Regulatory position paper on sulphur dioxide monitoring, comparing portable-instrumental techniques with manual methods

Similar documents
Method Implementation Document (MID 14181) EN 14181: Stationary source emissions Quality assurance of automated measuring systems

Monitoring Quick Guide 3

Technical Guidance Note (Monitoring) Medium Combustion Plant Directive and Generator Controls: monitoring point source emissions

EN14181 Stationary Source Emissions Quality Assurance of Automated Measuring Systems

Technical Guidance Note (Monitoring) Monitoring of stack gas emissions from medium combustion plants and specified generators

Breathing life into environmental and health & safety legislation.

Training for Personnel to the MCERTS standard Level 1 Technician and Level 2- Team Leader. Syllabus

Method Implementation Document for BS EN * - Application to manual stack emissions monitoring

National Physical Laboratory Hampton Road Teddington Middlesex United Kingdom TW11 0LW

QAL3 methods for different measurement principles from the point of view of AMS manufacturers, test houses and plant operators

The development of instrumentation for application to carbon capture and storage measurements

CASE STUDIES OF DUST AMS CALIBRATION AT EXTREME CONDITIONS

Performance Standards for Indicative Ambient Particulate Monitors Environment Agency July 2012 Version 2

The guidance represents the views of the Commission services at the time of publication. It is not legally binding.

Guido Baum Dr. Peter Wilbring

PRODUCT CONFORMITY CERTIFICATE

Deadline for tenders: 22 September 2016

Method Implementation Document (MID 1948)

PRODUCT CONFORMITY CERTIFICATE

PRODUCT CONFORMITY CERTIFICATE This is to certify that the

Performance Standards for Continuous Ambient Air Quality Monitoring Systems. Environment Agency Version 8 June 2012

PRODUCT CONFORMITY CERTIFICATE

Policy on Monitoring of Stack Emissions to Air at EPA Licensed Sites

Quality assurance of automated measuring systems

PRODUCT CONFORMITY CERTIFICATE

Continous Emission Monitoring: Methods, Regulations and QA/QC. Anders Arvidsson, Opsis AB, Sweden

PRODUCT CONFORMITY CERTIFICATE

An investigation into the use of EN for the determination of mass concentration of sulphur dioxide in stack emissions

the environmental future for monitoring systems and services

Application of EN14181 and EN15267 to Electronic Noses: Challenge or Provocation?

Presented By Maria Friedman National Environmental Monitoring Conference Washington, DC August 6, 2012

Measurement uncertainty implications for the enforcement of emission limits. Maciek Lewandowski (Environment Agency)& Michael Woodfield (AEAT) UK

Lars K. Gram. FORCE Technology. The Danish EPA's Reference Laboratory for Air Emission Monitoring.

PRODUCT CONFORMITY CERTIFICATE

National Air Quality Reference Laboratories and the European Network AQUILA:

PRODUCT CONFORMITY CERTIFICATE

PRODUCT CONFORMITY CERTIFICATE This is to certify that the

PRODUCT CONFORMITY CERTIFICATE This is to certify that the

CEM Istanbul, 14 th May 2014

PRODUCT CONFORMITY CERTIFICATE

Introduction and Agenda

PRODUCT CONFORMITY CERTIFICATE

TNP Regulatory Guidance

PRODUCT CONFORMITY CERTIFICATE

Hydrogen fuel Product specification. Part 2:

MINIMUM TECHNICAL COMPETENCE FOR AIR TIGHTNESS TESTING IN BUILDINGS FOR ENGLAND AND WALES

CEMS FOR COAL FIRED POWER STATIONS DEVICE CALIBRATION AND MAINTENANCE

PRODUCT CONFORMITY CERTIFICATE

(CEMS) Sanjeev K Kanchan. Centre for Science and Environment

Your ref.: Our ref.: Date: CDM Ref 1443 YKK/TRIKA/MLEH 19 January 2008

QUALITY ASSURANCE OF EMISSION MEASUREMENT

Technical TFTEI meeting of 24 June Agenda

PRODUCT CONFORMITY CERTIFICATE

General requirements for the competence of testing and calibration laboratories. In this presentation:

