Analysis of inorganic impurities in lubricating oils by ICP-MS

Similar documents
Analysis of wear metals and contaminants in engine oils using the 4100 MP-AES

Automating EPA 6020 Compliant Analysis with the Agilent 7900 ICP-MS and ESI prepfast Autodilution System

Trace Elements in Petrochemical Samples: Achieving Excellent Analytical Results

Benefits of running organic matrices using the Agilent 5100 ICP-OES fast, robust, high performance analysis

Direct Analysis of Photoresist by ICP-MS. Featuring the Agilent Technologies 7500s ICP-MS

Analysis of steel and its alloys using the GB/T standard and an Agilent 5100 ICP-OES in dual view mode

DETERMINATION OF ELEMENTS IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS 10500, & USING THE AGILENT 5100 ICP-OES

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES

VHG Labs. Standards and Supplies for Spectrochemical Analysis QUALITY. TRACEABILITY. DEPENDABILITY. WORLDWIDE

Application Note. Abstract. Authors. Environmental

Ultra-Trace Analysis of Beryllium in Water and Industrial Hygiene Samples by ICP-MS Application

Direct measurement of trace rare earth elements (REEs) in high-purity REE oxide using the Agilent 8800 Triple Quadrupole ICP-MS with MS/MS mode

Ultra-fast ICP-OES determinations of base metals in geochemical samples using next generation sample introduction technology

Determination of Elemental Impurities in Graphite-based Anodes using the Agilent 5110 ICP-OES

METHOD 3040A DISSOLUTION PROCEDURE FOR OILS, GREASES, OR WAXES

Table I: MCL and MRL Concentrations for Contaminants Monitored Under the Safe Drinking Water Act National Primary Drinking Water Regulations

The Agilent Atomic Spectroscopy portfolio for Environmental applications AA, MP-AES, ICP-OES, ICP-MS & ICP-QQQ

Increased Sample Throughput for ICP-OES Applied to U.S. EPA Method 200.7

Water Analysis Using ICP-OES with an Ultrasonic Nebulizer

Agilent Atomic Spectroscopy Solutions for Energy & Chemicals AA, MP-AES, ICP-OES, ICP-MS.

The Determination of Alloying Elements in High-Carbon Steel using Radially Viewed Inductively Coupled Plasma Optical Emission Spectrometry (ICP-OES)

ElvaX Light for Petrochemical analysis

Characterization of Surface Metals on Silicon Wafers by SME-ICP-MS. Featuring the Agilent Technologies 7500s ICP-MS

Radial, Axial or Dual View ICP: Which Do You Choose? Manny Almeida Teledyne Leeman Labs, Inc. Hudson, NH

High matrix tolerance of the Thermo Scientific icap 7000 Plus Series ICP-OES with the radial Ceramic D-Torch and the sheath gas adaptor

Determination of 22 Elements Following US EPA Guidelines with a New Megapixel CCD ICP-OES

Increase productivity for environmental sample analysis using the SVS 2+ Switching Valve System for Agilent 5100 SVDV ICP-OES

Composition Analysis of Animal Feed by HR ICP-OES

Multi-Element Analysis of Trace Metals in Animal Feed using ICP-OES

The Determination of Trace Elements in Stainless Steel by Forked Platform GFAAS

Amplifying ICPMS Productivity Using Discrete Sampling Technology

Analysis of Trace Metals in Surface and Bottled Water with the Optima 7300 DV ICP-OES

Analysis of Stainless Steel by Dual View Inductively Coupled Plasma Spectrometry

Multi-element determination in pharmaceutical preparations using the Thermo Scientific icap Q ICP-MS

ESI SC-FAST Sample Introduction Solutions for PerkinElmer ICP-OES and ICP-MS Systems

Analysis of Mineral and Heavy Metal Content in Beverages Using the Teledyne Leeman Labs Prodigy Plus ICP-OES

