Improved Method for Determination of Biofuel Sugars by HPAE-PAD

Similar documents
In-Cell and External Hydrolysis of Biomass for Carbohydrate Determinations Using Automated Solvent Extraction

Practical guidance for switching between anion and cation analysis systems

Determination of Inorganic Anions in Acid Rain Using a Dedicated High-Pressure Capillary Ion Chromatography System

Effect of Temperature on IC Signal Noise

The Effect of Working Electrode Gasket Thickness on the Sensitivity and Linearity of Carbohydrate Response by Pulsed Amperometric Detection

HPAE-PAD analysis of galactosyl-oligosaccharidecontaining samples using dual eluent generation cartridges

Thermo Scientific. Dionex CarboPac SA10. Column Product Manual. P/N: July Part of Thermo Fisher Scientific

Methods for Determining Sugars and Hydroxymethylfurfural in Biomass

Glycoprotein Monosaccharide Analysis Using HPAE-PAD with Eluent Generation

Dual EGC A simple solution for complex carbohydrate analysis

Monosaccharide and Sialic Acid Determinations in Biosimilars Using HPAE-PAD

Configuring a High-Pressure Dedicated Capillary IC System for Electrochemical Detection

Assay for Citrate and Phosphate in Pharmaceutical Formulations Using a High-Pressure Compact Ion Chromatography System

Determination of Trace Concentrations of Oxyhalides and Bromide in Municipal and Bottled Waters Using a Compact Ion Chromatography System

New Developments in Capillary Ion Chromatography Systems with Electrochemical Detection and Their Applications

Determination of Perchlorate in Drinking Water Using a Reagent-Free Ion Chromatography System

Implementation of a Walk-Up High- Pressure Capillary Ion Chromatograph for the Fast Separation of Pharmaceutical Relevant Inorganic Anions and Cations

HPAE-PAD analysis of Carbohydrates. Michael G. Hvizd Manager, Laboratory Operations Centers of Excellence Dionex Corporation Laboratory

Configuring the Dionex Integrion HPIC system for trace anion determinations in ultrapure water

Determination of Oxalate and Other Anions in Bayer Liquor Using Ion Chromatography

Determination of Total Nitrogen and Phosphorus in Environmental Waters

Thermo Scientific Dionex ICS-4000 Integrated Capillary HPIC System. Always ready, better results. capillary high-pressure ion chromatography

Errata. Product Manual for Dionex CarboPac PA20 Columns

Be a pioneer. Thermo Scientific Dionex ICS-6000 HPIC system the freedom to explore

Analysis of Fracking Flowback Water From the Marcellus Shale Using In-line Conductivity, Automated Dilution, and Ion Chromatography

with High Pressure Sponsored content from Presented in partnership with

Configuring the Dionex Modular HPIC system with an inline water purifier and inline calibration for trace anion determinations in ultrapure water

Dionex Eluent Generator Cartridges

Introduction. Yan Liu, Victor Barreto, Rosanne Slingsby, Chris Pohl Thermo Fisher Scientific, Sunnyvale, CA, USA

Determination of Manganese in Brine ( 30% NaCl)

CarboPac PA200 Columns For Dual EGC ICS-6000 For Research Use Only. Not for use in diagnostic procedures.

Determination of Cations in Hydraulic Fracturing Flowback Water from the Marcellus Shale

Determination of Glycoprotein Monosaccharide Composition by HPAE-PAD Using On-Line Electrolytically Generated Eluents

NISTmAb characterization with a high-performance RP chromatography column

Be a pioneer. Thermo Scientific Dionex ICS-6000 HPIC system the freedom to explore

Determination of Sulfite in Food and Beverages by Ion Exclusion Chromatography with Pulsed Amperometric Detection

Improved Determination of Trace Perchlorate in Drinking Water Using 2D-IC

Next generation suppressor for ion chromatography. Thermo Scientific Dionex ERS 500 Electrolytically Regenerated Suppressor

( ) IONPAC ATC-HC ANION TRAP COLUMN

Determination of Tobramycin in Crude and In-Process Production Samples During Manufacturing Using HPAE-IPAD

Determination of additives and byproducts in an acid copper plating bath by liquid chromatography

Determination of Perchlorate in Drinking Water by Ion Chromatography

Determination of Total and Potential Sulfate and Total Chloride in Fuel-Grade Butanol Per ASTM D

Direct Carbohydrate Analysis in Beverages and Foods Using Pulsed Amperometric Detection or Charged Aerosol Detection

