Disclosures. Laboratory Stakeholders. IVD vs. LDT. FDA Regulation of Laboratory Developed Tests 10/2/2015. FDA Regulation of LDTs

Similar documents
Regulatory Overview of Proposed LDT Framework. FDA Concerns. Background. FDA Proposed Regulatory Approach. By Ben Berg, Meaghan Bailey, RAC

Draft Guidance for Industry, Clinical Laboratories, and FDA Staff. In Vitro Diagnostic Multivariate Index Assays

Oversight of Laboratory Developed Tests

Perspective: Convergence of CLIA and FDA Requirements A Rational Shift in the Regulatory Paradigm

Impact of the Regulatory. Development and Innovation

FDA Regulation of Diagnostic Tests An Evolving World

Role of the WHO IVD Prequalification Programme in Light of National Regulatory Authority Approval

FDA PLANS TO REGULATE TESTS AS DEVICES. August 2010 SPECIAL REPRINT. By Ellen Flannery and Scott Danzis

Medidée Services SA. Nano-Tera.ch. 05 February 2015 part 8. PMA, 510k, IDE. Pierre-Alain Sommer

Regulatory & Reimbursement Policy Trends Affecting Market Access for Diagnostic Devices Emphasis on Validity and Quality

Re: Docket No. 2006D-0347, Federal Register: July 26, 2007 (Volume 72, Pages )

Copyright. Jeremiah J. Kelly (2015). All rights reserved. Further dissemination without express written consent strictly prohibited.

Statements on the Regulation of Laboratory Developed Tests

Molecular Diagnostics: The Shift to Complexity. Molecular Diagnostics Regulation: Where do we go from here? David W Feigal, Jr.

Docket No. FDA-2011-D-0215 Draft Guidance for Industry and FDA Staff on In Vitro Companion Diagnostic Devices

The Theranos 483s. What can we learn? David Amor, MS, CQA Principal, Medgineering

FDA s Proposed Regulatory Changes to Laboratory-Developed Tests (LDTs) By Roz Sweeney, Ph.D., Nerac Analyst

OMBU. Ombu Enterprises, LLC Ombu The Operational Excellence Company Ph: Fax: ENTERPRISES, LLC

Agenda. - Introduction Howard Birndorf. - ASR Issues and the Draft ASR FAQ Guidance Patrick Balthrop

FDA s Final Rule on Medical Device Data Systems Signals Active Regulation of Data Communication Technologies

Current Topics in NYS Clinical Laboratory Oversight

FDA Regulation of Testing

UNCONTROLLED DOCUMENT WHEN PRINTED

MEDICINES CONTROL COUNCIL

January 3, AtriCure Inc. Melissa Smallwood Associate Regulatory Affairs Specialist 7555 Innovation Way Mason, Ohio 45040

Advertising and Marketing Genetic Tests New Pathways or Old Roads?

U.S. Food and Drug Administration (FDA) Regulatory Pathways for Medical Devices

FDA's Initiative to Regulate Lab- Developed Tests (LDT) will Harm Patients and Academic Pathology

Introduction to the FDA. Historical Context. Changes over Time FDA - BE-200 9/23/2004. Copyright C. S. Tritt, Ph.D. 1

Paths to Market & FDA Product Lifecycle Regulation. Patricia Kaeding Design of Medical Devices University of Minnesota April 14, 2005

FDA Regulation of In Vitro Diagnostics: Current Perspectives and Initiatives

The In Vitro Diagnostic CRO

Developing a Companion Diagnostics in parallel to a medicinal product Legal and Regulatory requirements in Europe and in the US

DRAFT GUIDANCE. This guidance document is being distributed for comment purposes only. Document issued on: December 27, 2011

MCW Office of Research Standard Operating Procedure

AtriCure Inc. Jonathan McElwee Regulatory Affairs Manger 7555 Innovation Way Mason, Ohio 45040

So You Want to Develop a Test? Navigating the Regulatory Waters of Laboratory Developed Testing

CBER Regulation of Devices for Cell Therapy

Inspections, Compliance, Enforcement, and Criminal Investigations

DEPARTMENT OF HEALTH & HUMAN SERVICES

FDA Regulatory Hurdles: What is the Status Quo?

July 3, WARNING LETTER CHI-3-08

An Introduction to the Worldwide Regulatory Framework for Medical Devices. Elizabeth Malo M.S., Director, Regulatory Affairs

FDA s Evolving Policy on Personalized Medicine Tests

U.S. FDA CENTER FOR DEVICES AND RADIOLOGICAL HEALTH UPDATE. Jeff Shuren Director Center for Devices and Radiological Health

February 22, Dear Dr. Cox:

Overview of the FDA Approval Process for TB Diagnostics

Medical Device Postmarket Compliance in the United States

DRAFT GUIDANCE. This guidance document is being distributed for comment purposes only. Document issued on: July 14, 2011

Perspective: New European IVD Regulations New Concepts for Market Authorizations and Product Launch Schedules

A New Era of Clinical Diagnostics: How the Business Model is Changing.

