Analysis of Semivolatiles by GC/FID using the Ultra Inert Inlet Liners with Wool

Similar documents
Agilent Consumables Improve Performance with Shimadzu GC/MS

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water

Organohalide Pesticides in Water by GC/µECD with Agilent J&W DB-CLP1 and DB-CLP2

Semivolatile Analysis with Specially Designed Agilent J&W DB-UI 8270D Columns

Analysis of EPA Method 8270D Semivolatiles Using the Agilent J&W DB-UI 8270D Ultra Inert GC Column

Improving GC-MS Method Robustness and Cycle Times Using Capillary Flow Technology and Backflushing

Evaluating Inert Flow Path Components and Entire Flow Path for GC/MS/MS Pesticide Analysis

Contract Laboratory Program (CLP) Pesticide Analysis with 0.18 mm ID High-Efficiency GC Columns Utilizing Helium Carrier Gas. Application.

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph

Drinking Water Semivolatiles Analysis using the 6890N/5975B inert GC/MSD. Application. Author. Abstract. Instrument Operating Parameters.

A Direct Column-Performance Comparison for Rapid Contract Laboratory Program (CLP) Pesticide Analysis. Application. Authors. Abstract.

Methods for Increased Sample Throughput with Dirty Samples

Reduced carryover and increased reproducibility in the analysis of cold pressed pink grapefruit essential oil

Prefractionator for Reliable Analysis of the Light Ends of Crude Oil and other Petroleum Fractions

Tips on Making your GC System and Analysis more Robust. Agilent Restricted

Direct Injection of Fish Oil for the GC-ECD Analysis of PCBs: Results Using a Deans Switch With Backflushing. Application. Author.

The Use Of Automated Backflush on the 7890A/5975A GC-MS System. Application. Setup Example Using Biomarkers. Authors. Abstract.

Capillary GC Troubleshooting: a Practical Approach

GC Troubleshooting. Frank Michel, Katherine K. Stenerson T HCE

Large Volume Injection of Organochlorine Pesticide Extracts Using a Programmable Temperature Vaporizer in the Solvent Split Mode

Ghost Peaks in Gas Chromatography Part 2: The Injection Port

Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application

Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D*

Agilent Ultimate Plus Fused Silica Tubing

Chlorinated Solvents and Disinfection By-Product Analysis Using Agilent J&W HP-1ms Ultra Inert and DB-1301 Capillary GC Columns

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools

Gas Chromatography Assignment Chem 543/443

Analysis of 16 Pesticide Residues in Broccoli Using CarbonX dspe QuEChERS AOAC Kits Using GC/MS

Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run

Optimizing Productivity and Reliability for Monocyclic Aromatic Hydrocarbon Purity Analysis According to ASTM D7504 on the Agilent 8890 GC System

Validated Multi-Residue Method for Extraction and Analysis of Trace-Level Pesticides in Surface Water Application

Application Note. Abstract. Authors. Environmental

MARKET FOCUSED - environmental ORGANOCHLORINE PESTICIDES

Olefin Analyzer Solutions for Polymer Grade Ethylene and Propylene

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE.

Nitrosamines Analysis in Drinking Water Using GC/MS/MS Meeting Equivalence to EPA Method 521

Liners and consumables for OPTIC Inlet. Liner Selection Guide

Improving Productivity in Your Environmental and Food Testing Labs with Agilent 7890A/5975C GC/MS System

EPA Method 524 for Determination of VOCs in Drinking Water Using Agilent 5975T LTM GC/MSD with Static Headspace

Analysis of Atrazine in Drinking Water at the ppb Level Using New Agilent Reversed Phase LC Columns. Application. Author. Abstract.

