Oxygen Demand, Chemical

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

USEPA 1 Periodate Oxidation Method 2 Method to 20.0 mg/l Mn (HR) Powder Pillows

Phosphorus, Reactive (Orthophosphate)

Oxygen Demand, Chemical

Oxygen Demand, Chemical

Scope and application: For water, wastewater, seawater, brine solutions, produced waters and hydraulic fracturing waters.

Nitrogen, Simplified TKN (s-tkn )

USEPA 1,2 Bicinchoninate Method 3 Method 8506 (CuVer 1) and Method 8026 (CuVer 2) 0.04 to 5.00 mg/l Cu Powder Pillows or AccuVac Ampuls

Sulfate. USEPA 1 SulfaVer 4 Method 2 Method to 70 mg/l SO 4. Powder Pillows. Test preparation. Before starting.

1-(2-Pyridylazo)-2-Naphthol PAN Method 1 Method to mg/l Mn (LR) Powder Pillows

Color, True and Apparent, LR

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement

DR/4000 PROCEDURE. OXYGEN DEMAND, Chemical. Colorimetric Measurement, 0 to 40 mg/l COD

Hardness, Calcium/Magnesium

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Dimethylphenol Method Method LR (0.23 to mg/l NO 3 N or 1.00 to mg/l NO 3 ) TNTplus 835

Instrument Adapters Light shield DR 6000, DR 5000 DR 3900 LZV849 DR 3800, DR 2800 LZV646 DR or (A)

Calcium and Magnesium; Chlorophosphonazo Rapid Liquid Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Pour-Thru Cell

Method (0.1 to 8.0 mg/l Cu) TNTplus 860

Scope and application: For water and wastewater. Turbid samples are filtered during the analysis.

Phosphorus, Reactive (Orthophosphate)

3. Add 0.4 ml of. 7. Use a TenSette

Nitrogen, Simplified TKN (s-tkn ) DOC

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

3. Add 0.4 ml of. 7. Use a TenSette

Method 8017 (0.7 to 80.0 µg/l) Powder Pillows. Scope and Application: For water and wastewater; digestion is required to determine total cadmium.

DR/4000 PROCEDURE. IRON, Ferrous. Using Powder Pillows

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Phosphorus, Reactive (Orthophosphate)

Table 163 Instrument-specific information. Instrument Pour-thru Kit Cell orientation Adapter. DR 6000 LQV Arrow faces right

Method to 500 µg/l CH 2 O Powder Pillows

PHOSPHORUS, REACTIVE (0.00 to 5.00 mg/l PO 4

Sulfate DOC SulfaVer 4 Method 1. Method (0 70, and mg/l) Powder Pillows. Test procedure

Direct ISE Method Method to 4.00 mg/l NO 3 N Nitrate ISE

Direct ISE Method Method to 10.0 mg/l F Fluoride ISE

3. Fill a 50-mL. 2. Press: 1 ENTER The display will show mg/l, Al and the ZERO icon.

Amperometric Titrator assembly. Digital Titrator. Beaker, low-form, 250-mL. Stir bar, octagonal, Teflon-coated, 50.8 x 7.9 mm

PaSEC Colorimeter Instructions DR850

1,5-Diphenylcarbohydrazide Method* (Powder Pillows or AccuVac Ampuls) USEPA accepted for wastewater analyses** Using Powder Pillows

Oxygen Demand, Biochemical

TPTZ Method * (Powder Pillows or AccuVac Ampuls) Using Powder Pillows

USEPA Membrane Filtration Method Method m-tec. Scope and Application: For potable water, nonpotable water, recreation water and wastewater.

FerroVer Method (Powder Pillows or AccuVac Ampuls) USEPA approved for reporting wastewater analysis (digestion is required; see Section 2 * )

STP (off-line) Water Technologies & Solutions analytical procedure. colorimetric method DCA reagent 1 (5/5/1) 2 to 15 mg/l as STP

DR/4000 PROCEDURE OXYGEN SCAVENGERS. 4. Fill a second sample cell (the blank) with 25 ml of deionized water.

