TRS100 Rapid Assay System Pre-Installation Guide

Similar documents
RapID Raw Materials Verification System

! CAUTION: Read and follow these instructions.

Pre-installation Manual CONCEPT 301 D/401 D/501 D

INSTALLATION MANUAL MULTI DECK HOT MERCHANDISERS MD Versions. Solid back 3 level Pass through 3 level

Pre-installation Manual. Plate Processor 68/85 Thermal

Agilent Technologies 7620-AS Sample Changer Site Preparation Checklist Hardware Site Preparation Specification

Table of contents. 1. Introduction

DJ Switch 10F V2 ORDERCODE 50365

Installation plan. Washer-extractor PW 6107 EL. en - GB M.-Nr / 01

CRON HDI-920 Site Planning. Customer Responsibilities

Maintenance Bypass Breaker Cabinet Series Configuration 2. Installation and Operation Manual

E-Series Installing E-Series U Cabinet

Pre-Installation Guide and Checklist Fast T-Jet Blazer

XPM 2. Reflow Soldering / Curing Systems. Site Preparation & Installation

Low-profile, 11U high cabinets hold networking equipment, servers, and phone equipment.

MAN488 Issue 3 WBS/FBS MAINTENANCE BYPASSES AUTOMATIC VERSION

Maintenance Bypass Breaker Cabinet Series Configuration 5. Installation and Operation Manual

Agilent Technologies SP50 Operator Mode Manual

POWER PINNER BENCH HAND WELDER 7250 OPERATOR S MANUAL

RENA XPS-3000 Base. User Guide. SC2000 Media Transport System

Gerber. Scientific Products. Dear Gerber Sabre Customer:

EXPRESS5800/1080Xd. Installation Procedures

Calibrate Accessory Actuators Instructions

Spectrum Spotlight Installation Protocol Pre-Installation Details 16 November, 2001 Page Appendix 2. Related Documents

Installation and Operation Manual

Ground Fault Protection. Medium Voltage Artificial Neutrals

Saphyr System Safety Guide for Saphyr P/N 60325

Customer Responsibilities

Pre-Installation Guide

ISSUED: OCT. 11, 2005 REVISED: JAN. 1, 2007 HOSHIZAKI REFRIGERATED GLASS MERCHANDISER KD-90D MODEL INSTRUCTION MANUAL

Installation Instructions. Overview. Important Notes. LINCTM Enclosure EDB-WPM-GRX-6-CE

XAP Series Busduct Storage, Installation, and Maintenance

Gerber M Series Site Requirements

MNL, SEL DIMENSIONS, SITE AW PLANNING GUIDE ARTWORK SIZE A SHEET 1 OF 1

This document provides the Pre-Installation Checklist for the ADVANTAGE Router 600.

XSTi Standby UPS. 400VA, 600VA, 800VA Models. User & Installation Manual

AirBench EX Operation and Maintenance

Bedhead Service Systems. Instructions for Use. July 2016 ASF-18 (0)

INDeX Rack Installation Notes

Installation. Smart-UPS VT in Parallel kva 380/400/415 V 208/220 V 200/208 V

Gerber Scientific Products

C60 Standby UPS. 800VA Models. User & Installation Manual Xtreme Power Conversion Corporation. All rights reserved.

Bravo Platform Unpacking Guide

Operational Manual. Spectrophotometer Model: SP-830 PLUS. Metertech Inc. Version 1.07

Installation plan. Tumble dryer. PT 8807 EL (Electrically heated) en - US Mat. NR / 02

64C Amplification Block. Instruction Manual

Installation Manual. MONISAFE 400A System V ( ) Nautilus Hyosung, Inc. All Rights Reserved.