A New Horizon of Performance Agilent G1888 Network Headspace Sampler

Multiresidue Pesticide Analysis with the Agilent Intuvo 9000 GC and Agilent 7000 Series Mass Spectrometer

Sample Matrix Influence on GC/MS/MS Multiresidue Pesticide Analysis

ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER

Improved Gas Chromatograph Method for the Analysis of Trace Hydrocarbon Impurities in 1, 3-Butadiene

Determination of Asarinin in Xixin (Asari Radix Et Rhizoma)

Clarus SQ 8 GC/MS. The NEW

Interlaboratory Validation Study for the Analysis of Nitrosamines in Drinking Water using GC-MS/MS

Two Headspace Solutions for the Analysis of Dissolved Gases in Water by Method RSKSOP-175

Agilent 7693A Series Automatic Liquid Sampler Inject new performance into your gas chromatography. Our measure is your success.

CORESTA Recommended Method No. 57

Technical Article. Innovative Chromatography Solutions. The Rtx -CLPesticides Column Set Lets You Make the Choice

DETERMINATION OF ORGANIC PRIORITY POLLUTANTS AND EMERGING COMPOUNDS IN WASTEWATER BY MICRO SOLID PHASE EXTRACTION COUPLED TO GCMS

Environmental Volatiles using a Headspace-GC/MSD Analyzer

Standard Operation Procedure (SOP) for the Extraction and Instrument Analysis of Siloxanes from Cartridges

PerkinElmer Supplies/Parts. GC Accessories. Autosampler Syringes

2.1 This procedure provides steps and conditions for the determination of Vapor Space Organic Compounds in ground waters, and industrial waste waters.

Application Note. Abstract. Authors. Food Safety

Separation and Quantitation of a Mixture of Alcohols by Gas Chromatography

Chromatography Applications and Technology

Please enclose a item wise compliance report along with the technical bid.

Automated Solid Phase Extraction of PAH Compounds Utilizing the SPE-DEX 5000

MS Certified Vials. Pre-cleaned and certified vials for mass spectrometry

Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016

US EPA Method Atomx XYZ and Agilent 7890B GC/5977A MS

Benefits of running organic matrices using the Agilent 5100 ICP-OES fast, robust, high performance analysis

Productivity and accuracy enhancements

Determination of Volatile Organic Compounds (VOCs) in Various Drinking Water Sources by GC/MS

Application Note. Authors. Abstract. Introduction. Pharmaceutical

Varian Supplies/Parts. GC Accessories. Septum Nut. Merlin Microseal Kit

Analytical Method of Limit Test for Hexachlorobenzene. in Picloram TC and Method Validation Data*

Determination of 1,4-Dioxane in Drinking Water by Gas Chromatography/Mass Spectrometry (GC/MS) with Selected Ion Monitoring (SIM)

GC Method Development

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications

Confirmation of THC in Oral Fluids Using High-Resolution 2-D GC/MS Application Note

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

Low-level Detection of Ethanol, 1,4-Dioxane, and Other Oxygenates Using the Eclipse Purge-and-Trap Sample Concentrator

Gas Chromatography SSI-GC-003

CHROMATOGRAPHY/MASS SPECTROMETRY

Performance Characteristics of the Agilent 1220 Infinity Gradient LC system

ENVIRONMENTAL ANALYSIS. Solutions for Your Analytical Business Markets and Applications Programs. Solution Note

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Evaluation of the Mercury Concentration Accessory for US EPA Methodology

Application Note. Abstract. Authors. Environmental

Determination of Monocrotophos, Diazinon, Malathion, EPN, and Methamidaphos from Aqueous Samples Using Atlantic HLB SPE Disks

Content Page OPTIC Liners OPTIC 2 OPTIC 3 OPTIC-4 LINEX PHASER

Sampling Techniques Handbook. Agilent Automatic Liquid Samplers

FOOD PESTICIDE RESIDUES LABORATORY REFERENCE NO INSPECTOR DATE

SpeedExtractor E-916 Determination of Total Petroleum Hydrocarbons in Soil, Sediment, and Waste Samples using the SpeedExtractor E /2009

