Nitrogen Determination in Soils and Plants by Flash Combustion using Argon as Carrier Gas

Similar documents
Elemental Analysis: CHNS/O determination in coals

The Benefits of SOLAμ Technology in Sample Preparation. Jon Bardsley and Ken Meadows Thermo Fisher Scientific, Runcorn, UK

Isotope Ratio MS. Continuous Flow. Consumables and Spare Parts for Thermo Scientific Elemental Analyzers

Analysis of Fracking Flowback Water From the Marcellus Shale Using In-line Conductivity, Automated Dilution, and Ion Chromatography

Concentration of Human Hormones in Drinking Water Using Solid Phase Extraction and Analysis by High Performance Liquid Chromatography

Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

Automated Solid-Phase Extraction of Organochlorine Pesticides from Drinking Water. Pranathi R Perati Thermo Fisher Scientific, Sunnyvale, CA, USA

Jonathan R. Beck and Charles T. Yang; Thermo Fisher Scientific, San Jose, CA

Determination of 1,4-Dioxane in Drinking Water by Gas Chromatography/Mass Spectrometry (GC/MS) with Selected Ion Monitoring (SIM)

A Fast and Robust Linear ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge Variant Analysis

Claudio De Nardi, 1 Marta Kozak, 2 David Kasper 3 1. Thermo Fisher Scientific, Dreieich, Germany; 2 Thermo Fisher Scientific, San Jose, CA (USA); 3

Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease

Improving Retention Time Precision and Chromatography of Early Eluting Peptides with Acetonitrile/Water Blends as Solvent B

Thermo Scientific icap 7000 Plus Series ICP-OES. Gain more power. experience more performance

Thermo Scientific Solutions for Intact-Protein Analysis. Better, Faster Decisions for. Biotherapeutic Development

Lebedev A, 1 Damoc E, 2 Makarov A, 2 Samguina T 1 1. Moscow State University, Moscow, Russia; 2 ThermoFisher Scientific, Bremen, Germany

Analysis of Illegal Dyes in Food Matrices Using Automated Online Sample Preparation with Liquid Chromatography-Mass Spectrometry

TargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet

Thermo Scientific KIEL IV Carbonate Device

Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass

Sprint analysis of lubricating oils using the Thermo Scientific icap 7600 ICP-OES

A Novel Screening Method for Anthropogenic Sewage Pollutants in Waste Water, Ground Water and Drinking Water Samples by LC HRAM Analysis

Thermo Scientific EASY-nLC 1200 System. Leading in simplicity. and performance

Simple, fast and reliable analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS

Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease

ANALYTICAL STANDARDS AND REAGENTS

Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer. Extreme quantitative value. with unprecedented ease

Using HRAM Survey Analysis Combined with Rapid MS2 Data to Develop a Fragmentation Based Detection Workflow for Structure ID Acquisition

Global Commerce Review EMEA, Q2 2018

Extraction of Phthalates from Solid and Liquid Matrices

single-use fermentors

Global Commerce Review. Americas, Q2 2018

Thermo Scientific FLASH 2000 Series Organic Elemental Analyzers

Elemental Analysis (EA)

Quantitative Analysis of Wheat Flour Using FT-NIR

Analysis of Trace Elements in Seawater Using the Thermo Scientific icap 7000 Series ICP-OES Duo

GRI Sustainability Reporting Statistics Publication year By Report Services

integrated solutions in material handling

Chrom-Card Data System. A simple, chemist-oriented data handling system...productive from the first day!

for water and beverage analysis

Bringing custom solutions and contract manufacturing to life

A Novel ph Gradient Separation Platform for Monoclonal Antibody (MAb) Charge-Variant Analysis

EDS Phase Mapping of a Contact Metamorphosed Calc-Silicate Rock

Analysis of Glyphosate and AMPA in Environmental Water Samples by Ion Chromatography and On-Line Mass Spectrometry with Minimal Sample Preparation

Gasification of Biomass and Waste Recent Activities and Results of IEA Bioenergy Task 33

Development of Ultra-fast ph-gradient Ion Exchange Chromatography for the Separation of Monoclonal Antibody Charge Variants

Hillary B. Hewitson, Thomas E. Wheat, Diane M. Diehl INTRODUCTION

High-speed, High-resolution Oligonucleotide Separations Using Small Particle Anion-Exchangers

Chromatography Column Performance and Data Analysis Success Guide. Hints and Tips for Better Purifications

