Understanding ANSI/AAMI ST 91. Flexible and Semi-Rigid Endoscope Reprocessing in Healthcare Facilities

Similar documents
Low Temperature Processing. An Update on Guidelines, Standards and Requirements

ANSI/AAMI ST91: 2015

Best Practices for High Level Disinfection Part I

MDRAO Importance of Bedside & Manual Precleaning CONMED.CA

House Keeping. Introduction to ANSI/AAMI ST91:2015: Flexible and Semi-Rigid Endoscope Processing in Health Care Facilities. Welcome!

Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes Webinar FAQ

Administrative Policies and Procedures. Cross Reference: Date Issued: 2/14 Date Reviewed: 9/14 Date: Revised: 12/14 Attachment: None Page of 1 of 6

Disinfection & Sterilization

3/11/ M Sterile U Network 3M Sterile U Web Meeting March 11, Welcome! Topic: An Overview of ANSI/AAMI ST58: 2013

Best Practices for Reprocessing Endoscopes

TOTAL SOLUTION FOR ENDOSCOPE REPROCESSING

Christina Bradley Endoscope Decontamination A Webber Training Teleclass. Christina Bradley

Competency Guide: Care and Handling of Rigid Endoscopes

Slide 1. Slide 2. Slide 3

SUBJECT: Detection of Blood (hemoglobin) residue inside the biopsy channel of an endoendoscope

SUBJECT: Detection of residual organic soils inside various channels of lumened items.

ENDOSCOPY PRACTICE Manual cleaning and Disinfection of flexible endoscopes

This policy has been adopted by UNC Health Care for its use in infection control. It is provided to you as information only.

8/27/2015. Learning Objectives. Purpose of Sterilization. To destroy all living pathogenic microorganisms

ENDOSCOPE DECONTAMINATION- EDUCATION AND TRAINING COMPETENCY FRAMEWORK

How is Sterilization Validation Performed by the Medical Device Manufacturer?

Endoscopy What s New Up There? Frank Myers, MA, CIC, FAPIC

Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes

Endoscopy and OR Solutions SYSTEM 1 TM. Express STERILE PROCESSING SYSTEM SAFETY, ASSURANCE AND EFFICIENCY. One Integrated Approach to Healthcare

Is contamination of bronchoscopes really a problem?

Objectives. The Perioperative Nurse s Role 6/24/2016. Instrument Processing & Sterilization Chesapeake Perioperative Consortium 2016

MDRAO ONLINE COURSE FOURTH EDITION 2018

Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar

THE BASICS OF STERILIZATION

Dissecting Outbreaks of Multi-resistant Bacteria in Gastrointestinal Endoscopy

3/26/2014 SPAULDING CLASSIFICATION OBJECTIVES PRINCIPLES OF DISINFECTION AND STERILIZATION IN THE DENTAL SETTING PROCESSING SEMI CRITICAL OBJECTS

PRINCIPLES OF DISINFECTION AND STERILIZATION IN THE DENTAL SETTING OBJECTIVES

Manner of Assessment Code (check all that apply) & Surveyor Notes. Yes N/A. Yes. Yes N/A N/A. Yes N/A. N/A Yes N/A

Guidelines for Selection and Use of Disinfectants

Chemical Sterilants and High-Level Disinfectants: Managing Practical, Safe and Effective Use

8/23/2012. Continuing Education Contact Hours. How much of your formal education focused on sterilization or sterile processing?

SUBJECT: Detection of Blood residue inside the biopsy channel of a scope

Elevator (Scope Issues) Going Up! Frank Myers, MA, CIC Infection Prevention Clinical Epidemiology UCSD Health

Comparison of ANSI/AAMI ST-91 Standards, SGNA Standards and AORN Guidelines on Processing Endoscopes March 9, 2017

Monitoring Flexible Endoscope Channels to Assure Cleaning

REPROCESSING OF FLEXIBLE CYSTOSCOPES

Ideal Sterilization Method. Sterilization. Sterilization Practices in Healthcare Facilities

Facility: Sterile Processing Assessment Date: CRITERIA ANSI/AAMI COMPLIANT Number Y N N/A Point of Use Gross contaminant is removed with

ENDOSCOPE REPROCESSING AND INFECTION CONTROL FOR ENDOSCOPY

Loaner Program Guidance

Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy!

