Ultra-fast ICP-OES determinations of base metals in geochemical samples using next generation sample introduction technology

Similar documents
Analysis of steel and its alloys using the GB/T standard and an Agilent 5100 ICP-OES in dual view mode

Multi-Element Analysis of Trace Metals in Animal Feed using ICP-OES

Determination of 22 Elements Following US EPA Guidelines with a New Megapixel CCD ICP-OES

Analysis of Environmental Samples with the Agilent 730-ES Following US EPA Guidelines

Increase productivity for environmental sample analysis using the SVS 2+ Switching Valve System for Agilent 5100 SVDV ICP-OES

DETERMINATION OF ELEMENTS IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS 10500, & USING THE AGILENT 5100 ICP-OES

Determination of Elemental Impurities in Graphite-based Anodes using the Agilent 5110 ICP-OES

Analysis of wear metals and contaminants in engine oils using the 4100 MP-AES

Benefits of running organic matrices using the Agilent 5100 ICP-OES fast, robust, high performance analysis

The Determination of Alloying Elements in High-Carbon Steel using Radially Viewed Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES)

Automating EPA 6020 Compliant Analysis with the Agilent 7900 ICP-MS and ESI prepfast Autodilution System

The Agilent Atomic Spectroscopy portfolio for Environmental applications AA, MP-AES, ICP-OES, ICP-MS & ICP-QQQ

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES

Table I: MCL and MRL Concentrations for Contaminants Monitored Under the Safe Drinking Water Act National Primary Drinking Water Regulations

Composition Analysis of Animal Feed by HR ICP-OES

Application Note. Abstract. Authors. Environmental

Increased Sample Throughput for ICP-OES Applied to U.S. EPA Method 200.7

Analysis of Heavy Metals in Iron-based Fertilizers by HR ICP-OES

Accurate. Robust. Reliable. AGILENT 710 SERIES ICP-OES

Accurate. Robust. Reliable. AGILENT 710 SERIES ICP-OES

Atomic Spectroscopy Solutions for Environmental Applications

Water Analysis Using ICP-OES with an Ultrasonic Nebulizer

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

Achieving Optimum Throughput in ICP-MS Analysis of Environmental Samples with the Agilent 7500ce ICP-MS Application

Amplifying ICPMS Productivity Using Discrete Sampling Technology

Analysis of Toxic Trace Elements in Coffee Bean Products by HR ICP-OES

High matrix tolerance of the Thermo Scientific icap 7000 Plus Series ICP-OES with the radial Ceramic D-Torch and the sheath gas adaptor

Agilent 5100 ICP-OES DUAL VIEW ICP-OES MINUS THE WAIT

EU water analysis using the Thermo Scientific icap 7400 ICP-OES Duo

EU Water Analysis Using the Thermo Scientific icap 7400 ICP-OES Duo

Analysis of Stainless Steel by Dual View Inductively Coupled Plasma Spectrometry

Ultra-Trace Analysis of Beryllium in Water and Industrial Hygiene Samples by ICP-MS Application

Analysis of Mineral and Heavy Metal Content in Beverages Using the Teledyne Leeman Labs Prodigy Plus ICP-OES

Application Note PlasmaQuant PQ 9000 Elite. Analysis of Rare Earth Elements in Granite and Sandstone by HR ICP-OES. Challenge.

Analysis of Trace Elements in Seawater Using the Thermo Scientific icap 7000 Series ICP-OES Duo

Analysis of Stainless Steel by Dual View Inductively Coupled Plasma Spectrometry

The image part with relationship ID rid4 was not found in the file. Welcome

Analytically Monitoring the Effect of Fracturing Activity

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

Analysis of 14 Trace Elements in Stainless Steel

Analysis of inorganic impurities in lubricating oils by ICP-MS

Tender Document. National Centre of Excellence in Analytical. Chemistry, University of Sindh, Jamshoro. Inductive Coupled Plasma (ICP -OES) ISSUED BY:

Direct Analysis of Photoresist by ICP-MS. Featuring the Agilent Technologies 7500s ICP-MS

Common Challenges facing Laboratories doing Elemental Analysis Today

ICP Expert software. Technical Overview. Introduction

Analysis of Trace Metals in Surface and Bottled Water with the Optima 7300 DV ICP-OES

The Determination of Trace Elements in Stainless Steel by Forked Platform GFAAS

ESI SC-FAST Sample Introduction Solutions for PerkinElmer ICP-OES and ICP-MS Systems

Inductively coupled plasma (ICP) spectroscopy for online measurements of trace metal emissions at the 250 kw PACT furnace

Rubber and plastics gloves for food services Limits for extractable substances

Guidelines for Trouble Shooting and Maintenance of ICP-OES Systems. Eric Vanclay Product Manager Spectroscopy Consumables May 2, 2012

ICP-AES Determination of 15 Kind of impurity elements in the Vanadium-aluminum alloy

Fields of Application / Industry:

Functionality. Normal Operation

Trace Elements in Petrochemical Samples: Achieving Excellent Analytical Results

PROCUREMENT PROCEDURE OF CPRI (NON WORKS) Dt of Revision : Issue Dt. :

EXPAND YOUR RANGE EXTEND YOUR RESOURCES. Avio 500 ICP Optical Emission Spectrometer

ICP-OES. Varian Vista-MPX CCD Simultaneous ICP-OES

Avio 200 ICP-OES Flat Plate

The Determination of Toxic Metals in Waters and Wastes by Furnace Atomic Absorption

Agilent 4200 Microwave Plasma-Atomic Emission Spectrometer CHANGE IS IN THE AIR

Collaborative Efforts for Creating a Robust Trace Metals Analytical Procedure for Flue Gas Desulfurization Wastewaters by ICP-MS

INTRODUCTION. Journal of International Scientific Publications Ecology & Safety ISSN , Volume 9, 2015

Analysis of elemental impurities in drug products using the Thermo Scientific icap 7400 ICP-OES Duo

Automated Real-Time Determination of Bromate in Drinking Water Using LC-ICP-MS and EPA Method Application

Runs On Air. AGILENT 4100 MICROWAVE PLASMA ATOMIC EMISSION SPECTROMETER

Merits of EPA Method 1640:

The Determination of Trace Elements in Waters using the US EPA Method and the Thermo Scientific icap 6500 ICP-OES (Duo)

OPTIMA 4000 DV Series of ICP-OES Systems

Agilent Atomic Spectroscopy Solutions for Energy & Chemicals AA, MP-AES, ICP-OES, ICP-MS.

STUDY OF ELEMENT FINGERPRINTING IN GOLD DORÉ BY GLOW DISCHARGE MASS SPECTROMETRY

Analysis of Samples Related to Nuclear Submarine Operation

ElvaX ProSpector in Exploration & Mining

Multi-element determination in pharmaceutical preparations using the Thermo Scientific icap Q ICP-MS

Radial, Axial or Dual View ICP: Which Do You Choose? Manny Almeida Teledyne Leeman Labs, Inc. Hudson, NH

Optima 8x00 Series. ICP Optical Emission Spectrometers

ELEMENTS by ICP 7301 (Aqua Regia Ashing)

Optima 8x00 Series. ICP Optical Emission Spectrometers

Agilent 720/725 ICP-OES RELIABLE. PRODUCTIVE. ROBUST.

National Food Safety Standard

USABILITY BY DESIGN Every component of the Agilent 7900 ICP-MS is engineered for performance and reliability

DRPK-005

FOUNDRY-MASTER Pro. First class laboratory optical emission spectrometer for complete professional metal analysisoes

Analysis of Cast Iron Using Shimadzu PDA-7000

Agilent 5100 ICP-OES DUAL VIEW ICP-OES MINUS THE WAIT

Spectroscopy. Agilent Supplies for ICP-OES, ICP-MS & MP-AES. Agilent Supplies for AA. Agilent Supplies for UV/VIS

