Safety, Efficacy, and Microbiological Considerations. * For use with RAPICIDE PA outside of the USA, Canada and Australia

Similar documents
HIGH-LEVEL DISINFECTANT AND STERILANT

HIGH-LEVEL DISINFECTANT RAPICIDE. Safety, Efficacy, and Microbiological Considerations

High-Level Disinfectant and Sterilant. Rapicide SAFETY, EFFICACY AND MICROBIOLOGICAL CONSIDERATIONS

Rapicide High-Level Disinfectant and Sterilant. Safety, Efficacy, and Microbiological Considerations

Safety, Efficacy & Microbiological Consideration

Safety, Efficacy & Microbiological Consideration

ENVIROCIDE. Technical Bulletin

MAR COR RESEARCH REPORT. Fast, Safe, Reliable Surface Disinfection for the Life Sciences Market. Competent, Consistent, & Compliant

8,000 facilities count on it

Appendix 2. Revised 510(k) Summary

Safety Data Sheet SECTION 1. IDENTIFICATION OF THE SUBSTANCE/PREPARATION AND THE COMPANY/UNDERTAKING

SAFETY DATA SHEET. Identification of the material and the supplier. Rapid cleaning and grime removal. Wiri Auckland

DECON-PHENE Sterile Pharmaceutical Clean Room Formula Germicidal Detergent with Organic Soil Tolerance

Causes serious eye irritation Specific organ toxicity repeated exposure, Oral (Category 3), Central Nervous System

Glyphosate Technical

MATERIAL SAFETY DATA SHEET. In Case of Emergency, Call Syngenta Crop Protection, Inc. Post Office Box Greensboro, NC 27419

Imidacloprid Technical

Bayer Environmental Science

MSDS Number: VORTEX SEED TREATMENT FUNGICIDE MSDS Version 2.0

SAFETY DATA SHEET 1. PRODUCT AND COMPANY IDENTIFICATION. Product name: MontBrite 1240 solution

Issue Date: 29-May-2015 Revision Date 24-May-2015 Version IDENTIFICATION BC.HS-1041 UN1759. Built Laundry Detergent

MATERIAL SAFETY DATA SHEET

FREQUENTLY ASKED QUESTIONS FOR THE ADVANTAGE

SAFETY DATA SHEET FIXER HARDENER 1. CHEMICAL PRODUCT AND COMPANY IDENTIFICATION 2. HAZARDOUS IDENTIFICATION

SAFETY DATA SHEET. MANUFACTURER 24 HR. EMERGENCY TELEPHONE NUMBERS Nova Pressroom Products

SAFETY DATA SHEET SECTION 1 IDENTIFICATION. ALBA-RUST Expert Rust Remover

SAFETY DATA SHEET. Identification of the material and the supplier. Enviro Bathroom & Toilet Cleaner. Rapid cleaning and grime removal.

TOXICITY, EFFICACY, STABILITY AND CORROSION TEST RESULTS

Safety Data Sheet. Classified According to OSHA Hazard Communication Standard (HCS) Company: Ricca Chemical Company

MSDS Number: PONCHO 600 MSDS Version 2.1

CAS No.: & (R,S)-2-[(2,6-dimethylphenyl)-methoxyacetylamino]-propionic acid methyl ester

Safety Data Sheet DEPARTURE HERBICIDE

Safety Data Sheet InVade Bio Drain Release Date: 5/27/2015 Print Date: 5/28/2015 Version 1.0

LEGACYPRO ECO PRO HYPO WASH

BromMax 7.1 CONCENTRATED LIQUID BROMINE BIOCIDE

MATERIAL SAFETY DATA SHEET

2. COMPOSITION/INFORMATION ON INGREDIENTS

SECTION 2: COMPOSITION/INFORMATION ON INGREDIENTS

SAFETY DATA SHEET. Section 1: IDENTIFICATION OF THE MATERIAL AND SUPPLIER. HERITAGE MAXX Fungicide

SAFETY DATA SHEET. Section 1: IDENTIFICATION OF THE MATERIAL AND SUPPLIER. HERITAGE MAXX Turf Fungicide

MATERIAL SAFETY DATA SHEET

SECTION 1: PRODUCT IDENTIFICATION AND USE. with water and used for aerial application to control wildfire. Emergency Telephone Numbers: (24 Hours)

Shield-Brite PAA 5.0 SAFETY DATA SHEET 1. PRODUCT & COMPANY IDENTIFICATION

SERVIAN 75 WG Version 4 Revision Date Print Date

MATERIAL SAFETY DATA SHEET

SAFETY DATA SHEET according to regulation (EC) no 1907/2006 Caste off Version 2 (UK) Issue date 07/30/2008

