Corn Syrup Analysis E-1a- 1. ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

Similar documents
ACETALDEHYDE. The procedure is applicable to corn syrups including those containing fructose, and to most other starch hydrolyzates (Note 1).

Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D*

CORESTA Recommended Method No. 57

2.1 This procedure provides steps and conditions for the determination of Vapor Space Organic Compounds in ground waters, and industrial waste waters.

MERCAPTANS, METHYL-, ETHYL-, and n-butyl- 2542

ANILINE, o-toluidine, AND NITROBENZENE 2017

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

Formulas: See Table 1 MW: Table 1 CAS: Table 1 RTECS: Table 1. METHOD: 2555, Issue 1 EVALUATION: PARTIAL Issue 1: 15 March 2003

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE.

Separation and Quantitation of a Mixture of Alcohols by Gas Chromatography

Analytical Method of Limit Test for Hexachlorobenzene. in Picloram TC and Method Validation Data*

Course Outline. in Soil/Sediment&WasteSediment&Waste by. Overview of VOCs Analysis Sample Collection VOCs Analysis

IRON (Colorimetric) 2. Muffle Furnace: Equipped with pyrometer and capable of operating at controlled temperatures up to 600 C

Residual Solvents in medical cannabis analysis using the SRI 8610C FID GC June 2014

of Methane, Ethylene, Acetylene and Ethane in Water by Headspace-Gas Chromatography (HS-GC) with

STANDARD OPERATING PROCEDURES

HYDROCARBONS, BP C 1500

NWTPH-Gx. Volatile Petroleum Products Method for Soil and Water

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis

CORESTA Recommended Method No. 63

Analysis of styrene leached from polystyrene cups using GCMS coupled with Headspace (HS) sampler

Experiment 11. DNA Composition by HPLC 1,2

METHOD 310A - DETERMINATION OF RESIDUAL HEXANE THROUGH GAS CHROMATOGRAPHY

CORESTA RECOMMENDED METHOD N 61

Certified Reference Material - Certificate of Analysis

Capillary GC Troubleshooting: a Practical Approach

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis

Corn Syrup Analysis E-43-1 LEAD (CHELATING MEMBRANE CONCENTRATION AND GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY)

Certified Reference Material - Certificate of Analysis

THIOPHANATE-METHYL in air 5606

Optimization of Volatile Petroleum Hydrocarbons Analysis by Automated Headspace Using Method Development Tools

NPOC, VOC, TC-IC and TDN using Shimadzu TOC-V Protocols. Water Quality Analysis Laboratory at the University of New Hampshire

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations

EXTRACTION OF BIOLOGICAL TISSUES FOR AROMATIC AND CHLORINATED HYDROCARBONS

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Orono Spectral Solutions, Inc.

Volatile Organic Carbon Contaminated Site Assessment

Standard Operation Procedure (SOP) for the Extraction and Instrument Analysis of Siloxanes from Cartridges

EURL-SRM - Analytical Observations Report

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor

JAI RESEARCH FOUNDATION A Global Contract Research Organisation

Certified Reference Material - Certificate of Analysis

Analysis of Polychlorinated Biphenyls (PCBs) in Water Samples Collected February 28 March 1, 2000 from the Bayou Creek System. Wesley J.

GC Troubleshooting. Frank Michel, Katherine K. Stenerson T HCE

Organohalide Pesticides in Water by GC/µECD with Agilent J&W DB-CLP1 and DB-CLP2

Method of determination of phthalates by gas chromatography / mass spectrometry in wines

Partly O-(2-hydroxyethylated) cellulose. It may contain suitable ph-stabilisers such as phosphates.

Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016

Validation Data for the Analysis of Chlorotalonil in Leek Using Mini-Luke Method Followed by GC-QqQ-MS/MS

Certified Reference Material - Certificate of Analysis

Certified Reference Material - Certificate of Analysis

Standard Operating Procedure Title: Operation and Calibration of Sievers 820 TOC Analyser

Simplicity. Accuracy. Durability. Lotix. Automated High Temperature Combustion TOC Analyzer

Gas Chromatography SSI-GC-003

METHOD TRANSLATION OF HJ FOR INTUVO

FOOD PESTICIDE RESIDUES LABORATORY REFERENCE NO INSPECTOR DATE

Method Development for Rapid Extraction of Volatile Organic Compounds from Rock Samples Using Microwave Assisted Extraction

Validation of Sulfur Dioxide Using. SKC UME x 200 Passive Sampler Cat. No Page 1 of 14 Publication 1781 Rev 1607

Direct Injection of Fish Oil for the GC-ECD Analysis of PCBs: Results Using a Deans Switch With Backflushing. Application. Author.

