Multi-element determination in pharmaceutical preparations using the Thermo Scientific icap Q ICP-MS

Similar documents
Analysis of elemental impurities in drug products using the Thermo Scientific icap 7600 ICP-OES Duo

Simple, fast and reliable analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS

Analysis of elemental impurities in drug products using the Thermo Scientific icap 7400 ICP-OES Duo

EU Water Analysis Using the Thermo Scientific icap 7400 ICP-OES Duo

USP Chapters <232> and <233> Implementation Strategy. Horacio Pappa, Ph.D. Director - General Chapters U.S. Pharmacopeia

Dr. Jianfeng Cui, Thermo Fisher Scientific, Bremen, Germany APPLICATION NOTE AN43384

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES

Analysis of Trace Elements in Seawater Using the Thermo Scientific icap 7000 Series ICP-OES Duo

Agilent Resources for Workflow Compliance MEASURING ELEMENTAL IMPURITIES IN PHARMACEUTICAL MATERIALS

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

Functionality. Normal Operation

Analytically Monitoring the Effect of Fracturing Activity

CARBOWAX SENTRY Polyethylene Glycol (PEG) and Methoxypolyethylene Glycol (MPEG) Products

Collaborative Efforts for Creating a Robust Trace Metals Analytical Procedure for Flue Gas Desulfurization Wastewaters by ICP-MS

Automating EPA 6020 Compliant Analysis with the Agilent 7900 ICP-MS and ESI prepfast Autodilution System

Thermo Scientific icap 7000 Plus Series ICP-OES. Gain more power. experience more performance

Analysis of Pharmaceuticals by ICP-MS and WDXRF

Analysis of Toxic Trace Elements in Coffee Bean Products by HR ICP-OES

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

Increased Sample Throughput for ICP-OES Applied to U.S. EPA Method 200.7

EU water analysis using the Thermo Scientific icap 7400 ICP-OES Duo

Amplifying ICPMS Productivity Using Discrete Sampling Technology

High matrix tolerance of the Thermo Scientific icap 7000 Plus Series ICP-OES with the radial Ceramic D-Torch and the sheath gas adaptor

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

The image part with relationship ID rid4 was not found in the file. Welcome

Impact of New ICH Q3D and USP <232> Guidelines for Elemental Impurities Analyses

Analysis of Mineral and Heavy Metal Content in Beverages Using the Teledyne Leeman Labs Prodigy Plus ICP-OES

Elemental Impurities / Metal Impurities AFFINISOL HPMCAS (Hydroxypropyl Methylcellulose Acetate Succinate)

Direct Analysis of Photoresist by ICP-MS. Featuring the Agilent Technologies 7500s ICP-MS

Composition Analysis of Animal Feed by HR ICP-OES

Analysis of steel and its alloys using the GB/T standard and an Agilent 5100 ICP-OES in dual view mode

Characterization of Surface Metals on Silicon Wafers by SME-ICP-MS. Featuring the Agilent Technologies 7500s ICP-MS

Elementverunreinigungen in Pharmaprodukten Neue Limits und Verfahren zur Bestimmung nach USP

Comparing an instrumental technique based on its

TargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet

DETERMINATION OF ELEMENTS IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS 10500, & USING THE AGILENT 5100 ICP-OES

Application Note. Abstract. Authors. Environmental

Analysis of inorganic impurities in lubricating oils by ICP-MS

The Determination of Trace Elements in Waters using the US EPA Method and the Thermo Scientific icap 6500 ICP-OES (Duo)

Thermo Scientific Qtegra Intelligent Scientific Data Solution. Delivering quality. Driving productivity

US EPA Method using the XSERIES 2 ICP-MS

High Purity Acids Trace Elemental Analysis. Detect as low as 1 to 100 ppt

SECTIOn 4. Claritas PPT Multi-Element Standards for ICP-MS

Table I: MCL and MRL Concentrations for Contaminants Monitored Under the Safe Drinking Water Act National Primary Drinking Water Regulations

