PRINCIPLES OF DISINFECTION AND STERILIZATION IN THE DENTAL SETTING OBJECTIVES

Similar documents
3/26/2014 SPAULDING CLASSIFICATION OBJECTIVES PRINCIPLES OF DISINFECTION AND STERILIZATION IN THE DENTAL SETTING PROCESSING SEMI CRITICAL OBJECTS

Principles of Disinfection and Sterilization in the outpatient setting

Principles of Disinfection and Sterilization in the outpatient setting

Ideal Sterilization Method. Sterilization. Sterilization Practices in Healthcare Facilities

Disinfection & Sterilization

disinfectionandsterilization.org

Objectives. The Perioperative Nurse s Role 6/24/2016. Instrument Processing & Sterilization Chesapeake Perioperative Consortium 2016

Guidelines for Selection and Use of Disinfectants

THE BASICS OF STERILIZATION

Guidelines for the use of disinfectants in hospital settings. Eleni Tomprou Infection Control Nurse, MSc General Hospital of Athens Polykliniki

Sterilization, Disinfection and Antisepsis

Disinfection and Sterilization. Efficacy of Disinfection/Sterilization Influencing Factors

ARABIC CANADIAN MEDICAL CENTER CSSD POLICIES AND PROCEDURES

Low Temperature Processing. An Update on Guidelines, Standards and Requirements

3/11/ M Sterile U Network 3M Sterile U Web Meeting March 11, Welcome! Topic: An Overview of ANSI/AAMI ST58: 2013

Disinfection and sterilisation

Understanding ANSI/AAMI ST 91. Flexible and Semi-Rigid Endoscope Reprocessing in Healthcare Facilities

Christina Bradley Endoscope Decontamination A Webber Training Teleclass. Christina Bradley

Disinfection & Sterilization in the Ambulatory Surgery Center

Slide 1. Slide 2. Slide 3

Cleaning, Disinfection and Sterilization. Basics of Infection Prevention 2-Day Mini-Course October 2016

Disinfection, Sterilization and Antisepsis: An Overview

Sterilization, disinfection and reprocessing

Administrative Policies and Procedures. Cross Reference: Date Issued: 2/14 Date Reviewed: 9/14 Date: Revised: 12/14 Attachment: None Page of 1 of 6

8/27/2015. Learning Objectives. Purpose of Sterilization. To destroy all living pathogenic microorganisms

Best Practices for High Level Disinfection Part I

Sterilization and Disinfection

Inhibiting of Microbial Growth in vitro CLS 212

Medical Device Reprocessing Assessment - General (Cleaning, Disinfection, and/or Sterilization)

Agent Mechanisms of Action Comments Surfactants. Membrane Disruption; increased penetration Denature proteins; Disrupts lipids

Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar

UNIT Two. Lesson 1C Sterilization

Sterilization Policy. Georgia Regents Medical Center Policy Library. Policy Owner: Epidemiology POLICY STATEMENT

Bio-Burden Reduction in Biological Laboratories

Claire Kari Biosafety Specialist DEHS Biosafety October 2010

Hopefully you will learn.. Objectives. Contact Transmission. Pathway of Disease Transmission. Contact Precautions Direct Indirect

Guard by TSSU. October 4,2009

Manner of Assessment Code (check all that apply) & Surveyor Notes. Yes N/A. Yes. Yes N/A N/A. Yes N/A. N/A Yes N/A

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of

Inhibiting Microbial Growth in vitro. CLS 212: Medical Microbiology Zeina Alkudmani

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of

Infection Prevention and Control.

What s New in Disinfection and Sterilization of Patient Care Equipment A Webber Training Teleclass With Dr. William A. Rutala May 22, 2003 Slide 1

مادة االدوية املرحلة الثالثة أ.م.د. حسام الدين سامل

REPROCESSING OF FLEXIBLE CYSTOSCOPES

8/23/2012. Continuing Education Contact Hours. How much of your formal education focused on sterilization or sterile processing?