Drug Safety and FDA Revitalization Legislation

U.S. FDA CENTER FOR DEVICES AND RADIOLOGICAL HEALTH UPDATE

HUMANITARIAN USE DEVICES 3/8/2016

Reducing Unnecessary Medical Imaging Exposure

RE: Draft Local Coverage Determination Molecular Diagnostic Tests (MDT) (DL33599)

STATEMENT OF STEVEN GUTMAN, M.D., DIRECTOR

Impact of the IVD Regulations. Barbara Fallowfield Managing Director

Structure and Mandate of FDA

Office for Human Subject Protection. University of Rochester

fix Menu FDA Pharm 411 Medical Devices Overview of Device Regulations

Re: Comments on FDA s technical assessment of the Diagnostic Accuracy and Innovation Act

11.0 FDA-Regulated Research Research Involving Investigational Drugs and Biologics

FDA Update on Compounding

This letter corrects our substantially equivalent letter of November 25, 2014.

Distinguishing Medical Device Recalls from Product Enhancements and Associated Reporting Requirements

Customed, Inc 12/9/14

Guidance for Industry, Review Staff, and the Clinical Community

WARNING LETTER AUG 3, 2016

Compliance Program Guidance Manual Chapter 42 Blood and Blood Products

Topics. Safety First Initiative. Drug Safety Communications DRUG SAFETY INITIATIVES AT THE FOOD AND DRUG ADMINISTRATION

Confronting MedTech Start-Ups Three Biggest Challenges: Initiating Regulatory & Quality

11.0 FDA-Regulated Research Research Involving Investigational Drugs and Biologics

De Novo Classification Process (Evaluation of Automatic Class III Designation) Guidance for Industry and Food and Drug Administration Staff

A Risk-based Approach for In Vitro Companion Diagnostics Device FDA Approval Process Associated with Therapies that have Breakthrough Designation

COVERAGE ELIGIBILITY OF SERVICES ASSOCIATED WITH A CLINICAL TRIAL

Incorporating FDA Regulatory Thinking into Plans for Medical Devices

DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration New England District WARNING LETTER NWE-11-08W

ATELIER RÈGLEMENTAIRE DM US/CANADA

American Society of Cytopathology Core Curriculum in Molecular Biology

Clinical Trials Management for Molecular Diagnostics. April 2016

External Defibrillator Improvement Initiative

Health Industry Alert

The FDA/CDRH Perspective in the Regulation of Surgical Mesh

DEC DEPARTMENT OF HEALTH & HUMAN SERVICES

Inspections, Compliance, Enforcement, and Criminal Investigations

UT SOUTHWESTERN MEDICAL CENTER AT DALLAS INSTITUTIONAL REVIEW BOARD. Emergency Use of an Investigational Drug, Biologic or Device

May 22, Ranfac Corporation Mr. Christopher Whelan Senior Vice President 30 Doherty Avenue, P.O. Box 635 Avon, Massachusetts 02322

Medical Device Classification

The Institutional Review Board (IRB) Manual

AAMI Quality Systems White Paper: Comparison of 21 CFR Part 820 to ISO 13485:2016 1

November 9, AtriCure, Inc. Jonathan McElwee Regulatory Affairs Manager 7555 Innovation Way Mason, Ohio 45040

The Intersection of Genomics Research and the IDE Regulation

Topics. HCCA Research Compliance Confereence. May 31-June 3, Why are changes in Personalized Medicine Important to Me?

Guidance for Industry and FDA Staff Procedures for Handling Post-Approval Studies Imposed by PMA Order

GUIDANCE FOR SPONSORS, INDUSTRY, RESEARCHERS, INVESTIGATORS, AND FOOD AND DRUG ADMINISTRATION STAFF

Due diligence in the European medical devices industry

1 Purpose. 2 Procedure. Title: FDA-Regulated Research. SOP Number: 1301 Effective Date: June 2, Previous Version Dates:

Enhancing Laboratory Data Infrastructure to Access Real-World Evidence (RWE) for in vitro Diagnostics (IVDs): Three Models for RWE Use

Impact of the Cures Act on the Medical Device Industry

Transcription:

Disclosures FDA Regulation of Laboratory Developed Tests Beaumont Health System, 24 th Annual Symposium on Molecular Pathology September 16, 2015 Roger D. Klein, MD JD Director, Molecular Pathology Clinical Laboratory Improvement Advisory Committee Association for Molecular Pathology Chair, Professional Relations Committee Board of Directors Economic Affairs Committee Strategic Opportunities Committee College of American Pathologists Economic Affairs Committee Checklist Committee Pathology Coding Caucus American Medical Association Molecular Pathology Advisory Group CodeBridge Mapping Panel Laboratory Stakeholders Hospital and health system laboratories hospital outreach programs Small independent laboratories Large independent laboratories Advanced Diagnostic Laboratory Test providers In vitro diagnostic kit manufacturers IVD vs. LDT Laboratory Developed Test (LDT): A clinical laboratory procedure that is developed, validated, performed, continually monitored, and improved upon by and within a single laboratory for use with its own patients. Services regulated by Centers for Medicare and Medicaid Services (CMS) and States. In Vitro Diagnostic Test (IVD): A diagnostic test kit that is designed, developed, manufactured, packaged, sold and distributed in interstate commerce to many labs throughout the United States and the world. Products regulated by Food and Drug Administration. FDA Regulation of LDTs FDA has historically regulated medical device manufacturers On July 31, 2014, FDA notified Senate Committee on Health, Education, Labor and Pensions and House Energy and Commerce Committee of intent to issue draft framework for regulation of LDTs 2 draft guidance documents published on October 3, 2014 Public workshop on January 8 9, 2015 Public workshop on regulation of NGS tests on February 20, 2015 1

FDA proposing to regulate clinical laboratories as medical device manufacturers Clinical Laboratory Improvement Amendments (CLIA) FDA Regulation of LDTs Current medical device regulatory framework established by 1976 amendments to Food, Drug, and Cosmetic Act several subsequent modifications FDA claims authority to regulate LDTs Jurisdiction is disputed FDA claims to have exercised policy of enforcement discretion uncertainty regarding FDA s right to regulate LDTs statutory constraints against interfering with the practice of medicine resource limitations disagreement about the need for FDA intervention potential adverse impacts on patient access FDA Risk Based Device Classification Class I (low risk) subject to general controls applicable to all devices include adulteration and misbranding provisions, registration and listing, records and reports, and good manufacturing practices Class II (intermediate risk) subject to premarket notification under 510(k) FDCA FDA clears after demonstration of substantial equivalence to class I or class II predicate devices based on intended use may require special controls such as performance standards, postmarket surveillance, patient registries, guidelines, and recommendations. Class III (high risk) subject to premarket approval requires demonstration of safety and effectiveness typically requires data obtained from clinical trial New Device Introduction By default placed into Class III manufacturers can request immediate placement into Class I or Class II through de novo 510(k) process Relatively few cleared or approved molecular pathology tests dynamic nature of technologies low reimbursement for diagnostics tests effectiveness and ease of implementation of LDTs limited potential volumes for many assays costs and restrictions posed by FDA submission multiple possible specimens or indications for many tests, which if not mentioned in the label subjects laboratories LDT validation requirements FDA Requirements 2

LDTs for Unmet Needs Device meets definition of LDT No FDA cleared or approved IVD for specific intended use Device manufactured and used by healthcare facility for patient of facility or health system Premarket Review Premarket approval (PMA) Clearance through 510(k) process Premarket Approval (PMA) 510(k) Premarket Notification Process Class III devices Devices not substantially equivalent to predicate devices Must be supported by valid scientific evidence demonstrating safety and effectiveness of the device for its intended uses Typically include results of extensive clinical trials, bench tests, laboratory studies, animal studies, and references to any standards relevant to a device s safety or effectiveness Complete description of device and components Detailed description of the methods, facilities, and controls used to manufacture the device Proposed labeling and advertising literature Training materials Published and unpublished literature concerning prior uses Bibliography of published reports known not submitted regarding safety and effectiveness Involves comparison of new device to one or more legally marketed devices ( predicate device(s) ) Required to file 510(k) notification prior to: initial marketing of device making change or modification to cleared device that could affect safety or effectiveness making change or modification to intended use of previously cleared device Manufacturers submit at least 90 days prior to intended introduction into the market marketing requires FDA order stating that device is substantially equivalent to a legally marketed predicate device Substantial Equivalence Contents of 510(k) Same intended use as predicate; and either Same technological characteristics as predicate device; or Different technological characteristics, but information submitted must not raise new questions of safety and effectiveness and must demonstrate substantial equivalence Determination based on device s proposed labeling Device Name Identification Registration Number Classification Description Substantial Equivalence Comparison Software Standards Clinical Data Performance Biocompatibility Sterility Labeling Class III Certification and Summary 510(k) Summary Statement Truthful and Accuracy Statement 3