A SUPPLEMENTARY NOTE TO THE SUBSTITUTE FUELS PROTOCOL

PRODUCT CONFORMITY CERTIFICATE

Making the Confidence Connection: Conformity Assessment System Design

Protocol for intercomparison exercise, final version December 2008 AQUILA N 37

EPA vs. EN standards

CONTINUOUS EMISSION MONITORING SYSTEM. Copyright Adage Automation Pvt Limited

GLP/SC/01 Basic statistical tools for analytical chemistry (2 days)

IAF Mandatory Document. Accreditation Assessment of Conformity Assessment Bodies with Activities in Multiple Countries. Issue 2 IAF MD 12:2016

Guidelines for the assessment of the appropriateness of small interlaboratory comparisons within the process of laboratory accreditation

PRODUCT CONFORMITY CERTIFICATE

Raad voor Accreditatie (Dutch Accreditation Council RvA) Accreditation of Reference Material Producers

PRODUCT CONFORMITY CERTIFICATE

ISO and SANAS guidelines for use of reference materials and PT scheme participation. T & M Conference

data systems CEM International Conference and Exhibition on Emissions Monitoring

Moving from ISO 14001:2004 to ISO 14001:2015 Transition Guide

Overview of ASTM D7036: A Quality Management Standard for Emission Testing

IAF Mandatory Document

INDUSTRIAL MEASUREMENTS USING FTIR GAS ANALYZER

An Overview of Mercury Monitoring Options

EA-7/04 Legal Compliance as a part of accredited ISO 14001: 2004 certification

Establishment of an Accredited Reference Measurement Laboratory

A global monitoring systems offering. A world of advantages. Thermo Scientific Continuous Emissions Monitoring Systems

This document is a preview generated by EVS

Terms and conditions for laboratories appointed to undertake the testing of compost under the Compost Certification Scheme aligned to PAS 100 and the

Bord Na Móna Environmental Ltd trading as ANUA

Study into the loss of material from filters used for collecting particulate matter during stack emissions monitoring

Quality Manual ISO 9001:2015 Quality Management System

Laboratory Quality Assurance Manager & Laboratory Assessor RULES & HANDBOOK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Title Procedure for setting up the accreditation of new conformity assessment schemes

PRODUCT CONFORMITY CERTIFICATE

ILAC P10 TRACEABILITY

9001:2015, ISO 14001:2015 & ISO

IAF Mandatory Document. for the Audit and Certification of a Management System Operated by a Multi-Site Organization (IAF MD 1:2018)

Moving from ISO/TS 16949:2009 to IATF 16949:2016. Transition Guide

The new International Standard ISO 17034: general requirements for the competence of reference material producers

FTIR Analyser BROCHURE

Measurement uncertainty: Requirements set in the accreditation standards

BioTector System-C. Online TOC Analyzer for Clean Water Applications

General Accreditation Criteria Equipment assurance, in-house calibration and equipment verification

WELMEC European cooperation in legal metrology

Measuring Instruments Directive 2014/32/EU Assessment of Notified Bodies in Charge of Type Examination Presumption of Conformity based on EN 17065

Sampler with PM 2.5 pre-separator

Approved Document to support regulation

Guidance for Swedac s assessors on the assessment of laboratories based on the requirements set out in SS-EN ISO/IEC 17025:2005

ISO 13528:2015 Statistical methods for use in proficiency testing by interlaboratory comparison

Transcription:

Regulatory position paper on sulphur dioxide monitoring, comparing portable-instrumental techniques with manual methods Dr Rick Gould, Environment Agency, Technical Advisor, Certification and Standards, richard.gould@environment-agency.gov.uk; and Malcolm Beaver, Technical Advisor for Emissions Monitoring, malcolm.beaver@environment-agency.gov.uk; both within the Monitoring & Assessment team, Environment & Business Directorate, Environment Agency, PO Box 519, Preston, PR5 8BD, England. Summary This position paper describes the UK perspective and in particular, the Environment Agency s (EA) approach to specifying requirements for monitoring sulphur dioxide (SO2), describing the EA s experiences with both instrumental and manual methods. Many industrial processes have emission limit values for SO2. These industrial processes are diverse, with wide variations in emissions profiles and monitoring provisions. The EA s experience is that this situation requires both flexibility and pragmatism when monitoring SO2, in order to apply a risk-based approach and to ensure an appropriate level of quality of monitoring. Practically, this means achieving a measurement uncertainty that takes into account the environmental risks posed by the industrial process and the regulatory requirements. Using either manual or instrumental methods for SO2 monitoring provides this flexibility and pragmatism. Therefore the Environment Agency has always permitted test laboratories to use both types of technique to measure SO2, with the caveat that the test laboratories must be accredited to ISO/IEC 17025 and the applicable Monitoring Certification Scheme (MCERTS) performance standards, with appropriate monitoring standards and methods in their scope of accreditation. When examining the QAL2 and Annual Surveillance Test (AST) reports required by EN 14181, as well as periodic compliance-reports, there is a clear preference amongst test laboratories for instrumental methods, rather than using the wet-chemistry manual-method. The EA s strategy has been to require international and national standards for monitoring, using a hierarchy of standards that the European Commission now considers as Best Available Techniques (BAT) for monitoring. In addition to specifying standards for monitoring and quality assurance, the EA also routinely audits test laboratories using two types of audits. These are: unannounced field-audits, similar to those used by national accreditation bodies, and; desktop audits of monitoring reports, specifically those for the periodic compliance-monitoring of SO2 emissions as required by a permit, and QAL2/AST reports required by EN 14181. Our main findings are as follows: 18082017 Page 1

Instrumental techniques and manual methods for monitoring SO2 are both necessary and complementary. No single technique, method or instrument is better than any other, whilst a technique suited to one installation can be inconvenient for another. Instrumental monitoring was initially popular in the UK, as well as some other EU Member States, because of its perceived simplicity, advantages and convenience. That said, achieving the required level of uncertainty requires careful management of the technique, whilst audits of both test reports and test laboratories in the field reveals common problems that can be easily eliminated. Although instrumental techniques were initially very popular, there has been a resurgence in the use of the manual, wet-chemistry Standard Reference Method (SRM), EN 14791. As a result, most test laboratories use both the SRM and portable instrumental-techniques applied through Alternative Methods (AMs). That said, instrumental methods are by far the most popular. An analysis of audits of QAL2 and AST reports indicates that there was not a significant difference between the permanently-installed Automated Measurement Systems (AMS) results; and those from the SRM or AMs. The latter are the EA s methods described in Technical Guidance Note (TGN) M21 (Typically NDIR or UVbased portable instrumental-techniques) and TGN M22 (FTIR-based portable instrumental-technique). In approximately half the results, the AMS results were higher whereas this was reversed in the other half of the examples. The SRM results were higher as an average by approximately 10% across all sites, but this difference was not statistically significant. 1. The Environment Agency s strategy for monitoring, when applied to measurements of SO2 1.1 The Environment Agency s position on monitoring There are two facets of the EA s strategy for monitoring that are relevant to the SULF NORM project. Firstly, the EA sees a strong benefit in portable instrumental-techniques for periodic monitoring, such as real-time readings and the ability to characterise short-term variations in SO2 concentrations relative to changes in the industrial process. Secondly, the EA has always specified the standards hierarchy for monitoring, i.e. to use CEN standards where these are available and suitable, and then ISO, national, and other native national or international standards. The latter could include VDI and USEPA methods for monitoring, as well as the Environment Agency s TGN M21 and TGN M22. The EA has applied these standards through permit conditions and through MCERTS, which uses a framework of accreditation standards, standards for monitoring, and supplementary specifications to provide quality assurance and control. Additionally, if there were gaps in any standards such as personnel certification for the competency of testlaboratory staff performing emissions - then the EA developed and applied specific MCERTS standards until EN standards become available. Before CEN published the manual, wet-chemistry method EN 14791 for SO2 monitoring, test laboratories in the UK used the following two standard methods: 18082017 Page 2