TEST CODE DEFINITION STERRAD H202. Test Code Test Name Testing Description. Cycle. Efficacy. SE-200 STERRAD 200 Efficacy. Cycle. STD Cycle Efficacy

Guidelines for Disinfection and Sterilization in Healthcare Facilities, 2008 (Centers for Disease Control)

Quality Control EQUIPMENT CONTROL. Monitor critical performance characteristics of sterilization, decontamination or cleaning equipment

Competency Guide: Care and Handling of Rigid Endoscopes

Cleaning And Disinfectant Products

Best Practices for Environmental Cleaning. MODULE 3 Cleaning Products and Tools

Lecture Summary Microbial Control of Growth (CH5)

Issues and Controversies in Instrument Reprocessing

Important Terminology (pg )

Overview: High Level Disinfection

VHP Sterilization and Prion Inactivation

Infection Prevention in the Outpatient Setting

Definitions. BIOL 3702: Chapter 8. Control of Microbes in the Environment. Mechanical Removal Methods. Pattern of Microbial Death

Selection and Use of Disinfectants. Disinfection. Disinfection and Sterilization. Selection and Use of Disinfectants

Biosecurity Sanitation and Pest Control

SUBJECT: Detection of Blood residue inside the biopsy channel of a scope

Facility: Sterile Processing Assessment Date: CRITERIA ANSI/AAMI COMPLIANT Number Y N N/A Point of Use Gross contaminant is removed with

Disinfection and Sterilization

INFECTION PREVENTION PROCESSES 1

Definitions. BIOL 3702: Chapter 8. Control of Microbes in the Environment. Mechanical Removal Methods. Pattern of Microbial Death

Infection Control Manual. Table of Contents

8. Scrubbing or immersing the skin in chemicals to reduce the numbers of microbes on the skin is: A. disinfection B. sterilization C. antisepsis D.

CHECK YOUR for your CE certification of completion.

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces

Not to be reproduced - AGPAL & QIP

Draft Guidance for Industry and FDA Staff Processing/Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling

Foundations in Microbiology Seventh Edition. Talaro Chapter 11 Physical and Chemical Agents for Microbial Control

Ready-To-Use Cleaning and Disinfecting Fluid for Endoscopes & Medical Instruments Technical Information

SUBJECT: Detection of Blood (hemoglobin) residue inside the biopsy channel of an endoendoscope

Packaging & Sterilization

Chapter 7: Control of Microbial Growth

Important Terminology

Quality Control LOAD CONTROL. VERIFY Load Control Products. Use to monitor and release instrumentation with confidence RECORD KEEPING & LABELING

MDRAO Importance of Bedside & Manual Precleaning CONMED.CA

Sterilization & Disinfection

Sterilization and Disinfection: Getting It Right. Instructions

One such problem is blood residual inside the channels of a robotic arm.

Infection Prevention in Endoscopy, Keeping Patients and Practioners Safe

MDRAO ONLINE COURSE FOURTH EDITION 2018

Sterilization and Disinfection

3.03 Understand support services

ANSI/AAMI ST91: 2015

Sterilization Overview from the CDC

Endoscopy and OR Solutions SYSTEM 1 TM. Express STERILE PROCESSING SYSTEM SAFETY, ASSURANCE AND EFFICIENCY. One Integrated Approach to Healthcare

SOP BIO-006 USE OF AUTOCLAVE FOR STERILIZATION OF MATERIALS AND BIOLOGICAL WASTE

How is Sterilization Validation Performed by the Medical Device Manufacturer?

If All Else Fails Read the Directions. Disclosure. Manufacturer s Instructions. Presenter: Sue Lafferty CHICA Canada Conference, May 2010

Is Your Sterile Processing Department Ready for Survey?

The Gold Standard in Low Temperature Sterilization: Ethylene Oxide

Products made in Switzerland

CDWA3 / CDWA4 CDWAH. Page 4. BASKET ZONE: water pressure, spray system performance, detergent chemistry and concentration, temperature.

