Improved Gas Chromatograph Method for the Analysis of Trace Hydrocarbon Impurities in 1, 3-Butadiene

Similar documents
Fast Refinery Gas Analysis Using the Agilent 490 Micro GC QUAD

Prefractionator for Reliable Analysis of the Light Ends of Crude Oil and other Petroleum Fractions

A Rapid Method for Trace Analysis of Organophosphorus Pesticides in Drinking Water

Select Al2O3 MAPD : a new Inert Al2O3 column for the trace analysis of polar reactive hydrocarbons.

Agilent Gas Phase Solutions

The Analysis of 4-Vinyl-Cyclo-Hexene (VCH) in Butadiene with the DVLS LGI Injector

Alumina BOND/MAPD PLOT Columns!

Direct Injection of Fish Oil for the GC-ECD Analysis of PCBs: Results Using a Deans Switch With Backflushing. Application. Author.

The Use Of Automated Backflush on the 7890A/5975A GC-MS System. Application. Setup Example Using Biomarkers. Authors. Abstract.

Improving GC-MS Method Robustness and Cycle Times Using Capillary Flow Technology and Backflushing

Analysis of Permanent Gases and Extended Hydrocarbons in Natural Gas and Natural Gas Liquids by Capillary Gas Chromatography

Optimizing Productivity and Reliability for Monocyclic Aromatic Hydrocarbon Purity Analysis According to ASTM D7504 on the Agilent 8890 GC System

Large Valve Oven Analyzers for Refinery Gas Analysis. ASTS 2015 Houston Shannon Coleman Application Specialist Agilent Technologies

APPLICATION NOTE. UOP : Analysis of Trace CO and CO 2 in bulk H 2 and Light Gaseous Hydrocarbons by GC

Analysis of USP <467> Residual Solvents using the Agilent 7697A Headspace Sampler with the Agilent 7890B Gas Chromatograph

Olefin Analyzer Solutions for Polymer Grade Ethylene and Propylene

Hydrocarbon Processing Industry Gas Chromatographs. Standard GC Analyzer Solutions

Analysis of Suspected Flavor and Fragrance Allergens in Cosmetics Using the 7890A GC and Capillary Column Backflush Application

Fast Analysis of Extended Natural Gas using the New Transportable INFICON Micro GC Fusion

Mach-RGA. The Fast Refinery Gas Analyzer. Analyzer

Agilent Ultimate Plus Fused Silica Tubing

Fast On-Site Mine Safety Analysis by the Agilent 490 Micro GC

System Overview. Valve oven. Restrictor. Valve. Sample in and out

Agilent Consumables Improve Performance with Shimadzu GC/MS

Agilent 490 Micro GC. The Measure Anywhere Micro GC February 2011

SCION GC Series. The Gas Chromatographers Choice for Separations

SCION GC Series. The Gas Chromatographers Choice for Separations

Application Note. Abstract. Authors. Environmental

ANALYSIS OF PESTICIDE RESIDUES IN DRINKING WATER AS PER BUREAU OF INDIAN STANDARDS USING THE AGILENT 7000 GC/MS/MS WITH PESTICIDES ANALYZER

Using a Micro GC to measure, monitor, and optimize your process. ASTS, Houston June 2015 Shannon Coleman Agilent Technologies

Gas Chromatography SSI-GC-003

Turnkey configurations for gas chromatographic analysis, including instrumentation, methods, supplies and service. Analysis of Natural Gas

Analysis of Atrazine in Drinking Water at the ppb Level Using New Agilent Reversed Phase LC Columns. Application. Author. Abstract.

ADVANCES IN MICRO GAS CHROMATOGRAPHY

Capillary GC Troubleshooting: a Practical Approach

Improving Productivity in Your Environmental and Food Testing Labs with Agilent 7890A/5975C GC/MS System

Standard Test Methods for Aromatics in Mineral Spirits by Gas Chromatography 1

Performance characteristics of the 1260 Infinity Quaternary LC system

Agilent 490 Micro GC Biogas Analyzers

Expanding the Versatility of Micro GC in the Shale Gas Play; The Agilent 490 Micro GC offers unprecedented reliability and flexibility