ISO 7935 as a basis for a method employing a portable instrumental-technique. This ISO standard describes performance requirements for an instrumental system, together with requirements for use and quality assurance/quality control (QA/QC). The QA/QC requirements within ISO 7935 include functional tests and drift requirements. The scope of ISO 7935 can be applied to both permanently-installed AMS and portable instrument-techniques. When CEN published EN 15267-3, this conflicted with several parts of ISO 7935; therefore EN 15267-3 superseded ISO 7935. ISO 7934 as a wet chemistry method. This is very similar in principle to EN 14791 and was therefore superseded by the CEN standard. EN 14791 was clearly a direct replacement for ISO 7934, whilst this standard is still current in countries which are not members of CEN. However, although EN 15267-3 superseded ISO 7935, the ISO standard had a wider scope. In simple terms, EN 15267-3 specifies performance characteristics and test procedures for AMS, although before the development of EN 15267-4, it was adapted for testing portable instrumental-techniques. ISO 7935, on the other hand, includes performance characteristics and test requirements for instruments that measure SO2, but additionally includes procedural requirements for use, together with associated provisions for QA/QC. In other words, ISO 7935 readily provided a strong foundation for an AM using portable instrumental-techniques, whereas EN 15267-3 does not. This left a gap in available standards for the periodic monitoring of stack emissions using a portable instrumentaltechnique. Therefore the EA worked with an industry trade-association, the Source Test Association (STA, www.s-t-a.org ), to develop and produce an instrument-based AM for SO2, TGN M21. The STA s members include test laboratories, process operators and instrument manufacturers. TGN M21 is now one the EA s two AMs for the periodic monitoring of SO2 using portable instrumental-techniques. Additionally, in England and in other countries which apply the MCERTS standards, test laboratories using FTIR-based systems are required to apply TGN M22, which is now the EA s prescribed AM for periodic monitoring using transportable FTIR. When comparing TGN M21 and TGN M22, the former standard is specifically for monitoring SO2, whereas TGN M22 applies to SO2 and any other determinands that an FTIR can measure, as long as the measurements meet the performance requirements specified in TGN M22. The scope of the TGN M21 is not restricted to any particular instrumental-technique as TGN M21 specifies performance characteristics and QA/QC requirements for any instrumental-technique that the test laboratory chooses to use and that is capable of meeting the performance requirements. For example, common techniques for SO2 include NDIR and UV-based systems. If a test laboratory chooses to use a FTIR-based instrumental-system and apply TGN M21, then the test laboratory must also apply the QA/QC requirements of the ASTM or the applicable USEPA methods for FTIR, e.g. USEPA Method 320. 18082017 Page 3