SYSTEM 1E LIQUID CHEMICAL STERILANT PROCESSING SYSTEM

Product Application. Prestige Power Post in White Epoxy Coated Aluminium Finish

Symmetra PX 160 Classic Battery Cabinet

Quick-Start Operating Guide Document No Exhaust Fume Hood Copyright 2010 Terra Universal Inc. All rights reserved. Revised September 2010

Wealtec Corp. E-Centrifuge Microcentrifuge Operation Manual. Version 3.0 Item # *This instrument is intended for laboratory use only

INSTALLATION MANUAL COLD DELI 3/4/5 MERCHANDISERS

SECTION SECONDARY UNIT SUBSTATIONS

Operating Manual. Professional Weighing Equipment SCT SERIES Small Counting Scales. SCT 600g SCT 1200g

TECHCONNECT TC2 USBTP2 OWNERS MANUAL. TC2_USBTP2_manual_en.doc

Operation Manual. Professional Weighing Equipment. MRB Series Mid Resolution Balances. MRB 500g MRB 1200g MRB 2500g MRB 6000g MRB 10000g

Powerware 9390 Integrated Accessory Cabinet IAC-D Configuration Installation and Operation Manual. For use with kva and kva UPS Models

User Manual. Precision Balances XSR models

Quick-release pins swing the rack open from left to right to simplify installing and maintaining your equipment.

VPL-3200B Residential Vertical Platform Lift Technical Specifications

VPL-3300B Commercial Vertical Platform Lift Technical Specifications

MANUAL. ENGLISH Antenna Booster Ordercode: D1425. Highlite International B.V. Vestastraat EX Kerkrade the Netherlands

Site Preparation Guide. Scodix Ultra Press and Scodix Ultra-Pro Press and Scodix Ultra Foil Station

D-Series Warming Cabinet Unpacking and Setup

AP5K-C Precision AC Double-Pulse Spot Welding Machine User s Manual Shenzhen Will-Best Electronics Co., Ltd

Quick-Start Operating Guide Document No Process Gas Cooler Copyright 2010 Terra Universal Inc. All rights reserved. Revised October 2010

228 EXPRESS CASH RECYCLER 8, 10 OR 12 ROLLED STORAGE MODULES IN A UL SAFE (FRONT LOAD) PERSPECTIVE PLAN VIEW PAGE 1 OF 5

Agilent 2100 Bioanalyzer. Installation and Safety Guide

Continuous Wave Single Frequency Lasers. SCHERZO Series. 532nm 300mW

Professional Weighing Equipment. K-Series Scales. Operation Manual KHR 6000 KHR 123 KHR 502 KHR 3001

MD Power Supply -48Vdc +Bias Tee

PD6770 VANTAGEVIEW LOOP-POWERED PROCESS METER

Kelley HULK Dock Lifts and Ergonomic Lift Tables

Customer Responsibilities

Continuous Wave Single Frequency Lasers. SYMPHONY Series. 532nm 2000mW

Quick Start Guide. Corporation Rev. 1

Galaxy VM kva 400 V Receiving and Unpacking 06/

Pre-Installation Guide DWL 66 66FS. UV 355nm 250mW (Genesis)

EPS Receiving and Unpacking 06/

ISIMET. E Series Enclosure. Custom Assemblies Installation, Maintenance, and Operation Instructions

Manual Rotary Heat Sealers. Type: F108TX. List of content : Introduction 2. General description 3. Application 4. Safety precautions 5

Instruction Manual. Comet Assay Tanks Catalogue Numbers BCOM10 BCOM20 BCOM40 BCOM80

Certificate of Compliance

July P Wide Format Stacker User Guide

NEUTRAL GROUNDING TRANSFORMERS INSTALLATION AND MAINTENANCE INSTRUCTIONS

J60 Lithium Ion UPS 350VA Model User & Installation Manual Xtreme Power Conversion Corporation. All rights reserved.

Operation Manual INCUBATOR2 Dual Dry Block Incubator

Precision Range (All Models)

Site Preparation Guide. Stratasys J750 3D Printing System. DOC Rev. D1

Sensor unit MC2 for toxic gases and oxygen with analog output

HNC-120BA-L/R-S HNC-150BA-L/R-S HNC-180BA-L/R-S HNC-210BA-L/R-S

ZX70 & ZX80 SERIES PRECISION DECADE RESISTANCE BOX

INSTALLATION MANUAL AND OPERATING INSTRUCTIONS

SECTION to 5000 AMPERE PLUG-IN AND FEEDER BUSWAYS - SPECTRA. A. Low-Voltage Plug-in Busway to 5000 ampere, 600 VAC

Operation Manual. Dual Temperature Digital Dry Block Incubator. Catalog No. INCUBATOR Rev C. April 2016