Capillary Flow Technology: Backflush REDUCE RUN TIME AND INCREASE LABORATORY THROUGHPUT

SCION GC Series. The Gas Chromatographers Choice for Separations

Chem 321 Lecture 23 - Liquid Chromatography 11/19/13

Preparative Purification of Corticosteroids by HPLC; Scalability and Loadability Using Agilent Prep C18 HPLC Columns Application

Performance Characteristics of the Agilent 1220 Infinity Gradient LC system

Analysis of Semivolatiles by GC/FID using the Ultra Inert Inlet Liners with Wool

AGC NovaPRO Process Gas Chromatograph Refinery Applications

ADVANCES IN MICRO GAS CHROMATOGRAPHY

SIMPLIFY AND ACCELERATE GAS ANALYSIS. Micro GC Fusion GAS ANALYZER

Meet the airgc. Ambient air carrier. Top performance. catalytics accelerated

GC Method Development

Fast Detailed Hydrocarbon Analysis by Modified ASTM Method D6730

Ghost Peaks in Gas Chromatography Part 2: The Injection Port

7890A An Essential Partner for the Trace

Pesticide Analysis in a Complex Matrix using a 2D GC/MS System

Confirmation of THC in Oral Fluids Using High-Resolution 2-D GC/MS Application Note

GC Troubleshooting. Frank Michel, Katherine K. Stenerson T HCE

Accurately Identify and Quantify One Hundred Pesticides in a Single GC Run

ABB MEASUREMENT & ANALYTICS ANALYTICAL MEASUREMENT. Ethylene plant production excels with the PGC1000

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE.

CryoTrap. For Agilent Technologies 7890/6890/6850 GC Systems. CryoTrap

DETERMINATION OF ORGANIC PRIORITY POLLUTANTS AND EMERGING COMPOUNDS IN WASTEWATER BY MICRO SOLID PHASE EXTRACTION COUPLED TO GCMS

ADVANCES IN MICRO GAS CHROMATOGRAPHY

ADVANCES IN ON-LINE HYDROGEN MEASUREMENT

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Application Note. Author. Abstract. Pharmaceuticals. Detlef Wilhelm ANATOX GmbH & Co. KG. Fuerstenwalde, Germany mau

Determination of Diesel Range Organics (DROs) Using SPE and GC/FID by Method 8015D*

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

Thermal Desorption Technical Support Note 87: A cryogen-free method for monitoring trace greenhouse gases in air. Key Words:

Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1.

Gas Chromatography Assignment Chem 543/443

Two Headspace Solutions for the Analysis of Dissolved Gases in Water by Method RSKSOP-175

MARKET FOCUSED - environmental ORGANOCHLORINE PESTICIDES

METHOD TRANSLATION OF HJ FOR INTUVO

Faster Separations Using Agilent Weak Cation Exchange Columns

Analyze more samples. faster. while maintaining resolution. NEW Agilent J&W High Efficiency Capillary GC Columns

Protocol for Quantitative Determination of Residual Solvents in Cannabis Concentrates Prepared by: Amanda Rigdon, May 23 rd, 2016

The Importance of High Speed Data Acquisition in the Design and Implementation of GC/MS Applications

Using the Pulsed Flame Photometric Detector (PFPD) for Low-level Analysis of Organophosphorus Pesticides

Fast and High-Resolution Reversed-Phase Separation of Synthetic Oligonucleotides

A New Horizon of Performance Agilent G1888 Network Headspace Sampler

Fast Analysis of Environmental Phenols with Agilent Poroshell 120 EC-C18 Columns

High temperature Simdist analysis using the new MXT-1 HT Simdist phase. Jaap de Zeeuw and Gary Stidsen Restek corporation, Bellefonte, PA, USA

Chromatography Applications and Technology

ETHYLENE PLANT DRY CRACKED GAS Product Code 54801

The Use of High Resolution Accurate Mass GC/Q-TOF and Chemometrics in the Identification of Environmental Pollutants in Wastewater Effluents

SCOTTY Brand Transportables Convenient, Cost-Effective Calibration Gases

Reduced carryover and increased reproducibility in the analysis of cold pressed pink grapefruit essential oil

FINAL REPORT. Participants in the reported tasks 1. Ravi Subramanian 2. Manoranjan Misra 3. Kent Hoekman