Evaluation of the Mercury Concentration Accessory for US EPA Methodology

Kunikazu Ide and Yoshisuke Nakamura. 1. Introduction

SECTIOn 4. Claritas PPT Multi-Element Standards for ICP-MS

ELEMENTS by ICP 7301 (Aqua Regia Ashing)

Characterization of Surface Metals on Silicon Wafers by SME-ICP-MS. Featuring the Agilent Technologies 7500s ICP-MS

Flame Atomic Absorption Spectrometer (FAAS) Brand : PERKINALMER Model : PinAAcle 900T Custodian : PRADUP MESAWAT SOOKRUTHAI WONGSUBIN

Metal Aerosol Emissions from Biomass Combustion

Seamless Method Transfer between the Agilent Cary 8454 UV- Visible Spectrophotometer and the 8453 UV-Visible Spectrophotometer

icap 7000 Plus Series ICP-OES Streamlined performance and ultra-low ICP-OES detection limits for routine laboratories

ELEMENTS by ICP 7300 (Nitric/Perchloric Acid Ashing)

Eco Plasma Innovative Inductively Coupled Plasma Technology

Assaying and Analysis Choices in Practice

REQUEST FOR TENDER. Supply of a Dual View (radial & axial) ICP-OES instrument Issue date 04 July 2014 Briefing session Closing date and time

Transcription:

Ultra-fast ICP-OES determinations of base metals in geochemical samples using next generation sample introduction technology Application note Geochemistry, mining and metals Author John Cauduro Agilent Technologies Melbourne, Australia Introduction Mines, mineral processors and geologists involved in mineral exploration require accurate, precise analysis with rapid sample turn-around times. To remain viable in a competitive global environment, geochemical laboratories are analyzing increasing sample loads at lower and lower detection limits, making simultaneous ICP-OES the preferred analytical technique. This work describes the high speed analysis of certified reference materials using the Agilent 725 simultaneous ICP-OES with a switching valve system (Agilent SVS 2) and an inert multipurpose nebulizer (Agilent OneNeb). The SVS 2 provides greater sample throughput and the OneNeb provides a robust sample introduction system capable of handling high throughput of challenging geochemical digests. The reference materials were prepared using the popular and effective 4-acid digest comprised of a combination of hydrochloric acid (HCl), nitric acid (HNO 3 ), perchloric acid (HClO 4 ) and hydrofluoric acid (HF).

The new and innovative sample introduction system, termed the Switching Valve System 2 (SVS 2) comprises a software-controlled multi-port switching valve, a sample loop, and a high-speed, positive displacement pump. Using this system greatly reduces the sampleto-sample cycle, reducing argon consumption, while maintaining analytical accuracy. By limiting the amount of sample used during sample analysis, wear on components such as the nebulizer, torch and pump tubing is reduced, further minimizing operating costs. The Agilent OneNeb nebulizer offers superior performance for this application over other comparable nebulizers. The OneNeb increases nebulization efficiency by generating an extremely fine aerosol, which greatly improves sensitivity and signal-to-noise ratio, and lowers detection limits. This allows lower volumetric flow rates to be used for analysis, reducing sample consumption. Experimental Instrumentation An Agilent 725 simultaneous ICP-OES with radiallyviewed plasma was used for the analysis. The Agilent 725 ICP-OES features a custom-designed CCD detector, which provides true simultaneous measurement with comprehensive wavelength coverage from 167 to 785 nm for fast, precise measurements of all elements of interest. The optical system is housed within a thermally-stabilized environment at 35 ± 0.1 C and contains no moving parts, ensuring excellent longterm stability. For measuring elements that require low UV wavelengths (<190 nm) the optical path can be purged with either argon or nitrogen gas for improved performance, although this was not required for this analysis. The highly efficient RF generator produces a robust plasma, and the vertically-orientated radial plasma provides excellent solids handling, making it ideally suited to geochemical samples which typically have high concentrations of strong acids and high levels of total dissolved solids. The Agilent 725 ICP-OES was fitted with a mass flow controller and a five-channel peristaltic pump. The fivechannel peristaltic pump allows the sample, internal standard/ionization buffer solution, rinse solution and the waste to be simultaneously pumped. A conventional one-piece radial torch was used. The sample introduction system consisted of the OneNeb nebulizer and a double-pass glass cyclonic chamber. Agilent ICP Expert II software was used for instrument operation. A plasma gas flow of 10.5 L/min was used in the analysis, which reduced the overall argon consumption. This lower Ar consumption, coupled with the higher throughput with the SVS 2, offers significant cost savings. Solutions were automatically presented to the spectrometer using the Agilent SPS 3 Sample Preparation System, controlled by the operating software. Tables 1 and 2 list the operating conditions used for the ICP-OES and the SVS 2 during this analysis. Table 1. ICP-OES instrument operating parameters 2 Condition Power Plasma gas flow rate Auxiliary gas flow rate Spray chamber type Torch Viewing height Nebulizer type Nebulizer flow rate Pump tubing Pump speed Replicate read time Number of replicates 3 Sample uptake delay time Stabilization time Rinse time Fast pump Setting 1.2 kw 10.5 L/min 1.5 L/min Glass cyclonic (double-pass) Standard one piece quartz radial 10 mm OneNeb 0.8 L/min Rinse/instrument pump: Black-black tabs (0.76 mm ID) Waste: blue-blue tabs (1.65 mm ID) 20 rpm 3 s 0 s 20 s 0 s Off