Thermo Scientific Dionex DRS 600 Dynamically Regenerated Suppressor

HIGH QUALITY, RELIABLE ION ANALYSIS SOLUTIONS FOR EVERY IC CHALLENGE THERMO SCIENTIFIC DIONEX PRODUCT PORTFOLIO

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

Advanced Features New Advanced Automation Features Precise Performance Temperature Control Speed Application Flexibility

Direct Carbohydrate Analysis in Beverages and Foods Using Pulsed Amperometric Detection or Charged Aerosol Detection

ph gradient analysis of IgG1 therapeutic monoclonal antibodies using a 5 µm WCX column

Highlighted Capillary IC Applications

Minimizing Carbon Dioxide Contamination in Anion Reagent-Free Ion Chromatography

High-Resolution Charge Variant Analysis for Top-Selling Monoclonal Antibody Therapeutics Using a Linear ph Gradient Separation Platform

Continuously Regenerated Trap (CR-TC 500, Capillary)

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

A highly sensitive and robust 150 µm column to enable high-throughput proteomics

Thermo Scientific Dionex ICS HPIC Capillary and Analytical System. High resolution, fast analysis high-pressure ion chromatography

Performance characteristics of the 1260 Infinity Quaternary LC system

Options Eluent Regeneration (RFIC-ER ) Column Heater In-Line Vacuum Degas Full Chromeleon Chromatography Data System

Emulation of the Agilent 1100 Series LC Through Waters Empower Software Analysis of an Analgesic Mixture

Determination of halogens in coal using combustion ion chromatography

Carbohydrate Analysis in Beverages and Foods Using Pulsed Amperometric Detection or Charged Aerosol Detection

Determination of Trace Copper, Nickel, and Zinc in Boiling Water Reactors Using Nonsuppressed Conductivity Detection

HIC as a Complementary, Confirmatory Tool to SEC for the Analysis of mab Aggregates

Heparin Sodium - USP method

Determination of Morpholine, Ethanolamine, and Hydrazine in Simulated Nuclear Power Plant Wastewater

CX-1 ph Gradient Buffer

Thermo Scientific Dionex ICS-900 Ion Chromatography System

Sensitive Determination of Hexavalent. Brian De Borba, Lipika Basumallick, Jeffrey Rohrer

Dionex IonPac CS16-4µm and Dionex IonPac CS16-Fast-4µm

Performance Characteristics of the Agilent 1220 Infinity Gradient LC system

Dionex IonPac AS16 Column

Enjoy Industry-Leading Support

Fluorine and Chlorine in Cu concentrates and ores by Metrohm Combustion IC

Complete Customer Support

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

High-Resolution Separation of Intact Monoclonal Antibody Isoforms

Fast Ion Determinations and Efficient Separations of Complex Samples Using 4 µm Particle-Size Ion Exchange Columns

PRODUCT MANUAL. for the

Dionex DRS 600 Suppressor Dionex ERS 500e Suppressor Dionex ERS 500 Carbonate Suppressor

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Reducing Carbonate Interference in Anion Determinations with the Carbonate Removal Device (CRD)

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

Product Manual for the Dionex IonPac Ultra Trace Anion Concentrator (UTAC)

Thermo Scientific. Dionex IonPac AS9-HC. Column Product Manual. P/N: May Part of Thermo Fisher Scientific

Application Note 168

A HIGHLY VERSATILE IC SYSTEM CAPABLE OF EVERY CHALLENGE THAT COMES YOUR WAY THERMO SCIENTIFIC DIONEX ICS HPIC SYSTEM

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Biofuels Research at the University of Washington

Application Note. Author. Abstract. Pharmaceuticals. Detlef Wilhelm ANATOX GmbH & Co. KG. Fuerstenwalde, Germany mau

Dionex IonPac AS19 Columns

A Simple Rapid and Sensitive Method Development for Quantification of Quetiapine Fumarate in Bulk and Dosage Forms Using RP-HPLC

Advanced Biofuels (and Bio-products) Process Demonstration Unit with: BIO World Congress on Industrial Biotechnology July 22, 2015

Monosaccharides. Robust Applications, Fluidly Running. Introduction. Application Note Clinical/Pharma/ Food/Life science

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

Lifetime stability of size exclusion chromatography columns for protein aggregate analysis

Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC

Technical Guide. Thermo Scientific. Capillary High-Pressure Ion Chromatography. simplifying IC while increasing resolution and speed