GC Method Developement. Simon Jones GC Application Engineer GC Columns and Consumables

Analysis of an Acrylic Copolymer using Pyrolysis-GC/MS

Analysis of styrene leached from polystyrene cups using GCMS coupled with Headspace (HS) sampler

MicroSAM / Sitrans CV

Volatile Organic Carbon Contaminated Site Assessment

Multiple Detector Approaches to Protein Aggregation by SEC

A method to pack Agilent Prep C18 Media on Load & Lock Column

GC Inlets An Introduction

An Examination of the Presence, Formation, and Transformation of Volatile Halogenated Organic Species in Wastewater Extracts Using GC-ICP-MS

Super Clean Gas Filters Simply the best

A Direct Column-Performance Comparison for Rapid Contract Laboratory Program (CLP) Pesticide Analysis. Application. Authors. Abstract.

Transcription:

Improved Gas Chromatograph Method for the Analysis of Trace Hydrocarbon Impurities in 1, 3-Butadiene Application Note Hydrocarbon processing industry Author Chunxiao Wang Agilent Technologies (Shanghai) Co Ltd. 412 YingLun Road Waigaoqiao Free Trade Zone Shanghai 200131 P.R. China Abstract An improved gas chromatographic method based on an Agilent 7890 Series GC for the analysis of trace hydrocarbon impurities in monomer grade 1, 3-Butadiene has been developed. A High Pressure Liquid Injector (HPLI) allows sampling light petroleum without discrimination and provides high accuracy and precision. Post-column backflushing eliminates the long bake-off time for late-eluting compounds such as the dimer, 4-vinylcyclehexene. It not only saves the analytical cycle significant time, but also improves the retention time (RT) repeatability. The performances of different types of PLOT Alumina columns are also discussed. Introduction 1,3-Butadiene is an important petrochemical raw material. It is produced as a byproduct of the steam cracking process for the production of ethylene and other olefins. Most 1,3-Butadiene is used as a monomer in the production of synthetic rubber. The presence of trace hydrocarbon impurities especially propadiene and alkynes including (also known as m) can interfere with catalytic polymerization. Therefore, identifying and quantifying trace hydrocarbon impurities in1,3-butadiene is critical to producing high quality synthetic rubber products. ASTM standard method D 2593 uses packed columns for determination of hydrocarbon impurities in butadiene [1]. However, some C4 isomers cannot be separated by this method.

Alumina PLOT columns are commonly used for light hydrocarbon separation. The determination of polar hydrocarbon impurities at trace levels can be challenging and is highly dependent on the deactivation of the alumina surface. Alumina with KCL deactivation modifies the Al 2 surface making it less polar. This PLOT column is used in China s national standard method GBT-6017-2008 [2] and provides highly selective retention for both saturated and unsaturated light hydrocarbons, however, measurement of trace level will be impacted by overloading of 1,3-Butadiene. Other challenges include a long bake-out time required for the high molecular weight impurities, primarily dimer (4-vinylcyclohexene), to elute. Also, RT repeatability is affected by sample residues. In this application note, nonpolar Al 2 surface HP-AL/KCL and more polar surface GS-Alumina columns were tested. Backflushing was used to remove the high molecular weight impurities including dimer to the detector so that the analytical cycle time was shortened and RT repeatability was improved. The dimer, which normally requires a second sample injection on an alternative column [3], can be possibly detected. Experimental Two configurations based on a 7890 Series GC are used in this study. Both configurations include an HPLI, a split/splitless inlet, and an FID. Configuration A did not use backflushing. Configuration B used a six-port valve with both ends of a PLOT GS-Alumina column connected (particle traps are recommended for both column sides). By switching the valve, the column flow can be reversed for backflushing the high molecular weight impurities to the detector. Figures 2 and 6 show the valve drawing for configuration A and B respectively. Table 1 shows the typical GC conditions for impurities in 1,3-Butadiene analysis based on configuration B. Table 2 lists the gas standard mixtures used for method development. Table1. Gas chromatograph Injection source Typical GC Conditions for Impurities in 1,3-Butadiene Analysis Based on Configuration B with Backflushing Agilent 7890A Series GC High Pressure Liquid Injector (HPLI) Injection port Split/splitless, 120 C, split ratio: 20:1 (uses higher or lower split ratio according to the concentrations of hydrocarbons Sample size 0.5 µl (liquid) EPC flow for HPLI 30 ml/min Valve temperature 120 C Oven temperature program 60 C hold 3 minutes, to 5 C hold 3 minutes at 5 C/min, to 150 C hold 3 minutes at 5 C/min, to180 C hold 3 minutes at 30 C/min Column GS-Alumina 50 m 0.53 mm,.00 µm Column flow (He) 7.5 ml/min, constant flow mode FID Temperature 200 C H 2 flow 40 ml/min Air flow 400 ml/min Make up (N 2 ) 40 ml/min Table 2. Compound Propane Propylene iso-butane Butane Propadiene t-2-butene 1-Butene iso-butylene c-2-butene iso-pentane n-pentane 1,2-Butadiene 1,3-Butadiene Propyne Vinyl acetylene Ethyl acetylene Standard Sample Information Standard No. 1 Standard No. 2 Amt [% w/w] 1.8500e-2 1.1230e-1 7.2690e-2 6.2780e-2 8.52e-2 9.7800e-2 1.04e-1 1.2890e-1 3.2740e-2 3.0320e-3 5.01e-3 1.80e-1 Balance 1.5150e-2 1.8590e-2 1.0430e-2 Compound [ppm w/w] t-2-butene 120.000 c-2-butene 3400.000 n-pentane 50.000 1,2-Butadiene 370.000 1,3-Butadiene Balance Propyne 20.130 Vinyl acetylene 30.200 Ethyl acetylene 59.320 2

