Impurity Detection with a New Light Emitting Diode Induced Fluorescence Detector Coupled to the Agilent 7100 Capillary Electrophoresis System

Similar documents
Application Note. Biopharm. Author. Abstract. Christian Wenz Agilent Technologies, Inc. Waldbronn, Germany

Comparison of Different Brands of Carrier Ampholytes for Monoclonal Antibody Charge Heterogeneity Analysis by Capillary Isoelectric Focusing

Capillary Electrophoresis of Proteins

IgG Purity/Heterogeneity and SDS-MW Assays with High- Speed Separation Method and High Throughput Tray Setup

Rediscover the Strengths of Your PA 800 Plus - A Multi-Attribute Protein Analyzer

Application Note. Abstract. Author. Biopharmaceutical

Assay of IgG Purity and Heterogeneity using High-Resolution Sodium Dodecyl Sulfate Capillary Gel Electrophoresis

deltadot ANTIBODY ANALYSIS Application Note: Antibodies using LABEL FREE INTRINSIC IMAGING LONDON United Kingdom

Simplifying CE-SDS Data Processing

Application Note. Authors. Abstract. Biopharmaceuticals

High-resolution Analysis of Charge Heterogeneity in Monoclonal Antibodies Using ph-gradient Cation Exchange Chromatography

Separate and Quantify Rituximab Aggregates and Fragments with High-Resolution SEC

LabChip GXII: Antibody Analysis

Disulfide Linkage Analysis of IgG1 using an Agilent 1260 Infinity Bio inert LC System with an Agilent ZORBAX RRHD Diphenyl sub 2 µm Column

Automated and quantitative analysis of biologics

LIFluor EnhanCE. for use with dsdna 1000 and dsdna 20,000 Kits

Characterization of mab aggregation using a Cary 60 UV-Vis Spectrophotometer and the Agilent 1260 Infinity LC system

Automated and quantitative

Charge Heterogeneity Analysis of Rituximab Innovator and Biosimilar mabs

TECHNICAL INFORMATION

Unified Workflow for Monoclonal Antibody Charge Heterogeneity, Purity, and Molecular Weight Analysis

Optimization of cief to Resolve Challenging MAb Samples of Clinical Immunodiagnostic Relevance

Size Exclusion Chromatography of Biosimilar and Innovator Insulin Using the Agilent AdvanceBio SEC column

Capillary Electrophoresis in Quality Control PART II: CE-SDS: Method Development and Robustness

Separation of Recombinant Human Erythropoietin (rhepo) using the European Pharmacopoeia Method on the PA 800 plus Pharmaceutical Analysis System

Application Note. Author. Abstract. Biopharmaceuticals. Verified for Agilent 1260 Infinity II LC Bio-inert System. Sonja Schneider

AnaTag 5-FAM Protein Labeling Kit

E-Coli Cell Lysate By High Performance Capillary Electrophoresis

Analysis of Monoclonal Antibody Charge Variants by Capillary Zone Electrophoresis

Applications. Capillary Electrophoresis. Cell Lysate Analysis. Analysis of E. Coli Lysate. Technology

High-throughput and Sensitive Size Exclusion Chromatography (SEC) of Biologics Using Agilent AdvanceBio SEC Columns

CSE Separation of Cancer Cell Lysate

HT Pico Protein Express LabChip Kit, Frequently Asked Questions

SDS-CGE Method Development for Purified Protein Vaccine Antigens

Molecular Characterization of Biotherapeutics The Agilent 1260 Infi nity Multi-Detector Bio-SEC Solution with Advanced Light Scattering Detection

AnaTag HiLyte Fluor 750 Microscale Protein Labeling Kit

Analysis of Intact and C-terminal Digested IgG1 on an Agilent Bio MAb 5 µm Column

AnaTag HiLyte Fluor 647 Protein Labeling Kit

Reducing Cycle Time for Charge Variant Analysis of Monoclonal Antibodies

AdvanceBio HIC: a Hydrophobic HPLC Column for Monoclonal Antibody (mab) Variant Analysis

Separation of Native Monoclonal Antibodies and Identification of Charge Variants:

HPCE-5 2. Realising the potential of Capillary Electrophoresis. Why is the HPCE-512 so much better?