Medical Device Reporting (MDR) FDA currently enforces MDR provisions applicable to device user facilities against clinical laboratories must report deaths to FDA and manufacturer must report serious injuries to manufacturer FDA intends to enforced medical device reporting requirements on LDPs (21 C.F.R. 803 Part E) requires manufacturer to submit reports to FDA when become aware of information that reasonably suggests device may have caused or contributed to a death or serious injury or has malfunctioned and malfunction would be likely to cause or contribute to a reportable death or serious injury should it recur Serious Injury Injury or illness that: is life threatening; results in permanent impairment of a body function or permanent damage to body structure; or necessitates medical or surgical intervention to preclude permanent impairment of a body function or permanent damage to a body structure (21 C.F.R. 803.3) Caused or Contributed to a death or serious injury was or may have been attributed to a medical device, or that a medical device was or may have been a factor in a death or serious injury, including events occurring as a result of failure, malfunction, improper or inadequate design, manufacture, labeling, or user error (21 C.F.R. 803.3) Malfunction The failure of a device to meet its performance specifications or otherwise perform as intended (21 C.F.R. 803.3) Performance specifications include all claims made in the labeling for the device The intended performance of a device refers to the intended use for which the device is labeled or marketed MDR Procedures Clinical laboratories must develop, maintain, and implement written MDR procedures (21 C.F.R. 803.17) Include internal systems that provide instructions for: timely and effective identification communication evaluation of events that may be subject to MDR requirements standardized review process or procedure that assists in determining when an event meets MDR reporting criteria timely transmission of complete MDR reports to FDA Include instructions for addressing documentation and record keeping requirements for: information evaluated to determine if an event was reportable all medical device reports and information submitted to FDA information evaluated for the purpose of preparing submission of annual reports systems ensuring access to information that facilitates timely follow up and inspection Laboratories must establish and maintain MDR event files required by 21 C.F.R. 803.17 Quality System Regulations (QSR) Medical device amendments require promulgation of regulations requiring that the methods used in, and the facilities and controls used for, the manufacture, preproduction design validation (including a process to assess the performance of a device but not including the evaluation of the safety or effectiveness of a device), packing storage, and installation of a device conform to current good manufacturing practice, as prescribed in such regulations, to assure that the device will be safe and effective and otherwise in compliance with this Act (21 U.S.C. 360j(f)(1)(a)) Regulations require quality system in place for design, manufacture, packaging, labeling, storage, installation, and servicing of finished medical devices 4

GMP Required Procedures Timeline for implementation Management responsibility Quality audits Personnel (and training) Design Controls Document Controls Purchasing Controls Device Identification and Traceability Production and Process Controls Inspection, Measuring and Test Equipment Process Validation Receiving, In Process, and Finished Device Acceptance Nonconforming Product Corrective and Preventive Action (CAPA) Device Labeling and Packaging Device Handling, Storage, Distribution and Installation Device Recordkeeping Complaint Handling Servicing Statistical Techniques Comply with regulations applicable to operations Component manufacturers and third party servicers not covered Several guidance documents interpret Finalization of Guidance All new highest risk LDTs must have premarket review prior to clinical use 6 months Registration and listing / notification Begin adverse event reporting Implementation timeline of the guidance document 1 year 18 months 2 years 3 years 4 years 5 years Begin FDA to FDA to FDA to FDA to FDA premarket publish draft announce begin announce Completes review of guidance on priority list enforcing priority list premarket high risk risk for high risk premarket for review of LDTs on classification (Class III) review of moderate high risk market tests high risk risk (Class LDTs and LDTs II) tests begins review of moderate risk LDTs 9 years to fully implement the framework Next Generation Sequencing Effects of FDA Guidance The number of existing LDTs and lack of apparent exemptions suggest implementation would result significant financial consequences for large reference laboratories as written could cause discontinuance of many LDTs Hospital and health system laboratories could potentially avoid some regulation by using FDA kits when available and restricting testing to institution s patients Serious concerns about use of NGS for molecular oncology testing Advanced diagnostic laboratory tests (ADLT) undergo significant cost burdens potential impact on claims IVD manufacturers possibly stand to gain as strong incentives to use kit, but modifications of FDA cleared/approved kits? preservation of local testing? Legal Challenges? 5

Conclusions Potentially enormous change in regulatory oversight of clinical laboratories could eliminate key testing in laboratories across the country creating serious concerns about patient access major potential threat to continued advancement and innovation in molecular pathology serious concerns about use of NGS in oncology Stakeholders jockeying for position in order to gain competitive advantages poses additional threats Preservation of CLIA centric oversight framework likely continues to be best approach to ensure persistent patient benefits, innovation, and high quality laboratory testing 6