1 Monoclonal Antibody Characterization on Q Exactive and Oribtrap Elite Yi Zhang, Ph.D Senior Proteomic Marketing Specialist Oct. 12, 211
2 Outline Orbitrap Mass Spectrometer in mab Characterization Intact Protein Analysis on Q Exactive Using Protein Deconvolution Top-Down Analysis of mab Light Chain on Orbitrap Elite Conclusion 2
3 Mass Spectrometry in mab Characterization Monoclonal antibodies have become a major drug class. Highly specific Fewer side effects Quality control is challenging Big molecules (15kD) More things can go wrong. The role of mass spectrometry Mass measurement of intact mab Sequence confirmation Disulfide mapping PTM identification Glycan analysis 3
4 Challenges of Protein Detection in Orbitrap Analyzer Myoglobin (16.9 kd) z: +16 Fast Signal Decay: Collision of ions Fragmentation of ions De-phasing of ion packets Solutions: Higher Vacuum Hardware Innovation Software Innovation t, ms Enolase (46kD) z: t, ms Transferrin (78kD) z: t, ms 4
5 Hardware Inovations in New Orbitrap Analyzer Improved Detector Hardware Reduces Delay in Transient Acquisition Relative Intensity Relative Intensity ms <.6 ms Exactive Q Exactive & Orbitrap Elite ms <1ms ms Efficient Capture of Initial Transient 5 Transferrin (78kD) z: t, ms Stronger Signal Better Mass Accuracy See aslo ASMS Poster MP93 Oliver Lange et al.
6 Software Innovation Advanced Signal Processing Increases resolution by 2 fold for the same transient time Maximum Resolution: Q Exactive: 14K ( 2) 1K ( 4) Orbitrap Elite: 24K ( 4) R= R= R= R= R=1323 Enolase (46.64 kda), 47+, Orbitrap Elite Isotopically Resolved LC & HC Top-Down Analysis on LC Time Scale 6
7 Outline Orbitrap Mass Spectrometer in mab Characterization Intact Protein Analysis on Q Exactive Using Protein Deconvolution Top Down Analysis of mab Light Chain on Orbitrap Elite Conclusion 7
8 Intact Protein Mass Measurement on Q Exactive Relative Abundance Protein: ~2kD R: 14K z= z= z= z= z= z= z= z=9 Relative Abundance z= z= z=8 Isotopically Resolved Relative Abundance mab:~15kd R: 17.5K Isotopically Unresolved 8
9 Protein Deconvolution 1. Workflow software for intact protein mass determination Includes 2 deconvolution algorithms: Xtract for isotopically resolved proteins ReSpect TM for isotopically unresolved proteins (e.g. IgG) As of today, the software is feature complete and undergoing testing Target release date: Late October, early November 9
10 Protein Deconvolution 1. A Workflow Design Select Algorithm Load File Create/Select Method 1
11 Protein Deconvolution Method Parameters Straight-forward Parameter Setting Estimated Target Mass Automatic Resolution Detection 11
12 Protein Deconvolution Chromatogram Tab Averaged spectrum 12
13 Protein Deconvolution - ReSpect TM Deconvolution Fast Data Processing:1-2s Deconvoluted Spectrum Deconvolution Data 13
14 Protein Deconvolution - Report Comprehensive, Exportable Report 14
15 Isotopic Resolution for Accurate Mass Detection Relative Abundance Protein:~2kD R: 14K z= z= z= z= z= z=9 z= z= z= z=11 z= Relative Abundance Xtract Monoisotopic Mass (<5ppm) amu 1 Oxidation 15
16 Precise Mass Measurement of Intact Monoclonal Antibody Relative Abundance mab: ~15kD R: 17.5K ReSpect -.7 ppm GF+GF -7 ppm GF+G1F Relative Abundance ppm GF+G2F (or 2G1F) ppm G+GF 2xMan5 G+G -.9 ppm G1F+G2F G2F+G2F G1F+G2F+SA 16 Shiaw-Lin Wu, Barry Karger, Barnett Institute, Northeastern University
17 Accurate Mass Measurement from a Single Charge State GF/G1F Δ= 7. ppm 52+ GF/GF Δ=.1 ppm G1F/G1F or GF/G2F Δ = 1 ppm G/GF Δ= 1 ppm G1F/G2F Δ = 8 ppm
18 Reproducibility of Accurate MW Determination ReSpect TM Deconvolution of Q Exactive IgG Data ppm mass measurement errors RAW file Q Exactive G+GF GF+GF GF+G1F GF+G2F G1F+G2F /- 6.4 ppm mass tolerance for multiple measurements 18
19 Reproducibility of Relative Abundance Measurement ReSpect TM Deconvolution of Q Exactive IgG Data Relative Abundances RAW file Q Exactive G+GF GF+GF GF+G1F GF+G2F G1F+G2F CV 3.6% 1.7% N.A. 2.2% 4.9% GF+G1F GF+GF GF+G2F (or 2G1F) G+GF G1F+G2F 19
20 Summary of Q Exactive and Protein Deconvolution 1. Q Exactive accurately and reproducibly measures the mass of intact antibody to ~7ppm. The new Protein Deconvolution software enables fast and accurate calculation of the intact mass of mabs. Q Exactive is also a powerful HR/AM mass spectrometer for routine peptide identification and peptide mapping. This LC/MS-based workflow solution provides a robust QC solution for biopharmaceutical products. 2