SOP-C-102 Determination of Chemical Oxygen Demand

Application Notes for COD Analysis DETERMINATION OF CHEMICAL OXYGEN DEMAND (COD) IN WATER AND WASTE WATER.

Heterotrophic Bacteria

Why Hach Reagents? Alphabetical Listing of Reagents

Coliforms, Total and E. coli

Ground Water & UST Samples: Containers, Preservation and Hold Times Table

CEL212 Environmental Engineering

SOP_Chem_001 Chemical Analysis_ Chemical Oxygen Demand. Standard Operation Procedure Chemical Oxygen Demand Closed Reflux, Titrimetric Method

NACA Laboratory Equipment and Supplies

Approved for NPDES (Editorial Revision 1978) Silica, Dissolved (Colorimetric)

Analysis of dissolved nitrite (NO 2- )

1,10 Phenanthroline Method * (Powder Pillows or AccuVac Ampuls) Using Powder Pillows

Surface Water Samples: Containers, Preservation and Hold Times Table

Technical Process Bulletin

Auburn University Department of Civil Engineering Introduction to Environmental Engineering CIVL Laboratory #6 Chemical Oxygen Demand (COD)

METHOD #: Approved for NPDES (Issued 1971) Oxygen, Dissolved (Modified Winkler, Full-Bottle Technique) ANALYTE: CAS # O Oxygen

AUTOMATED WATER ANALYSIS

SUMMARY OF STANDARDS FOR WATER ANALYSIS ON COURSE WAT-E2120 SPRING 2017

3. Fill a 50-mL. 2. Press: 1 ENTER The display will show mg/l, Al and the ZERO icon.

DR 1900 PORTABLE SPECTROPHOTOMETER

Sampling Day Activities

METHOD 7550 OSMIUM (ATOMIC ABSORPTION, DIRECT ASPIRATION)

Municipal Wastewater. Introduction: Sample Matrix: Wastewater; various locations along the sewage treatment process.

BONDERITE NT-1 Phosphate-free, Regulated Heavy Metal-Free Pre-treatment

OBSERVATIONS OF AN OPERATOR IN THE LAB IMPROVING LABORATORY RESULTS. Shane Jordan-Hill. Merck Millipore

Professional Boiler Water Test Kit

Standard Methods for the Examination of Water and Wastewater

Most Probable Number (MPN) & Biological Oxygen Demand (BOD)

ENGI Environmental Laboratory. Lab #2. Solids Determination. Faculty of Engineering & Applied Science

Human CRP Precoated ELISA kit

METHOD #: Approved for NPDES and SDWA (Issued 1974) Mercury (Manual Cold Vapor Technique) ANALYTE: CAS # Hg Mercury

Human Myostatin, ELISA Kit (MSTN)

Cuvette Test LCK 555 BOD 5 /BOD [n]

Orono Spectral Solutions, Inc.

METHOD DETERMINATION OF TURBIDITY BY NEPHELOMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division

Ti-Pure TITANIUM DIOXIDE. CONCENTRATED ACID SOLUBILITY OF TiO 2 METHOD: T WP

Human Anti-Muscarinic Cholinergic Receptor 1 (M1)-Antibodies

Suspended Metals Digestion Procedure

Uranium Testing and Site Calibration for Water Testing Detection Range: 2 60ppb

SurTec 609 EC Pre-Treatment prior to E-Coat

Human breast cancer susceptibility protein 2, BRCA-2 ELISA Kit

Rev Experiment 10

Cadmium Testing and On Site Calibration for Water Testing Detection Range: ppm

ACRP. Fact Sheets. On-Site Monitoring Methods AIRPORT COOPERATIVE RESEARCH PROGRAM. Sponsored by the Federal Aviation Administration

CEE 311 Environmental Engineering I Fall 2015 Dr. Jagadish Torlapati BIOCHEMICAL OXYGEN DEMAND (BOD)

Measuring Dissolved Oxyg Instructions for the La Motte Field-Test Kit

BIOTECTOR TOC Analysers

Cuvette Test LCK 555 BOD 5 /BOD [n]