Transcription:

APPLICATION NOTE Improved Method for Determination of Biofuel Sugars by HPAE-PAD No. 1161 Sachin Patil 1, Akash Narani 2, Phillip Coffman 2, Todd Pray 2, Deepti Tanjore 2, and Jeffrey Rohrer 1, 1 Thermo Fisher Scientific, Sunnyvale, CA; 2 Advanced BioProcessing Demonstration Unit (AB-PDU), Lawrence Berkeley National Lab, Berkeley, CA Keywords Lignocellulosic biomass, Dionex Integrion HPIC system, Fermentable sugars, Biomass hydrolysis, Electrochemical detection, Biomass-to-biofuel conversion Goal To develop a fast ion chromatography (IC) method for the determination of sugars in biofuel samples using a Thermo Scientific Dionex CarboPac SA10-4µm column with electrolytically generated eluent and a high pressure IC system with electrochemical detection. Introduction Significant reduction in the supplies of underground fossil fuels and increased production of greenhouse gasses has led to a worldwide focus on alternative paths for energy generation from renewable sources, such as lignocellulosic biomass 1-2. The breakdown of cellulose and hemicellulose contained in lignocellulosic biomass feedstocks produces five- and six-carbon sugars that can be fermented into biofuels such as ethanol 3 and biodiesel 4. The efficiency of biomass hydrolysis into fermentable sugars is used to assess the overall efficiency of biomass-to-biofuel conversion, and is directly related to target biofuel yield and process economics. 5 The determination of carbohydrates in hydrolysates derived from lignocellulosic biomass is a crucial step in biofuel production. However, development of robust analytical methods remains a challenge. Current methods suffer from several drawbacks, such as low throughput, poor analyte resolution, and lack of applicability to a wider set of biomass samples 6. High performance anion-exchange chromatography with pulsed amperometric detection (HPAE-PAD) can be used to determine sugars from biomass hydrolysate samples. HPAE-PAD has been shown to deliver fast determinations of carbohydrates in biomass hydrolysate samples using the Thermo Scientific Dionex CarboPac SA10-4µm column. 7 Determination of carbohydrates in acid-hydrolyzed corn stover samples at 10-fold dilution was shown in Thermo Scientific Application Note 1089 (AN1089). 8 The method uses electrolytically generated hydroxide eluent and a 62 mil gasket in the electrochemical flow cell to resolve and detect the eight 1

common lignocellulosic biomass derived fermentable sugars, including xylose, sucrose, arabinose, galactose, glucose, mannose, fructose, and cellobiose. The current work updates the system used for the carbohydrate analysis in AN1089. 8 The new system combines flexibility and ease-of-use with high sensitivity and selectivity, bringing a higher level of convenience and cost effectiveness to simple sugar analysis. The improved Thermo Scientific Dionex IC PEEK Viper fitting reduces the dead volume in the flow path and enhances chromatographic performance. The method proposed here separates eight common carbohydrate sugars in less than eight minutes. The shorter run time allows for shorter sample turnaround times and reduced eluent consumption, thereby improving the overall process economics. Using this method, carbohydrates present in 10 individual biofuel samples were quantified. Conditions Column: Thermo Scientific Dionex CarboPac SA10-4µm, 4 250 mm (P/N 088233), CarboPac SA10-4µm Guard 4 50 mm (P/N 088234) Column Temp: 45 C Compartment Temp: 30 C Flow Rate: 1.5 ml/min Eluent: 1 mm KOH Eluent Source: Dionex EGC 500 KOH Eluent Generator Cartridge (P/N 075778) with Dionex CR ATC Continuously Regenerated Anion Trap Column (P/N 075550) Electrochemical Cell Gasket: 62 mil Reference Electrode: ph-ag/agcl Sampler Tray Temp: Ambient Injection Volume: 0.4 µl (Push_Full mode) Samples The biomass hydrolysate samples tested in this study were derived from corn stover, switchgrass, and/or energy cane. These biomass samples were pretreated either individually or as mixtures (mixed feedstocks) with any one of the reagents: acid, alkali, or ionic liquids. The pretreatments were followed by enzymatic hydrolysis with Cellic Ctec2 and Cellic Htec2 (Novozymes, CA) to yield near-complete hydrolysis of biomass oligosaccharides. All biomass hydrolysate samples were diluted 10-fold with DI water before use. Note: Lignocellulosic-biomass-derived samples have complex matrices. Highly retained contaminants in these samples will occupy a portion of the anion exchange sites, limiting the number of sites available for retention of analyte anions, especially after continued use. This may result in loss of column capacity and shortened retention times. If such column contamination is suspected then the Dionex CarboPac SA10-4µm column can be readily cleaned using 100 mm KOH or NaOH. Refer to Appendix A CarboPac SA10-4µm Column Care in the column manual (Document# 065579-02) for column washing procedure and additional information on column care. Carbohydrate Waveform Time (s) Potential (V) Integration 0.00 +0.10 0.20 +0.10 Begin 0.40 +0.10 End 0.41-2.0 0.42-2.0 0.43 +0.6 0.44-0.1 0.50-0.1 Equipment A Thermo Scientific Dionex Integrion HPIC system was used in this work. The Dionex Integrion HPIC system is an integrated ion chromatograph that includes: Dionex Integrion HPIC system pump Detector compartment temperature control Electrochemical detector (P/N 22153-60037) and cell (P/N 072044) ph-ag/agcl reference electrode (P/N 061879) Carbohydrate disposable Au working Electrode, pack of 6 (two 2.0 mil gaskets included) (P/N 066480) AS-AP autosampler (P/N 074925) with cooling tray option (recommended) and 1.5/0.3 ml vial tray (P/N 074936) Sterile assembled microcentrifuge tubes with screw cap, 1.5 ml (Sarstedt P/N 72.692.005) Nalgene Rapid-Flow 0.2 μm filter units, 1000 ml, nylon membrane, 90 mm diameter (Thermo Scientific P/N 164-0020) Consumables Part numbers are shown in Table 1 2