An Examination of the Presence, Formation, and Transformation of Volatile Halogenated Organic Species in Wastewater Extracts Using GC-ICP-MS

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

Quantitative Analysis of pesticides including Phenoxy-acids in surface and drinking water matrices using UPLC-ON-LINE-SPE-MS/MS

Autosampler Supplies for Agilent GC and HPLC Systems with CTC Autosamplers

Liner Style* Function Advantages Disadvantages Recommended For Straight Low surface area for less activity Least expensive Low activity

The Use of Deuterated Monitoring Compounds to Monitor Method Performance by the USEPA Contract Laboratory Program

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

Performance characteristics of the 1260 Infinity Quaternary LC system

Application Note. Abstract. Authors. Introduction. Environmental

CTC automated sample injectors for gas chromatography, distributed by Agilent Technologies

SEMI-VOLATILE ORGANICS IN DRINKING WATER using EPA Method optimized for greater sensitivity

Transcription:

Analysis of Semivolatiles by GC/FID using the Ultra Inert Inlet Liners with Wool Application Note Environmental Authors Limian Zhao and Bill Wilson Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808 USA David Mao Agilent Technologies, Inc. 91 Blue Ravine Road Folsom, CA 95630 USA Abstract With efficient deactivation on glass, the Ultra Inert liners with provide superior inertness for accurate analysis of semivolatile analytes by GC/FID, homogeneous sample mixing and evaporation, and maximum column and detector protection.

Introduction GC inlet liners are the centerpiece of the inlet system in which the sample is vaporized and mixed with the carrier gas, and then introduced to the capillary column. Inlet liners with have been used widely due to the benefits provided by glass. The inserted into liners can promote homogenous sample mixing and better quantitation. It provides a large surface area which aids the vaporization of liquid samples. It also acts as a trap to collect non-volatile residue in the sample, therefore protecting the GC column from the negative impacts of sample matrix. Liners with also prevent sample from hitting the bottom of the inlet before vaporization. However, the use of liners with has historically been greatly limited because of the high activity due to the large surface area and poor deactivations on glass, especially for the labile analytes, such as pesticides, active acidic and basic compounds. Traditional deactivation techniques usually cannot deactivate the glass surface area effectively. Active sites left on the can cause the degradation or adsorption of sensitive compounds in the liner before the analytes get to the column. This complicates quantitation, results in peak tailing or splitting, and can be particularly problematic for sensitive analytes. As a result, inlet liners with glass are usually not recommended for the analysis of active analytes. USEPA Method 8270 has been used widely to determine the concentration of semivolatile organic compounds in the environment. Samples using this method usually contain a mix of acids, bases, and neutrals that must be measured concurrently. This test is a challenge for GC/MS instrumentation due to the interaction of the analytes with the surface of sample flow path, where the inlet liner can be a significant contributor to system activity. The active sites on the liner surface can cause unwanted absorption of these compounds and lead to the loss of system response. The most active compounds, the nitrophenols, showed lower response factors (RFs) at the low concentrations causing poor calibration curve linearity and low analytes sensitivity. This effect can be magnified when using liners with glass. With the liners, the responses for the sensitive analytes at lower levels (< 20 ppm) may completely disappear, which leads to very poor calibration curve linearity and failure of the run. 2,4-dinitrophenol, 4-nitrophenol, 4,6-dinitro-2-methylphenol, 4-aminobiphenyl, pentachlorophenol, benzidine, 3,3-dichlorobenzidine, benzo[b]fluoranthene, benzo[k]fluoranthene. This test mix was used previously for the USEPA 8270 method improvements on the Agilent GC/MSD system. [1] All liner experiments were conducted using an FID to facilitate better comparisons because of the reproducibility of compound response over time. A calibration curve from 2 80 ng on column was used for linearity evaluation. In addition, the 10-day thermal stability at 330 C and 100 injections repeatability was investigated. Experimental Standards and Reagents EPA 8270 customized standard and internal standards were obtained from Ultra Scientific (North Kingstown, RI, USA). Ultra Resi-analyzed grade Methylene Chloride was from J. T. Baker (Phillipsburg, NJ, USA). Solutions and Standards The 8270 custom standard was purchased as 2000 µg/ml mixture in methylene chloride. Series calibration standards were prepared at 2, 5, 20, 40, and 80 µg/ml by appropriate dilution with methylene chloride. The 8270 semivolatile internal standard (IS) mixture at 4000 µg/ml in methylene chloride was spiked into standards with appropriate volume to generate constant 40 µg/ml IS concentration. Instrumentation All testing was done on an Agilent GC equipped with an Agilent 7683B autosampler and FID. Table 1 list the instrumental conditions used on this test and Table 2 lists flow path consumable supplies used in experiments. Table 1. Instrumental Conditions for Agilent GC/FID System used for Semivolatile Active Compounds Test Autosampler Agilent 7683B, 5 µl syringe (Agilent p/n 5181-5246), 1 µl injection volume. Preinj solvent A (methylene chloride) washes: 1 Sample pumps: 3 Postinj solvent B (methylene chloride) washes: 3 Agilent s Ultra Inert liner deactivation process significantly improves the efficiency and robustness of glass deactivation. The large glass surface area can be deactivated thoroughly. The Ultra Inert deactivation technique enables the use of Ultra Inert liners with for the analysis of sensitive semivolatile organic compounds using the EPA 8270 method. In this evaluation, the test mix includes representative difficult compounds in the 8270 method: N-nitrosodimethylamine, aniline, Carrier gas Inlet Analytical column Oven profile Detector Helium at 3 ml/min constant flow Splitless mode; 250 C, 30 ml/min purge flow at 1.5 min Ultra 2 column, 25 m 0.32 mm, 0.52 µm, (p/n 19091B-112) 40 C (1 min), 15 C/min to 310 C (0 min) FID, 250 C, H 2 /Air/Makeup N 2 : 40/450/45 ml/min 2