21 Outline Orbitrap Mass Spectrometer in mab Characterization Intact Protein Analysis on Q Exactive Using Protein Deconvolution Top-Down Analysis of mab Light Chain on Orbitrap Elite Conclusion 21
22 Orbitrap Elite: Velos Pro Iontrap + High Field Orbitrap Key Features for Intact Protein Analysis: Higher Resolution (24, at 1Hz) Faster Scan Speed 4X more scans at same resolution Multiple Fragmentation Mechanism CID HCD ETD Case Study: Top-Down Analysis of mab Light Chain 22
23 Step 1 Accurate MW Determination of mab Light Chain z= z=16 R: 24K z= z= z= z= z= z= z=18 Relative Abundance z= z= z=13 Relative Abundance z= z= z= z= z= z= z= z= Xtract Measured Mass = Target Mass = Dalton Mass Deviation 2 S-S? 23 Shiaw-Lin Wu, Barry Karger, Barnett Institute, Northeastern University
24 Step 2 - Top-Down Analysis with ETD at 12K Resolution Relative Abundance ETD R: 12K z= z= z= z= z= z= Relative Abundance z= z= z= z= z= z= z= z= z= z=? z= z=
25 Step 3 - Top-Down Analysis with HCD at 12K Resolution HCD R: 12K Relative Abundance [ ] x5 x5 x z= z= z= z= z= z= z= z= z= z= z= z=? z= z= z= z= z= z=? z= z=? z= z=
26 Step 4 ProSightPC Data Analysis 4 Dalton deviation is probably due to internal disulfides. 26
27 Summary of Top-Down Analysis on Orbitrap Elite Orbitrap Elite enables mab subunit analysis at isotopic resolution, allows accurate mass measurement (<5ppm), and quick identification of minor modifications. The combination of multiple fragmentation mechanisms in top-down analysis generates comprehensive sequence coverage, enables fast localization of modification with minimum sample preparation. Orbitrap Elite enables comprehensive characterization of intact antibody with its superior performance and versatility. 27
28 HR/AM Intact Mass Measurement Workflow Exact Mass Determination Solutions On-line Desalting: HPLC Separation Sample Analysis: Orbitrap Elite or Q Exactive High RP for resolution of proteins ~5k low RP for larger proteins Data Analysis: Protein Deconvolution 1. with ReSpector Xtract to zero charge accurate mass Relative Abundance Solutions RSLC nano/cap ProSwift RP columns Relative Abundance z=? R= R=6813 z=? z=? R=69867 z=? R=58977 z=? R=6784 z=? R= R=88597 z=? R=45346 z=? R=47666 R=43558 z=? z=? z=? Relative Abundance Solutions Orbitrap hybrid family Q Exactive Solutions ReSpect(unresolved spectra) Xtract (resolved spectra) 28
29 Conclusion Thermo Fisher provides complete LC/MS-based workflow solutions for intact protein analysis and comprehensive protein characterization. 29
30 Acknowledgement Q Exactive R&D Team Orbitrap Elite R&D Team San Jose Proteomics Marketing Team San Jose Demo Lab Barnett Institute, Northeastern University Shiaw-Lin Wu Barry Karger 3
31 31 Thank You!
A Complete Workflow Solution for Intact Monoclonal Antibody Characterization Using a New High-Performance Benchtop Quadrupole- Orbitrap LC-MS/MS
A Complete Workflow Solution for Intact Monoclonal Antibody Characterization Using a New High-Performance Benchtop Quadrupole- Orbitrap LC-MS/MS Zhiqi Hao, 1 Yi Zhang, 1 David Horn, 1 Seema Sharma, 1 Shiaw-Lin
More informationA High Resolution Bench-Top Orbitrap LC-MS Workflow Solution for Comprehensive Intact Monoclonal Antibody Characterization
A High Resolution Bench-Top Orbitrap LC-MS Workflow Solution for Comprehensive Intact Monoclonal Antibody Characterization Zhiqi Hao, 1 David M. Horn, 1 Shiaw-Lin Wu, 2 Fan Zhang, 2 and Patrick Bennett
More informationThermo Scientific Solutions for Intact-Protein Analysis. Better, Faster Decisions for. Biotherapeutic Development
Thermo Scientific Solutions for Intact-Protein Analysis Better, Faster Decisions for Biotherapeutic Development Quickly and Accurately Assess Product Quality and Safety Therapeutic proteins and monoclonal
More informationThe world leader in serving science. Jonathan L. Josephs and Aaron O. Bailey. Life Sciences Mass Spectrometry
Characterization of Monoclonal Antibodies Using UHPLC-HRAM MS The power of the Q Exactive BioPharma for online intact, subunit analysis, middle down, and peptide mapping Jonathan L. Josephs and Aaron O.