Table 1. Consumables for the Dionex Integrion HPIC system. Product name Description of High Pressure Device Part Number Thermo Scientific Dionex IC PEEK Viper fitting tubing assembly kits Dionex IC PEEK Viper fitting tubing assembly kit for the Dionex Integrion HPIC system: Includes one each of P/Ns: 088815-088821 088798 Guard to separator column: 0.007 4.0 in (102 mm) 088815 Dionex IC PEEK Viper fitting tubing assemblies Thermo Scientific Dionex AS- AP Autosampler vials 4-Port Injection Valve pod Thermo Scientific Dionex EGC 500 KOH Eluent Generator cartridge Thermo Scientific Dionex CR-ATC 600 Continuously Regenerated Anion Trap Column Valve to guard column: 0.007 5.5 in 088816 (140 mm) EGC Out to CR-TC Eluent In: 0.007 6.5 in (165 mm) 088817 Separator to ED Cell In: 0.007 7.0 in 088819 (178 mm), ED CR-TC Out to Degasser In: 0.007 9.5 in 088821 (241 mm) Package of 100, polystyrene vials, caps, septa, 0.3 ml 055428 Install in place of 6-port valve pod. The 4-port pod has an 074699 internal sample loop of 0.4 µl. Eluent generator cartridge when using 4 µm particle columns 075778 Continuously regenerated trap column used with Dionex EGC KOH 500 cartridge 088662 Dionex HP EG Degasser Degasser module 075522 Electrochemical Detector (ED) Included with Integrion ED. P/N if purchased separately. 072042 Electrochemical cell Includes knob and, support block 072044 ph-ag/agcl reference electrode Reference electrode 061879 Au on PTFE electrodes Working electrode, package of six 066480 High concentration Includes 62 mil gasket and modified spacer block 085324 carbohydrate analysis kit 62 mil gasket If purchased separately, package of two 075499 ph Buffer, ph 7 Reference electrode ph calibration standard SB115-500* ph Buffer, ph 10 Reference electrode ph calibration standard SB107-500* * Fisher Scientific P/N 3

