THE BASICS OF STERILIZATION

Similar documents
Objectives. The Perioperative Nurse s Role 6/24/2016. Instrument Processing & Sterilization Chesapeake Perioperative Consortium 2016

8/27/2015. Learning Objectives. Purpose of Sterilization. To destroy all living pathogenic microorganisms

Ideal Sterilization Method. Sterilization. Sterilization Practices in Healthcare Facilities

The Gold Standard in Low Temperature Sterilization: Ethylene Oxide

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of

6 Handling, collection, and transport of contaminated items General rationale Separation of waste and reusable items at point of

Sterilization Policy. Georgia Regents Medical Center Policy Library. Policy Owner: Epidemiology POLICY STATEMENT

Packaging & Sterilization

8/23/2012. Continuing Education Contact Hours. How much of your formal education focused on sterilization or sterile processing?

Disinfection & Sterilization

3/26/2014 SPAULDING CLASSIFICATION OBJECTIVES PRINCIPLES OF DISINFECTION AND STERILIZATION IN THE DENTAL SETTING PROCESSING SEMI CRITICAL OBJECTS

PRINCIPLES OF DISINFECTION AND STERILIZATION IN THE DENTAL SETTING OBJECTIVES

Low Temperature Processing. An Update on Guidelines, Standards and Requirements

ARABIC CANADIAN MEDICAL CENTER CSSD POLICIES AND PROCEDURES

Disinfection and sterilisation

Quality Control EQUIPMENT CONTROL. Monitor critical performance characteristics of sterilization, decontamination or cleaning equipment

FAQFrequently. Asked Questions. Education & Training. 3M Sterilization Assurance Techline

Welcome! How do I get a CE Certificate?

Responses to Questions for SYSTEM 1E Process Monitoring and Validation Webinar

STERILIZATION VALIDATION IN THIS SECTION REPROCESSING VALIDATIONS FOR REUSABLE MEDICAL DEVICES. Radiation Sterilization Validation

Choose a category. You will be given the answer. You must give the correct question. Click to begin. Instructions for using this template.

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces

HemoCheck TM Sample Policy. SUBJECT: Detection of blood residue on various surfaces

Instructions for Validation of Autoclaves

Manner of Assessment Code (check all that apply) & Surveyor Notes. Yes N/A. Yes. Yes N/A N/A. Yes N/A. N/A Yes N/A

UNIT Two. Lesson 1C Sterilization

HANDLING OF WRIGHT DISPOSABLE PROPHECY ANKLE INSTRUMENTS

Guidelines for Selection and Use of Disinfectants

Definitions. BIOL 3702: Chapter 8. Control of Microbes in the Environment. Mechanical Removal Methods. Pattern of Microbial Death

DEPARTMENT OF THE ARMY TECHNICAL BULLETIN

Sterilization, disinfection and reprocessing

Medical Device Reprocessing Assessment - General (Cleaning, Disinfection, and/or Sterilization)

Definitions. BIOL 3702: Chapter 8. Control of Microbes in the Environment. Mechanical Removal Methods. Pattern of Microbial Death

Administrative Policies and Procedures. Cross Reference: Date Issued: 2/14 Date Reviewed: 9/14 Date: Revised: 12/14 Attachment: None Page of 1 of 6

المملكة العربية السعودية وزارة الصحة اإلدارة العامة لمكافحة عدوى المنشئات الصحية

CAN STERILE REALLY BE STERILE?: EFFECTIVE AUTOCLAVING METHODS

Bio-Burden Reduction in Biological Laboratories

2. Describe correct actions to take to address sterilization process failures. 3. List possible causes of sterilization process failures.

