The injection of the 28 and 29 PCR cycle data was performed at 3kV for 10 and 5 seconds.

Similar documents
GlobalFiler TM Extra Cycle Evaluation Report

Artifacts Identified Post-Developmental Validation: AmpFLSTR Identifiler Plus PCR Amplification Kit

HID Validation: The 3500 Series Genetic Analyzers

Development and Validation of the 3500 Series Genetic Analyzer for Human Identification

Upgrading the Workflow: Enhancing Information Recovery from Forensic Evidence. The world leader in serving science

Direct Amplification of Reference Samples Using the GlobalFiler PCR Amplification Kit

Biology/DNA DNA Interpretation SOP. Biology/DNA Division

Technical Focus. Direct PCR Amplification of Buccal Samples on the NUCLEIC-CARD COLOR With the AmpFlSTR Identifiler Direct PCR Amplification Kit

Forensic Applications of CE and MPS

GeneMarker HID Software Review

Direct PCR Amplification of Blood Samples on Copan NUCLEIC-CARD TM Systems Using the AmpFlSTR Identifiler Direct PCR Amplification Kit

Investigator Quantiplex HYres Kit

First Look: Applied Biosystems NGM Detect PCR Amplification Kit

User Bulletin. Veriti 96-Well Thermal Cycler AmpFlSTR Kit Validation. Overview

GenPlex HID Training Class I

Procedure for Casework DNA Interpretation

Data Quality Worth Sharing

Evaluation of NGS technology for identification, ancestry prediction and phenotypic characters prediction with the Ion PGM System

Maximizing the Performance of Capillary Electrophoresis Systems The Maximizing Data Quality Series Part 4

PowerPlex Fusion 6C System

PowerPlex. Y System Validation

AmpFlSTR Identifiler PCR Amplification Kit

Get more information from challenging samples with next-generation sequencing of short tandem repeats

Generating Forensic DNA Profiles

Developmental validation of GlobalFiler PCR amplification kit: a 6-dye multiplex assay designed for amplification of casework samples

Design and Execution of Validation Studies for Establishing DNA Mixture Interpretation Procedures

TECHNICAL MANUAL. GenePrint 24 System. Instructions for Use of Product B1870 6/16 TM465

What is DNA? Deoxyribonucleic Acid The inherited genetic material that makes us what we are

Authenticating human cell lines using the Identifiler kits and capillary electrophoresis platforms

Procedure for GeneMapper ID-X for Casework

Human Identification Solutions Conference, Mar , Madrid Quantifiler Trio kit and forensic samples management: a matter of degradation

Procedure for GeneMapper ID-X for Casework

Comparison of the Promega Power Y and AB AmpFlSTR YFiler kits and the Subsequent Validation of YFiler.

Product Catalog 2012 AmplGen Ltd.

QIAGEN Validation Report

AmpF STR NGM PCR Amplification Kit - Overview

Developmental Validation of Amplicon Rx, a post-pcr Purification System

Applied Biosystems 3130/3130xl Genetic Analyzers Using Data Collection Software v3.0

Internal Validation of the Promega PowerPlex Fusion System with the Applied Biosystems 3130xl Genetic Analyzer

STR DNA Data Analysis. Genetic Analysis

STRmix Implementation and Internal Validation Erie County Central Police Services Forensic Laboratory

PowerPlex S5 System INSTRUCTIONS FOR USE OF PRODUCTS DC6951 AND DC6950.

Evaluating Noise and the Implications of Methodology in the Analytical Threshold Designation for Forensic Genetic Analysis

Investigator HDplex Handbook

Use of the MSI Analysis System with the Applied Biosystems 3500 Genetic Analyzer and POP-7 Polymer

PowerPlex CS7 System

Uniparental disomy (UPD) analysis of chromosome 15

Biotype Template Files for GeneMapper ID Software

GlobalFiler Kit Casework Test Site Results and Other Development Updates

STR amplification of DNA from buccal and blood samples stored long-term on Whatman FTA cards

Mentype Pentaplex ESS PCR Amplification Kit

Virginia and the New SWGDAM Guidelines

STR amplification of DNA from buccal and blood samples stored long-term on Whatman FTA cards

QIAGEN Validation Report

AmpFlSTR NGM SElect PCR Amplification Kit. User Guide

Mentype Nonaplex I PCR Amplification Kit

STRmix Implementation and Internal Validation-FUSION. Erie County Central Police Services Forensic Laboratory

!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"!"

