ISET. Agilent 1290 Infinity Binary LC with ISET, emulation of the Waters Alliance 2695 LC system analyzing aromatic acids. Application Note.

Similar documents
Agilent 1290 Infinity Binary LC System with ISET - Emulation of the Waters Alliance 2695 LC system analyzing b-blockers

Agilent 1290 Infinity Binary LC System with ISET Emulation of a Waters Alliance 2695 LC Applying Concave, Convex, and Linear Gradients

Application Note. Author. Abstract. Pharmaceuticals. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Technical Overview. Author. Abstract. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Comparing gradient transfer of isocratic hold and delay volume addition using the Agilent 1290 Infinity LC with ISET

Application Note. Environmental. Agilent 1290 Infinity Binary LC. with ISET and fine-tuning. Agilent 1290 Infinity Binary LC.

Emulation of the Agilent 1100 Series LC Through Waters Empower Software Analysis of an Analgesic Mixture

Agilent 1290 Infinity Quaternary LC Support of Columns with 2.1 to 4.6 mm ID to 1200 bar

Agilent 1290 Infinity LC with ISET under Waters Empower control Emulation of Agilent 1100 Series LC

Method Transfer from an Agilent 1200 Series LC to an Agilent 1260 Infinity II LC

Fast Method Development Using the Agilent 1290 Infinity Quaternary LC System with Column Selection Valve

Application Note. Author. Abstract. Pharmaceuticals. Detlef Wilhelm ANATOX GmbH & Co. KG. Fuerstenwalde, Germany mau

Method Transfer from an Agilent 1260 Infinity LC to an Agilent 1260 Infinity II LC

Technical Overview. Author. Abstract. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

Analysis of amoxicillin and five impurities on the Agilent 1220 Infinity LC System

Quality-by-Design-Based Method Development Using an Agilent 1290 Infinity II LC

Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC

Application Note. Author. Abstract. Small Molecule Pharmaceuticals. Sonja Krieger Agilent Technologies, Inc. Waldbronn, Germany

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

Automated Scouting of Stationary and Mobile Phases Using the Agilent 1290 Infinity II Method Development Solution

Performance characteristics of the 1260 Infinity Quaternary LC system

Application Note. Author. Abstract. Small Molecule Pharmaceuticals & Generics. A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

Performance Characteristics of the Agilent 1220 Infinity Gradient LC system

Protein Separation with ph Gradients Using Composite Buffer Systems Calculated by the Agilent Buffer Advisor Software

Scale-up with the Agilent SD-1 Purification System analytical and preparative runs on a single system

Analytical Instrument Qualification According to USP <1058>: Requirements and Examples for the Agilent 1290 Infinity LC System

Improved Sensitivity with the Agilent 1290 Infinity Evaporative Light Scattering Detector (ELSD)

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Determination of Asarinin in Xixin (Asari Radix Et Rhizoma)

Technical Overview. Author. Abstract. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Authors. Abstract. Introduction. Pharmaceutical

Automated alternating column regeneration on the Agilent 1290 Infinity LC

High-Throughput LC/MS Purification of Pharmaceutical Impurities

New Advances in UHPLC -Resolution, Speed & Sensitivity

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs

Simplifying Methods Transfer: Novel Tools for Replicating Your Established Methods on an ACQUITY Arc System

Method Translation in Liquid Chromatography

Application of Agilent AdvanceBio Desalting-RP Cartridges for LC/MS Analysis of mabs A One- and Two-dimensional LC/MS Study

Fast Analysis of Environmental Phenols with Agilent Poroshell 120 EC-C18 Columns

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Developing Purification Strategies for the Agilent 1260 Infinity II Preparative LC/MSD System

Quantification of genotoxic "Impurity D" in Atenolol by LC/ESI/MS/MS with Agilent 1200 Series RRLC and 6410B Triple Quadrupole LC/MS

Fast, Low Pressure Analysis of Food and Beverage Additives Using a Superficially Porous Agilent Poroshell 120 EC-C18 Column

Molecular Characterization of Biotherapeutics The Agilent 1260 Infi nity Multi-Detector Bio-SEC Solution with Advanced Light Scattering Detection