Similarly, EN/TS 17021 includes provisions for all instrumental-techniques that meet the performance requirements. If a test laboratory uses FTIR-based systems, then the laboratory must also satisfy the QA/QC requirements of the CEN TS which is being developed by CEN TC 264 WG26. 1.3 Terminology It is worth noting that different standards and methods use varying terminology, depending on their scope. For example, a portable instrumental-technique is equivalent to the term Portable Automated Measuring System (P-AMS), used in EN 15267-4. Additionally, the term permanently-installed instrumental-techniques and Continuous Emissions Monitoring Systems (CEMs) are equivalent to the term Automated Measuring System (AMS), EN 15267-4 and EN 15267-3 provide for type-testing and certification for P-AMS and AMS, respectively. For example, in the case of P-AMS for carbon monoxide (CO), EN 15267-4 specifies performance requirements and test procedures used for type-testing and certification of the P-AMS. Following this, if an instrument meets the performance requirements in EN 15267-4, then the SRM for CO, EN 15058, provides for the ongoing QA/QC throughout the operational lifetime of a P-AMS for CO. EN 15058 specifies functional tests that a test laboratory must perform during each measurement campaign and also at specified intervals such as annually, in order to give confidence that the P-AMS is performing to the tolerances specified on its type-testing certificate. In the case of SO2 monitoring, the scope of EN15267-4 would apply to techniques such as NDIR, UV-based systems and FTIR used as P-AMS, whilst ongoing QA/QC would then be carried out under EN/TS 17021, TGN M21 or TGN M22 as appropriate. Similarly, an AMS for SO2 AMS would be tested and validated under EN 15267-3; following this, the ongoing QA/QC would be performed according to EN 14181, together with either the SRM for SO2, or an accredited AM. 1.4 Application through permits according to Article 14 of the Industrial Emissions Directive (2010/75/EU) All permits for installations in England regulated under the Industrial Emissions Directive (IED) state that monitoring must be MCERTS accredited or certified where applicable. Therefore for test laboratories measuring SO2 with instrumental methods, this means that the test laboratories scope of accreditation must include the following: Accreditation to EN ISO/IEC 17025, EN ISO/TS 15675, EN 15259 and the MCERTS performance standards for manual stack-emissions monitoring for test laboratories and personnel. TGN M21 and/or TGN M22 must be within the scope of accreditation. Portable instrumental-techniques must be certified to at least Annex F of the MCERTS performance standards for continuous emission monitoring systems, or to all the requirements for permanently installed AMS (i.e. meet the requirements of EN 18082017 Page 4

15267-3). In the next two years, the MCERTS performance standards will be revised to provide for the requirements of EN 15267-4, which refers to portable instrumentaltechniques as Portable-AMS. 1.5 TGN M21 for monitoring SO2 using instrumental methods TGN M21 is very similar to the instrumental SRMs such as EN 15058, produced by CEN TC 264 WG16. TGN M21 itself is based on EN 15058 for measuring carbon monoxide. As such, it includes performance specifications for the transportable AMS used in TGN M21, procedures for performing the measurements, provisions for quality assurance and quality control, and an uncertainty budget. TGN M21 does not specify any particular type of technique for the transportable AMS, but instead specifies performance requirements for the complete transportable AMS. Therefore although test laboratories typically use NDIR-based instruments when using TGN M21, there are some test laboratories which use NDUV-based transportable AMS. TGN M21 has been accepted by the EA as equivalent to EN 14791 on the basis of test data acquired in accordance with the Technical Specification for equivalency of AMs current at the time, CEN/TS 14793. 2. Experiences with TGN M21 When TGN M21 was published, test laboratories rapidly adopted it, such that it became the dominant method for SO2 monitoring. Also, as a growing number of test laboratories were purchasing and using transportable FTIR, the EA, STA and National Physical Laboratory collectively produced TGN M22 to specify performance requirements for the FTIR, together with procedures for monitoring, and quality assurance/control. Today, test laboratories commonly use all three methods, i.e. TGN M21, TGN M22 and EN 14791 for measuring SO2. 2.1 Results of field audits of test laboratories Every MCERTS-accredited test laboratory must have at least one audit by the United Kingdom Accreditation Service (UKAS) annually. Additionally, the EA performs unannounced audits too. As these audits can happen anywhere at any time, the EA has found that these additional audits help to maintain the quality of the monitoring. Regarding typical non-compliances, the most common of those found has been that test laboratories use an inappropriate concentration of span gas during the QA/QC checks of the transportable AMS. However, one test laboratory began to routinely use a gas blender so that the span gas was always at the appropriate concentration. Therefore the EA recommends that test laboratories should also use a gas blender for span checks. Another observation is that according to TGN M21, test laboratories typically use a dry gas when calibrating the AMS and when performing span checks. Although a wet gas is not required, it can improve the quality of the calibration when there are potential losses in the sampling system. This can result in a negative bias and this will be explored in the final subsection. 18082017 Page 5