Overview: High Level Disinfection

*Slides updated to include newer guidelines in February 2016 by Shelby Lassiter, BSN, RN, CPHQ, CIC

Endoscopy Reprocessing: Manual versus Automated

Sterilization Policy. Georgia Regents Medical Center Policy Library. Policy Owner: Epidemiology POLICY STATEMENT

Best Practices for High Level Disinfection - Revisited. Objectives. The Bible for Chemicals ST-58 FDA. What is Available? 9/6/2015

Principles of Disinfection and Sterilization in the outpatient setting

Principles of Disinfection and Sterilization in the outpatient setting

BIOTECH-GERMANDE ENDOSCOPE REPROCESSING WITH BRUSHLESS CLEANING PROCESS CONDITIONS. Lionel PINEAU PhD

St Boniface Research Centre Winnipeg, Manitoba Canada

Is Your Sterile Processing Department Ready for Survey?

Loaner Programme Guidance

Loaner Programme Guidance

Bio-Burden Reduction in Biological Laboratories

Cleaning, Disinfection and Sterilization. Basics of Infection Prevention 2-Day Mini-Course October 2016

Sterilization, disinfection and reprocessing

ANSI/AAMI ST91:2015 Flexible and semi-rigid endoscope processing in health care facilities

Amanda Billings, MPH, CLS, CIC

CDWA3 / CDWA4 CDWAH. Page 4. BASKET ZONE: water pressure, spray system performance, detergent chemistry and concentration, temperature.

Endoscope Microbial Surveillance Testing Made Easy

5 Updates to Flexible Endoscope Reprocessing

Medical Device Reprocessing Assessment - General (Cleaning, Disinfection, and/or Sterilization)

ADVANTAGE PLUS Pass-Thru

AUTOMATED ENDOSCOPE REPROCESSOR. OER-Pro Easy. Fast. Reliable.

FREQUENTLY ASKED QUESTIONS FOR THE ADVANTAGE

THE TERMS VALIDATION AND VERIFICATION ARE FREQUENTLY

Obtaining Manufacturer s IFU: Challenges and Barriers

Update on Duodenoscope Related CRE Transmission. Disclosures. Outline 9/2/2016. None

Infection Prevention in Endoscopy, Keeping Patients and Practioners Safe

disinfectionandsterilization.org

Guidelines for the use of disinfectants in hospital settings. Eleni Tomprou Infection Control Nurse, MSc General Hospital of Athens Polykliniki

Agent Mechanisms of Action Comments Surfactants. Membrane Disruption; increased penetration Denature proteins; Disrupts lipids

Transmissi (ERCP) visit. by multiple. sites. it. perform the to this safety. time to help. provided by a. Interim. b. Interim. Method: c.

The Decontamination, Storage and Use of Flexible and Rigid Endoscopes

Monitoring the disinfection of flexible endoscopes using ATP bioluminescence

The Evolution of Liquid Chemical Sterilization. For High-Speed Device Turnaround

Facility: Audit Date Auditor Name/s

Sterilization, Disinfection and Antisepsis

Sterilization and Disinfection: Getting It Right. Instructions

The purpose of this DOPs form is to provide a universal training and assessment tool for continuity when training in manual cleaning processes.

DEFENDO. Sterile Single-use Valves. Procedure

Disinfection & Sterilization in the Ambulatory Surgery Center

The Gold Standard in Low Temperature Sterilization: Ethylene Oxide

Guard by TSSU. October 4,2009

#003. Liquid Chemical Sterilization Story. The. A Continuing Education Study Guide presented by

Endoscope Reprocessing and Infection Prevention

STERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation

Draft Guidance for Industry and FDA Staff Processing/Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling

Controversies in Reprocessing Flexible Endoscopes: High Level Disinfection or Terminal Sterilization????

