Minimizing Carbon Dioxide Contamination in Anion Reagent-Free Ion Chromatography

Similar documents
Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography

Introduction. Yan Liu, Victor Barreto, Rosanne Slingsby, Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA

Thermo Scientific Dionex ICS-900 Ion Chromatography System

Effect of Temperature on IC Signal Noise

Determination of Manganese in Brine ( 30% NaCl)

Implementation of a Walk-Up High- Pressure Capillary Ion Chromatograph for the Fast Separation of Pharmaceutical Relevant Inorganic Anions and Cations

Analysis of Fracking Flowback Water From the Marcellus Shale Using In-line Conductivity, Automated Dilution, and Ion Chromatography

Options Eluent Regeneration (RFIC-ER ) Column Heater In-Line Vacuum Degas Full Chromeleon Chromatography Data System

Complete Customer Support

Determination of Inorganic Anions in Acid Rain Using a Dedicated High-Pressure Capillary Ion Chromatography System

Practical guidance for switching between anion and cation analysis systems

Determination of Trace Concentrations of Oxyhalides and Bromide in Municipal and Bottled Waters Using a Compact Ion Chromatography System

Analysis of Glyphosate and AMPA in Environmental Water Samples by Ion Chromatography and On-Line Mass Spectrometry with Minimal Sample Preparation

Thermo Scientific Dionex DRS 600 Dynamically Regenerated Suppressor

New Developments in Capillary Ion Chromatography Systems with Electrochemical Detection and Their Applications

Next generation suppressor for ion chromatography. Thermo Scientific Dionex ERS 500 Electrolytically Regenerated Suppressor

columnsi o n P a c C S 1 6 C a t i o n - E x c h a n g e C o l u m n

Fast Ion Determinations and Efficient Separations of Complex Samples Using 4 µm Particle-Size Ion Exchange Columns

Dionex Eluent Generator Cartridges

( ) IONPAC ATC-HC ANION TRAP COLUMN

Configuring the Dionex Modular HPIC system with an inline water purifier and inline calibration for trace anion determinations in ultrapure water

Passion. Power. Productivity. Corporate Headquarters

A Novel ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge-Variant Analysis

In-Cell and External Hydrolysis of Biomass for Carbohydrate Determinations Using Automated Solvent Extraction

Dionex IonPac AS16 Column

Thermo Scientific. Dionex IonPac AS9-HC. Column Product Manual. P/N: May Part of Thermo Fisher Scientific

columns IonPac ICE-AS1 Ion-Exclusion Column Superior Chromatographic Performance Separates weakly ionized acids by differences in pk a

Application Note 168

Determination of Perchlorate in Drinking Water Using a Reagent-Free Ion Chromatography System

Dionex IonPac AS19 Columns

Dionex IonPac AS14 Columns

PRODUCT MANUAL FOR THE. ELECTROLYTIC ph MODIFIER (EPM) EPM, P/N

Configuring the Dionex Integrion HPIC system for trace anion determinations in ultrapure water

Thermo Scientific Dionex ICS-4000 Integrated Capillary HPIC System. Always ready, better results. capillary high-pressure ion chromatography

Improved Determination of Trace Perchlorate in Drinking Water Using 2D-IC

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions

Enjoy Industry-Leading Support

Determination of Total Nitrogen and Phosphorus in Environmental Waters

Highlighted Capillary IC Applications

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

Continuously Regenerated Trap (CR-TC 500, Capillary)

Capillary Ion Chromatography and High-Pressure Ion Chromatography: Applications and Solutions

High-speed, High-resolution Oligonucleotide Separations Using Small Particle Anion-Exchangers

For new orders of the following parts discussed in this manual, please use the updated part numbers listed below.

Thermo Scientific Dionex ICS HPIC Capillary and Analytical System. High resolution, fast analysis high-pressure ion chromatography

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Eluent Generator Cartridges

Product Manual for Dionex IonPac AS24A and AG24A Columns

Determination of Cations in Hydraulic Fracturing Flowback Water from the Marcellus Shale

Reducing Carbonate Interference in Anion Determinations with the Carbonate Removal Device (CRD)

Optimize Your Process Operations by Improving Process Monitoring

Dual EGC A simple solution for complex carbohydrate analysis

Dionex IonPac AS11-HC-4µm

Blank Optimization Using Ultrapure Water Suitable for Trace Ion Analysis.

