QIAGEN Validation Report

Similar documents
QIAGEN Validation Report

QIAGEN Validation Report

Developmental validation of the Investigator Argus X-12 QS Kit

Direct amplification of FTA and swab samples using the Investigator Argus X-12 QS Kit

PowerPlex. Y System Validation

QIAGEN Supplementary Protocol

QIAGEN Supplementary Protocol

Rotor-Gene Q quick-start guide

Investigator Quantiplex HYres Kit

Validation of a Multiplexed System for Quantification of Human DNA and Human Male DNA and Detection of PCR Inhibitors in Biological Samples

Comparison of the Promega Power Y and AB AmpFlSTR YFiler kits and the Subsequent Validation of YFiler.

QIAGEN Supplementary Protocol

Microbial DNA qpcr Array Urinary Tract Infections

QIAGEN Supplementary Protocol

The STAR Q Swab AS Instrument for highthroughput PCR setup of QIAGEN s Investigator STR GO! Kits

EpiTect PCR Control DNA Handbook

GlobalFiler TM Extra Cycle Evaluation Report

AmpF STR NGM PCR Amplification Kit - Overview

FastLane Kits from Sample Direct to Result

Product Catalog # Description List Price (JPY) Primer Assays (desalted)

PowerPlex ESI 17 Pro System

PowerPlex ESX 17 Fast System for Use on the Applied Biosystems Genetic Analyzers

TruSight Tumor 15 Protocol Guide

Introducing: The AmpF STR NGM SElect Express Kit From Sample to Result in 2 Hours*

PowerPlex ESI 16 Fast System for Use on the Applied Biosystems Genetic Analyzers

Applied Biosystems models 7500 (Fast block), 7900HT (Fast block), StepOnePlus, ViiA 7 (Fast block)

Dedicated to Molecular Diagnostics

QIAsymphony SP Protocol Sheet

QIAsymphony RGQ Application Sheet

Quantifiler Human DNA Quantification Kit Quantifiler Y Human Male DNA Quantification Kit

mirna QPCR Master Mix

Mentype Nonaplex I PCR Amplification Kit

User Bulletin. Veriti 96-Well Thermal Cycler AmpFlSTR Kit Validation. Overview

RT 2 Easy First Strand Handbook

PowerPlex Y23 System for Use on the Applied Biosystems Genetic Analyzers

Guidelines and Considerations for Evaluating or Validating Extraction Chemistry

mirna QPCR Master Mix

PowerPlex ESX & ESI Fast Systems

Examination of Proposed Manufacturing Standards and Implementation of ISO 18385

PowerPlex Y23 Developmental Validation

RT 2 HT First Strand Handbook

Mentype Nonaplex I PCR Amplification Kit

QIAsymphony SP Protocol Sheet

PrimePCR Pricing and Bulk Discounts

Internal Validation of the Promega PowerPlex Fusion System with the Applied Biosystems 3130xl Genetic Analyzer

PrimePCR Pricing and Bulk Discounts

PrimePCR Pricing and Bulk Discounts

Upgrading the Workflow: Enhancing Information Recovery from Forensic Evidence. The world leader in serving science

Automated purification of genomic DNA from large-volume samples

Investigator HDplex Handbook

Forensic News. FAS Corner April The Maximizing Data Quality Series Part 2: DNA Quantitation

Technical Focus. Direct PCR Amplification of Buccal Samples on the NUCLEIC-CARD COLOR With the AmpFlSTR Identifiler Direct PCR Amplification Kit

STR amplification of DNA from buccal and blood samples stored long-term on Whatman FTA cards

The Evolution of Direct Amplification: From Sample to Result in 2 Hours*

Data Quality Worth Sharing

STR amplification of DNA from buccal and blood samples stored long-term on Whatman FTA cards

TECHNICAL MANUAL. PowerPlex 16 System. Instructions for Use of Products DC6530 and DC6531. Revised 5/16 TMD012

QIAcube HT Your Purification Expert

PowerPlex 5C Matrix Standards, 310

Microbiology Enlightened

Mentype Pentaplex ESS PCR Amplification Kit

PowerPlex ESX 16 System

PowerPlex ESX 17 System

THUNDERBIRD SYBR qpcr Mix

Thermo Scientific Canine Genotypes Panel 2.1

QIAsymphony SP Protocol Sheet

Mentype Chimera Manual

Cat. # RR430S RR430A. For Research Use. SYBR Fast qpcr Mix. Product Manual. v201610da

Use of the GenePrint 10 System with the Applied Biosystems 3500 and 3500xL Genetic Analyzers and POP-7 Polymer

Mentype DIPscreen Manual

PowerPlex 5-Dye Matrix Standards, 310

Latest Tools for Human Identification

PowerPlex Fusion 6C System

3color RT HS-PCR Mix SYBR Ready-to-use mix for real-time Hot Start PCR with SYBR Green. Dedicated for white reaction tubes and plates.