Table 2. Flow Path Supplies Vials Amber screw cap (Agilent p/n 5182-0716) Vial caps Blue screw cap (Agilent p/n 5182-0717) Vial inserts 150 µl glass w/ polymer feet (Agilent p/n 5183-2088) Septum Advanced Green Non-Stick 11 mm (Agilent p/n 5183-4759) Ferrules 0.5 mm id, 85/15 Vespel/graphite (Agilent p/n 5062-3514) O-rings Non-stick Flip-Top Liner O-ring (Agilent p/n 5188-5366) Inlet liners Agilent Ultra Inert deactivated single taper splitless liner (5190-2293) Inlet seal Gold plated inlet seal with washer (Agilent p/n 5188-5367) Results and Discussion The purpose of these tests is to evaluate the Ultra Inert liners with for the analysis of EPA 8270 semivolatile analytes. Although liners with can provide protection to the column and detector system from complicate environmental samples, they are usually not recommended for sensitive analytes analysis due to un-controlled activity of glass caused by large surface area and inefficient liner deactivation. The Agilent Ultra Inert deactivation process enables the application of Ultra Inert deactivated liners with for the analysis of sensitive semivolatile organic compounds. FID was used purposely as the detector to eliminate any activity contributed from the mass spectrometer. A previously established test mix [1] was used, which includes four phenols, several bases, and several neutral components. These compounds were selected to not only represent 8270 active analytes, but also to be resolved easily, and unambiguously detected by GC/FID. Figure 1 shows a sample chromatogram of the test mix. The feasibility of using Ultra Inert liners with was determined by the linearity of the practical calibration curve, the reproducibility among liners, the thermal stability at high temperature, and the precision of replicate injections. In parallel, competitor s equivalent liners with were tested for calibration curve linearity comparison. Calibration curve linearity and liner to liner reproducibility A linear calibration curve for target analytes is critical to achieve accurate and reliable quantitative analysis results. The EPA 8270 method does not specify a calibration range, yet traditionally a dynamic range of 20 to 160 ng on column has been widely used in USEPA Contract Lab Program (CLP). With the increased sensitivity of newer GC/MS systems, users are moving toward lower detection limits. Therefore, a calibration range of 2 to 80 ng on column was chosen for this test. The 2 ng on column quantitation limit gave satisfactory response and peak shape on FID with S/N ratio greater than 20. In order to evaluate the compounds linearity over the calibration curve range, the response factors (RFs) at each calibration level were calculated as Equation 1. The overall RF values across the curve were used to calculate the relative standard deviation (RSD). Peak Area Analyte Concentration Internal Standard RF = Peak Area Internal Standard Concentration Analyte (Equation 1) The higher the RF values, the higher the analyte responses and better peak intensity; and the lower the RSD value across the calibration range, the better the calibration curve linearity. According to USEPA 8270 method requirement [2], the RSD for each target analyte should be less than 20%. As an FID was used as the detector instead of a MSD, RFs are not identical for MSD and FID due to inherent response differences. Figure 2 shows the average FID RF values across the calibration range of 2-80 ng on column and RSDs for the selected representative active 8270 analytes. Each of the bars shows the average RF of seven tested liners for the corresponding analyte at each level as described in the legend in Figure 2. The order is 2 ng at the top increasing to 80 ng at the bottom. Above each bar is listed the average RF and RSD values across the calibration range of 2 80 ng on column. 3,3-Dichlorobenzidine data is not available due to its complete co-elution with IS peak 5, chrysene-d12 (Figure 1). It has to be mentioned that longer splitless time (1.5 min) with 30 ml/min of purge flow was used for the test. This longer splitless time allows a more complete transfer of target analytes onto the column at a fixed flow. Figure 1. A sample chromatogram of 8270 short test mix with peaks identification. 3