AnaTag HiLyte Fluor 555 Protein Labeling Kit

Monoclonal Antibody Charge Variants Identification by Fully Automated Capillary Isoelectric Focusing Mass Spectrometry

Improved Sensitivity with the Agilent 1290 Infinity Evaporative Light Scattering Detector (ELSD)

Size Exclusion BioHPLC Columns

Journal of Chemical and Pharmaceutical Research, 2017, 9(3): Research Article

Microfluidic Assays for High Throughput Screening of Protein Quality in Bioprocess Development

AnaTag HiLyte Fluor 488 Microscale Protein Labeling Kit

mab Titer Analysis with the Agilent Bio-Monolith Protein A Column

Update on Heparin Contamination: Assessment of Optimized Capillary Electrophoresis, Proton NMR and Ion-Exchange HPLC Methodologies

Automated and Quantitative Analysis of Biologics

PA 800 plus Catalog. Systems/Software

Clone Selection Using the Agilent 1290 Infinity Online 2D-LC/MS Solution

Rapid UHPLC Analysis of Reduced Monoclonal Antibodies using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) 300SB-C8 Column

APPLICATION INFORMATION

Performance characteristics of the 1260 Infinity Quaternary LC system

SCIEX Anion Analysis Kit For P/ACE MDQ and P/ACE MDQ plus Capillary Electrophoresis Systems. Instruction Guide

AnaTag HiLyte Fluor 647 Microscale Protein Labeling Kit

Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC

"VUPNBUFE BOE RVBOUJUBUJWF

Knock! Knock! Who s there? Maurice Who? SungAe Suhr Park, Mee Ko, Eleanor Le, Janice Chen Amgen Inc.

A Comprehensive Workflow to Optimize and Execute Protein Aggregate Studies

Repeatability of C100ht Biologics Analyzer, for High Throughput Glycan Screening

NISTmAb characterization with a high-performance RP chromatography column

HPCE-5 2. A Revolution in Capillary Electrophoresis. High Performance Capillary Electrophoresis

Maximizing Chromatographic Resolution of Peptide Maps using UPLC with Tandem Columns

Bio-Monolith Protein G Column - More Options for mab Titer Determination

Application Note. Biopharma. Authors. Abstract. James Martosella, Phu Duong Agilent Technologies, Inc Centreville Rd Wilmington, DE 19808

Alternative to 2D gel electrophoresis OFFGEL electrophoresis combined with high-sensitivity on-chip protein detection.

CE Oligonucleotide Analysis Kit Instruction Manual

Quantification of genotoxic "Impurity D" in Atenolol by LC/ESI/MS/MS with Agilent 1200 Series RRLC and 6410B Triple Quadrupole LC/MS

Fast and High Resolution Analysis of Intact and Reduced Therapeutic Monoclonal Antibodies (mabs)

High-Throughput Glycan Screening to Remove Biopharmaceutical Development Bottlenecks

IgG Purity and Heterogeneity Assay Kit For the PA 800 Plus Pharmaceutical Analysis System. Application Guide

Capillary electrophoresis of heparin and related impurities using highly concentrated buffers in a 25 µm bubble cell capillary

Characterize mab Charge Variants by Cation-Exchange Chromatography

Best Quantification Practices with the Agilent 5200 Fragment Analyzer System

Developing Robust and Efficient IEX Methods for Charge Variant Analysis of Biotherapeutics Using ACQUITY UPLC H-Class System and Auto Blend Plus

Fraction Analysis of Cysteine Linked Antibody-Drug Conjugates Using Hydrophobic Interaction. chromatography. Agilent 1260 Infinity II Bio-Inert System

Reversed-phase Separation of Intact Monoclonal Antibodies Using Agilent ZORBAX Rapid Resolution High Definition 300SB-C8 1.

Performance Characteristics of the Agilent 1220 Infinity Gradient LC system

Quantitative analysis of PCR fragments with the Agilent 2100 Bioanalyzer. Application Note. Odilo Mueller. Abstract

High Throughput Sub-4 Minute Separation of Antibodies using Size Exclusion Chromatography

Separating proteins with activated carbon

Report. Comparison of magnetic beads coating protocols for immunoaffinity capillary electrophoresis. Natalia Gasilova, Hubert H.

LavaDigest Protease Monitoring Kit

Electronic Supplementary Information for The effect of protein concentration on the viscosity of a recombinant albumin solution formulation

Agilent AdvanceBio SEC Columns for Aggregate Analysis: Instrument Compatibility

XactEdit Cas9 Nuclease with NLS User Manual

One-stop cief and CE-SDS. One-day method dev. Results in a snap. FDA, PDQ.

