Kilovoltage X-ray dosimetry in the clinic Philip Mayles The Clatterbridge Cancer Centre
Kilovoltage Codes of Practice Germany DIN 1988 and 1996 UK IPEMB (IPEM) 1996 and 2005 IAEA TRS 277 2nd edn 1997 Holland NCS 1997 USA AAPM TG61 2001 IAEA TRS 398 2004
Formula for Medium Energy Measure at 2 cm deep IPEM COP Klevenhagen et al Phys Med Biol 41 2605-2625 1996 In phantom
Formula for Low Energy Measure in air IPEM COP Klevenhagen et al Phys Med Biol 41 2605-2625 1996 In air
Formula for Low Energy Measure in air AAPM COP Ma et al Med Phys 28 868-893 2001 In air
Formula for Low Energy Measure in air IPEMB recommends use of the in air protocol if the prescription is to the max dose for medium energy IPEMB Addendum Aukett et al Phys Med Biol 50 2739-2748 2005 In air
IPEM code for Very Low Energy Measure at surface of phantom Note that k ch is based on measurement comparing to in air measurement IPEMB COP Klevenhagen et al Phys Med Biol 41 2605-2625 1996 IPEMB Addendum Aukett et al Phys Med Biol 50 2739-2748 2005 In phantom
Grenz ray chamber PTW23342 In phantom
Grenz ray chamber PTW23342 In phantom
How are Low and Medium Energy defined? IPEMB based on HVL Medium energy 0.5-4 mm Cu (>160 kv) Low energy 1 8 mm Al (50 160 kv) Very low energy 0.035-1 mm Al (8 50 kv)
How is Energy defined? Penetrative power Depth dose First Half Value Layer Energy spectrum Generating potential Filtration
Measurement of HVL AAPM COP Ma et al Med Phys 28 868-893 2001 In air
HVL mm Al HVL mm Al Specification of Beam Quality 3 2.5 2 Standards Lab TECDOC 1455 Clatterbridge 4.5 4 3.5 3 Standards Lab TECDOC 1455 Clatterbridge 1.5 2.5 2 1 1.5 0.5 1 0.5 0 0 20 40 60 80 100 120 0 0 50 100 150 200 250 300 kv kv Low Energy Medium Energy
Does Beam Quality Matter? N k HVL mm Al Adapted from TRS398
Factors in Calculation: M Electrometer reading corrected for temperature, pressure, ion recombination, polarity effect and electrometer accuracy AAPM COP Ma et al Med Phys 28 868-893 2001 In air In phantom
Factors in Calculation: M Electrometer reading corrected for temperature, pressure, ion recombination, polarity effect and electrometer accuracy IPEM COP Klevenhagen et al Phys Med Biol 41 2605-2625 1996 In air In phantom
Factors in Calculation: M Recombination correction V H and V L are the high and low voltages and M H and M L are the meter readings Important for short SSDs AAPM COP Ma et al Med Phys 28 868-893 2001 In air In phantom
Factors in Calculation: μ en /ρ air AAPM and IPEMB agree to better than 0.3% but the sources of data were not independent In air
Dose is tissue dependent μ en /ρ air AAPM COP Ma et al Med Phys 28 868-893 2001 In air
Factors in Calculation: μ en /ρ 2cm 1 10 100 AAPM COP Ma et al Med Phys 28 868-893 2001 In phantom
Factors in Calculation: B w Since the backscatter factor is fundamentally a water-kerma ratio, reliable measurements are non-trivial. Therefore for the application of this protocol backscatter factors should not be measured in the clinic AAPM COP Ma et al Med Phys 28 868-893 2001 In air
Factors in Calculation: B w Since the backscatter factor is fundamentally a water-kerma ratio, reliable measurements are non-trivial. Therefore for the application of this protocol backscatter factors should not be measured in the clinic Largest difference AAPM:IPEMB is 0.4% AAPM COP Ma et al Med Phys 28 868-893 2001 In air
Factors in Calculation: P stem 1.0 if field size for calibration is same as measurement Likely to be important only for plane parallel chambers Need to know the value for the reference chamber In air
Factors in Calculation: k ch AAPM: IPEM: P Qcham P sheath k ch In phantom
Factors in Calculation: k ch AAPM: IPEM: P Qcham P sheath k ch TRS277: k u p u In phantom
Phantom Materials Compared Solid Water (RMI Gammex) and Plastic Water (Computerized Imaging Reference Systems) to H 2 O Percentage Depth Dose Solid water: up to 2.4% difference (75kVp beam 40mm deep) Plastic water: up to 23.2% difference (75kVp beam 2mm deep) Hill et al Phys Med Biol 50 N331-N344 2005 In phantom
Phantom Materials Compared Solid Water (RMI Gammex) and Plastic Water (Computerized Imaging Reference Systems) to H 2 O Percentage Depth Dose Solid water: up to 2.4% difference (75kVp beam 40mm deep) Plastic water: up to 23.2% difference (75kVp beam 2mm deep) Absolute dose measurement: Hill et al Phys Med Biol 50 N331-N344 2005 In phantom
Are In Air and In Water measurements equivalent? 300kV 3.67mm Cu Ma et al Med Phys 25 2376-2384 1998 In air In phantom
Are In Air and In Water measurements equivalent? 1.22 0.964 300kV 3.67mm Cu Ma et al Med Phys 25 2376-2384 1998 In air In phantom
Issues with Backscatter Method Thanks to Rhydian Caines and Richard Clements
Depth Dose measurement Measurement of doses close to the surface is difficult Ma et al recommend using an NACP chamber but found differences of 2.4% With a diode the differences were up to 10% In phantom
New Devices New uses are being proposed Zeiss Intrabeam Ariane Papillon 50
Methodology for New Devices Ebert et al have carried out a thorough evaluation of the dosimetry However, the uncertainties in dosimetry probably place the priority on consistency between centres An intercomparison jig has been proposed by Armoogum and Watson Ebert et al Med Phys 30 2424-2431 2003 Armoogum and Watson Z Med Phys 18 120-127 2008
Uncertainty Estimates
Summary Protocols give reliable results at the point of measurement Accurate measurement of percentage depth doses is problematic A protocol should be chosen which measures the dose close to the prescription point
Kilovoltage Codes of Practice Germany DIN 1988 and 1996 UK IPEMB (IPEM) 1996 and 2005 IAEA TRS 277 2nd edn 1997 Holland NCS 1997 USA AAPM TG61 2001 IAEA TRS 398 2004
TRS 398 Calibrate chamber in absorbed dose to water is a chamber specific factor which corrects for differences between the reference beam and the actual beam In phantom
Is TRS 398 the answer? Only PTB offers an absorbed dose to water calibration Other standards labs may offer a calibration based on one of the protocols Jurado et al have used it with calibrations from PTB IAEA TECDOC 1455 shows agreement within 3% at low energies Jurado et al BJR 78 1-12 2005 In phantom