Analysis of Heavy Metals in Iron-based Fertilizers by HR ICP-OES

Accurate. Robust. Reliable. AGILENT 710 SERIES ICP-OES

Accurate. Robust. Reliable. AGILENT 710 SERIES ICP-OES

The image part with relationship ID rid4 was not found in the file. Welcome

Common Challenges facing Laboratories doing Elemental Analysis Today

The Determination of Toxic Metals in Waters and Wastes by Furnace Atomic Absorption

EU water analysis using the Thermo Scientific icap 7400 ICP-OES Duo

Atomic Spectroscopy Solutions for Environmental Applications

EU Water Analysis Using the Thermo Scientific icap 7400 ICP-OES Duo

Analysis of Environmental Samples with the Agilent 730-ES Following US EPA Guidelines

Analysis of Trace Elements in Seawater Using the Thermo Scientific icap 7000 Series ICP-OES Duo

Determination of Elemental Bad Actors (As, Hg and Si) in Petrochemical Samples Using ICP-AES and ICP-MS

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

Avio 200 ICP-OES Flat Plate

Analysis of Toxic Trace Elements in Coffee Bean Products by HR ICP-OES

High performance radio frequency generator technology for the Thermo Scientific icap 7000 Plus Series ICP-OES

Achieving Optimum Throughput in ICP-MS Analysis of Environmental Samples with the Agilent 7500ce ICP-MS Application

US EPA Method using the Thermo Scientific icap 7600 ICP-OES Duo

TECHNIQUES OF INDUCTIVELY COUPLED PLASMA OPTICAL EMISSION SPECTROMETER S. KARTHIKEYAN SSA CENTRAL POLLUTION CONTROL BOARD SOUTH ZONE OFFICE BANGALORE

Analysis of Stainless Steel by Dual View Inductively Coupled Plasma Spectrometry

ISO INTERNATIONAL STANDARD. Iron ores Determination of various elements by X-ray fluorescence spectrometry Part 1: Comprehensive procedure

Metal Aerosol Emissions from Biomass Combustion

ANALYSIS OF OILS WITH THE SPECTROFLAME MODULA

Analytically Monitoring the Effect of Fracturing Activity

Background Statement for SEMI Draft Document 5134 Revision to SEMI C , Specifications for Hydrofluoric Acid

Analysis of Samples Related to Nuclear Submarine Operation

Tender Document. National Centre of Excellence in Analytical. Chemistry, University of Sindh, Jamshoro. Inductive Coupled Plasma (ICP -OES) ISSUED BY:

Collaborative Efforts for Creating a Robust Trace Metals Analytical Procedure for Flue Gas Desulfurization Wastewaters by ICP-MS

Functionality. Normal Operation

ICP-OES. Varian Vista-MPX CCD Simultaneous ICP-OES

Background The document will be revised for its five year review and updated per the committee agreed upon changes.

The Determination of Trace Elements in Waters using the US EPA Method and the Thermo Scientific icap 6500 ICP-OES (Duo)

Background Statement for SEMI Draft Document 4795 (Yellow Ballot) NEW STANDARD: SPECIFICATIONS FOR NITRIC ACID, USED IN PHOTOVOLTAIC APPLICATIONS

Simple, fast and reliable analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS

Agilent 7900 ICP-MS RAISE YOUR EXPECTATIONS WITH THE NEXT GENERATION OF ICP-MS

Agilent 7900 ICP-MS RAISE YOUR EXPECTATIONS WITH THE NEXT GENERATION OF ICP-MS

ABSTRACT. This study investigates the use of inductively coupled plasma-optical emission spectrometry

STUDY OF ELEMENT FINGERPRINTING IN GOLD DORÉ BY GLOW DISCHARGE MASS SPECTROMETRY

PROCUREMENT PROCEDURE OF CPRI (NON WORKS) Dt of Revision : Issue Dt. :