More informationPrecise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF
Precise Characterization of Intact Monoclonal Antibodies by the Agilent 6545XT AdvanceBio LC/Q-TOF Application Note Author David L. Wong Agilent Technologies, Inc. Santa Clara, CA, USA Introduction Monoclonal
More informationA comprehensive CE/ESI-MS solution for BioPharma applications
A comprehensive CE/ESI-MS solution for BioPharma applications Thermo Scientific Q Exactive hybrid quadrupole-orbitrap mass spectrometers with ZipChip system Fast CE Separation Nanospray Sensitivity HRAM
More informationWorkflows for the Characterization of Glycan structure on Biotherapeutics
New Innovations in UHPLC and LC-MS Workflows for the Characterization of Glycan structure on Biotherapeutics Ken Cook EU Bio-Separations Support Expert April 2015 1 The world leader in serving science
More informationThermo Scientific Q Exactive HF Orbitrap LC-MS/MS System. Higher-Quality Data, Faster Than Ever. Speed Productivity Confidence
Thermo Scientific Q Exactive HF Orbitrap LC-MS/MS System Higher-Quality Data, Faster Than Ever Speed Productivity Confidence Transforming discovery quantitation The Thermo Scientific Q Exactive HF hybrid
More informationPut the PRO in Protein Characterization
Put the PRO in Protein Characterization Aaron Boice LC/Q-TOF Product Manager 5-June-2017 Agenda The Right Tools for the Job Tackling Intact Proteins Break it Down Peptides and PTMs Sweet Success Released
More informationADVANCING ATTRIBUTE CONTROL OF ANTIBODIES AND ITS DERIVATIVES USING HIGH RESOLUTION ANALYTICS
ADVANCING ATTRIBUTE CONTROL OF ANTIBODIES AND ITS DERIVATIVES USING HIGH RESOLUTION ANALYTICS Henry Shion 1, Jing Fang 1, William Alley 1, Barbara Sullivan 2, Nick Tomczyk 3, Liuxi Chen 1, Ying-Qing Yu
More informationProMass HR Applications!
ProMass HR Applications! ProMass HR Features Ø ProMass HR includes features for high resolution data processing. Ø ProMass HR includes the standard ProMass deconvolution algorithm as well as the full Positive
More informationProteins. Patrick Boyce Biopharmaceutical Marketing Manager Waters Corporation 1
Routine Characterization of mabs and Other Proteins Patrick Boyce Biopharmaceutical Marketing Manager Europe and India 2011 Waters Corporation 1 Agenda Why? What scientific challenges? Technology Example
More informationCharacterization of intact monoclonal antibody with microfluidic chip electrophoresis mass spectrometry
POSTER NOTE 65 Characterization of intact monoclonal antibody with microfluidic chip electrophoresis mass spectrometry Authors Chien-Hsun Chen, 1 Andreas Krupke, 2 Stephane Houel, 1 Aaron Bailey, 1 Aran
More informationSimple, Robust, High Quality Intact Mass Analysis A Biosimilars Case Study
3 Conjugation Site (ADC) 4 34 35 1 1 35 34 CASE STUDY 3 National Institute for Bioprocessing 3 4 Research and Training 1 35 31 5 31 3 6 9 3 9 5 6 7 7 6 8 8 3 1 19 16 17 19 18 7 9 11 11 5 17 9 7 3 1 9 9
More informationAnalysis of Monoclonal Antibodies and Their Fragments by Size Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer
Analysis of Monoclonal Antibodies and Their Fragments by Size Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer Shanhua Lin, 1 Hongxia Wang, 2 Zhiqi Hao, 2 Patrick Bennett, 2 and Xiaodong
More informationSeparation of Native Monoclonal Antibodies and Identification of Charge Variants:
Separation of Native Monoclonal Antibodies and Identification of Charge Variants: Teamwork of the Agilent 31 OFFGEL Fractionator, Agilent 21 Bioanalyzer and Agilent LC/MS Systems Application Note Biosimilar
More informationLTQ Orbitrap XL Hybrid FT Mass Spectrometer Unrivaled Performance and Flexibility
m a s s s p e c t r o m e t r y LTQ Orbitrap XL Hybrid FT Mass Spectrometer Unrivaled Performance and Flexibility Part of Thermo Fisher Scientific LTQ Orbitrap XL OFFERING OUTSTANDING MASS ACCURACY, RESOLVING
More informationIsotopic Resolution of Chromatographically Separated IdeS Subunits Using the X500B QTOF System
Isotopic Resolution of Chromatographically Separated IdeS Subunits Using the X500B QTOF System Fan Zhang 2, Sean McCarthy 1 1 SCIEX, MA, USA, 2 SCIEX, CA USA Analysis of protein subunits using high resolution
More informationEnsure your Success with Agilent s Biopharma Workflows
Ensure your Success with Agilent s Biopharma Workflows Steve Madden Software Product Manager Agilent Technologies, Inc. June 5, 2018 Agenda Agilent BioPharma Workflow Platform for LC/MS Intact Proteins
More informationMore Capabilities, New Possibilities in Mass Spectrometry
More Capabilities, New Possibilities in Mass Spectrometry Boston, Oct 14, 2011 Iain Mylchreest, Ph.D San Jose, CA Increasing App s in Life Sciences Research and Applied Markets Proteomics Biomarker discovery
More informationThermo Scientific Peptide Mapping Workflows. Upgrade Your Maps. Fast, confident and more reliable peptide mapping.
Thermo Scientific Peptide Mapping Workflows Upgrade Your Maps Fast, confident and more reliable peptide mapping. Smarter Navigation... Peptide mapping is a core analytic in biotherapeutic development.