Procine CRP ELISA Kit

Application Note AN1501

CHEMetrics, Inc. Indigo Ozone Vacu-vials Technical Data Summary

Engineered Systems & Designs TURBIDITY METER

:00 am 6:00 pm Mon-Thu (MT) 6:00 am 5:00 pm Fri (MT)

Transcription:

Oxygen Demand, Chemical DOC316.53.01101 Manganese III Reactor Digestion Method (with chloride removal) Method 10067 30 to 1000 mg/l COD Mn Test N Tube Vials Scope and application: For water and wastewater Test preparation Instrument-specific information Table 1 shows all of the instruments that have the program for this test. The table also shows adapter and light shield requirements for the instruments that use them. To use the table, select an instrument, then read across to find the applicable information for this test. Table 1 Instrument-specific information for test tubes Instrument Adapters Light shield DR 6000, DR 5000 DR 3900 LZV849 DR 3800, DR 2800, DR 2700 LZV646 DR 1900 9609900 (D 1 ) DR 900 4846400 Cover supplied with the instrument Before starting Install the instrument cap on the DR 900 cell holder before ZERO or READ is pushed. DR 3900, DR 3800, DR 2800 and DR 2700: Install the light shield in Cell Compartment #2 before this test is started. To find if the sample contains chloride, use Quantab Titrator Strips for low range chloride. If the sample COD is expected to exceed 1000 mg/l, dilute the sample. Refer to Multiplication factors for sample dilutions on page 6. Run one blank with each set of samples. Run all tests (the samples and the blank) with the same lot of vials. The lot number is on the container label. The reagent blank vial can be used for multiple tests. Fill a clean COD vial with deionized water and use this vial to zero the instrument, then measure the absorbance of the reagent blank vial. The absorbance value should be approximately 1.41 1.47. Prepare a new reagent blank vial when the absorbance is outside of this range. To oxidize resistant organics, samples can be digested for up to four hours. Digest the blank for the same time period as the samples. Make sure that the filter disc is not in the center of the vial during the Zero and Read steps. Make sure that the filter disc is more than 20 mm (0.8 in.) or less than 10 mm (0.4 in.) from the bottom of the vial. If necessary, move the filter disk by gently swirling or by lightly tapping the vial on a table top. The Chloride Removal Cartridge can be used only once. If the sample boils during the digestion, the vial is not properly sealed. Test results will be invalid. Spilled reagent will affect test accuracy and is hazardous. Do not run tests with spilled vials. The maximum range of the VPD gauge is 40 inches of water; it will not indicate the full vacuum level obtained. Full vacuum is 20 25 inches of mercury; this can be measured at the vacuum pump with a gauge calibrated for inches of mercury. 1 The D adapter is not available with all instrument versions. 1

Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective equipment. Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory agencies for further disposal information. Items to collect Description Quantity Blender 1 DRB200 Reactor 1 Forceps, extra fine 1 Light shield and adapter (For information about sample cells, adapters or light shields, refer to Instrument-specific information on page 1.) 1 Manganese III COD Reagent Vials, 20 to 1000 mg/l COD 1 Pipet, TenSette, 0.1- to 1.0-mL, with pipet tips 1 Pipet, TenSette, 10.0 to 10.0-mL, with pipet tips 1 Sulfuric Acid, concentrated ACS 1 ml Test tube rack 1 Vacuum pretreatment device 1 Vacuum pump 1 Vial, glass, for sample and acid 2 Water, deionized varies Refer to Consumables and replacement items on page 7 for order information. Sample collection and storage Collect samples in clean glass bottles. Use plastic bottles only if they are known to be free of organic contamination. Test biologically active samples as soon as possible. Homogenize samples that contain solids to get a representative sample. To preserve samples for later analysis, adjust the sample ph to less than 2 with concentrated sulfuric acid (approximately 2 ml per liter). No acid addition is necessary if the sample is tested immediately. Keep the preserved samples at 2 6 C (36 43 F) for a maximum of 28 days. Correct the test result for the dilution caused by the volume additions. 2 Oxygen Demand, Chemical, Manganese III Method with chloride removal (1000 mg/l)