Analysis of Heavy Metals in Iron-based Fertilizers by HR ICP-OES

Multi-Element Analysis of Trace Metals in Animal Feed using ICP-OES

icap 7000 Plus Series ICP-OES Streamlined performance and ultra-low ICP-OES detection limits for routine laboratories

Determination of Elemental Impurities in Graphite-based Anodes using the Agilent 5110 ICP-OES

ICP-AES Determination of 15 Kind of impurity elements in the Vanadium-aluminum alloy

Elemental Impurities Regulations View from a CRO

Analysis of Environmental Samples with the Agilent 730-ES Following US EPA Guidelines

Agilent Solutions for achieving 21 CFR 11 Compliance on Atomic Spectroscopy Instrumentation

Ultra-fast ICP-OES determinations of base metals in geochemical samples using next generation sample introduction technology

Class 1 Elements should be essentially absent Known or strongly suspected human toxicants Environmental hazards

Selenium in Natural Waters by Graphite Furnace Atomic Absorption using EPA Method 200.9

Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass

Assaying and Analysis Choices in Practice

Thermo Scientific icap Q ICP-MS

ESI SC-FAST Sample Introduction Solutions for PerkinElmer ICP-OES and ICP-MS Systems

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Analysis of Fracking Flowback Water From the Marcellus Shale Using In-line Conductivity, Automated Dilution, and Ion Chromatography

The Determination of Alloying Elements in High-Carbon Steel using Radially Viewed Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES)

The Agilent Atomic Spectroscopy portfolio for Environmental applications AA, MP-AES, ICP-OES, ICP-MS & ICP-QQQ

Trace Metals in Waters by GFAAS, in Accordance with U.S. EPA and Health Canada Requirements

Thermo Scientific icap RQ ICP-MS. Simplicity, productivity and robustness for routine labs

Correlation coefficient

Analysis of Stainless Steel by Dual View Inductively Coupled Plasma Spectrometry

Isotope Ratio MS. Continuous Flow. Consumables and Spare Parts for Thermo Scientific Elemental Analyzers

Nitrogen Determination in Soils and Plants by Flash Combustion using Argon as Carrier Gas

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

Thermo Scientific icap 7000 Series ICP-OES Accessories Guide

The Benefits of SOLAμ Technology in Sample Preparation. Jon Bardsley and Ken Meadows Thermo Fisher Scientific, Runcorn, UK

CERTIFICATE OF ANALYSIS

Trace Elements in Petrochemical Samples: Achieving Excellent Analytical Results

Institute for Reference Materials and Measurements Community Reference Laboratory for Feed Additives

á232ñ ELEMENTAL IMPURITIES LIMITS

Selenium in Natural Waters by Graphite Furnace Atomic Absorption using EPA Method 200.9

The Determination of Toxic Metals in Waters and Wastes by Furnace Atomic Absorption

Why Q3D? Elemental Impurities: Is Your Company Prepared for ICH Q3D

Merits of EPA Method 1640:

Jonathan R. Beck and Charles T. Yang; Thermo Fisher Scientific, San Jose, CA

Determination of trace and ultratrace elements by ICP-SFMS

Implementation of a Walk-Up High- Pressure Capillary Ion Chromatograph for the Fast Separation of Pharmaceutical Relevant Inorganic Anions and Cations

CERTIFICATE OF ANALYSIS REVISED REPORT

Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease

for water and beverage analysis

Thermo Scientific EASY-nLC 1200 System. Leading in simplicity. and performance

Determination of 22 Elements Following US EPA Guidelines with a New Megapixel CCD ICP-OES

Analysis of Stainless Steel by Dual View Inductively Coupled Plasma Spectrometry

CERTIFICATE OF ANALYSIS 1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION

ElvaX Light for Petrochemical analysis

High-Purity Acids & Reagents

Analysis of Triazine Herbicides in Drinking Water Using LC-MS/MS and TraceFinder Software

High-Purity Acids & Reagents

Analysis of Triazine Herbicides in Drinking Water Using LC-MS/MS and TraceFinder Software