Instructions for use. Reusable surgical instruments. Hersteller:

Transcription:

Module F PRINCIPLES OF DISINFECTION AND STERILIZATION IN THE DENTAL SETTING OBJECTIVES State the principles of disinfection and sterilization List the current methods for disinfection and sterilization per CDC guideline recommendations 1

SPAULDING CLASSIFICATION Spaulding Classification of Surfaces: 1. critical Objects which enter normally sterile tissue or the vascular system and require sterilization 2. semi critical Objects that contact mucous membranes or non intact skin and require high level disinfection, which kills all but high levels of bacterial spores 3. non critical Objects that contact intact skin but not mucous membranes, and require low level disinfection ORDER OF RESISTANCE OF MICROORGANISMS TO DISINFECTANTS/STERILANTS Hardest to Kill Easiest to Kill Prions (Creutzfeldt Jakob Disease (CJD), mad cow disease) Spores (C. difficile) Mycobacteria (Tb) Non enveloped viruses (norovirus) Fungi (Candida) Vegetative bacteria (MRSA, VRE) Enveloped viruses (HIV, HBV) 2

PROCESSING CRITICAL INSTRUMENTS Penetrate or enter normally sterile tissue or spaces, including the vascular system Surgical instruments (elevators, bone files, rongeurs, forceps, etc.) Must be sterilized between uses or used as singleuse disposable devices Goal: Sterility = devoid of all microbial life PROCESSING SEMI CRITICAL OBJECTS Contact mucous membranes and non intact skin Mouth mirrors, cheek retractors, handpieces Must be sterilized or immersed in high level disinfectant Goal: High level disinfection = free of all microorganisms except low numbers of bacterial spores 3

HIGH LEVEL DISINFECTANTS Germicide Concentration Glutaraldhyde (Cidex ) 2.0% Ortho phthaladehyde (Cidex OPA ) 0.55% Hydrogen Peroxide* (Sporox TM ) 7.5% Hydrogen Peroxide and peracetic acid* (Peract TM ) 1.0% / 0.08% Hydrogen Peroxide and peracetic acid* (Endospor TM +) 7.5% / 0.23% Hypochlorite (free chlorine)* (Sterilox) 650 675 ppm Accelerated hydrogen peroxide (Resert TM XL) 2.0% Peracetic Acid (Steris 20 TM ) 0.2% Glutaraldehyde and Isopropanol (Aldahol III ) 3.4% / 26% Glutaraldehyde and phenol/phenate (Sporicidin ) 1.21% / 1.93% Exposure time 8 45 min (US) and temperature 20 25 C; *May cause cosmetic and functional damage NON CRITICAL INSTRUMENTS AND DEVICES Contact intact skin BP cuffs, electrocardiogram (EKG) leads, stethoscopes Disinfect using a low level disinfectant Goal: Kill vegetative bacteria, fungi, viruses 4

LIQUID DISINFECTANTS Disinfectant Agent Use Concentration Ethyl or isopropyl alcohol 70% 90% Chlorine (bleach) 100ppm Phenolic UD Iodophor UD Quaternary ammonium UD compound (QUAT) Improved hydrogen peroxide 0.5%, 1.4% Exposure time 1 minute UD = Manufacturer s recommended use dilution STERILIZATION PROCESS 5

FACTORS INFLUENCING THE EFFICACY OF DISINFECTION AND STERILIZATION PROCESSES Cleaning of the object Organic and inorganic load present Type and level of microbial contamination Concentration and exposure time to the disinfectant/sterilant Nature of the object Temperature, ph, and water hardness WHERE ARE YOU PROCESSING YOUR INSTRUMENTS? Decontamination Holding area Clean instruments, handpieces Rinse Dry Enter Exit Closed storage of sterile packages BI incubator and record keeping Sterilization Sterilization and storage Arrange Instruments Add chemical monitors, BIs, supplies Packaging Wrap cassettes, package items, seal, label 6

CLEANING INSTRUMENTS Soak in enzymatic or non enzymatic detergent Wear the appropriate PPE Keep instruments submerged in solution and use a longhandled brush when manually cleaning instruments AUTOMATED CLEANING Ultrasonic cleaner Instrument washer Washer disinfector Regulated by FDA Household dishwasher NOT recommended 7