COLLEGE OF PHARMACY STERILE PRODUCTS PHT 434. Dr. Mohammad Javed Ansari, PhD. Contact info:

3/11/ M Sterile U Network 3M Sterile U Web Meeting March 11, Welcome! Topic: An Overview of ANSI/AAMI ST58: 2013

Inhibiting of Microbial Growth in vitro CLS 212

Sterilization - validation, qualification requirements. Sterilization - Overview

Quality Control LOAD CONTROL. VERIFY Load Control Products. Use to monitor and release instrumentation with confidence RECORD KEEPING & LABELING

Understanding ANSI/AAMI ST 91. Flexible and Semi-Rigid Endoscope Reprocessing in Healthcare Facilities

Quality Assurance with Rigid Containers Katrina Simpson, M.A., CST, CSPDT

Sterilization and Disinfection: Getting It Right. Instructions

They can be used on washer-disinfectors, ultrasonic cleaners, surgical instruments and endoscopes. Page 5

Your Sterile Processing Partner. Duraholder IPS

Sterilization, Disinfection and Antisepsis

Lets Talk Indicators Challenging the process Sterilization. ICAN 5 th November Peter Newson

Your Sterile Processing Partner. Duraholder IPS

TEST CODE DEFINITION STERRAD H202. Test Code Test Name Testing Description. Cycle. Efficacy. SE-200 STERRAD 200 Efficacy. Cycle. STD Cycle Efficacy

Best Practices for High Level Disinfection Part I

Monitoring your decontamination and reprocessing cycle

Facility: Sterile Processing Assessment Date: CRITERIA ANSI/AAMI COMPLIANT Number Y N N/A Point of Use Gross contaminant is removed with

SUBJECT: Detection of Blood residue inside the biopsy channel of a scope

Low temperature sterilization with hydrogen peroxide vapour; V-PRO TM applications & best practice

Steam Sterilization Monitoring in the New CSA Standard (Z )

Putting Confidence in Your Hands.

Inhibiting Microbial Growth in vitro. CLS 212: Medical Microbiology Zeina Alkudmani

Christina Bradley Endoscope Decontamination A Webber Training Teleclass. Christina Bradley

Disinfection & Sterilization in the Ambulatory Surgery Center

Cross Infection Control

THE TERMS VALIDATION AND VERIFICATION ARE FREQUENTLY

Standard Operating Procedure Safe Autoclave Operations

Guide for Cleaning, Sterilization and Storage of Introducers for Caldera Medical Implants

Is Your Sterile Processing Department Ready for Survey?

BIO-MEDICAL WASTE MANAGEMENT- CHALLENGES IN INDIA

QUALITY ASSURANCE IN AN MDRD

3M s Sterilization Tech Line FAQs Sandra Velte, BA, CSPDT

Sterilization & Chemical Resistance of Healthcare Polymers

Options for load release

Specialty Testing of Steam Sterilizers

Best Practice Monitoring for Sterile Processing Professionals

MDRAO ONLINE COURSE FOURTH EDITION 2018

Guide for Cleaning, Sterilization and Storage of Reusable Introducers

Ethylene Oxide Sterilant Alternatives

One such problem is blood residual inside the channels of a robotic arm.

3.03 Understand support services

3.03 Understand support services

FACT SHEET : Using Autoclaves Safely

Autoclave Operation and Performance Verification

SUBJECT: Detection of Blood (hemoglobin) residue inside the biopsy channel of an endoendoscope

Creating the Improved Standard in Healthcare Sterilization

What Every Infection Preventionist Should Know About the Critical Issues in the Steam Sterilization Process

ANSI/AAMI ST79 A2:2009 Key Changes in the 2009 Amendment

5.1.1 & Changes started in 2008 and were mainly driven under the Chairmanship of Prof Hans van Doorne 11/10/2017

How is Sterilization Validation Performed by the Medical Device Manufacturer?

Competency Guide: Care and Handling of Rigid Endoscopes

Workshop on Medical Device Cleanliness: How Clean is Clean Enough? Sponsored by ASTM Committee F04 on Medical and Surgical Materials and Devices

Animal cell and tissue culture. Lab 1

Contact: To sterilize or decontaminate uniformly, superheated steam must contact all areas of the load.