Mentype Nonaplex I PCR Amplification Kit

Y chromosomal STRs in forensics

EVALUATION OF THEPRECISIONID SYSTEM FOR TARGETED SEQUENCING OF DNA AND RNA MARKERS FOR HUMAN IDENTITY AND BODY FLUID IDENTIFICATION

Multiplex analysis of 10 STR loci plus Amelogenin

TECHNICAL MANUAL. PowerPlex 16 System. Instructions for Use of Products DC6530 and DC6531. Revised 5/16 TMD012

DNA Typing for the Identification of Korean War Victims

Developmental validation of the Investigator Argus X-12 QS Kit

Use of the GenePrint 10 System with the Applied Biosystems 3500 and 3500xL Genetic Analyzers and POP-7 Polymer

Canine ISAG STR Parentage Kit (2014)

Report of Analyzing Short Tandem Repeats for Parentage Testing

PowerPlex ESI 17 Pro System

AmpFlSTR NGM PCR Amplification Kit

Forensic DNA collection: extraction of molecular information from buccal cells using direct amplification

Thermo Scientific Equine Genotypes Panel 1.1

Relative Fluorescent Quantitation on Capillary Electrophoresis Systems:

Thermo Scientific Bovine Genotypes Panel 2.2

VALIDATION OF PROMEGA S POWERPLEX 16 FOR PATERNITY TESTING

Thermo Scientific Canine Genotypes Panel 2.1

PowerPlex ESI 16 System

DNA 101. DNA 101:Kate Philpott and Jennifer Friedman

PowerPlex ESX 17 System

PowerPlex ESX 16 System

PowerPlex ESX 16 System

Capillary Electrophoresis and 5-Channel LIF Detection of a 26plex Autosomal STR Assay for Human Identification

Development of a Global STR Multiplex for Human Identification Analysis

TECHNICAL MANUAL. SwabSolution Kit. Instructions for Use of Product DC8271. Revised 9/16 TMD037

PowerPlex 4C Matrix Standard

QIAGEN Validation Report

STUDY OF VNTR HUMAN POLYMORPHISMS BY PCR

PowerPlex Y23 System for Use on the Applied Biosystems Genetic Analyzers

AmpFlSTR Identifiler Plus

AmpFlSTR NGM PCR Amplification Kit. User s Guide

PowerSeq 46GY System Eases the transition to MPS for forensic laboratories

Naming Convention and Laboratory Methods: PROVEDIt Database

Mentype Chimera Manual

Development of a Global STR Multiplex for Human Identification Analysis. Julio Mulero, Ph. D. R&D Group Human Identification, Life Technologies

Capillary Electrophoresis in DNA Analysis

Multiplex analysis of 19 STR loci plus Amelogenin. User manual

AmpFlSTR COfiler. PCR Amplification Kit. User Bulletin

PowerPlex ESI 16 System

Validation Studies on DNA Mixtures Using the PowerPlex Fusion 6C System

Transcription:

TECHNICAL NOTE Evaluation of a decreased cycle number (28 cycles) and decreased injection time (5 seconds) when the GlobalFiler PCR Amplification Kit is run on a 3130xl Genetic Analyzer The performance of the GlobalFiler PCR Amplification Kit (PN 4476135 and 4482815) run on the 3130xl Genetic Analyzer was recently reassessed. Internal testing shows that, depending on instrument sensitivity, samples amplified with the GlobalFiler kit and run under default HID injection conditions on the 3130xl can demonstrate off-scale peak heights for homozygous peaks when greater than or equal to 1 ng of DNA is targeted. This elevated peak height and the resulting spectral artifacts may impact data interpretation. In order to produce data that is within the linear dynamic range of the 3130xl instrument in this situation, a decreased PCR cycle number and a decreased injection time were evaluated and compared to the default PCR cycle number and injection time conditions. The following summarizes the internal studies executed at Thermo Fisher Scientific to verify the performance of the GlobalFiler kit at 28 and 29 cycles and injected on the 3130xl at 3 kv for 10 and 5 seconds. Summary of Internal Studies Three DNA samples (2 male, 1 female), positive control DNA 007, and GlobalFiler Allelic Ladder were processed on a single 3130xl Genetic Analyzer with Data Collection v4.0 with 6-Dye Module v1. Amplification was performed at 28 and 29 PCR cycles with the GlobalFiler kit. A sensitivity series was prepared and total DNA inputs of 2, 1, 0.5, 0.25, 0.125, and 0.063 ng were targeted. The data from the 1ng and the 0.125 ng samples are presented. The injection of the 28 and 29 PCR cycle data was performed at 3kV for 10 and 5 seconds. The Peak Amplitude Threshold (PAT) was set at 50 RFU for all studies. Performance was evaluated based on the following criteria: o Intra-color balance (ICB) 40% o Allele peak height ratio or intra-locus balance (ILB) 70% o No drop-out for the 0.125 ng DNA input samples For Research, Forensic or Paternity Use Only. 1