Direct PCR Amplification of Blood Samples on Copan NUCLEIC-CARD TM Systems Using the AmpFlSTR Identifiler Direct PCR Amplification Kit

Fully Automated Library Quantification for Illumina Sequencing on the NGS STAR

Thermal cycler amplification robustness: a comparison of several models

TECHNICAL MANUAL. SwabSolution Kit. Instructions for Use of Product DC8271. Revised 9/16 TMD037

Evaluation of the PowerQuant System on the QuantStudio 5 Real-Time PCR System

Cat. # RR391A. For Research Use. Probe qpcr Mix. Product Manual. v201610da

AmpFlSTR COfiler. PCR Amplification Kit. User Bulletin

Amplify Your Expectations With Bio-Rad s Full Range of PCR Solutions. Amplification: Conventional PCR

PowerPlex CS7 System

PowerPlex ESI 16 System

Development and Validation of the 3500 Series Genetic Analyzer for Human Identification

AutoLys workstation: Validation of a fully automated forensic sample lysis solution

NRAS Codon 61 Mutation Analysis Reagents

The injection of the 28 and 29 PCR cycle data was performed at 3kV for 10 and 5 seconds.

PowerPlex ESI 16 System

PowerPlex ESX 16 System

Validation of a Multiplexed System for Quantification of Human DNA and Human Male DNA and Detection of PCR Inhibitors in Biological Samples

!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"

A comparative performance evaluation of illustra Ready-To-Go GenomiPhi V3 and illustra GenomiPhi V2 DNA amplification kits

For in vitro Veterinary Diagnostics only. Real-Time RT-PCR Detection Kit for Avian Nephritis Virus

Brilliant III Ultra-Fast SYBR Green QPCR Master Mix

QIAcube Pure Efficiency

High-Specificity mirna QPCR Core Reagent Kit

Absolute Human Telomere Length Quantification qpcr Assay Kit (AHTLQ) Catalog # reactions

Transcription:

QIAGEN Validation Report Developmental validation of the Investigator HDplex Kit The QIAGEN Investigator HDplex Kit is used for multiplex PCR in human identification and forensic casework. It was developed for the reliable generation of DNA profiles from blood and buccal swabs. The Investigator HDplex Kit was validated for various reaction conditions and situations that can arise during its storage and transport, as well as during the processing of forensic samples. Optimal reaction conditions were established, and the effects of variations in those conditions were assessed. The kit was tested in house and in independent external forensic laboratories. A range of thermal cyclers and genetic analyzers were used to demonstrate the robustness of the assay (page 2 and page 3). Its sensitivity to different cycle numbers (page 4), and to variations in PCR annealing temperature (page 5) were also assessed, as was its sensitivity to serial dilutions (page 6). The stability of the components was validated with regard to repeated freezing and thawing (page 8), long-term storage (page 9), and transport (page 10). Specific issues that can arise during forensic casework were investigated, including cross-reactivity with nonhuman DNA (page 11) and inhibition due to contaminants introduced during extraction (page 12). The reproducibility of the results was also verified (page 14). The validation of the Investigator HDplex Kit showed that it yields robust and reproducible results within the normal range of conditions expected in forensic casework, as well as in long-distance transport and long-term storage. Note: All of the electropherograms shown were generated on an Applied Biosystems 3130 Genetic Analyzer. The standard conditions specified in the Investigator HDplex Handbook were used for the electrophoresis. Sample & Assay Technologies