Endoscope Decontamination. A Tube in a Hole??

A GLIMPSE AT THE TRUE COST OF REPROCESSING ENDOSCOPES:

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of

Safety, Efficacy & Microbiological Consideration

Products made in Switzerland

Transcription:

Understanding ANSI/AAMI ST 91 Flexible and Semi-Rigid Endoscope Reprocessing in Healthcare Facilities

Continuing Education Contact Hours Participants must complete the entire presentation/seminar to achieve successful completion and receive contact hour credit. Partial credit will not be given. All of the presenters are employees of STERIS Corporation and receive no direct compensation other than their normal salaries for participation in this activity. STERIS Corporation is an approved provider of continuing nursing education by the California Board of Registered Nursing, provider number CEP 11681 for 1 contact hour along with IAHCSMM and CBSPD. STERIS Corporation is providing the speakers and contact hours for this activity. However, products referred to or seen during this presentation do not constitute a commercial support by the speakers.

Learning Objectives Upon completion of this course, you will be able to: Provide an overview of the applicable standards and guidelines for flexible endoscope reprocessing Understand ANSI/AAMI ST 91, a comprehensive guide to flexible and semi-rigid endoscope reprocessing in healthcare facilities Define the key points in the reprocessing of flexible and semi-rigid endoscopes

Processing Processing (or reprocessing) is a process carried out on a reusable medical device to allow its subsequent safe use, which can include cleaning, disinfection, sterilization and related procedures.

Spaulding Classification System Spaulding Classification of Devices or Items Critical (Penetrates the vascular system or sterile areas of the body) Ex: Surgical Instruments Level of Disinfection Sterilant (Sporicidal) Interim Disinfection (To be followed with separate subsequent decontamination steps) Interim Disinfection can be utilized Semi-Critical (To be used on mucous membranes or non-intact skin) Ex: Flexible Endoscopes Sterilant/High Level (Sporicidal or High Level = 100% kill of all vegetative organisms) Interim Disinfection can be utilized Non-Critical (Contact intact skin) Ex: Blood Pressure Cuff Intermediate (TB-cidal) Interim Disinfection can be utilized

Typical Endoscope Reprocessing Cycle

Infection and Device Risks Device damage Mishandling, inadequate processing Toxic substances Chemicals Infections Endogenous - Colonized patient s flora Spread from GI tract to bloodstream/adjacent organs during procedure Exogenous sources Inanimate environmental surfaces Medical equipment, devices Medications and injection equipment

CLEANR Study Direct Observation Manual cleaning n = 69; p = 0.001 Ofstead et al., Gastroenterology Nursing, 2010

Current risks are outdated and inaccurate (Ofstead et al, 2013; Dirlam-Langlay et al, 2013) Most outbreaks not investigated or published Lack of surveillance for post-endoscopic infection No central repository for reporting Reviews of reprocessing practices show widespread lapses in essential steps Risks are greater than just infections (e.g., toxicity with aldehydes)

Regulations, Standards and Guidelines Regulations Rule or directive made or maintained by an authority Mandatory Standards Requirements and specifications consistency and fit for purpose Voluntary but can become mandatory Guidelines, Recommended Practices, Technical Information reports Technical guidance, information or preferred procedures regarding a given topic Voluntary but with interpretation

Examples

AAMI/ANSI ST 91 Contains comprehensive information and direction for processing devices and accessories Includes Flexible GI endoscopes Flexible bronchoscopes Surgical flexible endoscopes Semi-rigid operative endoscopes

Information in ST 91 Definitions Design of endoscope reprocessing areas Personnel considerations Cleaning High level disinfection Automated endoscope reprocessors (AERs) Liquid chemical sterilization Gaseous chemical sterilization Processing accessories Storage and transportation to site of use Quality control Quality process improvement Informational annexes