Reagents and Standards Deionized (DI) water, Type I reagent grade, 18 MΩ-cm resistivity or better L( )-Fucose (Sigma, P/N F2543) D-Galactose (Sigma, P/N G0625) D(+)-Mannose (Sigma P/N M6020) D-Fructose (Sigma P/N F25X43) D-Xylose (Aldrich P/N X-10705) Sucrose (Sigma P/N 84097) D-Glucose (P/N 1910-01) D(-)-Arabinose (Sigma P/N A3131) D(+)-Cellobiose (Sigma P/N C7252) Results Separation Figure 1 shows a representative chromatogram obtained using a standard mix containing eight common biofuel sugars at 0.6 g/l each, analyzed on Dionex CarboPac SA10-4µm column. Fucose was used as an internal standard. All of the peaks were well resolved and the separation was completed within 8 min. 70 nc 1 2 3 4 5 6 25 0 2 4 Minutes 6 8 10 Column: Dionex CarboPac SA10-4µm Peaks: 1. Fucose Analytical, 4 250 mm 2. Sucrose Eluent: 1 mm KOH 3. Arabinose Flow Rate: 1.5 ml/min 4. Galactose Inj. Volume: 0.4 µl 5. Glucose Temp: 45 C 6. Xylose Detection: PAD (Au) 7. Mannose Gasket: 62 mil 8. Fructose 9. Cellobiose 7 8 9 Linearity and Precision Using this method, calibration curves were generated for the eight biofuel sugars. The results are included in Table 2. The method is linear for more than two orders of magnitude from 0.005 to 2 g/l for all eight carbohydrates. This wide dynamic range is an important feature required of any method designed to measure sugars present in biomass hydrolysates, as it is ideal for determination of the sugars that are typically present in divergent ratios. The retention time and peak area relative standard deviations (RSD) are less than 1.26% across all sugars, indicating excellent method precision. Accuracy To determine method accuracy, 10 field biomass hydrolysate samples were procured. The samples were collected at various stages of biomass processing and conversion of lignocellulosic biomass to ethanol. Method accuracy was evaluated by first quantifying the sugars present in the biomass hydrolysate samples. From this set, five samples were randomly selected for recovery studies. Representative chromatograms for the selected five biofuel samples are shown in Figure 2. The major sugars present in these samples are glucose and xylose. 200 nc 1 3 2 4 5 6 7 8 9 0.6 g/l standard A 20 0 2 4 6 8 10 Minutes Column: Dionex CarboPac SA10-4µm Analytical, 4 250 mm Eluent: 1 mm KOH Flow Rate: 1.5 ml/min Inj. Volume: 0.4 µl Temp: 45 C Detection: PAD (Au) Gasket: 62 mil E D C B Peaks: 1. Fucose 2. Sucrose 3. Arabinose 4. Galactose 5. Glucose 6. Xylose 7. Mannose 8. Fructose 9. Cellobiose Figure 2. Representative chromatograms of five lignocellulosic biomass based biofuel samples A-E. Figure 1. Separation of biofuel sugars on the Dionex CarboPac SA10-4µm column using proposed method (0.6 g/l standard). 4

All of the biofuel sugars that are present above the lowest calibration standard were spiked with 50 150% of the original amount found in the biofuel sample. The original amounts present, spiked amounts, and percent spike recoveries are included in Table 2. For all four samples studied here, excellent recoveries were obtained for all spiked biomass sugars. Robustness Due to the complex matrices that are characteristic of biomass samples, it is critical to ensure minimal matrix effect on retention time and peak area stability during a long sequence run. The method robustness was studied by injecting the standard sugar mix at frequent intervals during a 200 injection sequence run of the biofuel samples. The results included in Tables 3 and 4 show excellent retention time and peak area reproducibility respectively across the four injections of standard sugar mix located at different positions in the sequence. Table 2. Calibration (at 11 levels between 0.006 to 2 g/l of each of the biofuel sugars) and precision data (0.6 g/l, n=7) for eight common biofuel sugars. Parameter Sucrose Arabinose Galactose Glucose Xylose Mannose Fructose Cellobiose Coe. Of Determination 0.996 0.995 0.999 0.993 0.992 0.992 0.997 0.999 RT RSD (%) 0.12 0.08 0.08 0.10 0.06 0.10 0.05 0.04 Area RSD (%) 0.28 0.47 0.51 0.62 0.66 0.69 1.26 0.28 Table 3. Biofuel sugar recovery studies (n=7). Sample Name Sugar Avg. Amount (g/l) Spike (g/l) Recovery (%) RSD (%) Arabinose 0.02 0.02 100 0.1 A Glucose 1.17 0.8 106 0.1 Xylose 0.5 0.5 114 0.0 Cellobiose 0.03 0.04 98.7 0.0 Arabinose 0.01 0.10 94.1 0.0 B C Galactose 0.07 0.05 97.6 0.0 Glucose 1.07 0.8 89.4 0.0 Xylose 0.8 0.5 95.6 0.0 Arabinose 0.09 0.10 91.2 0.0 Galactose 0.05 0.05 129 0.1 Glucose 1.03 0.8 93.8 0.0 Xylose 0.7 0.5 104 0.0 Arabinose 0.04 0.02 88.3 0.3 D Galactose 0.00 0.01 117 0.1 Glucose 0.72 0.8 97.7 0.0 Xylose 0.4 0.5 111 0.0 Arabinose 0.03 0.02 95.3 0.1 E Galactose 0.01 0.02 99.1 0.0 Glucose 0.55 0.8 100 0.0 Xylose 0.4 0.5 108 0.0