Results of developmental validation Effect of different cycler types Multiple PCR thermal cyclers were tested with the Investigator HDplex Kit to demonstrate its robustness. 500 pg Control DNA XY5 was used as a PCR template. The reaction took place under standard conditions and was performed with the following thermal cyclers. GeneAmp PCR System 9700 with Aluminum 96-Well Block (Applied Biosystems Inc., Foster City, CA, USA) GeneAmp PCR System 9700 with Silver or Gold-plated Silver 96-Well Block (Applied Biosystems Inc., Foster City, CA, USA) DNA Engine PTC-200 Peltier Thermal Cycler (Bio-Rad Laboratories GmbH, Munich, Germany) Techne TC-512 Thermal Cycler (biostep GmbH, Jahnsdorf, Germany) T1 Thermal cycler (Biometra biomedizinische Analytik GmbH, Göttingen, Germany) Eppendorf Mastercycler ep (Eppendorf AG, Hamburg, Germany) The electropherograms in Figure 1 show comparable mean peak heights for all of the tested PCR cyclers. No imbalance, dropouts, or preferential amplification for the STR systems was observed on any of the thermal cyclers. If using the GeneAmp PCR System 9700 with an Aluminum Block use Std Mode. If using a Silver 96-Well Block or Gold-plated Silver 96-Well Block, use 9600 Emulation Mode. Do not use Max Mode. GeneAmp 9700 Alu GeneAmp 9700 Silver Bio-Rad PTC-200 Techne TC-512 Biometra T1 Eppendorf Mastercycler ep Figure 1. Effect of different cycler types on mean peak height. Investigator HDplex Kit Validation Report 11/2011 2

Effect of different genetic analyzers The Applied Biosystems 3130 Genetic Analyzer and ABI PRISM 310 Genetic Analyzer were tested with the Investigator HDplex Kit in order to demonstrate its robustness. 500 pg Control DNA XY5 was used as a PCR template. The reaction took place under standard conditions and the PCR products were tested on both genetic analyzers in parallel. The electropherograms in Figure 2 show comparable mean peak heights for both of the tested genetic analyzers. No imbalance, dropouts, or preferential amplification for the STR systems was observed on either genetic analyzer. Applied Biostytems 3130 ABI PRISM 310 Figure 2. Effect of different genetic analyzers on mean peak height. Investigator HDplex Kit Validation Report 11/2011 3

Effect of different cycle numbers Altering the number of cycles can improve amplification when working with lowcopy-number DNA or a large amount of template DNA (e.g., >1 ng). To test the effect of changing the number of cycles, 1 ng Control DNA XY5 was used for 28 cycles, while 500 pg was used for 30 cycles, 100 pg was used for 32 cycles, and 50 pg was used for 34 cycles, respectively (Figure 3). Complete profiles were generated under all of the tested conditions. It should be noted that an increase in cycle number to more than 30 may also increase the chance of unspecific products. Furthermore, because of stochastic effects, peak imbalance or dropouts may be observed for low-copy-number samples (with 100 pg or less of template DNA), regardless of any increase in cycle numbers. 28 cycles 1 ng 30 cycles 500 pg 32 cycles 100 pg 34 cycles 50 pg Figure 3. Effect of different cycle numbers on mean peak height. To better visualize smaller peak heights, the y-axis scales were adjusted for best fit. Investigator HDplex Kit Validation Report 11/2011 4

Effect of PCR annealing temperature variations Specificity and sensitivity are critical in forensic casework, and both are affected by the annealing temperature (T m ). Since the actual T m may vary depending on cycler conditions, the assay was validated in a range surrounding the optimal T m for the Investigator HDplex reaction (60 C). Annealing temperatures of 58 C, 59 C, 60 C, 61 C, and 62 C were applied to the amplification of 500 pg Control DNA XY5. The PCR was performed on an Eppendorf Mastercycler ep and analyzed on an Applied Biosystems 3130 Genetic Analyzer. As shown in Figure 4, no allelic dropout was detected within ±2 C of the optimal T m of 60 C. This shows there was no loss in sensitivity under these conditions. Furthermore, no unspecific peaks occurred, indicating good specificity in this T m range. 58 C 59 C 60 C 61 C 62 C Figure 4. Effect of variations in the PCR annealing temperature. Investigator HDplex Kit Validation Report 11/2011 5