More informationAnswers, Simple and Streamlined. Solutions for Biotherapeutic Intact Mass Analysis
Answers, Simple and Streamlined Solutions for Biotherapeutic Intact Mass Analysis Expert Functionality for Mass Spec Non-Experts SCIEX OS on the X500B QTOF system dramatically improves the user experience,
More informationCharacterization of a fusion protein under native and denaturing conditions with maxis II
Characterization of a fusion protein under native and denaturing conditions with maxis II Abstract The maxis II UHR-QTOF was used to characterize key attributes of a fusion protein in its native conformation
More informationPLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC
PLRP-S Polymeric Reversed-Phase Column for LC/MS Separation of mabs and ADC Analysis of Intact and Fragmented mabs and ADC Application Note Biotherapeutics and Biologics Author Suresh Babu C.V. Agilent
More informationApplication Note # ET-20 BioPharma Compass: A fully Automated Solution for Characterization and QC of Intact and Digested Proteins
Application Note # ET-20 BioPharma Compass: A fully Automated Solution for Characterization and QC of Intact and Digested Proteins BioPharma Compass TM is a fully automated solution for the rapid characterization
More informationEnabling routine characterization of proteins. Agilent MassHunter BioConfirm software
Enabling routine characterization of proteins Agilent MassHunter BioConfirm software Bioconfirm a key component of your major biopharma workflows Intact protein analysis BioConfirm will rapidly deconvolute
More informationImproved Analysis of Biopharmaceutical Samples Using an MS-only Orbitrap Mass Spectrometer
Improved Analysis of Biopharmaceutical Samples Using an MS-only Orbitrap Mass Spectrometer Olaf Scheibner; Eugen Damoc; Eduard Denisov; Jan-Peter Hauschild; Oliver Lange; Frank Czemper; Alexander Kholomeev;
More informationENHANCED PROTEIN AND PEPTIDE CHARACTERIZATION
Agilent MassHunter BioConfirm Software ENHANCED PROTEIN AND PEPTIDE CHARACTERIZATION Spectrum Algorithm Peak Modeling Deconvolution Mass 16951.778 Sequence Name Myoglobin (horse) Target mass 16951.673
More informationSolutions for the Core and Protein Laboratory: MassHunter Walkup
Solutions for the Core and Protein Laboratory: MassHunter Walkup Jade C. Byrd MS Software Product Manager Page 1 Common Challenges in Characterizing Biomolecules Molecular size (insulin is 5 kda, mabs
More informationProfiling Glycosylation of Monoclonal Antibodies at Three Levels Using the Agilent 6545XT AdvanceBio LC/Q TOF
Application Note Biotherapeutics and Biosimilars Profiling Glycosylation of Monoclonal Antibodies at Three Levels Using the Agilent 64XT AdvanceBio LC/Q TOF Author David L. Wong Agilent Technologies, Inc.
More informationFast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles
Fast and Efficient Peptide Mapping of a Monoclonal Antibody (mab): UHPLC Performance with Superficially Porous Particles Application Note Biotherapeutics and Biosimilars Authors James Martosella, Alex
More informationLC/MS Based Quantitation of Intact Proteins for Bioanalytical Applications
LC/MS Based Quantitation of Intact Proteins for Bioanalytical Applications Alex Zhu, Ph.D. Agilent Technologies Wilmington, DE ASMS,2017-06 1 Outline Introduction on intact protein quantitation Agilent
More informationAntibody subunit analysis workflow on a quadrupole-orbitrap mass spectrometer: from optimized sample preparation to data analysis
APPLICATION NOTE 72854 Antibody subunit analysis workflow on a quadrupole-orbitrap mass spectrometer: from optimized sample preparation to data analysis Authors Kai Scheffler, Thermo Fisher Scientific,
More informationLigand Binding Mass Spectrometric Immunoassay (LB-MSIA ) Workflow with Deglycosylation for Therapeutic Antibodies
Ligand Binding Mass Spectrometric Immunoassay (LB-MSIA ) Workflow with Deglycosylation for Therapeutic Antibodies Kwasi Antwi, Amanda Ribar, Urban A. Kiernan, and Eric E. Niederkofler, Thermo Fisher Scientific,
More informationRapid Peptide Mapping via Automated Integration of On-line Digestion, Separation and Mass Spectrometry for the Analysis of Therapeutic Proteins
Rapid Peptide Mapping via Automated Integration of On-line Digestion, Separation and Mass Spectrometry for the Analysis of Therapeutic Proteins Esther Lewis, 1 Kevin Meyer, 2 Zhiqi Hao, 1 Nick Herold,
More informationThermo Scientific Orbitrap Fusion Lumos Tribrid Mass Spectrometer. Breakthrough Gains for Quantitative Biology. Sensitivity Transformed
Thermo Scientific rbitrap Fusion Lumos Tribrid Mass Spectrometer Breakthrough Gains for Quantitative Biology Sensitivity Transformed The Most Advanced Instrument Built for the Most Challenging Experiments
More informationAnalysis of Monoclonal Antibodies and Their Fragments by Size-Exclusion Chromatography Coupled with an Orbitrap Mass Spectrometer
Analysis of Monoclonal Antibodies and Their Fragments by ize-exclusion Chromatography Coupled with an Orbitrap Mass pectrometer hanhua Lin, Hongxia Wang, Zhiqi Hao, and Xiaodong Liu Thermo Fisher cientific,
More informationThe Overview of LC-MS Quantitative Solutions for Biotherapeutic Analysis
The Overview of LC-MS Quantitative Solutions for Biotherapeutic Analysis Featuring SCIEX Instrumentation and Technologies Lei Xiong, Elliott Jones SCIEX, Redwood City, California, USA Figure 1. Summary
More informationFourier-Transform MS for Top Down Proteomics
Fourier-Transform MS for Top Down Proteomics Proteomics Team @ Northwestern University Thermo User Meeting, December 9 th, 2015 Hosted by Thermo Fisher Scientific Odense, Denmark On the Use of Top Down
More informationFaster, easier, flexible proteomics solutions
Agilent HPLC-Chip LC/MS Faster, easier, flexible proteomics solutions Our measure is your success. products applications soft ware services Phospho- Anaysis Intact Glycan & Glycoprotein Agilent s HPLC-Chip
More informationrapiflex Innovation with Integrity Designed for Molecules that Matter. MALDI TOF/TOF
rapiflex Designed for Molecules that Matter. Innovation with Integrity MALDI TOF/TOF rapiflex TM The first MALDI-TOF/TOF that adapts to your needs. The rapiflex is the most advanced MALDI TOF/TOF system
More informationStructure Characterization and Differentiation of Biosimilar and Reference Products Using Unique Combination of Complementary Fragmentation Mechanisms
Structure Characterization and Differentiation of Biosimilar and Reference Products Using Unique Combination of Complementary Fragmentation Mechanisms Zhiqi Hao, 1 Chen Li, 2 Shiaw-Lin Wu, 2,3 David M.