1.611 Figure 2. The 8270 representative active analytes RFs (FID) over calibration range (2-80 ng on column) using Agilent Ultra Inert deactivated liners with (p/n 5190-2293). (# liners = 7) All compounds meet the EPA method requirement with less than 15% RSD, including the most problematic compound, 2,4-DNP which had an average RF value of 0.277 with 13.1% RSD, across seven of the Ultra Inert liners. The liner to liner performance is reproducible with excellent consistency. Table 3. 2,4-DNP Average RF (on FID) and RSD Values for Seven Replicates of Agilent Ultra Inert Deactivated Liners with Wool (p/n 5190-2293) from Four Different Lots Average RF over Liner Lot 2 80 ng on column RSD (%) UI liner 1 0.271 14.7 Lot 1 UI liner 2 0.266 13.8 UI liner 3 Lot 2 0.274 14.9 UI liner 4 0.279 10.3 UI liner 5 Lot 3 0.286 9.5 UI liner 6 0.283 11.8 UI liner 7 Lot 4 0.269 10.3 Table 3 shows the average RF and RSD values for seven replicates of Ultra Inert liners from four different lots. Multi-injection repeatability and thermal stability at high temperature Multi-injection repeatability was tested by continuously injecting 1 µl of 5 µg/ml standard samples for 100 injections. Data were collected and RF values were calculated after every 10 injections. Figure 3 shows repeatability over 100 injections with 5 ng on column. The results indicate that excellent repeatability can be achieved with Ultra Inert liners with, with less than 6% RSD RF values over 100 injections. Two calibration curves were run before and after 100 injections. The two curves match each other, and no curve diversion was observed for any target active compound. Figure 4 shows representative 2,4-DNP calibration curves before and after 100 injections. 4