Thermo Scientific icap 7000 Plus Series ICP-OES. Gain more power. experience more performance

Analysis of Trace Metals in Surface and Bottled Water with the Optima 7300 DV ICP-OES

Determination of 1,4-Dioxane in Drinking Water by Gas Chromatography/Mass Spectrometry (GC/MS) with Selected Ion Monitoring (SIM)

Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease

Transcription:

Multi-element determination in pharmaceutical preparations using the Thermo Scientific icap Q ICP-MS Dr. Simon Nelms, Thermo Fisher Scientific, UK Application Note 43100 Key Words icap Q, Pharmaceutical preparations, ICP-MS, USP 232, USP 233 Goal To demonstrate the capability of the icap Q ICP-MS to detect and quantify metal impurities in medicinal preparations according to the proposed USP 232/233 methods. Introduction Monitoring and control of metal impurities in medicinal preparations is of importance to the pharmaceutical industry as drug production and formulation processes often involve either direct addition of metals (as catalysts) or non-intentional addition via contaminated reagents or contact of the pharmaceutical ingredients with metal surfaces during production. Historically, qualitative methods based on subjective tests such as visually detecting metal sulphides via colorimetry (the US Pharmacopeia 231 method (USP 231)) were used, but these have been recognized to be inaccurate and insufficiently comprehensive in terms of detecting all metals of concern. In August 2008, at a workshop organized by and conducted at the Institute of Medicine (IOM) of the US National Academy of Sciences, plans were put in place to develop improved methods and to harmonize analytical approaches to these measurements. The result was a revised impurities list (USP 232) and a new analysis procedure (USP 233), that described the use of ICP-OES and ICP-MS for determining metal impurities in pharmaceutical products. Following comments and feedback from the pharmaceutical industry, both USP 232 and USP 233 were further revised in 2011 to improve their clarity and flexibility and these documents are now in the process of being officially implemented. and electronic signatures that persons create, modify, maintain, archive, retrieve or transmit. This means that the control software of analytical instruments used in pharmaceutical production must include tools to maintain the integrity of the analysis method and results, allow audit trailing and electronic signatures as well as provide security features to ensure that alterations cannot be made without clear indication of what has been changed, who changed it and why. This note describes the application of a new quadrupole ICP-MS, the Thermo Scientific icap Q, to the detection and quantification of the 16 target elements specified in USP 232 in accordance with the ICP-MS procedures described in USP 233. In order to comply with the requirements described above for 21 CFR Part 11 compliance, the Thermo Scientific Qtegra software suite of the icap Q ICP-MS has been specifically developed to provide comprehensive features for the pharmaceutical industry, such as audit trails, support for electronic signatures and tools for integrated data management. In addition to the requirements described in the USP documents, any method used for the analysis of pharmaceuticals must also comply with the US Food and Drug Administration s 21 CFR Part 11 regulations regarding electronic records and validation of electronic signatures. These regulations are concerned with ensuring the integrity and authenticity of any electronic records