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

Analysis of Nucleosides Using an Agilent Infinity II High Speed UHPLC with the 6130 Single Quadrupole Mass Selective Detector

Practical Applications of Method Translation Using the Agilent Method Translation Tool. eseminar and Workshop

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

Analysis of TCM injections using the Agilent 1290 Infinity LC system

Agilent 1260 Infinity II Prime LC ACHIEVE CONVENIENCE EVERY DAY IN EVERY WAY

Automated QbD-Based Method Development and Validation of Oxidative Degraded Atorvastatin

Peptide Mapping: A Quality by Design (QbD) Approach

Agilent 1290 Infinity II LC SET NEW BENCHMARKS IN EFFICIENCY WITH THE NEXT GENERATION OF UHPLC

ultra why micro-lc? product note introducing the ExpressLC -Ultra system: UHPLC with all the advantages of micro-lc

Application Note. Authors. Abstract. Biopharmaceuticals

Application Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany

Agilent 218 Purification System. Purify your way

EPA Method 543: Selected Organic Contaminants by Online SPE LC/MS/MS Using the Agilent Flexible Cube

Qualification of High-Performance Liquid Chromatography Systems

The Agilent 1260 Infinity BioInert Quaternary Pump. Scope of a low-pressure mixing UHPLC pump with Bio-Inert Capabilities

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

Agilent 1290 Infinity II 2D-LC Solution Biopharmaceutical Polymer Analysis. WCBP Jan 2017 Washington, DC

Agilent 1290 Infinity II LC MORE EFFICIENCY MORE FREE TIME. #EfficientUHPLC

Benefits of 2D-LC/MS/MS in Pharmaceutical Bioanalytics

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

Agilent 1260 Infinity LC & LC/MS Purification Systems. Purify your way

Agilent SD-1 Purification System. Purify your way SD-1

Techniques for Improving the Efficiency of Large Volume Loading in Preparative Liquid Chromatography

Agilent Automated Card Extraction Dried Blood Spot LC/MS System

Optimizing Productivity and Reliability for Monocyclic Aromatic Hydrocarbon Purity Analysis According to ASTM D7504 on the Agilent 8890 GC System

A Modular Preparative HPLC System for the Isolation of Puerarin from Kudzu Root Extracts

Agilent OpenLAB CDS MatchCompare for the Comparison of Peptide Mapping Samples

Automated fraction re-analysis does it really make sense? Application. Udo Huber. Abstract

Performance characteristics of the Agilent 1100 Series capillary LC system using diode-array UV and MS for detection. Technical Note.

Analytical HPLC solutions

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Agilent 1220 Infinity II LC GET ON THE FAST TRACK TO HIGHEST EFFICIENCY AT A PRICE YOU CAN AFFORD

ACQUITY UPLC. I-Class PLUS WHAT SEPARATES YOU FROM EVERYONE ELSE

Confident Pesticides Analysis with the Agilent LC/Triple Quadrupole and TOF/QTOF Solutions

Application Note. Author. Abstract. Environmental. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

NISTmAb characterization with a high-performance RP chromatography column

Agilent 1290 Infinity 2D-LC Solution. Selectable separation performance

10. Validated Normal Phase HPLC Method for the Determination. Fulvestrant is primarily used in the treatment of hormone receptor

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

Characterize Fab and Fc Fragments by Cation-Exchange Chromatography

Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants

Characterize mab Charge Variants by Cation-Exchange Chromatography

Minimizing Sample Carryover Using the Multiwash Function of the Agilent 1290 Infinity II Multisampler

Automated Purification Software for Agilent OpenLAB CDS

Agilent 1260 Infinity Purification Systems. Infinitely better purity and recovery

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Ultra High Performance Liquid Chromatograph. Nexera XR C196-E082C

Ultra High Performance Liquid Chromatograph. Nexera XR C196-E082D

A Guide To Speeding Up Your Separation

easy. HPLC has never been so Agilent 1120 Compact LC Our measure is your success.

