High Throughput Sub-4 Minute Separation of Antibodies using Size Exclusion Chromatography

Similar documents
Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis

Size Exclusion BioHPLC Columns

TSKgel G2000SWXL Columns for the Reproducible Analysis of Bovine Serum Albumin

Separation of Monoclonal Antibodies Using TSKgel HPLC Columns

Column for High Performance,High-Binding Capacity Ion Exchange Chromatography:TSKgel SuperQ-5PW and Its Applications

TSK-GEL BioAssist Series Ion Exchange Columns

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

R R Innovation Way P/N SECKIT-7830 Newark, DE 19711, USA Tel: Fax: Website: Published in November 2013

CHARACTERIZATION & PURIFICATION

Size Exclusion BioHPLC columns Ion Exchange BioHPLC columns

Validation of a Dual Wavelength Size Exclusion HPLC Method with Improved Sensitivity to Detect Aggregates of a Monoclonal Antibody Biotherapeutic

4/4/2013. BioHPLC columns. Paul Dinsmoor Biocolumn Technical Specialist. April 23-25, Size Exclusion BioHPLC Columns

BioHPLC columns. Tim Rice Biocolumn Technical Specialist

TSKgel Size Exclusion Chromatography Columns

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Mobile Phase Optimization in SEC Method Development

PRINCIPLES OF CHROMATOGRAPHY SEC COLUMNS

Application Note. Authors. Abstract. Biopharmaceuticals

Analysis of biomolecules by SEC and Ion-Exchange UPLC

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies

The New Standard in High Resolution Size Exclusion

ION EXCHANGE KIT FOR MAB SEPARATIONS

Superdex 200 Increase columns

Superdex 200 Increase columns

Higher Order mab Aggregate Analysis using New Innovative SEC Technology

TOYOPEARL GigaCap Series

PRINCIPLES OF CHROMATOGRAPHY SEC COLUMNS

ProSEC 300S. Protein Characterization columns

2012 Waters Corporation 1

NISTmAb characterization with a high-performance RP chromatography column

EAG.COM MATERIALS SCIENCES APPLICATION NOTE. By J. Tompkins 1, T. Spurgeon 1, R. Tobias 1, J. Anders 1, E. Butler-Roberts 2, and M.

Lifetime stability of size exclusion chromatography columns for protein aggregate analysis

TSKgel Size Exclusion Chromatography Columns

TSKgel STAT Columns for High Performance Ion Exchange Chromatography

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

Data File. HiLoad Superdex 30 prep grade HiLoad Superdex 75 prep grade HiLoad Superdex 200 prep grade. Pre-packed gel filtration columns

Size Exclusion Chromatography

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

Fast protein separations with ion exchange columns

Protein-Pak Hi Res HIC Column and HIC Protein Standard

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1.

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Size exclusion chromatography for Biomolecule analysis: A "How to" Guide

Understanding the Effects of Common Mobile Phase Additives on the Performance of Size Exclusion Chromatography

Bivalirudin Purification:

Gentle Bioanalysis of Proteins

TOSOH BIOSCIENCE. Hydrophobic Interaction Chromatography. How hydrophobic do you want to go? Separations Business Unit

Fast Agilent HPLC for Large Biomolecules

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808

Columns for Biomolecules BioLC Column Lines

Superdex 200 HR 10/30 INSTRUCTIONS

Purification of Nucleic Acids

Preparative HPLC is still the

Multiple Detector Approaches to Protein Aggregation by SEC

YMC-BioPro NEW. porous and nonporous IEX columns. For the analysis and separation of peptides, proteins and biomolecules

Choosing the Right Calibration for the Agilent Bio SEC-3 Column

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

Characterization of IgG monomers & their aggregates

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Molecular Characterization of Biotherapeutics The Agilent 1260 Infi nity Multi-Detector Bio-SEC Solution with Advanced Light Scattering Detection

Technical Data. Analytical Data

Developing Quantitative UPLC Assays with UV

A universal chromatography method for aggregate analysis of monoclonal antibodies

ADVANCE ACCURACY AND PRODUCTIVITY FOR FASTER ANALYSIS

mab and ADC Analysis Shanhua Lin, Ph.D. The world leader in serving science

Application Note. Rapid optimisation and development of an automated two-step purification procedure for monoclonal IgG antibodies.

