Clinical Evaluation of a New In-Line Continuous Blood Gas Monitor

Similar documents
Technical Compendium CDI. Blood Parameter Monitoring System 500

Abbott istat for Blood Gas Analysis: The Halifax Perfusion Experience

Platelet Preservation:

Heparin Use in Pediatric Bypass Empirical Regimen (ACT) vs. Heparin Concentration: A Multicenter Trial

Clinical Evaluation of the Hemotec Heparin Dose Response Cartridge

Validation of a quality assessment system for blood gas and electrolyte testing

The Advantages of Automatic Quality Control (AQC) for Blood Gas Testing QC

COATING SYSTEMS BIOLINE COATING

This document is a preview generated by EVS

The Rheological Effects of X-Coating with Albumin and Hetastarch on Blood During Cardiopulmonary Bypass

3/17/2016. Antithrombin and heparin dose effects on measures of anticoagulation in first five days of infant-pediatric ECLS.

The Rheological Effects of X-Coating with Albumin and Hetastarch on Blood during Cardiopulmonary Bypass

Design and implementation of a microchemistry analyzer

The Heater Cooler as a Source of Infection from Nontuberculous Mycobacteria

Affinity NT. Clearly the Choice for Consistent Performance. and Reliable Results

Affinity NT. Clearly the Choice for Consistent Performance. and Reliable Results

Run critical care tests reliably and economically in low-volume settings

INSPIRE Oxygenator Family

A Dedicated Perfusion Electronic Medical Record with Discrete Epic Integration

Proficiency Testing on the i-stat System

autolog Autotransfusion System

9/7/17. Implementing the ARK Immunoassay for Therapeutic Drug Monitoring of High-Dose Methotrexate. Disclosure. Learning Objectives

Closing the Gap in Pre-analytical Specimen Quality

Lung-gas composition and transfer analysis: O 2 and CO 2 diffusion coefficients and metabolic rates

AUTOLOG. Autotransfusion System. A simple, yet sophisticated system.

Successful performance in proficiency testing (PT) is required

PROTOCOL DRAFTING GUIDE

510(k) SUMMARY. Masimo Corporation 40 Parker Irvine, CA FAX

TECHNICAL BULLETIN. ACT Test Result Calibration Options: PREWARMED vs. NON-PREWARMED Result Calibration BACKGROUND

Radiology Staffing Survey 2010

i-stat iccnet CHSA Edition 1.2 Page 1 of 7 i-stat Rosy Tirimacco / Paul Simpson

Human Research Protection Program Guidance for Human Research Determination

DEPARTMENT OF HEALTH & HUMAN SERVICES

Plan Subject Index Number Section Subsection Category Contact Last Revised References Applicable To Detail PRINCIPLE:

Evidence-Based Algorithm for Heparin Dosing Before Cardiopulmonary Bypass. Part 1: Development of the Algorithm

Statement of the. Society for Cardiovascular Angiography & Interventions. Presented by. Augusto D. Pichard, MD, FSCAI. Before the

How Advances in Molecular Diagnostics Will Further Drive Consolidation of All Lab Testing. Executive War College 2013 New Orleans, LA

Interventional radiology suite or hybrid operating room: Which is the best for lung nodule localization?

Department of Laboratories St. Louis, MO ISSUE DATE: February 2002 REVISION DATE: August 2011 REVIEWED DATE: May 2017 PRINCIPLE:

Comparison of istat and EPOC Blood Analyzers

Total Analytic Error From Concept to Application

Tough Love: Managing your lab customers to improve relationships and outcomes. Brian Jackson, MD, MS

What makes the technology unique?