Technical Overview. Authors. Abstract. Verified for Agilent 1260 Infinity II LC Bio-inert System

Transcription:

Agilent 129 Infinity Binary LC with ISET, emulation of the Waters Alliance 2695 LC system analyzing aromatic s Application Note Food Testing & Agriculture Author A.G.Huesgen Agilent Technologies, Inc. Waldbronn, Germany Absorbance [mau] 2 175 15 125 1 75 5 25 Agilent 129 Infinity Binary LC with ISET Waters Alliance 2695 ISET 3 4 5 6 7 8 9 1 Time [min] Abstract The Intelligent System Emulation Technology (ISET) using the Agilent 129 Infinity Binary LC was introduced to emulate Agilent 12 Infinity Series and 12 Series LCs. Now the emulation of non-agilent LCs has been introduced. This Application Note shows the emulation of the Waters Alliance 2695. Based on a Waters Application Note, the transfer of the analysis of aromatic s onto the 129 Infinity Binary LC using ISET was evaluated. The agreement of retention times and resolution was determined.

Introduction Seamless instrument-to-instrument method transfer is often a strong demand for most industries. Changing established methods is expensive and time consuming and comparison with reference data is difficult, but equipment has to be replaced from time to time. The Agilent 129 Infinity Binary LC, in combination with ISET, offers the possibility to emulate older non-agilent instrumentation, like the Waters Alliance 2695 LC systems. ld methods from non-agilent equipment can be transferred to the 129 Infinity Binary LC with ISET maintaining the same retention times results and, typically, receiving better resolution. Using the 129 Infinity Binary LC, UHPLC methods can be applied for future use. This Application Note analyzes aromatic s on the 129 Infinity Binary LC using a chromatographic method developed on the Waters Alliance 2695 LC system. Seamless method transfer, with excellent correlation of retention times, and typically better resolution, is shown. Experimental The following instruments were used, see Table 1. Acquisition and Evaluation Software penlab CDS ChemStation version C.1.4 and ISET Agilent 129 Infinity Binary LC Waters Alliance Agilent 129 Infinity Binary Pump: G422A 2695 Agilent 129 Infinity Autosampler: G4226A ALS cooler: G133B Column compartment: G1316C Diode array detector: G4212A VWD (Dual absorbance detector) Table 1 Instrumentation used Chromatographic conditions Column: Atlantis T3 4.6 15, 5 µm Mobile phases: A = 1 mm NH 4 C (.771 g/l) ph = 3, B = ACN Gradient: at minutes % B, at 1 minutes 25% B, at 11 minutes % B, at 15 minutes % B, Flow rate: 2 ml/min Injection volume: 1 µl Column temperature: 3 C Detection: 285 nm, 5 Hz Analyzed compounds H 3 C Vanillic Syringic CH 3 CH 3 H 4-Hydroxybenzoic Protocatechuic H Gallic The analyzed aromatic s are plant ingredients used, for example, in traditional plant medicine. Gallic has antiviral and antifugal activities. Protocatechuic has antioxidant and anti-inflammatory activities. 4-Hydroxybenzoic is a metabolite of catechins after intake of green tea. Vanillic is used as flavoring agent. Syringic occurs in special plant oil. 2

In the ISET parameter screen of the 129 Infinity Binary LC, enter the following parameter to select the instrument to be emulated, see Figure 1. Results and Discussion The following experiments were performed to prove the seamless method transfer from the Waters Alliance 2695LC system to the 129 Infinity Binary LC in combination with ISET. Selection of manufacturer Selection of instrument Selection of mobile phases Selection of autosampler configuration Experiments Analysis of organic s on the Alliance 2695, based on a Waters Application Note 1 Transfer of the developed method onto the 129 Infinity Binary LC with and without applying ISET and ISET with fine tuning Evaluated parameters Precision of retention times for both LC systems Determination of the deviation of retention times; specified allowed maximum deviation is <±5% Determination of the resolution, typically better on the 129 Infinity Binary LC; specified allowed maximum deviation is < 5% Figure 1 Selection of the instrumentation to be emulated. 3