Application Note. Author. Abstract. Pharmaceuticals. Detlef Wilhelm ANATOX GmbH & Co. KG. Fuerstenwalde, Germany mau

Small-scale purification and analysis of protein samples by size exclusion chromatography (SEC)

ACQUITY UPLC Protein BEH SEC Columns and Standards

AdvanceBio HIC: a Hydrophobic HPLC Column for Monoclonal Antibody (mab) Variant Analysis

Next Generation Zirconia-Based Antibody Purification Media

Optimizing Protein Separations with Agilent Weak Cation-Exchange Columns

Thermo Scientific MAbPac HIC Columns. Novel Hydrophobic Interaction HPLC Columns. Designed for Monoclonal Antibody Analysis

Separation of Recombinant Human Erythropoietin (repo) Using Agilent Bio SEC-3

Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor

Biotherapeutic Method Development Guide

SEC Column Selection Guide for the Separation of Water-soluble Polymers

Superdex 75 Increase columns

Enhanced Analytical Development for Complex Antibody Formats. Markus Wild, F. Hoffmann - La Roche Ltd., Switzerland

Phenyl Membrane Adsorber for Bioprocessing

TOSOH FACT The Tosoh logo symbolizes the corporate philosophy of Tosoh s vision of the ideal. The curved lines represent the realization of

Faster Separations Using Agilent Weak Cation Exchange Columns

CX-1 ph Gradient Buffer

Biotherapeutic Method Development Guide

Econo-Pac High Capacity Ion Exchange Cartridges, 5 ml Instruction Manual. Catalog Numbers , ,

HYDROPHOBIC INTERACTION CHROMATOGRAPHY TOSOH BIOSCIENCE

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

Agilent 1290 Infinity II 2D-LC Solution Biopharmaceutical Polymer Analysis. WCBP Jan 2017 Washington, DC

Effective Scaling of Preparative Separations Using Axial Compression Columns

Fundamentals and Techniques of Preparative HPLC. Parto Zist Behboud Tel: 42108

Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC

An Automated System for At-Line Process Analysis of Biopharmaceutical Fermentation Reactions

HPLC Columns HIC PH-814 MANUAL. Shodex HPLC Columns Europe, Middle East, Africa, Russia. For technical support please use contact details shown below:

Transcription:

High Throughput Sub-4 Minute Separation of Antibodies using Size Exclusion Chromatography TSKgel APPLICATION NOTE Introduction Gel Filtration Chromatography (GFC) is a powerful analytical tool in the separation of antibodies. Traditionally, GFC columns with a dimension of 7.8 mm ID 30 cm are used for analytical purposes. The longer column dimension, however, leads to longer run times and sample dilution, as well as substantial solvent waste. Alternatively, a column of smaller dimension provides high throughput separation with shorter run times, high resolution and minimal solvent waste when used on a conventional HPLC system. This Application Note demonstrates the use of a 4.6 mm ID 15 cm TSKgel SuperSW mab HTP SEC column for the highly reproducible separation of antibodies in less than 3.5 minutes by using a moderate flow rate of 0.75 ml/min. The SuperSW mab HTP column provides excellent stability for high speed, sub-4 minute separations of monoclonal antibodies. Materials and Methods Column: Instrument: Mobile phase: TSKgel SuperSW mab HTP, 4 µm, 4.6 mm ID 15 cm Agilent 1100 HPLC system 100 mmol/l phosphate/100 mmol/l sulfate buffer, Figure 1. Separation of Monoclonal Antibodies on TSKgel SuperSW mab HTP Column under High Flow Conditions Detector response (mau) 2000 1500 1000 500 mab 01 0 0 0.5 1 1.5 2 2.5 3 3.5 4 Retention time (minutes) Mouse IgG Human IgG mab 02 Figure 2. Retention Time Reproducibility of the TSKgel SuperSW mab HTP Column under High Flow Conditions Flow rate: Detection: Temperature: Injection vol.: Samples: ph 6.7 + 0.05% NaN 3 0.75 ml/min UV @ 280 nm ambient 5 μl PABA, 0.01 mg/ml mab 01, 4.6 mg/ml mab 02, 4.6 mg/ml human IgG, 4.6 mg/ml mouse IgG, 4.6 mg/ml Results and Discussion Figure 1 shows the separation of four different monoclonal antibodies in less than three minutes using the TSKgel SuperSW mab HTP column at a flow rate of 0.75 ml/min. High resolution separation of the monomer, dimer, and fragment peaks of the mouse IgG sample are clearly shown under these conditions. The separation of these IgG-based proteins within 3 minutes using the TSKgel SuperSW mab HTP column corresponds to a 3.75-fold decrease in analysis time relative to conventional (7.8 mm ID 30 cm) SEC columns. Sustained pressure from operating at elevated flow rates can lead to voids within the column, generating poor peak shapes and drifting retention time. As shown in Figure 2, 540 consecutive injections of mab 02 and PABA separated on the TSKgel SuperSW mab HTP column at 0.75 ml/min show good reproducibility with no discernible drift in retention. Additionally, no significant loss in resolution between the mab monomer and dimer was observed on the TSKgel SuperSW mab HTP column operated at 0.75 ml/min, yielding a %RSD < 3 (Figure 3). Figure 3. Resolution Stability of the TSKgel SuperSW mab HTP Column under High Flow Conditions Resolution 1.4 1.2 1 0.8 0.6 0.4 0.2 0 Rs (monomer/dimer) 0 100 200 300 400 500 600 Number of Injections