Chapter 2 Coated Capillaries for CE-MS of Therapeutic Protein

Application Note. Author. Abstract. Pharmaceuticals. Detlef Wilhelm ANATOX GmbH & Co. KG. Fuerstenwalde, Germany mau

Quantification of Host Cell Protein Impurities Using the Agilent 1290 Infinity II LC Coupled with the 6495B Triple Quadrupole LC/MS System

Lifetime stability of size exclusion chromatography columns for protein aggregate analysis

Analysis of amoxicillin and five impurities on the Agilent 1220 Infinity LC System

Novel instrumentation to gain insight into the aggregation state of proteins by Taylor Dispersion Analysis: a comparison with Dynamic Light Scattering

Technical Note. Introduction

This is not unique to CE. It took for instance about 25 years for HPLC to establish itself properly in the pharmaceutical industry.

CEofix NTMP. Determination of Proteins and Peptides with Capillary Electrophoresis. Introduction. Methods and Materials.

Transcription:

Impurity Detection with a New Light Emitting Diode Induced Fluorescence Detector Coupled to the Agilent 7100 Capillary Electrophoresis System Application Note Biologics & Biosimilars, Proteomics & Protein Science Author Christian Wenz Agilent Technologies, Inc. Waldbronn, Germany Abstract Light emitting diodes could present alternatives to lasers as light sources for fl uorescence detectors in applications requiring high sensitivity such as impurity detection in purifi ed monoclonal antibody preparations by capillary gel electrophoresis. This Application Note describes the performance of a detector for diode induced fl uorescence from Picometrics coupled to the Agilent 7100 Capillary Electrophoresis System. It shows that a small spiked-in protein impurity in a monoclonal antibody sample derivatized with 3-(2-furoyl)-quinoline-2- carboxaldehyde can be quantifi ed with good accuracy and precision down to a level of 0.0 %. The extrapolated limit of detection was 0.02 %, a value that is close to limits of detection reported for laser-based systems.

Introduction Capillary gel electrophoresis (CGE) is an important tool for biopharmaceutical product development and quality control 1. It is used to monitor product purity, detect minor product or process related impurities, and confirm batch-to-batch consistency of therapeutic proteins, especially monoclonal antibodies (mab). In addition to UV detection 2, protein derivatization combined with laser-induced fluorescence (LIF) is often used to enhance the sensitivity of the method. Examples of popular derivatization reagents are -carboxytetramethylrhodamine succinimidyl ester 3 and 3-(2-furoyl)- quinoline-2-carboxaldehyde (FQ) 4,. However, it has been shown that light emitting diodes (LEDs) can be used as alternative light sources for fluorescence detection in CGE. A recently published comparative study of LED and laser-based systems showed a very comparable signal-to-noise (S/N) for an antibody sample derivatized with the above-mentioned reagents, provided that a similar excitation wavelength was used in both cases 6. Potential benefits of LED in comparison to lasers are lower energy consumption, increased baseline stability, and longer lifetime. This Application Note describes the performance of a detector for LED-induced fluorescence (LEDIF) from Picometrics coupled to the Agilent 7100 Capillary Electrophoresis system for the detection of low-level impurities in an mab sample. Impurities were simulated by artificially contaminating a purified mab preparation with small standard proteins. Prior to CGE analysis, samples were derivatized with FQ. Derivatization with this reagent is convenient because it becomes fluorescent only after reaction with a primary amine and, therefore, purification of the derivatized protein is not necessary. Experimental Instrumentation Agilent 7100 Capillary Electrophoresis System (G7100A) Materials The mab used was provided by a biotechnical company. Standard proteins, citrate-phosphate buffer concentrate, N-ethylmaleimide (NEM), and sodium dodecyl sulfate (SDS) were purchased from Sigma Aldrich (St. Louis, MO, USA). Potassium cyanide (KCN) and FQ were purchased from Life Technologies (Paisley, UK). SDS Gel Buffer was purchased from Beckman Coulter (Fullerton, CA, USA). Protein concentrations were measured with the Qubit assay (Life Technologies, Paisley, UK). The Zetalif LED detector equipped with a 480 nm LED and a high pass 10 nm emission filter was obtained from Picometrics (Toulouse, France). All other materials and instrumentation were from Agilent Technologies, Inc. (Waldbronn, Germany). Sample preparation A stock solution of 40 mm FQ in methanol was prepared and stored at 20 C. Immediately before use, an aliquot of this stock solution was diluted with water to 1 mm. NEM solutions were prepared fresh daily. Prior to derivatization, the mab sample was diluted with citrate-phosphate buffer (90 mm sodium citrate, 77 mm sodium phosphate, ph 6.) to a protein concentration of 2 mg/ml. For derivatization with FQ, the following reagents were mixed in 0.-mL microcentrifuge vials: 3 µl of diluted mab 7.0 µl of 10 % SDS 4.0 µl of 100 mm NEM 0. µl of 100 mm KCN 2 µl of 1 mm FQ Final concentrations in the reaction mixture were 1 mg/ml mab, 1 % SDS,.6 mm NEM, 0.7 mm KCN, and 0.3 mm FQ. The solution was incubated for minutes at 7 C and then cooled down on ice. After the addition of 30 µl 10 % SDS, the solution was transferred into 100-µL CE sample vials. For reducing analysis, samples were additionally incubated with 0 mm DTT for 10 minutes at 70 C. CE conditions The LEDIF detector was connected to the CE instrument through the integrated analog-to-digital converter. The range used was 0 /V. The rise time and photomultiplier high voltage were set to 0. s and 610 V, respectively. Fifty µm id bare fused silica capillaries with a total length of 33 cm and an effective length 21 cm were used. Once a day, capillaries were conditioned as follows: 1. High-pressure flush at 2 bar with 0.1 N NaOH for 10 minutes, with 0.1 N HCl for minutes, and with water for 2 minutes 2. High-pressure flush at 4 bar with SDS Gel Buffer for 10 minutes 3. Water dip for both electrodes 4. Voltage equilibration at 16. V for 10 minutes with minutes ramping. Prior to every run, capillaries were conditioned as follows: 1. High-pressure flush at 4 bar with 0.1 N NaOH for 3 minutes, with 0.1 N HCl for 1 minute, with water for 1 minute and with SDS Gel Buffer for 10 minutes 2. Water dip of both electrodes Samples were electrokinetically injected from the outlet vial position by applying kv for 20 seconds and, after a water dip of the outlet electrode, separated by applying 16. kv (00 V/cm) for 30 minutes. Two bar pressure was applied to both inlet and outlet home vials during the run. After use, capillaries were conditioned as follows: 2