Precision of Retention Times Starting at % for the organic channel and at 1% for the aqueous channel are demanding gradient conditions for any pump. The precision, especially for the early eluting peaks, might be significantly lower than for the later eluting peaks. The 129 Infinity Binary LC provided excellent precisions even for these demanding conditions, see Figure 2. The precision for the first peak was.2% RSD using the 129 Infinity Binary LC and.23% RSD for the Alliance 2695 LC. The precision for all other peaks was ~.2% RSD for both instruments. The 129 Infinity Binary LC system performed approximately 1 times better for the first peak. AU Waters Alliance 2695.1 RSD RT =.23%.8.6.4.2 3.5 4 4.5 5 5.5 mau Figure 2 Precision of retention times applying demanding gradient conditions. min 12 1 8 6 4 2 Agilent 129 Infinity Binary LC RSD RT =.3% 3 3.5 4 4.5 5 5.5 min 4

Determination of the Deviation of Retention Times The chromatographic conditions developed on the Alliance 2695 were transferred onto the 129 Infinity Binary LC. The best correlation of retention times with the original data was obtained using the 129 Infinity Binary LC with ISET and fine tuning, see the red trace in Figure 3. Fine tuning means in this case an additional delay volume of 4 µl was added in the ISET parameter screen by enabling the manual fine tuning parameter see Figure 4. Without the fine tuning step, the agreement of retention times using the 129 Infinity Binary LC was already close, see the orange trace in Figure 3. AU 25 2 15 1 5 Gallic Protocatechuic 4-Hydroxybenzoic Vanillic Syringic 3 4 5 6 7 8 9 Agilent 129 Infinity Binary LC without ISET Agilent 129 Infinity Binary LC with ISET Agilent 129 Infinity Binary LC with ISET + fine tuning (+4 µl delay) Waters Alliance 2695 Figure 3 verlay of chromatograms of all used LC instruments and Agilent 129 Infinity Binary LC configurations. min Without enabling the ISET tool, the 129 Infinity Binary LC showed less retention for all peaks as expected, see the blue trace in Figure 3. Adding or subtracting delay volume Figure 4 Fine tuning parameters within ISET. 5

Figure 5 summarizes the retention time deviations obtained on the different 129 Infinity Binary LC configurations. ptimum results were achieved using the ISET tool with fine tuning. The maximum retention time deviation was +.93%. With ISET and no fine tuning, the maximum deviation was 4.4%. This was also within the allowed specified range. Without ISET, the retention time deviation on the 129 Infinity Binary LC was 11.8%. This was mainly due to the significant lower delay volume of the 129 Infinity Binary LC compared to the Alliance LC system. The allowed deviation of retention times for the 129 Infinity Binary LC with ISET is ±5%. Deviation (%) 1 5 _ 5 _ 1 _ 15 Gallic Protocatechuic Deviation of retention times 4-Hydroxybenzoic Agilent 129 Infinity Binary LC w/o ISET Agilent 129 Infinity Binary LC with ISET Agilent 129 Infinity Binary LC with ISET and fine tuning Vanillic Syringic Figure 5 Deviation of retention times as percentage for all Agilent 129 Infinity Binary LC configurations. Maximum allowed deviation <±5% Determination of the Deviation of the Resolution Data Figure 6 summarizes the resolution data for the different 129 Infinity Binary LC configurations. With ISET and ISET and fine tuning, the resolution had improved for all peaks. Without ISET, the resolution for the second peak was decreased significantly. Deviation (%) 2 15 1 5 _ 5 _ 1 _ 15 _ 2 _ 25 Deviation of resolution Maximum allowed deviation ~5% Agilent 129 Infinity Binary LC w/o ISET Agilent 129 Infinity Binary LC with ISET Agilent 129 Infinity Binary LC with ISET and fine tuning _ 3 Protocatechuic 4-Hydroxybenzoic Vanillic Syringic Figure 6 Deviation of resolution data for all Agilent 129 Infinity binary LC configurations. 6

Conclusion The Agilent 129 Infinity Binary LC, in combination with the ISET, enables the emulation of older non-agilent LCs, such as the Waters Alliance LC system, with excellent precision of retention times. The allowed deviation of retention times is <±5%. In this example, the deviation of retention times was <1% using the 129 Infinity Binary LC with ISET and fine tuning. The resolution was improved for all peaks. The allowed maximum deviation for the resolution is < 5% Reference 1. Waters Application note, Analysis of aromatic s using Atlantis T3, 29. 7

www.agilent.com/chem/iset Agilent Technologies, Inc., 213 Published in the USA, April 1, 213 5991-219EN