Determination of Sulfite in Food and Beverages by Ion Exclusion Chromatography with Pulsed Amperometric Detection

for IonPac AG9-HC IonPac AS9-HC

Thermo Scientific. Dionex IonPac CS17. Product Manual. P/N: November Part of Thermo Fisher Scientific

Dionex DRS 600 Suppressor Dionex ERS 500e Suppressor Dionex ERS 500 Carbonate Suppressor

Dionex IonPac AS11-HC-4µm

Determination of Perchlorate in Drinking Water by Ion Chromatography

Analysis of Disinfection Byproducts by Ion Chromatography

A HIGHLY VERSATILE IC SYSTEM CAPABLE OF EVERY CHALLENGE THAT COMES YOUR WAY THERMO SCIENTIFIC DIONEX ICS HPIC SYSTEM

PRODUCT MANUAL. IONPAC AG14A GUARD COLUMN (4 x 50 mm, P/N ) (3 x 30 mm, P/N )

Assay for Citrate and Phosphate in Pharmaceutical Formulations Using a High-Pressure Compact Ion Chromatography System

Fast mass transfer Fast separations High throughput and improved productivity Long column lifetime Outstanding reproducibility Low carryover

Monosaccharide and Sialic Acid Determinations in Biosimilars Using HPAE-PAD

Dionex IonPac CS19 Columns

Application Note 171 INTRODUCTION

Identification of RNA Linkage Isomers by Anion-Exchange Purification with ESI-MS of Automatically-Desalted Phosphodiesterase-II Digests

Determination of Trace Copper, Nickel, and Zinc in Boiling Water Reactors Using Nonsuppressed Conductivity Detection

Ion Chromatography Coupled to MS, a Powerful Approach for Polar Pesticides Determination

Chromatography Column Performance and Data Analysis Success Guide. Hints and Tips for Better Purifications

Errata. Product Manual for Dionex IonPac Cryptand Columns

[ Care and Use Manual ] Waters AccellPlus QMA and CM Bulk Media. I. calculating bulk media needs. II. packed column use. IIi.

electrochemical detection

Nano LC at 20 nl/min Made Easy: A Splitless Pump Combined with Fingertight UHPLC Nano Column to Boost LC-MS Sensitivity in Proteomics

Be a pioneer. Thermo Scientific Dionex ICS-6000 HPIC system the freedom to explore

Advanced Features New Advanced Automation Features Precise Performance Temperature Control Speed Application Flexibility

Determination of halogens in coal using combustion ion chromatography

Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water

Product Manual for Dionex IonPac AS4A-SC and AG4A-SC Columns

Dionex IonPac CS16-4µm and Dionex IonPac CS16-Fast-4µm

Sensitive Determination of Hexavalent. Brian De Borba, Lipika Basumallick, Jeffrey Rohrer

Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water. Pranathi R Perati Thermo Fisher Scientific, Sunnyvale, CA, USA

Product Manual for the Dionex IonPac Ultra Trace Anion Concentrator (UTAC)

Alex J. Zhang, Pranathi Perati, and Monika Verma; Thermo Fisher Scientific, Sunnyvale, CA, USA

HPAE-PAD analysis of galactosyl-oligosaccharidecontaining samples using dual eluent generation cartridges

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

Importance of pure water in modern ion chromatography

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates

User Manual. Dionex Guardcap H Revision 02 March For Research Use Only. Not for use in diagnostic procedures.

Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants

Investigations into Method Improvements for Hexavalent Chromium. Presented by Aaron Rose Thermo Fisher/Dionex Products For 2012 Watercon

Be a pioneer. Thermo Scientific Dionex ICS-6000 HPIC system the freedom to explore

Nuvia HR-S Strong Cation Exchange Media Instruction Manual

High-Resolution Separation of Intact Monoclonal Antibody Isoforms

High-Resolution Charge Variant Analysis for Top-Selling Monoclonal Antibody Therapeutics Using a Linear ph Gradient Separation Platform

High-Resolution Analysis and Purification of Oligonucleotides with the DNAPac PA-100 Column

HIGH QUALITY, RELIABLE ION ANALYSIS SOLUTIONS FOR EVERY IC CHALLENGE THERMO SCIENTIFIC DIONEX PRODUCT PORTFOLIO

Quantification of Drug Metabolites in Early-Stage Drug Discovery Testing

Transcription:

Minimizing Carbon Dioxide Contamination in Anion Reagent-Free Ion Chromatography Archava Siriraks, 1 John M. Riviello, 2 and Christopher A. Pohl 3 1 Trovion Pte Ltd., Singapore; 2 Trovion Co., Campbell, CA, USA; 3 Thermo Fisher Scientific, Sunnyvale, CA, USA