2 and calibration solution preparation Three pharmaceutical samples were selected for this work, namely: A - Soluble aspirin product (in tablet form) B - Cold and flu remedy (in powder form) C - Children s cough syrup (in liquid form) For the purposes of this work, the Target Limit was taken as the limit per dose, based on 4 doses of the selected pharmaceuticals per day, as 4 was the maximum number of doses prescribed for each medicine per day. With this Target Limit taken into account, and as the samples were diluted by 1000x from the original sample, the calibration solutions were prepared at the concentrations given in Table 2. Each sample (5.00 ± 0.05 g) was dissolved into 500 ml of a diluent containing 1 % (v/v) HNO 3, 0.5 % (v/v) HCl and 200 ppb Au, prior to a further 1:10 dilution in the same diluent for analysis. The samples were measured using an external calibration approach against calibration solutions prepared in the same diluent as the samples. The calibration solutions contained all 16 of the elements listed in USP 232, Elemental Impurities - Limits. Internal standardization was used, with Ga, In and Tl as internal standards at 10, 5 and 5 ppb respectively, added on-line via a T-piece. Method analysis was carried out in accordance with the requirements described in USP 233, Elemental Impurities Procedures. This document specifies that the elements to be measured should be calibrated at a level of blank, 0.5 J and 2 J where J = the concentration (w/w) of the element(s) of interest at the Target limit, appropriately diluted to the working range of the instrument. For this work, a blank and three standards (at 0.5 J, J and 2 J) were measured. The Target limits (in the original, undiluted samples) for each element are presented in Table 1. Element Calibration Calibration Calibration solution 1 solution 2 solution 3 Arsenic 1.875 3.75 7.5 Cadmium 0.625 1.25 2.5 Lead 1.25 2.5 5 Mercury 1.875 3.75 7.5 Chromium 31.25 62.5 125 Copper 312.5 625 1250 Manganese 312.5 625 1250 Molybdenum 31.25 62.5 125 Nickel 31.25 62.5 125 Palladium 12.5 25 50 Platinum 12.5 25 50 Vanadium 31.25 62.5 125 Osmium 3.125 6.25 12.5 Rhodium 3.125 6.25 12.5 Ruthenium 3.125 6.25 12.5 Iridium 3.125 6.25 12.5 Element Daily Dose PDE* Target Limit (J) (µg/day) (per dose at 4 doses per day) (µg) Inorganic arsenic 15 3.75 Cadmium 5 1.25 Lead 10 2.5 Inorganic mercury 15 3.75 Chromium 250 62.5 Copper 2500 625 Manganese 2500 625 Molybdenum 250 62.5 Nickel 250 62.5 Palladium 100 25 Platinum 100 25 Vanadium 250 62.5 Osmium Rhodium 100 (Combination 100 (Combination not to exceed) not to exceed) Ruthenium Iridium * PDE = permitted daily exposure based on a 50 kg person Table 2: Calibration solution concentrations (ppb). USP 233 specifies that spike recoveries should be performed on each of the pharmaceutical preparations under test at the Target limit (J) and at 80 % of the Target limit (0.8 J). These spike tests are required to be carried out on a minimum of three separate samples of each preparation at both spike levels. Three samples of each of the three materials under test were therefore prepared at spike levels of 0.8 J and J accordingly, and analyzed against the calibration solutions. USP 233 also requires that the analysis be precision tested. For this, the protocol specifies that six independent samples of the material under test, spiked with the target elements at the indicated levels should be measured. As the protocol does not explicitly state what the indicated levels means, it was decided in this work to measure one of the materials (sample B the cold and flu remedy) at spike level J as described earlier. Finally, instrument detection limits for all the USP 232 target elements (based on 3x the standard deviation of the calibration blank) and method detection limits (based on 3x the standard deviation of the mean of 5 consecutive independent blanks) were determined. Table 1: Target Limit (J) for the elements specified in USP 232.