Acidified sample preparation 1. Set the DRB200 Reactor power to on. Preheat to 150 C or set to the COD program. 2. Put 100 ml of sample in a blender. Blend for 30 seconds or until homogenized. If suspended solids are present, continue to mix the sample while the sample is moved to the mixing cell for the prepared sample. 3. Prepare the blank: Use a pipet to add 9.0 ml of deionized water to a glass mixing cell. 4. Prepare the sample: Use a pipet to add 9.0 ml of homogenized sample to a second glass mixing cell. 5. Use a pipet or dispenser to add 1.0 ml of concentrated Sulfuric Acid to both mixing cells. 6. Close the mixing cells tightly. Invert several times. The cells get hot during mixing. Let the mixing cells cool to room temperature. Go to the Vacuum pretreatment procedure on page 4. Oxygen Demand, Chemical, Manganese III Method with chloride removal (1000 mg/l) 3

Vacuum pretreatment procedure 1. Attach the vacuum pretreatment device (VPD) to a vacuum pump that can create a vacuum of 20 25 inches of mercury. Do not use an aspirator type vacuum. 2. Write a sample identifier on each vial. Insert the Mn III COD vials in the numbered holes in the VPD base. Remove the caps from the vials. 3. Put the VPD top on the base. Insert a fresh Chloride Removal Cartridge (CRC) directly above each Mn III COD Reagent Vial. Close any open holes in the VPD with the supplied stoppers. 4. Start the vacuum pump. Adjust the vacuum regulator valve on top of the VPD until the internal gauge reads 20 inches of water. 5. Prepare the blank: Use a pipet to add 0.60 ml of acidified blank into the CRC. It should take 30 45 seconds to pull the liquid through the CRC into the vial. Note: If the liquid does not flow through the CRC, increase the vacuum until the flow starts, then reduce the vacuum back to 20 inches of water. 6. Prepare the sample: Use a pipet to add 0.60 ml of each acidified sample into the CRC. 7. Close the vacuum regulator valve completely to achieve full vacuum. After one minute of full vacuum, move the VPD back and forth several times to remove drops that cling to the CRC. 8. Open the VPD regulator to release the vacuum. Turn the pump off. Remove the VPD top and set it aside. Dispose of the used Chloride Removal Cartridges. Go to Sample preparation and measurement on page 5. 4 Oxygen Demand, Chemical, Manganese III Method with chloride removal (1000 mg/l)

Sample preparation and measurement 1. Use forceps to remove the filter from the top of the CRC. If the sample does not contain suspended solids, it is not necessary to transfer the filter to the digestion vial. 2. Insert each filter into the corresponding Mn III COD vial. Use the numbers on the VPD as a guide. Use a clean towel or deionized water to clean the forceps between samples. 3. Remove the Mn III COD vial from the vacuum chamber. Put the original caps on and close tightly. Invert the vials several times to mix. 4. Insert the vials in the DRB200 Reactor at 150 C. Close the cover. Digest the samples for one hour. Start 5. After an hour, remove the vials from the DRB200. Let the vials cool in a rack for two minutes. If the solution develops a colorless upper layer and a purple lower layer, invert the vials several times to mix. 6. Let the vials cool to room temperature in a cool water bath, or hold under running tap water for several minutes. 7. Start program 432 COD Mn III. For information about sample cells, adapters or light shields, refer to Instrument-specific information on page 1. 8. Clean the blank vial. Zero 9. Insert the blank vial into the 16 mm cell holder. Make sure that the filter disc does not block the instrument light beam. Refer to Before starting on page 1. 10. Push ZERO. The display shows 0 mg/l COD Mn. 11. Clean the sample vial. 12. Insert the sample vial into the 16 mm cell holder. Make sure that the filter disc does not block the instrument light beam. Refer to Before starting on page 1. Oxygen Demand, Chemical, Manganese III Method with chloride removal (1000 mg/l) 5