Introduction The ubiquitous presence of carbon dioxide in air (typically 390 ppm and increasing 1 ) results in dissolved carbon dioxide being the most common anionic contaminant in laboratory water. In deionized water, dissolved carbon dioxide exists as nonionic carbonic acid and bicarbonate. Carbonic acid is both a blessing and a curse in suppressed ion chromatography (IC). When it was recognized that solutions of sodium carbonate and sodium bicarbonate were useful eluents and could be suppressed to low conductance carbonic acid, 2 carbonate-based eluents became the mainstay for suppressed anion IC. With the advent of hydroxide-selective, anion-exchange columns and the subsequent development of Reagent-Free IC (RFIC ) systems and electrolytic eluent generation using water as the pumped phase, carbonate became a contaminant, instead of an eluent. Because potassium hydroxide suppresses to water, the presence of carbonate in the eluent water or ambient carbon dioxide absorbed into the basic eluent causes an increase in the suppressed background conductivity. Because most anion RFIC system separations using potassium hydroxide are based on gradient elution, trace carbonate accumulates at lower eluent concentration and elutes at higher eluent concentrations. The elution of the carbonate causes a baseline shift as well as a carbonate peak. In samples, dissolved carbon dioxide may constitute the highest concentration analyte and compromise separation and detection of trace analytes. The carbonate peak, whether from the eluent or the sample, can be quite large relative to analytes and thus can interfere with integration. This work describes processes for minimizing carbonate contamination in gradient anion RFIC systems. Devices for in-line purifi cation of water and eluent are described as well as decarbonation techniques for post-suppressor carbonate removal. Examples demonstrating the benefi ts of reducing carbon dioxide contamination in anion RFIC system separations are shown. Experimental Instrumentation Water Purification Trovion iwp Electrolytic Water Purification system consisting of: Reagent Pump Controller CIRA EP or CIRA 10 Electrolytic Water Purifier Thermo Scientifi c Dionex ICS-2100 RFIC system with: Dionex EGC II KOH Eluent Generator Column (EGC) Dionex Continuously Regenerated Anion Trap Column (CR-ATC) Dionex CRD 300 Post Suppression Carbonate Removal Device (CRD) Thermo Scientifi c Dionex IonPac AEP-ER1 (for water recycle) Chromatography Conditions Columns: Dionex IonPac AS18, AS19, or AS20, as specified Flow Rate and Gradients: as specified Suppressors: Thermo Scientifi c Dionex ASRS 300 Anion Self-Regenerating Suppressor (4 mm) Detection: Conductivity Results and Discussion Anion Analysis of Deionized Water The primary source of carbonate is from the deionized water used for electrolytic eluent generation eluent. A Dionex ICS-3000 system was used to analyze deionzed water from a bench top lab water system. A Dionex IonPac AS18 separator column was used isocratically (25 mm KOH) with a large loop (1.5 ml) injection method. The deionized water was injected as a sample every 30 min and for approximately 21 h (42 runs). Next the deionzed water was passed through the Trovion iwp system prior to analysis (Figure 1). The Trovion iwp system is a point-of-use electrolytic water purifi cation system that produces ionically pure water for direct feed to a Dionex RFIC system. 3 Figure 2 compares the anion chromatogram for the deionzed water sample and the same water passed through the iwp system. Not only was the carbonate reduced 90%, but the iwp system also reduced other sub-ppb anions such as fl uoride acetate, formate, chloride, and sulfate. Figures 3a and 3b show the trend in carbonate over the 21 h. Note for the deionzed water, the carbonate peak continues to rise as carbon dioxide from the air is absorbed into the water. With the iwp system, the carbonate peak is not only an order of magnitude smaller, but continues to decrease as the CIRA EP electrolytic water purifi er becomes cleaner as a result of continuous electrolytic regeneration. 2 Minimizing Carbon Dioxide Contamination in Anion Reagent-Free Ion Chromatography

FIGURE 1. Trovion iwp System with the CIRA EP shown at top. FIGURE 2. Comparison of deionized water and Trovion iwp water. (Conditions: Dionex IonPac AS18 separator, [4 mm] 25 mm KOH at 1.0 ml/min, Dionex ASRS 300 suppressor.) FIGURE 3. A) Measured carbonate (peak height) by anion RFIC system in deionized water (20 h). B) Measured carbonate (peak height) by anion RFIC system in iwp water (20 h). Conditions used as in Figure 2. Thermo Scientific Poster Note LPN2989-01_e 11/11SV 3