Conclusions Separations of IgG-based proteins using the TSKgel SuperSW mab HTP column at 0.75 ml/min yielded highly reproducible results with high resolution and moderate back pressure within 3 minutes. This corresponds to a 3.75-fold decrease in analysis time relative to traditional SEC methodology. Additionally, due to the smaller dimension of the TSKgel SuperSW mab HTP column, minimal solvent waste is observed even at increased flow rates, making this a cost effective and green method for protein separations when compared to that of traditional 7.8 mm ID 30 cm SEC columns. The TSKgel SuperSW mab HTP column operated at 0.75 ml/min for 540 injections of monoclonal antibody showed no drift in retention and good reproducibility. These results demonstrate that the TSKgel SuperSW mab HTP, 4 µm, 4.6 mm ID 15 cm column clearly has a competitive advantage in fast assay and high throughput analysis of antibodies using a conventional HPLC system. Tosoh Bioscience and TSKgel are registered trademarks of Tosoh Corporation. 3604 Horizon Drive, Suite 100 King of Prussia, PA 19406 Tel: 800-366-4875 email: info.tbl@tosoh.com www.tosohbioscience.com AN77 1214

Physical Stability of a Silica- Based Size Exclusion Column for Antibody Analysis Atis Chakrabarti and Roy Eksteen Tosoh Bioscience LLC, King of Prussia, PA 19406 TP161 1

Abstract There are many answers to the question of how long a column lasts. The most trivial reply is still the most appropriate - it all depends on what you do with the column. Particle structure, bonded phase composition, number of injections, mobile phase, ph, flow rate, sample type/mass/volume, temperature, pressure, storage, cleaning protocol, et cetera - all play a more or less important role in determining when to write the column s inevitable obituary. When it comes to gel filtration chromatography, particle stability and thus column stability is even more of an issue than in reversed phase chromatography for the simple reason that the pore volume of SEC columns is maximized to provide optimal mass resolution, and the higher the pore volume the more fragile the particle. From a user s perspective, it would be most desirable if no precautions would be required to obtain a long column lifetime. From a manufacturer s perspective, such a column would be impossible to make and even if it could be made, such a column would be too costly to sell. In this initial report of our ongoing study about the stability of the TSKgel G3000SW XL column for quality control of monoclonal antibodies, we evaluated three columns each of three bonding lots. Preliminary indications are that retention times and efficiency did not change over at least 1,000 injections of a 10µL mixture of protein standards into a 100µL injection loop. During the lifetime studies the analytical column was neither protected by a guard column nor by inline filters other than the ones that are a standard fixture in the Agilent 1100 HPLC system. 2

Introduction Column lifetime is a critical factor in all modes of liquid chromatography but particularly so for silica-based SEC columns as the benefit of increased particle porosity has to be balanced with the requirement to maintain physical integrity of the particles. While the physical stability of reversed phase HPLC columns is taken for granted by its practitioners, the same cannot be said for the physical stability of polar bonded phase columns. Although polar-bonded phase columns can display comparable efficiency and peak symmetry as RPLC columns, polar bonded phase columns are sometimes plagued by a sudden loss of performance; usually from the development of a void due to resettling of the packing at the top of the column. 3

Factors playing a more or less important role in determining column lifetime Particle structure Bonded phase composition Number of injections Sample type, sample mass, and sample volume Mobile phase conditions (ph, buffer, salt) Pressure Temperature Column cleaning Storage In this study we report on variations in column efficiency, peak symmetry, and retention when making 1,000 injections of a 10uL standard mixture of globular proteins on five (5) TSKgel G3000SW XL columns operated under standard operating conditions. 4

Objective This study was conducted to determine the long term physical stability of TSKgel G3000SW XL columns as a function of operating conditions. In this initial study we report on experiments performed using a common mobile phase composition, a standard flow rate, and a small injection volume of globular protein standards. Characteristics of TSKgel SW columns Particle size: 5µm Pore size: 250Å Bonded phase: diol-containing ligands Protein calibration range: 10,000 500,000Da 5

Material and methods: Columns TSKgel G3000SW XL, 5µm, 7.8mm ID x 30cm Bonding Lot 08R o S1237 o S1238 o S1239 Bonding Lot 09R o S1261 o S1262 o S1263 Bonding Lot 30P o S6210 o S6211 o S6212 Silica Lot A Silica Lot B In this poster we report on the stability of the highlighted columns prepared from three bonding lots, made from two silica lots. 6