Read 13. Push READ. Results show in mg/l COD Mn. Multiplication factors for sample dilutions If the sample COD is expected to exceed 1000 mg/l, dilute the sample as shown in Table 2. For other dilutions not shown, add the sample volume and the deionized water and divide by the sample volume to obtain the multiplication factor. All dilutions require that the ratio of sample to sulfuric acid remain at 9:1. Note: Mixing concentrated sulfuric acid and water is not additive. Adding 1.0 ml of concentrated sulfuric acid to 9.0 ml of sample does not result in a final volume of 10.0 ml. This factor is built into the calibration curve. Table 2 Multiplication factors Sample (ml) Deionized water (ml) Range (mg/l COD) Multiplication factor 6.0 3.0 30 1500 1.5 3.0 6.0 60 3000 3 1.0 8.0 180 9000 9 0.5 8.5 360 18,000 18 Interferences For best results, use a minimum of 0.5 ml sample for the dilution. If the sample values exceed 18,000 mg/l COD, use a separate sample dilution, then start the sample chloride removal procedure. Example: Dilute the sample to a range of 90 4500 mg/l COD. Sample Volume (2.0 ml) + Deionized water (7.0 ml) = Total Volume (9.0 ml) Multiplication factor = (total volume)/(sample volume) = 9.0 ml/2.0 ml = 4.5 Standard test range is 50 to 1000 mg/l COD. Example test range = 4.5(50) to 4.5(1000) = 225 to 4500 mg/l COD Inorganic materials can also be oxidized by trivalent manganese and cause a positive interference when present in significant concentrations. Chloride is the most common interference and is removed by sample pretreatment with the Chloride Removal Cartridge. Refer to the Oxygen Demand, Chemical procedure that uses the Manganese III Reactor Digestion Method (with chloride removal). If chloride is known to be absent or present in insignificant concentrations, the pretreatment is not necessary. A simple way to determine if chloride has an effect on the test results is to run routine samples with and without the chloride removal, then compare results. Other inorganic interferences (e.g., nitrite, ferrous iron, sulfide) are not usually present in significant amounts. If necessary, determine the concentration of these interferences with separate methods, then adjust the final COD test results accordingly. Ammonia nitrogen is known to interfere in the presence of chloride. Ammonia nitrogen does not interfere if chloride is absent. 6 Oxygen Demand, Chemical, Manganese III Method with chloride removal (1000 mg/l)

Accuracy check Standard solution method Use the standard solution method to validate the test procedure, the reagents and the instrument. Items to collect: COD standard solution, 800 mg/l (use 0.60 ml in place of the sample) 1. Use the test procedure to measure the concentration of the standard solution. 2. Compare the expected result to the actual result. Note: The factory calibration can be adjusted slightly with the standard adjust option so that the instrument shows the expected value of the standard solution. The adjusted calibration is then used for all test results. This adjustment can increase the test accuracy when there are small variations in the reagents or instruments. Method performance The method performance data that follows was derived from laboratory tests that were measured on a spectrophotometer during ideal test conditions. Users can get different results under different test conditions. Program Standard Precision (95% confidence interval) Sensitivity Concentration change per 0.010 Abs change 432 600 mg/l COD 576 624 mg/l COD 8 mg/l COD Estimated detection limit The EDL for program 432 is 4 mg/l COD. The EDL is the calculated lowest average concentration in a deionized water matrix that is different from zero with a 99% level of confidence. Summary of method Chemical Oxygen Demand (COD) is defined as a measure of the oxygen equivalent of the organic matter content of a sample that is susceptible to oxidation by a strong chemical oxidant (APHA Standard Methods, 19th ed., 1995). Trivalent manganese is a strong, non-carcinogenic chemical oxidant that changes quantitatively from purple to colorless when it reacts with organic matter. It typically oxidizes about 80% of the organic compounds. Studies have shown that the reactions are highly reproducible and test results correlate closely to Biochemical Oxygen Demand (BOD) values and hexavalent chromium COD tests. None of the oxygen demand tests provide 100% oxidation of all organic compounds. A calibration is provided which is based on the oxidation of Potassium Acid Phthalate (KHP). A different response may be seen in analyzing various wastewaters. The KHP calibration is adequate for most applications. The highest degree of accuracy is obtained when test results are correlated to a standard reference method such as BOD or one of the chromium COD methods. Special waste streams or classes will require a separate calibration to obtain a direct mg/l COD reading or to generate a correction factor for the precalibrated KHP response. The sample digestion time can be extended up to four hours for samples that are difficult to oxidize. Test results are measured at 510 nm in spectrophotometers and 520 nm in colorimeters. Consumables and replacement items Required reagents Description Quantity/test Unit Item no. Manganese III COD Reagent Vials 1 25/pkg 2623425 Chloride Removal Cartridge (CRC) 1 25/pkg 2661825 Oxygen Demand, Chemical, Manganese III Method with chloride removal (1000 mg/l) 7