More informationAnalysis of Monoclonal Antibody (mab) Using Agilent 1290 Infinity LC System Coupled to Agilent 6530 Accurate-Mass Quadrupole Time-of-Flight (Q-TOF)
Analysis of Monoclonal Antibody (mab) Using Agilent 9 Infinity LC System Coupled to Agilent Accurate-Mass Quadrupole Time-of-Flight (Q-TOF) Application Note Authors Ravindra Gudihal and Suresh Babu CV
More informationA highly sensitive and robust 150 µm column to enable high-throughput proteomics
APPLICATION NOTE 21744 Robust LC Separation Optimized MS Acquisition Comprehensive Data Informatics A highly sensitive and robust 15 µm column to enable high-throughput proteomics Authors Xin Zhang, 1
More informationThe Direction of Thermo Fisher Scientific
The Direction of Thermo Fisher Scientific Iain Mylchreest Ph.D, VP GM Life Sciences MS Thermo Scientific Mass Spectrometry Users Meeting Indianapolis September 22 th, 2011 The World Leader in Serving Science
More informationAdvanced Characterization of Antibody Drug Conjugates (ADCs) by Liquid Chromatography and Mass Spectrometry (LC/MS) John Gebler, Ph.D.
Advanced Characterization of Antibody Drug Conjugates (ADCs) by Liquid Chromatography and Mass Spectrometry (LC/MS) John Gebler, Ph.D. Director www.waters.com/adc 2015 Waters Corporation 1 2015 Waters
More informationNext Generation Technology for Reproducible and Precise Proteome Profiling
7 th Czech Mass Spectrometry Conference April 11th, 2018 Next Generation Technology for Reproducible and Precise Proteome Profiling Lars Kristensen, Ph.D. Application and training specialist The world
More informationComparability Analysis of Protein Therapeutics by Bottom-Up LC-MS with Stable Isotope-Tagged Reference Standards
Comparability Analysis of Protein Therapeutics by Bottom-Up LC-MS with Stable Isotope-Tagged Reference Standards 16 September 2011 Abbott Bioresearch Center, Worcester, MA USA Manuilov, A. V., C. H. Radziejewski
More informationAn Improved Approach for Glycan Structure Identification from HCD MS/MS Spectra
Introduction Method Experiments An Improved Approach for Glycan Structure Identification from HCD MS/MS Spectra Weiping Sun, Yi Liu, Gilles Lajoie, Bin Ma and Kaizhong Zhang Department of Computer Science,
More informationOverview of Waters Solutions for Biopharmaceutical Analysis and Characterization
Overview of Waters Solutions for Biopharmaceutical Analysis and Characterization Denis Calnan Biopharma Business Development Manager (Northern Europe) 2011 Waters Corporation 1 About Waters Facts and Figures
More informationChromatographic Workflows for Biopharmaceutical Characterization
Chromatographic Workflows for Biopharmaceutical Characterization Dr. Ken Cook European Bioseparation Sales Support Expert Thermo Fisher Scientific, Hemel Hempstead/Germany The world leader in serving science
More informationHighly Confident Peptide Mapping of Protein Digests Using Agilent LC/Q TOFs
Technical Overview Highly Confident Peptide Mapping of Protein Digests Using Agilent LC/Q TOFs Authors Stephen Madden, Crystal Cody, and Jungkap Park Agilent Technologies, Inc. Santa Clara, California,
More informationLC-MS Analysis of Digested mabs Using a Reversed-Phase Desalting Cartridge and High-Resolution, Accurate Mass Spectrometry (HRAMS)
LC-MS Analysis of Digested mabs Using a Reversed-Phase Desalting Cartridge and High-Resolution, Accurate Mass Spectrometry (HRAMS) Martin Samonig,¹ Shane Bechler,² and Kelly Flook² ¹Thermo Fisher Scientific,
More informationautoflex max Practical Power, Maximum Utility Innovation with Integrity MALDI-TOF MS
autoflex max Practical Power, Maximum Utility Innovation with Integrity MALDI-TOF MS Wide application versatility with enhanced dynamic range The autoflex max series is based on Bruker s robust and efficient
More informationVyužití cílené proteomiky pro kontrolu falšování potravin: identifikace peptidových markerů v mase pomocí LC- Q Exactive MS/MS
Využití cílené proteomiky pro kontrolu falšování potravin: identifikace peptidových markerů v mase pomocí LC- Q Exactive MS/MS Michal Godula Ph.D. Thermo Fisher Scientific The world leader in serving science