While it is important to remove carbonate from the deionzed feed water used in the Dionex RFIC system for electrolytic eluent generation, there are others sources of carbonate in the system. Once the potassium hydroxide is formed in the eluent generator, the basic eluent is even more effective at absorbing carbonate than deionzed water. The Dionex CR-ATC column (Figure 4) is designed to remove anionic contamination from the potassium hydroxide eluent. 4 The task of removing ionic contamination such as carbonate from the eluent is more difficult than removing the carbonate or other anionic contamination from the feed water. Thus, by using electrolytic water purifi cation for the feed water, the carbonate load in the eluent is reduced which improves the performance of the Dionex CR-ATC. Further reduction in carbonate that originates after electrolytic eluent generation (including sample injection) can be accomplished with the Dionex CRD 200 or 300 device (Figure 5). The Dionex CRD is a postsuppressor device, 5 which removes carbonic acid (as carbon dioxide) from the suppressed eluent. In gradient operation, trace carbonate in the potassium hydroxide concentrates in the early portion of the gradient, and then elutes as a peak in the later portion of the gradient. The large carbonate peak compromises integration of analytes (sulfate and phosphate) in the region of the carbonate and also affects analyte response since the analytes are being measured on a higher carbonate background rather than a water background. FIGURE 4. Dionex CR-ATC schematic. FIGURE 5. Dionex CRD 200 device schematic. Controlling Carbonate in an Anion RFIC System A Dionex ICS-2100 RFIC system was optimized for controlling carbonate contamination. Figure 6 shows the major components of the RFIC system and components used for minimizing carbonate. The Dionex ASRS suppressor was operated in the external water mode at 1.0 ml/min. To minimize ambient contamination of the feed DI water, the 2 L eluent bottle was pressurized with 7 10 psi nitrogen. A Trovion CIRA 10 purifier was placed between the pressurized eluent bottle and the analytical pump. The CIRA 10 is a singlechannel electrolytic water purifi er which consists of an anion and composite ion-exchange layer as shown in Figure 7. In the CIRA 10, the majority of the anionic contamination is retained by the homogenous anion-exchange layer and any residual anionic contamination is removed in the composite layer. From the CIRA 10, the water enters the Dionex ICS-2100 pump degasser where dissolved gasses such as nitrogen and oxygen are removed so that the analytical pump does not lose prime. 4 Minimizing Carbon Dioxide Contamination in Anion Reagent-Free Ion Chromatography

The in-line degasser has a volume 480 µl, so the residence time of the eluent is about 30 seconds at 1 ml/min. This residence time is suffi cient for effective removal of nitrogen and oxygen, but not carbon dioxide. At this point, the feed water is essentially ionically pure and from the pump, enters the Dionex EGC II column where the potassium hydroxide eluent is formed. Next, the potassium hydroxide passes to the Dionex CR-ATC column which acts as a fi nal anionic polisher before the eluent enters the analytical column. Removing the majority of the carbonate from the feed water prior to eluent generation ensures that the Dionex CR-ATC column can remove residual anionic impurities from the eluent. FIGURE 6. Dionex ICS-2100 RFIC system optimized for controlling carbonate contamination. FIGURE 7. Cross-section drawing of the CIRA 10 electrolytic water purifier. From the Dionex CR-ATC column, the eluent passes through the Dionex EGC degasser which removes the electrolytically formed hydrogen from the eluent generation process. The eluent then passes to the injection valve and analytical column. From the Dionex ASRS suppressor, the suppressed eluent (water) passes into the CRD 300 where residual carbon dioxide (carbonic acid) from the eluent and sample is reduced prior to conductivity detection. Using the system described above, a gradient from 5.0 to 50 mm potassium hydroxide was run over 20 minutes at 0.70 ml/min as shown in Figure 8. The baseline shift in this gradient was approximately 160 ns/cm. The carbonate peak is from residual carbonate in the eluent. This experiment demonstrates the diffi culty of controlling carbonate in gradient anion Reagent-Free IC. To further investigate reduction of carbonate, suppressed eluent was recycled back to the feed water container. The Trovion Reagent Pump Controller (RPC) was used to deliver deionized water for the Dionex ASRS regenerant (SRS external water mode). From the conductivity cell, the suppressed eluent (water) was directed back to the nitrogen pressurized eluent bottle. The 2 L eluent bottle was replaced with a 250 ml bottle (filled with 200 ml of DI water) to decrease the recycle time for the eluent (about a fi ve hour recycle time). A Dionex IonPac AEP-ER1 trap column was placed between the pump and the eluent generator to reduce ionic organic contamination. This closed loop system not only has the advantage of recycling the eluent feed water, but multiple passes of the feed water through the CIRA 10 should further reduce carbonate contamination. Sample anions are removed by the CIRA 10 prior to electrolytic eluent generation. Thermo Scientific Poster Note LPN2989-01_e 11/11SV 5