Consumables and replacement items (continued) Description Quantity/test Unit Item no. Sulfuric Acid, concentrated, ACS 75 ml 2.5 L 97909 Water, deionized varies 4 L 27256 Required apparatus Description Quantity/test Unit Item no. Blender, 2-speed, 120 VAC option 1 each 2616100 Cap, with inert Teflon liner, for mixing bottle varies 12/pkg 2401812 Cell, mixing 2 each 2427700 DRB 200 Reactor, 110 VAC option, 15 x 16-mm wells 1 each LTV082.53.40001 DRB 200 Reactor, 220 VAC option, 15 x 16-mm wells 1 each LTV082.52.40001 Forceps, extra fine point 1 each 2669600 Pipet, TenSette, 1.0 10.0 ml 1 each 1970010 Pipet tips, for TenSette Pipet, 1.0 10.0 ml varies 50/pkg 2199796 Pipet, TenSette, 0.1 1.0 ml 1 each 1970001 Pipet tips, for TenSette Pipet, 0.1 1.0 ml 2 50/pkg 2185696 Test tube rack 1 each 1864100 Vacuum Pretreatment Device (VPD) 2 1 each Vacuum Pump, 1.2 CFM 115 V 1 each 2824800 Recommended standards Description Unit Item no. COD Standard Solution, 800-mg/L 200 ml 2672629 Oxygen Demand Standard (BOD, COD, TOC), 10-mL ampules 16/pkg 2833510 Potassium Acid Phthalate (KHP), ACS 500 g 31534 Wastewater Influent Standard Solution, Mixed Parameter, for NH 3 -N, NO 3 -N, PO 4, COD, SO 4, TOC 500 ml 2833149 Optional reagents and apparatus Description Unit Item no. Dispenser, automatic, 1.0 5.0 ml each 2563137 Titrator Strips, Quantab, for low range chloride 40 tests 2744940 Finger cots 2/pkg 1464702 Paper, ph, 0 14 ph range 100/pkg 2601300 COD Standard Solution, 300-mg/L 200 ml 1218629 COD Standard Solution, 300-mg/L 500mL 1218649 COD Standard Solution, 1000-mg/L 200 ml 2253929 Standard Methods for the Examination of Water and Wastewater (current edition) each 2270800 2 The Vacuum Pretreatment Device is obsolete. There is no replacement. 8 Oxygen Demand, Chemical, Manganese III Method with chloride removal (1000 mg/l)

Optional reagents and apparatus (continued) Description Unit Item no. Pipet tips for TenSette Pipet, 1.0 10.0 ml 250/pkg 2199725 Pipet tips for TenSette Pipet, 0.1 1.0 ml 1000/pkg 2185628 Oxygen Demand, Chemical, Manganese III Method with chloride removal (1000 mg/l) 9

FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free 800-227-4224 Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web www.hach.com; E-mail techhelp@hach.com HACH COMPANY WORLD HEADQUARTERS Telephone: (970) 669-3050 FAX: (970) 669-2932 Hach Company/Hach Lange GmbH, 1989 2016. All rights reserved. 12/2016, Edition 10