RFs on FID 1.800 1.600 1.400 1.200 1.000 0.800 0.600 0.400 0.200 0.000 100 Injections Repeatability for Agilent Ultra Inert Deactivated Liner with Wool 5 10 20 30 40 50 60 70 80 90 100 Aniline (2.7%) 2,4-DNP (5.3%) 4-NP (3.3%) 4,6-DN-2MP (3.2%) 4-ABP (1.4%) PCP (2.0%) Benzidiene (2.3%) Figure 3. The 100 injections repeatability for Agilent Ultra Inert deactivated liner with (p/n 5190-2293), less than 6% RSD achieved for all of active 8270 analytes over 100 injections with 5 ng on column. Peak area ratio of 2,4-DNP to IS 0.7 0.6 0.5 0.4 0.3 2,4-DNP calibration curves before and after 100 injections using Agilent Ultra Inert deactivated liner with 1st curve 2nd curve 1st curve before 100 injections, y = 0.0075x 0.0071 R 2 = 0.9996 0.2 2nd curve after 100 injections, y = 0.0078x 0.0058 0.1 R 2 = 0.9999 0 0 10 20 30 40 50 60 70 80 Injection amount on column (ng) Figure 4. The 2,4-DNP calibration curve before and after 100 injections using Agilent Ultra Inert deactivated liner with (p/n 5190-2293). The thermal stability of Ultra Inert liners was tested by keeping the liner in the inlet at 330 C for 10 days. On each day, two 1 µl injections of 5 µg/ml standards were run, data were collected and RF values were calculated. Ten day RF variability for active analytes is shown in Figure 5. Results indicate that the Ultra Inert deactivated liner with is thermally stable at an inlet temperature of 330 C with RSD less than 7% for active analytes. RFs on FID 1.80 1.60 1.40 1.20 1.00 0.80 0.60 0.40 0.20 0.00 Thermal Stability of Agilent UI Liner with Wool at 330 C Day 1 Day 2 Day 3 Day 4 Day 5 Day 7 Day 8 Day 9 Day 10 Aniline (2.9%) 2,4-DNP (4.4%) 4-NP (3.1%) 4,6-DN-2MP (4.3%) 4-ABP (2.3%) PCP (1.5%) Benzidiene (6.4%) Figure 5. 10-day thermal stability at 330 C for Agilent Ultra Inert deactivated liner with (p/n 5190-2293), less than 7% RSD achieved for all of active 8270 analytes over 10-day test at 5ng on column. The results of the multiple injections and thermal stability test demonstrate the excellent stability of Ultra Inert deactivated liner with allowing for more samples to be run over a longer time period, and reducing downtime for maintenance. With better protection to the analytical column and detector, the Ultra Inert liner with supports the longevity of the whole system for quantitative analysis. Performance comparison against competitors equivalent liners The performance comparison was conducted by comparing the Ultra Inert liners with with similar competitors liners and equivalent configuration, including Restek Siltek, Restek IP SemiVolatile, and Restek Sky liners. To show the liner performance differences, besides 2,4-DNP, several other compounds were compared in detail including 4,6-dinitro-2-methylphenol, 4-aminobiphenyl, pentachlorophenol, and benzidine. These compounds are problematic compounds mentioned in EPA 8270 method [2]. 5