More informationTop reasons to upgrade to a Q Exactive HF-X hybrid quadrupole-orbitrap mass spectrometer
GRANT APPLICATION RESOURCE QExactive HF-X hybrid quadrupole-orbitrap mass spectrometer Top reasons to upgrade to a HF-X hybrid quadrupole-orbitrap mass spectrometer Keywords Proteomics, protein identification,
More informationAgilent Technologies. Carl Myerholtz, Ph.D. Senior Director of QTOF Technology. FAPESP/Agilent Workshop August 30,
Agilent Technologies Carl Myerholtz, Ph.D. Senior Director of QTOF Technology FAPESP/Agilent Workshop August 30, 2017 1 FAPESP/Agilent - Call for Research Proposals Focus Areas Biopharma Metabolomics In
More informationLC/MS Analysis of the Monoclonal Antibody Rituximab Using the Q Exactive Benchtop Orbitrap Mass Spectrometer
LC/MS Analysis of the Monoclonal Antibody Rituximab Using the Q Exactive Benchtop Orbitrap Mass Spectrometer Martin Samonig 1,2, Christian Huber 1,2 and Kai Scheffler 2,3 1 Division of Chemistry and Bioanalytics,
More informationInnovations for Protein Research. Protein Research. Powerful workflows built on solid science
Innovations for Protein Research Protein Research Powerful workflows built on solid science Your partner in discovering innovative solutions for quantitative protein and biomarker research Applied Biosystems/MDS
More informationIdeS-cleaved mab subunit analysis with LC-HRAM-MS: a quick and accurate comparison of biosimilar and originator biotherapeutics
APPLICATION NOTE 2186 IdeS-cleaved mab subunit analysis with LC-HRAM-MS: a quick and accurate comparison of biosimilar and originator biotherapeutics Authors Sara Carillo, Craig Jakes, Izabela Zaborowska,
More informationHigh-Throughput Peptide Mapping with the Vanquish UHPLC System and the Q Exactive HF Mass Spectrometer
High-Throughput Peptide Mapping with the Vanquish UHPLC System and the Q Exactive HF Mass Spectrometer Martin Samonig 1, Kai Scheffler 2, Remco Swart 1, and Jonathan Josephs 3 1 Thermo Fisher Scientific,
More informationIMPORTANCE OF CELL CULTURE MEDIA TO THE BIOPHARMACEUTICAL INDUSTRY
IMPORTANCE OF CELL CULTURE MEDIA TO THE BIOPHARMACEUTICAL INDUSTRY Research, development and production of biopharmaceuticals is growing rapidly, thanks to the increase of novel therapeutics and biosimilar
More informationBioanalysis of Intact Therapeutic Monoclonal Antibodies from Non-Human Primates Using MSIA D.A.R.T. S Technology
APPLICATION NOTE Bioanalysis of Intact Therapeutic Monoclonal Antibodies from Non-Human Primates Using MSIA D.A.R.T. S Technology An Integral Part of Ligand Binding-Mass Spectrometric Immunoassay Workflow
More informationThermo Scientific Proteomics & Metabolomics Seminar Tour
Proteomics & Seminar Tour Ignite your workflows with our with Cutting Edge LC-MS Technology Stop by to learn the latest advances in mass spectrometry workflows, technologies launched at 2012 ASMS and talk
More informationClearly better LC/MS solutions
Agilent 6000 Series LC/MS Clearly better LC/MS solutions Our measure is your success. products applications soft ware services Agilent: the clearly better choice for qualitative and quantitative LC/MS.
More informationHigh Resolution Accurate Mass Peptide Quantitation on Thermo Scientific Q Exactive Mass Spectrometers. The world leader in serving science
High Resolution Accurate Mass Peptide Quantitation on Thermo Scientific Q Exactive Mass Spectrometers The world leader in serving science Goals Explore the capabilities of High Resolution Accurate Mass
More informationApplication Note LCMS-87 Automated Acquisition and Analysis of Data for Monitoring Protein Conjugation by LC-MS Using BioPharma Compass
Application Note LCMS-87 Automated Acquisition and Analysis of Data for Monitoring Protein Conjugation by LC-MS Using BioPharma Compass Abstract Here we describe the development and use of an automated
More informationChapter 1. Primary Structure. See for options on how to legitimately share published articles.