Table 2. SVS 2 operating parameters Condition Loop uptake delay Uptake pump speed refill Uptake pump speed inject Sample loop size Time in sample Bubble inject time Setting 5.0 s 500 rpm 108 rpm 1.0 ml 4.5 s 4.9 s Inter-element corrections (IEC) One of the main challenges when analyzing geochemical digests is that they often contain a diverse range of elements at varying concentrations. This leads to unpredictable spectral interferences that may not be resolved by the high-resolution polychromators found on today s ICP-OES instruments. A reliable method for analyzing geochemical samples must overcome these interferences. The continuous wavelength coverage and comprehensive wavelength database in ICP Expert II makes finding a wavelength free of interferences easy, but when this is not possible the IEC feature in ICP Expert II provides an easy-to-use solution to spectral interferences. An example is shown in Figure 1. Silver (Ag) at 328.068 nm has a spectral interference from zirconium (Zr). Manganese and molybdenum also present possible interferences, but are not shown here. Hence IECs were used for this and other lines, giving excellent results particularly when close to the limit of detection (LOD). Off-peak background correction is the recommended form of background correction when using IECs. Samples and sample preparation Ore-grade Certified Reference Materials (Geostats) were prepared by taking 0.4 g and performing a 4-acid digestion using a combination of HF, HClO 4, HCl, and HNO 3 at temperatures up to 260 C. Hydrofluoric acid is used in to dissolve silicates, resulting in a neartotal digestion. The 4-acid digestion is a very effective dissolution procedure for multi-element analysis for a wide range of elements. However, the method is not suitable for volatile elements such as B, As, Pb, Ge, Sb or rare earth elements (REE). Figure 1. Zirconium interference shown to the right of the silver line at 329.068 nm Calibration solutions were prepared from custom-grade single-element solutions, supplied by Merck, ACR chemical reagents and Inorganic Ventures Inc. The calibration standards and calibration blank solutions were prepared in deionized water (> 18 MΩ cm) supplied from a Millipore system and stabilized with 30% v/v HCl (Merck). Table 3 lists the selected elemental wavelengths and concentration range covered by the calibration standards for each element. Sensitivity, linear dynamic range and freedom from spectral interferences were taken into consideration during wavelength selection. Results and discussion Certified concentrations of base metals in the respective ore reference materials are reported in Table 4. Analyses were performed in various combinations and the error reported for each result represents the largest variation from the mean value. The LOD is based on three sigma of ten replicate analysis of the blank solution during the analytical run. Table 5 shows the detection limits (in the solid) for the analysis. Detection limits are well below those required for the analysis, with the exception of As in some ore materials. 3