FIGURE 8. Typical anion RFIC system gradient blank obtained when controlling carbonate contamination. A gradient from 5 to 100 mm potassium hydroxide was run over 45 min. After approximately two cycles of the feed water, the gradient blank in Figure 9 was obtained. Not only was the gradient drift reduced (30 ns/cm) as a result of the recycle, but the carbonate blank peak was also greatly reduced. Using the water recycle method, Figure 10 shows a typical run is shown for an 8 to 50 mm gradient. Note the extremely small carbonate peak and the gradient drift is less than 25 ns/cm. Figure 11 shows the analysis of trace bromate in drinking water using a small sample loop (25 µl) with the water recycle system of Figure 6. FIGURE 9. Typical background of gradient elution of 5 100 mm KOH using the water recycle method. FIGURE 10. Optimized anion RFIC system gradient separation. 6 Minimizing Carbon Dioxide Contamination in Anion Reagent-Free Ion Chromatography

FIGURE 11. Determination of bromate (4 ppb) in drinking water using a small loop injection (25 µl) with water recycle. System as shown in Figure 6. Conclusions While carbon dioxide is the most common contaminant in deionized water, processes to minimize the impact of carbonate contamination in gradient anion RFIC system are readily available to the IC practitioner. No single process can remove all carbonate contamination. In anion RFIC systems, controlling carbonate in the feed water is critical (CIRA electrolytic water purifi er). The use of other devices such as the Dionex CR-ATC column to remove carbonate from the eluent and the Dionex CRD device to remove carbonic acid from the suppressed eluent are all important components for reducing carbonate effects. Pressurizing the eluent container with an inert gas (nitrogen) and recycling the eluent water furthers reduces carbonate contamination. Minimizing carbonate contamination results in lower background conductivity, reduced baseline drift, and lower noise. A reduced carbonate peak improves quantitation of analytes which elute in the region of carbonate. References 1. Volker, Tyler. CO 2 Rising: The World s Greatest Environmental Challenge, Chapter 5, ISBN 978-0-0262-22083-5, Massachusetts Institute of Technology, 2008. 2. Small, H.; Stevens, T.S.; Bauman, W.C.; Novel Ion Exchange Chromatographic Method Using Conductometric Detection, Anal. Chem. 47, 1801 1809 (1975). 3. Riviello, J. and Siriraks, A., Electrolytic Water Purification for RFIC, International Ion Chromatography Symposium 2010, Cincinnati, OH, poster number P-105. 4. Riviello, J. and Siriraks, A., Automated In-line Sample Pretreatment Using an Integrated Electrolytic Water Purifier, Pittsburgh Conference 2010, Orlando FL, paper number, 2250 4. 5. Srinivasan, K; Lin, R.; Saini, S.; Pohl, C.; Avdalovic, N.; A New Continuously Regenerated Trap Column for Ion Chromatography, PittCon 2003, Orlando, FL, March 13, 2003, Paper 2020 2. Thermo Scientific Poster Note LPN2989-01_e 11/11SV 7

www.thermoscientific.com/dionex Thermo Scientific Dionex products are designed, developed, and manufactured under an ISO 9001 Quality System. 2011 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. U.S./Canada (847) 295 7500 Brazil (55) 11 3731 5140 Austria (43) 1 616 51 25 Benelux (31) 20 683 9768 (32) 3 353 42 94 Denmark (45) 36 36 90 90 France (33) 1 39 30 01 10 Germany (49) 6126 991 0 Ireland (353) 1 644 0064 Italy (39) 02 51 62 1267 Sweden (46) 8 473 3380 Switzerland (41) 62 205 9966 United Kingdom (44) 1276 691722 Australia (61) 2 9420 5233 China (852) 2428 3282 India (91) 22 2764 2735 Japan (81) 6 6885 1213 Korea (82) 2 2653 2580 Singapore (65) 6289 1190 Taiwan (886) 2 8751 6655