Results and Discussion Sample injection Gas chromatographic sampling and representative analysis of highly volatile liquefied hydrocarbons with high precision and accuracy can be challenging. A unique sample injection device, HPLI, based on a unique needle interface and high pressure liquid rotary valve has been designed for sampling light petroleum matrices with broad boiling point distributions without the danger of carryover. Various pressurized liquid samples have been tested on this device with high accuracy and precision in a previous application note [4]. The HPLI was selected as the best sample introduction device for this work. Column selection Several types of Alumina PLOT columns with the deactivation of the Al 2 surface are available. Alumina with KCL deactivation gives a relatively nonpolar Al 2 surface. Alumina bond Na 2 SO 4 produces a midpolar Al 2 surface and GS-Alumina is the most polar Al 2 surface with proprietary deactivation. For impurity analysis, the affect of overloading must first be determined. Normally, overloading phenomena will produce asymmetric peak shapes with a significant tail. Alumina PLOT columns based on gas-solid chromatography show peak shape of an overloaded component that will sharply raise then start to tail. In this case, with higher concentrations of 1,3-Butadiene, the peak elutes earlier and the peak tail remains the same as shown in Figure 1. Therefore, when injecting a large amount of 1,3-Butadiene, its peak may overlap the component which elutes just in front. 8000 6000 4000 2000 0 Figure 1. With higher concentrations of 1,3-Butadiene Peak elutes earlier Split ratio 2:1 2.5 5 7.5 12.5 15 17.5 20 22.5min Impact of overloading on peak broadening on Alumina PLOT column. To identify a suitable column for the impurities analysis, HP-AL/KCL(Alumina with KCL deactivation) and GS-Alumina columns were tested. Configuration A without backflushing was used for the experiments. 5:1 :1 20:1 30:1 50:1 1,3-Butadiene Peak end remains the same Sample out Restrictor Valve W S Filter Sample in C P EPC flow (PCM B-1) FID Carrier gas Needle Split vent Figure 2. Configuration A without backflush. 3