RTV SILICONE CONFORMAL COATING SECTION 1- CHEMICAL PRODUCT AND COMPANY IDENTIFICATION

SAFETY DATA SHEET. Section 1: IDENTIFICATION OF THE MATERIAL AND SUPPLIER. SUBDUE MAXX Fungicide ABN Section 2: HAZARDS IDENTIFICATION

Material Safety Data Sheet

CHEMICAL & FERTILIZER, LLC Safety Data Sheet

1. CHEMICAL PRODUCT AND COMPANY IDENTIFICATION

Safety Data Sheet SECTION 1. IDENTIFICATION OF THE SUBSTANCE/PREPARATION AND THE COMPANY/UNDERTAKING

Page 1 of 6 AdcoEco Laundry Detergent US GHS SAFETY DATA SHEET

MATERIAL SAFETY DATA SHEET

INGREDIENT NAME: Inert Ingredients: identity withheld as Trade Secret CONCENTRATION: % CAS NUMBER: N/A OSHA-PEL 8HR TWA: N/L

DRANO MAX GEL CLOG REMOVER

Safety Data Sheet Hydrogen Fluoride


ULTRASONIC COUPLANT POWDER. Bright Yellow Fine Powder. Water Soluble forms a thick yellow gel

MATERIAL SAFETY DATA SHEET


Bayer Environmental Science

ORTIVA TOP Version 3 Revision Date Print Date

MSDS Number: MERIT 0.25%G INSECTICIDE MSDS Version 1.0

CAS No.: & (R,S)-2-[(2,6-dimethylphenyl)-methoxyacetylamino]-propionic acid methyl ester

Safety Data Sheet Per GHS Standard Format

MATERIAL SAFETY DATA SHEET

Safety Data Sheet according GB

CAS No.: [2-chloro-4-(trifluoromethyl)phenoxy]-N-(methylsulfonyl)-2-nitrobenzamide, sodium salt

MATERIAL SAFETY DATA SHEET

SAFETY DATA SHEET according to applicable EC directive

INGREDIENT NAME: Inert/Other Ingredients Proprietary Information CONCENTRATION: 99.20% CAS NUMBER: N/A OSHA-PEL 8HR TWA: N/L

SAFETY DATA SHEET. Avalanche. DISTRIBUTOR 24 HR. EMERGENCY TELEPHONE NUMBERS Simplex Products

MATERIAL SAFETY DATA SHEET #NMS700 KNOCKDOWN. CLASS A FIRE FIGHTING FOAM Concentrate. Section 1. CHEMICAL PRODUCT/COMPANY IDENTIFICATION

SAFETY DATA SHEET OASIS PRO CLEAN ESCAPE

MATERIAL SAFETY DATA SHEET Hydrogel

Hydrosonic Solvent 1. IDENTIFICATION

Duro-Last. PRODUCT AND COMPANY IDENTIFICATION Product Name: Duro-Last Version: 2 Plasticized PVC Membrane, all colors

MATERAL SAFETY DATA SHEET

REScue Concentrate One Step Disinfectant Cleaner & Deodorizer

Bayer Environmental Science

SAFETY DATA SHEET. SSS Power Play Neutral Floor Cleaner

NOMIX ENVIRO MATERIAL SAFETY DATA SHEET HILITE Page 1 of 5

NETWORK PRO ONE VERSOL RED FOUNTAIN SOLUTION

Version 0.1 Revision Date 05/29/2015 Print Date 11/10/2015

MSDS Number: GAUCHO 75 ST FS INSECTICIDE MSDS Version 1.3

SAFETY DATA SHEET 1. PRODUCT AND COMPANY IDENTIFICATION. Product name: BoroSpec K Potassium Borohydride Powder

CULTAR Version 3 Revision Date Print Date

Component 2-Propenoic acid, homopolymer, sodium salt

MATERIAL SAFETY DATA SHEET

SAFETY DATA SHEET. Virbac New Zealand Limited. Intracillin 300 Injection. Section 1: Identification of the Substance and Supplier

PEAK 75 WG Version 12 Revision Date Print Date

Safety Data Sheet. Company: Ricca Chemical Company. 448 West Fork Drive Arlington, TX USA

Safety Data Sheet Duro-Last, Inc. Duro-Last Accessories

I. Product and Company Information. II. Hazard Identification. III. Composition / Information on Ingredients. IV. First Aid Measures

ACGIH TLV. Ethylene Glycol (< 10%) Not Established 100 mg/m³ (ceiling) [aerosol] Glycerin 15 mg/m³ TWA (total); 5 mg/m³ TWA (respirable)

MATERIAL SAFETY DATA SHEET

SAFETY DATA SHEET. Exclamation mark. H314: Causes severe skin burns and eye damage. H302: Harmful if swallowed. H401: Toxic to aquatic life.

