Mechanical Characterization of Sol-Gel Coatings Using a Nano Indenter G200

Similar documents
Indentation Rules of Thumb

High Temperature Mechanical Characterization of Tin (Sn) Using Nanoindentation

Evaluation of Failure in Low-k Films Using Stiffness Mapping and Dynamic Imaging

Keysight Technologies Indentation Rules of Thumb Applications and Limits

Keysight T150 UTM. Universal Testing Machine. Data Sheet

Keysight Technologies Laser Heated Tip and High Temperature Stage for G200 Nano Indenter. Data Sheet

Keysight Technologies Nanoindentation, Scratch, and Elevated Temperature Testing of Cellulose and PMMA Films. Application Note

Nanoindentation and nanoscratch behaviors of DLC coatings on different steel substrates

ISO Standardized Nanoindentation

Tennessee State University College of Engineering and Technology Department of Mechanical Engineering

Keysight Technologies High-Temperature Nanoindentation on Pure Titanium (Ti) Application Note

CONTROLLED HUMIDITY NANOINDENTATION OF POLYMER FILMS

NANOINDENTATION CREEP MEASUREMENT

HONEYCOMB MECHANICAL BEHAVIOR USING MACROINDENTATION

AMMONITE FOSSIL MECHANICAL PROPERTIES USING NANOINDENTATION

Nanoindentation Behaviour of Mg Based Hybrid Composites with Graphite nanofiber/alumina Fiber

M3 Review Automated Nanoindentation

YIELD & TENSILE STRENGTH OF STEEL & ALUMINIUM USING MICROINDENTATION

ELASTIC-PLASTIC RESPONSE OF POLYMER COATING UNDER REPETITIVE IMPACT

Scratch Testers. Coating Adhesion, Scratch and Mar Resistance. ::: World Leader in Scratch Testing

Keysight Technologies Scratch Testing of Multilayered Metallic Film Stacks. Application Note

Dynamic Nanoindentation Characterization: nanodma III. Ude Hangen, Ph.D

Available online at ScienceDirect. Procedia Engineering 79 (2014 )

Evaluation of Mechanical Properties of Hard Coatings

NANOINDENTATION OF SILICON CARBIDE WAFER COATINGS

Nanoindentation. Kathy Walsh, Ph.D. Frederick Seitz Materials Research Laboratory University of Illinois at Urbana-Champaign

Center for Tribology, Inc. COMPREHENSIVE MATERIALS TESTING FOR MECHANICAL AND TRIBOLOGICAL PROPERTIES

Agilent 8490G Coaxial Attenuators

Nano Puncture Resistance Using Nanoindentation

Small-Scale Mechanical Testing at Michigan Technological University Houghton, MI

INTEGRATED TRIBO-SPM FOR NANO-TRIBOLOGY

CORROSION HARDNESS EFFECT OF STEEL SCREWS BY NANOINDENTATION

NanoForce -short note - Ilja Hermann October 24,

CB x 33 x 70 cm

INSTRUMENTED INDENTATION TEST FOR HARDNESS AND MATERIALS PARAMETER FROM MILLINEWTONS TO KILONEWTONS

SUPER HARDENING OF W/NbN NANOLAYERS UNDER SHALLOW NANOINDENTATION. Brian Michael Ennis. BSME, University of Pittsburgh, 2002

Mechanical properties of optical glass fibers damaged by nanoindentation and water ageing

Full Nanomechanical Characterization of Ultra-Thin Films

Doctoral School on Engineering Sciences Università Politecnica delle Marche

Vickers Berkovich Knoop Conical Rockwell Spherical Figure 15 a variety of different indenter's shapes and sizes

EDGE CHIPPING RESISTANCE USING MACROINDENTATION TESTING

Installation, Commissioning and Acceptance Testing of a High Temperature Ultra Nanoindentation Tester