Eco Plasma Innovative Inductively Coupled Plasma Technology

Background Statement for SEMI Draft Document 4701B New Standard: GUIDE FOR HAFNIUM AMIDES

Analysis of 14 Trace Elements in Stainless Steel

EDX 600. Atomic Absorption Spectrometry

Standard Solutions (Traceable to NIST)

Background Statement for SEMI Draft Document 4701D New Standard: GUIDE FOR HAFNIUM AMIDES

Evaluation of methods for analysis of lead in air particulates: An intra-laboratory and interlaboratory

Standard Solutions (Traceable to NIST)

Guidelines for Trouble Shooting and Maintenance of ICP-OES Systems. Eric Vanclay Product Manager Spectroscopy Consumables May 2, 2012

ICP-AES Determination of 15 Kind of impurity elements in the Vanadium-aluminum alloy

Assaying and Analysis Choices in Practice

SEMI C SPECIFICATIONS AND GUIDELINES FOR AMMONIUM HYDROXIDE

ICP Expert software. Technical Overview. Introduction

icap 7000 Plus Series ICP-OES Streamlined performance and ultra-low ICP-OES detection limits for routine laboratories

Introducing the new Agilent 7800 Solution-Ready ICP- MS. Agilent Technologies 2015

Analysis of elemental impurities in drug products using the Thermo Scientific icap 7400 ICP-OES Duo

Inductively coupled plasma (ICP) spectroscopy for online measurements of trace metal emissions at the 250 kw PACT furnace

CERTIFICATE OF ANALYSIS 1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION

Analysis of Metals in Water, Stream Sediments and Floodplain Soils Collected March 21-23, 2005 from the Bayou Creek System. David J.

Metals Analyzer. OES 6000 Optical Emission Spectrometer. fast and accurate metal analysis

prepfast Autodilution Autocalibration Auto QC Dilution Elemental Scientific

USABILITY BY DESIGN Every component of the Agilent 7900 ICP-MS is engineered for performance and reliability

Analysis of Trace Metals in Tobacco and Tobacco Ash using Atomic Emission Spectroscopy

Improved Antimony Recoveries in Soil Matrices by 3050B/6010B. DataChem Laboratories, Inc. Robert P. Di Rienzo Jeffery S. Ward John T.

Rubber and plastics gloves for food services Limits for extractable substances

Transcription:

Analysis of inorganic impurities in lubricating oils by ICP-MS Application note Authors Energy and Fuels Junichi Takahashi, Kei Kasahara Agilent Technologies, Japan Introduction The most widely used techniques for monitoring metals in petrochemical products have been atomic absorption spectrometry (AAS), x-ray fluorescence spectrometry (XRF) and inductively coupled plasma optical emission spectrometry (ICP-OES) [1]. These techniques give sufficient detection limits for the common applications so, while the importance of metals in the products has not been ignored, not much attention has been given to updating analytical methods to take advantage of new technology such as ICP-MS. Official analytical methods issued by ASTM from 1991 to 2005 were based upon AAS [2,3,4] and ICP-OES [5,6]. Papers published during these years have mainly discussed sample preparation such as emulsification [7], solvent extraction [8], and so on with AAS or ICP-OES. Although the superior sensitivity of ICP-MS is well known and ICP-MS is