Sensitivity to serial dilutions The Investigator HDplex Kit is designed to detect a range of DNA quantities. The optimal amount of input DNA to yield good quality STR profiles is 500 pg. Figure 5 (page 7) shows a serial dilution of Control DNA XY5 from 500 pg to 32 pg. The optimal linear dynamic range of the assay is in the range of 500 to 125 pg. Full profiles (26 alleles) were consistently obtained at 125 pg using the standard conditions specified in the Investigator HDplex Handbook. Issues with very high amounts of DNA An amount of DNA above the optimal range (>1 ng) may give inaccurate or unusable data. The fluorescence intensity may cause peaks to go off the scale. Such data cannot be quantified accurately based on peak height and area. As a secondary effect, irregular stutter-peak heights may occur. Off scale peaks are often accompanied by pull up peaks. This effect hinders the multi-component analysis, and is caused by an inaccurate spectral separation. Finally, split peaks may occur as a result of incomplete +A nucleotide addition. Poor STR profiles resulting from high DNA concentration can be improved by reamplifying a sample using less DNA. Issues with very low amounts of DNA An amount of DNA in the PCR below the optimal range (<125 pg) may lead to incomplete profiles, where partial profiles lacking alleles occasionally occur. Furthermore, low allele copy numbers in the PCR can result in an unbalanced amplification of the alleles due to stochastic fluctuation. These effects can be handled by either performing multiple analyzer runs of the same sample to complete the partial profiles, or by adding the maximum volume of the DNA template to the PCR. The Investigator HDplex Kit was developed for use with reference samples, especially for applications in paternity testing, rather than casework samples that may involve low amounts of DNA. Investigator HDplex Kit Validation Report 11/2011 6

500 pg 250 pg 125 pg 63 pg 32 pg Figure 5. Effect of serial dilution of control DNA XY5 during amplification. The y-axis scales were adjusted for best fit. Investigator HDplex Kit Validation Report 11/2011 7

Stability 20 freeze/thaw cycles In a forensic lab, it is possible that the maximum number of reactions of a kit will not be performed in a single day. Therefore, the Investigator HDplex components were tested to prove they would yield stable results after multiple rounds of freezing and thawing. Regardless of these results, we do not recommend repeated freezing and thawing of the kit contents. Figure 6 shows the electropherograms obtained by amplifying 500 pg Control DNA XY5 with fresh kit components (No Freeze/Thaw) and with kit components stressed by 20 rounds of freezing and thawing (20x Freeze/Thaw). The overall kit performance was not compromised under the chosen conditions: comparable peak heights were obtained before and after 20 rounds of freezing and thawing. No Freeze/Thaw 20x Freeze/Thaw Figure 6. Results of a simulated freeze/thaw stability test of kit components. Investigator HDplex Kit Validation Report 11/2011 8

Long-term stability Forensic kits must be viable after long-term storage. Investigator HDplex Kit components were stored for 1 year at 20 C. After 1 month, 2 months, 3 months, 6 months, and 12 months, the kit performance was tested by amplifying 500 pg control DNA XY5 under standard PCR conditions. The electrophoresis was performed on an Applied Biosystems 3130 Genetic Analyzer. Over the course of the experiment, the overall kit performance was found to be stable (Figure 7). 1 month 2 months 3 months 6 months 12 months Figure 7. Effects of long-term storage. The kit components were tested at the indicated times. Investigator HDplex Kit Validation Report 11/2011 9

Simulated shipment condition on dry ice Investigator HDplex Kits are shipped on dry ice. To assess the performance of the kit after such transportation, a 5-day storage on dry ice and at 20 C was simulated. Kits were stored for 16 h on dry ice, then transferred to 20 C for 8 h. This cycle was repeated for 5 days. Each day, components from these kits were used to amplify 500 pg Control DNA XY5. The electrophoresis was performed on an Applied Biosystems 3130 Genetic Analyzer. The results indicate that the performance before and after storage on dry ice was comparable (Figure 8). Day 0 Day 1 Day 2 Day 3 Day 4 Figure 8. Prolonged storage of Investigator HDplex components on dry ice. Investigator HDplex Kit Validation Report 11/2011 10

Species specificity Non-human DNA can be present in forensic casework samples. It is critical that assays show no cross reactivity between species. To verify Investigator HDplex species specificity, 2.5 ng of DNA from dog, cat, rabbit, cow, and pig were each tested following the standard assay protocol, with 500 pg Control DNA XY5 as a positive control. The presence of any amplified peaks in the electropherograms would have suggested cross reactivity with DNA from the non-human species. None of the tested DNA yielded any detectable product peaks under the chosen conditions, as shown in Figure 9. Human Dog Cat Rabbit Cow Pig Figure 9. Results of the species specificity assessment. Investigator HDplex Kit Validation Report 11/2011 11