Multi-element determination in pharmaceutical preparations using the Thermo Scientific icap Q ICP-MS

PRODUCT INFORMATION. DWS Pallet MEASURING AND WEIGHING PALLETS AND OTHER BULKY FREIGHT ITEMS. Track and trace systems

FRACTIONLYNX APPLICATION MANAGER

Thermo Scientific Dionex ICS-900 Ion Chromatography System

ALLIANCE SYSTEM FOR CARBAMATE ANALYSIS

Overview. Tools for Protein Sample Preparation, 2-D Electrophoresis, and Imaging and Analysis

Thermo Scientific Qtegra Intelligent Scientific Data Solution. Delivering quality. Driving productivity

EU Water Analysis Using the Thermo Scientific icap 7400 ICP-OES Duo

Travel Flash Report. Winter 2018

Product Quality Control of a HME Co-Extrudate Using a Raman Imaging Microscope

Thermo Scientific icap 6000 Series ICP Emission Spectrometer

BIOANALYTICAL STRATEGY FOR IN VITRO METABOLITE SCREENING WITH EXACT MASS USING THE Q-Tof micro. Jose M. Castro-Perez 1, Carina Leandersson 2

Solution Partner Program Global Perspective

DOW PuraGuard Propylene Glycol USP/EP. Your Security for Purity

Liner Style* Function Advantages Disadvantages Recommended For Straight Low surface area for less activity Least expensive Low activity

Simplifying Complex Multi-Residue Pesticide Methodology in GC-MS/MS

Analysis of elemental impurities in drug products using the Thermo Scientific icap 7600 ICP-OES Duo

PrimePCR Pricing and Bulk Discounts

Welcome to the Bank Statement Processing introduction topic. During this training, we use the acronym BSP for Bank Statement Processing.

Chemical Compatibility of the Pall QPoint TM Filter Capsule - Rose Outlet with Systemic Water Treatments

Siemens Partner Program

PrimePCR Pricing and Bulk Discounts

PrimePCR Pricing and Bulk Discounts

Method Optimisation in Bottom-Up Analysis of Proteins

Thermo Scientific EASY-Spray Technology. Plug-and-spray with. State-of-the-art performance

MODULAR SERVICE CONTRACTS THE SERVICE STRATEGY FOR PROCESS AUTOMATION. SICK LifeTime Services

Automatic Determination of Greenhouse Gases by GC

no matter the culture we have a solution

safetyiq: THE NEW DIMENSIONS OF SAFETY DISCOVER INTELLIGENT PROTECTION FOR MORE PRODUCTIVITY Forward-thinking safety products, systems and services

PRODUCT INFORMATION MCS100E HW/PD/CD MULTICOMPONENT ANALYZER SYSTEMS EMISSION AND RAW GAS MONITORING WITH HOT MEASUREMENT.

2009 Talent Shortage Survey Results

High-Throughput Oligonucleotide Analysis on a Small Particle Pellicular Anion-Exchanger

WORKFORCE METRICS BENCHMARK REPORT

5 REASONS TEKTRONIX OSCILLOSCOPES ARE RELIABLE AND TRUSTED TECHNICAL BRIEF

IFS FIELD SERVICE MANAGEMENT

Quality Control Measures for Routine, High-Throughput Targeted Protein Quantitation Using Tandem Capillary Column Separation

Highlights. Figure 1. World Marketed Energy Consumption by Region,

ENSURING QUALITY THROUGH COMPLIANCE [ COMPLIANCE ]

Multidrop. The most precise and reliable reagent dispensing solution. Analyze Detect Measure Control TM

Imaging and Electrophoresis. Gel Doc EZ Imager. Accelerate your research. Push-button imaging in just 10 seconds.

2009 Talent Shortage Survey Results

protein interaction analysis bulletin 6299

Optimize Your Process Operations by Improving Process Monitoring

Key results from the survey on organic. Organic Agriculture Worldwide. agriculture worldwide 2012

CRYOGENIC. B a l l V a l v e s

CXP GROUP. SITSI - your source for exceptional analyst insight

Energy & Climate Change ENYGF 2015

TECHNICAL INFORMATION

GLOBAL VIDEO-ON- DEMAND (VOD)

Transcription:

Nitrogen Determination in Soils and Plants by Flash Combustion using Argon as Carrier Gas Liliana Krotz, Walter Galotta and Guido Giazzi Thermo Fisher Scientific, Milan, Italy

