received by all products (measured with dosimeters) quantitative measurement number, location, within calibration time-limit Radiation sensitive colour discs applied to packaging procedures to distinguish irradiated and nonirradiated materials Variation in density of packaging should be addressed during validation 45 Other Sterilization Processes Gasses and Fumigants e.g. ethylene oxide, hydrogen peroxide vapour Only when no other method is suitable Must demonstrate that process does not adversely affect product Packaging must be able to permit ingress of gas and humidity Ensure product load is adequately heated and humidified prior to sterilization (called conditionning ) need to take into account validation performed in summer or winter 46 23

Other Sterilization Processes 47 Gasses and Fumigants (2) Temperature distribution is acceptable Concentration of sterilant gas is sufficient Use of biological indicators is important (Bacillus atropheus) Half cycles If cycle of half normal time destroys biological indicators (10 6 organisms), double time will achieve SAL of 10-6 Aeration Ventilated conditions Defined limits of residuals Process included in validation Safety and toxicity issues considered Useful Publications ISO/EN 17665 - Sterilization of health care products - Moist Heat (Parts 1 and 2) ISO/EN 11135 - Sterilization of health care products - Ethylene Oxide (Parts 1 and 2) ISO/EN 11137 - Sterilization of health care products - Radiation (Parts 1,2 and 3) Validation of Moist Heat Sterilization Processes: Cycle Design, Development, Qualification and ongoing Control. PDA Technical Report No. 1 Revised 2007 48 24

Questions? 49 25