SAFETY DATA SHEET. Section 1: IDENTIFICATION OF THE MATERIAL AND SUPPLIER ADVION ANT GEL. For the control of Ants ABN

Transcription:

Safety, Efficacy, and Microbiological Considerations * For use with RAPICIDE PA outside of the USA, Canada and Australia 1

RAPICIDE, MEDIVATORS, ADVANTAGE PLUS and DSD EDGE are registered trademarks of Medivators Inc. RAPIDAER is a registered trademark of Cantel Medical (UK) Limited. 50097-467 Rev C 2015 Medivators Inc. All rights reserved. This publication is protected by copyright. Copying, disclosure to others, or the use of this publication is prohibited without the express written consent of Medivators Inc. MEDIVATORS reserves the right to make changes in the specifications shown herein without notice or obligation. Contact your MEDIVATORS representative or MEDIVATORS customer service for more information. 2

Contents Product Description... 4 Intended Use... 4 Safety... 5 Efficacy... 6 Materials Compatibility... 8 Toxicity... 10 Biocompatibility... 12 Compatibility with endoscopes (500 cycles)... 13 Packaging Configuration... 13 3

PRODUCT DESCRIPTION RAPICIDE PA Solutions A and B are combined to produce a single use-solution for High- Level-Disinfection (HLD) of endoscopes and related accessories with peracetic acid (PAA) as the active agent. The characteristics of the single use-solution are: Active Ingredient: Peracetic Acid PAA Concentration: 1000-1300 ppm Minimum Recommended Concentration (MRC): 850 ppm Reuse Period: Single use High-Level Disinfection Claim: 5 Minutes at room temperature (minimum of 20 o C) Sterilization Claim: 10 minutes at room temperature (minimum of 20ºC) Dilution Required: 1:1 to 48 parts water Activation Required: No Corrosion Inhibitors: Yes RAPICIDE PA is a single use liquid chemical germicide solution with PAA in the concentration range of 1000-1300 ppm. RAPICIDE PA also contains corrosion inhibitors and water conditioning agents. INTENDED USE RAPICIDE PA is intended for high-level disinfection of endoscopes and related accessories in MEDIVATORS ADVANTAGE PLUS, DSD EDGE or RAPIDAER automated endoscope reprocessors (AER). RAPICIDE PA is intended for single use at room temperature (minimum of 20 C, may also be used at 30ºC). In appropriate AER s, RAPICIDE PA High-Level Disinfectant Test Strips may be used to verify that the disinfectant concentration remains above the minimum recommended concentration (MRC). ADVANTAGE PLUS and DSD EDGE AER s contain a sample port where the RAPICIDE PA use solution can be dispensed for monitoring of the MRC. MEDIVATORS RAPIDAER reprocessor utilizes built in monitoring where test strips are not required. Read and follow the directions for use and precautionary statements contained in the package insert before using. 4

SAFETY In handling RAPICIDE PA, the following precautions should be followed: RAPICIDE PA parts A & B are corrosive. Always wear proper eye protection, and protective, chemically-resistant gloves when handling RAPICIDE PA. Avoid contact with eyes, skin and clothing. In case of contact with eyes or skin, flush with copious amounts of water for a minimum of 15 minutes. Seek immediate medical attention in case of eye contact. Avoid breathing fumes or vapors. Prepare and use RAPICIDE PA only in well ventilated areas. Harmful if swallowed. If swallowed, drink large amounts of water. Do not attempt to induce vomiting. Call a physician immediately. Do not store RAPICIDE PA near or with food. Refer to product insert and manufacturer s instructions for reprocessing of medical devices in AER units. STORAGE Store RAPICIDE PA at temperatures of 25 o C (77 o F) or lower. Protect from freezing. NEUTRALIZATION AND DISPOSAL Dispose of this product in accordance with all applicable Federal, State and Local regulations. If allowed by federal, state or local regulatory authority, flush spilled solution of RAPICIDE PA (Solutions A & B combined) to the drain using cold water. Run cold water for 15 minutes. Do not reuse the empty containers. Triple rinse the containers with water and dispose of in an incinerator or in accordance with local, state or federal requirements. For more information on spills and disposal refer to the RAPICIDE PA spill cleanup and disposal document. 5