On the HP-AL/KCL column, elutes just before 1,3-Butadiene. A 0.5-µL solution of standard No.1 is injected by the HPLI at a split ratio of 200:1. Figure 3 shows that 150 ppm of is masked partly by the main peak of 1,3-Butadiene. Reducing the sample size to 0.06 µl with split ratio :1 allows to be separated from 1,3 butadiene as shown in Figure 4. In this case, 20 ppm of was detected with a signal-to-noise ratio of 2.9. 40 35 Propylene iso-butylene 1,3-Butadiene 30 25 20 15 Propane iso-butene Butane Propadiene 1-Butene t-2-butene c-2-butene 1,2-Butadiene 150 ppm of masked partly by main peak of 1,3-Butadiene VA iso-pentane n-pentane Propyne EA 2.5 5 7.5 12.5 15 17.5 20 min GC conditions Column HP-AL/KCL, 50 m, 0.53 mm, 15 µm Carrier gas He: 4 ml/min Sample size 0.5 µl (liquid), EPC flow for HPLI: 15 ml/min Inlet spit ratio 200:1 Oven 80 C (hold minutes), to 0 C at 5 C/min, to 1 C at 3 C/min, to 180 C (hold 30 minutes) at 15 C/min Figure 3. Chromatograms on HP-AL/KCL column with the sample size of 0.5 µl. 4

14 13 12 11 9 8 7 6 Standard No.1 propane propylene propadiene butane 1-butene iso-butylene iso-pentane pentane 1,3-butadene Propyne: 150 ppm w/w% S/N: 65 11 9 8 7 6 Standard No.2 5 15 20 min pentane 1,3-butadene Propyne: 20 ppm w/w% S/N: 2.9 5 15 20 min GC conditions Column HP-AL/KCL, 50 m, 0.53 mm, 15 µm Carrier gas He: 5.6 ml/min Sample size 0.06 µl (liquid), EPC flow for HPLI: ml/min Inlet split ratio :1 Oven program 60 C (hold 3 minutes) to 5 C (hold 4 minutes) at 5 C/min to 150 C at 5 C/min, to 190 C (hold 30 minutes) at 20 C/min Figure 4. Chromatograms on HP-AL/KCL column with sample size of 0.06 µl. 5

On the GS-Alumina column elutes after 1,3-Butadiene. The large amount of 1,3-Butadiene has less impact on determination. As shown in Figure 5, a larger sample size of 0.5 µl with a split ratio of 20:1 for standard No.1 and :1 for standard No. 2 are used. 20 ppm of can is separated from 1,3-Butadiene with a S/N of 95. In this case, levels lower than 20 ppm of can be analyzed. Other hydrocarbon impurities including propadiene and 1,2-Butadiene which just elutes before 1,3 Butadiene are also well separated. 140 120 propylene 1,3-butadiene Standard No.1 Inlet split ratio: 20:1 0 80 60 isobutane butane propadiene 1-butene isobutylene Propyne: 150 ppm S/N: 356 40 20 isopentane n-pentane 5 15 20 25 min 14 1,3-butadiene Standard No.2 Inlet split ratio: :1 12 n-pentane Propyne: 20 ppm S/N: 95 8 6 GC condition 5 15 20 25 min Column GS-Alumina 50 m, 0.53 mm,.00 µm Carrier gas Sample size Oven program He: 6.89 ml/min 0.5 µl (liquid), EPC flow for HPLI: 30 ml/min 60 C (hold 3 minutes) to 5 C (hold 3 minutes) at 5 C/min to 150 C (hold 2 minutes) at 5 C/min, to 180 C (hold 35 minutes) at 30 C/min Figure 5. Chromatograms on GS-Alumina column with sample size of 0.5 µl. 6

Post-column backflushing 1,3-butadiene is a reactive molecule which is typically stored with an inhibitor to prevent polymerization during storage. However, even in the presence of an inhibitor, small amounts of 1,3-Butadiene dimer (4-vinylcyclohexene) will be formed over the long term. Typically, alumina PLOT columns cannot be used to analyze the dimer and it requires over 30 minutes at high temperature to bake dimer and other heavier impurities out from the column. Otherwise, the residuals will affect RT repeatability in subsequent runs. Sample out Restrictor Valve W S C P Filter EPC flow (PCM B-1) Sample in FID Carrier gas Needle Split vent 5 4 3 Switching the valve to reverse the column flow to backflush the heavier impurities to the detector 6 1 2 Figure 6. Configuration B - with backflushing. 7