Chapter 1 Downloaded via 148.251.232.83 on January 29, 2019 at 07:45:49 (UTC). See https://pubs.acs.org/sharingguidelines for options on how to legitimately share published articles. Determination of the
More informationZipChip TM. Microfluidic CE-ESI Biotherapeutics CE-ESI-MS Applications. Please visit us at
ZipChip TM Microfluidic CE-ESI Biotherapeutics CE-ESI-MS Applications Please visit us at http://www.908devices.com/ // Actionable Intelligence HPMS & On-board Analytics / Slide 1 June 2016 // ZipChip CE-ESI
More informationNew Approaches to Quantitative Proteomics Analysis
New Approaches to Quantitative Proteomics Analysis Chris Hodgkins, Market Development Manager, SCIEX ANZ 2 nd November, 2017 Who is SCIEX? Founded by Dr. Barry French & others: University of Toronto Introduced
More informationUnified Workflow for Monoclonal Antibody Charge Heterogeneity, Purity, and Molecular Weight Analysis
Procedures Unified Workflow for Monoclonal Antibody Charge Heterogeneity, Purity, and Molecular Weight Analysis Separation and Online Detection of Intact mab Variants and Impurities using CESI-MS Bryan
More informationBioConfirm B Presentation
BioConfirm B.08.00 Presentation 1 Major Enhancements to Increase Your Productivity The Latest Version of our BioConfirm Product, B.08.00 BioConfirm is protein and peptide characterization software using
More informationHost Cell Protein Analysis Using Agilent AssayMAP Bravo and 6545XT AdvanceBio LC/Q-TOF
Application Note Biotherapeutics and Biosimilars Host Cell Protein Analysis Using Agilent AssayMAP Bravo and 6545XT AdvanceBio LC/Q-TOF Authors Linfeng Wu, Shuai Wu and Te-Wei Chu Agilent Technologies,
More informationAntibody Analysis by ESI-TOF LC/MS
Application Note PROTEOMICS METABOLOMICS GENOMICS INFORMATICS GLYILEVALCYSGLUGLNALASERLEUASPARG CYSVALLYSPROLYSPHETYRTHRLEUHISLYS Antibody Analysis by ESI-TOF LC/MS Authors Lorenzo Chen, Merck & Company,
More informationCharacterize Fab and Fc Fragments by Cation-Exchange Chromatography
Characterize Fab and Fc Fragments by Cation-Exchange Chromatography Application Note Biologics and Biosimilars Authors Isabel Vandenheede, Emmie Dumont, Pat Sandra, and Koen Sandra Research Institute for
More informationA Unique LC-MS Assay for Host Cell Proteins(HCPs) ) in Biologics
A Unique LC-MS Assay for Host Cell Proteins(HCPs) ) in Biologics Catalin Doneanu,, Ph.D. Biopharmaceutical Sciences, Waters September 16, 2009 Mass Spec 2009 2009 Waters Corporation Host Cell Proteins
More informationLiver Mitochondria Proteomics Employing High-Resolution MS Technology
Liver Mitochondria Proteomics Employing High-Resolution MS Technology Jenny Ho, 1 Loïc Dayon, 2 John Corthésy, 2 Umberto De Marchi, 2 Antonio Núñez, 2 Andreas Wiederkehr, 2 Rosa Viner, 3 Michael Blank,
More informationThe clearly better choice
Agilent Accurate-Mass 6500 Series Q-TOF and 600 Series TOF LC/MS Systems The clearly better choice in TOF and Q-TOF LC/MS Our measure is your success. products applications soft ware services The unmatched
More informationBiotherapeutic Non-Reduced Peptide Mapping
Biotherapeutic Non-Reduced Peptide Mapping Routine non-reduced peptide mapping of biotherapeutics on the X500B QTOF System Method details for the routine non-reduced peptide mapping of a biotherapeutic
More informationMonoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column
Monoclonal Antibody Analysis on a Reversed-Phase C4 Polymer Monolith Column Shane Bechler 1, Ken Cook 2, and Kelly Flook 1 1 Thermo Fisher Scientific, Sunnyvale, CA, USA; 2 Thermo Fisher Scientific, Runcorn,
More informationApplication Note. Kwasi Antwi, Amanda Ribar, Urban A. Kiernan, and Eric E. Niederkofler Thermo Fisher Scientific, Tempe, Arizona
Analysis of an Antibody Drug Conjugate using MSIA D.A.R.T. S. Technology, an Integral Part of Ligand Binding Mass Spectrometric Immunoassay (LB-MSIA) Workflow Application Note Kwasi Antwi, Amanda Ribar,
More informationSGS Delivering Excellence in Compliant Extractables and Leachables Testing
SGS CASE STUDY SGS Delivering Excellence in Compliant Extractables and Leachables Testing A case study in unknown impurity identification & quantification SGS is a world leading inspection, verification,
More informationThermo Scientific Ligand Binding Mass Spectrometric Immunoassay (LB-MSIA)
INSTRUCTIONS FOR USE OF PRODUCTS Thermo Scientific Ligand Binding Mass Spectrometric Immunoassay (LB-MSIA) Demonstration of the LB-MSIA Protocol for the Analysis of Intact Adalimumab from Mouse Plasma
More informationSimultaneous Quantitation of a Monoclonal Antibody and Two Proteins in Human Plasma by High Resolution and Accurate Mass Measurements
Simultaneous Quantitation of a Monoclonal Antibody and Two Proteins in Human Plasma by High Resolution and Accurate Mass Measurements Paul-Gerhard Lassahn 1, Kai Scheffler 2, Myriam Demant 3, Nathanael
More informationFPO. BioPharma Compass 2.0. Innovation with Integrity. Accelerate Biopharmaceutical Analysis. Software