Table 4. Retention time robustness. Inj. No. Retention Time (min) Fucose Sucrose Arabinose Galactose Glucose Xylose Mannose Fructose Cellobiose 1 2.66 3.16 3.53 3.74 4.13 4.56 4.83 5.10 7.63 75 2.68 3.18 3.55 3.75 4.13 4.58 4.83 5.12 7.63 100 2.66 3.16 3.53 3.73 4.12 4.56 4.82 5.10 7.60 200 2.66 3.15 3.53 3.73 4.10 4.53 4.79 5.08 7.55 Table 5. Peak area robustness. Inj. No. Area (nc * min) Fucose Sucrose Arabinose Galactose Glucose Xylose Mannose Fructose Cellobiose 1 0.46 2.09 2.91 3.08 3.10 3.48 2.99 2.78 2.20 75 0.44 2.03 2.85 3.03 3.01 3.35 2.88 2.74 2.16 100 0.44 2.05 2.88 3.05 3.04 3.37 2.93 2.75 2.20 200 0.44 2.03 2.84 3.03 3.02 3.33 2.91 2.72 2.21 Conclusion An improved method for quantification of biomass sugars that can be converted to biofuels is presented here. The method is able to separate all eight key biomass sugars within 8 min. The response data for all biomass sugars studied here showed excellent correlation with concentration. The method provides excellent peak area reproducibility, as well as retention time precision within the concentration levels studied. Spike recovery studies performed to demonstrate method accuracy showed good recovery from all of the spiked samples, suggesting that the method is accurate. Finally, the method is robust with retention time and peak area remaining stable when subjected to over 200 biofuel sample injections. In summary, the method proposed here is a convenient, precise, and robust way to quantify biomass sugars in complex hydrolysate samples and will improve the reliability of determining biomass-to-biofuel efficiency calculations. References 1. Hill, J.; Nelson, E.; Tilman, D.; Polasky, S.; Tiffany, D. Environmental, economic, and energetic costs and benefits of biodiesel and ethanol biofuels. PNAS, 2006, 103(30), 11206 11201. 2. Ragauskas, A.J.; et al. The Path Forward for Biofuels and Biomaterials. Science, 2006, 311, 484-489. 3. Goldemberg, J. Ethanol for a sustainable energy future. Science, 2007, 315, 808 810. 4. Alonso, D. M.; et al. Catalytic conversion of biomass to biofuels. Green Chem. 2010, 12, 1493 1513. 5. Wyman C.E., Ethanol from Lignicellulosic biomass: Technology, economics and opportunity. Bioresour. Technol., 1994, 50, 3 16. 6. Sluiter, J.B.; et al. Compositional Analysis of Lignocellulosic Feedstocks. 1. Review and Description of Methods. J. Agric. Food Chem., 2011, 58(16), 9043 9053. 7. Thermo Scientific Application Note 192. Carbohydrate Determination Biofuel Samples. https://appslab. thermofisher.com/app/2277/au192-carbohydratedetermination-biofuel-samples (Accessed July 22, 2016). 8. Thermo Scientific Application Note 1089. Determination of Carbohydrates in Acid Hydrolysates of Wood. https://tools.thermofisher.com/content/sfs/brochures/ AN-1089-Carbohydrates-Acid-Hydrolysates-Wood- AN70941-EN.pdf (Accessed July 22, 2016).

Acknowledgements We thank the BioEnergy Technologies Office (BETO) at the Department of Energy (DOE) division of Energy Efficiency and Renewable Energy (EERE) for the funding required to perform the deconstruction studies. We also thank Drs. Allison Ray and Chenlin Li from the Biofuels and Energy Technologies Department at the Idaho National Laboratory for providing us with the lignocellulosic biomass feedstocks that were studied in this project. The Cellic Ctec2 and Cellic Htec2 enzymes used here were graciously provided for research purposed by Novozymes, CA. Find out more at thermofisher.com/chromatography 2016 Thermo Fisher Scientific Inc. All rights reserved. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of 7 these products in any manners that might infringe the intellectual property rights of others. AN1161-EN 1216S