Experiments Performed A. Effect of injection time on signal intensity: Ten replicate runs of the GlobalFiler Allelic Ladder were run at 3kV for 10 and 5 second injection times on the 3130xl. Summary of Results Effect of injection time on signal intensity The channel-average peak height intensities of the GlobalFiler allelic ladder runs at 10 second and 5 second injection times are shown in Figure 1 and Table 1. Signal intensities decreased by an average of 43% when the injection time was decreased from 10 seconds to 5 seconds. Figure 1. Channel-average allelic ladder peak heights from 10 replicate runs each at 10-second and 5-second injection times. Table 1. Change in the channel-average signal intensity of the GlobalFiler allelic ladder upon decrease of the injection time from 10 seconds to 5 seconds. The % change is relative to the signal at 10 seconds. Channel Average peak height % change 10-seconds σ N-1 5-seconds σ N-1 FAM 1345 272 734 104 45% VIC 1059 220 591 89 44% NED 824 148 475 69 42% TAZ 899 167 507 74 44% SID 949 167 557 83 41% Average 1015 203 573 101 43% For Research, Forensic or Paternity Use Only. 2

B. Effect of changing the number of PCR cycles on signal intensity: The effect of changing the number of PCR cycles on signal intensity was evaluated using DNA samples F1, M1, and positive control at 1 ng/reaction in GlobalFiler STR assays with 28 and 29 PCR cycles. The injection time was set to 10 seconds for all three samples. Three replicates of samples F1 and M1 were run at both PCR cycling conditions. There were 6 replicate runs of the positive control at 28 cycles, and 9 replicate runs at 29 cycles. Summary of Results Effect of PCR cycles on signal intensity The channel-average peak height intensities of the three DNA samples are shown in Table 2. There is DNA-to-DNA differences in the relative change; F1 and M1 were more impacted by reduction in the number of cycles than the positive control, showing an average relative decrease in the average peak height of 33% and 38% respectively, while the positive control showed an average relative decrease of 26%. Table 2. Change in the channel-average signal intensity of PCR fragments from the GlobalFiler assay of three DNA samples at 28 and 29 cycles; the injection time was kept constant at 10 seconds. The % change is relative to the signal at 29 cycles. Sample F1 Channel Average peak height 28 cycles σ N-1 29 cycles σ N-1 % change FAM 2408 502 3544 692-32% VIC 3257 516 4684 292-30% NED 1987 250 3095 125-36% TAZ 2641 276 3933 293-33% SID 2145 281 3205 489-33% Average 2488 498 3692 644-33% M1 FAM 2051 485 3346 599-39% VIC 2321 346 3598 370-35% NED 2109 207 3872 553-46% TAZ 2597 590 4095 635-37% SID 2003 377 2956 325-32% Average 2216 245 3573 446-38% POS FAM 3315 382 4472 632-26% VIC 2902 382 3986 551-27% NED 1873 183 2871 244-35% TAZ 3817 485 4851 667-21% SID 2530 376 3310 495-24% Average 2887 742 3898 813-26% For Research, Forensic or Paternity Use Only. 3