Additional Standards References

Work Flow Design Physical separation from patient care Adequate space for all functions Unidirectional - dirty clean (HLD/ sterilization) storage Pass-through window (recommended) Hand hygiene Emergency eyewash

Decontamination Area Sink requirements Two of adequate size Leak testing, manual cleaning Rinsing only Third decontamination sink for treated water rinsing Flushing capabilities Third clean sink for rinsing disinfected items

Physical Facilities Environmental surfaces Floors, horizontal surfaces cleaned at least daily Lighting Electricity Clean compressed air Ventilation Negative in decontamination Positive in clean area 10 exchanges/hour Temperature (60-73 F), monitored Relative humidity 20%-60%

Personnel Education, Training, Competency Verification Written policies and procedures in place Endoscopes, accessories Guidelines and standards Safety requirements OSHA, standard precautions, hand hygiene, PPEs Education and training Upon hire, annually, designated intervals, new products or devices introduced Certification recommended Competency verification Each processing task

Pre-Cleaning at Point of Use Appropriate PPE Exterior surfaces wiped with non-lint sponge/cloth Flush channel(s) with water/cleaning solution Follow manufacturer IFUs Separate endoscope from other instruments Avoid damage Attach fluid resistant cap Prepare for transport

Pre-Cleaning at Point of Use Bedside procedure Reduce levels of microorganisms and soil Prevent soil drying Reduce risk of biofilm development

Transport Closed, labeled transport containers Biohazard label

Detects damage Leak testing Defects harbor microorganisms Perform prior to immersion to prevent fluid invasion Allow adequate time to identify leaks Scope loosely coiled Manipulate knobs and buttons Follow manufacturer IFUs Multiple types of testing processes Manual, mechanical Mechanical AER Document for records

Manual Cleaning Fresh cleaning solution at correct temperature Submerge endoscope and accessories to prevent splashing Clean surfaces with soft, line-free cloth or sponge Attach irrigators to flush all channels Select appropriate brush size Clean all channels, valves, buttons, cylinders and elevators until no visible soil is seen Flush, brush, flush until clean Additional mechanical cleaning may be used

Endoscope Accessories Reusable accessories, valves, tubing processed per manufacturer IFUs Disassembled and cleaned Inspect for integrity AER manufacturer validates processing

Cleaning Chemistries Labeled for endoscopes Typically neutral detergents May or may not contain enzymes Follow IFUs for water temperature, dilution rates Essential features Optimum cleaning performance Material/device protection Water quality Toxicity validation

Manual Rinse Thoroughly rinse with copious amounts of potable water Follow AAMI TIR 34 for quality of water Rinse all external and internal surfaces Remove residual soils Remove cleaning chemicals Purge channels with air to remove residual water Dry exterior with lint-free cloth Keep detachable valves, buttons together with endoscope as a set

Definitions High level disinfectant A product that should inactivate all microbial pathogens, except large numbers of bacterial spores, when used according to labeling A HLD is often a liquid chemical sterilant (LCS) used for a shorter exposure time than that required to pass an FDA-defined spore inactivation test High Level Disinfection Product that should kill all microbial pathogens but not necessarily high numbers of bacterial spores Disinfection and rinsing need to be controlled Liquid chemical sterilant Product validated to provide microbial kill adequate to obtain FDA clearance for a sterilization label claim Sterilization Validated process used to render a product free from viable microorganisms, including bacterial spores. Liquid or gaseous process

High Level Disinfectant/Sterilant Examples Chemicals Sterilant High Level Disinfection Notes 3.4% glutaraldehyde 20.1% isopropanol 8 hours at 20 o C 10 mins at 20 o C Requires activation 3 rinses following exposure 2% hydrogen peroxide 6 hours at 20 o C 8 mins at 20 o C No activation 1 rinse 0.575% OPA No claim (passes sporicidal test at 32 hours at 20 o C) 10 mins at 20 o C No activation 3 rinses 2.4% glutaraldehyde 10 hours at 25 o C 45 mins at 25 o C Requires activation 3 rinses following exposure 3.4% glutaraldehyde 10 hours at 25 o C 90 mins at 25 o C Requires activation Thoroughly rinse following exposure www.fda.gov