accepted in a variety of industrial fields, ICP-MS has not been widely utilized for routine applications in the petrochemical industry. There are various reasons for this, including the fact that petroleum manufacturing and refining sites do not require the superior sensitivity offered by ICP-MS. Another factor is that the high purity organic reagents needed for ultratrace analysis are not commonly available in the commercial market. For example, metallic standards for oil analysis are manufactured by very few vendors. Although 72 elements can be determined by ICP-MS, the number of elements included in commercially available organic standards is around 30 at most. That number is sufficient for most existing methods, because the other elements are typically too low to be determined using ICP-OES. Regardless of the detection mode, analysis of petroleum-based samples presents a significant challenge for ICP instrumentation. It is very difficult to sustain an ICP when introducing organic solvents, particularly volatile ones. For ICP analysis, a liquid sample is typically converted to an aerosol by a nebulizer; at the same time, a large amount of solvent vapor is created and transported to the ICP torch, even when a chilled spray chamber is used. Sample aerosol droplets and particles coming out of the central tube of the torch pass through the center of the plasma. However, organic vapor cannot do so because the high temperature argon plasma has a much higher viscosity than the low temperature organic vapor. If the linear velocity of the carrier gas is not high enough to penetrate the plasma, the vapor surrounds the plasma and, as a result, cuts off the electromagnetic coupling between the load coil and ICP. The use of a narrower bore injector such as 1.0 mm internal diameter increases the linear gas velocity and allows the sample aerosol to punch through the plasma, so enabling organic solvent introduction, but the sensitivity is reduced by approximately 30% compared to a typical injector used with aqueous samples. For analysis by ICP-MS, the high concentration of carbon matrix in the organic samples could also lead to soot deposits on the sampling cone and torch. In order to avoid such deposits, ICP-MS requires the addition of O 2 to the carrier gas to decompose the carbon in the plasma. In order to ignite the ICP with O 2 gas present, a very robust radio frequency (RF) generator circuit is required. Successful analysis of petroleum samples using ICP-MS must overcome these challenges. The all solid state RF generator for the Agilent 7700 Series ICP-MS is so robust that a 1.5 mm injector torch can be used to introduce most organic solvents. Moreover, the ignition of the ICP is possible with organic solvent and oxygen present. In this report, the analysis of lubricant oil using the Agilent 7700x ICP-MS is described. Experimental Equipment An Agilent 7700x ICP-MS was used with a standard glass concentric nebulizer, quartz spray chamber and connector tube. Sample uptake was by self aspiration, using a 70 cm long sample capillary tube with 0.25 mm internal diameter. The instrument s peristaltic pump was used only for draining the spray chamber, using solvent resistant peristaltic pump tubing. The optional platinum (Pt) tipped sampling cone and skimmer cone were used, as the plasma with O 2 addition is much more reactive and would quickly degrade the standard nickel cones. An Agilent I-AS autosampler was used to automate sample introduction over a long sequence. Standards and samples All metallic standards were prepared from Conostan Oil Analysis Standards. Conostan S-21 pre-mixed standard containing 21 metals at 100 ppm was used as the stock standard and by serial dilution of S-21, a working standard solution with the same concentration for each metal was easily obtained. This is very convenient for trace and ultratrace analysis, but when concentrations of some elements are significantly higher than others, S-21 must be modified. In order to determine minor and ultratrace elements simultaneously, single element standards were mixed with S-21. A yttrium standard 2