26/01/2016. A Day in the Life Steam Sterilization Reprocessing. Welcome! House Keeping. From the GoToWebinar page:

HyperGrip Series Product Sterilization White Paper

gke Steri-Record Testsets and PCDs for validation, process and batch monitoring of sterilization processes

HANDLING OF WRIGHT MEDICAL DISPOSABLE PROPHECY TM ANKLE INSTRUMENTS

Effective Use of Autoclaves

Sterilization & Disinfection

What Dentistry Can Learn from AAMI ST79. Best Practices for device and instrument reprocessing

Nitrogen Dioxide (NO 2 ) PDA: A Global. Sterilization and. Association. Decontamination Systems. November 4, PDA Europe Parenterals 2014

Transcription:

THE BASICS OF STERILIZATION Objectives State the importance of sterilization to patient care Review three key essentials of the sterilization process Describe sterilization methods used, application, limitations, and risks of each Discuss the parameters associated with each method of sterilization Review monitoring and record keeping requirements Introduction The proper sterilization of medical devices, surgical instruments, supplies and equipment utilized in direct patient care and surgery is a critical aspect of the modern health care delivery system and directly impacts patient safety. The Association for the Advancement of Medical Instrumentation (AAMI) defines sterilization as: A process designed to remove or destroy all viable forms of microbial life, including bacterial spores, to achieve an acceptable sterility assurance level. Sterility is measured by probability expressed as sterility assurance level (SAL). It is generally accepted that a SAL of 10-6 is appropriate for items intended to come in contact with compromised tissue, which has lost the integrity of natural body barriers. This would include sterile body cavities, tissues and the vascular system. A sterility assurance level SAL of 10-6 means that there is less than or equal to one chance in a million that a particular item is contaminated or unsterile following a sterilization process. Reusable medical devices, including surgical instruments that enter normally sterile tissue or the vascular system require sterilization before each use. Improperly sterilized or contaminated medical devices utilized in patient care can contribute to surgical site infection and pose a serious risk to the patient s safety and welfare and can result in a serious lifethreatening infection or even death.

Sterilization is a complex process. It is imperative that healthcare workers and technicians involved in the reprocessing and sterilization of medical devices have a thorough knowledge and understanding of the scientific principles and methods of sterilization utilized in today s health care settings. These standards and principles apply and are a critical component of infection control regardless of where patient care is provided, be it an acute care hospital, ambulatory surgical center, out-patient facility, dental or physician s office. Key Essentials of Sterilization Process The efficacy of any sterilization process is contingent on the following three essentials: 1) Conditions must be present to effectively destroy living organisms. In other words, the sterilant and sterilizing equipment must be validated and appropriate in design and operation to achieve the correct pressure, time, temperature and sterilant combination to be lethal to microorganisms. 2) Devices to be sterilized must be thoroughly cleaned and decontaminated to reduce bioburden (soil), in order to ensure the effectiveness of the sterilization process. The higher the bioburden, the greater the challenge to the sterilization process. If bioburden is too great, the established sterilization parameters may not be adequate rendering the sterilization process ineffective. 3) There must be direct, complete and adequate contact between the sterilant and all surfaces and crevices of the device to be sterilized. Achieving sterilization is not only contingent on achieving these three essentials but it is also important that medical devices are properly inspected, assembled, packaged, and loaded into the sterilizer in accordance with established procedures and techniques. It is imperative to obtain and follow all manufacturers instructions for use and reprocessing of medical devices. This includes patient care equipment, instruments, sterilants, sterilization systems and sterilizers. All methods of sterilization are intended to kill microorganisms; therefore, one must be mindful that sterilants and sterilizing equipment can be hazardous to the health care worker. The inherent risks associated with various sterilants and equipment can be greatly minimized through staff

education, observing appropriate precautions, adhering to established policies and procedures and following the manufacturers instructions for use. Sterilization Methods and Parameters Sterilization involves the use of a physical or chemical procedure to destroy all microbial life, including highly resistant bacterial spores. The major sterilizing agents commonly used in healthcare facilities today are a) saturated steam, b) hydrogen peroxide gas plasma, c) ethylene oxide gas, d) ozone and e) liquid chemicals. Dry heat is also used, although less commonly. Steam Saturated steam, under pressure, is the oldest and most widely used, economical, effective and reliable method of sterilization available to health care facilities. The steam sterilizer consists of a pressurized chamber. Increasing the pressure in the chamber elevates and holds the temperature. The sterilizer chamber and all contents must be free of any air entrapment to ensure direct contact of the steam to all surfaces to be sterilized. Steam sterilizers are designed to eliminate all air from the chamber during the conditioning phase in the sterilization cycle. Steam is vaporized water and serves as the conduit to rapidly permeate packaging delivering high-temperature, moist heat to all contents and destroying microorganisms. Steam kills microorganisms by coagulating and denaturing the cell protein. Due to the very high temperatures and moisture associated with steam sterilization it may only be used with heat and moisture stable medical devices, instruments and compatible materials. The different types of steam sterilization processes include: Dynamic air removal, also known as pre-vacuum, high-vacuum, or mechanical air removal. Dynamic air removal steam sterilizers utilize a mechanical vacuum pump to suck air from the sterilizer chamber during the conditioning phase of the sterilization cycle. Since the removal of air is mechanical the total cycle time is less than that of a gravity cycle. Gravity, a downward air displacement process, relies on gravity to remove air from the sterilizer chamber. Since steam is lighter than air, as it enters