1. High-pressure flush at 4 bar with 0.1 N NaOH for 1 minutes 2. High-pressure flush at 3. bar with 0.1 N HCl for minutes, and with water for 10 minutes. Flushes for daily preconditioning and conditioning after use were done in forward direction (that is, pressure was applied to the inlet vial). Prerun flushes were done in backward direction. The capillary temperature was kept at 40 C. For all reagents, 2-mL glass vials were used. The fill volume was 1.2 ml, except for the water dip vials that contained 1.6 ml water and the waste vials that contained 0.6 ml of water. Three separate waste vials were used for the collection of 0.1 N NaOH, 0.1 N HCl/water, and SDS Gel Buffer, respectively. SDS Gel Buffer containing inlet and outlet home vials were exchanged after every sequence of eight runs. Results and Discussion Sensitivity of CGE with LEDIF detection The analysis of a FQ-derivatized mab sample by CGE with LEDIF detection is shown Figure 1. Two detector parameters were optimized in pilot experiments. Firstly, the photomultiplier high voltage was adjusted to a value that resulted in main peak maxima in the range of 30 40. This was done to take advantage of the full measuring range without risking the rise of the signal beyond the 0 limit. Secondly, the baseline noise reduction parameter rise time was varied between 0 and seconds and then set to a value of 0. seconds that resulted in an acceptable background noise level without affecting the peak shape of small mab impurities (data not shown). The peak present in all runs at approximately 7. minutes might represent a buffer contaminant 4 (Figure 1). The background noise was calculated according to the ASTM over time intervals of 2 minutes around the migration times of the peaks of interest and was between 2 2. m. To estimate the sensitivity of the method, the mab sample was spiked with four standard proteins at a level of 0.2 % (Figure 1). Under nonreducing conditions, where all standard proteins were clearly resolved from mab derived peaks, peak heights of approximately 100 m were observed for three out of four protein spikes, corresponding to an S/N of 40 and an extrapolated limit of detection (LOD) of approximately 0.02 % (S/N > 3). This value is close to LODs reported for protein impurities in FQ-derivatized mab samples obtained with laser-based systems 4,. 3 30 Intact IgG 9.1 9.0 * * * 2 8.9 * 20 8.8 1 8.7 10 8.6 8. 3 30 2 Light chain Heavy chain 9.0 8.9 8.8 * * * * 20 8.7 1 8.6 10 8. 8.4 Figure 1. Sensitivity of CGE with LEDIF detection. Samples were analyzed under nonreducing (top) and reducing conditions (bottom). Shown are full-scale electropherograms (left) and an expanded view (right) of the original mab sample (dark blue), the mab sample artificially contaminated with standard proteins at a concentration of 0.2 % (blue) and a control sample without protein (light blue). Molecular weights of standard proteins were 12, 18, 29, and 4 kda, respectively. Standard protein peaks are labeled (*). 3