Get a Closer Look. Continuing Support

Quality Management of the Total Examination Process. Elvar Theodorsson

Menu and Throughput to Accommodate Your High-volume Testing Needs

MITRAL VALVE REPAIR BACKGROUND

INFORS HT Product Guide: Visiferm DO Optical Oxygen Sensor

A CRITICAL REVIEW ON APPLICATION OF BIOMEDICAL SIGNAL PROCESSING IN MEDICINE

WEQAS: what is it and what does it mean?. Celia Critchley Point of Care Co-ordinator Warrington and Halton Hospitals NHS Foundation Trust

S1725 Method Validation and Verification: Case Studies and Laboratory Challenges

APB PRINCIPLE SPECIMEN. REF (300 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

The Journal of Thoracic and Cardiovascular Surgery

The portable laboratory: an evaluation of the accuracy and reproducibility of i-stat

TERRA NOVA Wind + Solar Energy Premium Kit

Proxima. pco 2 and Calculated values for HCO 3 - and Base Excess on Proxima

The future in hemodynamic monitoring

EFPIA-PHRMA PRINCIPLES FOR RESPONSIBLE CLINICAL TRIAL DATA SHARING REPORT ON THE 2016 MEMBER COMPANY SURVEY

DETERMINATION OF OPERATING ROOM REQUIREMENTS USING SIMULATION. Julie C. Lowery Jennifer A. Davis

Harmonization of red blood cell distribution width (RDW): an attainable target?

Guidance for Industry and FDA Staff. Replacement Reagent and Instrument Family Policy

Goals. Create a tool to manage a ubiquitous, complex clinical process. Deploy the tool in the ICUs and ED. Evaluate changes in clinical practice

The evolution of TQIP Best Practices for Massive Transfusion

Implementation of massive transfusion protocols to guide practice in severe bleeding- is your institution ready?

Enrollment Challenges in Cardiothoracic Surgical Clinical Trials

Effect of Autotransfusion on Fibrinolysis in Open Heart Patients

ABB Measurement & Analytics Monitoring cement plant stack emissions using FTIR

Establishing Quality Control Target Values and Standard Deviations for Hematology Instrumentation

Type of intervention Primary prevention. Economic study type Cost-effectiveness analysis.

Low-dose and High-resolution Cardiovascular Imaging with Revolution* CT

Incorporating FDA Regulatory Thinking into Plans for Medical Devices

Is the learning curve for video-assisted thoracoscopic lobectomy affected by prior experience in open lobectomy?

Clinical trial design issues and options for study of rare diseases

DOCUMENT DESPATCH ADVICE DRAFT IN WIDE CIRCULATION MHR 06 /T- 50/51/52/53/ TECHNICAL COMMITTEE: MHR 06

The Effect of Hypothermia on Heparin Anticoagulation as Measured by Activated Clotting Time and Factor Xa Inhibition

DIAGNOSTICS ASSESSMENT PROGRAMME

Point-of-care tests (POCT) and Laboratory Information Systems (LIS)

Point of Care Testing in Clinical Trials

Meets Standards AFTER REVIEW ( 2.5 pts) Fail to meet the standard safety precautions, counseled one time

ITC Hemochron Signature Elite

Benjamin Wei, M.D., Assistant Professor of Cardiothoracic Surgery. The Journal of Thoracic and Cardiovascular Surgery

Proxima. pco 2 and Calculated values for HCO 3 - and Base Excess on Proxima

Modified Microsample ACT Test for Heparin Monitoring

TOP 10 INVESTIGATOR RESPONSIBILITIES WHEN CONDUCTING HUMAN SUBJECTS RESEARCH

Improving Outcomes of Cardiac Surgery with a Point-of-Care Based Transfusion Algorithm

AAG PRINCIPLE SPECIMEN. REF (150 tests) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

HST-330 TM and Total Laboratory Automation

Tailoring best practices for the evaluation of endpoints in clinical device trials

Me- of Monitoring Heparin Therapy for

Indian Journal of Medical Research and Pharmaceutical Sciences July 2016; 3(7) ISSN: ISSN: DOI: /zenodo Impact Factor: 3.

Teleflex Incorporated (NYSE: TFX) Investor Presentation

HEMOCHRON. Whole Blood Coagulation Systems

Variability of Laboratory Test Results

Case Studies: Sharing Lessons Learned

Laboratory Developed Tests. William Castellani, MD Inter-regional Commissioner Clinical Pathologist, Penn State Hershey Medical Center

Biomedical Optics Instrumentation

pump multi-indication pump

The Journey Towards Standardization in Validation of Fluorescent Cell-Based Assays

MIS Pancreatic Surgery The Journey. Michael L. Kendrick, MD Mayo Clinic, Rochester, Minnesota

Transcription:

THE JOURNAL OF EXTRA-CORPOREAL TECHNOLOGY Original Article Clinical Evaluation of a New In-Line Continuous Blood Gas Monitor Rhonda Southworth, MHA, CCP*; Robin Sutton, MS, CCP*; Sam Mize, CCP*; Alfred H. Stammers, MSA, CCP'\ Lance W. Fristoe, CCPA; David Cook, MOt; Dan Hostetler, CCPt; Wayne E. Richenbacher, MD** *Medical University of South Carolina, Charleston, South Carolina; AUniversity of Nebraska Medical Center, Omaha, Nebraska; tmayo Clinic, Rochester, Minnesota; and **University of Iowa Hospitals and Clinics, Des Moines, Iowa Keywords: cardiopulmonary bypass, in-line blood gas monitor, device evaluation ABSTRACT Two methodologies for obtaining accurate blood gas and electrolyte values during cardiopulmonary bypass (CPB) are traditional laboratory analyzers, which use an electrochemical technology, and continuous in-line monitoring systems, which use a fluorometric and/or spectrophotometric technology. The purpose of the present study was to evaluate the accuracy of a new continuous in-line monitor, the 3M CDI Blood Parameter Monitoring System 500, which provides continuous in-line measurements of ph,, potassium (K+), oxygen saturation, hematocrit, hemoglobin, and temperature, during partial or complete CPB. Study parameters included arterial ph,, and K+ values. Overall performance was analyzed by calculating the mean difference (expressed as the bias) between the CDI system 500 and the laboratory analyzer for each parameter. The accuracy of the arterial ph,, and K+ values provided by the CDI system 500 was then evaluated using target values established in the acceptable performance standards for laboratory analyzers from the Clinical Laboratory Improvement Act of 1988 (CLIA '88). The accuracy of the value provided by the CDI system 500 was evaluated using a target value of± 10% of the reference, or laboratory analyzer, value. A prospective multi-center trial was conducted following Institutional Review Board approval. A total of 75 cases was included in the analyses, with over 200 data points from 4 clinicallocations. Results for ph,, and K+ were within the target values established by CLIA '88. ph bias was 0.00 ± 0.02 ph units. bias was -0.3 ± 3.3 mm Hg. K+ bias was approximately+ 0.12 ± 0.31 mmole/l. Results for were within 10% of the reference value. bias was 7.5 ±13.8 mm Hg. The results of this clinical trial show that the CDI System 500 continuous in-line monitoring system provides values that meet the accuracy standards for laboratory analyzers for arterial ph,, and K+. Address correspondence to: AI Stammers, MSA, CCP Division of Clinical Perfusion Education University of Nebraska Medical Center 600 South 42nd Street, Box 985155 Omaha, NE 68198-5155 166