INDIAN INSTITUTE OF TECHNOLOGY BOMBAY MATERIALS MANAGEMENT DIVISION

Application of nanoindentation technique in microelectronics

Guidelines for high temperature nanoindentation

MICROINDENTATION & FRACTURE OF MINERAL ROCK

Characterization of acoustic emission signals from particulate filled thermoset and thermoplastic. polymeric coatings in four point bend tests

XPM: High Speed Nanoindentation and Mechanical Property Mapping. Eric Hintsala, Ph.D

Nanoindentation Study on Tourmaline

STRUCTURAL AND MECHANICAL CHARACTERIZATION OF SPARK PLASMA SINTERED TUNGSTEN

Agilent Beam Lead Attachment Methods. Application Note. Introduction. Attachment Methods

Substrate Effect, Pile-Up and Sink-In Factors during Thin Film Nanoindentation. Kevin Russell Schwieker

Nanoindentation of La-Cr-O Thin Films

Interfaces friction effect of sliding contact on nanoindentation test

A review of nanoindentation continuous stiffness measurement technique and its applications

NANOMECHANICAL TESTING OF AN a-c:n NANOLAYER PREPARED BY ION BEAM ASSISTED DEPOSITION ON TI6AL4V ALLOY

Indentation of Coatings at micro/nano scale: Crack formation and deflection

Application of Atomic Force Microscopy (AFM) in Polymer Materials

3. MECHANICAL PROPERTIES OF STRUCTURAL MATERIALS

Tribomechanical Properties of DLC Coatings Deposited by Magnetron Sputtering on Metallic and Insulating Substrates

Using nano-indentation and microscopy to obtain mechanical properties

Multilayered Systems of Nanostructured TiB 2 Thin Films and its Mechanical Properties

Application Note #1007 All-Inclusive Scratch Testing of Coatings and Thin Films Using Bruker's Universal Test System

Agilent Engineering Services for PNA Series MW Network Analyzers

AN ABSTRACT OF THE THESIS OF

Title: COMPLEX STUDY OF MECHANICAL PROPERTIES OF a-si:h AND a-sic:h BORON DOPED FILMS

Annealing effects on microstructure and mechanical properties of chromium oxide coatings

Hysitron TI 980 TriboIndenter. Innovation with Integrity. Tribology and Mechanical Testing

Hysitron TI 980 TriboIndenter. Innovation with Integrity. Tribology and Mechanical Testing

UMT TriboLab TM for Automotive Applications Suresh Kuiry, Ph.D.

Noise Figure Analyzers

FORCES IN BIOLOGY NANOINDENTER

Agilent N4420B S-Parameter Test Set

Mechanical Nanocharacterization of Gold for MEMS Contact Switches

Agilent Series Portable Logic Analyzers

Nano-indentation Study of the Surface of. Ion-Exchanged Lithium Silicate Glass

Ceramic Processing Research

CSM Nano-Tribometer *

Nanoindentation, Scratch and nanodma : Innovations for Atomic Force Microscopes. Ryan Stromberg

Agilent CSA Spectrum Analyzer N1996A

Tablet Coating Failure Using Micro Scratch Testing

Nanoindentation stress-strain for Fracture Analysis and computational modeling for hardness and modulus

NATURAL ORGANIC-INORGANIC COMPOSITE SYSTEMS. Aleš PETRÁŇ, Dora KROISOVÁ

Combination of Instrumented Nanoindentation and Scanning Probe Microscopy for Adequate Mechanical Surface Testing

High strain point glass substrate for photovoltaic solar cell

SURFACE PROTECTION. ForceField 2700 Series Low gel surface protection film Coextruded adhesive based product GENERAL PRODUCT INFORMATION:

NANO SCRATCH TESTING OF THIN FILM ON GLASS SUBSTRATE

EFFECT OF CRYSTALLOGRAPHIC ORIENTATION ON MECHANICAL PROPERTIES OF STEEL SHEETS BY DEPTH SENSING INDENTATION

UV15: For Fabrication of Polymer Optical Waveguides

Supplementary Figures

Chapter 6: Mechanical Properties: Part One

Techniques to Improve Coating Adhesion of Superhard Coatings

Analyzing Time on Sample During Nanoindentation

CHARACTERISATION OF INTERFACIAL CRACKING IN MICROELECTRONIC PACKAGING

Direct Measurement of the Nanoscale Mechanical Properties of NiTi Shape Memory Alloy

Boost Your Yield Get more out of Inspection Inspection-Systems for Plastic Film and Sheets 100% Optical Web Inspection.