PREPARATION AND PACKAGING Critical and semi critical items that will be stored should be wrapped before heat sterilization Hinged instruments opened and unlocked Place a chemical indicator inside the pack Wear heavy duty, puncture resistant utility gloves Peel packs Rigid containers Self seal roll stock Sterile wraps woven and non woven Compatible with sterilization method Must be FDA approved PACKAGING 8

LOADING Place items/packages correctly and loosely into the sterilizer so as not to impede penetration of the sterilant Peel packs and non perforated containers (e.g., basins) should be placed on their edge STERILIZATION The complete elimination or destruction of all forms of microbial life by either physical or chemical processes. 9

METHODS OF STERILIZATION High Temperature Methods: Steam sterilization Dry Heat Steam formaldehyde (chemiclave) Low Temperature Methods: Hydrogen peroxide gas plasma Vaporized hydrogen peroxide Ethylene oxide Advantages Non toxic STEAM STERILIZATION Cycle easy to control and monitor Inexpensive Rapidly microbicidal Rapid cycle time Least affected by organic/inorganic soils Penetrates medical packing, device lumens 10

Disadvantages STEAM STERILIZATION Deleterious for heat labile instruments Inappropriate for heat sensitive instruments Inappropriate for moisture sensitive instruments Dulling Rusting Potential for burns STEAM STERILIZATION Steam under pressure (autoclaving) Gravity displacement Pre vacuum 11

PROCESS TIMES FOR PACKAGED ITEMS Method Time Exposure (minutes) Temperature Dry Range, Time C ( F) Does (minutes) not include Range (minutes) Steam autoclavedrying time Steam Gravity autoclave Gravity Prevacuum Prevacuum 30 304 15 4 121 o C 132121 123 o C C (250 254 F) 132 135 o C (270 275 F) 132 135 o C (270 275 F) DRY HEAT STERILIZATION Transfers heat energy from air inside the oven to the instruments Requires higher temperatures 160 190 C (320 o 375 o F) Good for items that are likely to dull or rust in the autoclave, Good for powders, cellulose and ink Packaging must be able to withstand high temperatures 12

DRY HEAT STERILIZER TYPES Static Air Heating coils in bottom of chamber Natural convection as hot air rises in chamber Takes 1 2 hours for sterilization at 160 C Forced Air Rapid heat transfer Circulates heated air at high velocity throughout chamber Reduces time to 6 12 minutes at 190 C UNSATURATED CHEMICAL VAPOR STERILIZATION Hot chemical vapors (under pressure) kill microorganisms Active ingredient: formaldehyde+ Vapors extremely irritating to eyes and lungs Purge system and good ventilation essential Harvey sterilizer or Chemiclave No corrosion is instruments are dry Must use packaging designed for chemical vapor sterilizers 13

LIQUID CHEMICAL STERILANT/DISINFECTANTS Only for heat sensitive critical and semicritical devices Powerful, toxic chemicals raise safety concerns http://www.fda.gov/cdrh/ode/germlab.html http://www.fda.gov/medicaldevices/deviceregulationandguidan ce/reprocessingofsingle-usedevices/ucm133514.htm STORAGE OF STERILE ITEMS Ensure the sterile storage area is a well ventilated area that provides protection against dust, moisture, and temperature and humidity extremes. 8 inches from the floor8 5 inches from ceiling 18 inches from ceiling if sprinkler head present 2 inches from outside walls 10 inches from floor Sterile items should be stored so that packaging is not compromised. Label sterilized items with a load number that indicates the sterilizer used, the cycle or load number, and the date of sterilization. 14

STORAGE OF STERILE ITEMS Event related shelf life recognizes that the product remains sterile until an event causes it to become contaminated (e.g. moisture). Packages should be evaluated before use for loss of integrity. Repack and reprocess if compromised. RECOMMENDATIONS METHODS OF STERILIZATION Cleaning removes salts and proteins and MUST precede sterilization. Steam is preferred for critical and semi critical items not damaged by heat Follow the operating parameters recommended by the manufacturer Use an FDA cleared container, wrapping or packaging system that is compatible with the type of sterilization process used Use low temperature sterilization technologies for reprocessing critical and semi critical items damaged by heat Immediately use critical items that have been sterilized by liquid sterilants (e.g. peracetic acid) immersion process (no long term storage). 15