INTRODUCTION Obtaining accurate blood gas and electrolyte values has always been critically important to the clinician during cardiopulmonary bypass (CPB), thereby prompting the investigation of different methodologies to obtain them. The traditional laboratory analyzer, using an electrochemical technology to provide intermittent samples and/or on demand, has long been considered a "gold standard" for accuracy. Continuous in-line monitoring systems, by contrast, are based on a fluorometric and/or spectrophotometric technology that continuously monitor and display results. One reservation about the acceptance of continuous in-line monitors is the perception that they do not meet the same accuracy standards for measuring blood gas and electrolyte parameters as laboratory analyzers. Because continuous in-line monitors use a different technology from that of conventional laboratory analyzers, comparing the accuracy of the two methods is more complicated than if both used the same technology. In both systems, however, the goal is the same: to provide clinically useful values for the blood gas parameters being measured to ensure the best available patient care will be delivered in a costeffective way (1,2). The purpose of the present study was to evaluate the accuracy of a new continuous in-line monitor, the 3M CDI Blood Parameter Monitoring System 500", by comparing it with the laboratory analyzers used at 4 clinical locations. ries 1620c, and Chiron 865ct. The CDI system 500 is an AC-powered, microprocessorbased monitor which uses optical fluorescence technology with m~crosensors to measure ph, PC0, 2 P0, 2 and K+. The m1crosensors are contained in a disposable shunt sensor that can be placed in any shunt or purge line with continuous flow (Figure 1). A minimum blood flow of 35 ml/min is recommended for optimal performance of the shunt sensor. Blood gas parameters are displayed at either actual temperature or adjusted to 37 C. MATERIALS AND METHODS CLINICAL SITES Four clinical sites participated in a prospective 2 month clinical trial using the CDI System 500 to monitor blood parameter values during CPB procedures. The ph, PC0, 2 P0, 2 and K+ values for the CDI System 500 were compared with the values recorded by the standard laboratory analyzer used at each site. The 4 sites were St. Mary's Hospital, Mayo Clinic, Rochester, MN; Medical University of South Carolina, Charleston, SC; University of Nebraska Medical Center, Omaha, NE; and the University of Iowa Hospital and Clinics, Iowa City, la. MONITOR AND ANALYZERS The following analyzers were used at the 4 sites: Nova Stat Profile 5b, Instrumentation Laborato- Figure 1. CDI system 500 circuit diagram STUDY PROTOCOL Following Institutional Review Board approval, and obtainment of informed patient consent where necessary, a prospective study was performed. Adult patients (18 years or older) undergoing cardiac surgery with CPB expected to last 1 hour or more were included. A minimum of 3 blood samples were collected with each patient serving as his or her own control. The microsensors were calibrated according to the manufacturer's instructions. The ph, PC0, 2 and PO microsen- 2 sors were automatically calibrated before each case using tonometered 2-point gases. The K+ microsensor was calibrated using the patient's blood K+ value as measured by the laboratory analyzer. This first K + measurement was not included as a data point in the study results. When a blood sample was withdrawn, the blood parameter values displayed by the CDI System 500 were recorded. These recorded values were then coma 3M Health Care Cardiovascular Systems. Ann Arbor, Michigan 48103 b Nova Biomedical Corporation, Boston Massachusetts 02154 c Instrumentation Laboratories, Lexington, Massachusetts 02173 d Chiron Diagnostics, Medfield, Massachusetts 02052 167

pared with the laboratory analyzer values. tablished in CLIA '88 (Table 2). The CLIA standards represent a commonly recognized method to monitor the proficiency of laboratory analyzers (3). To date, no acceptable standard has been published for comparing the variance between a laboratory analyzer and a con- STATISTICS The overall performance of the CDI System 500 was analyzed by calculating the mean difference between the CDI System 500 values and the laboratory analyzer values. This difference is expressed as the bias. All data is expressed, unless otherwise specified, as the mean± the standard deviation of the mean. Clinically useful target values for Mean Difference (Bias) the parameters were established using Standard Deviation the acceptable performance standards Number of Data Points from the Clinical Laboratory Improvement Act of 1988 (CLIA '88). Table 1: Statistical results of blood gas values for the CDI System 500 compared with laboratory analyzers ph (units) 0.00 0.02 263 (mm Hg) (mm Hg) K+ (mmole/1) -0.3 7.5 0.12 3.3 13.8 0.31 263 262 190 RESULTS A total of 75 patients from the 4 clinical sites was included in the analyses. Data from all 4 sites, which included over 200 data points, were analyzed together. Table 1 shows a statistical summary of the data: the mean differences and standard deviations for ph,, and K+ of the CDI System 500 compared with the laboratory analyzers. Figures 2 through 5 display the results of the CDI System 500 blood parameter values compared with the lab analyzers for all 4 clinical sites. The CDI System 500 value for ph showed no bias when compared with the laboratory analyzer. The standard deviation was 0.02. The bias for was -0.3 mm Hg, with a standard deviation of 3.3. The bias for K+ was approximately + 0.12 mmole/1, with a standard deviation of 0.31. The bias for was 7.5 mm Hg, with a standard deviation of 13.8. DISCUSSION The results of this clinical trial show that the CDI System 500 continuous in-line monitoring system provides values that meet the accuracy standards for laboratory analyzers for arterial ph,, and K+. The bubbleplots for ph,, and K+ (Figures 2-4) display clinically useful target values, based on standards es- Figure 2: ph sensor bias chart ph Saneor Bias Chart Each point represents the difference between the CDI system 500 value and the laboratory analyzer value (y-axis) referenced against the laboratory analyzer value (x-axis). Upper and lower limits of CLIA mean target values are represented by the pair of parallel lines across the data field. Figure 3: C0 2 sensor bias chart C02 Seneor Bias Chart 12 +~~~~-.. '!'~~~~""~~'-~~ ~~-~ -~- ~ ~-~- --~~~-- ~~~... -~.- ----~-- - --- -12-15 + 10 1s ~ ~ ~ ~ ~ a M ~ ~ C02 Analyzer Value rrmhg value (y-axis) referenced against the laboratory analyzer value (x-axis). Upper and lower limits of CLIA mean target values are represented by the pair of parallel lines across the data field. 168