C. Comparison of the ICB and ILB at 28 and 29 PCR cycles, and 10 and 5 second injection times: Three reference DNA samples (2 male, 1 female) with known genotypes and the positive control 007 DNA were tested in this study. Assay performance was tested at 1 ng of input DNA for all samples and 0.125 ng of input DNA for the three reference DNA samples. The ICB is calculated as the ratio of the minimum average peak height within a dye channel to the maximum average peak height within the same dye channel. It provides an indication of the marker-to-marker color balance within a dye channel. The target specification is an ICB 40%. For a particular dye channel, MIN (Averagepeak height) ICB = X100 MAX(Averagepeak height) Average peak height at a marker is genotype dependent and calculated according to the formulas below. Heterozygous allele: Peak height(allele1) Peak height(allele2) Average peakheight 2 Homozygous allele: Average peak height Peak height(allele1) 2 The ILB (also referred to as peak height ratio) provides a measure of the balance between alleles within a locus. The target specification is the peak height ratio between 2 alleles of the same marker must be 70%. The ILB is calculated as the ratio of the minimum peak height to maximum peak height between two heterozygous alleles, as shown in the calculation below. MIN (Allele 1, Allele 2) ILB for each marker = X100 MAX(Allele 1, Allele 2) In the example shown in Table 2, the ILB values were calculated from positive control data in the VIC dye channel. For Research, Forensic or Paternity Use Only. 4

Table 3. Calculation of the ILB for the VIC channel from DNA positive control data at 1 ng load, 29 cycles PCR, and 10-second injection time. Marker Allele 1 Allele 2 Peak Height 1 (RFU) Peak Height 2 (RFU) Average peak height (RFU) Yindel 2 2 4212 4212 2106 100% AMEL X Y 5331 4728 5030 89% D8S1179 12 13 4443 4525 4484 98% D21S11 28 31 3827 3760 3794 98% D18S51 12 15 4353 3984 4169 92% DYS391 11 11 5249 5249 2625 100% ILB The ICB for the VIC dye channel is the ratio of the minimum to the maximum average peak heights of AMEL, D8S1179, D21S11, and D18S51. Using the data in Table 2, ICB = 3794/5030 = 0.754 (75%) In the calculation of the VIC dye channel ICB, the Y-indel and DYS391 peak heights are excluded because these markers are haploid markers, and will typically give lower average signal intensities relative to other markers in the VIC dye channel. Summary of results Intra-color balance (ICB) and intra-locus balance (ILB) at 28 and 29 PCR cycles, and 10 and 5 second injection times: At 1 ng of total input DNA, the ICB does not change significantly between 28 and 29 cycles, at 5 and 10 second injection times in all the DNA samples tested (Figures 2 5). Figure 2. ICB of the positive control DNA sample at 1 ng load. The data includes 28 and 29 cycles and 5 and 10 second injection times. The ICB at all dye channels meets the target specification, and there is no significant difference between these values at all conditions tested. For Research, Forensic or Paternity Use Only. 5

Figure 3. ICB of the male sample M1 at 1 ng load. The data includes 28 cycles/10 second injection time and 29 cycles/5 and 10 second injection times. There is one outlier at 28 cycles/10 seconds in the FAM dye channel; however the average of the three replicates is greater than the target specification. Figure 4. ICB of the male sample M2 at 1 ng load. The data includes 28 cycles/5 second injection time and 29 cycles/5 and 10 second injection times. The ICB at all dye channels meets the target specification, and there is no significant difference between these values at all conditions tested. For Research, Forensic or Paternity Use Only. 6

Figure 5. ICB of the female sample F1 at 1 ng load. The data includes 28 cycles/10 second injection time and 29 cycles/5 and 10 second injection times. The ICB at all dye channels meets the target specification, and there is no significant difference between these values at all conditions tested. Figures 6 9 show the ILB for the four DNA samples tested in this study. At 1 ng of total input DNA, the ILB does not change significantly when reducing the number of cycles from 29 to 28 or the injection time from 10 to 5 seconds. Figure 6. ILB of the positive control 007 DNA at 1 ng input. Some replicates have outlier points with ILB <70%. However, the average ILB is > 70% and met the performance criteria specified. Similar observations were made for the two male samples (M1, M2) and the female sample (F1) in this study. For Research, Forensic or Paternity Use Only. 7

Figure 7. ILB of the male sample M1 at 1 ng input. The average ILB is > 70% Figure 8. ILB of the male sample M2 at 1 ng input. The average ILB is > 70%. Figure 9. ILB of the female sample F1 at 1 ng input. The average ILB is > 70%. For Research, Forensic or Paternity Use Only. 8