Key Points with Disinfection Label claims can vary Safety, preparation, contact time, numbers or rinsing etc Request specifics from manufacturers (e.g., rinse thoroughly ) Single use or multiple use Use of solution test strips to verify minimum recommended concentration (MRC) Multiple use disinfectants Closely follow label claims, including maximum reuse life NO topping off All surfaces of the device should be in contact Rinsing Correct water quality (bacteria-free; AAMI TIR34) Fresh water for every rinse (by immersion) Correct number of rinses

High Level Disinfection Steps Preparation (if applicable, e.g., activation) Immersion/Flow (ensure all parts/lumens exposed) Remove Device, Purge all Lumens Rinsing (each rinse timed and repeated as necessary in fresh water)

Manual and Automated Cleaning / Reprocessing

Automated Endoscope Reprocessors Review claims and instructions for use carefully FDA clearance Note limitations Provide More efficient process Less user exposure to chemicals Ensure standardized results Repeat cycle if interrupted Correct use of cleaning detergents and/or disinfectants Adequate ventilation Rinse water control (filtered) Routine maintenance

Liquid Chemical Sterilization One system cleared (in the USA) as a liquid chemical sterilant processing system for heat sensitive devices Liquid chemical sterilization with a peracetic acid sterilant process Controlled rinsing with extensively treated water Removal of bacteria, viruses, fungi, protozoa Quick cycle time After processing, devices should be used immediately

Endoscope Drying Essential to prevent bacterial/fungal growth or biofilm development over time Hanging to dry or drip dry is NOT effective Compressed air Most AERs purge water from the endoscope lumens do not dry Air pressure and quality of air through channels Alcohol flush facilitates drying Dry all removable parts and keep with endoscope

Sterilization Essentials Sterilization is dependant on adequate cleaning, rinsing and device preparation Drying may also be essential Packaging considerations (if applicable) Modality claims are product specific, controlled processes Steam (AAMI ST 79) Ethylene oxide (AAMI ST 41) Vaporized Hydrogen peroxide (AAMI ST 58) Ozone (AAMI ST 58) Correct equipment installation, maintenance

Terminal Sterilization Processes for flexible and semi-rigid endoscopes Recommended for devices entering sterile body cavities Required for all endoscope accessories that penetrate mucosa Biopsy forceps, sphincterotomes, etc. Claims are product specific Controlled processes Monitoring required for all systems

Ethylene Oxide Devices are clean and dry ETO is sensitive to the presence of residual soil Low pressure (vacuum) systems Venting cap required Sterilization parameters validated by endoscope manufacturer Conditioning, sterilization and aeration Post-sterilization aeration is essential Processing time typically >18 hours May have a limited number of cycles before requiring extensive repair

Hydrogen Peroxide Gas Processes with and without plasma Vacuum processes require device venting Packaging requirements Claims (lumen length and diameter restrictions) product specific Generally critical flexible endoscopes Some systems include claims for single/multiple lumen devices, with or without an additional load and bronchoscopes No GI endoscope claims Typical sterilization time ~30 mins Devices clean and dry before sterilization

Protective Microbial Barriers Sheaths are designed as an accessory for certain types of endoscopes and vary in design Two categories Intended to reduce the level of soiling Intended to prevent endoscope soiling Follow instructions for use carefully Requirement for periodic reprocessing Application and removal Inspection (e.g., if compromised)

Storage General considerations Dry prior to storage Prevent coiling or kinking (hanging preferred) Well ventilated, clean area No risk of contamination Angulation locks in free position All removable parts detached, kept with endoscope Label or tag affixed to identify endoscope has been processed Tracking and traceability