As shown in Table 4, the Ultra Inert liners with usually provide the lower RSD and higher average RF values through the calibration range, indicating the excellent linearity of calibration range and better system sensitivity. The Ultra Inert liners perform much better than Restek Siltek deactivated liner, and equivalent or slightly better than Restek IP deactivated SemiVolatile and Restek Sky liner. Comparisons between UI deactivated liner with and liner without It was demonstrated previously that the single taper splitless liners without showed excellent performance for the active semi-volatile compounds analysis.[1,3] The direct connect liners has been recommended for clean samples also with minimal inlet activity.[1] However, use of these liners potentially subjects the column to more degradation and the detector to more contamination from dirty samples. The addition of in the liner can efficiently protect the column and detector (MS source) by trapping the high boiling material and other interferences from sample matrix. However, the inertness of liners is always a concern due to the large surface area of glass and inefficient deactivation on. The Agilent Ultra Inert liner deactivation technique provides efficient and robust deactivation on the glass liner body as well as glass, enabling the use of liners for active compounds analysis. Figure 6 shows parallel comparison results between Ultra Inert liners with and liners without but with the same liner configuration (single taper, splitless). The results demonstrate that the Ultra Inert liners with provide equivalent performance as liners without to support quantitative analysis of active semivolatile analytes. With other benefits provided by liners such as homogeneous sample mixing and efficient liquid vaporization, and protection to column and MS source, Agilent Ultra Inert liners with can be the best choice for samples with complicated matrices. Table 4. Liners Performance Comparison with Competitors Equivalent Liners for Calibration Range of 2-80 ng on Column Using GC/FID Wool Liners Compounds Agilent Ultra Inert Deact. (n=7) Restek Siltek Deact. (n=4) Restek IP Deact. SV (n=4) Restek Sky (n=3) 2,4-DNP 4,6-DN-2-MP 4-ABP PCP Benzidine Average RF 0.277 0.230 0.270 0.285 RSD (%) 13.1 29.7 15.7 14.6 Average RF 0.370 0.342 0.375 0.377 RSD (%) 5.2 14.6 6.8 7.6 Average RF 0.906 0.768 0.925 0.904 RSD (%) 3.1 20.1 5.8 1.0 Average RF 0.301 0.320 0.323 0.300 RSD (%) 9.1 13.7 15.3 6.9 Average RF 0.616 0.458 0.619 0.594 RSD (%) 5.9 25.1 9.1 10.1 6

Performance comparison of Ultra Inert liners with and without 9.000 8.000 1.527, 3.6 1.513, 1.1 7.000 Average RF, RSD (%) 6.000 RFs on FID 5.000 4.000 3.000 2.000 0.277, 13.1 0.288, 15.6 0.465, 0.8 0.457, 3.9 0.370, 5.2 0.382, 8.5 0.906, 3.1 0.913, 1.1 0.301, 9.1 0.299, 6.1 0.616, 5.9 0.619, 4.8 2 5 20 40 1.000 80 0.000 w/ w/o w/ w/o w/ w/o w/ w/o w/ w/o w/ w/o w/ w/o Aniline 2,4-DNP 4-NP 4,6-DN-2MP 4-ABP PCP Benzidine Figure 6. Performance comparison of Agilent Ultra Inert deactivated liner with (p/n 5190-2293) and Ultra Inert deactivated liner without (p/n 5190-2292). Conclusion Agilent Ultra Inert liners with have shown excellent inlet inertness to support the quantitative analysis of EPA 8270 semi-volatile active analytes. The linearity of the calibration curve meets the EPA method requirement of <20% RSD for all of active analytes even down to 2 ng on column. The thermal stability at 330 C of the deactivated liner was evaluated with great consistency over 10 days. Multi-injection repeatability over 100 injections for all active compounds was excellent with less than 6% RSD. With efficient and robust deactivation of the, Agilent Ultra Inert liners with provide excellent inertness for accurate quantitative analysis of semivolatile active analytes. Because liners with provide other benefits like homogeneous sample mixing and evaporation, non-volatile residue trapping, and column and detector protection, Ultra Inert liners with are the best choice for the analysis of active compounds in dirty sample matrix. References 1. M. Szelewski, B. Wilson, P. Perkins, Improvements in the Agilent 6890/5973 GC/MSD system for Use with USEPA Method 8270. Agilent Technologies publication 5988-3072EN. 2. USEPA 8270D method, http://www.epa.gov/osw/hazard/testmethods/sw846/pd fs/8270d.pdf 3. L. Zhao, A. Broske, D. Mao, Evaluation the New Agilent Ultra Inert No Wool Liners for Active Compounds Analysis by GC. Agilent Technologies pubilication 5990-7380EN. For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. 7

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 2011, 2017 Printed in the USA April 24, 2017 5990-7381EN