Mechanical Behavior and Characterization of Stern-shaft Mechanical Sealing Device. Yongjin Lu a *, Rui Lin

Assessment of the Nanohardness of a-sic: H Film by Cyclic Nanoindentation

Your Supplier for Fused Silica

Repetition: Adhesion Mechanisms

Transcription:

Mechanical Characterization of Sol-Gel Coatings Using a Nano Indenter G200 Application Note Jennifer Hay Agilent Technologies Introduction This application note presents the results of nanomechanical tests in which an Agilent Nano Indenter G200 was used to measure the Young s modulus and hardness of three sol-gel coatings. The same instrument was also utilized to perform controlled scratch tests to evaluate adhesion. The samples provided for testing are described in Table 1. Indentation Tests Twenty indentation tests were performed on each sample using a Nano Indenter G200 (XP head) fitted with a Berkovich indenter tip. The Agilent Continuous Stiffness Measurement (CSM) option was used to measure properties as a continuous function of penetration depth. Figure 1 shows a typical load-time history for a single test. Each indentation test consists of the following test segments: 1. Approach the surface until contact is detected. 2. Load such that the loading rate divided by the load remains constant at 0.5/sec until the penetration depth reaches 3000 nm (3 µm). 3. Hold the peak force constant for 10 seconds. 4. Reduce the force on the indenter until the applied force becomes 10% of the peak force. 5. Hold the force on the indenter constant for 100 seconds. 6. Withdraw the indenter completely. Reference AG5T5R82 AG5T5R111 VR82 Description 50% organic, 50% inorganic 50% organic, 50% inorganic 50% organic, 50% inorganic Thickness (µm) Substrate 6±1 aluminum 12±1 aluminum 6±1 glass Table 1. Samples for testing. Figure 1. Example of load-time history for a single indentation test. (Note: since loading terminated at a prescribed displacement, peak forces varied from test to test.)

Figure 2. Exemplary data analysis: Young s modulus for this test on this film is calculated as the average of all values between the markers M and N. The positions of markers M and N are controlled by the user. The loading process (P /P = 0.05/sec) causes a constant strain rate to be applied to the test material. When testing materials for which the measured hardness depends on strain rate, this loading process must be used. Let us consider a bulk sample (not a thin film) of strain-rate-sensitive material. Only a loading process that holds P /P constant will return a hardness that is constant with penetration depth. Indentation tests that hold the loading rate or displacement rate constant will measure a hardness that appears to decrease with indentation depth, even on a bulk sample with uniform properties. This is because the imposed strain rate is decreasing. 1 The fifth test segment is for the purpose of evaluating the change in displacement due to thermal expansion and/or contraction of the sample and/or instrument. Data acquired in the last 50 seconds of this test segment are used to calculate a thermal drift rate, which is then used to make a small correction to displacements measured throughout the test. This common procedure is explained in more detail elsewhere. 2 The CSM option works by imposing a small oscillation on the indenter during loading. Typically, the amplitude of the force oscillation is controlled (increased) such that the amplitude of the displacement oscillation remains constant at 2 nm. The amplitude of the force oscillation relative to the amplitude of the displacement oscillation allows one to calculate stiffness as a continuous function of penetration depth. This, in turn, allows the calculation of Young s modulus and hardness as a continuous function of indentation depth. 3 Therefore, each indentation test returns Young s modulus and hardness as a function of indentation depth. The properties of the film alone are calculated by averaging over an appropriate displacement range for these films, the range was set to 150 200 nm. For example, the measured modulus from the first test on sample AG5T5R82 is shown in Figure 2. The value of Young s modulus for this test, E1, is reported as the average of all data between the markers M and N. The same procedure is repeated for every test to get E2, E3, etc. Finally, we calculate an average and standard deviation for all n valid tests to get 2