3 Instrument configuration The Thermo Scientific icap Qc ICP-MS was used for all measurements. The sample introduction system used consisted of the standard Peltier cooled, quartz cyclonic spray chamber, PFA concentric nebulizer and demountable quartz torch with a 2.5 mm ID quartz injector. Standard Ni sample and skimmer cones were also used. The instrument was operated in a single collision cell mode with kinetic energy discrimination (KED), using pure He as collision gas. All samples were presented for analysis using a SC4 DX autosampler from Elemental Scientific (Omaha, NE, USA). General analytical conditions Parameter Value Forward power 1550 W Nebulizer gas 0.89 L/min Auxiliary gas 0.8 L/min Cool gas 14.0 L/min Collision cell gas He at 7 ml/min uptake/wash time 45 s each Dwell times Optimized per analyte Number of points per peak 1 Number of repeats per sample 3 Total acquisition time 3.5 mins Table 3: Instrument operating parameters. Results and Discussion 1. Calibration performance. Linear calibrations with low (sub-ppb) blanks were obtained for all 16 of the elements specified in USP 232 as shown by the examples of V and 208 Pb in Figure 1 and by the calibration data presented in Table 4. (a) V. Isotope R BEC V 0.99994 0.012 52 Cr 0.99993 0.102 55 Mn 0.99993 0.102 60 Ni 0.99999 0.034 63 Cu 0.99997 0.070 75 As 0.99993 0.009 98 Mo 0.99983 0.088 102 Ru 0.99985 0.001 Rh 1.00000 0.001 106 Pd 0.99989 0.022 111 Cd 0.99968 0.003 189 Os 0.99998 0.008 Ir 0.99979 0.020 195 Pt 0.99991 0.001 202 Hg 0.99989 0.019 208 Pb 0.99996 0.003 Table 4: Calibration correlation coefficient (R) and BEC (ppb) data. Table 4 shows that for most of the target elements, low ppt BEC (background equivalent concentration) values were obtained. However, traces of Cr, Mn and Cu were detected, as a result of trace contamination of the blank reagents. The very low V and 75 As blanks obtained demonstrate the excellent collision cell performance of the icap Qc ICP-MS in removing the 35 Cl 16 O and 40 Ar 35 Cl interferences generated on V and 75 As respectively by the HCl used in this analysis. (b) 208 Pb. Figure 1: Example calibrations.

2. analysis results. The concentration determined for each target element in the pharmaceutical materials under investigation (three repeat analyses per sample) is shown in Table 5. The results shown here have been corrected for the 1000x sample dilution, and show the concentrations measured in the original 5g sample (in µg). V 52 55 60 Cr Mn Ni 63 75 Cu As 98 Mo 102 Ru Soluble n.d. 0.04 n.d. n.d. n.d. n.d n.d. n.d aspirin #1 Soluble n.d. 0.03 n.d. n.d. n.d. n.d. n.d. n.d. aspirin #2 Soluble n.d. 0.03 n.d. n.d. n.d. n.d. n.d. n.d. aspirin #3 Cold and 0.012 0.03 n.d. n.d. 0.01 n.d. n.d. n.d. flu remedy #1 Cold and 0.012 0.03 n.d. n.d. 0.02 n.d. n.d. n.d. flu remedy #2 Cold and 0.015 0.02 n.d. n.d. 0.01 n.d. n.d. n.d. flu remedy #3 Cough n.d. n.d. 1.00 n.d. 0.05 n.d. n.d. n.d. syrup #1 Cough n.d. n.d. 0.97 n.d. 0.05 n.d. n.d. n.d. syrup #2 Cough n.d. n.d. 1.01 n.d. 0.06 n.d. n.d. n.d. syrup #3 Rh 106 Pd 111 Cd 189 Os Ir 195 Pt 202 Hg 208 Pb Soluble n.d. n.d. n.d. n.d. n.d. n.d n.d. n.d aspirin #1 Soluble n.d. n.d. n.d. n.d. n.d. n.d. n.d. n.d. aspirin #2 Soluble n.d. n.d. n.d. n.d. n.d. n.d. n.d. n.d. aspirin #3 Cold and n.d. n.d. n.d. n.d. n.d. n.d. n.d. n.d. flu remedy #1 Cold and n.d. n.d. n.d. n.d. n.d. n.d. n.d. n.d. flu remedy #2 Cold and n.d. n.d. n.d. n.d. n.d. n.d. n.d. n.d. flu remedy #3 Cough n.d. n.d. n.d. n.d. n.d. n.d. n.d. 0.019 syrup #1 Cough n.d. n.d. n.d. n.d. n.d. n.d. n.d. 0.017 syrup #2 Cough n.d. n.d. n.d. n.d. n.d. n.d. n.d. 0.018 syrup #3 n.d. = not detected Table 5: analysis results (µg in the original 5g sample). Table 5 shows that, in all three samples, most of the target elements were either present at low concentrations or not detected at all (i.e. present at concentrations lower than the detection limit). The cough syrup however, was found to contain around 1 µg of Mn, 0.05 µg of Cu and almost 0.02 µg of Pb (in the 5 g sample dose), while 0.03 to 0.04 µg of Cr was detected in the soluble aspirin and cold / flu remedy samples (also in 5 g of sample). These levels, although easily detectable with the icap Qc ICP-MS, are far below the Target Limit values for these elements (i.e. Cr 62.5 µg, Mn 625 µg, Cu - 625 µg and Pb 2.5 µg per 5 g dose, with 4 doses per day for each of the three samples tested in this work). 3. Detection limits. The instrumental detection limits (I.d.L) for the target elements, calculated for the original undiluted sample are presented in Table 6. These limits are calculated from 3x the standard deviation of the calibration blank. 52 55 60 63 75 98 102 Element V Cr Mn Ni Cu As Mo Ru I.d.L 0.0006 0.009 0.01 0.01 0.01 0.003 0.01 0.0005 106 111 189 195 202 208 Element Rh Pd Cd Os Ir Pt Hg Pb I.d.L 0.0001 0.0007 0.001 0.002 0.003 0.001 0.003 0.0008 Table 6: Instrument detection limits (in µg, relative to the original 5 g sample). Table 6 shows that the icap Qc ICP-MS provides exceptionally low detection capability for all 16 of the USP 232 target elements. These results give an indication of the instrumental detection limits achievable with the instrument but, as such, are not indicative of what would be practically achieved on a routine basis. To determine the detection limits that would be routinely achievable for this analysis, method detection limits have to be calculated. Therefore, method detection limits based on 5 separate consecutive blanks were measured. These were calculated from 3x the standard deviation of the mean of the 5 blanks and are presented in Table 7. Isotope V 52 Cr 55 Mn 60 Ni 63 Cu 75 As 98 Mo 102 Ru M.d.L 0.011 0.01 0.03 0.03 0.01 0.01 0.07 0.002 Isotope Rh 106 Pd 111 Cd 189 Os Ir 195 Pt 202 Hg 208 Pb M.d.L 0.005 0.03 0.001 0.01 0.07 0.01 0.03 0.007 Table 7: Method detection limits (in µg, relative to the original 5 g sample). The method detection limits shown in Table 7 provide a robust indication of the icap Qc s performance for the USP 233 method as they allow for blank to blank variation, caused by random contamination in the sample tubes or during sample preparation. As shown in this Table, sub-µg method detection limits relative to the original, undiluted samples were obtained for all 16 target elements, demonstrating that the icap Qc far exceeds the performance required to meet the analytical criteria required by USP 233. 4