the sterilizer chamber, it rises to the top. As the chamber fills with steam, the steam forces the air down and out of the chamber drain. An abbreviated steam sterilization cycle, known as Immediate Use Steam Sterilization (IUSS), can be accomplished in either a pre-vacuum or gravity displacement sterilizer. IUSS sterilization is intended for emergency use when instrumentation is urgently needed, and time does not allow for standard routine processing. These items are intended for immediate use and should not be stored for later use. Steam Parameters The essential conditions for steam sterilization are temperature, saturated steam, time and pressure. Recommended minimum sterilization times for wrapped instruments: Prevacuum: 270 o F; 4 min exposure; 20-30 min drying time Gravity displacement: 250 o F; 30 min exposure; 45 min drying time Recommended minimum immediate use sterilization times for non-porous items only (i.e., no lumens) Prevacuum: 270 o F; 3 min exposure; no drying time High-speed gravity 270 o F; 3 min exposure; no drying time Recommended minimum IUSS sterilization times for metal items with lumens and porous items (e.g., rubber, plastic) sterilized together Prevacuum: 270 o F; 4 min exposure; no drying time High-speed gravity 270 o F; 10 min exposure; no drying time When sealed containers are used for IUSS sterilization, the manufacturer s written instructions for exposure times and drying times should be followed, as the times may be extended. Ethylene Oxide Ethylene Oxide (EtO) is a low temperature sterilization method that has been used for many years. The low temperature process makes it suitable for the sterilization of heat and moisture sensitive medical devices that

cannot tolerate the high temperatures and moisture associated with steam sterilization. EtO is a colorless gas, which destroys microorganisms by a process called alkylation. The EtO penetrates the cells membrane and reacts with the nuclear material rendering it unable to metabolize and reproduce. EtO does have some disadvantages, including a relatively long sterilization cycle and the need for aeration for a specified amount of time, after sterilization. EtO is toxic and porous materials exposed to EtO will absorb the toxic gas requiring aeration following the sterilization process to allow the residual gas to escape from the medical devices before they are handled and used. EtO is extremely flammable in its pure form and requires special precautions be taken in its storage and use. Furthermore, EtO is also a known carcinogen. EtO presents a work place hazard and the Occupational Safety and Health Administration (OSHA) has developed permissible exposure limits and requires continuous monitoring of the work area and personnel. The Environmental Protection Agency (EPA) has also established regulations to protect the environment from EtO exposure. Sterilizers must be designed to address certain requirements for ventilation and exhaust and users must monitor the process to ensure that safety and environmental standards are followed properly. Ethylene Oxide Parameters There are four primary conditions/requirements to effectively sterilize instruments with EtO (1) gas concentration (2) humidity, (3) temperature and (4) time. A change in any one of these conditions will require an adjustment of the other conditions. For example, if the temperature is decreased the exposure time must be extended. Because of the vast differences in the configuration, density, design and permeability of products routinely sterilized in the hospital setting, standard cycle parameters have been established to accommodate the most challenging of circumstances. EtO sterilizers are designed and programmed to operate at the required concentration and humidity levels for the established routine hospital cycles parameters.