Figure 3. Applied force as a function of scratch distance for a single scratch test. To simplify the presentation of properties as a function of penetration depth (and to facilitate comparison among samples), we also average the channels of Young s modulus and hardness into small, discrete displacement windows. For example, for sample AG5T5R82, we take all data from all valid tests for which the penetration depth is between 0 nm and 5 nm and report a single average and standard deviation for this window. Then we take all data from all valid tests for which the penetration depth is between 5 nm and 10 nm and report an average and standard deviation for this window. And so on. The traces in Figure 6 were generated in this way. Scratch Tests Six scratch tests were performed on each sample. Lateral forces were not measured, so the Agilent Lateral Force Measurement (LFM) option is not necessary for duplication of this work, although it may be advantageous for future testing. Figure 3 shows the applied force as a function of scratch position. Each scratch test consists of the following test segments: 1. A first profile of the surface is realized under a very small load (100 µn), in order to have the original morphology of the surface before the scratch. 2. Then, along the same path, the normal load is increased from 0 mn to 150 mn over a distance of 200 microns. 3. A last profile is realized under a very small load (100 µn) to measure the residual deformation in the groove. If the scratch causes failure in the material, the critical load is calculated as the load corresponding to the position at which the final scan reveals a significant increase in roughness. In other words, if the final scan (segment 3) reveals a significant increase in roughness at a position of 98 microns from the origin of the scratch, then the critical load is taken to be the load that was applied during the scratch (i.e., during segment 2) at that same position. 3

Results Indentation Table 2 summarizes the film properties measured in this work. Images of residual impressions are provided in Figure 4. Sample VR82 exhibited adhesive failure, as evidenced by a halo around the residual impression. The point of failure is clear in the indentation data as illustrated in Figure 5. This failure occurred at an applied force of 87.5 mn ± 11.8 mn. The two samples on aluminum substrates did not exhibit this failure. Sample E Average Over Defined Range σ(e) Table 2. Summary of sol-gel film properties. H Average Over Defined Range σ(h) GPa GPa GPa GPa AG5T5R82 6.24 0.18 0.515 0.014 20 AG5T5R111 5.68 0.10 0.514 0.016 16 VR82 5.66 0.07 0.442 0.009 19 n Figure 4. Residual impressions; halo around VR82 dents indicates fracture at the interface. Figure 5. Example of adhesive failure during an indentation test on VR82. Point of failure, F, was identified automatically, not manually. 4

a) Figure 6 shows the evolution of hardness and modulus as a function of penetration depth for all samples. Finally, Figure 7 shows corresponding results for the fused silica tested as a reference material immediately before and after the sol-gel samples. b) Results Scratch Sample VR82 also failed during scratch testing. The critical load at failure was 43.7 mn ± 1.2 mn. This is much lower than the load causing failure during indentation, probably because lateral forces contributed to the failure. Neither of the coatings on aluminum substrates exhibited failure during scratch testing. Figure 8 shows the residual scratches on sample VR82. Figure 6. Properties of sol-gel films as a function of penetration depth. a) b) Discussion and Conclusions Indentation tests to measure Young s modulus and hardness were prescribed to go to a maximum depth of 3000 nm (3 µm). After reviewing the resulting traces of modulus and hardness as a function of displacement, the displacement range of 150 200 nm was selected for reporting film properties. The lower limit of this range, 150 nm, was sufficiently deep to get past surface effects, whereas the upper limit was sufficiently shallow to avoid significant influence from the substrate. On future testing of similar samples, it would be better experimental practice to limit the depth of the test to something only a little greater than the upper limit of the range that will be used to report properties. Limiting the indentation depth to something like 250 nm would minimize the influence of thermal drift on the results. Figure 7. Properties of fused silica reference material, before and after testing sol-gel samples. 5