5 4. Spike recoveries. The spike recoveries for each repeat of all three samples at the 0.8 J and J spike levels are shown in Figures 2 and 3. USP 233 states that the acceptance criteria for this test are recoveries of between 70 and 150 % for the mean of the three repeat analyses of each sample at both spike levels. Figures 2 and 3 show that these criteria (indicated by the red lines) are easily achieved using the icap Qc ICP-MS, with average recoveries at both spike levels ranging from 91 to 113 %. Figure 2: Recoveries (in %) at the 0.8 J spike level Figure 3: Recoveries (in %) at the J spike level. 5. Repeatability. Results for the analysis of 6 independent aliquots of sample B (the cold / flu remedy) spiked at the Target Limit, J, with the 16 USP 232 elements are shown in Table 8 below. V 52 Cr 55 Mn 60 Ni 63 Cu 75 As 98 Mo 102 Ru Cold / flu 65.2 63.6 642 62.6 627 4.03 63.9 6.11 remedy run 1 Cold / flu 64.8 63.3 636 61.5 613 4.04 64.1 6.28 remedy run 2 Cold / flu 65.8 64.1 6 62.3 620 4.10 66.7 6.31 remedy run 3 Cold / flu 66.3 64.8 649 62.9 627 4.09 66.2 6.46 remedy run 4 Cold / flu 66.4 65.1 656 63.1 634 4.07 65.0 6.39 remedy run 5 Cold / flu 66.0 64.7 655 62.2 623 4.01 63.5 6.21 remedy run 6 Mean 65.8 64.3 648 62.4 624 4.06 64.9 6.29 Std. Dev. 0.62 0.70 8 0.56 7.06 0.03 1.32 0.12 %RSD 0.9 1.1 1.2 0.9 1.1 0.8 2.0 2.0 Rh 106 111 189 Pd Cd Os Ir 195 202 208 Pt Hg Pb Cold / flu 6.07 23.9 1.24 6.40 6.41 25.6 3.66 2.49 remedy run 1 Cold / flu 6.13 24.2 1.27 6.17 6.39 25.3 3.68 2.43 remedy run 2 Cold / flu 6.30 25.1 1.18 6.19 6.45 25.5 3.75 2.46 remedy run 3 Cold / flu 6.35 25.5 1.27 6.21 6.48 26.2 3.77 2.43 remedy run 4 Cold / flu 6.24 24.6 1.22 6.10 6.33 25.2 3.71 2.37 remedy run 5 Cold / flu 6.04 24.2 1.22 6.27 6.54 26.0 3.78 2.48 remedy run 6 Mean 6.19 24.6 1.23 6.22 6.43 25.6 3.73 2.44 Std. Dev. 0.13 0.62 0.03 0.10 0.07 0.38 0.05 0.05 %RSD 2.0 2.5 2.8 1.6 1.1 1.5 1.3 1.9 Table 8: Repeatability test results for the cold / flu remedy sample (spiked at the Target Limit, J). All concentrations are shown in µg. Table 8 shows that excellent repeatability (< 3 % RSD for all 16 elements and < 2 % RSD for 14 of them) for 6 separate analyses of the spiked cold / flu remedy sample was obtained, illustrating the robustness and reliability of the tested method using the icap Qc ICP-MS.