Table 3. Element wavelength, CCV concentration and IEC concentration ranges for the elements Element Wavelength (nm) Calibration range CCV IEC Ag 328.068 0.1 4.0 2 10 Al 237.312 10 500 200 200 500 As 193.696 2.0 100 50 500 Ba 233.527 1.0 50 20 50 500 Cd 226.502 0.5 25 10 25 Analyte Co 237.863 2.0 100 50 10 50 Cr 267.716 1.0 50 20 20 50 Cu 327.395 2.0 500 50 200 500 Fe 239.563 20 1000 500 500 500 Mn 257.61 2.0 100 50 100 500 Mo 284.824 10 500 100 100 500 Ni 231.604 2.0 500 50 500 500 Pb 405.781 2.0 100 50 500 Zn 206.200 2.0 100 50 500 Zr 343.823 0.5 10 10 2 10 Figure 2. Time plot. Number of samples per hour. Although a small amount of precipitate was observed in the digestion, the overall measured concentrations were in good agreement with the certified data. Using the SVS 2 significantly improved throughput, cutting the sample-to-sample time by half, from 1 minute and 30 seconds per sample, to 45 seconds. This increases sample throughput from 40 samples an hour to 80 samples an hour, as shown in Figure 2. As an added benefit, the daily argon consumption was greatly reduced, as shown in Figure 3. Figure 3. Daily argon consumption Conclusion This work has demonstrated the ability of the Agilent 725 simultaneous ICP-OES with radially-viewed plasma to accurately measure base metals in challenging geochemical samples. The highly efficient RF generator of the 725 ICP-OES transfers RF energy efficiently into the plasma to produce a robust, stable plasma even at low plasma gas flows of 10.5 L/min Ar. The low Ar plasma flow and efficient spectrometer purge use a total argon consumption of only 15.5 L/min, and when combined with the doubling of sample throughput from 40 samples an hour to 80 samples an hour using the SVS 2, running costs are slashed. The powerful and easy-to-use IEC feature in ICP Expert II means a reliable method can be set up to handle the wide range of unpredictable spectral interferences often encountered in geochemical samples. 4

Table 4. Comparison of results with Certified Base Metal Reference Material CRM Element Units Ag As Co Cu Ni Pb Zn GBM398-4 Certified mg/kg 48.7 12 1974 3891 4071 11714 5117 Triplicate analysis SD 5.1 9 125 195 187 776 229 Measured mg/kg 50.14 <LOD 1877 3662 3780 11566 5188 SD 0.44 26 104 33 157 43 Recovery % 103.0 <LOD 95.1 94.1 92.8 98.7 101.4 GBM399-5 Certified mg/kg 24.2 320 46 29424 24412 21173 9493 Triplicate analysis SD 1.5 27 5 1446 1248 1402 504 Measured mg/kg 24.60 297 45.3 26608 22899 21147 8752 SD 0.27 11 2.2 286 216 160 90 Recovery % 101.6 92.7 98.5 90.4 93.8 99.9 92.2 GBM908-14 Certified mg/kg 303.7 NA NA 23715 NA 32955 42716 Single analysis SD 21.9 704 1185 1503 Measured mg/kg 333.3 228 119 23727 15.0 35736 Recovery % 109.7 NA NA 100.1 NA 108.4 101.5 Table 5. Limit of detection results in mg/kg of sample Element Ag As Co Cu Ni Pb Zn LOD 0.75 37 1.3 43 12 99 66 References 1. Hoobin, D. Ultra-fast ICP-OES determinations of oil and plant material using next generation sample introduction technology. 2. Hoobin, D. Ultra-fast ICP-OES determination of trace elements in water, conforming to US EPA 200.7 and using next generation sample introduction technology. 3. Kovachev, N. Development of a new liquid sample introduction system for ICP-OES and ICP-MS systems (poster). 5

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc. 2012 Published August 27, 2012 Publication number: 5991-0876EN