Overview Purpose: To show nitrogen determination of soils and plants by Organic Elemental Analysis (OEA) using argon as the carrier gas. Methods: Soils and plants were analyzed by an elemental analyzer with an automatic autosampler, using argon as the carrier gas. Results: Data collected of nitrogen from different soils and plants are discussed to assess the performance of the elemental analyzer using argon as the carrier gas. Introduction Nitrogen in soils is important for the evaluation of organic matter and the calculation of the amount of fertilizer to be added by providing information regarding the deficiency or excess of nutritional elements important for plant growth. Nitrogen content is important in determining the quality of various types of crops for feeding and processing, as well as for N-cycle and N-fixation monitoring in agricultural and environmental research. For this reason the use of an accurate instrumental analytical technique is required, as the demand for improved sample throughput, reduction of operational costs and minimization of human errors is becoming every day more notable, it is essential to have a simple and automatic technique which allows fast analysis with excellent reproducibility. The Thermo Scientific TM FLASH 2000 analyzer (Figure 1), which typically uses helium as its carrier gas and is based on the dynamic flash combustion of the sample, copes effortlessly with the wide array of laboratory requirements such as accuracy, day to day reproducibility and high sample throughput. However as in recent years there have been worldwide shortages and large increases in the cost of helium, and it has been necessary to test an alternative gas, argon, which is readily available. This paper presents data on nitrogen determination in soil and plant reference materials with varying nitrogen concentrations to show the performance of the system using argon as the carrier gas, and the reproducibility of the results obtained. 2 Nitrogen Determination in Soils and Plants by Flash Combustion using Argon as Carrier Gas

FIGURE 1. FLASH 2000 Elemental Analyzer Methods Samples are weighed in tin capsules and introduced into the combustion reactor via the Thermo Scientific TM MAS TM 200R autosampler together with the proper amount of oxygen. After combustion, the produced gases are carried by an argon flow to a second reactor filled with copper, then swept through CO 2 and H 2 O traps, a GC column, finally being detected by a thermal conductivity detector (TCD). The analytical configuration as well as the TCD detector are the same as those used with helium as the carrier gas (Figure 2). A complete report is automatically generated by the Thermo Scientific TM Eager Xperience data handling software and displayed at the end of the analysis. The Eager Xperience software provides a new AGO (Argon Gas Option) function which allows the modification of the argon carrier flow during the run to optimize the analysis. FIGURE 2. FLASH 2000 Nitrogen configuration. Analytical conditions Combustion Furnace Temperature: 950 C Reduction Furnace Temperature: 840 C Oven Temperature: 50 C (GC column inside the oven) Argon Carrier Flow: 60 ml/min Argon Reference Flow: 60 ml/min Oxygen Flow: 300 ml/min Sample Delay: 10 sec Run Time: 10 min Note: The oxygen amount necessary for the complete combustion of samples is calculated automatically by the Thermo Scientific TM OxyTune TM function present in the Eager Xperience software. Thermo Scientific Poster Note PN42209_PITTCON 2014_E_03/14S 3

Results To evaluate the performance of the system for soil and plant analysis, relative reference materials with different nitrogen concentrations were chosen. Instrument calibration was performed with approximately 50-70 mg of Aspartic acid (10.52 %N) standard using K factor as calibration method. Table 1 shows the nitrogen data obtained of the analysis of Thermo Scientific TM Soil Reference Material. The sample weight was 250 350 mg. The certified N % is 0.21 and the accepted range according to the technical specification is 0.19 0.23 N%. The average of the 12 runs is 0.19 %N with a RSD % of 2.55. TABLE 1. Nitrogen data of Thermo Scientific Soil Reference Material. Weight (mg) N % 250.224 0.20 260.746 0.20 270.494 0.19 300.714 0.19 310.893 0.19 320.423 0.19 270.794 0.19 280.101 0.19 290.134 0.19 300.844 0.19 320.242 0.20 350.896 0.20 Four Soil Reference Materials from 0.13 to 0.35 N % were chosen to correlate the experimental results to the expected values. Table 2 shows the certified N % and the uncertainty. Table 3 shows the weight of sample used and the experimental N % obtained which was very satisfactory. TABLE 2. Expected N % of Soil Reference Materials. Sample Description Specification N % Uncertainty (±) Low Organic Content Soil 0.133 0.023 Medium Organic Content Soil 0.27 0.02 Loamy Soil 0.27 0.02 Chalky Soil 0.35 0.02 TABLE 3. Experimental Nitrogen data of Soil Reference Materials. Sample Description Weight (mg) Experimental N % RSD % Low Organic Content Soil 0Reference Material 351.931 307.521 340.210 0.117 0.113 0.112 2.321 Medium Organic Content Soil 0Reference Material 285.120 336.816 322.010 0.264 0.258 0.265 1.443 Loamy Soil Reference Material 322.020 326.020 340.567 0.265 0.271 0.268 0.119 Chalky Soil Reference Material 254.416 326.987 350.471 0.337 0.365 0.372 5.173 4 Nitrogen Determination in Soils and Plants by Flash Combustion using Argon as Carrier Gas