Performance with simulated inhibition While the Investigator HDplex has been optimized for use with reference samples in applications such as paternity testing, the kit has been tested with various concentrations of inhibitors as sometimes seen in difficult samples. If the DNA extraction from forensic casework samples is done using inappropriate methods, Investigator HDplex assay performance may be compromised. QIAGEN sample preparation technology is recommended for extraction, as it yields pure DNA free of inhibitors. Humic acid, a principle component of humic substances, has an inhibitory effect on PCR. It is often co-purified and co-extracted from forensic samples collected from soil. To test the robustness of the kit in the presence of typical inhibitors, the assay was run in the presence of 0, 20, and 30 ng/µl humic acid under standard conditions (500 pg Control DNA XY5, 30 cycles). Hematin is formed by the oxidation of heme, the main component of blood. It has been identified as a PCR inhibitor in DNA samples extracted from bloodstains. Its interfering effect is related to the inhibition of polymerase activity. Investigator HDplex Kit performance was assessed in the presence of increasing concentrations of hematin: 0, 100, and 150 µm under standard conditions (500 pg Control DNA XY5, 30 cycles). Full profiles (>100 RFU) were obtained in the presence of 30 ng/µl humic acid and 100 µm hematin. The results for these maximum inhibitor concentrations from both studies are shown in Figure 10 (page 13). Investigator HDplex Kit Validation Report 11/2011 12

No inhibitor 30 ng/µl Humic acid 100 µm Hematin Figure 10. Impact of hematin and humic acid on performance. The y-axis scales were adjusted for best fit. Investigator HDplex Kit Validation Report 11/2011 13

Reproducibility Concordance test with an internal DNA pool The reproducibility of results generated by the Investigator HDplex was validated under standard conditions. This included a test of automatic allele calling based on the allelic ladder, and a concordance analysis of the allele correlation compared to control DNA that had been pre-validated in additional forensic assays. These assays covered of the STR systems of interest partially or completely. The following kits were used: Investigator ESSplex SE (QIAGEN GmbH, Hilden, Germany) Investigator Nonaplex ESS (QIAGEN GmbH, Hilden, Germany) AmpFlSTR SEfiler Plus (Applied Biosystems Inc., Foster City, USA) PowerPlex 16 (Promega Corporation, Madison, USA) Mentype Chimera (Biotype GmbH, Dresden, Germany) A total of 80 DNA samples (40 male, 40 female) of different origin (blood, saliva) were quantified, and profiles were generated with each assay in quadruplicate. All of the assays generated full profiles with steady peak heights of >50 rfu. The profiles generated with Investigator HDplex kits concurred with those generated by the listed assays.. Investigator HDplex Kit Validation Report 11/2011 14

Ordering Information Product Contents Cat. no. Investigator HDplex Kit (25) Primer mix, reaction mix, DNA Polymerase, Control DNA, allelic ladder, DNA size standard, and nuclease-free water 381213 Investigator HDplex Kit (100) Primer mix, reaction Mix, DNA Polymerase, Control, allelic ladder, DNA Size Standard, and nuclease-free water 381215 For up-to-date licensing information and product-specific disclaimers, see the respective QIAGEN kit handbook or user manual. QIAGEN kit handbooks and user manuals are available at www.qiagen.com or can be requested from QIAGEN Technical Service or your local distributor. Trademarks: QIAGEN, Investigator (QIAGEN Group); ABI PRISM, AmpFlSTR, Applied Biosystems, GeneAmp, SEfiler Plus (Applied Biosystems Inc.); DNA Engine (Bio-Rad Laboratories GmbH); Techne (biostep GmbH); Chimera, Mentype (Biotype Diagnostics GmbH); Eppendorf, Mastercycler (Eppendorf AG); PowerPlex (Promega Corporation).Registered names, trademarks, etc. used in this document, even when not specifically marked as such, are not to be considered unprotected by law. 2011 QIAGEN, all rights reserved. www.qiagen.com France 01-60-920-930 The Netherlands 0800 0229592 Australia 1-800-243-800 Austria 0800/281010 Belgium 0800-79612 Canada 800-572-9613 China 021-51345678 Denmark 80-885945 Finland 0800-914416 Germany 02103-29-12000 Hong Kong 800 933 965 Ireland 1800 555 049 Italy 800 787980 Japan 03-5547-0811 Korea (South) 1544 7145 Luxembourg 8002 2076 Norway 800-18859 Singapore 65-67775366 Spain 91-630-7050 Sweden 020-790282 Switzerland 055-254-22-11 UK 01293-422-911 USA 800-426-8157 1070374 11/2011 Sample & Assay Technologies