LIGHT CONTROLLER LC 500

Laboratory Refrigerator Model: ,-01,-02,-03,-04,-05,-06,-07,-08, -09,-10,-11,-12,-13,-14

Transcription:

TRS100 Rapid Assay System Pre-Installation Guide NOTICE: This document contains references to Cobalt or Cobalt Light Systems. Please note that Cobalt Light Systems was purchased by Agilent in July 2017. For more information about these products and support, go to www.agilent.com/en/products/raman-spectroscopy-/ramanspectroscopy-systems 1 of 9 P a g e

Table of Contents 1. Introduction... 1 1.1. Delivery... 1 1.2. Unpacking... 2 1.3. TRS100 Mechanical Installation... 2 2. Specifications... 5 2.1. Environmental Requirements... 5 2.2. Minimum Computing Requirements (if customer supplied PC)... 5 2.3. Electrical Services... 6 2.4. Laser Safety and Room Interlocks... 6 2.5. Consumables... 6 3. Support Contact Details... 7 3.1. Contact Details... 7 List of Figures Figure 1 TRS100 Shown in Flight Case Base with Lid to Side... 1 Figure 2 TRS100 Overall Dimensions... 2 Figure 3 Maintenance and Facilities Clearances... 3 Figure 4 TRS100 on Optional Scissor Lift Trolley Shown at Maximum Height... 4 2 of 9 P a g e

1. Introduction 1.1. Delivery When the system arrives, check that there is no visible damage, with the delivery driver present. If damage has occurred contact the carrier and Cobalt immediately. Make a photographic record where appropriate. Check that shock-watch and tilt indicators, if fitted to the outside of the packing cases, have not been activated. If the indicators have been activated notify Cobalt immediately. The system is supplied as a single unit in a number of possible of packaging types depending on customer location: Overseas customers requiring airfreight; the system will be externally crated in accordance with ISPM15 UK WMMP. The case dimensions will be approximately 1.45 x 0.91 x 0.89m. The weight of the case will be approximately 240kg. UK and some European customers may have the unit delivered by the installation engineers. This will be secured and protected either with simple re-usable plastic bubble wrap and pallet wrap if mounted on the optional trolley, or within a dedicated hard flight case as shown in Figure 1. Figure 1 - TRS100 Shown in Flight Case Base with Lid to Side A forklift, or similar, will be required to off load palletised or crated shipments and position the system near to its final installation position. Note: If the system has to be unpacked before being transported to its final installation position, e.g. if it is to be installed in a clean room, a sturdy trolley, skates or other suitable mechanism will be required to move the system over a distance. WARNING The packing crates are heavy and could cause serious injury and / or damage to the equipment if not handled correctly. Use suitable lifting equipment and procedures. Only lift the packing cases from the base. 3 of 9 P a g e

1.2. Unpacking Do not remove the TRS100 from its packaging if it has been freight shipped for a minimum of 24 hours to allow the unit to acclimatise. CAUTION Do not remove the equipment from the packing crates until they have been moved to their designated installation site. The equipment has been carefully packed to protect the equipment from damage in transit. Removal of the packing equipment could make the equipment vulnerable to damage during transit. The TRS100 should only be lifted from its packaging case base using the lifting handle frames. These mount into the four M10 tapped holes on each dark grey end cover at either end of the TRS100. A four man lift is recommended to manoeuvre and carry the TRS100 into its final installation position and the installation engineer(s) may require some assistance from the customer for this operation. 1.3. TRS100 Mechanical Installation The TRS100 weighs approximately 160kg and has the dimensions shown in Figure 2. Figure 2 - TRS100 Overall Dimensions It is recommended that the TRS100 is positioned on a stable bench or table which is capable of supporting loads up to 300kg. A standard cantilever style table should not be used. Certain routine maintenance activities require the front (laser enclosure) and occasionally the top (electrical enclosure) doors to be open as shown in Figure 3. It is recommended that this space is left clear. 100mm minimum should be left at the rear of the system to provide clearance for the cooling fans and cable entry, 160mm minimum should also be left at the right of the TRS100 to allow clearance for the right air input and access to the electrical isolation switches. 4 of 9 P a g e