Storage Liquid chemical sterilization process Drying with cleaned pressurized air Reduce risks of contamination Follow manufacturer IFUs Gaseous sterilization sterilized endoscopes stored in container in which they were sterilized Sterile storage conditions (AAMI ST 58) Hang time Perform risk assessment (facility specific) Develop policy and procedure

Risk Assessment for Hang Time Perform assessment to determine length of storage time Consider: Complexity of device Condition after processing Transportation methods Conditions of storage environment Handling during storage Manufacturer IFUs Professional organization guidelines Current research studies Develop policy and procedure

Storage, Terminal Sterilization Storage conditions are monitored according to ANSI/AAMI ST79 Endoscopes are rotated according to policy Endoscopes are identified and labeled

Quality Control Plan Policies follow standards and guidelines, manufacturer IFUs May conflict or be periodically updated Reprocessing policy Facility Reprocessing area design Essential requirements Department Detailed work instructions Safety and essential steps Recalls Staff training and demonstrated competency Periodic auditing of practices Documentation

Quality Control Plan Verification points* Physical: exposure times, temperature, volumes, shelf-life, number of uses, etc Chemical: test strips and chemical indicators Biological (when appropriate): spore test strips and biological indicators Standards recommend Monitoring every device/load Routine (equipment) testing Following installation or major repair Periodic (load) testing *Specific for each product/process

Quality Control Plan, continued Cleaning Preparation (dilution, temperature) and use (contact time) Visual and another method (e.g., protein, ATP detection) Disinfection Preparation and use (contact time, reuse criteria) Solution test strips (indicators) Sterilization Pressure, temperature, time Chemical, biological and spore test indicators Documentation

Quality Control Program Cleaning Verification Retained soil causes patient infections! Manufacturers required to validate cleaning instructions Visual inspection combined with other verification methods Biochemical testing Cleaning efficacy of mechanical equipment Monitoring cleaning parameters (e.g., temperature) Repeat cleaning if soil present

Cleaning Verification ATP an energy molecule found in all living things Test Enzyme Solution + ATP = Light Luminometer measure Relative Light Unit output Higher RLU= Higher ATP Hemoglobin Tests for blood residues Too specific and can be too sensitive TYPES OF TESTS Carbohydrate Tests for sugar and starches Currently in combination with protein and hemoglobin testing in one test Protein Tests for protein residues

Quality Control Plan Microbiological Surveillance Testing Limited information for guidance of surveillance cultures Surveillance cultures to assess adequacy of reprocessing Culture results delayed 2-3 days Quarantine endoscopes/accessories/aer? Is there a cutoff to define proper disinfection? What endoscopes are cultured/sampling? Areas cultured? Protocol for culturing correctly Real time testing

Risk Analysis and Mitigation Stage 1: Risk identification Stage 2: Risk assessment Stage 3: Risk mitigation Stage 4: Maintenance (quality control) and Quality Process Improvement

Risk Identification: Examples

Action Items Obtain copies of applicable standards and guidelines for flexible endoscope reprocessing Understand ANSI/AAMI ST91, a comprehensive guide to flexible and semi-rigid endoscope reprocessing in healthcare facilities Develop appropriate policy and procedures the equipment used.

Questions

university.steris.com Go to: http://university.steris.com Playing a part in your professional development today To help you achieve your career vision for tomorrow

Useful References AAMI/ANSI ST 58. Chemical sterilization and high-level disinfection. 3rd edition (2013) AAMI/ANSI ST91. Flexible and semi-rigid endoscope processing in health care facilities (2015) AAMI/ANSI ST 41. Ethylene oxide sterilization 4th edition (2010) AAMI TIR34 Water for the reprocessing of medical devices (2014) ASGE/SHEA/SGNA/APIC: Multi-society guideline on reprocessing flexible gastrointestinal endoscopes (2011) Kovaleva et al (2014). Transmission of infection by flexible gastrointestinal endoscopy and bronchoscopy. Clin Micro Rev 26: 231-254. McDonnell & Sheard (2012). A Practical guide to decontamination in healthcare. Wiley-Blackwell.