blended in blank oil was used as the internal standard. Analytical grade O-xylene or kerosene (from Kanto Chemicals, Japan) was used as the blank solvent for dilution of the oil samples. Lubricating oil samples were purchased from various suppliers. Procedure The plasma was ignited while aspirating iso-propyl alcohol (IPA). A 1 ppb tuning solution, prepared by diluting SPEX mixed inorganic standard in IPA, was used for the optimization of plasma and ion lens parameters. In Table 1, the major operating parameters used in this experiment are shown. Most lens parameters were optimized using the auto-tuning function of the instrument. Hydrogen reaction and helium collision modes were used and, apart from the cell gas and flow rate, the operating parameters were identical in all modes. Table 1. ICP-MS operating parameters Mode H 2 He LoHe Stabilization time sec 5 30 30 Plasma RF power W 1600 Sampling depth mm 8 Carrier gas L/min 0.6* Makeup gas L/min 0 Option gas** L/min 0.4 Spray chamber temp C -5 Lenses Extract 1 V 0 Extract 2 V -135 Cell entrance V -30 Cell exit V -60 Octopole bias V -20 Q-pole bias V -16 Cell Gas H 2 flow ml/min 7 0 0 He flow ml/min 0 5 4 Approximately 2 g of each oil sample was taken, weighed precisely and dissolved in kerosene (w/w). The final dilution factor should be 100 or lower. Samples were analyzed using an external calibration method with 300 ppb yttrium as the internal standard. Yttrium was added manually to each sample solution (w/w) prior to analysis. An example calibration set is shown in Table 2. Table 2. Target concentrations of calibration standards Unit CALBLK CALSTD1 CALSTD2 CALSTD3 CALSTD4 Na ppm 0 0.005 0.02 0.1 0.5 Mg ppm 0 0.5 2 10 50 P ppm 0 0.5 2 10 50 Ca ppm 0 2 8 40 200 Zn ppm 0 0.5 2 10 50 Mo ppm 0 0.1 0.4 2 10 Al ppb 0 0.2 0.8 4 20 Fe ppb 0 0.2 0.8 4 20 Cd ppb 0 0.2 0.8 4 20 Ti ppb 0 0.2 0.8 4 20 V ppb 0 0.2 0.8 4 20 Cr ppb 0 0.2 0.8 4 20 Mn ppb 0 0.2 0.8 4 20 Ni ppb 0 0.2 0.8 4 20 Cu ppb 0 0.2 0.8 4 20 Ag ppb 0 0.2 0.8 4 20 Sn ppb 0 0.2 0.8 4 20 Ba ppb 0 0.2 0.8 4 20 Pb ppb 0 0.2 0.8 4 20 *Kerosene uptake was 70 µl/min using self aspiration. **20% oxygen balanced with 80% argon was used. 3

Results and discussion Selection of diluent solvent Xylene is frequently used as a diluent solvent of oil samples, and particularly some crude oil samples are dissolved most easily in xylene. However, considering the stability of metallic standards at concentrations as low as 1 ppb, xylene is not always the best choice for the diluent solvent. Therefore, the analytical results of NIST SRM 1848 were compared between xylene and kerosene, which is available with low trace element impurity levels. Table 3 does not show any significant differences in quantitative results so, in this experiment, kerosene was chosen as the diluent solvent for lubricating oils. Table 3. Example of analysis of NIST Standard Reference Material 1848 (Lubricating Oil Additive Package) Kerosene Xylene Certified Value % (uncertainty) Calcium 0.36 0.36 0.359 (0.011) Magnesium 0.78 0.83 0.821 (0.038) Phosphorus 0.78 0.88 0.788 (0.028) Zinc 0.87 0.93 0.873 (0.022) Detection Limit (DL) and Background Equivalent Concentration (BEC) DLs and BECs calculated from the calibration curve are shown in Table 4. As expected, the figures of merit obtained by ICP-MS were significantly lower than is achievable using ICP-OES, by around 1 or 2 orders of magnitude for both DL and BEC. These values are impressive despite the significant contribution from the elemental impurities present in the diluent solvent. In order to simplify the analytical procedure, commercial reagents were used without further purification. Table 4. DL and BEC of elements in kerosene m/z mode DL (ppb) BEC (ppb) B 10 H 2 0.85 2.1 Na 23 H 2 0.41 2.8 Mg 24 H 2 0.59 0.66 Al 27 LoHe 0.096 0.092 P 31 He 58 95 Ca 40 H 2 0.42 0.68 Ca 44 H 2 0.63 0.60 Ti 49 LoHe 0.14 0.036 V 51 He 0.013 0.004 Cr 52 H 2 0.045 0.016 Fe 54 H 2 0.11 0.078 Mn 55 He 0.031 0.008 Ni 60 He 0.045 0.037 Cu 63 He 0.06 0.17 Zn 66 He 0.12 0.14 Zn 67 He 0.47 0.20 Mo 95 He 0.087 0.11 Mo 97 He 0.049 0.096 Ag 107 He 0.012 0.01 Cd 114 He 0.015 0.008 Sn 118 He 0.064 0.062 Ba 137 H 2 0.018 0.008 Pb 208 He 0.015 0.03 4