Figure 8. Residual scratches on sample VR82. Only the sol-gel film on the glass substrate (sample VR82) exhibited adhesive failure. This is probably due to the fact that the rougher aluminum substrate provided a substantially greater surface area available for bonding between the film and substrate. The diameter of the halo around the residual indents can be used to calculate interfacial fracture toughness. The fact that the normal load causing failure was much lower for scratch testing is not surprising, because lateral forces probably contribute to the failure. Although the LFM option was not used in this testing, it may be used to measure the lateral force on the indenter during scratching. We were not expecting sample VR82 to fail during indentation; however, when failure was evident, Agilent NanoSuite Explorer software was utilized to quickly define an algorithm that would automatically pick out the critical indentation load identified by point F in Figure 5. Technology and Applications The Nano Indenter G200 is powered by electromagnetic actuation to achieve unparalleled dynamic range in force and displacement. The instrument s unique design avoids lateral displacement artifacts, while software compensates fully for any drift in force. Using the G200, researchers can measure Young s modulus and hardness in compliance with ISO 14577. Deformation can be measured over six orders of magnitude (from nanometers to millimeters). Applications of the G200 include semiconductor, thin films, and MEMs (wafer applications); hard coatings and DLC films; composite materials, fibers, and polymers; metals and ceramics; and biomaterials. 6

7

Nanoindentation instruments from Agilent Technologies conform to the ISO 14577 standard, delivering confidence in test accuracy and repeatability. These state-of-the-art solutions ensure reliable, high-precision measurement of nanomechanical properties for research and industry. References 1. B.N. Lucas and W.C. Oliver, Indentation power-law creep of high-purity indium, Metall. Mater. Trans. A, 30 (3), 601-610 (1999). 2. J.L. Hay and G.M. Pharr, Instrumented Indentation Testing, book chapter in ASM Handbook Volume 8: Mechanical Testing and Evaluation (10th edition), edited by H. Kuhn and D. Medlin, pp. 232-243, ASM International, Materials Park, OH (2000). 3. W.C. Oliver and G.M. Pharr, An Improved Technique for Determining Hardness and Elastic Modulus Using Load and Displacement Sensing Indentation Experiments, Journal of Materials Research, 7, pp. 1564-1583 (1992). Nanomechanical Systems from Agilent Technologies Agilent Technologies, the premier measurement company, offers highprecision, modular nano-measurement solutions for research, industry, and education. Exceptional worldwide support is provided by experienced application scientists and technical service personnel. Agilent s leading-edge R&D laboratories ensure the continued, timely introduction and optimization of innovative, easy-to-use nanomechanical system technologies. www.agilent.com/find/nanoindenter www.agilent.com For more information on Agilent Technologies products, applications or services, please contact your local Agilent office. The complete list is available at: www.agilent.com/find/contactus Phone or Fax United States: (tel) 800 829 4444 (fax) 800 829 4433 Canada: (tel) 877 894 4414 (fax) 800 746 4866 China: (tel) 800 810 0189 (fax) 800 820 2816 Europe: (tel) 31 20 547 2111 Japan: (tel) (81) 426 56 7832 (fax) (81) 426 56 7840 Korea: (tel) (080) 769 0800 (fax) (080) 769 0900 Latin America: (tel) (305) 269 7500 Taiwan: (tel) 0800 047 866 (fax) 0800 286 331 Other Asia Pacific Countries: tm_ap@agilent.com (tel) (65) 6375 8100 (fax) (65) 6755 0042 Product specifications and descriptions in this document subject to change without notice. Agilent Technologies, Inc. 2009 Printed in USA, August 6, 2009 5990-4514EN