The Configuration B uses a six-port valve with both ends of a PLOT GS-Alumina column connected. With this design, the column flow can be reversed by switching the valve after the last component of analytical interest elutes. In this way, the heavier impurities including dimer are back-flushed to the detector. The analytical cycle time is shortened significantly as shown in lower chromatogram in Figure 7. Also, this makes it possible to analyze dimer. Without backflushing, it requires an additional 31 minutes at 180 ºC to allow dimer to elute from the column as shown in upper chromatogram in Figure 7. Since all the residuals are backflushed, the RT repeatability is improved. Table 3 demonstrates excellent repeatability. 250 200 150 0 50 propane propylene butane propadiene 1-butene iso-butylene iso-pentane n-pentane Without backflushing, additional 31 minutes at 180 C is required to allow dimer (4-vinylcyclohexene-1) to elute from column 20 30 40 50 60 min 250 dimer 200 150 0 50 propane propylene butane propadiene 1-butene iso-butylene iso-pentane n-pentane Valve switching for backflush After the last peak of interest elutes, switch the valve to reverse column flow, dimer is backflush to the detector GC conditions are listed in table 1 Sample Standard no.1 20 30 40 50 60 min Figure 7. Chromatograms on GS-Alumina column with and without backflushing. Table 3. Repeatability with Backflushing Area n=19 RT n=19 Compounds Average STDEV RSD% Average STDEV RSD% t-2-butene 29.178 0.115 0.39 13.816 0.007 0.05 c-2-butene 744.490 1.991 0.27 15.707 0.008 0.05 n-pentane 11.029 0.034 0.31 16.561 0.008 0.05 1,2-Butadiene 76.324 0.232 0.30 17.408 0.007 0.04 1,3-Butadiene 211309.669 563.114 0.27 17.672 0.009 0.05 Propyne 4.346 0.033 0.77 20.597 0.006 0.03 VA 5.865 0.028 0.47 24.255 0.006 0.03 EA 13.501 0.045 0.33 26.878 0.009 0.03 Dimer 13772.325 33.166 0.24 29.771 0.018 0.06 Sample: Standard No.1 8

Mixed C4 from a petrochemical producer is also tested. The dimer is detected as a symmetric peak with other light hydrocarbon impurities as shown in figure 8. In the experiment, the sample cylinder is pressured by 0 psig of nitrogen. Helium could also be used. 120 0 80 1,3-Butadiene concentration: about 40% 60 40 No backflush 20 Dimer 20 30 40 50 60 min 160 140 120 0 80 60 40 20 propadiene butane 1-2-butene 1-butene n-pentane 1,3-butadiene C5= dimer Dimer Backflush 20 30 40 50 60 min Figure 8. Plant sample, mixed C4 stream analysis with and without backflushing. 9

Conclusion An Agilent 7890 Series GC configured with a High Pressure Liquid Injector (HPLI) with column backflushing mode improves the analysis of the trace hydrocarbon Impurities in 1,3-Butadiene. A GS-Alumina column (50 m 0.53 mm,.00 µm) with 0.5 µl (liquid) sample size was identified as an ideal method for trace impurities analysis. The method has the following advantages: Low level hydrocarbon impurities including in monomer grade 1,3-Butadiene can be analyzed The HPLI provides excellent peak area repeatability Post-column backflushing eliminates the long bake off time for late-eluting compounds including dimer and reduces the analytical cycle time significantly Backflushing improves the RT repeatability The late-eluting compounds are eliminated Baselines remain clean with no carryover Peak positions remain stable References 1. ASTM D2593 93 (2009): Standard Test Method for Butadiene Purity and Hydrocarbon Impurities by Gas Chromatography, Annual Book of Standards, Volume 05.01, ASTM, 0 Barr Harbor Drive, West Conshohocken, PA 19428 USA 2. GBT-6017-2008: Butadiene for industrial use- Determination of purity and hydrocarbon impurities Gas chromatographic method 3. D2426 93 (Reapproved 2009): Standard Test Method for Butadiene Dimer and Styrene in Butadiene Concentrates by Gas Chromatography, Annual Book of Standards,Volume 05.01, ASTM, 0 Barr Harbor Drive, West Conshohocken, PA 19428 USA 4. High-Pressure Injection Device for the Agilent 7890A and 6890 Series Gas Chromatographs, Agilent Technologies publication 5989-8037EN, February 25, 2008 For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 2012 Printed in the USA November 29, 2012 5991-1499EN