EFFICACY High-Level Disinfection Summary RAPICIDE PA will high-level disinfect with an exposure time of 5 minutes at room temperature (minimum 20 o C) at a concentration equal to or greater than the minimum recommended concentration (MRC) of 850 ppm PAA. Sterilization Summary RAPICIDE PA will sterilize (>99.9999% reduction of bacterial endospores) with an exposure time of 10 minutes at room temperature (minimum 20ºC) at concentration equal to or greater than the minimum recommended concentration (MRC) of 850 ppm PAA. Efficacy Testing RAPICIDE PA was tested at 850 ppm PAA under case conditions according to AOAC methods and other applicable standards for microbiological testing for germicidal efficacy. Tests demonstrated sporicidal, bactericidal, fungicidal, tuberculocidal and virucidal efficacy. 6

Table 1: SUMMARY OF EFFICACY TESTING RAPICIDE PA 5 minutes at 20ºC TESTS ORGANISM RESULTS Sporicidal B. subtilis >4.5 log reduction* Tuberculocidal M. bovis M. terrae >4 log reduction Virucidal Adenovirus type 5 >6 log reduction Fungicidal Yeast C. albicans Total Kill Bactericidal P. aeruginosa Total Kill Bactericidal S. aureus Total Kill Bactericidal E. feacalis E.hirae Total Kill Simulated Use M. terrae >6 log reduction * >6 log reduction at 10 minutes 7

MATERIAL COMPATIBILITY RAPICIDE PA High-Level Disinfectant is compatible with the following reusable devices and materials: flexible endoscopes, rubber, plastic, many types of metals (such as stainless steel, carbon steel, and aluminum). RAPICIDE PA High-Level Disinfectant is recommended for use with medical devices made from materials shown in Table 2 below. These devices may be damaged if cleaned with highly alkaline detergent, poorly rinsed after disinfection, stored wet, or dried at temperatures exceeding 40 C (105 F). TABLE 2: List of Compatible Materials Metals Plastics Elastomers Hastelloy Polyethylene Aflas 301 Stainless steel Polyvinylchloride Silicone Rubber 304 Stainless steel Teflon (PTFE) Viton 316 Stainless steel Polyurethane Santaprene Inconnel PolysuIfone EPDM Polyester Polycarbonate Nylon PVDF Polypropylene PETG Acetal The compatibility of RAPICIDE PA has been examined both by running intensive scope exposure tests, and also by doing materials compatibility testing versus the "use" solution (as well as the pure part A and part B). In addition, biocompatibility testing has been done in order to demonstrate the rinsing of the chemical components from the scopes. 8

The endoscope compatibility testing was run using Olympus model TJF-1 and model CF-100l scopes, subjecting them to 500 cycles on the ADVANTAGE PLUS reprocessor. Each cycle consisted of a 2 minute exposure to detergent, rinsing, a 6 minute disinfection cycle (worst case), followed by rinsing and alcohol. The scopes were inspected before and after the 500 HLD cycle run. The inspection was done by Surgical Repair Technologies (Eagan, MN), involving functional evaluation of angulation, control, biopsy, insertion tub, light guide, suction, light guide connector, eyepiece, image, light illumination, air/water function, and leak testing. No change was observed in physical appearance or functionality, and the scopes passed all leak tests. We performed chemical exposure lifetime testing on exposed components of the ADVANTAGE PLUS reprocessor. The exposure to the Use Solution (after mixing and diluting the parts A & B to the use dilution) was performed on metal, plastic and elastomer wetted components. The metal components included several types of stainless steel (SS 304, 301, 316), Inconnel springs, ceramic and glass. All components passed the testing. The plastic and elastomer component included EPDM, PETP, nylon, PFA, PP, PTFE, PVC, PVDF, Viton, silicone, santoprene, PS, and proprietary materials such as Iglidur and Gelcoat. All components passed the testing. Chemical exposure lifetime testing was also performed on materials that would be exposed to Part A alone (undiluted) and also part B alone (undiluted). These materials included stainless steel (316) and Hasteloy, PVC, PP, EPDM and Aflas. All components passed the testing, except for slight effects on two EPDM components (which are part of preventative maintenance and were replaced). The chemical exposure testing for both the Use Solution and Part A or Part B alone included dimensional, weight, visual and infrared analysis. In addition to the chemical compatibility testing, a study was done to test the toxicity of rubber components (from Olympus parts) that were exposed to 100 HLD cycles. These materials are from the distal end of an endoscope, which will make patient contact. It was concluded that the RAPICIDE PA had a non-toxic effect on the rubber parts, based on results of cytoxicity testing after the 100 cycles. This test is governed by ANSI/AAMI/ISO 10993-5:1999. Biocompatibility testing was also done to examine the residuals on endoscopes after 1 and 100 cycles of HLD exposure. It was concluded that the ADVANTAGE PLUS reprocessor adequately rinses detergent and disinfectant residues from the surfaces of the endoscope, making it safe for both patients and hospital staff. The above testing indicates the RAPICIDE PA is safe to use on endoscopes and the exposed materials. No effect was observed on the endoscope components or functionality. 9