Relative humidity 50 75% (items should be completely dried prior to packaging and being placed in the sterilizer) Temperatures range from 85-145 o F depending on the cycle chosen, chamber size, and gas blends (manufacturers instructions must be followed) Exposure Time is usually 2 hours or longer depending on load capacity, density and porosity Aeration times and temperatures are 8 hours at 140 o F and 12 hours at 122 o F. Hydrogen Peroxide Gas Plasma H 2 O 2 Gas Plasma is a low temperature sterilization alternative that has been available for several years and is suitable for many heat-sensitive and moisture-sensitive or moisture-stable medical devices. Unlike EtO sterilization, gas plasma sterilization is devoid of the occupational, environmental and patient safety concerns. Gas plasma is also less expensive and total cycle times are significantly less than using EtO. Medical devices are ready for use following sterilization without the need for aeration. This method disperses hydrogen peroxide in a vacuum chamber, creating a gas plasma cloud. Reactive species are generated from the hydrogen peroxide that are reactive with microorganisms. Gas Plasma sterilizes by oxidizing key cellular components of the microorganisms, which inactivates and destroys them. There are some restrictions on lumened devices validated for sterilization by this method, based on the gauge and length of the device. Users should obtain documentation from the manufacturer for appropriate use and restrictions. Gas plasma is not compatible with highly porous absorbers, such as cellulose and paper products and cannot be utilized to process liquids. Hydrogen Peroxide Gas Plasma Parameters Temperature range varies but is maintained between 104-131 F. Total cycle time will range between 28-75 minutes, depending on the sterilizer model and size.

Liquid Chemical Sterilization Liquid chemical sterilization is utilized for the sterilization of heat-sensitive devices that can be immersed. This method employs the use of a germicidal solution and requires the complete immersion of items in the solution for a prescribed period of time to kill microorganisms. Peracetic acid is a liquid chemical sterilant used in conjunction with a selfcontained automated processor designed for this method of sterilization. It is commonly used for flexible endoscopes and components. Devices that are sterilized by this method are intended for just-in-time use and have no shelf life. Glutaraldehyde is also a liquid chemical sterilant, when used according to the manufacturer s directions for sterilization; however, it is usually used as a high-level disinfectant. There are several disadvantages of using glutaraldehyde as a sterilant. These disadvantages include toxic fumes, long exposure time, potential for contamination of sterile devices during rinsing and transfer to the point of use and no method to biologically monitor the sterilization process. A well-ventilated work environment is required and personal protective equipment such as gloves, goggles, masks, aprons and gowns are necessary to protect the staff performing the process. Chemical sterilants can pose a risk to health care workers and patients, if they come in contact with tissue, skin, mucous membranes or eyes. Liquid Chemical Sterilization Parameters Items must be totally immersed in the chemical, allowing direct contact to all surfaces for a prescribed period of time to achieve sterilization. Peracetic acid comes in a single use, sealed container at a precise concentration. Peracetic acid requires 12 minutes of exposure and a temperature of 122 o F. Glutaraldehyde is activated once a container is opened by adding the premeasured activating chemical, which comes with each bottle. Glutaraldehyde requires 10 hours of exposure.

Dry Heat Dry heat sterilization should be used to sterilize anhydrous (waterless) items that can withstand high temperatures. Dry heat sterilizers are not commonly found in healthcare facilities today. If used, it is generally to sterilize talcum powder for surgical procedures. Dry heat sterilization may be used to sterilize sharp instruments, such as dental instruments, burrs, and reusable needles that would be damaged by the moisture of steam. Dry heat sterilization is accomplished by conduction where heat is transferred from molecule to molecule or from the exterior surface of an item to its internal parts. The destruction of organisms occurs by oxidation, which is a slow burning up process of coagulating the cell s protein. It is a long sterilization process due to the length of time it takes for objects to reach required temperatures. Unlike steam sterilization, there is no moisture present, which speeds up heat penetration. Dry Heat Parameters There are great variations in times and temperatures based on volume, density, packaging and sterilization apparatus. It is essential to follow the manufacturer s instructions for the items being sterilized and the sterilizer being utilized. Settings range from 30-minute exposure at 356 F to 6-hours exposure at 250 F. These are the exposure times only and do not include the time required for all parts to reach the desired temperature. Ozone Ozone sterilization is the newest low-temperature sterilization method and is suitable for many heat-sensitive and moisture-sensitive or moisturestable medical devices. Ozone is a strong oxidizer and is generated within the sterilizer using medical grade oxygen, water and electricity. Ozone is inexpensive, non-toxic and environmentally friendly.