A Sandy Soil Reference Material at 0.07 N % (700 ppm N) was chosen to evaluate the method at trace level. Table 4 shows the weight of sample used, the experimental N % obtained, the certified N % and the uncertainty. The data obtained was very satisfactory. TABLE 4. Nitrogen data of Sandy Soil Reference Material. Sample FLASH 2000 Specification Weight (mg) Experimental N % 408.698 0.0700 402.849 0.0711 398.636 0.0703 409.065 0.0707 RSD % N % Uncertainty (±) 408.650 0.0691 1.8848 0.07 0.01 412.241 0.0718 403.313 0.0694 401.651 0.0692 400.316 0.0718 400.420 0.0732 Three Plant Reference Materials from 0.15 to 3.01 N % were chosen to correlate the experimental results to the expected values. Table 5 shows the certified N % and the uncertainty. TABLE 5. Expected Nitrogen values of Plant Reference Material. Sample Description Specification N % Uncertainty (±) Birch Leaves 2.12 0.06 Orchard Leaves 2.28 0.04 Alfalfa 3.01 0.20 Table 6 shows the experimental N% obtained which were very satisfactory. The nitrogen average of the ten runs is inside the uncertainty. TABLE 6. Experimental Nitrogen data. Sample Birch Leaves Orchard Leaves Alfalfa Weight (mg) N % Weight (mg) N % Weight (mg) N % 203.77 2.11 100.46 2.24 120.889 3.06 162.55 2.12 120.53 2.24 121.957 3.07 162.55 2.11 120.21 2.26 119.196 3.09 165.10 2.13 140.23 2.30 122.920 3.04 165.08 2.12 140.32 2.25 118.133 3.03 169.53 2.14 140.34 2.28 110.219 2.99 170.05 2.13 150.60 2.31 105.062 2.97 150.80 2.11 140.39 2.26 120.706 3.08 153.99 2.11 140.20 2.30 119.268 3.09 153.99 2.11 139.63 2.29 118.429 3.05 Av. N % 2.12 2.27 3.05 RSD % 0.519 1.156 1.34 Thermo Scientific Poster Note PN42209_PITTCON 2014_E_03/14S 5

Figure 3 shows a typical calibration with Aspartic acid using K factor as calibration method FIGURE 3. Typical calibration curve. Figure 4 shows a typical nitrogen chromatogram FIGURE 4. Typical Nitrogen chromatogram. Conclusion Good repeatability, accuracy and precision was obtained with the FLASH 2000 Nitrogen Analyzer using argon as the carrier gas. No memory effect was observed when changing the type of sample, indicating complete combustion and detection of the element. 6 Nitrogen Determination in Soils and Plants by Flash Combustion using Argon as Carrier Gas

www.thermoscientific.com 2014 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientific, Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa +43 1 333 50 34 0 Australia +61 3 9757 4300 Austria +43 810 282 206 Belgium +32 53 73 42 41 Canada +1 800 530 8447 China 800 810 5118 (free call domestic) 400 650 5118 Denmark +45 70 23 62 60 Europe-Other +43 1 333 50 34 0 Finland +358 9 3291 0200 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9494 Italy +39 02 950 591 Japan +81 45 453 9100 Latin America +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Norway +46 8 556 468 00 Russia/CIS +43 1 333 50 34 0 Thermo Fisher Scientific, San Jose, CA USA is ISO 9001:2008 Certified. Singapore +65 6289 1190 Spain +34 914 845 965 Sweden +46 8 556 468 00 Switzerland +41 61 716 77 00 UK +44 1442 233555 USA +1 800 532 4752 PN42209_E 03/14S