METHOD 25D - DETERMINATION OF THE VOLATILE ORGANIC CONCENTRATION OF WASTE SAMPLES. NOTE: Performance of this method should not be

SpeedExtractor E-916 Determination of Total Petroleum Hydrocarbons in Soil, Sediment, and Waste Samples using the SpeedExtractor E /2009

NH3 Sampling Protocol Using The Ogawa Sampler*

STANDARD OPERATING PROCEDURES SOP: 1502 PAGE: 1 of 11 REV: 0.0 DATE: 10/24/94

Analysis of Semivolatiles by GC/FID using the Ultra Inert Inlet Liners with Wool

National Food Safety Standard

EPA Method 524 for Determination of VOCs in Drinking Water Using Agilent 5975T LTM GC/MSD with Static Headspace

METHOD 7550 OSMIUM (ATOMIC ABSORPTION, DIRECT ASPIRATION)

Determination of Inorganic Anions in Acid Rain Using a Dedicated High-Pressure Capillary Ion Chromatography System

TurboMatrix Headspace and Headspace Trap Samplers

ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER

CTC automated sample injectors for gas chromatography, distributed by Agilent Technologies

EDB July ETHYLENE DlBROMlDE. F. Analytical Quantitation... I. Determinative Method. Contents. A. Introduction I3. Equipment...

A Direct Column-Performance Comparison for Rapid Contract Laboratory Program (CLP) Pesticide Analysis. Application. Authors. Abstract.

Determination of Volatile Organic Compounds (VOCs) in Various Drinking Water Sources by GC/MS

Determination of Table 3 Compounds by LC/MS/MS Chemours Fluoroproducts Analytical Method Revision date: 4/26/2018

Updating Solid Phase Extraction Methods: Tips and Tricks for Improving Existing and New SPE Methods using Method Development

Application Note. Abstract. Authors. Food Safety

Methane baseline concentrations and sources in shallow aquifers from the shale. gas-prone region of the St. Lawrence Lowlands (Quebec, Canada)

Polyvidone Polyvinylpyrrolidone H 2 C H C N

CX-1 ph Gradient Buffer

Field Analytical Method Development for C 6 -C 9 Range Hydrocarbon Measurement

LogD Measurement by Agilent HPLC with ichemexplorer. Michael Lopez Reaction Analytics Inc.

Methods and Materials Manual Of Ion Chromatography Using DX-100 IC. Collecting Samples. LDPE bottle with cap (pre-rinsed with DI water)

Agilent Consumables Improve Performance with Shimadzu GC/MS

Orono Spectral Solutions, Inc.

Application Note AN1501

Oxygen Demand, Chemical

Contract Laboratory Program (CLP) Pesticide Analysis with 0.18 mm ID High-Efficiency GC Columns Utilizing Helium Carrier Gas. Application.

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph

Determination of additives and byproducts in an acid copper plating bath by liquid chromatography

INTERNATIONAL RESEARCH JOURNAL OF PHARMACY

2008 Acid catalyzed esterification of propionic acid with 1-butanol to propionic acid 1-butyl ester

ATMOSPHERIC CHEMISTRY NITROGEN DIOXIDE Version 1.1 (previously AC Protocol, updated Jan 2001)

Provläsningsexemplar / Preview. First edition

Performance Characteristics of the Agilent 1220 Infinity Gradient LC system

MS Certified Vials. Pre-cleaned and certified vials for mass spectrometry

Transcription:

Corn Syrup Analysis E-1a- 1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and heat. The liberated compounds are then measured using a headspace gas chromatographic system equipped with a flame ionization detector. Isovaleraldehyde and acetaldehyde results are reported at 11% D.S. for finished sweeteners. It is important to keep the headspace volume the same or it will have an effect on the final concentrations. Because of this, the volume of the preparation vials is important to maintain. SCOPE This method is applicable to the analysis of finished HFCS, 95 Dextrose, and most starch hydrolyzates (Note 1). SPECIAL APPARATUS 1. Single column gas chromatograph equipped with a flame ionization detector and headspace autosampler (Note 2). 2. Capillary GC Column, Supelco Supel-Q Plot fused silica capillary column (30m, 0.53mm ID), or equivalent. 3. Sample vials (20 ml), caps, seals and crimping tools. 4. Repipettor capable of reproducibly delivering aqueous solutions to within ± 0.1%. 5. Analytical balance capable of measuring 0.0001 g. 6. Transfer pipettes. 7. Steam or water bath set at boiling. Analytical Methods of the Member Companies of the Corn Refiners Association, Inc. Accepted 10-9-09

Corn Syrup Analysis E-1a- 2 REAGENTS 1. Ultra-pure filtered water (Note 3). 2. Phosphoric Acid Solution: In a 1 L volumetric flask add 500 ml of Ultrapure water. Then add 12.171 g of concentrated phosphoric acid (85%). Mix thoroughly. When mixed, fill to the mark with Ultra-pure water and mix. 3. Blank syrup standard (Note 4). 4. Isovaleraldehyde spiking solution: Accurately weigh 107 mg of 98% isovaleraldehyde into a clean 100 ml volumetric flask. Dilute to volume with Ultra-pure water. This is Stock 1 (1,050 ppm). Pipette 1.0 ml of Stock I into a clean 100 ml volumetric flask. Dilute to volume with Ultrapure water. This is Stock II (10.5 ppm). Pipette 1.0 ml of Stock II into a clean 100 ml volumetric flask. Dilute to volume with Ultra-pure water. This is the Spiking Solution (105 ppb). 5. Acetaldehyde spiking solution: Purchase a certified solution of 5.62 ppm 4,6-0-ethylidene-D-glucose or weigh 56.2 mg of 4,6-0-ethylidene-D-glucose (Aldrich catalogue #E3275-4 or Pfaltz and Bauer catalogue #E11450) into a beaker, and transfer to a one liter volumetric flask using Ultra-pure water, dilute to volume. Then prepare a 5.62 ppm solution by diluting 100 ml of this solution to one liter with Ultra-pure water. Store capped in the refrigerator. GC AND HEADSPACE PARAMETERS GC Parameters 1. Injector port temperature: 150 C 2. Detector temperature: 250 C 3. Oven temperature ramp profile:

Corn Syrup Analysis E-1a- 3 Table 1 Oven Temperature Ramp Profile Step Ramp Rate C/min Next Temp ( C) Hold Time (min) Run Time (min) Initial 50 0 0 1 7.5 190 1.5 20.2 4. Gas pressure: 45 ml/min H 2, 450 ml/min air. Headspace Sampler Parameters for capillary column 1. Carrier gas (He): 20mL/min at 12 psi. This will provide passive pressure/flow to the GC. 2. Needle temperature: 125 C 3. Transfer line temperature: 125 C 4. Oven temperature: 80 C 5. Pressurization time: 0.7 minutes 6. Injection time: 0.3 minutes 7. Thermo time: 60 minutes 8. GC cycle time: 23.5 minutes 9. Needle withdraw time: 0.5 minutes PROCEDURE Standard Preparation: 1. Prepare five vials with caps. 2. Weigh 1.5000 ± 0.0050 g of blank syrup into each vial.

Corn Syrup Analysis E-1a- 4 3. Pipette purified water, the spiking solution of acetaldehyde and IVA, and phosphoric acid into each vial in the quantities shown in Table 2. The phosphoric acid is added last into each vial. 4. Immediately cap the vial and crimp it tightly. These vials should not be left for any period of time with stock solution in them unless they are capped. Sample Preparation: For 42 FX samples, weigh 1.6000 ± 0.0050 g of sample into a vial, and to this pipette 7.9 ml of deionized water, then 1.0 ml of phosphoric acid; cap and crimp tightly. Table 2 summarizes the 42 FX sample preparation. For 55 FX samples, weigh 1.5000 ± 0.0050 g of sample into a vial, and to this pipette 8.0 ml of deionized water, then 1.0 ml of phosphoric acid; cap and crimp tightly. Table 2 summarizes the 55 FX sample preparation. All vials are then mix thoroughly and then placed in a boiling water bath for 60 minutes. The samples are then transferred to the headspace analyzer and are ready to be analyzed on the GC. CALCULATIONS Results are calculated using a 5-point calibration curve. The calibration standards consist of the blank syrup and each of the four prepared standards. The blank is subtracted from each standard and then a line equation is created with the corrected standards (x-axis is the standard concentration, the y-axis is the peak area). The concentration of components, expressed in parts per billion of 11% D.S. finished product samples, can be determined by the following equation: Y= mx+b where: Y= peak area x= concentration of acetaldehyde or IVA in ng/g (ppb) of 11% ds syrup m = slope b = y-intercept

Corn Syrup Analysis E-1a- 5 Tables Table 2 - Combined IVA/ Acetaldehyde GC Method Standards and Sample Composition Syrup (g) Purified Water (ml) Acetaldehyde Spiking Solution (ml) IVA Spiking Solution (ml) Phosphoric Acid (ml) Blank 1.5 8.0 -- -- 1.0 Std #1 Std #2 1.5 1.5 7.6 7.2 0.2 0.4 0.2 0.4 1.0 1.0 Std #3 1.5 6.8 0.6 0.6 1.0 Std #4 1.5 6.0 1.0 1.0 1.0 55 FX 1.5 8.0 -- -- 1.0 42 FX 1.6 7.9 -- -- 1.0 Table 3 - Combined IVA/ Acetaldehyde GC Method Standard Concentrations Standard Acetaldehyde Concentration (ppb) IVA Concentration (ppb) Blank 0 0 #1 22.9 2 #2 45.7 4 #3 68.6 6 #4 114.3 10 NOTES AND PRECAUTIONS 1. Sulfur dioxide levels in excess of 10 ppm may interfere. 2. Different GC s and headspace units may use different parameters. 3. High quality water (over 18 mega-ohms) must be used to prepare all reagents and blanks to avoid contamination. 4. The blank syrup should have low or non-detectable levels of IVA and acetaldehyde. 5. This method typically reports slightly higher acetaldehyde results than CRA Method E-1. 6. The variability of IVA and acetaldehyde results obtained from this method are similar and typically better than those reported from methods ISBT 12.0 and CRA method E-1 respectively.

Corn Syrup Analysis E-1a- 6 REFERENCES 1. CRA Acetaldehyde Method E-1. CRA Method E-1 Acetaldehyde : Revision Dec. 19, 2006. 2. ISBT Isovaleraldehyde Method 12.0. Isovaleraldehyde (IVA) : Revision March 2003.