Chromatographic conditions stability study TSKgel G3000SW XL, 5µm, 7.8mm ID x 30cm Silica lot A o Bonding lot 08R: columns S1237, S1238, S1239 o Bonding lot 09R: Columns S1261, S1262, S1263 Silica lot B o Bonding lot 30P: columns S6210, S6211, S6212 Mobile Phase: 0.1mol/L KH 2 PO 4 /Na 2 HPO 4, ph 6.7, + 0.1mol/L Na 2 SO 4 + 0.05% NaN 3 Flow rate: 1.0mL/min Detection: UV@280nm Temperature: ambient Injection vol.: 10µL Samples TSKgel SW XL test mixture: thyroglobulin, γ-globulin, ovalbumin, ribonuclease A, p-aba (see next slide for more details) Monoclonal antibody: BI-MAb-2 (Boehringer-lngelheim), 4.5g/L in glycine/sodium phosphate, ph 6.0 7

Chemicals Protein test mixture Notes: Sample solvent: 100mmol/L phosphate buffer, ph 6.7, unless mentioned otherwise. The small molecular weight compound p-aminobenzoic acid is used to mark the mobile phase volume in the column, while the efficiency and peak symmetry of the paba peak are indicators of the integrity of the packed bed. All chemicals and standards were obtained from Sigma-Aldrich and were of electrophoretic or analytical grade. High purity HPLC grade solvents were used for the preparation of stock standards, samples and mobile phases. 8

Column protection Although it is recommended practice to protect the column from potential sources of contamination no such precautions were made in this study. Standards and mobile phases were not filtered through a 0.45μm syringe filter. A frit filter was not used between injector and column. TSKgel G3000SW XL columns were not protected by a guard column in any of these studies. Note: the PTFE Purge Valve Frit on the high pressure side of the Agilent 1100 HPLC system was replaced after about 3,000 injections (~ 45L mobile phase). Dirty Frit New Frit 9

Results Tables (I) N, AF and t R at 1 st, 500 th and 1000 th injection on S1261-09R (II) %RSD values of N, AF and t R during 1000 injections Figures 1. Precision: column S1261-09R 2. Precision: columns from bonding lot 08R 3. Precision: columns from bonding lots 08R + 09R 4. Precision: between silica lots A and B 5. Sample load protein standards 6. Sample load monoclonal antibody 10

Table I: Retention, efficiency and peak symmetry at the 1 st, 500 th and 1000 th injection Table II: %RSD (n=100) over 1000 injections 11

Injection-to-injection reproducibility: within column High degree of reproducibility from injection to injection is illustrated by superimposed chromatograms of the 1 st and 1000 th injection on column S1261-08R. 12

Column-to-column reproducibility within bonding lot High degree of reproducibility within a bonding lot is illustrated by superimposed chromatograms of the 250 th injection on three columns from bonding lot 08R. 13

Reproducibility between bonding lots High degree of reproducibility between bonding lots is illustrated by superimposed chromatograms of the 1000 th injection on one column each of bonding lots 08R and 09R. 14

Reproducibility between silica lots High degree of reproducibility between silica lots is illustrated by superimposed chromatograms of the 1 st and 1000 th injection on one column each of silica lots A and B. 15

Loading capacity of protein standard Efficiency of γ-globulin declined slightly, while peak shape and retention time were not affected when injecting 100µL sample containing 6mg protein. 16

Loading capacity of monoclonal antibody Peak shape and efficiency were not affected when injecting 400µg of a monoclonal antibody preparation. 17

Preliminary conclusions 1 Based on five TSKgel G3000SW XL columns tested, each column was stable for at least 1,000,10µL injections of a protein standard mixture in 0.1mol/L, ph 6.7 phosphate buffer containing 0.1mol/L sodium sulfate and 0.05% sodium azide, operated at 1mL/min at room temperature. Efficiency (N) and peak symmetry (AF) values showed minimal variation from the first to the last (1000 th ) injection indicating that the integrity of the packed bed was maintained throughout the experiment. Retention times of globular protein standards did not change while pumping 15L mobile phase through the column, indicating that the integrity of the diol-bonded phase was maintained throughout the experiment. Plate numbers, peak symmetry values and retention times for each protein studied showed minimal variation for TSKgel G3000SW XL columns prepared from: The same bonding lot 3 columns tested. Different bonding lots one column each of 3 lots tested. Different silica lots one column each of two lots tested. 18

Preliminary conclusions 2 Efficiency, peak symmetry and retention times did not change when injecting a 100µL of a standard protein mixture containing 6mg total protein. Chromatographic parameters were also not affected when injecting 0.4mg of a monoclonal antibody preparation. 19

Future studies When performing column stability studies one first needs to understand the effect of common chromatographic conditions on the integrity of the packed bed before one can study how real samples influence column lifetime. In the next phase of this study we plan to investigate the role of injection volume and the role played by the nature of the injection solvent on column stability. Another topic of interest is the role of mobile phase additives, such as organic modifiers and detergents. 20