6. Internal standard stability. The robustness of the icap Qc ICP-MS in meeting the USP 232 and 233 requirements was assessed by monitoring the absolute suppression and relative drift of the internal standards throughout the analysis. The response (in %) of all three internal standards relative to the initial calibration blank for all blanks, standards and samples analyzed in this work is shown in Figure 4. Application Note 43100 Figure 4: Internal standard response throughout the analysis. Figure 4 shows that no drift was observed throughout the duration of the run (2.75 hours), with the final sample giving internal standard responses of 106 %, 101 % and 97 % for 71 Ga, 115 In and 205 Tl respectively. The range in response of the internal standards was 89 % to 109 %, illustrating the robustness of the icap Q for this analysis. Conclusions. This application note has shown that the Thermo Scientific icap Qc ICP-MS is an ideal tool for elemental determination in pharmaceutical preparations. With its high sensitivity and robustness, the instrument is easily capable of accurately and precisely measuring all 16 of the specified elements in a single analysis mode at the Target Limits currently listed in USP 232 and in accordance with the analytical performance criteria described in USP 233. Finally, the range of security features, data management and audit trailing tools included in the advanced and flexible Qtegra software provide the necessary support to meet the demands of 21 CFR Part 11 compliance for the highly regulated pharmaceutical industry. Thermo Fisher Scientific (Bremen) GmbH Management System Registered to ISO 9001:2008 thermoscientific.com 2012 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organisation. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa-Other +27 11 570 1840 Australia +61 2 8844 9500 Austria +43 1 333 50 34 0 Belgium +32 53 73 42 41 Canada +1 800 530 8447 China +86 10 8419 3588 Denmark +45 70 23 62 60 Europe-Other +43 1 333 50 34 0 AN43100_E 03/12C Finland /Norway/Sweden +46 8 556 468 00 France +33 1 60 92 48 00 Germany +49 6 408 1014 India +91 22 6742 9434 Italy +39 02 950 591 Japan +81 45 453 9100 Latin America +1 608 276 5659 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 South Africa +27 11 570 1840 Spain +34 914 845 965 Switzerland +41 61 716 77 00 UK +44 1442 233555 USA +1 800 532 4752