Ozone sterilization is compatible with stainless steel instruments however, there are some lumen restrictions. Currently, ozone sterilization is not cleared to process flexible endoscopes, glass or plastic ampules, liquids or implants. Information regarding device compatibility should be obtained from the manufacturer of the device and the sterilizer manufacturer. Packaging materials that can be used for ozone sterilization include anodized aluminum containers with non-cellulose disposable filters, polyethylene pouches and uncoated nonwoven wrap. Ozone Parameters The cycle time is approximately 4.5 hours at a temperature of 85 o F - 94 o F. Quality Assurance A high degree of sterility assurance is of critical importance for the sterilization process. For safe, effective sterilization practice it is essential that staff performing the sterilization function have an in-depth-knowledge and understanding of the scientific principles and methods of sterilization, parameters, applications, and risks associated with each method. Staff must be adequately trained to perform the sterilization function and demonstrate their competency, annually. Process controls including the healthcare facility s policies and procedures for sterilization and the sterilizer manufacturer s written instructions for sterilizer operation should be readily available and consistently followed. Each sterilization cycle should be monitored with mechanical, chemical, and biological indicators to verify that parameters required for sterilization have been met. Mechanical indicators such as recording charts and computerized printouts show the conditions of the sterilizer during the cycle. An air removal test (Bowie-Dick Test) of prevacuum steam sterilizers should be performed daily in an empty chamber. The appropriate indicators should be placed on the outside of packages to be sterilized, to provide evidence of sterilization process completion (e.g., autoclave tape, load cards, etc.). Chemical indicators or integrators should be placed inside each package being sterilized, to provide results after sterilization as to

whether the condition(s) measured for sterilization were met on a packageby-package basis. The appropriate biological and chemical indicator should be used in steam, ethylene oxide, hydrogen peroxide gas plasma, dry heat, peracetic acid, and ozone sterilizers. The biological indicator (BI) spore, Geobacillus Stearothermophilus should be used for testing steam, gas plasma, paracetic acid, and ozone sterilizers. Bacillus atrophaeus should be used for ethylene oxide and dry heat sterilizers. A negative BI result is indicative that conditions for sterilization to occur were met. The frequency of biological monitoring should be based on current regulations, recommended practices and the sterilizer manufacturers recommendations: Steam sterilizers- daily and preferably with every load. Gas plasma, paracetic acid, dry heat, and ozone - daily Ethylene oxide and implants every load Additionally, record keeping is required to document essential information such as testing, cycle runs, load contents, chemical and biological indicator results, and the sterilizer operator. This information is typically recorded on a sterilization log. Record keeping is the key to accountability and provides an audit trail. Sterilization records are necessary if there is a recall of items sterilized in-house or if sterilization records are subpoenaed for a lawsuit. Summary Knowing and understanding the scientific principles and various sterilization methods and their proper applications, as well as the monitoring and record keeping requirements, can help to ensure effective sterilization, patient safety, and cost-effectiveness.

References American Society for Healthcare Central Service Professionals, Training Manual for Central Service Technicians, Fifth Edition, 2006, Chapter 7. Association for the Advancement of Medical Instrumentation, Guidelines for the selection and use of reusable rigid sterilization container systems for ethylene oxide sterilization in health care facilities, ANSI/AAMI ST33:1996. Association for the Advancement of Medical Instrumentation, Steam sterilization and sterility assurance in health care facilities, ANSI/AAMI ST46:2002. Association of perioperative Registered Nurses, Standards, Recommended Practices, and Guidelines. Recommended Practices Sterilization, 2006. Central Service Technical Manual, 7th Edition, published by International Association of Healthcare Central Service Materiel Management, 2008 Chapters 9 and 10. Perkins, John J., Principles and Methods of Sterilization in Health Sciences, 2 Ed., Charles C. Thomas, 1983. Reichert, Marimargaret and Young, Jack, Sterilization Technology for the Health Care Facility, Aspen Publishing 1993. Tilton, Gayla and Kaufman, Myrna, Sterilization a Review of the Basics, Managing Infection Control, June 2004