Impurity quantification The performance of the method for low-level impurity quantification was evaluated with mab samples spiked with a single standard protein at concentrations of 0.0 0.2 % (Figure 2). Samples were derivatized and analyzed on two days using different CE instruments. Peak integration was done automatically with the same set of parameters (Table 1). The maximum deviation of the measured standard protein % area from the target value was 12 %, and the average deviation was 7.1 %. A precision of 10 RSD % or better was observed for standard protein concentrations 0.1 %. Only for the lowest concentration was the precision 1 RSD %. An S/N of 8 was calculated for the 0.0 % sample, which confirms the above-mentioned LOD estimation. 4 40 3 30 2 20 1 10 12 kda LC HC 8.4 8 9 10 8.7 8.6 8. 12 kda Figure 2. Limit of quantification. A 12 kda standard protein was added at concentrations between 0.0 0.2 % to the mab sample and analyzed under reducing conditions. Shown is an overlay of full-scale electropherograms of mab without (bottom) and with 12 kda protein added at concentrations of 0.0, 0.10, 0.1, 0.20, and 0.2 % (top) with a zoom on the 12 kda protein peaks in the inset. Table 1. Accuracy and precision of low level impurity detection by CGE with LEDIF detection. Samples of mab with different amounts of added 12 kda protein (cf. Figure 2) were derivatized and analyzed on two different days with two CE instruments and four runs per instrument (n = 8). Shown is the relative time corrected area (% Area) and the S/N for the 12 kda protein. % Area S/N Target Average SD RSD % %Deviation from target Average SD RSD % 0.00 0.048 0.007 1.0 4.0 7.9 0.9 12.0 0.100 0.092 0.008 8.9 8.1 14. 1.8 12.6 0.10 0.131 0.010 8.0 12.4 22.7 3.0 13.2 0.200 0.181 0.009 4.8 9.6 29.7 3.4 11. 0.20 0.247 0.02 10.2 1.3 37.0 3.6 9.6 4

Conclusion This work shows that the Agilent 7100 CE instrument equipped with the Picometrics LED-based fluorescence detector is well suited for low-level impurity detection by CGE with FQ-derivatized mab samples. The method allowed the detection of standard proteins spiked into an mab preparation down to a level of 0.02 %, a value that is close to published LODs obtained with laser-based systems. Good accuracy and precision was observed for the quantification of a single protein spike at concentrations in the range of 0.0 0.2 %. Given the similar performance combined with lower energy consumption, increased stability, and longer lifetime of LEDs, the described approach may present an attractive alternative to CE systems with laser-based fluorescence detectors. References 1. A. Guo, et al. Capillary electrophoresis methods for pharmaceutical analysis; Vol. 9 of separation sciences and technology series Ahuja, S. and Jimilar M. I. (Eds.), Academic Press, Chapter 14, (2008). 2. C. Wenz Performance of commercially available gels for protein characterization by capillary gel electrophoresis with UV detection on the Agilent 7100 CE system Agilent Technologies application note, publication number 990-7976EN. 3. O. Salas-Solano, et al. Optimization and validation of a quantitative capillary electrophoresis sodium dodecyl sulfate method for quality control and stability monitoring of monoclonal antibodies, Anal. Chem. 78, 683-94, (2006). 4. D.A. Michels, et al. Fluorescent derivatization method of proteins for characterization by capillary electrophoresis-sodium dodecyl sulfate with laser-induced fluorescence detection Anal. Chem. 79, 963-71, (2007).. D.A. Michels, M. Parker, and O. Salas-Solano Quantitative impurity analysis of monoclonal antibody size heterogeneity by CE-LIF: example of development and validation through a quality-by-design framework Electrophoresis 33, 81-26, (2012). 6. A. Rodat-Boutonnet, et al. A comparative study of LED-induced fluorescence and laser-induced fluorescence in SDS-CGE: application to the analysis of antibodies Electrophoresis 33, 1709-14, (2012).

www.agilent.com/chem/ce This information is subject to change without notice. Agilent Technologies, Inc., 2014 Published in the USA, March 1, 2014 991-4043EN