Recovery test In order to validate the analytical method proposed here, a recovery test was carried out. A partially synthetic oil sample was diluted and spiked. The results are shown in Table 5. Recoveries for P and Ca were adversely affected by the low concentration of the spikes compared to the unspiked concentrations. Table 5. Spike recovery test Concentration spiked Unspiked Measured Spiked B 50 ppb 4.3 47 85 Na 0.1 ppm 0.023 0.12 99 Mg 10 ppm 0.11 10 101 Al 4 ppb 17 21 107 P 10 ppm 25 36 116 Ca 10 ppm 32 44 121 Ti 4 ppb 4.9 8.7 97 V 4 ppb 1.9 5.8 96 Cr 4 ppb 0.64 5.6 123 Mn 4 ppb 4.4 8.8 108 Fe 4 ppb 1.9 6 103 Ni 4 ppb 1.2 5.5 107 Cu 4 ppb 0.098 4.5 109 Zn 10 ppm 29 39 101 Mo 2 ppm 3.3 5.4 108 Ag 4 ppb 0.029 4.6 113 Cd 4 ppb 3.2 7.5 107 Sn 4 ppb 0 3.9 101 Ba 4 ppb 0.97 5.4 111 Pb 4 ppb 7.2 11 99 Recovery rate (%) Analysis of commercial lubricating oils From an evaluation of the full mass spectra, it was clear that the concentration variation between elements was significant. Some elements must be determined at relatively high concentrations, while others are present at very low concentrations. The proposed calibration strategy was sufficient to cover this range after a dilution factor of 100. The analytical results for minor and trace elements are shown in Table 6 and ultratrace elements are shown in Table 7. Long term stability An oil sample spiked with 19 elements was measured for 5 hours to assess long-term signal stability. Considering the most difficult case, a 10 times diluted oil sample (partially synthetic oil) was used. 100 ppb S-21 standard was spiked into the sample and the spiked concentrations were monitored. The normalized concentrations were plotted in Figure 1 with RSD%. Considering the very low dilution factor (high matrix level), excellent stability was obtained. Normalized concentration 1.40 1.30 1.20 1.10 1.00 0.90 0.80 0.70 Na 0.8% P 2.3% V 1.8% Mn 1.9% Zn 0.8% Mg 0.4% Ca 0.7% Cr 4.5% Ni 2.0% Mo 0.5% Al 2.6% Ti 5.5% Fe 4.0% Cu 2.0% Ag 1.7% Cd 1.6% Ba 1.8% Pb 2.6% 0.60 0 20 40 60 80 100 120 140 160 180 Elapsed time, min Figure 1. Long term stability of spiked elements in 10 times diluted oil sample 5