Monitoring the effectiveness of your sterilization equipment is essential! STERILIZATION MONITORING Physical cycle time, temperature, pressure Chemical heat or chemical sensitive inks that change color when germicidal related parameters reached Biological Bacillus spores that directly measure sterilization 16

MONITORING OF STERILIZERS Internal Chemical Indicator Validates the sterilant penetrated the pack or tray Detect local problem BIOLOGICAL MONITORS Steam Geobacillus stearothermophilus Dry heat B. atrophaeus (formerly B. subtilis) Ethylene oxide (ETO) B. atrophaeus 17

RECOMMENDATIONS MONITORING OF STERILIZERS Monitor each load with physical and chemical (internal and external) indicators. Use biological indicators to monitor effectiveness of sterilizers at least weekly with spores intended for the type of sterilizer. Use biological indicators for every load containing implantable items 18

RECOMMENDATIONS MONITORING OF STERILIZERS Following a single positive biological indicator from steam sterilization: Remove the sterilizer from service and review sterilizer instructions. Retest the sterilizer. If spore test negative, put the sterilizer back in service. If the spore test is positive: do not use until it has been inspected; and recall (to the extent possible) all items processed since the last negative spore test; challenge in three consecutive empty sterilization cycles. Single positive biological indicator (BI) from other than steam sterilization (e.g., dry heat sterilization) Treat as non sterile all items back to last load tested with negative indicator. Retrieve items and reprocess. Record Keeping 19

FAILURE TO FOLLOW DISINFECTION AND STERILIZATION PRINCIPLES WHAT DO YOU DO? Scenario: Dental Office A discovered that for the past 3 days all surgical instruments were exposed to steam sterilization at 132 o C for 0 minutes rather than the intended 4 minutes. A technician turned the timer to 0 minutes in error. American Journal of Infection Control 41 (2013) S67 S71 20

STEPS IN THE EVENT OF DISINFECTION AND STERILIZATION FAILURE 1. Confirm failure of disinfection or sterilization reprocessing 2. Immediately embargo any possibly improperly disinfected/sterilized items 3. Do not use the questionable disinfection/sterilization unit (sterilizer, automated endoscope reprocessor) until proper functioning has been assured. 4. Inform key stake holders (risk management, management, lawyers) 5. Conduct a complete and thorough evaluation of the cause of the disinfection/sterilization failure 6. Prepare a line listing of potentially exposed patients 7. Assess whether the disinfection/sterilization failure increases a patient s risk for infection 8. Inform an expanded list of stakeholders of the reprocessing issue 9. Develop a hypothesis for the failure and initiate corrective actions 10. Develop a method to assess potential adverse patient events 11. Consider notification of appropriate state and federal authorities (health department, FDA) 12. Consider patient notification 13. If patients are notified, consider whether such patients require medical evaluation for possible post exposure therapy with appropriate anti infectives, as well as followup and detection of infections (HIV, Hepatitis B and C) if warranted. 14. Develop a detailed plan to prevent similar failures in the future 15. Write after action report. RECOMMENDATIONS FOR QUALITY CONTROL Provide comprehensive and intensive training for all staff assigned to reprocess medical/surgical instruments To achieve and maintain competency: Staff receive hands on training Work with supervision until competency is documented Competency testing should be conducted at commencement of employment and regularly Review written reprocessing instructions to ensure compliance 21

RECOMMENDATIONS FOR QUALITY CONTROL Conduct infection control rounds periodically Establish a maintenance contract and record of service. Ensure protocols equivalent to guidelines from professional organizations Consult Association for the Advancement of Medical Instrumentation (AAMI) and/or manufacturer for preparation and packing of items RESOURCES 22

Thank You 23