tinuous in-line monitor. The CLIA '88 standards provide an industry-acceptable method to evaluate the variance between two laboratory analyzers. Therefore, we chose these standards as clinically useful target values with which to evaluate the CDI System 500 performance for ph, PC0, 2 and K+. For ph, PC0, 2 and K+, the variance between the CDI System 500 values and the laboratory analyzer values was within the target values established by CLIA '88. For values, the CLIA target value (see Table 2) could not be used because it represents a variable range (± 3 standard deviations) as opposed to a fixed interval (eg. ± 5 mm Hg). We therefore used ± 10% of the reference, or laboratory analyzer, value to evaluate the performance of the sensor. This value is based on information from the study by Mark et al, who concluded that a bias as high as 15% above arterial P02 values is of minor clinical importance (4). A bias of 7.5 mm Hg at typical arterial P02 values is less than 10% of the reference value. The overestimation in the results is apparent. In earlier studies, it has been shown that at higher values, the laboratory analyzer tends to underestimate values (4,5). Therefore, in our study, the overestimation of results may have less to do with the accuracy of the CDI System 500 than with the methodology used by the laboratory analyzer. In conclusion, the results of this study provide evidence that values obtained by continuous in-line monitoring of the CDI System 500 are comparable in accuracy with those obtained by traditional laboratory analyzers. This study may therefore provide clinically useful comparisons of the two methodologies for clinicians evaluating the utility of continuous in-line monitoring during CPB. ACKNOWLEDGEMENTS Figure 4: K+ sensor bias chart.11 -... 7...,..,,.......\)011115, ~ 20.poklta....,...,... value (y-axis) referenced against the laboratory analyzer value (x-axis). Upper and lower limits of CLIA mean target values are represented by the pair of parallel lines across the data field. Figure 5: 0 2 sensor bias chart 02 Sensor Bias Chart... value (y-axis) referenced against the laboratory analyzer value (x-axis). The range of target values, defined by 10% of the laboratory analyzer value, is represented by the diverging lines across the data field; as P02 increases, the acceptable variance from the laboratory analyzer value increases. Table 2: Comparison between CLIA '88 standards for laboratory analyzers and CDI System 500 results Blood Gas Parameters CLIA '88 Mean Target Values ±.04 ph units ±5 mmhg ± 3 standard deviations ± 0.5 mmole/l CDI System 500 Bias 0.00 ph units -0.3 mmhg NN 0.12 mmole/l This study was funded in part by 3M Health Care Cardiovascular Systems, Ann 'See Table 1. bnot applicable. See text for explanation of the comparable target value used for P0 2 169

Arbor, Michigan. Financial support consisted of supplying 3M CDI Blood Parameter Monitoring disposables for study purposes, funding for laboratory work, and in some cases, support for educational programs. REFERENCES 1. Rubsamen DS. Continuous blood gas monitoring during cardiopulmonary bypass-how soon will it be the standard of care? J Cardiothorac Vase Anes 1990;4: 1-4. 2. Stammers AH. Monitoring controversies during cardiopulmonary bypass: how far have we come? Perfusion 1998; 13:35-43. 3. Code of Federal Regulations Title 42, Volume 3, Subpart I, Section 493.931 "Routine Chemistry." 4. Mark JB, Fitzgerald D Fenton T, et al. Continuous arterial and venous blood gas monitoring during cardiopulmonary bypass. J Thorac Cardiovasc Surg 1991;102:431-39. 5. Feenstra L, Dexer JR, Hodgkin JE. Methodology of arterial blood gas analysis. In: Burton GG, Hodgkin JE, eds. Respiratory Care: A Guide to Clinical Practice. 2nd ed. Philadelphia: JB Lippincott, 1984: 978-91. 170