Table 6. Concentration of minor and trace elements measured in a series of commercial lubricating oils (mg/kg, ppm) 23 Na 24 Mg 27 Al 31 P 44 Ca 54 Fe 66 Zn 97 Mo 114 Cd A Mineral 1.4 11 0.75 1100 3000 0.23 1100 110 0.096 B Mineral 2.6 20 1.1 1900 5100 0.42 1700 200 0.17 C Mineral 1.9 18 1.1 1100 2900 0.33 1100 960 0.78 D Partially synthetic 0.21 2 0.11 5.4 360 0.042 13 1.3 0.0013 E Partially synthetic 0.51 1800 0.6 1800 1200 0.33 1900 2.1 0.0048 F Mineral 2.2 30 0.14 690 1600 0.16 750 0.11 0.0051 G Mineral 9 1100 1.6 1300 1400 0.84 1400 0.051 0.022 H Mineral 2.3 7.8 0.39 780 2000 0.2 870 230 0.19 I Partially synthetic 0.99 13 0.22 780 2200 0.17 840 78 0.069 J Mineral 0.13 8.4 0.69 780 2100 0.09 870 81 0.075 K Synthetic 0.96 9 0.29 750 2100 0.16 810 78 0.075 L Partially synthetic 0.9 3.6 0.54 810 1100 0.16 990 110 0.11 M Mineral 0.69 720 0.48 780 1200 0.21 870 81 0.075 N Partially synthetic 0.99 8.4 0.39 780 2200 0.17 840 81 0.072 O Partially synthetic 0.96 7.5 0.39 720 2100 0.15 780 78 0.069 Table 7. Concentration of ultratrace elements measured in a series of commercial lubricating oils (µg/kg, ppb) 49 Ti 51 V 52 Cr 55 Mn 60 Ni 63 Cu 107 Ag 118 Sn 137 Ba 208 Pb A Mineral 39 63 51 130 29 5.1 n.d. 5.1 11 19 B Mineral 51 100 66 210 45 9.3 n.d. 5.1 20 30 C Mineral 81 54 21 75 26 6 0.45 7.5 10 19 D Partially synthetic 8.7 5.7 2.6 12 2.7 n.d. n.d. n.d. 1.5 4.5 E Partially synthetic 220 51 45 170 150 6.3 0.81 n.d. 17 33 F Mineral 45 17 16 66 36 45 1.1 2.5 6600 75 G Mineral 84 120 48 570 90 20 1.7 6.6 9.9 99 H Mineral 27 9.3 33 93 54 8.7 n.d. 3.9 17 20 I Partially synthetic 120 12 29 87 21 3.6 n.d. 1.8 9.6 14 J Mineral 39 19 17 60 39 6 n.d. n.d. 11 14 K Synthetic 480 8.7 27 87 17 3.6 n.d. n.d. 8.4 14 L Partially synthetic 150 63 20 63 36 3.9 n.d. n.d. 33 240 M Mineral 39 42 20 81 24 3 n.d. 33 4.2 9 N Partially synthetic 24 39 30 87 21 2.7 n.d. n.d. 10 11 O Partially synthetic 21 33 33 84 23 3 n.d. n.d. 7.2 9.3 6

Conclusions By applying ICP-MS to the analysis of metals in lubricating oils, the concentration range of detectable elements could be significantly extended compared to ICP-OES, namely from 0.5 ppb to 5000 ppm. From the standpoints of robustness and stability as well as sensitivity, ICP-MS has demonstrated significantly better performance than either ICP-OES or AAS. References 1. J. R. de Souza, E. F. dos Santos, C. B. Duyck and T. D. Saint Pierre, Spectrochim. Acta, Part B, 66, 368-372 (2011) 2. ASTM D3237-12, Standard Test Method for Lead in Gasoline by Atomic Absorption Spectroscopy, originally approved in 1973, currently approved in 2012. 5. ASTM D5185-09, Standard Test Method for Determination of Additive elements, Wear Metals, and Contaminants in Used Lubricating Oils, originally approved in 1991, currently approved in 2009. 6. ASTM D7111-11, Standard Test Method for Determination of Trace Elements in Middle Distillate Fuels, approved in 2005. 7. D. M. Brum, C. F. Lima, N. F. Robaina, T. C. O. Fonseca and R. J. Cassella, Spectrochim. Acta, Part B, 66, 338-344 (2011) 8. M. Rahmani and M. Kaykhaii, Microchim. Acta, 174, 413-419 (2011) 3. ASTM D4628-05, Standard Test Method for Analysis of Barium, Calcium, Magnesium and Zinc in Unused Lubricating Oils by Atomic Absorption Spectrometry, originally approved in 1986, currently approved in 2011. 4. ASTM D5863-00a, Standard Test Methods for Determination of Nickel, Vanadium, Iron and in Crude Oils and Residual Fuels by Flame Atomic Absorption Spectrometry, originally approved in 1995, currently approved in 2011. 7

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc. 2013 Published October 16, 2013 Publication number: 5991-3263EN