TOXICITY RAPICIDE PA has been extensively tested for potential toxicity to both user (environmental) and patient. Summary of Environmental Toxicity Standard toxicological testing was performed to evaluate the possible toxicological effects from inhalation, oral consumption, eye contact, skin contact, dermal irritation, dermal sensitization and genotoxicity. Table 3 summarizes the environmental test results. Table 3: Summary of Environmental Toxicity Tests Test Test Results Acute Oral Toxicity (Rat) Acute Inhalation Toxicity (Rat) Primary Eye Irritation Acute Dermal Toxicity (Rabbit) Repeat Dermal Toxicity (Rat) Primary Skin Irritation (Rabbit) Skin Sensitization (Guinea Pigs) Genotoxicity (Ames) LD 50 > 5 g/kg LC 50 > 2.1 mg/kg Corrosive to ocular tissue May produce systemic effects at LD 50 > 2 g/kg RAPICIDE PA is not associated with observed signs of toxicity at 1000 mg/kg Not an irritant Not an allergen Non-mutagenic Summary of Patient Toxicity Standard toxicological tests were performed to evaluate the possible toxicological effects by cytotoxicity, mucosal irritation and hemolysis. Table 4 summarizes the patient test results. 10

Table 4: Summary of Patient Toxicity Tests Test Cytotoxicity (RAPICIDE PA) Cytotoxicity (A-rubber) Hemolysis Mucosal Irritation Acute Intravenous Toxicity Test Results Non-toxic at 1:500 dilution Non-toxic Non-Hemolytic at 1:1000 dilution Minimal irritant LD 50 > 5 g/kg 11

BIOCOMPATIBILITY - A study was conducted to determine the amount of peracetic acid (PAA) and hydrogen peroxide (HP) residue that remain on endoscopes after reprocessing in an AER with RAPICIDE PA High-Level Disinfectant and Sterilant, and rinsing with water. The study used RAPICIDE PA at the maximum possible use-solution concentration of PAA and exposed three pediatric colonoscopes to the longest labeled exposure time, a five hour cycle. The pediatric colonoscope was used in the study because it is representative of an endoscope that is most difficult to rinse due to its long narrow channels and large surface area. Following a five hour cycle in the ADVANTAGE PLUS AER, the three endoscopes were immediately and exhaustively extracted for 24 hours to remove all PAA and HP residues. This was followed by an additional 4 hour residue extraction. From the extractions, a worst case of human observable exposure dose was calculated to be 0.9 mg/kg and 1.11 mg/kg for PAA and HP, respectively. The worst case human observable exposure dose limits was compared to the reported rat Repeat Dermal No Observed Effect Level (NOEL) doses of 4.4 mg/kg and 5.4 mg/kg for PAA and HP, respectively. The risk to humans is shown to be close to five fold lower than to the rat Repeat Dermal NOEL. These extractions were performed over a 24 hour period, which is a great deal longer than a typical endoscopy procedure of 15 minutes to two hours. In relating these residual values to realistic patient exposure, we conclude that these PAA and HP levels are not toxic if introduced into the human body. 12

COMPATIBILITY WITH ENDOSCOPES (500 CYCLES) An ADVANTAGE PLUS reprocessor was operated with RAPICIDE PA in two types of endoscopes for a minimum of 500 cycles. The AER was set to 32º ± 2º C for worst case simulation. At the end of the cycles the endoscopes were examined by Surgical Repair Technologies. No damage was detected, and the endoscopes functioned properly, demonstrating that RAPICIDE PA High Level Disinfectant is compatible with flexible endoscopes (MEDIVATORS QP 201745). PACKAGING CONFIGURATION RAPICIDE PA is available in the following packaging configuration. One package contains 20 liters of use-solution. Packaging: 4 Bottles per Case (20 Liters) 2 Bottles of Part A X 5 Liters 2 Bottles of Part B X 5 Liters each bottle weighs approximately 10.5 lbs. 13