The left cover can be removed by a Cobalt engineer for maintenance and spectrograph / CCD calibration purposes. 250mm is the recommended minimum space to allow for removal of this cover. Figure 3 - Maintenance and Facilities Clearances The TRS100 can be supplied on an optional scissor lift trolley, the approximate dimensions are shown in Figure 4. Note: The optional trolley adds a further 60kg to the total system weight. 5 of 9 P a g e

Figure 4 - TRS100 on Optional Scissor Lift Trolley Shown at Maximum Height 6 of 9 P a g e

2. Specifications 2.1. Environmental Requirements The system should be installed in an environment as detailed here. It is essential that the climate of the laboratory is controlled to ensure the stability of the system and to ensure that the maximum range of CCD detector temperature control is available. Typically air-conditioning would be installed to maintain the temperature and humidity within the ranges listed below. Recommended ambient temperature during 21 ± 3 C operation Recommended stability of ambient temperature 1 C during operation Storage temperature >10 C <40 C Relative humidity <80 % non condensing Location Clean, dust free environment, free of substantial sources or shock and vibration >2m from air conditioning ducts or heating units Floor strength Able to bear the system weight of about 160kg (220kg including trolley). Table 1 - Environmental Requirements Note: Temperature fluctuations can affect the electrical stability of various components such as the laser. Accuracy and repeatability of measurements is achieved by: allowing the instrument a warm up period prior to use; reducing the ambient temperature fluctuations below the limits described above. Systematic fluctuations from heating and cooling ducts should be reduced by moving the TRS100 further away from these sources. 2.2. Minimum Computing Requirements (if customer supplied PC) Processor Memory Hard Disk OS i5 processor 8Gb. 64 bit 500Gb Windows 10 Pro Table 2 - Minimum Computer Requirements 7 of 9 P a g e

2.3. Electrical Services The electrical system has been designed meet BS EN 61010-1 2001 standards and checked against US regulations. All installed wiring and electrical components have been selected to comply with European standards and are UL recognized. The mains electrical input, laser, camera CCD, motion control system, DC power supply units inputs and outputs are all protected by separate magnetic circuit breakers (MCB). There are no high voltage circuits accessible in the sample measurement area, and all low voltage circuits, such as interlocks and door locks, in this region are double insulated or enclosed in trunking. Number of outlets required. 1 x single-phase outlet for the TRS100 Additional outlets may be required for computer control system and monitor Location of inlet on Rear panel towards top centre of TRS100 TRS100 Type of electrical inlet IEC C13 An IEC 10A mains lead with either a BS1363 (fused British plug), a CE22 (European plug) or CEE22 (USA plug) as required. Voltage requirements Between 220-240VAC and 110VAC 10A Current Requirements 10A maximum Line Frequency 50/60Hz Voltage fluctuation < 10 % Table 3 - Electrical Supply Requirements 2.4. Laser Safety and Room Interlocks The TRS100 contains a Class IV laser within an interlocked Class I enclosure specified by BS EN 60825-1 2007 Safety of Laser Products. A four-pin XLR socket can be found at the rear of the system; this provides one part of the override to the dual channel interlock circuit. Interlocks should only be overridden in a laser-safe environment and the TRS100 should have both interlock channels connected to the room and door interlock. If only a single channel room interlock is available the second channel can be shorted: this will not provide a level of redundancy and will cause the system to fault if only one channel is open. Note: Routine and annual maintenance does not require that the system interlocks are overridden. Repair or realignment of the system requiring interlocks to be overridden would normally require the system to be returned to Cobalt or authorised service centres unless appropriate facilities are available. 2.5. Consumables The system does not require replacement of any consumable items, however it is recommended that a supply of reagent grade Isopropyl Alcohol and/or Methanol and lens tissues is used to maintain cleanliness of the laser aperture window 8 of 9 P a g e

3. Support Contact Details 3.1. Contact Details United Kingdom Address: Agilent Technologies LDA UK Ltd. 174 Brook Drive Milton Park Abingdon Oxfordshire OX14 4SD United Kingdom Tel: +44 (0)1235 856 555 e-mail: cobalt.support@agilent.com END OF DOCUMENT 9 of 9 P a g e