Based on test reports for routine periodic monitoring and for QAL2/AST exercises, a quick scan of the data, coupled with simple statistical-tests, suggests that test laboratories can usually meet the uncertainty requirements specified by legislation. This is reflected, for example, in both the data and the graphs of the AMS results versus test-laboratory measurements. Work elsewhere under SULF NORM is examining proficiency testing-data to better understand the real-world capabilities of EN 14791 and TGN M21; therefore it is recommended that further work explores this correlation further, and quantifies the uncertainties of the SRM and EN/TS 17021. 2.2 Audits of QAL2/AST reports The Environment Agency examined QAL2 and AST reports for the past two years. Within these reports, test laboratories predominantly used instrumental methods, i.e. FTIR, NDIR and NDUV. Regardless of the instrumental technique used, the average test laboratory results were slightly higher than those of the AMS, but not significantly so; the overall average was 10% higher for the SRM results when considered as a group, but the difference was not significant. In half of the reports examined, the AMS results were higher than those of the SRMs and lower in the remaining reports. That said, when there has been a significant difference in individual reports, the SRM results are typically lower than those of the AMS. On a further examination, audits have suggested errors in applying the instrumental methods, which in turn can lead to a negative bias in the SRM readings. 3. Recommendations Test laboratories should use gas blenders when performing calibration and span checks, in order to produce the ideal concentrations of test gases. Additionally, the benefits of using moistened test-gases needs consideration. The SULF NORM project team should perform a comparison of span checks through the sampling system of a transportable AMS, using both dry gas and wet gas, in order to determine whether there are losses of SO2 in the sampling system when the test gas is in the presence of moisture. The SULF NORM project team should determine whether spiking with SO2 during measurements has an effect on the calibration of the transportable AMS. The SULF NORM project team should determine the uncertainties of the portableinstrumental techniques when using EN/TS 17021, and compare these with those of the SRM EN 14791, in order to determine whether there is a robust correlation between the SRM and AM. References: ISO 7934:1989 Characterization of air quality. Stationary source emissions. Method for the determination of the mass concentration of sulphur dioxide: hydrogen peroxide/barium perchlorate/thorin method. 18082017 Page 6

ISO 7935:1992 Characterization of air quality. Stationary source emissions. Determination of the mass concentration of sulfur dioxide. Performance characteristics of automated measuring methods. EN 15267-3:2007. Air quality. Certification of automated measuring systems. Performance criteria and test procedures for automated measuring systems for monitoring emissions from stationary sources. EN 15267-4:2017. Air quality. Certification of automated measuring systems. Performance criteria and test procedures for automated measuring systems for periodic measurements of emissions from stationary sources. BS EN 14791:2017. Stationary source emissions. Determination of mass concentration of sulphur oxides. Standard reference method. EN 15058:2017. Stationary source emissions - Determination of the mass concentration of carbon monoxide - Standard reference method: non-dispersive infrared spectrometry. EN/TS 17021:2017. Stationary source emissions. Determination of the mass concentration of sulphur dioxide by instrumental techniques. EN 14181:2014. Stationary source emissions. Quality assurance of automated measuring systems. USEPA Method 320 - Vapor Phase Organic and Inorganic Emissions by Extractive FTIR. Environment Agency Technical Guidance Note (TGN M22) - Measuring stack gas emissions using FTIR instruments Available from www.mcerts.net. Environment Agency Technical Guidance Note (TGN M21) - Stationary source emissions A procedure to use an Alternative Method for measuring emissions of sulphur dioxide, using instrumental techniques. Available from www.mcerts.net. MCERTS Performance Standards and Test Procedures for Continuous Emission Monitoring Systems. For gaseous, particulate and flow-rate monitoring systems. Environment Agency. Available from www.mcerts.net. ASTM D6348-3: 2003 Standard Test Method for Determination of Gaseous Compounds by Extractive Direct Interface Fourier Transform Infrared (FTIR) Spectroscopy.. 18082017 Page 7