FPO BioPharma Compass 2.0 Accelerate Biopharmaceutical Analysis Innovation with Integrity Software Developed with the Biopharma Industry Award-winning biopharmaceutical analysis platform Commercial and
More informationQuantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass
Application Note: 529 Quantitative LC-MS Analysis of 14 Benzodiazepines in Urine Using TraceFinder 1.1 Software and High Resolution Accurate Mass Xiang He, Marta Kozak; Thermo Fisher Scientific, San Jose,
More informationApplication Note. Author. Abstract. Biotherapeutics and Biologics. Sonja Schneider Agilent Technologies, Inc. Waldbronn, Germany
D-LC/MS Characterization of Charge Variants Using Ion Exchange and Reversed-Phase Chromatography Multiple Heart-Cutting D-LC Analysis of Innovator versus Biosimilar Monoclonal Antibodies Application Note
More informationQuantification of Host Cell Protein Impurities Using the Agilent 1290 Infinity II LC Coupled with the 6495B Triple Quadrupole LC/MS System
Application Note Biotherapeutics Quantification of Host Cell Protein Impurities Using the Agilent 9 Infinity II LC Coupled with the 6495B Triple Quadrupole LC/MS System Authors Linfeng Wu and Yanan Yang
More informationThermo Scientific EASY-Spray Technology. Plug-and-spray with. State-of-the-art performance
Thermo Scientific EASY-Spray Technology Plug-and-spray with State-of-the-art performance Effortless nano electrospray ionization Integrated design EASY-Spray source Nano-flow LC-MS relies critically on
More informationProteomics-Based Biomarker Discovery Study in Human and non-human Plasma Using a Two-Pass Workflow
Proteomics-Based Biomarker Discovery Study in Human and non-human Plasma Using a Two-Pass Workflow Maryann S. Vogelsang 1, David A. Sarracino 1, Janin Schulte 2, Bryan Krastins 1, Michael Athanas 3, Amol
More informationAsish Chakraborty, St John Skilton, Weibin Chen and John C. Gebler Waters Corporation, Milford, MA, U.S. INTRODUCTION METHODS. Sample preparation
Com p r e h ensiv e a n d Rou t in e C ha r ac t e riz at io n o f P rot eins a n d P e p t i d e s by LC/MS and LC/MS E Asish Chakraborty, St John Skilton, Weibin Chen and John C. Gebler Waters Corporation,
More informationEnhanced Analysis of Host Cell Proteins Using the Agilent 6545XT AdvanceBio LC/Q-TOF
Enhanced Analysis of Host Cell Proteins Using the Agilent 6545XT AdvanceBio LC/Q-TOF Linfeng Wu, Ph.D. LC/MS Application Scientist ASMS 2017 1 Introduction Residual Host Cell Protein (HCP) are process
More informationTony Mire-Sluis Vice President, Corporate, Product and Device Quality Amgen Inc
The Regulatory Implications of the ever increasing power of Mass Spectrometry and its role in the Analysis of Biotechnology Products Where do we draw the line? Tony Mire-Sluis Vice President, Corporate,
More informationThermo Scientific EASY-nLC 1000 System. Effortless, split-free nanoflow UHPLC Top performance in LC/MS
Thermo Scientific EASY-nLC 1000 System Effortless, split-free nanoflow UHPLC Top performance in LC/MS Effortless split-free, nanoflow UHPLC The Thermo Scientific EASY-nLC 1000 is a fully integrated, split-free,
More informationAdvanced QA/QC characterization MS in QC : Multi Attribute Method
Advanced QA/QC characterization MS in QC : Multi Attribute Method Global BioPharma Summit The world leader in serving science A Complex Problem: Drug Safety and Quality Safety Is the product safe to use?
More informationAgilent 6545XT AdvanceBio LC/Q-TOF System DESIGNED FOR BIOPHARMA
gilent 6545XT dvancebio LC/Q-TOF System DESIGNED FO BIOPHM ONE INSTUMENT, MULTIPLE WOKFLOWS Characterizing proteins requires multiple approaches to access multiple levels of information; you can examine
More informationSpectrum Mill MS Proteomics Workbench. Comprehensive tools for MS proteomics
Spectrum Mill MS Proteomics Workbench Comprehensive tools for MS proteomics Meeting the challenge of proteomics data analysis Mass spectrometry is a core technology for proteomics research, but large-scale
More informationE f f ic i ent ly Com pa ring Bat c h e s o f a n Intac t Mono c lo na l A n t ibo dy using t h e
E f f ic i ent ly Com pa ring Bat c h e s o f a n Intac t Mono c lo na l A n t ibo dy using t h e BioPharmaLynx Software Package Scott J. Berger and Weibin Chen Waters Corporation, Milford, MA, U.S.A INTRODUCTION
More informationLC/MS/MS Solutions for Biomarker Discovery QSTAR. Elite Hybrid LC/MS/MS System. More performance, more reliability, more answers
LC/MS/MS Solutions for Biomarker Discovery QSTAR Elite Hybrid LC/MS/MS System More performance, more reliability, more answers More is better and the QSTAR Elite LC/MS/MS system has more to offer. More
More informationConjugated protein biotherapeutics
B i o P r o c e s s TECHNICAL Mass Spectrometric Conjugate Characterization Process Qualification of Recombinant Protein Hapten Conjugation Miao-Fang Lin, Margo Wilson, Michael V. Murray, and Greg W. Adams
More information