D. Combined effect of changing the number of cycles and injection time on CE signal intensity: Data from the entire sensitivity series (2, 1, 0.5, 0.25, 0.125, and 0.063 ng of DNA) were evaluated for the three reference samples at 28 and 29 PCR cycles and with 10 and 5 second injection times. With 2 ng DNA input, the GlobalFiler assay showed high signal intensities under all conditions tested such that the peak heights of several markers were off-scale. At 0.063 ng of DNA, some of the replicates showed partial profiles (PAT = 50 RFU) at both 28 and 29 cycles and at both 5 and 10 second injection times. The data analysis described below focuses on the 1 ng and 0.125 ng DNA input samples. Summary of Results Combined effect of changing PCR cycle number and inject time on CE signal intensity A variability gauge analysis was performed in this study. This is a method of data analysis designed to evaluate the relative contributions of changes in the assay parameters (cycle number and injection time) to the total variation in the results. All four of the DNA samples (3 reference and positive control 007) were included for the 1ng DNA input at 28 and 29 PCR cycles and 10 and 5 second injection times (Figure 10). The same analysis was done for the three reference samples at 0.125ng DNA input, 28 and 29 PCR cycles, and 10 and 5 second injection times (Figure 11). As indicated below, in both DNA template concentrations, a 5 second reduction in injection time has a greater contribution to signal decrease than reducing the number of PCR cycles by one. In other words, there is a more significant decrease in the signal intensity by decreasing the injection time to 5 seconds while keeping 29 PCR cycles, compared to decreasing the number of PCR cycles to 28, while keeping the injection time at 10 seconds. Figure 10. Variability gauge analysis at 1ng DNA input for all four DNA samples. Injection time has a more significant effect on signal intensity (45%) than changing the number of PCR cycles from 29 to 28 (17%). Some of the positive control peaks are off-scale at 29 PCR cycles/10 second injection. The peak amplitude threshold (PAT) was set to 50 RFU. For Research, Forensic or Paternity Use Only. 9

Figure 11. Variability gauge analysis at 0.125ng DNA for the three reference samples. The results are very similar to that obtained at 1ng DNA load. The injection time has a more significant effect on signal intensity (48%) than changing the number of PCR cycles from 29 to 28 (9%). The peak amplitude threshold (PAT) was set to 50 RFU. The correct allele calls were made for all four DNA samples at 1ng DNA load. 27 runs were generated from the three samples using a 0.125ng DNA input. Within this sample set there were two instances, of allele dropout where the signal intensities were less than 50 RFU. These instances both occurred at 28 cycles with a 5 second injection time (one was SE33 in the female sample, and the other D16S539 in the second male sample), Conclusions For 1 ng and 0.125 ng DNA input amounts, this study showed that there is a more significant decrease in the signal intensity by changing the injection time on the 3130xl Genetic Analyzer from 10 seconds to 5 seconds rather than decreasing the number of PCR cycles from 29 to 28. In the four DNA samples tested, there is no apparent negative impact of reducing the number of cycles from 29 to 28 or the injection time of the 3130xl Genetic Analyzer from 10 seconds to 5 seconds on the ICB or the ILB of the 1 ng DNA template samples. Instrument-to-instrument variability exists and some laboratories will have capillary electrophoresis instruments with higher sensitivity. Changing the injection time to reduce the signal intensity may be necessary in order to avoid signal saturation. Furthermore, laboratories can incrementally evaluate a range of injection times to optimize peak heights. Laboratories should perform internal validation studies in accordance with internal standard operating procedures to ensure that modifications to the default conditions are delivering sensitivity required to generate usable and reliable information. If changing the injection time is not sufficient in reducing signal saturation, reducing the number of cycles from 29 to 28 is another technique that may be investigated. For Research, Forensic or Paternity Use Only. 10

Applicability to Other Chemistries and Instruments The content of this technical note and the data generated was specific for the GlobalFiler chemistry run on a 3130xl Genetic Analyzer. Given that the mechanisms of PCR and CE across STR kits and CE instrumentation are the same, the results obtained and conclusions generated are highly likely to be applicable to other HID STR chemistries and capillary electrophoresis instruments. As noted previously, laboratories should perform the necessary internal validation studies to ensure that the selected PCR and injection parameters perform as expected in the laboratory system. For Research, Forensic or Paternity Use Only. 11